0 ratings0% found this document useful (0 votes) 448 views82 pagesUT Ruane Notes
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
ULTRASONIC
INSPECTIONcee Y
BASIC PRINCIPLES .....
Introduction to the basic concept
The nature of sound.
The acoustic spectrum
‘THE PROPAGATION OF SOUND
The ultrasonic beam
Side lobes
The ultrasonic pulse
Resolution
Pulse repetition frequency (PRF)...
Modes of propagation
Boundary waves... :
Factors affecting the propagation of ultrasound.
Acoustic impedance
Couplant
Attenuation
The decibel (4B).
SOUND GENERATION snnmnnnmnnmnnninnnnnnnnnnnnn
The piezo electric effect
Reflection, refraction and Snell’s law
Mode conversion
Diffraction
Critical angles
EQUIPMENT
Probes...
ee UT4-4
Probe selection UT4-5
‘The ultrasonic flaw detector (flow diagram of a typical A-scan flaw detector) UT4-6
Calibration blocks and their uses... UT48
Block No.1, A2, V1, DINS4/120 or dutch block nnn UT4-8
0° compression probe uses Ur4s
UT49
‘Shear probe uses... .
Block no2, Ag, V2, DINS4/122 o kidney block snmnnensnnnnnnsnnnsee STAD
Compression probe uses...
Shear probe uses.
Institute of Welding (OWYAS block.
Equipment checks
0° PROBE SCANNING
Catia : Urs.
To calibrate a 0° probe to a range of 0 to 100 mm........ bosses UTS-1
Calibration exercises UTS-2
Accurate measurement. ne = see UTS-2
Multiple back wall method. — UTS-3
PeeTABLE O
[roy
Defect detection,
Sensitivity :
Graphs and DAC curves.
Scanning patterns 0° probe
Sizing methods 0° probe
ANGLE PROBE SCANNING..
Calibration
Angle probes test sensitivity
Scanning patterns
Skip factors ssn .
‘The ratio ofthe sides ofthe triangles inthe three most common probe angles... UT6-5
The irradiation factor.
Plotting systems.
Sizing methods angle probes...
TESTING TECHNIQUES
A,B& C scanning
Pulse echo systems
Through transmission testing
‘The tandem technique
Immersion
ULTRASONIC THICKNESS SURVEYING
Acceptteject criteria
Reporting...
ULTRASONIC WROUGHT PLAT
Technique
UT9-2
uTIO
UTIO-1
Defects in plate material .
ULTRASONIC WELD TECHNIQUE
Technique so
Defect signal interpretation UTI0-3
ULTRASONIC TESTING OF FORGINGS senso UTA
General UTIL
‘Technique UTH-L
Defects in forgings UTIH-2
Accept and reject criteria UTIL-4
Reporting. nen uri
ULTRASONIC TESTING OF CASTINGS . neem UTD
General... “ : ae seo UTIQL
‘Technique nnn oe UTI2-1
Defects in castings UTI22
Accept and reject criteria uTi2-5
Reporting, UTI2-5
BRITISH STANDARDS. = APPENDIX A.
British Standards relating to ultrasonic testing smn APPA-1
Peo NedNTEN
BL
APPENDIX B
. APPENDIX C
FORMULAE USED IN ULTRASONIC TESTING essen
TABLE OF ACOUSTICAL VELOCITIES
Table of acoustical velocities in different materials....... “ APPC-1
TABLE OF ACOUSTIC IMPEDANCES APPENDIX D
Table of acoustic impedances for different materials... -APPD-1
ATTENUATION FACTOR APPENDIX E
Example method for determining the attenuation factor of a material APPE-1
EXAMPLE CALCULATIONS sn me APPENDIX F
Example calculations used in ultrasonics APPF-1
Buane &
one TP ON TPONe!0
20
Note I: Modern digital flaw ©
detector use more recent
Aspley technologies suchas
plasma or LCD soreens.
0
0
Noe 2 The echo at At ithe
result of sound energy
Teflecting back of the front
‘uface ofthe specimen
logether with the ringing of
the eral andthe inital
pe all merged ino one
Signal emslope
00
BASIC PRINCIPL
BASIC PRINCIPLES
INTRODUCTION TO THE BASIC CONCEPT
The most common technique used in ultrasonic testing is the pulse echo technique.
This makes use of the phenomenon that sound waves travel in straight lines and are
reflected by an obstacle placed in their path
‘The mechanism is just the same as audible sound waves bouncing off a brick wall and
an echo being received. The strength of the echo is controlled by the size of the wall
Also, if the time lapse between sending and receiving the echo is measured, it is
possible to determine the distance to the wall.
Given the required instrumentation we can pass sound waves through solid materials
and receive echoes from the back wall of the material. If a defect is present in the
‘material then the sound energy would be reflected back from it and give an echo earlier
than that from the back wall because the sound has not travelled as far, The strength or
amplitude of this echo will be an indication of the size of the defect and the distance
travelled by the sound will tellus its depth.
This then is the basis of ultrasonic testing.
