0% found this document useful (0 votes)
61 views79 pages

LactoscanSP Eng PDF

The document describes an ultrasonic milk analyzer called LACTOSCAN SP. It has the following key capabilities: 1. It can measure fat, SNF, protein, lactose, water content, temperature, freezing point, salts, total solids, and density of various milk types like cow, sheep, goat, and buffalo milk. 2. It has measurement ranges for the different parameters and an accuracy of +/- 0.1-0.3% depending on the parameter. 3. It operates under normal ambient conditions of 10-40°C temperature and 30-80% humidity.

Uploaded by

Simeone Mamani
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
61 views79 pages

LactoscanSP Eng PDF

The document describes an ultrasonic milk analyzer called LACTOSCAN SP. It has the following key capabilities: 1. It can measure fat, SNF, protein, lactose, water content, temperature, freezing point, salts, total solids, and density of various milk types like cow, sheep, goat, and buffalo milk. 2. It has measurement ranges for the different parameters and an accuracy of +/- 0.1-0.3% depending on the parameter. 3. It operates under normal ambient conditions of 10-40°C temperature and 30-80% humidity.

Uploaded by

Simeone Mamani
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Ultrasonic milk analyzer

Milkotronic Ltd

LACTOSCAN SP
MILK ANALYZER
Wide LCD display – 4 lines x 16 characters
Plastic box

Operation manual

Operation manual 04.04.19 1/79


Ultrasonic milk analyzer

Switching Adapter

 Input: 100 - 240 V ~1.6 A max.


50-60 Hz
 Output: +12 V 3 А min.
 Output power: 36 - 42 W

Measurement modes

 cow milk
 sheep milk
 UHT milk
 goat milk
 buffalo milk
 cream
 whey
 recovered milk
 other /pasteurized milk/

CAUTION!
Keep the switching adapter dry!
Please, read and follow strictly all the instructions in the manual.

Due to continuous improvement in the device, information contained in this


manual is subject to change without notice. Contact the company-producer
for revisions and corrections

4, Narodni Buditeli Str.


8900 Nova Zagora
BULGARIA
Phone/Fax: + 359 457 67082
e-mail: [email protected]
www.lactoscan.com
www.milkotronic.com

Operation manual 04.04.19 2/79


Ultrasonic milk analyzer

SAFETY INSTRUCTIONS

1. Read this manual carefully and make sure that you understand all the
instructions.

2. For safety purposes the device is equipped with grounded power


cable. If there is no grounded electrical outlet where the device will
be used, please, install such before using the device.

3. Place the device on leveled and stable plate. In case it falls or is


severely shocked it may be damaged.

4. Connect to the electrical network in such a way that the power cable
to stay away from the side for accessing the device and not to be
stepped on.

5. Every time before cleaning the device switch it off and unplug it from
the electrical outlet. The device has to remain unplugged till the
cleaning completion.

6. Do not disassemble the unit in order to avoid possible electrical


shock. In case of malfunction contact your local dealer.

7. Handle the liquids the device works with carefully, following all the
instructions for their preparation.

8. Place the switching adaptor in such a way as to be protected from


overflow and spillage of liquids.

Operation manual 04.04.19 3/79


Ultrasonic milk analyzer

PARTS AND ACCESSORIES


In the table below the standard delivery configuration of the milk analyzer is
listed:
№ Description Item № pcs

1. Ultrasonic portable milk analyzer LSSP001 1


60 sec
1 sample measurement time 30 sec
2. Operation manual LSSP002 1
3. Plastic sample holder LSSP003 2
4. Spare Pipes LSSP004 2
5. 12 V DC Power Supply Cable LSSP005 1
6. Alkaline cleaning solution Lactodaily 100 g 1
7. Acidic cleaning solution Lactoweekly 100 g 1

In the table below the milk analyzer spares and accessories, which are
delivered on customers request are listed:
№ Description Item № pcs
a) included in the set: /
b) not included in the set (may be
additionally bought):
8. RS232 Interface Cable - Analyser-IBM PC LSSP006

9. Service Pack - CD LSSP007

10. pH measuring system LSS009 1


11. pH probe with cable and holder LSS010 1
12. Buffer solution Ph 60 ml LSS011 1
(pH7.00±0.01/20˚C)
13. Buffer solution pH 60 ml LSS012 1
(pH4.00±0.01/20˚C)
14. Milk conductivity measuring system LSS013 1
15. Buffer solution conductivity 50 ml (5.02 LSS014 1
(±5%) mS/cm (18±0.1˚C)
16. Real time clock LSS015 1
17. ECS POS Serial Printer LSS017 1
18. 12 V Serial Printer Power Supply Cable LSS018 1
19. RS232 Interface Cable - Milk Analyser – LSS019 1
Serial Printer

Operation manual 04.04.19 4/79


Ultrasonic milk analyzer

20. Plug type 1

21. Spare O-ring for the pH probe 1

Operation manual 04.04.19 5/79


Ultrasonic milk analyzer

FUNCTION

The function of the milk analyzer is to make quick analysis of milk on fat
(FAT), non-fat solids (SNF), proteins, lactose and water content
percentages, temperature (oС), freezing point, salts, total solids, as well
as density of one and the same sample directly after milking, at
collecting and during processing.

Operation manual 04.04.19 6/79


Ultrasonic milk analyzer

2. TECHNICAL PARAMETERS

2.1. Working modes characteristics:


The program of the milk analyzer has 5 (five) working modes.

2.1.1. Measurement mode milk / dairy product – first type

2.1.2. Measurement mode milk / dairy product – second type

2.1.3. Measurement mode milk / dairy product – third type

These modes have been calibrated on customers’ request for 3 milk types
from the following: cow, sheep, UHT, buffalo, goat, camel milk, cream, ice
cream mixtures, whey, recovered milk, etc. before leaving the production
facilities and the text on the display will be for the corresponding types, as is
indicated on page 2 Measurement modes.

2.1.4. Cleaning
2.1.5. Printing

Operation manual 04.04.19 7/79


Ultrasonic milk analyzer

2.2. Measuring range:

Fat ………………..................... …………………...from 0.01% to 25%


SNF ……………………………… ...…………………....from 3% to 15%
Density * ………………………. ……...…….from 1015 to 10 40 kg/m3
Proteins …………………………. ………………………..from 2% to 7%
Lactose ………………………... …………………..from 0.01 % to 6 %
Water content ………………… ……………...……..from 0 % to 70 %
Temperature of milk …………… ……………………..from 1oC to 40oC
Freezing point ………………… …………..from – 0,400 to – 0,700oC
Salts ……………………………. ……………...………from 0,4 to 1,5%
PH*..……………………………... …………………..………from 0 to 14
Conductivity *..………………… ……………...…from 3 to 14 [mS/cm]
Total Solids* .…………………… ………………………...from 0 to 50 %
* Option, on customers’ request

** Density data are shown in an abbreviated form. For example 27.3 have to
be understood as 1027.3 kg/m3. To determine the milk density, write down
the result from the display and add 1000.
Example: result 21,20; density = 1000 + 21,20 = 1021,2 kg/m3

The abbreviated form of the density is used also when entering data for
samples in working mode Recalibrate, for example:
If the measured sample density is 1034.5 kg/m3, then in the menu for
entering the samples parameters used for calibration, across the parameter
Den = , you have to enter 34.5.

**** Please, carefully read Appendix Freezing Point.

2.3. Accuracy:
Fat …..……………………………….. …...………………….…... 0.10%
SNF ………………………………….. …………...………….…... 0.15%
Density ……………………………… ……………………….. 0.3 kg/m3
Proteins ……………………………... ……...……………….…... 0.15%
Lactose ……………………………… .…………………………... 0.20%
Water content ………………………. .……………………………. 3.0%
Temperature of milk ……………….. .……………………………… 1oC
Freezing point……………………….. ..……...………………… 0.001oC
Salts ………………………………... …..…………...…………... 0.05%
PH …………………………………. ……………………………±0.05%
Conductivity ………………………. ……………………………...±0.05

Operation manual 04.04.19 8/79


Ultrasonic milk analyzer

Total solids …………………………. …………………………… 0.17%

The difference between two sequent measurements of one and the same
milk could not exceed the maximum permissible absolute error.

2.4 Correct ambient conditions:

Maximum permissible absolute error is guaranteed in case of normal ambient


conditions:

Air temperature………………………… ….………….from 10oC to 40oC


Relative humidity ……………………… ……………...from 30% to 80%
Power supply ………………………… …………………...220V (110V)
Extent of contamination at normal environmental conditions…………..2

Maximum permissible absolute error values in point 2.3 are in dependence


on the correctness of the corresponding chemical method, used for
component content determination. In point 2.3. are used the following
reference methods: Gerber – for fat, gravimetric – for SNF, Kjeldahl – for
protein. The boundary for maximum variation of repeatability when the power
supply voltage is from +10 to – 15% from the nominal voltage values (220 V)
have to be no more than 0.8 accuracy according point 2.3. The analyzer is
used in conditions free of outer electrical and magnetic fields (except the
magnetic field of the Earth) and vibrations.

2.5. Dimensions:
………………………….175/175/150 mm, mass 1,5 kg

2.6. Continuous working time:


…….…….………………………….…...……....non-stop

2.7 Milk sample volume per one measurement:


………………………………..………..15 cm3 (= 25 ml)

Operation manual 04.04.19 9/79


Ultrasonic milk analyzer

Fig.1 Front panel

1. Upper cover 6. Output pipe


2. Wide Display 7. Input pipe
3. Button “Down” 8. Sample holder
4. Button “Enter” 9. pH probe (option)
5. Button “Up”

Operation manual 04.04.19 10/79


Ultrasonic milk analyzer

fig. 2 Back panel

1 Switching adapter 5. 12 V input


2. Power switch 6. pH probe (option)
3. 12 V printer output 7. Serial interface (RS232/ printer)
4. pH probe input (option) 8. USB (option)

Operation manual 04.04.19 11/79


Ultrasonic milk analyzer

Fig. 4 Peripherals connection

Operation manual 04.04.19 12/79


Ultrasonic milk analyzer

Fig. 4 Cable Description

90-1801-0008
RS232 Interface Cable - Milk Analyser – Serial Printer/ IBM PC

The connection of the analyser with other devices via RS232 Interface is
made through the coupling DB9 on the back panel of the device. In it can be
plugged only the cable, supplied by the producer with the analyser (or
produced by the customer as per above scheme). The coupling from the
cable, marked on the scheme with „TO LACTOSCAN” must be plugged in the
computer. If due to some reason is used an external cable USB to RS232
Converter (usually when working with a laptop the customers buy such
cables from the market), it must be plugged in the coupling “TO PC” of the
original cable, not directly in the device.

