BactoScanTM FC+
The approved rapid method for determination of
total bacteria in raw milk
Based on Individual Bacteria Count (IBC), BactoScan™ FC+ gives an
accurate determination of the hygienic quality of raw milk with a ca- Sample type
pacity of testing up to 200 samples per hour. BactoScan is the indus- Cow, goat, sheep and buffalo milk
trial standard in many countries all over the world, and in the EU more
than 75% of all milk supplies are paid based on BactoScan results. Parameters
Individual Bacteria Count (IBC/ml)
Superior accuracy through Individual Bacteria Count
BactoScan FC+ measures the hygienic quality of milk by analysing bac- Technology
teria in raw milk. As the automated BactoScan FC+ counts the total Flow cytometry technology that enables
bacteria as single cells and not clusters, it gives a uniform high accu- precise and instant milk bacteria analysis
racy of results.
Approvals
High capacity with confidence
BactoScan FC+ is a FDA/NCIMS and MicroVal
High performance capacity enables a high throughput of milk sam-
ples. Reduce labour time and get instant results for fast response to (EURL) approved rapid electronic counting
farmers. The repeatability and reproducibility of BactoScan FC+ is ex- method and complies with ISO/IDF guidelines
cellent compared to plate count methods, while a self-cleaning pro-
gram minimises carry-over contamination from previous samples.
Optimise instrument management with FOSS digital services
Ensure consistent performance of all instruments in your network and
avoid downtime by making upgrades and adjustments while instru-
ments continue to run. Protect your data with automatic back up.
Specifications
Performance
Measuring range: 5.000 - 20.000.000 IBC/ml
Repeatability
Range (IBC/µl) SR (log-units) Typical SR (log units)
10 – 50 0.07 0.06
51-200 0.05 0.04
>200 0.04 0.02
Entire range 0.05
Reproducibility (between instruments):
Range (IBC/µl) SR (log-units) Typical SR (log units)
10 – 50 0.11 0.08
51 –200 0.07 0.06
>200 0.06 0.04
Carry-over effect < 0.5 % (uncompensated)
Sample type Cow, sheep, goat and buffalo
Speed* 65, 130, 200 samples per hour
Working factor Standard 300, (optionally: 95, 600 and 1200)
Accuracy Typical Sy.x < 0.25 log units in the entire measuring range
Reference or anchor method: Standard Plate Count (SPC) (IDF Standard100B:1991)
Application data
Analysis time 9 minutes
Sample intake approx. 4.5 ml
Sample temperature 2 – 42°C (35.6-107.6 F)
Raw milk of normal composition and good quality.
Sample quality
Unpreserved or preserved with azidiol
Instrument management
Networking software FossManagerTM
*BactoScan™ FC+ Semiautomatic is only available at a capacity of 65 samples per hour. This version does not include Conveyor and Stirrer system.
2
Standards and Approvals
BactoScan™ FC+ is CE-labelled and complies with the following directives and regulations:
• EMC (ElectroMagnetic Compatibility) Directive 2004/108/EC
• LVD (Low Voltage) Directive 2006/95/EC
• Machinery Safety Directive 2006/42/EC
• Regulation (EC) 1272/2008 on classification, labelling and packaging of substances and mixture, CLP (EC)
• WEEE Directive 2002/96/EC
• Packaging and packaging waste Directive 94/62/EC
• REACH 1907/2006/EC
BactoScan™ FC+ technology complies with:
• FDA/NCIMS approved
• MicroVal (EURL) approved
• ISO/IDF standards, AOAC Number of national approvals e.g.:
Germany: Bundesanstalt für Milchforshung, France: Ceca Lait.
Data output
Real-time display/print-out, storage on hard disk.
Host transmission (RS232) and PC network transmission (TCP/IP).
Data export using CSV files, CS83 protocol or XML.
Standard equipment
Basic analyser incl. table and reagent containers, PC*, software.
Optional equipment
Printer, extra reagent containers*, ID bar code laser scanner*, 2D reader*, ID bottle rotation*, Conveyor 4000*,
Conveyor 5000 basic*, Conveyor 7*, Conveyor extensions*, Output buffer*, Sample racks*.
*Not included in semi-automatic version.
3
FOSS
Tel.: +45 7010 3370
info@[Link] · [Link]
GB, September 2018