ONDOLOG ANALYZER Rev.
00
USER GUIDE
Contents
Contents .................................................................................................................................... 1
1. Download and install software ................................................................................................ 2
2. Connect to Data logger .......................................................................................................... 4
3. Settings .................................................................................................................................. 5
4. Read Logger .........................................................................................................................10
Chart .....................................................................................................................................10
Report summary ....................................................................................................................12
View data ..............................................................................................................................13
5. General information...............................................................................................................14
6. Calibration .............................................................................................................................15
7. Real Time .............................................................................................................................16
8. Error messages.....................................................................................................................17
Setting ...................................................................................................................................17
Date/Time ..........................................................................................................................17
Description .........................................................................................................................17
Alarm .................................................................................................................................17
Password ...........................................................................................................................17
Real Time ..............................................................................................................................18
Calibration .............................................................................................................................18
Chapter: Contents
1
ONDOLOG ANALYZER Rev.00
USER GUIDE
1. Download and install software
- Requirements :
Windows OS : Win 7 , Win 8.1, Win 10
Microsoft Visual C++ 2012 Redistributable
Microsoft .Net Framework 4.5
- Download software at:
[Link]
- Two files [Link] and [Link] will be downloaded
- Install ONDOLOG ANALYZER by following these steps:
Open file [Link]
Choose Everyone to install for all users or Just me to install for yourself. Then click Next,
Chapter: 1. Download and install software
Wait till the installation is completed,
Click Close to complete Setup.
After installing, ONDOLOG ANALYZER shortcut is shown in Desktop.
The background will be :
2
USER GUIDE
ONDOLOG ANALYZER
Rev.00
3
Chapter: 1. Download and install software
ONDOLOG ANALYZER Rev.00
USER GUIDE
2. Connect to Data logger
Connect data logger to PC via micro USB port .
On the logger, there are a Start/Stop key to Start or Stop the logger. Press the key 5 seconds to
Start and press 5 seconds again to Stop logger.
There are two LEDs to indicate, the green LED indicates the logger is recording, the red LED if
there is alarm.
Note 1 : if the Stop key is set No in Settings, the key will be disabled and user can’t stop the
logger by the [Link] this case the logger could be stopped only by software.
Note 2 : if Auto start is used, Start key will be disabled. If user presses the key 5s, the logger
will be stopped.
Chapter: 2. Connect to Data logger
4
ONDOLOG ANALYZER Rev.00
USER GUIDE
3. Settings
Chapter: 3. Settings
5
ONDOLOG ANALYZER Rev.00
USER GUIDE
Logger Information
Logger S/N and Model are read from non erasable memory when connecting logger to
computer.
Non – Stop Recording
If this feature is enabled, the logger will continue writing the data when the memory is full. The
oldest data will be replaced by the latest data.
Start/Stop Button
When YES is chosen, the logger can be stopped by pressing START/STOP button. If NO is
chosen, the logger only stops when the recording is finished or stopped by software. If no-stop
recording, the logger could be stopped only by software.
TimeZone
Display day and time. Time Zone could be selected according to user’s geographical region.
The default setting is done according the time & date settings of the computer.
Description
Chapter: 3. Settings
Description is written down by user. It’s limited to 20 characters and special characters not
available for use.
6
ONDOLOG ANALYZER Rev.00
USER GUIDE
Auto Start
There are two options for this feature.
1. If YES is enabled, then the logger will start and write the data to memory only at the setting
time. Pressing the start/stop button is not available. Display is as follows :
Example, if you set the time at [Link], the day on 25-Apr-16 the logger will auto start at
[Link], 25-Apr-16.
Note : The Date/Time you set must be enough late for the logger to complete settings.
2. If NO is enabled, the logger is started by pressing the Start/Stop button. In this case, data
recording will start only after the START DELAY set by user. The display now is
Start Delay is from 0 to 60 minutes to set the time delay to start to write data to the non erasable
memory of the logger.