90-1801-0009
DC 12V Power Supply Milk Analyzer Cable
1. GND
2. No connection
3. No connection
4. 12V DC

There’s an option in the analyser – embedded USB interface (USB to


RS232 Serial Converter) for connection with computer type IBM PC. It is

Operation manual 04.04.19 13/79


Ultrasonic milk analyzer

intended for connecting computers from laptop type, which have no RS 232
interface (COM Ports – DB9 Connector). This option may be available
together with the standard RS232 – connector DB9, which always exists in
the analyzers. At one and the same time, connection analyzer – computer,
can be established only through one of the couplings – either only DB9, or
only USB, i.e. analyzer could not be connected at one and the same time,
using one of the interfaces with one and using the other interface to another,
second computer. The coupling for this option (Printer Type) is at the back of
the analyzer, next to the standard RS232 DB9 connector. Other USB devices
like printers, keypads etc COULD NOT be connected to this coupling.
USB interface is based on the element MCP2200 of the company
Microchip Technology Inc. – site: http://www.microchip.com. For establishing
a connection through this interface a driver has to be installed for MCP2200
in the correspondent computer, which will be connected with the analyzer.
Please, follow the procedure, from the Internet site of the producer:
http://www.microchip.com/wwwproducts/devices.aspx?ddocname=en546923,
by choosing the suitable for your operation system driver.
After installing the driver, choose the СОМ port, which will be used for
the real communication. For Win XP it is done by executing the following
commands: Start -> Settings -> Control panel -> System -> Hardware ->
Device Manager -> Ports (COM and LPT) – Right Click -> Properties -> Port
Settings -> Advanced -> COM Port Number. Choose Number 1 or if another
number is chosen, for example 3, then, when working with the software tools,
in the field COM Port (upper right corner), you have to set the new number of
the port, in this case 3.

Operation manual 04.04.19 14/79


Ultrasonic milk analyzer

3. QUALIFICATION OF RAW MILK, THERMALLY TREATED


MILK, OTHER DAIRY PRODUCTS AND DERIVATIVES

3.1. Taking samples and preparation for analysis

In order to receive reliable results in qualification of milk, dairy products and


derivatives are needed: precise samples taking; correct samples storing (in
need to be preserved); correct preparation before making measurement. The
rules and requirements for this are described in details in Appendix Preparing
Samples.

3.2. Making the measurement.

3.2.1. Preparing the analyzer for working mode

3.2.1.1. Put the analyzer on the working place, providing good ventilation and
not in the vicinity of heat providing devices or sources. The temperature in the
premises has to be in the boundaries 10-30oC.

3.2.1.2. Check if the power switch is in "0" position and that the outlet voltage
complies with the voltage indicated on the rating plate of the analyzer.
Connect the power supply cable to the electrical outlet.

3.2.1.3. Switch on the “POWER” button, which starts the identification


procedure. For a short time the display shows the number of the software
versions, for example:

Milkanalyzer xxx
MB vers yy
Ser. N. xxxx

where:

Milkanalyzer xxx is the time for measurement.


MB vers YY is the motherboard software version.
Ser. N. xxxx is the serial number – written on the rear panel of the analyzer.
These data are called analyzer’s Identity

Operation manual 04.04.19 15/79


Ultrasonic milk analyzer

If in the process of exploitation there is a need to ask a question the


company-producer, you have to send the data, written on the display during
the above described initialization procedure i.e. the analyzer’s identity.

3.2.1.4.Till the analyzer is prepared for work (at about 5 minutes) the
following message is written on the display: “Getting ready”. Above pointed
time is in dependence of the environmental temperature and increases with
decreasing the temperature.

3.2.1.5. When the device is ready for work the display shows: “Ready to
start”.
The analyzer is ready to make analysis in mode 1 (normally Cow)

3.2.1.6. If you want to pass to another mode press the button Enter and hold
it pressed. The following message appears on the display:

Release button to
start menu

Release the button Enter. The display shows the possible working modes:

Milk selector
Cal1 – Cow
Cal2 – Sheep
Cal3 – UHT
------------------------------
Cleaning
Printing

Using “up”▲and ”down”▼buttons, choose the working mode and press Enter
in order to start it.

3.2.2. Making analysis


To start measurement: pour the preliminary prepared sample in the sample
holder of the analyzer; put the sample holder in the recess of the analyzer;
press the button Enter.

Operation manual 04.04.19 16/79


Ultrasonic milk analyzer

The analyzer sucks the milk, makes the measurement and returns the milk in
the sample-holder. During the measurement the temperature of the sample is
shown on the display.
Ignore the results received immediately after switching on the analyzer and
after measuring distilled water. Make a second measurement with new
portion of the same sample.

3.2.3. Displaying the results

3.2.3.1. When the measurement is finished, the sample returns in the


sample-holder and the display shows the results. For example:

Results:
F=ff.ff S=ss.ss
D=dd.dd P=pp.pp
L=ll.ll W=ww.ww

Where:

F= ff.ff - measured FAT in percentage;


S= ss.ss - measured SNF in percentage;
D= dd.dd - measured density in percentage;
P= pp.pp - measured protein in percentage;
L= ll.ll - measured lactose in percentage;
W= ww.ww - measured sample’s added water in percentage.

By pressing the button” Down” ▼ the display shows the second page,
containing the results:

Page 2 Results:
T=tt.tC pH=pp.pp
FP=-0.fff
s=0.sss A=aa.aa

Where:

tt.tC - sample’s temperature;


pp.pp - sample’s pH result – if there is a pH probe connected;
-0.fff - measured sample’s freezing point;
0.sss - measured solids values;
aa.aa - measured total solids

Operation manual 04.04.19 17/79


Ultrasonic milk analyzer

By pressing the button”Up” ▲ display shows the third page with results:

Page 3 Results:
L=l.ll

Where:
L= ll.ll - measured Lactose in %;

By pressing the buttons “up”▲ and ”down”▼, the operator has the possibility
to pass from one page result to another.

If the device has an embedded option “Conductivity” and "conductivity


measurement" started, the result is shown on the display, showing the basic
results replacing lactose results in the following way:
C=xx.xx
In this case the Lactose result is shown on a new page - Page 3 Results.
xx.xx is the measured milk sample’s conductivity in [mS/cm]. If the results are
outside the limits for this type of sample (see table from the Appendix
Conductivity measurement), the cursor flashes after the letter С, reminding
that the sample is not correct. On the printout it is printed as !!!.
If the conductiity value is outside measuring range (2-14 mS/cm), the
following message appears on the display:
C=OutRg (Out of Range), and on the printout there isn’t any line with
conductivity value.

3.2.3.2. Write down the results in the form. The results remain on the display
till a new measurement is started. If the analyzer is connected to a computer
or a printer, it sends the data to the computer or prints them.

Mode Printing
Serves for control of the printing. There are 2 variants:
-after switching on the power supply of the device. Then the analyser’s
parameters are printed (Identity).
-after completed measurement. Prints out the results from the last
measurement.

Operation manual 04.04.19 18/79


Ultrasonic milk analyzer

4. CLEANING THE ANALYZER

4.1. Periodically cleaning (rinsing) the analyzer

It is done in the process of routine work of the analyzer. Its aim is to prevent
drying up and adhesion of different milk components in the milk analyzer’s
measuring system.

The company-producer recommends usage of the chemicals, supplied with


the analyzer – alkaline and acidic (Lactodaily and Lactoweekly). You may
order them separately or together with the analyzer. Try to use only these
chemicals for cleaning the analyzer.

In case you missed to order these chemicals, the alternative is to use alkaline
and acidic cleaning solutions for dairy equipment by one the companies,
producing such chemicals, as for example:
http://www.diversey.com
http://www.ecolab.com
http://www.calvatis.com

Do not use chemicals not intended for usage in the milking systems or
vessels in the dairy sector. Pay special attention to the concentration of the
acidic chemical. Increased concentration may damage the measuring
sensor.

4.1.1. Periodical cleaning frequency.

It is easy to understand what is the period on which the rinsing could be done
as the analyzer reminds you when it is necessary. This is done by a sound
signal in 1-second cycle after the set time intervals elapse:
 55 min. after switching on the power supply of the analyser, but idle
work;
 15 min. after the last measurement of real milk sample.

*Idle Mode is that part of the standard working mode, when the analyser is not making
measurements. There’s embedded in the analyser system for measurement of the idle
time. The idle time is measured starting from the last action of the operator. In
dependence of it (what the operator last did), are taken decisions regarding the cleaning.

Operation manual 04.04.19 19/79


Ultrasonic milk analyzer

There are 2 options:


Option A: If the analyser:
1. Was only switched on but was not started in measurement mode,
2. Or the last action was cleaning,
3. Or the last action was measuring sample with very low Fat (similar to water)
Then the signal for cleaning is started after 55 min.
Option B: If the last thing done with the analyser was measurement of normal milk sample,
the signal for cleaning is started after 15 min.

After cleaning completion, new measurement takes place in above described


time intervals.
The following message appears on the display:

Time to start
cleaning

4.1.2. Making the rinsing


After above message is received put in the recess of the analyzer a glass
filled with 120 ml water (in case 1 from p.4.1.1.) or alkaline cleaning solution
(in case 2 from p.4.1.1.).
Press Enter to start the rinsing mode.
In this mode the analyzer makes 3 cycles and stops.
Already used solution is poured out of the analyser. Now the device is ready for
the next measurement. In case of doubt that the analyzer is still not well
cleaned, the procedure Cleaning may be executed repeatedly.

4.2. Complete cleaning

4.2.1. Complete cleaning frequency


This cleaning is done after finishing the work with the analyzer at the end of
the working day or if it is obvious that the measuring system of the analyzer is
contaminated in case of intensive work with it. It is done with alkaline cleaning
solution.
Preparation of 3 % alkaline solution of Lactodaily for circulation cleaning in
the milk analyzer:
1. Take the package 100 g concentrated chemical Lactodaily
2. In appropriate vessel (for example bucket) pour 1 l water.
3. Add the powder and then again water up to 3 l.

Operation manual 04.04.19 20/79


Ultrasonic milk analyzer

For a single cleaning cycle you need only 25 ml cleaning solution. We


recommend you to prepare working solutions of cleaning chemicals, enough
for normal work for 1 week, because, during their stay unused, the working
solutions loose their strength and also is difficult to store them.

Then follow the instruction for milk analyzer cleaning.


4.2.2. Cleaning

4.2.2.1.Rinsing milk residues


Fill in the glass with water. Put it in the recess of the analyser and start
command Cleaning from the main menu. After finishing it pour out the
contaminated water.
4.2.2.2.Cleaning with alkaline cleaning solution
Fill in the glass with warm (50-60 С) alkaline cleaning solution. Put it in the
recess of the analyser and start the command Cleaning from the main menu.
After finishing it, pour out the contaminated liquid.
4.2.2.3.Rinsing with water
Fill in the glass with water. Put it in the recess of the analyser and start
command Cleaning from the main menu. After finishing it pour out the
contaminated water. Now the device is ready for work.

4.2.2.4. Cleaning with acidic solution


It is recommended to be done every day.

Preparation of 3 % acidic solution of Lactoweekly for circulation cleaning in


the milk analyzer:
1. Take the package 100 g concentrated chemical Lactodaily
2. In appropriate vessel (for example bucket) pour 1 l water.
3. Add the chemical and then again water up to 3 l.

Fig. 5 Labels for the cleaning chemicals

Operation manual 04.04.19 21/79


Ultrasonic milk analyzer

The following procedure is executed:


1. Rinsing the milk residues:
Fill in the glass with water. Put it in the recess of the analyser and start
command Cleaning from the main menu. After finishing it pour out the
contaminated water.
2. Cleaning with acidic solution
Fill in the glass with warm (50-60 С) acidic cleaning solution. Put it in the
recess of the analyser and start the command Cleaning from the main menu.
After finishing it, pour out the contaminated liquid.
3. Rinsing with water
Fill in the glass with water. Put it in the recess of the analyser and start
command Cleaning from the main menu. After finishing it pour out the
contaminated water. Now the device is ready for work.

