MSE Type-2SterlizationProcedure

This document outlines two sterilization procedures for medical devices: 1. A manual cleaning method involving soaking devices in an enzymatic cleaner solution for 5 minutes, scrubbing with a brush for 15 seconds, and rinsing under flowing water for 1 minute. 2. An autoclave sterilization method placing packaged devices in a sterilizer for 15 minutes at 121°C and allowing them to dry for 30 minutes. The document also lists common medical components, noting that some are single-use only and others are reusable, as well as precautions for pre-sterilized implants.

Uploaded by

GeorgeJC
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
67 views1 page

MSE Type-2SterlizationProcedure

This document outlines two sterilization procedures for medical devices: 1. A manual cleaning method involving soaking devices in an enzymatic cleaner solution for 5 minutes, scrubbing with a brush for 15 seconds, and rinsing under flowing water for 1 minute. 2. An autoclave sterilization method placing packaged devices in a sterilizer for 15 minutes at 121°C and allowing them to dry for 30 minutes. The document also lists common medical components, noting that some are single-use only and others are reusable, as well as precautions for pre-sterilized implants.

Uploaded by

GeorgeJC
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

MSE TYPE-2 STERILIZATION PROCEDURE

1 MANUAL CLEANING METHOD


A. Prepare to Clean
> Prepare a cleaning solution by mixing an enzymatic manual cleaner (<55°C),
or similar cleaning solution with tap water using the recommended concentration
(4ml/l), following manufacturer’s instructions.
> Prepare this solution in a container large enough to fully submerge the device.
B. Soak & Scrub
> Soak the device in the container of prepared detergent solution for a minimum
of 5 minutes.
> Scrub the device for a minimum of 15 seconds with a soft nylon-bristled brush
and/or pipe brush. Scrub the device below water line to ensure contact with
enzymatic cleaner. Articulate moving parts of the device to ensure all surfaces
are scrubbed.
C. Rinse
> Remove the device from the enzymatic solution and thoroughly rinse under
flowing tap (utility) water for a minimum of 1 minute.
> Allow the device to dry.
> After rinsing, inspect the device for visible soil residue. If present, repeat this
procedure until no visible soil remains.

2 AUTOCLAVE METHOD • GRAVITY MOIST HEAT STERILIZATION


A. Clean
> Place each device in appropriately sized sterilization pouches.
> Place all packaged devices into the sterilizer and set for 15 minutes at 121°C.
> Allow devices to dry for 30 minutes.

Component Key:
1. MSE Expander • Single patient use only
2. Micro Implants • Pre-sterilized by gamma ray irradiation**
3. Activation Keys • Single patient use only
4. Mini Hand Driver • Reusable
5. Ratchet Wrench Driver • Reusable
6. Short Engine Blade • Reusable
7. Safety Leashes • Single patient use only
SMPI163Rev081021

**Caution: All micro implants have been pre-sterilized by


the manufacturer. No additional sterilization is required.