‘The instrument that produces the sound energy is called the probe and the echoes are
shown on a cathode ray tube! (CRT) within a flaw detector.
cer
Probe a
Al
A
. BL
‘Sound waves VAY
Vvvw \
TITAS eT TOD
vy
Sound energy is twansmitted from the probe into the test specimen at surface "A"
producing an echo at Al?, Some of the sound is reflected by the defect at "B" and the
resulting echo appears at BI, The remainder of the sound continues through the
specimen to be reflected by the back wall "C", the echo from the back wall appearing
aC
If the screen is calibrated from a test block of known thickness then the depth of the
4efect from the specimen surface (A to B) can be read off the screen
‘THE NATURE OF SOUND
Sound is caused by mechanical vibrations.
In order for sound to pass there must be a medium that will support mechanical
vibrations therefore SOUND CANNOT TRAVEL IN A VACUUM.
‘The particles (molecules) within the medium vibrate passing on energy fiom one t0
another giving the effect of sound movement through the material,
Inet 060103 vuTi4BOS
The density and elasticity ofa
‘medi are aso the rain
factors tha affect the 10
velocity
20
co
50
o
ote: Velocity is sometimes
denoted bythe eter ‘c
n
Note: The maximum
Freguene the human ear cary
tletect reduces with age. [148
‘generally acceped that most
People wll have heard all
"he high frequency sounds
‘hat they are lable to
encounter y the time they
reach ten years of ae.
Ley PANS Com auNetat
‘The ability to support sound depends on the elasticity and density of the medium.
Since these properties will vary, from one material to another, some materials will pass
sound more easily than others,
Sound follows a waveform:
Wavelength
3
2 Timelitance
veLocrry isthe distance moved in unit time
WAVELENGTH ithe distance benveen successive peaks ofa wave
PERIOD isthe time taken for one complete cycle
FREQUENCY is the umber of yees per second
Leycle persecond = 1 Hertz (Hz)
1 Kilohertz (KHz) 1,000 Hz
1 Megahertz (MHz) 1,000,000 Hz.
Wavelength
Wavelengths a function of frequency and velocity
Wovelengh= Yeloeiy gg aw
quency 7
Two ve fa and t= %
THE ACOUSTIC SPECTRUM.
Manual contact testing range
« . steals
Infrasonic Sonic (audible)| Ultrasonic = 1MHz ‘SMHz
(VO Ann Ahnu
« ST
Frew” \ I Tt Oey
J AY VV
1 i
tz okt: SOOKE 2sMz
“Normal test range->
a uraPPO
K jcors:
Extrome (D6 intensity) edge = 1.22
50% eget dB = 056
10% edge20 dB = 1.08
»
30
90
7
100
TO Cee at
PROPAGATION OF SOUND
THE PROPAGATION OF SOUND
‘THE ULTRASONIC BEAM.
dno eam edge
rari ‘Vistang ot reece
> beam cee
(0008 ines)
eal
beam edge
fez (om inca)
‘The dead zone
Seen on the CRT as an extension of the initial pulse, the dead zone is the ringing time
of the crystal and is minimised by the damping medium behind the crystal. Flaws or
other reflectors, lying in the dead zone region of the beam will not be detected. The
dead zone can be seen at the start of the trace on a CRT displaying A-scan, but only
‘with single crystal probes,
The dead zone increases when the probe frequency decreases.
The near or fresnel zone
In this region of the beam, the sound intensity is variable owing to wave interference,
therefore, reflectors or flaws lying in this zone may appear smaller or larger than theit
actual size. The signal heights displayed on the CRT are unpredictable so it is
desirable to keep the near zone length to a minimum,
‘The near zone length can be calculated using the following formula:
Dp? Dxt
Near zone length (mm) = 2— or D_*
sath (mm) =
Where: D = crystal diameter (mn)
2.= wavelength (mm)
f= probe frequency (Hz)
st material velocity (mm/s)
It can be seen from the formula that the near zone can be decreased by decreasing the
crystal diameter or decreasing the probe frequency.
‘The far or fraunhoffer zone
Beyond the near zone the far 2one exists. In the far zone the beam diverges resulting
in a decay in sound intensity as the distance from the erystal is increased, just as a
beam of light from a torch gets weaker the further it travels
‘The amount of beam divergence depends upon the crystal size and the wavelength as
shown in the following formu:
sing K4 gp KEY
Sine DE
Whee: @ — ~ thehalfangle D = crystal diameter (mm)
K = aconstant = probe frequency (Hz)
= wavelength mm) v= atleral velocity (rm)
tee 8083 ur21Peel
2»
30
0
0
co
0
TON eRe HLT D
Te) FEL M ECONO eLe
It may be seen from the above beam spread formula, that the beam divergence can be
decreased by increasing the crystal diameter or by increasing the probe frequency.
Jnfortunately this will extend the length of the near zone. So in probe design there is a
compromise to obtain a minimal beam spread and a short near zone.
beam extreme e486 96 intensity
10% intensity edge
half angle
e 50%
tensity edge
‘beam centre 100% intensity
‘near zone. 50% intensity edge
crystal
10% intensity edge
beam extreme edge 0% intensity
In the far zone of the ultrasonic beam there is no wave interference therefore the sound
intensity in this zone is predictable.
‘The sound intensity reduces from 100% in the centre to 0% at the edge of the beam ,
therefore when the centre of the beam hits a reflectoriflaw the amplitude of the signal
‘on the CRT will be at its maximum.