Please, pay attention that, when the analysers gives a signal for need of
cleaning 15 min after the last measurement of real milk samples or 55 min.
after being powered and not used, cleaning is made ONLY with alkaline
solution in concentration 1-3%.
During the basic/final cleaning sequence is: alkaline solution – water – acidic
solution - water

IMPORTANT
THE MAIN REASON FOR MALFUNCTIONING OF THE
DEVICE IS THE BAD CLEANING OF THE SYSTEM AFTER
MAKING ANALYSIS.
In case of malfunction due to the bad cleaning of the
analyser your guarantee is not valid anymore and any
repair has to be paid.

Operation manual 04.04.19 22/79


Ultrasonic milk analyzer

4.3. Peristaltic pump service


Fig.6 Peristaltic pump

Operation manual 04.04.19 23/79


Ultrasonic milk analyzer

5. POSSIBLE MALFUNCTIONS AND ERROR MESSAGES,


TROUBLESHOOTING

In the table below are described the possible malfunctions during the milk
analyzer’s exploitation and ways for their repair/remedy. If the problem
persists after all recommended measures are taken, please, connect the
nearest service center for help. Do not forget to tell the analyser’s identity.

To receive the analyzer’s identity, refer to point 3.2.1.3.

Error Possible Repair/remedy


message problem
/cause
2 MA Immediately switch off the analyzer.
overheated Pay attention the analyzer to be situated
Overheated
Accompanied away from direct sunlight or heating devices.
milk
by a Wait 5-10 minutes the device to cool down or
analyzer
continuous to be normalized the ambient temperature
sound signal and switch it on again.
The analyzer is ready to measure the next
sample. In order to avoid the future
appearance of the same error message,
please, check the following:
- The sample is prepared according the
instructions and there aren’t air bubbles
in it.
Insufficient - There is a real suction of the sample
quantity of after starting measurement, i.e. it is
the milk obvious that the level of the milk sample
3 Empty sample in the sample holder decreases. In other
Camera sucked in case – there is damage in the suction
the system system.
or air in the - Avoid the end of the suction pipe to be
sample above the surface of the liquid (not
dipped enough).
- Avoid curdling of the milk sample. Clean
immediately if there is a sample curdled
in the system.
- In mode Measurement, after starting the
measurement, remove the sample
holder and see if there is no milk poured

Operation manual 04.04.19 24/79


Ultrasonic milk analyzer

back in the sample holder.


The analyzer is ready to measure the next
sample. In order to avoid the future
appearance of the same error message,
please, check the following:
-The sample is prepared according the
Sucked
4 Sample instructions and its temperature does not
overheated
Overheat exceed the maximum permissible sample’s
sample
temperature.
-Complete the procedure for checking the
analyzer in case of error message Empty
Camera.

Operation manual 04.04.19 25/79


Ultrasonic milk analyzer

6. MAKING CORRECTIONS AND RECALLIBRATION OF THE


DEVICE

In the process of work with the analyser there is a possibility the results to
start differing between the data for some of the measuring parameters when
measured with the milk analyzer and the corresponding reference method of
analysis (Gerber for fat, Kjeldhal for proteins etc). In order to establish the
possible discrepancy and to correct the readings of the milk analyser do the
following:

6.1. Taking samples and preparation of samples for checking the


accurracy of the milk analyser, making corrections and
recalibration

This is a basic moment for the correct checking the accuracy of the analyser
and for making correct and precise correction and calibration. It is
accomplished according Appendix Sampling and preparation of samples for
verification the accuracy of the milk analyzer, making corrections and
recalibration.

6.2. Determination the type of the discrepancy:

6.2.1. Making measurements

Make measurements with different samples (not less than 3) with known
values of a separate parameter (for example fat content), determined by the
known reference methods of analysis (for example Gerber's method for
determination of fat content). For more accuracy it is recommended among
these samples to be also such with values, close to the lowest and highest
bounds for the measured parameters.
Make 5-time measurement for each of the samples. Calculate the average
value for each sample parameter, without taking into consideration the first
measurement for each sample.

6.2.2. Analysing the measurement results

Make comparison between the values of the parameter from the reference
sample and measured with the analyser. Make analysis of the difference
received.

6.2.2.1. If the received differences are relatively constant value for samples
with different content of the analysed parameter, it is necessary to make
correction.

Operation manual 04.04.19 26/79


Ultrasonic milk analyzer

For example
М% of the reference samples: 2,20 3,00 3,80 4,60 5,20
М%average when measuring
with the analyser: 2,38 3,17 4,01 4,79 5,42
Difference: 0,18 0,17 0,21 0,19 0,22

Conclusion: the difference is relatively constant value and correction is


possible to be done with – 0,2 % (see Corrections, p6.3.3)

6.2.2.2. If the differences are not a constant value it is necessary recalibration


to be done.

For example.
М% of the reference samples: 2,20 3,00 3,80 4,60 5,20
М% when measured with the
analyser: 2,02 2,93 3,76 4,75 5,44
Difference: -0,18 -0,07 -0,04 0,15 0,24
Conclusion: It is obvious that the difference is variable value and
recalibration have to be done (See Recalibration, p.6.4).

6.3. Making corrections

6.3.1. Possible corrections, limits and changing steps

Every parameter from each calibration may be separately corrected. Below is


the table with possible corrections, limits and changing steps:

Parameter Increasing Decreasing Step


FAT 0.95% 0.95% 0.01%
SNF 4.75% 4.75% 0.05%
Density 4.75% 4.75% 0.05%
Lactose 0.95% 0.95% 0.01%
Salts 0.95% 0.95% 0.01%
Proteins 0.95% 0.95% 0.01%
Added water 9.00% 9.00% 1.00%
Sample’s temperature 9.90oС 9.90oС 0.1oС

Operation manual 04.04.19 27/79


Ultrasonic milk analyzer

6.3.2. Preparing the analyzer for mode Corrections


6.3.2.1. Press the button Enter and without releasing it switch on the power
supply of the device, wait for the starting identification messages and release
the button after the following message appears on the display:

Release button
to start setup

After releasing the button on the display is shown:

Setup Menu

followed by possible to be entered by the operator menus:

Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Accessories
Exit

6.3.2.2. By using buttons “up” ▲ and “down” ▼position on Corrections and


press Enter.

6.3.3. Making correction

6.3.3.1 Determining the correction mode


When starting Corrections, the following appears on the display:

Corrections:
Measurement
Temperature
Cond measure
Exit

Operation manual 04.04.19 28/79


Ultrasonic milk analyzer

Position on Measurement and press Enter. By using buttons “up”▲


and”down”▼position on the corresponding calibration (for example
Correction 1 – cow) and press Enter.

6.3.3.2. Choosing correction parameter


After choosing calibration mode the display shows the following:

Cal1 Cow
Param:Fat
Correct=00.00

Edit OK Next

Using the buttons “up”▲ and ”down”▼position on the action you want to take
(for example Edit) and press the button Enter.

6.3.3.3. Making correction


After choosing parameter (for example fat) the display shows the following:

Cal:….
Param:….
Correct= 00,00
- OK +

Using the buttons “up”▲ and ”down”▼is possible to increase or decrease the
value of the measured parameter in the above pointed limits. Leaving this
mode means saving the correction value and activating it.

6.3.3.4. Making verification


After the corrections are made put the milk analyser in working mode and
make several times measurement of reference samples with known values of
the corrected parameter. If the difference between the values of the
parameter from the reference methods and milkanalyser are in the limits for
the parameter it may be considered that the correction is successfully made.
If the discrepancy between the measurements from the milk analyser and
classical methods is bigger than is necessary to make second correction
according above described way.
If after the second correction the results are unsatisfactory we recommend
making a calibration of the analyser. In dependence of the conditions and
your requirements you may make the calibration using a personal computer

Operation manual 04.04.19 29/79


Ultrasonic milk analyzer

type IBM PC and the company's calibration program or autonomous - by


recalibration.

When making corrections or calibrations be 100% sure in the accuracy of the


reference methods result.

6.4. Recalibrating the milk analyser

6.4.1. Running the analyser in mode Recalibrate

6.4.1.1. Press the button Enter and without releasing it switch on the power
supply of the device, wait for the starting identification messages and release
the button after the following message appears on the display:

Release button
to start setup

After releasing the button on the display is shown:

Setup menu

Followed by the possible to be entered by the operator menus:

Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Accessories
Exit

6.4.1.2. By using buttons “up” ▲ and “down” ▼ position on Settings and


press button Enter.

6.4.1.3. Analogically, position on Recalibrate and press the button Enter.

Operation manual 04.04.19 30/79


Ultrasonic milk analyzer

6.4.2. Making recalibration

6.4.2.1. Choosing the calibration mode


After starting Recalibrate, the display shows the following:

Cal: 1 Cow

Prev OK Next

You can choose the type of milk to be calibrated. By pressing ▼ (Next) you can
switch between Cal: 1, Cal: 2 or Cal: 3.
By pressing the button ОК you are choosing the type of calibration.

6.4.2.2. Entering values for the separate sample parameters


The following menu is displayed:

Cal1 Cow
High
FAT=f.ff

Edit OK Next

In this display the results, received by using the corresponding reference


methods from Appendix Methods for high-fat milk analysis must be
entered.
In this menu, with button ▼ (Edit) must be entered the values of the high fat
milk sample

For example:
FAT=05.29

Cal1 Cow
High
FAT=f.ff

- OK +

With buttons ▼(-),▲(+) set the needed value. With next pressing of Enter the
cursor is moved to the next number. After needed value entering completion for
FAT, press Enter (ОК) and you are going back to the previous menu:

Cal1 Cow
High

Operation manual 04.04.19 31/79


Ultrasonic milk analyzer

FAT=05.29

Edit OK Next
With the button ▲(Next), choose SNF and in the same as above
described procedure, enter the value for SNF. After it is finished, press
“Enter” (OK) and you are going back to the previous menu. With button
▲ (Next), choose DEN (density) and enter the value for density; the rest
of the parameters are entered in the same manner – LAC (lactose), SOL
(salts), PRO (protein)

Cal1 Cow
High
PRO=f.ff

Edit OK Next

If you miss to enter some of the parameters of milk, the following warning
message will appear:

You Must Enter


Values > 00.00
Try Again

Then you must press the button Enter (ОК) and enter the missed parameters.
After all the parameters are entered, press Enter (ОК).

You must enter values for all the measured milk parameters!!!

The screen for entering the results, received with the corresponding
reference methods (See Appendix Methods) for the low-fat milk is displayed:

Cal1 Cow
Low
FAT=f.ff

Edit OK Next

In the same way the values of milk with low fat sample are entered.

Operation manual 04.04.19 32/79


Ultrasonic milk analyzer

You must enter values for all the measured milk parameters!!!

In other case the calibration will not be correct.

6.4.2.3. Making recalibration with the available samples


After entering the values for the separate parameters of the sample, pressing
Enter (OK) will display the following menu:

Cal: Cow
Put sample High
5 times

which reminds us to put 5 times the sample with high FAT.

The sample has to be with temperature in the boundaries 15-25°С.