Example, if Start Delay is 30, the logger will start to write data 30 minutes after pressing the
button.
Sampling frequency
The interval time is from 2 s up to 32 min. that corresponds to the recording duration of 18 hours
to 365 days.
Security
Chapter: 3. Settings
Tick the box to select password that is used to protect the logger from unwanted user. Then
write down the password and confirm. The password contains 4 characters.
7
ONDOLOG ANALYZER Rev.00
USER GUIDE
Sensors
Below window shows the type of sensor in the logger.
Function tab
Write tab : Write the settings to the non erasable memory of the logger.
Read tab : Read the current settings of the logger.
Clear tab : Clear the settings of the logger
Stop tab : Stop the logger if it is still running.
If pressing the Write or Clear tab when logger is recording, the below message is displayed
Chapter: 3. Settings
If pressing the Stop tab when logger is recording, the below message is displayed
8
ONDOLOG ANALYZER Rev.00
USER GUIDE
Event Log
Here is displayed the status of settings and device.
There are 5 status of device : Running (logger is recording),
Stopped (logger is stopped),
Set (logger is set and in standby mode),
Delaying (logger is waiting for starting),
No setting (logger has no setting).
Chapter: 3. Settings
9
ONDOLOG ANALYZER Rev.00
USER GUIDE
4. Read Logger
Chart
The below chart is displayed after reading the logger.
Icon Description
Open a file
Save
Print screen
Chapter: 4. Read Logger
Report Summary
View Data
Horizontal
Show/ hide Horizontal grid
10
ONDOLOG ANALYZER Rev.00
USER GUIDE
Vertical
Show / hide Vertical grid
Change Color
Edit Comment
Limited in 200 characters, allow special characters
Show/Hide legend
Edit Title
Limited in 50 characters, including special characters
Change Unit
Elapsed Time
Show Value
Point to chart to see value
Chapter: 4. Read Logger
11
ONDOLOG ANALYZER Rev.00
USER GUIDE
Report summary
In report, the chart and summary information of logger are shown. To access Report summary,
click icon
Chapter: 4. Read Logger
12
ONDOLOG ANALYZER Rev.00
USER GUIDE
View data
Recorded data are shown in this file. To access View data use icon
To export file use button and then choose the file type.
Chapter: 4. Read Logger
13
ONDOLOG ANALYZER Rev.00
USER GUIDE
5. General information
Chapter: 5. General information
In general information, user can check the status of logger. There are 5 status : Running (logger
is recording), Stopped (logger is stopped), Set (logger is set and in standby mode), Delaying
(logger is delaying to start), No setting (logger has no setting).
14
ONDOLOG ANALYZER Rev.00
USER GUIDE
6. Calibration
Calibration is used to adjust the temperature or humidity sensors drift.
To calibrate following these steps :
1. Choose the sensor to calibrate by check to Calibration Temp and Calibration Hum
2. Choose unit for temperature (℃ or ℉)
3. Input Reference value
4. Press Write to complete calibration
5. Press Reset to reset the current offset.
Note : The logger must be stopped before calibration.
Chapter: 6. Calibration
15
ONDOLOG ANALYZER Rev.00
USER GUIDE
7. Real Time
In Real Time mode chart, the sensor values are updated every 1s.
Chapter: 7. Real Time
16
ONDOLOG ANALYZER Rev.00
USER GUIDE
8. Error messages
Setting
Date/Time
Date/Time set is smaller than the Date/Time of current clock. The Date/Time must be later than
current clock.
Description
No input data for description. Description could not be empty.
Alarm
Minimum level is higher than maximum level. Click OK and write again the correct value.
Chapter: 8. Error messages
Password
Password could not be empty if security box is checked. Click OK and enter the password.
17
ONDOLOG ANALYZER Rev.00
USER GUIDE
Real Time
Real Time mode can run only if the logger is started.
Calibration
To start calibration, the logger must be stopped.
Reference value is not entered.
Chapter: 8. Error messages
18