‘The sound intensity will also decrease with a greater distance (in the range axis) to a
reflector or flaw.
In the far zone the amplitudes of reflected sound from large and small reflectors follow
different laws.
LARGE REFLECTORS (larger than the width of the ultrasonic beam) follow the
INVERSE LAW - The amplitude is inversely proportional to the distance, ie. if the
distance is doubled then the signal amplitude is halved (ic... reduced by 64B).
SMALL REFLECTORS (smaller than the width of the beam) follow the INVERSE
SQUARE LAW - The amplitude is inversely proportional to the square of the distance,
ic. if the distance is doubled then the amplitude from the second reflector is one
quarter of the amplitude of the nearer (124B less).
‘Small reflectors
ke
Large reflectors
kk
of af
nee _ m
th
woemem
8
aaa UT22Buane &
Peon
20
«0
©
”
”
SIDE LOBES
Side lobes are secondary lobes to the primary ultrasonic beam or main lobe that are
formed at the face of a transducer and radiate away from the main lobe. They
represent areas of high and low acoustic intensities and may cause unwanted echoes to
be received by the probe, especially on rough surfaces, which may be mistaken for
flaws on the CRT.
For shear wave probes, the minimum refracted beam angle in steel is approximately
33° to 35°, but at these relatively acute angles, side lobes may be formed which,
although usually negligible, may cause spurious indications on the CRT. For this
reason itis usually safer to set the minimum beam angle for shear wave probes in steel
at 40,
‘The narrower the main lobe, ie. the smaller the half-angle of the beam, the weaker and
‘more numerous the side lobes.
saci oer
THE ULTRASONIC PULSE
In a modern ultrasonic pulse echo flaw detector the pulse of ultrasound is created by
charging a capacitor in the circuitry then suddenly releasing this charge of electrical
energy, about IKv to 2Kv, into the probe. This electrical energy is converted into a
mechanical vibration by the piezo electric crystal in the probe. The ultrasonic
vibrations are formed by the collapse of the crystal after the electrical energy has been
removed, The behaviour of the crystal, on collapse, can be likened to the behaviour of
a spring when it is stretched then released. The spring, will return to its former shape
then shorten then stretch, etc., until it finally comes to rest in its original shape. This
cycle of expansion and contraction is what forms the ultrasonic pulse.
Maximum,
expansion.
Maximum|
contraction
tee #064108 ur23Buyano &
TPONe!
Boo
0
0
0
00
IT UT2+ THE Coo)
Pulse length
This length of pulse is unacceptable since in order to show separate, clear reflected
signals on the CRT then the pulses of sound must be short and sharp. To shorten the
pulses the ultrasonic crystal must be damped with a backing medium which absorbs the
Sound energy (in much same way as a shock absorber fitted to a spring on a motor
vehicle dampens the vibration of the suspension). In this way the pulse length can be
reduced to between 3 and 5 cycles.
gn cycle
Damped pulse
Amplitude Time/distance
The ideal pulse length would be approximately two cycles but such levels of damping
are difficult to achieve with conventional backing mediums and commercially
available crystals.
DAMPING, then controls PULSE LENGTH (the number of cycles x wavelength).
‘The other factor that controls pulse length is probe frequency. The higher the
frequency the shorter the wavelength, i.c. the length of cach cycle in the pulse and
hence the shorter the pulse length (containing the same number of cycles).
ENGTH controls RESOLUTION.
PULS
Perret ur4Note: PAE. is sometimes
‘called tinebose frequency.
Oe
ATION OF SOUND
RESOLUTION
Resolution is the ability to separate on the timebase two or more reflectors that are
0
close together in terms of beam path length,
Consider two reflectors within the beam with a beam path, length, difference of 3mm.
If the pulse length was greater than 3mm then the signals from the two reflectors
would be contained within the same envelope, as in (a). If the pulse length was less
than 3mm then, in practical terms, the signals would be separated, as in (b).
20
x0 : :
O1234567891 O12345678 510
40 @ )
The above therefore demonstrates that the shorter the pulse length, the better the
resolution.
PULSE REPETITION FREQUENCY (P.R.F.)
The pulse repetition frequency (p.r-f.) or pulse repetition rate ([Link].) is the number of
pulses of ultrasonic energy that leave the probe in a given time (usually per second),
Each pulse of energy that leaves the probe must return before the next pulse leaves
“otherwise they collide causing "ghost" or spurious echoes to appear on the CRT. The
time taken for the pulse to travel from the probe and return is known as the transit time.
«| ‘The time between pulses leaving the probe is known as the clock interval. ‘Therefore it
ccan be stated that the transit time must be shorter than the clock interval or ghosting
‘occurs. Practically speaking the clock interval should be around five times the transit
time.
so
NCE TRAVELLED (mm
TRANSIT TIME (usec) = DISTANCE TRAVELLED (nm)
» VELOCITY
CLOCK INTERVAL
1
SK INTERVAL (600-1 Minimum = TRANSIT TIME
coe P.R-F. (MHz) Practical = 5 x TRANSIT TIME.