Before each milk measurement stir 2-3 times the milk sample by pouring it
from one vessel to another. The needed quantity is poured in the sample-
holder and it is put in the recess of the analyser. Start the measurement by
pressing the button Enter. The sample is sucked. Appears the following
menu:

Cal: Cow
Put sample: High
5 times
Temp=….

After the sample is measured, appears the following menu:

Cow
High
N1=….. 2=…….

Cal meas=1/5

Operation manual 04.04.19 33/79


Ultrasonic milk analyzer

which reminds us to make the next measurement. Before each measurement


the milk is stirred by pouring it 2-3 times from vessel to vessel. Continue the
procedure till the 5th measurement.
After 5th measurement completion automatically appears the menu, which
reminds us to place the Low fat milk sample:

Cal: Cow
Put Sampl: Low
5 times

Stir 2-3 times the milk sample before each measurement by pouring it from
one vessel to another. The needed quantity is poured in the sample-holder
and it is put in the recess of the analyser. Start the measurement by pressing
the button Enter. The sample is sucked. Appears the following menu:
Make 5 times measurement of the low FAT sample.
After 5th measurement completion automatically appears the menu:

Cal: Cow
Put sample: Water
5 times

Which reminds for 5-times water measurement.


After the 5th measurement appears the menu:

Recalibrated
Power Off-On

This means that the calibration was completed successfully and the analyzer
is recalibrated for cow milk, marked as “Cal: Cow”.
Switch off the power supply of the device and switch it on again.
The device is ready to work with the new calibration.

Next time when the analyser is switched on, it will be ready for work with
those milk types it was just calibrated with.
If calibration with another milk type is needed, do not forget to change the
calibration number for the new type of milk.
Calibration for Sheep milk will be saved as second calibration, UHT – as
third. This sequence may not be followed and calibrations can be saved in
whichever order is needed. Calibration can be done with different liquid dairy
products using 2 representative samples.

Operation manual 04.04.19 34/79


Ultrasonic milk analyzer

Checking the calibration


1. Switch on the calibrated device.
2. Make sure it shows the same serial number as this already calibrated.
For checking, use the third sample with medium FAT content.
3. Measure the milk 5 times in the mode you’ve calibrated it.
In case that the device is not connected towards printer write down the
results.
4. Ignore the first two results.
The rest three could not differ more than 0,05% FAT, 0,07% SNF, 0,7%
Density one from another

Operation manual 04.04.19 35/79


Ultrasonic milk analyzer

7. STARTING THE DEVICE IN A SERVICE TEST/SETUP


OPERATIONAL MODE. MENUS DESIGNATION

7.1. Starting the device in a service Test/Setup operational mode.

In order to start the Setup of the device the operator has to press the button
Enter and without releasing it to switch on the power supply of the device, to
wait for the starting identification messages and to release the button after
the following message appears on the display:

Release button
to start setup

After releasing the button on the display is shown:

Setup Menu

Followed by possible to be entered by the operator menus:

Special modes
Corrections
Settings
------------------------------
Tests
Exit

You may move in the menus by using buttons “up” ▲ and “down” ▼.
If by pressing the button Enter you choose a menu, each menu offers new
points/submenus. When Exit is chosen the device leaves the Setup mode
and returns to normal work.

Operation manual 04.04.19 36/79


Ultrasonic milk analyzer

Due to the continuous improvements made in the milk analyser or due to the
type of the ordered product, it is possible some of the options in the device to
be not active. In this case, if you try to enter the corresponding menu, the
following message will appear: Not available option.

7.2. Menus Function:

7.2.1. Special modes.

Serve for choosing special (technological) working modes. After starting it the
following appears on the display:

Special modes
Calibration
Cycle
Fast cycle
Exit

This mode is normally used in production conditions.

7.2.1.2. Calibration mode


In mode Calibration the analyzer is ready to make measurement and to send
the received results towards the technological milk analyzers calibration
system. For this purpose you need personal computer type IBM PC,
company’s calibration system LSC.EXE and methods for calibration of milk
analyzers (see the corresponding documents). To start measurement in this
mode, the operator has to put a sample-holder containing milk sample in the
recess of the analyzer and to press the button Enter.

7.2.1.2. Cycle mode / Fast Cycle mode


Mode Cycle serves for training the analyzers. When you start this mode, the
analyzer, without additional commands, sucks the sample, makes the
measurement, pours the sample out in the sample-holder and displays the
received results cyclically.
Note:
This mode is normally used in production conditions. It is recommended the
customer to calibrate the device using the embedded function Recalibrate (i.e
autonomous, without computer)

7.2.2. Corrections

Operation manual 04.04.19 37/79


Ultrasonic milk analyzer

Serves for entering corrections in the measured data. Detailed description in


point 6.3.2 and 6.3.3.

7.2.3. Settings.

Serve for assigning different working parameters (modes).

7.2.3.1. Net number.


Serves for assigning the device network number when connecting it in the
production network. The possible numbers are from 0 to 15 including.
After starting this function the display shows the following:

Net number
0
- OK +

By using the button “up”▲ the operator has the possibility to increase the
number, showing the channel’s number, and by button “down”▼, to decrease
it. Pressing the button Enter saves the chosen channel and exits the function.

When connected in the production network each device has to have a unique
number.

7.2.3.2.Recalibrate.
Serves for changing definite calibration. Methods are described in point 6.4.

7.2.3.3. Save/Rest Cal.


Through this menu you may save the new calibration in the device or to
restore the old one (factory) calibration. This is necessary in case that you’ve
calibrated the device for cow milk, but after that the device is not measuring
correctly and you decide to restore the factory calibration settings. Position
the cursor across “Restore calibration” and press “Enter”

Possibilities:
Save calibration – saves the chosen calibration in an internal buffer.
Restore calibration – restores the chosen calibration from the internal
buffer.
The procedure Save/Restore is done for each calibration separately.

Operation manual 04.04.19 38/79


Ultrasonic milk analyzer

Current working Internal storing


calibration (1,2 or 3) Save  buffer

Current calibration content is not changed, the analyzer continues using it,
but there is a reserve copy in an internal buffer.

Current working Internal storing


calibration (1,2 or 3) Restore  buffer

The current calibration is replaced with the calibration from the internal buffer
and the analyzer starts working with it. The content of the internal buffer is not
changed.

If after recalibration “Save calibration” is pressed the new calibration


settings will be saved over the factory settings. After that is impossible
to restore the factory settings of the calibration. Save the newly made
calibration only if you are sure about its correctness.

7.2.3.4. Settings Page 2.


After this menu is started the display shows the following:

Settings Page2
Set Calibr Name
Select High Fat
HFSpeed for Cal
Reslt Precision
PCB Main Identi
Larg Res En/Dis

Exit

Now there is a possibility one of the following options to be set:

Operation manual 04.04.19 39/79


Ultrasonic milk analyzer

7.2.3.4.1.Larg Res En/Dis.


The format of the measurement data send towards the computer is set.
If the option Large Disable, is chosen, then only the main results are send to
the computer – Fat, SNF, Density, Lac, Proteins, Added Water, sample
temperature, device serial number and calibration number.
If the option Enable, is chosen, except the above mentioned parameters also
data for Salts, Freezing Point, pH, Conductivity will be send to the computer. In
this case is necessary the software in the computer to be conformable to the
format of the sent data. After starting the menu, the display shows (for example):

Large Results
No

No OK Yes

7.2.3.4.2. Select High Fat


If the analyzer has embedded function for measurement of high fat products, by
this menu the calibration, with which this measurement to be made is chosen.
What is seen when this option is turned on is an obvious slowdown of the
sample’s suction speed.

7.2.3.4.3. HFSpeed for Cal


If the analyzer has embedded function for measurement of high fat products,
and a new calibration for high fat measurement is needed, before starting the
new calibration the operator has to start this menu. What is seen when this
option is turned on is an obvious slowdown of the sample’s suction speed during
calibration.
Please, pay attention to the fact that switching off the power supply cancels this
command action.

7.2.3.4.4. Reslt Precision


Serves for setting the precision of the measured results shown on the display. It
is set separately for every parameter, the precision could be 0.01 (standard) or
0.1.

7.2.3.4.5. PCB Main Identi


Gives information about the type and the version of the analyser (LS Identity).

Operation manual 04.04.19 40/79


Ultrasonic milk analyzer

7.2.3.4.6. Set Calibr Name.


Sets the names of the separate calibrations. The name could be chosen from
the group of predefined calibrations names or to edit a new one. When editing
the new name there is a possibility all ASCII codes to be used, as letters (caps
and normal), numbers and punctuation marks and popular symbols. The
calibration name consists of 8 symbols.

Example:
When it is suitable to us this possibility of the analyser? For example if you have
a device factory calibrated for Cow milk, Sheep Milk and UHT milk, but you need
often to measure camel milk. Using the methods, explained in details in
Appendix Methods you may make a new calibration without need to send the
analyser back to the producer for calibration. Using this procedure you may
make calibrations for most often analysed milk and to write down the exact
calibration name, which will be shown on the display and printed on the printer.

After starting this menu the display shows:

Cal:1 Cow

\Prev OK Next

There are the following possibilities:


With button OK – to start editing the name of the chosen calibration.
With button Prev – to choose the previous calibration, chosen for setting the
calibration name.
With button Next – to choose the next calibration name for editing.
If the operator has chosen and confirmed calibration for change of the name, the
display shows (example):

Cal:Cow

PreDef Exit Edit

There are the following possibilities:


With button PreDef – to choose a calibration name from the list of preliminary
given names.
With button Exit – to leave the menu.
With button Edit – to edit the new calibration name.

Operation manual 04.04.19 41/79


Ultrasonic milk analyzer

If a name from the preliminary given names list is chosen, the display shows:

Cal1: Cow
New Cow

Exit Yes Next

There are the following possibilities:


With button Exit – to leave the menu.
With button Yes – to confirm the chosen from the list calibration name.Now the
program returns to the beginning of the menu for setting calibration names.
With button Next – to show the next calibration name from the list.

If it is decided a new calibration name to be edited, the display shows:

Cal1: Cow
New Cow
Prev Set Next

There are the following possibilities:


With button Prev – to display the previous ASCII symbol.
With button Set – to confirm the ASCII symbol, shown on the display and
passes to editing the next symbol from the calibration name.
With button Next – to show the next ASCII symbol.

The editing finishes by entering the eighth symbol from the name of the
calibration.

7.2.3.5. Set Base Frpnt.


Serves for editing the Basic Freezing Point. It is used by the customer
according the Appendix for calculation of the added water and determination
of the Freezing point of the sample.

7.2.4. Tests.

Start different tests. Possibilities:

Operation manual 04.04.19 42/79


Ultrasonic milk analyzer

7.2.4.1. Test pump.


Starts pump’s test. The number of the completed suction/display cycles is
indicated.

7.2.4.2. Ultrasound.
Test for the ultrasonic system. Used in production conditions.

7.2.4.3. Set Amplitude.


Serves for ultrasound amplitude adjustment. It is used under production
conditions or by the customer (after sensor change) according the
instructions in the document SetCell.pdf.

7.2.4.4. RS232 COMPort

7.2.5. pH meter & Co meter

Please, use this menu only after reading the above pointed document
SetCell.pdf

7.2.6. Exit

By pressing the button you may leave the program and pass towards another
menu.

