Hi Pavan, Have a look at this thread: https://sourceforge.net/p/vscapture/discussion/general/thread/5c39a04b08/ MR400 uses a protocol different from Intellivue. If the waveform protocol is available from the manufacturer, the support can be added. Regards,
Hi Juan, The Servo-U ventilator numerics and waveforms are fully supported throught the updated VSCaptureCLI app (available in the home page or files section of this site). Details about the complex protocol may be available online by searching the term 'Servo Communication Interface' (SCI) protocol. Regards, John.
Hi Nuri, If VSCapture was useful in your commercial product, please consider acknowledging the project in your site and supporting the project through the donation link. Your solution seems quite interesting. Regards, John.
Thank you. Does your embedded system use new drivers for this? Are there any demo videos or sites about it?
I have only enabled Mindray HL7 numeric data export, just curious how you got the ECG waveforms, is this your own app or VSCaptureCLI?
Hi Nicolas, From the output (in CSV & JSON url export mode) , it seems the app is not able to connect to the monitor port. Ensure the monitor is set to the correct IP address (check service menu) on the same network & subnet or if the port is not blocked for any other reason. Have you tried running the app with just JSON file export and then uploading the file in another process with tail -f? Alternatively, you can try MQTT export but test the local file export first. Please email xeonfusion at yahoo...
Hi Kai, Are the waveforms from Mindray HL7? I plan to add Mindray waveform support in a future version. Regards, John.