»
*
100
ts UT2SPBK
Sony
Noe: Compression waves
fare produced in see ifthe
Incident angle ofthe ean In
Derspes less than
pprovimately 27.4%
20
30
50
Noe: Shear wowes only are
produced in steel the
Incident angle ofthe beam in
perspec between
‘proximately 28° and S6™
o
80
©
100
PROPAG.
MODES OF PROPAGATION
Compression or longitudinal waves
Probes that produce compression waves will normally have an incident and refracted
angle of, or close to, 0°
‘These waves travel through a medium
ceasing the particles of te material to
‘oscillate parallel to the direction of wave
propagation and consist of altemate
compression and dilation pressure waves.
4
sso
0 00m00d 0 6 0,0 0 0
$14 ponerse ° empen Sos)
an
‘ Congreve velo in sel = 5960 mi
at Compresion wares can propane
Chrno tigen pene
OG ‘el pare bonding (s ondion dt
Dirctcn af ropseton ‘only exists in solids) isnot essential.
Shear or transverse waves
Particles vibrate at 90° to the direction of
propagation and have @ whip like action
aero)
Shear velocity in steel = 3240 mvs
Shear waves can only propagate in sali
rigid particle bonding being a pre-requisite.
= Incident angle
tenet en uT26Ruane &
TPONe!
0
so
80
100
BOUNDARY WAVES
‘These forms of propagation can only occur when a solid to gas interface is present. If
the objects were immersed, these modes would be fully attenuated
Surface or rayleigh waves
Surface waves are formed when shear waves refract to 90°. The whip-like particle
vibration of the shear wave is converted into an elliptical motion by the particles
changing direction atthe interface with the surface.
These waves are not often used in industrial ND-T. although they do have some
applications in the aerospace industry. Their mode of propagation is elliptical along
the surface of a material, penetrating to a depth of one wavelength. They will follow
the contour of a surface and they travel at approximately 90% the velocity of shear
Depth of Direction of propagation
penetration
(one wavelength)
Direction of particle motion
Where sharp changes in contour occur, such as a comer edge, reflected energy will
return to the probe.
Plate or lamb waves
Plate waves are formed by the introduction of surface waves into thin plate material
‘They are a combination of compression and surface or shear and surface waves
‘causing the plate material to flex by totally saturating the material,
There are two types of plate waves:
DODDDDID ~ suerrene
Symmetrical “<> <> <> <> <> Congitudinal wave)
platc waves
Surface wave
v v
~*~ ___* = --
Plate di f ’
t t
fora) NSF XSL Shear wave)
ee
rtm fy
ee
ewe W088) um7Buane &
TP ON!
"anisotropic: The erin are
‘random i rtetaion and 10
Dave diferent ease
properties in diferent
‘directions.
x0
0
Nove: Velocity is sometimes
denoted bythe eter “e
{constan veloc)
0
LGM PER ELL ocean OaSe os)
FACTORS AFFECTING THE PROPAGATION OF ULTRASOUND
‘The propagation of ultrasonic waves in a material is dependant on the density and
clastic properties of that material and the type of wave transmitted
‘The practical considerations which will affect propagation will include:
‘the test material's grain size
+ attenuation (absorption and scatter effects)
+ acoustic impedance ofthe test material
+ characteristic impedance of inclusions
+ diffraction
+ lack of homogeneity
* anisotropic! materials
ACOUSTIC IMPEDANCE
Acoustic impedance (Z) is the resistance of a material to the passage of ultrasound. It
is the product ofthe material density (p) and sound velocity (¥).
ie Z = pv
Itis the acoustic impedance difference between two different materals/mediums which
governs the intensity of ultrasound reflected from the interface between them.
Conversely, the amount of ultrasound passing from one material to another depends on
this difference between the two materials, This difference is expressed as the acoustic
impedance rato.
‘Theoretically if an ultrasonic wave was passed through two materials, with the same
acoustic impedance (1:1 ratio), in intimate contac, then no reflection would occur, i.
100% transmission of sound would occur. In practice itis very dificult to achieve
intimate contact without a coupling medium (see next section). ‘The couplant would
have a different acoustic impedance to the material and so would affect the amount of
sound reflected
‘The amount of energy reflected at an interface can be calculated with the following,
formula:
Where Z1 and 22 are the
% Reflected energy =( 21°22) x 100 respective acoustic
Zia 22
impedances of the two
materials.
It can be seen from the formula that:
HIGH ACOUSTIC IMPEDANCE RATIO (eg, 20:1). = MORE REFLECTED ENERGY
LOW ACOUSTIC IMPEDANCE RATIO (eg, I:1) = MORE TRANSMITTED ENERGY
It can also be seen from the formula that the same amount of energy is reflected,
regardless of which direction the sound is travelling across the interface.
we 064 ur2-8Ruane &
TPONe!
NOTES
Nove: The deal acoustic 10
Impedance ofcouplant
‘shouldbe in becom the
couse impedance ofthe
probe andthe acoustic
Impedance ofthe test
‘material The ideal thickness
ofthe layer of couplant
should be one quarter of th?