Operation manual 04.04.19 43/79


Ultrasonic milk analyzer

7.2.7. Milk analysers’ setup menu structure

Analyzer Setup
Special
modes
Calibration
Cycle
Corrections
Measurement
Temperature
Settings
Net number
Recalibrate

Save Calibr
Restore Calibr
Settings Page2
Set Calibr Name
Select High Fat
HFSpeed for Cal
Reslt Precision
PCB Main Identi
Set Base FrPnt
Tests
Test pump
Ultrasound
Set Amplitude

Operation manual 04.04.19 44/79


Ultrasonic milk analyzer

8. ADDITIONAL POSSIBILITIES OF THE ANALYZER

8.1. Connecting to 12 V DC power supply.


If there is a need the analyzer to work on place without electrical supply
available, then it could be powered by car battery or other 12 V DC external
power supply. Use the 12 V power supply cable, supplied with the analyser.

8.2. Connecting to IBM PC


The milk analyser can be connected to IBM PC using the RS232 interface
cable. In order to make the connection: switch off both milk analyser and
PC.Connect the RS 232 cable towards Serial interface and towards the
computer. Turn on both milk analyser and PC. Now the milk analyser is ready
to communicate with IBM PC. This is enough for starting the program for
collection and archive of the measurement results. For more detailed
information regarding milk collection data program see the file resLS.

8.3. Connecting a printer (option).


In order to print out the measurement results, a serial printer could be
connected to the device – for example ESC/POS Serial printer, production of
Datecs or Seiko. The interface connector for the printer is on the rear panel of
the device – “Serial printer output”. The printer (if it is Datecs), should be
connected to the “12 V printer output” on the device rear panel. Connect it via
cables, delivered by the company-producer. If the printer is connected directly
to the electrical network, then the analyzer and the printer should be
connected to one and the same electrical phase.
Communication parameters: 9600 bps, No parity, 8 bits, 1 stop bit. It’s one-
way communication (uses one line) – the analyzer only sends and the printer
only accepts data.

8.4. Measuring high fat samples (option).

The standard device measures samples up to 25% fat.


On customer’s request, the device could be produced with possibility to
measure samples up to 50% fat. The customer can choose which calibration
to have this possibility and which not, as well as during the process of
exploitation to change the measuring mode i.e. to pass from measuring
normal fat percentage towards high and vice versa.
What the operator sees during these passes is the difference in the speed of
sucking the sample. For that purpose, the high-fat sample has to be
preliminary heated up to 30С +- 3С.
To choose the mode, follow the sequence below:

Operation manual 04.04.19 45/79


Ultrasonic milk analyzer

Setup->Settings->Settings
Page2->Option
Select->SelPumpSpeed->Speed for Cal x

After which the display shows:

Calibr x
Pump Speed

Normal OK HiFat

By pressing the correspondent buttons the operator can choose the type of
measurement and to exit the menu.
When changing the type of measurement on a calibration is necessary a new
calibration of the device on the new speed to be done. When calibrating
measuring high fat sample, before starting the calibration procedure, the
operator has to choose from the menu:

Setup->Settings->Settings
Page2->Option
Select->SelPumpSpeed-> HFSpeed for Cal

By which the device passes in a mode of measuring high fat samples.


This calibration mode is active till the power supply of the device is switched
off i.e. it has to be always set if the device will be calibrated for high fat
measurement.

8.5. Embedded real time clock (option).

On customer’s request, a real time clock could be embedded in the device,


showing astronomical time and date. The clock is powered by battery, so it is
independent on power supply of the device. The advantage of this option is
that on the print out with the measurement results are shown also the exact
time, when the measurement is made, for example:

Operation manual 04.04.19 46/79


Ultrasonic milk analyzer

The embedded clock is controlled by the device’s Setup, from the main menu
Accessories, submenu RT Clock. When chosen, the display shows the
following:

Display Time
Adjust Time
Adjust Date
Exit

Using these menus, the operator has the possibility to show on the display
the current time and date, and if necessary – to adjust them.
The time is shown in the format:

hh:mm:ss
dd:mm:yy

Exit

where:
hh - the current hour
mm - current minutes
ss - current seconds
dd - current day
mm - current month
yy - current year

Operation manual 04.04.19 47/79


Ultrasonic milk analyzer

APPENDICES
APPENDIX 1: PREPARATION OF SAMPLES FOR
MILKANALYSERS’ CALIBRATION

For calibration are needed samples of cow milk with the following
parameters:

Low Fat High Fat Middle


1 Cow 2,2% 5,2% 3,6%

For the calibration are needed:


1. Distilled water
2. Min. 3 milk samples with known values for fat, SNF, protein, density,
lactose, salts.
Calibration samples have to be with low, middle and high values of the
analyzed components. Samples have to be representative for given milk type.
Volume of the sample has to be enough for making min 5 measurements for
each sample – not less than 1,00 l. Changes in the analyzed parameters in
the samples have, if possible, to cover the whole measuring range – i.e. used
samples to be with low, middle and high content of the analyzed components.

Methods of milk samples preparation for calibration.


For milk sample with middle value of the analysed components we
recommend to use milk taken from not less than 10 animals from most
common in the region breed.
Sample with low and high value are prepared on the following way:
1. Pour the fresh milk with FAT at about 3.7% in a separating funnel.
2. Leave the funnel with the milk in refrigerator for 12 hours at temperature
+5-+8 ° C.
3. Draw the substratum of the separated milk in a vessel, mix it well, pour
it and heat it in water-bath up to 20°C.
4. Pour the upper layer in another vessel.
5. Determine the concentration of the measured components (FAT,
protein, SNF, density, lactose, solids) by using certified methods.

The analyser’s accuracy depends only on the correctness of the chemical


analysis of the components in the samples and the normal acidity during
calibration!

It is recommended the first cow milk sample with low fat content to be with
the following parameters:

Operation manual 04.04.19 48/79


Ultrasonic milk analyzer

2-2,3% FAT; 8.7-9% SNF; 3,3-3,5 % Protein; 4,8-4,9% Lactose; 0,75 Salts;
1030-1033 kg/m3 Density.
The second cow milk sample with high fat content to be with the following
parameters:
5-5,3% FAT; 8.4-8,79% SNF; 3,1-3,2% Protein; 4,6-4,7% Lactose; 0,7 Salts;
1028-1029 kg/m3 Density.
If, after milk’s separation you do not obtain samples in the requested range,
then, by adding milk with high fat value into the low fat milk sample you can
obtain necessary value-2,3%
Analogous to this, by adding low fat milk sample into a milk sample with high
fat value you may receive 5,3%
Samples with medium values are received by mixing low fat and high fat
samples in necessary proportion.
If there is a need of longer sample storing they have to be preserved; the
most commonly used preservative is potassium dichromate (K2Cr2O7) - 1 g
for 1 000 ml.
When using samples, stored shortly, preliminary pour the sample from one
vessel to another in order to distribute the milk components evenly paying
attention not to form foam in the sample.
When the samples are stored for a longer period it is recommended to warm
it up to 35-45 °С, and the vessel to be shaken carefully. In case that there is a
cream stuck on the vessel’s surfaces – remove it. The sample is poured from
vessel to vessel several times and is cooled down (advisable to 20 °С /.

If there is separated liquefied fat or white particles with irregular form on the
vessel’s walls reliable results could not be received.

Because it is very difficult both lactose and salts to be measured but they are
substantial and influence in great extend when determine added water. That’s
why it is better both lactose and salts to be calculated by using SNF results.
The milk must be for sure without added water.

If you are unable to make the analysis of milk in certified methods in a pinch
you can use the following formulas:

DETERMINATION OF THE BASIC PARAMETERS IN THE MILK SAMPLE


BY USING FORMULAS IS NOT AS PRECISE AS USING THE ARBITRARY
METHODS, BUT IS SUITABLE FOR USAGE IN FIELD WORK.

Operation manual 04.04.19 49/79


Ultrasonic milk analyzer

1. Determination some of the parameters by formulas

There is dependence between the different parameters in milk and its


density, which may be expressed with mathematical equation. On this base
different formula, tested and confirmed by the classical laboratory methods
for analysis, are developed. We recommend the following:

2. SNF determination.

For determination of SNF the correlation dependence exists between the


milk’s density, fat and SNF in the milk. When the density and the fat are
known, the SNF can be calculated.
There are several formulas with different applicability.

А/ When the Total Solids and fat are known


SNF is calculated by subtracting the fat percentage from the Total solids.
SNF = Total Solids – F (%)
Where
Total Solids in (%),
F – fat content in (%),
This formula is used for determination of SNF in whey, buttermilk, and cream.

B/ Known quantity of fat and density (most commonly used method when
maximum accuracy is needed).
We recommend the following formula:

0,075  F %  100  100 / density


SNF 
0,378

This is a universal formula and actual for milk of almost all kind of cows and
sheep all over the world.

3. Determination of lactose content


We recommend the following formulas:

А/ for cow milk


Lact. = SNF * 0,55 (% )
Where
SNF – content of SNF in percentages (%),
0,55 – constant coefficient.

B/ for sheep milk


Lact. = SNF* 0,45 (% )

Operation manual 04.04.19 50/79


Ultrasonic milk analyzer

Where
SNF –solids-non-fat content in percentages (%),
0,45 – constant coefficient.

This is an actual coefficient for sheep breeds on the territory of the Balkan
Peninsula.

4. Determination of salts content


We recommend using the following formulas:

А/ for cow milk


Salts = SNF* 0,083 (% )
Where
SNF – solids-non-fat content in percentages (%),
0,083 – constant coefficient.

B/ for sheep milk

Salts = SNF * 0,075 (% )


Where
SNF – solids-non-fat content n percentages (%),
0,075 – constant coefficient.

This is an actual coefficient for sheep breeds on the territory of the Balkan
Peninsula.

5. Determination of total proteins content


We recommend using the following formulas:

А/ for cow milk


Protein = SNF * 0,367 (% )
Where
SNF - solids-non-fat content in percentages (%),
0,367 – constant coefficient.

B/ for sheep milk


Protein = SNF * 0,475 (% )

Operation manual 04.04.19 51/79


Ultrasonic milk analyzer

Where
SNF – solids-non-fat content in percentages (%),
0,475 - constant coefficient.

This is an actual coefficient for sheep breeds on the territory of the Balkan
Peninsula.

Operation manual 04.04.19 52/79


Ultrasonic milk analyzer

APPENDIX 2 FREEZING POINT DETERMINATION

1. Methods for determination.

The milk analyzer determines the freezing point of each sample and the
quantity of added water. The milk analyser does not measure the freezing
point, but calculates it from the components it depends on. The basic
components in the milk are water, solids, lactose, FAT, proteins, minerals
(salts) and acids. The freezing point depends only on the diluted in the milk
components and quantity of the solvent (in the milk it is water). The ultrasonic
technology allows direct measurement of FAT, proteins, lactose + salts (the
soluble components, only influencing the freezing point), and the quantity of
the solvent in % is determined by 100 % – total solids %, total solids = lactose
% + FAT % + proteins % + salts % + acids %.
Without understanding the meaning of the freezing point – determined or
shown from the milk analyzer added water result easily may lead to a mistake
for the value of this parameter.

2. The basic freezing point.

Milk freezes at lower temperature than water. The average freezing point of
the raw milk in the most regions is at about -0,540С. The average reading
for your region is called “basic” freezing point.
The freezing point of milk is a “physiological constant”. This does not mean
that it will not vary. In fact feed, breed, season, time of lactation, climate,
whether the sample is taken at the beginning, middle or end of lactation – all
these factors will have an effect on the freezing point of the individual sample.
This means that there is an average value of all these numbers. The more
samples used in obtaining this average, the more reliable it is as a base. Or
the basic freezing point is an average of freezing points of milk, taken from
many cows. When a laboratory checks a producer, it is only comparing the
average of the producer’s cows against a larger area average.
The Health authorities establish the basic freezing point or agriculture
departments in some regions, sometimes by universities, separate dairy
producers, or their associations. Frequently, tolerances have been
established on top of a basic freezing point to allow some variations in the
milk as well as device or operator variations.
Without mentioning the basic freezing point, the Association of Official
Analytical Chemists now recommends an upper limit freezing point at -
0,525C (2,326 standard deviations above the most recently determined
North American average of –0,5404C), below which there will be at 95%

Operation manual 04.04.19 53/79


Ultrasonic milk analyzer

confidence that will show 99% of all freezing point determinations on


unwatered milk:
“if the freezing point is –0,525C or below, milk may be presumed to be free
of water or may be confirmed as water free by tests, specified below. If the
freezing point is above –0,525C, milk will be designated as “presumptive
added water” and will be confirmed as added water or added water free by
tests specified below. Evaluate extreme daily fluctuations in the freezing point
of herd, pooled herd, or processed milk for presence of added water”.
“Presumed added water”, as described above, must be “confirmed” by means
of tests on authentic milk samples obtained as specified in the АОАС
METHODS.
After determination the freezing point of your sample via the milk analyzer,
the added water is calculated using the following formula:

FrPo int Base  FrPo int Calc


AddedWater   100[%]
FrPo int Base

Where:
FrPointBase is the basic freezing point
FrPointCalc is measured freezing point

Note:
If the freezing point is not correctly determined, the result for the added water
is not valid. In this case results for FrPoint and AddWater are not shown on
the display and on the printout from the printer. If the density of the measured
sample is 0, the result for AddWater is not valid and is also not shown on the
display and the printouts.

Sample:
First variant
If you’ve entered for milk analyzer basic freezing point -0.520C (according
article 5.9 of the EU Milk Hygiene Directive 92/46/ЕЕС), measured freezing
point –0.540C, using the above pointed formula you’ll receive –3,8%.
Because it is not possible the added water to be negative value, the milk
analyzer indicates 0% added water. The reason for this is the tolerance in the
basic freezing point, reasons for which are described below.
If in the same milk we add 3,8% water, and the basic freezing point is the
same, the milk analyzer will measure freezing point –0.520C, and will
indicate again 0% added water.

Operation manual 04.04.19 54/79


Ultrasonic milk analyzer

Second variant
If you’ve entered for the device basic freezing point –0.540C, measured
freezing point –0.540C, the milk analyzer will indicate 0%. When you add
3,8% water, the device will indicate 3,8%-added water.
From the above mentioned follows that it is very important to enter correct
basic freezing point in the device.

The device’s results for added water may give information about doubt of
added water in the milk and the exact value of this added water may be
determined after a “cowshed sample” is taken and the result for the freezing
point, measured by the milk analyzer of the “cowshed sample” is entered as
basic freezing point in the formula for calculation of added water.
Then the result from this formula will give us the absolute value of the added
water for the corresponding milk supplier.

Operation manual 04.04.19 55/79


Ultrasonic milk analyzer

APPENDIX 3 PH MEASURING

1. General information

PH probe is a unit, measuring the solution acidity or alkalinity degree. It is


measured on scale of 0 to 14. The term pH is derived from "p", the
mathematical symbol for the negative logarithm, and "H", the chemical
symbol of Hydrogen. The formal definition of pH is the negative logarithm of
the Hydrogen ion activity.

2. pH Electrode

For pH measurement the milk analyzer needs a combination electrode,


compatible with most pH electrodes that have BNC connectors and zero
potential (the pH where the mill volt output of the electrode equals 0) near 7
pH.

2.1. Electrode part


The electrode is the most important part of the pH measurement. The
electrode glass membrane is fragile and must be handled with care. To
protect the glass membrane and to maintain activation, a protective rubber
cap containing a suitable storage solution covers the glass membrane.

2.2. Electrode care & Electrode maintenance


pH Electrodes are susceptible to dirt and contamination and need to be clean
regularly depending on the extent and condition of use. At no time should one
touch or rub the glass bulb as this causes the build-up of electrostatic charge.

2.3. Storage
For best results, always keep the pH bulb wet. An optimal storage solution for
combination electrode is pH 4 buffer with 225 grams of KCl per liter. Table
salt, NaCl, can be used if KCl is not really available. Other pH buffers or tap
water are also acceptable storage media, but avoid storage in de-ionized
water. The protective rubber cap filled with the buffer solution provides ideal
storage for long periods.

2.4. After Use


After measurement is completed, follow the sequence below for storage.
- Wash the electrode and reference junction in de-ionized water.
- Close the refilling hole by returning its rubber sleeve or stopper cap.
(Necessary for only refillable electrode).
- Store the electrode as mentioned above (see section Storage).

Operation manual 04.04.19 56/79


Ultrasonic milk analyzer

2.5. Electrolyte Replacement (for refillable electrode only).


The reference electrolyte needs to be refilled when the electrode has been
used for a long period, or when the internal electrolyte has dried up. To
accomplish this, follow the procedure described below.
- Remove the protective rubber cap or sleeve;
- Remove the protective rubber sleeve to expose the filling port of the
electrode;
- Remove the old reference electrolyte with a syringe;
- Fill the new reference electrolyte.

2.6. New electrolyte preparation:


- Open the KCl container;
- Add in de-ionized water until it reaches the level of 20 ml;
- Close the container and shake it to dissolve the KCl;
- Add in fresh electrolyte until it reaches the level of the refilling port. The
reference electrolyte used should be 3M(Mol) KCl;
- Replace the rubber sleeve.

2.7. Re-use the electrode.


- Rinse the liquid junction with de-ionized water.

If these steps fail to restore normal electrode response, you may attempt to
rejuvenate it (See: Electrode Rejuvenation).

2.8. Electrode cleaning


Electrodes which are mechanically intact can often be restored to normal
performance by one or combination of the following procedures.

- Salt deposits:
Dissolve the deposit by immersing the electrode in tap water for ten to fifteen
minutes. Then thoroughly rinse with de-ionized water. Wash the electrode pH
bulb in a little detergent and water. Rinse electrode tip in with de-ionized
water.

- Oil/Grease films:
Wash electrode pH bulb in a little detergent and water. Rinse electrode tip
with de-ionized water.

- Clogged Reference Junction:


pH electrodes have junction, which allows the internal fill solution of the
measuring electrode to leak out into the solution being measured. The

Operation manual 04.04.19 57/79


Ultrasonic milk analyzer

junction can become clogged by contamination in the solution. If a clogged


junction is suspected it is best to clear the junction.
Heat up the diluted KCl solution to 60-80°C. Place the sensing part of the pH
electrode into the heated KCl solution for approximately 10 minutes. Allow the
electrode to cool while immersed in some unheated KCl solution.

- Protein Deposits
Prepare 1% pepsin solution in 0.1 M HCI. Allow the electrode to stand in this
solution for five to ten minutes. Rinse the electrode with de-ionized water.

2.9. Electrode activation


Generally, if the procedure of storage and maintenance had been closely
followed, the electrode can
be used immediately. However, should the electrode response become
sluggish, it may be possible that the bulb has dehydrated.
The bulb can be dehydrated by immersing the electrode in an ideal storage
solution (e.g. buffer pH 4 solution) for 1-2 hours. If this fails, the electrode
may require re-activation. If the above procedure does not reactivate the
electrode to acceptable status, try rejuvenation the electrode by following the
procedure outlined below.

2.10. Rejuvenation Procedure


Dip and stir the electrode in freon or alcohol for 5 minutes.
Leave the electrode in tap water for 15 minutes.
Dip and stir the electrode in concentrated acid (HCI, H2S4) for 5 minutes.
Leave the electrode in tap water for 15 minutes.
Dip and stir in strong base (NaOH) for 5 minutes.
Leave the electrode in tap water for 15 minutes.
Test with standard calibration solution.
Finally, test with standard calibration buffer solution to see if the electrode
yields acceptable results. You may repeat again for better response
(maximum 3 times). If the response does not improve, then the electrode has
completed its useful life. Replace with a new electrode.

2.11. Electrode Lifespan


pH electrodes have a finite lifespan due to their inherent properties. How long
a pH electrode will last will depend on how it is cared and the solution it is
used to measure. Even if an electrode is not used it still ages. Electrode
demise can usually be characterized by a sluggish response, erratic readings
or a reading, which will not change. When this occurs an electrode can no
longer be calibrated. pH electrodes are fragile and have a limited lifespan.
How long an electrode will last is determined by how well is maintained and
the pH application. The harsher the system, the shorter the lifespan. For this

Operation manual 04.04.19 58/79


Ultrasonic milk analyzer

reason it is always a good idea to have a back-up electrode on hand to avoid


any system down time.

3. Buffer Solutions

Buffers are solutions that have constant pH values and the ability to resist
changes in that pH level. They are used to calibrate pH measurement
system.

PH buffer solution description ( Pharmacopoeia standard)

Use only this types standard buffers for calibration!

Description pH 7.00±0,01/20ºC pH 4.00±0,01/20ºC


Composition Potassium dihydrogen Borax, Sodium
phosphate, Di-sodium hydroxide solution
hydrogen phosphate
Temperature 10ºC - 7.06 10ºC - 4.00
parameters 25ºC - 6.99 25ºC - 4.00
20ºC - 7.00 20ºC - 4.00
30ºC - 6.98 30ºC - 4.00
40ºC - 6.95 40ºC - 4.00
50ºC - 6.91 50ºC - 4.05

4. pH Electrode Calibration

pH Electrodes are like batteries; they run down with time and use. As an
electrode ages, its glass changes resistance. For this reason, electrodes
need to be calibrated on a regular basis. Calibration in pH buffer solution
corrects for this change.
Calibration is an important part of electrode maintenance. This assures not
only that the electrode is behaving properly but that the system is operating
correctly.
Usually pH meters require calibration at 3 specific pH values. One calibration
is usually performed at pH 7, second and third are typically performed at pH 4
and pH 10.

It is best to select a buffer as close as possible to the actual pH value of the


sample to be measured. Use standard calibration buffers that the
temperature and the sample solution are the same.
Use the operation manual for the corresponding pH meter.