‘wavelength of sound dhrough
Some recently developed
irasonc stems se nO
‘couplant these are kxowm as
‘tr coupled systems and thes
te very powerful
‘aplication and sensive
‘received eieuy
0
0
70
PONT PEM ULB OO oer
‘COUPLANT
Because of the very high acoustic impedance ratio of air to a solid material almost
100% of the energy is reflected at an interface between them (the basis of flaw
detection). Therefore to enable the sound energy to transmit more readily into the test
specimen we have to exclude any air that may be present between the probe and test
surface. This is achieved by substituting the air with a material that has a closer
acoustic impedance ratio to the probe and test material. This is known as a couplant.
Common couplants are: water, oil, grease, polycell, swarfega and glycerine
‘The selection of couplant is sometimes based on the post-test use of the material being
tested, e.g. water based couplants may cause rusting or corrosion but are easier to clean
off in preparation for painting or coating when compared to oil or grease, which may
actually protect the material from corrosion.
Viscosity of the couplant may also be a consideration, ideally rough surfaces require a
more viscous couplant to effectively fill the air gaps more uniformly. Whatever
couplant is used for calibrationisetting the search sensitivity, this must be used
throughout the subsequent inspection,
ATTENUATION
Attenuation is defined as the loss in intensity of the ultrasonic beam as it passes
through a material and is dependant upon the physical properties of the material.
‘The wo main causes of attennation are SCATTER and ABSORPTION,
Scatter
This is the major cause of attenuation and is the redirection of the sound waves
reflecting off grain boundaries, porosity and non-metallic inclusions, etc., and becomes
‘more apparent on the inspection when the size of grains become of the wavelength of
the search unit being employed,
Absorption
‘As the sound travels through a material a small amount of the energy is used up by the
interaction ofthe particles, as they vibrate, causing friction which is dissipated as heat.
AAs the frequency of the sound is increased the attenuation increases duc to more
particle vibration (absorption) and increased sensitivity to small reflectors (scatter from
rain boundaries, porosity and inclusions) which is related to the wavelength of the
sound.
Materials such as castings and austenitic stainless steel are highly attenuative due to
their coarse grain structures, etc. ‘The attenuation factor of a material can be measured
and is expressed in dB/mm (sce the appendices for an example).
Natural attenuation also occurs due to the divergence of the beam in the far zone,
i.e, assuming compression probe use, the amplitude of the backwall echo will be
halved (-6dB) every time the distance from the probe is doubled,
- UT2-90
so
@
0
80
0
THE PROP.
THE DECIBEL (DB)
“The decibe isa logarithmic base unit used to compare sound intensities.
Because we do not know the actual energy being transmitted by a probe, we can only
compare sound intensities being received and express them as a ratio, eg. twice as
much, ten times as much ee
‘A change in sound intensity, expressed in dB, can be measured by comparing signal
heights on a calibrated CRT. The change in dB is given by the formula:
a= 2010p Where Hi and H2 are the respective sigal heights
By transposing the formula it is possible to determine the ratio of the signal heights
when the dB difference is known,
‘The gain/attenuator controls on a conventional ultrasonic flaw detector are calibrated
in decibels, ic. if we reduce the intensity of ultrasound by 6dB any signal on the CRT
‘will drop to half its original height. If we reduce or increase the intensity by 204B then
the signal will reduce to a tenth or increase by ten times its original height respectively.
It is important to note that on certain flaw detectors, if reject or suppression is used to
remove small unwanted signals from the display, then the linearity of the amplifier,
and hence the other signals, will be adversely affected, ie. a 6dB drop will not reduce
the signal by 50%,
Table of approximate dB drops:
oo 2 Drop
0 10% 0%
4 20% 80%
| 2 25%. Eee
10 33% 61%
is 50% 5096
2 80% 20% }
tae 06008 ur21020
%
0
”
109
SOUND GENERATION
‘THE PIEZO ELECTRIC EFFECT
This is defined as the property of certain crystals to convert electrical energy into
mechanical energy and vice versa, ‘These crystals maybe naturally occurring,
artificially manufactured or grown in solution.
Electrical energy in. > => Blectrical energy out
Piezo electric crystals
These crystals may be X-cut or Y-cut depending on which orientation they are sliced,
from the erystal material. The crystals used in ultrasonic testing are X-cut due to the
rode of vibration they produce (compressional). This means thatthe crystal is sliced
with its major plane (the erystal face) perpendicular to the X axis of the crystal
material
‘Typical eryst lectreal eqnnections
Pulse width
control resistor
Long distance to
avoid static discharge|
Original exystal
Gold or silver condoctors (silvering)
reinforced with chrome for wear resistance
The frequency of the crystal is determined by its thickness and its acoustical velocity
and can be calculated with the formula:
pr =X, Whow FF = Froese From the formula itcan be
2 = Cet mei eecny scen thatthe thinner the
eee ‘crystal, the higher the
frequency.
Piezo electric erystal materials
Natural “Artificially grown Manufactured ceramics
Barium Titanate (BaTiO,)
- Lead Zirconate (PbZrO) |
Quartz Tourmaline | Lithium Sulphate (iSO) | {55 Zirconate Titanate (PT) |
Lead Metaniobate (PbNb;0,)
eesti uT34Buane &
TPONe!