Operation manual 04.04.19 59/79


Ultrasonic milk analyzer

For Sensorex pH electrodes, originally supplied with Lactoscan read the


following information:

Temperature compensations
The output of pH electrodes varies with temperature in manner, predicted by
theory. When needed, Sensorex can supply electrode holders with build-in
automatic temperature compensators. The need of automatic compensation
depends on the temperature variation, the pH value being measured. At pH
of about 7 there is no error due to temperature and, of course, at a constant
temperature there is no error. As shown in the following table, the pH error
due to temperature is a function of both the temperature and the pH value
being measured. At a pH of about 7 there is no error due to temperature and,
of course, at a constant temperature there is no error. The more the
temperature changes from the ambient calibration temperature and the more
the pH departs from 7 the greater is the pH error.

pH temperature error table

C pH
2 3 4 5 6 7 8 9 10 11 12
5 .30 .24 .18 .12 .06 0 .06 .12 .18 .24 .30
15 .15 .12 .09 .06 .03 0 .03 .06 .09 .12 .15
25 0 0 0 0 0 0 0 0 0 0
35 .15 .12 .09 .06 .03 0 .03 .06 .09 .12 .15
45 .30 .24 .18 .12 .06 0 .06 .12 .18 .24 .30
55 .45 .36 .27 .18 .09 0 .09 .18 .27 .36 .45
65 .60 .48 .36 .24 .12 0 .12 .24 .36 .48 .60
75 .75 .60 .45 .30 .15 0 .15 .30 .45 .60 .75
85 .90 .72 .54 .36 .18 0 .18 .36 .54 .72 .90
0 pH Error Range Less than .1 pH Error Range

5. PH helpful hints

For greatest accuracy in pH measurement, follow these guidelines:


Use the same technique to measure samples, which was used for calibration.
Be consistent with stirring rates, times and conditions.
Calibrate with buffers, which are close in temperature to that of the sample.
Calibrate the pH electrode regularly, e.g. once an hour for accuracy to within
0.01 pH, or once a day for accuracy to within 0.1 pH.
Use fresh buffers for calibrations. Avoid contamination of the stock buffer
solution and do not use it beyond the expiry date.
Keep all connections dry.

Operation manual 04.04.19 60/79


Ultrasonic milk analyzer

Immerse the electrode far enough into the solution to insure the reference
junction is below the surface.
Allow adequate time for the electrode to stabilize in standards and samples
before taking a reading.
Clean the electrode periodically. Allow more time for aged electrodes.
Do not use the pH electrode in solutions of fluoride ion at low pH. This will
etch the glass membrane.
Sulphide vapors can permeate the electrode wick and contaminate the
reference element. Minimize contact in such environments and change the
reference electrolyte frequently.

1. Preparation for pH measurement


When the analyzer is with pH measuring option, it is received from the
customer with pH probe packed separately and there’s a stopper on its place.
If you need to measure pH follow the procedure below:
1. Loosen the nut anti-clockwise.
2. pull up the stopper
3. Carefully place the pH probe paying attention not to remove the sealing O-
ring.
1. Place the probe with the nut in the hole and tighten it.

Operation manual 04.04.19 61/79


Ultrasonic milk analyzer

Fig. 7 Placing the stopper

It is very important to close the nut tightly, paying attention not to allow air to
enter the system.

Operation manual 04.04.19 62/79


Ultrasonic milk analyzer

Fig. 8 Placing the probe

Operation manual 04.04.19 63/79


Ultrasonic milk analyzer

If you are working with the analyser regularly (each day) do not remove the
probe after work..

If you ‘ll not use the analyser more than 2 days, you must take out the probe
and to place the stopper back.
The pH probe must be stored separately as per the instructions of point 2.3
Appendix 3.

7. PH measuring.

Measuring pH is an additional feature of the analyser and is optional.


Remove the protective rubber cap of the pH electrode. Take care to handle it
appropriate in order not to be damaged. Use de-ionized or distilled water to
rinse the electrode before usage. Fill in the sample holder with milk, put it in
the recess of the analyser and dip the pH electrode into the milk sample,
ensuring complete dip of the electrode in the sample. Stir gently for
homogenization of the sample.
Measuring can be done in two modes:

Operation manual 04.04.19 64/79


Ultrasonic milk analyzer

Off line by starting the menu pH & Co Meter | Measuring, when the
analyser works only as a pH meter.

On line automatic pH measuring, when measuring the rest of the sample’s


parameters.

When starting work with рН meter first connect the probe/sensor, and then
the power supply of the device.

Having in mind the characteristics of the process of pH measuring it is


necessary to dip the рН probe in the sample and then to press the button
Enter.

After starting the menu pH & Co Meter the following message appears on
the display:

pH Calibration
pH Measuring
pH En/Disable
pH U Display
-------------------------
pH Test
Co Meter Calibr
Co Meter Test
Co Meter En/Dis
-------------------------
Exit

Operation manual 04.04.19 65/79


Ultrasonic milk analyzer

8.pH Calibration.
Fig. 9 Calibration buffers

Take out the sensor. You may not place the nut, because the analyzer do not
sucks during calibration.
Serve for pH meter’s calibration. For this purpose 2 sample buffers are used,
shown on the display as Low buffer (for example 3.00 pH) and High buffer
(for example 7.00 pH). Follow the procedure:

Please, use this procedure only in case that you have enough quantity
calibration buffers, as they could not be used second time.
If you have no enough buffers, then place the pH probe in the vessels near
the analyser.

1. Start the Calibration menu.


2. Put the probe in the Low buffer.
3. Using the analyzer’s buttons enter the exact buffer value. The following is
shown on the display:

pH Calibr
Put Izopot buff
Buf=xx.xxx

Operation manual 04.04.19 66/79


Ultrasonic milk analyzer

4. The operator has to enter the buffer’s value, when the probe is in its
isopotential point and to press the button Enter.
After that the display shows:

pH Calibr
Put Izopot buff
Buf=xx.xxx
V=x.xxxV Set

Where x.xxxxV is the measured in the probe voltage.


1. Press the button Set when the readings stop moving. The analyzer
automatically takes out the calibration liquid and the display shows the
following:

Put filled with


Water glass
And Enter press
To continue

2. The operator has to place a glass filled with distilled water and to press
Enter, in order to clean the probe from previously used calibration
liquid. After the cleaning the analyzer is ready to start working with the
next calibration liquid.

3. Repeat the procedure with the Next buffer. The following message
appears on the display:

pH Calibr OK

Which shows that the calibration procedure was completed successfully. The
calibrated device is ready for making measurements.
8. The device automatically passes in mode pH measuring.
9. Check the correctness of the calibration by measuring buffer solution 7.00.

Operation manual 04.04.19 67/79


Ultrasonic milk analyzer

9. pH Measuring.
After starting this menu the measurement is done in mode off line, i.e. the
analyser works only as a pH meter. The operator has to dip the probe in the
sample and on the display the following is shown:

pH measuring
x.xxxV
y.yy pH
Exit

Where:
x.xxx – measured by the probe voltage
y.yyy – measured probe’s pH
By pressing the button Exit, the operator may exit the program and to pass
towards another menu.

10. pH En/Disable
Serves for enabling/disabling the pH measuring during normal work of the
analyser - On line. After starting it the display shows:

pH Measuring
XXX
No OK Yes

Where:
XXX is the current situation of the working mode. By pressing the buttons
below the corresponding inscriptions it could be changed, as Yes – means
that during normal work of the analyser – measuring the rest of the
parameters, pH will also be measured. If No is chosen, then pH is not
measured.

11. pH U Display
Serves for allowing/forbidding the value of the pH probe voltage during pH
parameter measuring. After starting it the display shows:

PHUDisplay
XXX

No OK Yes

Operation manual 04.04.19 68/79


Ultrasonic milk analyzer

Where XXX is the current state of the displaying mode. By pressing the
buttons below the inscriptions it could be changed, as Yes – means that
during pH measuring the voltage of the pH probe will be shown.
If No is chosen, it will not be shown. It refers to both of the measuring modes.
12. pH test
serves for testing the measuring system in production mode.

Operation manual 04.04.19 69/79


Ultrasonic milk analyzer

APPENDIX 4 CONDUCTIVITY MEASURING

1. Method of determination.
Conductivity (or Electrolytic Conductivity) is defined as the ability of a
substance to conduct electrical current. It is the reciprocal of the resistance.

In a healthy animal*, the mean value of electric conductivity is:

Milk type Conductivity values


Cow milk between 4 to 6 mS/cm (18°C);
Sheep milk between 3 to 5 mS/cm (18°C);
Buffalo between 2,5 to 5 mS/cm (18°C);

*These values depend on the geographical region, the breed and on other
factors.

Milk conductivity changes on the concentration of ions in the milk:

Added water, sugar, Decrease the ion's concentration. Milk conductivity


proteins, insoluble decreases.
solids
Added salts Increase the ion's concentration. Milk conductivity
increases. Increase the ion's concentration. Milk
conductivity increases. Often the milk is falsified
by adding salt: towards milk with good
characteristics: fat 4%, SNF 8,8, conductivity 4,5
are added salt and water. Then the results are
changed to 3,2 and 8,8, conductivity 10. In other
words adding water regulates the increased value
of SNF and density till normal (within the
boundaries/parameters) and even the fat is
normal. By the values of these parameters may be
determined if the sample is falsified, but the only
characteristic, proving this is conductivity, which is
out of boundaries nevertheless added water. But
be careful, as the falsification is not the only
possible reason for conductivity increasing. The
other possibility is mastitis that’s why we
recommend using another (chemical) method for
checking it.
Significantly extreme Should indicate the development of mastitis.
value (6,5 - 13,00 Infections damage the tissue of the udder. This

Operation manual 04.04.19 70/79


Ultrasonic milk analyzer

mS/cm (18°C) allows sodium and chlorine ions from the blood to
be released into the milk. The concentration of
ions in the milk is thereby raised, and it can more
easily conduct an electrical current - the
conductivity of the milk increases.

Milk conductivity can be used as tests for degree of water evaporation in


condense milk production.
Milk conductivity change notifies of powder (dry) milk solution rate.

2. Conductivity measurement
Conductivity measurement is additional possibility of the analyser and is
delivered on customers request/

3. Co Meter Calibr
Serves for conductivity measuring system calibration. Clean the analyzer
before starting conductivity measurement. (see p. 4.1). You need a standard
buffer with conductivity 5.02[mS/cm] (you may order it for delivery together
with the analyser), with temperature 18oС. After starting this mode, the
analyzer makes preparation for measurement and when it is ready, the
following message is displayed:

Co Meter Value
Base= 5.02

The basic value of the buffer solution is shown. We can use buffer solution
with another conductivity (from 4 to 5 mS/cm) or refer to the Note at the end
of this point. The value of the used buffer must be changed or confirmed.
Pressing ENTER confirms, + or – increase or decrease the value. After 3rd
pressing of ENTER the following message appears on the display:

CoMeter Calibr
Put new sample
And press Enter

The buffers' temperature is indicated during measurement. After finishing the


measurement the following message appears on the display:

Operation manual 04.04.19 71/79


Ultrasonic milk analyzer

Co meter Calibr
Put new sample
And press Enter
ADC=xxxxx/1

Where xxxx is the result from the first calibration measurement. The operator
have to put a new buffer, N.B. do not use one and the same buffer more than
once! Then start the next measurement. This procedure has to be repeated 5
times. At the end the following message appears on the display:

CoCalibr-OK
xxxx xxxx xxxx
Xxxx xxxx Diff=xxxx
Power Off/On

Now the operator has to switch off the power supply of the analyzer. After
switching it on again, the analyser has to be cleaned again with water, which
ends the calibration of the conductivity measurement system calibration.

Note:
Another possibility for calibration of analyzer’s conductivity measurement
function.
You need conductivity meter. First measure milk with normal acidity with
conductivity meter and use it as sample for calibrating the analysers
conductivity measurement function.

4. Co Meter Test.
Serves for testing the working mode of the milk's sample conductivity
measurement system. It is used in the production conditions. After this menu
is chosen, the analyser executes the procedure for sample's measurement
and the display shows the data, used for obtaining the samples conductivity.

Co Meter Test
CoADC= xxxx
Power Off - Stop

Operation manual 04.04.19 72/79


Ultrasonic milk analyzer

5. Co Meter En/Dis.
Enables or disables the conductivity measurement system. The following
message appears on the display:

Cond Measuring
Yes

No OK
Yes

6.Corrections in conductivity measurement

It is done by starting the menu Corrections -> Cond measure. You have the
possibility to increase/decrease the measured conductivity value from – 1.00
till +1.00, with step 0.01. After starting this function the display shows the
following:

Con Meter
-1.0<=Corr>=1.0
Co Corr=+0.00
Edit – Up/Down

The cursor is positioned below the +. By using buttons Up/Down, the


operator has the possibility to change the value (number). By pressing the
button Enter, the operator confirms the chosen value and moves to the next
position for editing it. After the last position is edited, if the correction value is
within allowed boundaries, the following is displayed: Co Corr Saved, which
means, that the correction is entered and saved. On the contrary – it returns
at the beginning and expects valid correction.

7. Conductivity calibration buffer preparation

In order a standard buffer for conductivity measuring to be prepared follow


the instruction below:
1. Take the packet with the powder buffer.

2. Carefully shake the packet in order to gather the powder at the bottom.

3. Cut one end of the packet.

Operation manual 04.04.19 73/79


Ultrasonic milk analyzer

4. Empty its content in a measuring mug with 1 l volume, paying attention


all its content to be emptied.

For standard buffer: 5,02 ms – 3,056 г

5. Add 600-700 ml distilled water, which was preliminarily deaerated in


vacuum dryer or boiled and then cooled down to 20 ˚С.

6. Shake the mug till the powder is fully dissolved.

7. Add distilled water to the mark.

Operation manual 04.04.19 74/79


Ultrasonic milk analyzer

Contents:
FUNCTION ....................................................................................................................................................................... 6

2. TECHNICAL PARAMETERS ................................................................................................................................... 7

2.1. WORKING MODES CHARACTERISTICS: ........................................................................................................ 7


2.1.1. MEASUREMENT MODE MILK / DAIRY PRODUCT – FIRST TYPE ................................................................................ 7
2.1.2. MEASUREMENT MODE MILK / DAIRY PRODUCT – SECOND TYPE ............................................................................ 7
2.1.3. MEASUREMENT MODE MILK / DAIRY PRODUCT – THIRD TYPE ............................................................................... 7
2.1.4. CLEANING ............................................................................................................................................................. 7
2.1.5. PRINTING .............................................................................................................................................................. 7
2.2. MEASURING RANGE: ............................................................................................................................................ 8

2.3. ACCURACY: ............................................................................................................................................................. 8

2.4 CORRECT AMBIENT CONDITIONS: .................................................................................................................. 9

2.5. DIMENSIONS: .......................................................................................................................................................... 9

2.6. CONTINUOUS WORKING TIME: ........................................................................................................................ 9

2.7 MILK SAMPLE VOLUME PER ONE MEASUREMENT: .................................................................................. 9


FIG. 4 PERIPHERALS CONNECTION ....................................................................................................................... 12
3. QUALIFICATION OF RAW MILK, THERMALLY TREATED MILK, OTHER DAIRY PRODUCTS AND
DERIVATIVES .............................................................................................................................................................. 15

3.1. TAKING SAMPLES AND PREPARATION FOR ANALYSIS ......................................................................... 15

3.2. MAKING THE MEASUREMENT. ....................................................................................................................... 15


3.2.1. PREPARING THE ANALYZER FOR WORKING MODE ............................................................................................... 15
3.2.2. MAKING ANALYSIS ............................................................................................................................................. 16
3.2.3. DISPLAYING THE RESULTS .................................................................................................................................. 17
MODE PRINTING ........................................................................................................................................................... 18
4. CLEANING THE ANALYZER ................................................................................................................................ 19

4.1. PERIODICALLY CLEANING (RINSING) THE ANALYZER ........................................................................ 19


4.1.1. PERIODICAL CLEANING FREQUENCY. .................................................................................................................. 19
4.1.2. MAKING THE RINSING ......................................................................................................................................... 20
4.2. COMPLETE CLEANING ...................................................................................................................................... 20
4.2.1. COMPLETE CLEANING FREQUENCY ..................................................................................................................... 20
4.2.2. CLEANING ........................................................................................................................................................... 21
FIG. 5 LABELS FOR THE CLEANING CHEMICALS............................................................................................................. 21
4.3. PERISTALTIC PUMP SERVICE................................................................................................................................... 23
FIG.6 PERISTALTIC PUMP .............................................................................................................................................. 23
5. POSSIBLE MALFUNCTIONS AND ERROR MESSAGES, TROUBLESHOOTING ..................................... 23

6. MAKING CORRECTIONS AND RECALLIBRATION OF THE DEVICE ....................................................... 26

6.1. TAKING SAMPLES AND PREPARATION OF SAMPLES FOR CHECKING THE ACCURRACY OF
THE MILK ANALYSER, MAKING CORRECTIONS AND RECALIBRATION ................................................ 26

6.2. DETERMINATION THE TYPE OF THE DISCREPANCY: ............................................................................ 26


6.2.1. MAKING MEASUREMENTS ................................................................................................................................... 26

Operation manual 04.04.19 75/79


Ultrasonic milk analyzer

6.2.2. ANALYSING THE MEASUREMENT RESULTS .......................................................................................................... 26


6.3. MAKING CORRECTIONS ................................................................................................................................... 27
6.3.1. POSSIBLE CORRECTIONS, LIMITS AND CHANGING STEPS ...................................................................................... 27
6.3.2. PREPARING THE ANALYZER FOR MODE CORRECTIONS ........................................................................................ 28
6.3.3. MAKING CORRECTION ......................................................................................................................................... 28
6.4. RECALIBRATING THE MILK ANALYSER ..................................................................................................... 30
6.4.1. RUNNING THE ANALYSER IN MODE RECALIBRATE .............................................................................................. 30
7. STARTING THE DEVICE IN A SERVICE TEST/SETUP OPERATIONAL MODE. MENUS
DESIGNATION .............................................................................................................................................................. 36

7.1. STARTING THE DEVICE IN A SERVICE TEST/SETUP OPERATIONAL MODE. ................................... 36

7.2. MENUS FUNCTION: ............................................................................................................................................. 37


7.2.1. SPECIAL MODES. ................................................................................................................................................. 37
7.2.2. CORRECTIONS ..................................................................................................................................................... 37
7.2.3. SETTINGS. ........................................................................................................................................................... 38
7.2.4. TESTS.................................................................................................................................................................. 42
7.2.5. EXIT .................................................................................................................................................................... 43
7.2.6. MILK ANALYSERS’ SETUP MENU STRUCTURE ...................................................................................................... 44
8. ADDITIONAL POSSIBILITIES OF THE ANALYZER ....................................................................................... 45

8.1. CONNECTING TO 12 V DC POWER SUPPLY. ................................................................................................ 45

8.2. CONNECTING TO IBM PC .................................................................................................................................. 45

8.3. CONNECTING A PRINTER (OPTION). ............................................................................................................. 45

8.4. MEASURING HIGH FAT SAMPLES (OPTION)............................................................................................... 45


8.5/ EMBEDDED REAL TIME CLOCK (OPTION). ............................................................................................................... 46
APPENDICES ............................................................................................................................................................. 48
APPENDIX 2 FREEZING POINT DETERMINATION ........................................................................................... 53
1. METHODS FOR DETERMINATION. .............................................................................................................................. 53
2. THE BASIC FREEZING POINT. ..................................................................................................................................... 53
1. GENERAL INFORMATION ........................................................................................................................................... 56
2. PH ELECTRODE ......................................................................................................................................................... 56
2.1. ELECTRODE PART .............................................................................................................................................. 56

2.2. ELECTRODE CARE & ELECTRODE MAINTENANCE ................................................................................ 56

2.3. STORAGE ................................................................................................................................................................ 56

2.4. AFTER USE ............................................................................................................................................................. 56

2.5. ELECTROLYTE REPLACEMENT (FOR REFILLABLE ELECTRODE ONLY). ....................................... 57

2.6. NEW ELECTROLYTE PREPARATION: ........................................................................................................... 57

2.7. RE-USE THE ELECTRODE. ................................................................................................................................ 57

2.8. ELECTRODE CLEANING .................................................................................................................................... 57

2.9. ELECTRODE ACTIVATION ............................................................................................................................... 58

2.10. REJUVENATION PROCEDURE ....................................................................................................................... 58

Operation manual 04.04.19 76/79


Ultrasonic milk analyzer

2.11. ELECTRODE LIFESPAN.................................................................................................................................... 58


3. BUFFER SOLUTIONS .................................................................................................................................................. 59
4. PH ELECTRODE CALIBRATION .................................................................................................................................. 59
5. PH HELPFUL HINTS ................................................................................................................................................... 60
1. PREPARATION FOR PH MEASUREMENT ................................................................................................................ 61
FIG. 7 PLACING THE STOPPER ............................................................................................................................... 62

FIG. 8 PLACING THE PROBE.................................................................................................................................... 63


7. PH MEASURING. ........................................................................................................................................................ 64
8.PH CALIBRATION. ...................................................................................................................................................... 66
FIG. 9 CALIBRATION BUFFERS ............................................................................................................................. 66
9. pH Measuring. .................................................................................................................................................... 68
10. pH En/Disable ................................................................................................................................................... 68
11. pH U Display .................................................................................................................................................... 68
12. pH test ............................................................................................................................................................... 69
1. METHOD OF DETERMINATION. .................................................................................................................................. 70
2. CONDUCTIVITY MEASUREMENT ...................................................................................................................... 71

3. CO METER CALIBR ................................................................................................................................................ 71

4. CO METER TEST. .................................................................................................................................................... 72

5. CO METER EN/DIS. ................................................................................................................................................. 73

6.CORRECTIONS IN CONDUCTIVITY MEASUREMENT ................................................................................... 73

7. CONDUCTIVITY CALIBRATION BUFFER PREPARATION .......................................................................... 73

Operation manual 04.04.19 77/79


Ultrasonic milk analyzer

GUARANTEE CARD

LACTOSCAN SP

Standard model Wide Display

Guarantee period is 1 (one) year after purchasing date.


Improper handling, transport and storage will invalidate the guarantee.
Guarantee is void if warranty labels are removed.

Serial № Date of purchase:

Password:

Distributor:

Signature:

Stamp:

Operation manual 04.04.19 78/79


Ultrasonic milk analyzer

GUARANTEE CARD

Purchaser:

Service report:

Service Delivery
Damage Signature
entry date date

Covers:
Lactoscan SP
Last edited: 29.03.2019

HEAD OFFICE:

4, Narodni Buditeli Str.


8900 Nova Zagora
BULGARIA
Phone/Fax: + 359 457 67082
www.lactoscan.com
www.milkotronic.com

Operation manual 04.04.19 79/79

You might also like