The limitations of modern
ceramic onsial material are
‘ar they have ow
‘mechanical strength Le they
Are bre, a thay ave a
tendency to age. The
ledeantage however i that
they are excellent generators
of ulrasound
2»
30
The Curie temperanre for
Barium Titanate around
100°C 19 120°C although the
piezo lec propertes of
arsum Titanate il tart 0
dlograde at temperatures 2f
0PC and above
re primary reason standard
probes re nt usualy used
fn materials above SO°C
because ofthe possibility
of degradation of the crystal
50
The seconary reason is due
tothe probe shoe
characterises beginning 0
change altering velocity and
therefore she Beam angle
tn shear wave probes
| Speeder: Mirror-ie
»
0
iD GENERATION
Properties of piezo electrie materials
Gyeal mari Tivantgs Lato
Tati Foor pian aes
Quartz Good wear resistance properties
Best received and easily
damped
thium sulphate | Soluble in water
Best transmitter and good |
Som piezo electric properties ee
aaa May be preformed to focus | Te™Perature entical
beam
Good piezoelectric
Lead Zirconate
properties -
Good transmitter and all os
Lead Zirconate Titanate | Fond properties Poor silvering
The polarisation of ceramics
In their natural state the polycrystalline ceramic material's crystals are randomly
orientated and the piezo electric properties cancel each other out. To polarise these
ceramics they are heated up to their Curie temperature and subjected to an clectrostatic
field. The crystals align themselves with the direction of the field, which is maintained
during cooling. This polarised ceramic material then behaves as a piezo electric
transducer until heated again to i's Cure temperature
‘The most common crystal materials in use are Barium Titanate and Lead Zirconate
Titanate,
REFLECTION, REFRACTION & SNELL'S LAW
Reflection
Ultrasonic waves are reflected by objects or interfaces placed in their path. When
striking a specular! reflector the angle at which this reflection takes place is governed
by the law of reflection, which states:
Angle of incidence = Angle of reflection
Refraction
‘This describes what happens to an ultrasonic beam when it passes from one medium to
another where the two media have different acoustical velocities, e.g. from perspex to
steel. The beam changes direction or angle inthe vertical plane.
Incident angle
@
Reflected angle Iocideat angle
Cy) i Medivon 1
Medium
a Refracted sound
Reflected sound Refracted angle
®
7 ursBuane &
TPONe!
2
30
“0
0
0
”
©
0
100
Snell's law
The relationship between the incident angle and refracted angles is governed by Snell's
Tw that states:
Sing VI Where: a = incident angle
Sin V2 B= refracted angle
V1 = velocity in medium 1
V2. = velocity in medium 2
‘
MODE CONVERSION
A change in wave-form from one to another, together with the accompanying change
in velocity, due to reflection or refraction at an interface. An example of mode
conversion that we make use of is when the compression wave, generated by the
crystal in a shear wave probe's perspex shoe, crosses an interface between the shoe and
steel test piece and converts to a shear wave.
‘Another example of mode change that we do not want to occur, ¢.g. shear waves
changing to compression waves. This occurs fairly regularly when carrying out a
critical root scan on a single v fully penetrated weld with a shear wave probe, i.e. some
of the ultrasound entering the root bead can be reflected vertically up to the weld cap
and if a critical angle is exceeded, the wave mode will change from shear to
compression. Accordingly on its return path to the probe, the received spurious signal
displayed on the time base will represent an indication that appears to plot on ful! skip
just outside the weld side wall on the opposite side from the scanning surface.
‘The operator however will not be able to confirm this from the opposite side of the
weld on half skip which, if it had been an actual flaw, he would have expected to do
so. AS this type of mode change/spurious indication gives a fairly characteristic signal
display, an experienced operator would be expected to interpret this effect fairly easily
It is also possible though that mode conversions and/or spurious indications can be
misinterpreted as flaws, particularly if not investigated carefully,
DIFFRACTION
‘This occurs when sound waves pass the tip of a narrow reflector. Some of the sound
scatters off the tip causing waves in different directions that reinforce or cancel out the
original waves. This results in a series of high and low intensity waves radiating out
from the tips, giving the impression of sound bending around the edges of the defect.
tz <— Diffracted energy
sa
——1k
Sound waves hs
Le
oS
—
ZS20
20
“0
0
o
0
80
OUND GENER:
CRITICAL ANGLES
TThese are the incident angles in the first medium at which the refracted angles in the
second medium change over from one wave-form to the next. The first critical angle is
where the refracted compressional wave is just about to disappear leaving only shear
waves in the second medium, The second critical angle is where the refracted shear
‘wave has changed to a surface wave.
‘The critical angles can be calculated using Snell's law.
Ist critical angle
Critical angles perspex to stect
In diagram (1) a compression wave (c) is incident on the boundary between perspex
and steel at an angle of O°, At the interface some energy reflects (1) and some is
transmitted across (c) continuing through at 0°. Tn (2) as we increase the angle (a) of
the incident wave (c), inthe perspex ({e, less than first incident critical angle), the
‘wave in the steel (6) refracts (8°), due to the difference in the velocities of perspex
(V1) and steel (v2). However a5 well as the refracted compression wave there will also
bbe a weak shear wave mode (s)- see sketch (2). In (3) if we increase the incident angle
(Ge until the refracted wave (c) reaches 90° (B°(e)), then the incident angle (a) has
reached what we term the frst critieal angle, Following behind the refracted
compression wave isa shear wave (s) and in (4) as we increase the incident angle (c)
sill further, the compression wave internally reflets (1) leaving only shear waves (8)
inthe stel,At (5) if the incident angle (a) is further increased then the shear wave
refacts until it reaches 90° (B°(3). This is what we term the second critical angle.
‘At this point the shear wave bounding along the interface has changed into a surface
‘wave (Gu). Therefore we can sce that (i) compression and shear waves exist in the
second medium when the incident angle is between 0° and the first critical angle. (ii)
Only shear waves exist when the incident angle is between the first and second critical
angles and (ii) surface waves exist only at the second critical angle. Beyond the
second critical incident angle, asin diagram (6), all conventional modes of propagation,
ate reflected internally
ss uT3420
”
0
Calculation of the critical angles for a perspex to stee! interface.
= incident angle B= refracted angle
v1 = compressional velocity in perspex = 2740 mis
v2 = velocity in steel, compressional= 5960 m/s
shear 3240 mis,
1st eritical angle:
wl 2140 mis
sina = 1 « sin sina = 240 ms , sin go°
vie a 5960 m/s
459731543 x 1
a= 278°
2nd critical angle:
vt 2740 mis
sina = “1 x sin sina = 740 ms sin 90°
vis a © = 3240 mis
Sina = 0.845679012 x 1
@ = 577"
At the first critical angle compression and shear waves co-exist, so the lowest angle for
shear waves only in practical use, is just beyond the first critical angle, at an incident
angle of 29°, which gives a refracted shear angle of 35°,
sing = 2240.78 y sin 29°
20 mis
Atthe second critical angle surface waves exist so the highest incident angle we use for
shear waves is 56° that gives an 80° shear wave
So the range of shear wave probe angles in steel (for practical purposes) are 35° to 80°,
produced from incident angles of 29° to 56° in perspex.
tae won UT3sBuane &
Peon
20
+ May aso be referred 100s
vim eral probe.
tal pro 30
“0
50
"Dead sone: Ringing tine 60
oferta
70
EQUIPMENT
PROBES
The angle of a probe used in ultrasonic testing is measured from a line drawn,
perpendicular to the test surface. This line is known as the normal. A 0° probe then is
fone which transmits sound at 90° to the test surface. Also known as a normal probe,
this probe usually transmits compressional or longitudinal waves. A 60° angle probe
would transmit sound at 60° to the normal, i.e. 30° from the surface. ‘The most
common angle probes transmit shear waves (although angled compression probes do
‘exist for special applications) and the manufacturers quote the angle of the probe for
‘use on mild steel
0° combined double* probe
Double probes have two crystals, one transmits and the other receives ultrasound. The
Electrical connections
Backing, Casing
medium
Crystals
‘Cork separator
Perspex shoes
cork separator in between the shoes prevents "cross-talk" or "chatter" between the
crystals. Using oil as a couplant may eventually break down the acoustic barrier and,
produce spurious standing echoes on the display. Having separate crystals eliminates
the dead zone! on the display, enabling the detection of near-surface defects. These
probes are therefore useful for testing thin sections, e.g. thickness gauging and
‘examining for near surface flaws. The crystals may be focused to give a focal point at
the ideal beam path range to be examined.
Single crystal angle probe
Electrical connection Casing
Damping
‘material
Caystal fh
Index point Perspex shoe
‘Single crystal probes have one crystal that transmits and receives ultrasound. The flaw
detector controls the process by transmitting a pulse of energy then switching the
circuit to receive, listening for any returning sound, in between pulses, ‘The circuitry
can be switched quicker than the crystal can be damped. So the receiver picks up the
last few vibrations of the crystal, as it switches in, and displays them on the screen as
the dead zone. This eliminates the possibility of detecting near-surface defects.
ens uratRuane & LON PER LULA GY
TP ON
Angle probes have a perspex shoe, on which the erystal sits, that can be machined to
any angle. The angle of the wedge determines the angle that the ultrasound strikes the
interface (incident angle). This in turn, according to Snell's law, controls the angle that
the sound will propagate through the test material (refracted angle). Damping material
10 | on the back of the crystal (also known as a backing slug) controls the length of the
ultrasonic pulses by absorbing the sound energy, producing short sharp pulses. The
Tength of the pulse is the main factor in determining the resolution of the equipment.
‘The most common damping/backing medium is Tungsten Araldite
SHORT PULSE LENGTH/WIDTH/DURATION MEANS GOOD RESOLUTION,
20
Soft nosed probe
Electrical conneotion
x0 i
Damping Casing
sug
Clamping) coal
ring,
“
Couplant” —\Soft diaphragm
‘This has a soft diaphragm mounted on the front of the crystal, clamped in place by 2
threaded ring, the space in between the diaphragm and the crystal being filled with
so | couplant to expel any air. The soft diaphragm follows the contour of the surface under
test, making this probe ideal for rough or uneven surfaces, e.g. castings or rough
‘machined components.
‘Water gap or gap scanning probe
«
Electrical water
connection in
Probe
70
Nozzle. Column of
water acts as
couplant
» “Test material
v,
Sound path
‘This consists of a water jacket with a nozzle at the end and a probe inside. Water is fed
into the jacket and flows out through the nozzle, forming a column of water, to the test
surface, through which the sound can travel, Because of the flexibility of the coupling
50 | medium, (water) the probe can be used on rough or uneven surfaces. These probes
are usually used in automated ultrasonic scanning systems and can be set up, using a
guide wheel to follow the contour of a component. They can also be used in atrays to
scan a wider area
100
teem) UTs2Ruane &
TPONeI
20
30
0
0
70
30
%0
100
Wheel type probe
Blectrical
connection
Spring loaded
joint
Soft tyre, solid
‘or water filled
Crystal (within axle)
In this probe the crystal is within the axle of the wheel and the sound travels through,
the soft tyre into the test material. The spring loaded joint allows the probe to follow
the contour of the surface so it can be used on rough or uneven surfaces. It is used in
a similar way to the water gap probe. The main advantage of this type of probe is that
it removes the requirement of externally applied couplant, mainly used in aerospace
industries.
Delay line probe
Electrical connection
Damping Casing
= }
Clamping: Crystal
“ye Re 3
onepse
ee
a
The delay line probe is very similar in construction to the soft nosed probe. The
difference is that it has a long perspex shoe clamped in instead of a diaphragm. The
length of the shoe extends the time taken for the echo from the front surface, of the
‘material under test, o return to the crystal. This places the front surface echo (FSE)
further along the timebase, ie. beyond the dead zone, This enables near surface
defects to be located or thin plate to be tested using a single crystal probe. These
probes are usually high frequency probes (which means they have a small dead zone),
but high frequency = long near zone, therefore, to use them for near surface flaw
detection/sizing, the long shoe is used to contain the near zone in the probe not in the
test material,
Ine 06003 UT43Buane & ca} Tayo
TPONe
Magnetostrictive transducers
20
30
Used for detecting defective bar stock, the transducer coil has a magnetic field that is
switching at ultrasonic frequency. This field causes the bar stock to vibrate at an
40| ultrasonic frequency and the vibrations travel along the length of the bar. When the
vibrations reach the other end of the bar, they reflect back and are then picked up by
the transducer (in receive mode) and register on the detector. The equipment is
calibrated off a defect free piece of bar stock to register a specific value on the detector
and defective bar stock is recognised by a change in this value
so
PROBE FREQUENCY, BANDWIDTH & DAMPING
‘An ultrasonic probe transmits sound at a range of frequencies, not just at the stated
Tre freaieney sate nite, | frequency, this is known as the bandwidth. For example a SMHz probe may produce a
eee frequency range of 4 to 6MHz. ‘The bandwidth is also an indication of the damping
Thi he reer oft ae
eee Broad Band Probes ‘Narrow Band Probes
“They are highly damped [ They have low damping |
Have a short pulse length - A longer pulse length
a (typically 1 102 cycles) _ (ypically 3 or 4 cycles)
TA short ringing time (dead 2000) Tong ringing ime (dead zone)
Better resolving power I Poor resolution
Poor penetration Good penetration |
0
%0
100
twee 0608 uresRuane & LONG M CES XO UD
TPONe
PROBE SELECTION
‘The selection of probes for ultrasonic inspection is influenced by various aspects of the
test and the particular material under test. These may include; the type and size of
0
defect being sought, the type of material under test and the distance the sound has to
travel through the material. Probe angle is another consideration when searching for
defects at different orientations throughout the material
Below is a table of properties of probes using the two criteria that we ean select,
ie frequency and diameter
»
Effect of frequency
Low Frequency High Frequeney
wo Tong wavelength _ ‘Short wavelength
More beam spread Less beam spread |
Shorter near 2one “Longer near zone
Better penetration Less penetration
Tess attenuation ~___ More attenuation
fa Longer dead zone L Shorter dead zone -
Less sensitivity Tligher sensitivity
Effects of Diameter
oy | Large Diameter ‘Small Diameter,
~_Less beam spread ‘More beam spread
Longer near zone Shorter near zone
Better penetration Less penetration
co | | Less attenuation (due to beam spread) More attenvation
Difficult coupling on curved surfaces | Easier coupling on curved surfaces
More coverage on flat surfaces Less coverage on flat surfaces
Another consideration is whether to use a single crystal or a combined double crystal
70 | probe. The advantages of a single erystal probe are; better penetration, for the same
size probe as a double, because the effective transmitter crystal diameter is larger, no
focal point, ic. it works effectively over a longer range and cost (cheaper). The main
advantage of a double erystal probe, is that there is no dead zone on the screen, this
‘means better near surface resolution can be achieved.
It can be seen from the tables that higher frequency probes have a higher sensitivity.
In this context, sensitivity refers to the ability to detect small defects. The higher the
probe frequency the smaller the wavelength and the smaller the size of reflector the
probe can detect. It is generally accepted that the smallest reflector a probe can
detect is half the probe's wavelength. So a probe with a long wavelength (low
frequency) will not detect small reflectors, such as small defects or grain boundaries
and so the sound will penetrate further through the material because it is not reflected
at these small interfaces.
0
0
100
few en88 uT4s: