Artis zee
AX
Replacements of Parts
Real Time Controller
Valid for:
- FD-RTC Core Duo: Part No. 10280904 modified for Sys-
tems beginning with System SW VC13
- FD RTC V4: Part No. 10281168 beginning with System
SW VC13 © Siemens 2008
The reproduction, transmission or use
of this document or its contents is not
10281168
10280904
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Product-specific Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Use and Purpose of this Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Additionally Required Documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Required Measuring Equipment, Tools and Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
SIRC, Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Replacing the Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
When performing the work steps and checks, observe the general safety information for
medical products. (TD00-000.860.01 / General Safety Notes)
After completing repair work and before turning the system over to the customer, the
ground wire resistance of the unit has to be tested.
CAUTION [hm_Serv_Training]
M WARNING AXIOM Artis dFCM and dBCM Systems with Magnetic Navi-
gation (NIOBE):
The two permanent magnets are installed by the Stereotaxis
Company. Magnetic objects can be attracted by the two per-
manent magnets and can fly through the air like shots.
¹ The safety information in the technical documents
from the Stereotaxis Company absolutely must be
observed.
WARNING [hm_serv_WarnSetUPSPowerMayBePresent]
WARNING [hm_serv_LabelPowerPresent]
Even after switching off the system (Shutdown), power is still
present at the control cabinet and generator.
If not observed, death or serious bodily injury can occur.
¹ Observe the warning signs in the cabinets.
¹ Switch off the power supply to the X-ray system with
the on-site on/off switch.
WARNING [hm_serv_CheckBypassImage]
Bypass image is not displayed on exam room display.
This can cause to an Interruption of interventional proce-
dure.
¹ Perform check of the bypass signal after service proce-
dure. With transition to bypass the analog video signal
is fed as second source to the live display in the exam
room. The display must switch in less than 2 seconds
to the analog signal (prio).
WARNING [hm_serv_FixCablesAfterService]
Unwanted disconnection of electrical cables.
This can cause to an interruption of interventional proce-
dure.
¹ Check fixation of connectors and cables after service
procedures.
General Information 0
Screenshots and graphics used in this document should be viewed as examples only. They
cannot describe all possible combinations; they are informational in nature and are used
only for better orientation.
Prerequisites 1.1
Procedure 0
Repair and replacement of parts may be performed only by trained service technicians.
All tools, test equipment and aids, with the exception of the "standard service and installa-
tion tools" and the keyboard, are listed and specified in the Service Tool Catalogue from
SIEMENS MED SPC (Spare Parts Catalogue on CD-ROM or on the Intranet).
• Multimeter
• USB keyboard
• Service or installation tool kit n.a.
• Antistatic Kit (TKD Static Dissipative Field Service Kit) Part No.: 97 02 606
• Fiber optic repair kit Part No.: 8397569
• Fiber optic testing device (for testing fiber optic cables) Part No.: 8899200
Board Hardware boards in the RTC PCI slots (from left to right)
SIRC System Interface & Radiation Control PCI board with:
• ECG Interface Component (EIC) (option)
• X40 ribbon cable to the X4 FDR board
• X41 connector (2-pole cable; RTC Reset)
• X50, 51, 52 connectors (connection cable between the SIRC PCI
board and the X60, X61, X62 connector front panel)
• X42, 43 connectors; fan connections 1 and 2
• X44, 45 connectors; temperature sensors 1 and 2
• X47 connector; encoder signals
• X48 connector; extension signals (not used)
FDR Flat panel Detector Receiver board
• X2, X3 connectors; ribbon cable
• X703 coaxial connector (to the bypass monitor); coaxial cable
• U600 connector (Gigalink to the IAS); fiberglass cable
• X6, X7 connectors (RS232); plastic fiber optic cable
• X9, X10 connectors (frame request, X-ray enable); plastic fiber
optic cables
• X11 connector (Reset); plastic fiber optic cable
• X8 connector (Gigalink from the FD); fiberglass cable
COPRA1 COmmon PRocessing Architecture (D62)
• X2, X3 connectors; ribbon cable
COPRA2 COmmon PRocessing Architecture
• X2, X3 connectors; ribbon cable
COPRA3 COmmon PRocessing Architecture
• X2, X3 connectors; ribbon cable
There is a 50-pole ribbon cable in the accessory pack of a new COPRA board.
This is a description of which HW is included in the shipment for the particular RTC replace-
ment part.
Requirements 0
Safety Information
• See (Notes / p. 5)
Used for
• Artis zee systems with System SW ≥ VC13
Replacement
• n.a.
Procedure 0
9. Click on Go in the bottom command bar. The backup routine will start and a window
with a scrolling backup log ("Start Backup....”) will open.
Pack the defective RTC in the original packaging in exactly the same way as the new RTC
was shipped.
Make sure there is exactly the same seating in the foam plastic parts, otherwise it is pos-
sible that the RTC will be damaged during transport because of its great weight. In this hap-
pens, there will be no credit issued for the returned part!
Fig. 7: Configuring the MAC address for the LAN bootable components, "old" MAC address
6. Click on Reset for the particular RTC, A or B.
7. Click on OK.
Fig. 8: Configuring the MAC address for the LAN bootable components
8. Click on Configure for the particular RTC, A or RTC B.
b) Connect the USB keyboard to the front of the RTC. Alternatively, the PS/2 key-
board can also be connected to the violet connector on the top of the RTC.
c) Click on Configure for the particular RTC (A/B) in the "Configure MAC Address..."
window.
d) When the "Switch RTC off/on..." window appears, switch the RTC off and back
on. The following step must be performed immediately following switch-on of
the RTC.
e) Press the space bar on the keyboard for approx. 5 sec.
f) Then remove the keyboard.
12. Instead of the old "MAC address" or "NOT_DEFINED", the MAC address for the
"new" RTC is displayed.
16. If the message “... correction could be made .. .” appears, confirm this with
OK.
20. Highlight the checkbox for the corresponding "Download RTC A or B").
24. If applicable, click on Go and wait until the following window appears.
Fig. 19: Artis RTC Software Manager (Download Mode): SIRC/ECG Download
27. If the nominal and the actual firmware version are the same, exit the window with Fin-
ish. Continue with Point 33.
If not, select all checkboxes, in which the nominal and actual firmware versions are
not the same. Continue with Point 28.
28. Click on Go to start the download.
¹ There must be an ok in the “Status” column for all firmware components following
the download.
Fig. 20: Message box: Switch the RTC Off/On after the SIRC/ECG download
31. Switch the RTC(s) off and back on.
32. Wait until the 7-segment display on RTC A “2A” or on RTC B "2b appears.
33. Click on OK to close the window.
Fig. 21: Artis RTC SW Manager (Download Mode), after the SIRC/ECG download
34. Click on Exit to close the “Artis Software Manager (Download Mode)” window.
Requirements 0
Safety Information
• See (Notes / p. 5)
Used with
• FD RTC Core Duo modified for System SW VC13, Part No. 10280904.
This RTC was upgraded during the SW update from VC12 to VC13 with the new V3
SIRC PCI board (Part No. 10096748) and with the new ECG interface component (Part
No. 10096666).
• V4 FD RTC, Part No. 10281168
Replacement:
• n.a.
Procedure 0
NOTE When plugging the ribbon cables into the X2 and X3 front
connectors on the COPRA or FDR2 board, extreme care must
be used to prevent the pins from getting damaged.
Do not plug in the cables using force or at an angle!
The SIRC is comprised of the SIRC-Connector-Panel and the SIRC-PCI Platine that are
connected to each other by three ribbon cables (X60/61/62).
The SIRC-Connector-Panel is the central connection point for the system data and signal
cables.
Two 7-segment displays, V45/44, display the RTC boot routine. The status of voltages and
signals can be read from the status LEDs. Test points on the X73 measuring strip are used
to check important input and output signals and voltages. The S1 switch is used to enable
radiation using the software for service purposes. In the normal mode, it is set to OFF.
Requirements 0
Used with
• FD RTC Core Duo modified for System SW VC13, Part No. 10280904.
This RTC was upgraded during the SW update from VC12 to VC13 with the new V3
SIRC PCI board (Part No. 10096748) and with the new ECG interface component (Part
No. 10096666).
• V4 FD RTC, Part No. 10281168
Replacement
• n.a.
Procedure 0
E 5. The old SIRC-Connector-Panel ausbauen. Dazu muß die auf der linken Seite befindli-
che Rändelschraube gelöst werden.
NOTE When plugging the 3 ribbon cables (X60, X61, X62) into the
connectors on the SIRC PCI board, proceed extremely cau-
tiously so that the pins do not get damaged.
Do not plug in the cables using force or at an angle!
10. Check whether all connectors are correctly engaged and, if applicable, threaded in
place.
E 11. Switch on the system and let it boot.
12. Check of the Bypass Mode:
a) Click on Options > Service > Local Service to open the ’Service Software’ win-
dow.
Requirements 0
Used with:
• FD RTC Core Duo modified for System SW VC13, Part No. 10280904.
This RTC was upgraded during the SW update from VC12 to VC13 with the new V3
SIRC PCI board (Part No. 10096748) and, if applicable, with the new ECG interface
component (Part No. 10096666).
• V4 FD RTC, Part No. 10281168
Replacement
• n.a.
Procedure 0
E 8. Carefully remove the SIRC-PCI Platine from the PCI slot and place it down on an ESD
mat.
Fig. 31: Artis RTC Software Manager (Download Mode): SIRC/ECG Download
30. If the nominal and the actual firmware version are the same, exit the window with Fin-
ish.
Continue with Point 35.
If not, select all checkboxes, in which the nominal and actual firmware versions are
not the same.
Continue with Point 30.
31. Click on Go to start the download.
¹ There must be an ok in the “Status” column for all firmware components.
32. Click on Finish.
33. Wait until the following window appears.
Fig. 32: Message box: Switch the RTC Off/On after the SIRC/ECG download
34. Switch the RTC(s) off and back on.
35. Click on OK to close the window.
36. Wait until the 7-segment display on RTC A shows “2A” and “2b” on RTC B and the fol-
lowing window appears:
Fig. 33: Artis RTC SW Manager (Download Mode), after the SIRC/ECG download
37. Click on Exit to close the “Artis Software Manager (Download Mode)” window.
38. Click on Next or on Config Check.
¹ The Config check starts. No errors may occur.
39. Click on Exit to close the Config Check window.
40. Click on Next or on Version Check.
¹ The Version check starts.
If SW conflicts occur, click on Refresh.
To install or remove the EIC board, the SIRC PCI board must first be removed. The ECG
Interface Component is comprised of:
• EIC board (D21), which is plugged in on the SIRC PCI board.
• Slot panel wit 2x 15-pole Sub-D connectors and connection cable. This cable is
plugged in on the D21.X4 EIC board.
Requirements 0
Safety Information
• See (Notes / p. 5)
Estimated Completion Time and Number of Persons
• 0.5 hrs. / 1 person
Used for:
• FD RTC Core Duo modified for System SW VC13, Part No. 10280904.
This RTC was upgraded during the SW update from VC12 to VC13 with the new V3
SIRC PCI board (Part No. 10096748) and, if applicable, with the new ECG interface
component (Part No. 10096666).
• V4 FD RTC, Part No. 10281168
Replacement
• n.a.
Jumpers, bridges, switches, buttons
• S2, S3 buttons
• D21: S1 DIPP switches:
Switch positions: S1.1 and S1.2 always on "OFF".
Procedure 0
board can be removed. After disengaging the mechanism, also remove the board
securing strip prior to pulling out a board. To do this, (16/Fig. 4 / p. 12)remove the two
screws and pull out the mount towards the front.
E 9. Carefully remove the SIRC-PCI Platine from the PCI slot and place it down on an ESD
mat.
E 10. Remove the four mounting screws (1/Fig. 27 / p. 37) from the ECG board (D21) and
carefully pull the board up and out of the two connector strips by working it alternately
on both sides. Do not pull up on it sharply and do not bend it.
E 11. Plug in the new ECG board (D21) on the SIRC PCI board and secure it in place. When
doing this, pay attention to correct positioning of the new D21 board when plugging it
into the connector strips!
12. D21 board: Check the S1 DIPP switch positions:
Set all switches to “OFF”.
E 13. Install the SIRC-PCI Platine (with the new D21 ECG board) in the same PCI slot. When
doing this, plug the three ribbon cables back in on the rear connector panel.
14. Reinstall the slot panel with the 2 Sub-D connectors. Plug in the ribbon cables on
D21.X4.
15. Reconnect all cables on the SIRC PCI board.
16. Reinstall the mechanical board securing mechanism.
17. Manually move the board securing mechanism until the boards are engaged in place.
Then retighten the wing nut.
18. Connect all previously removed cables on the top of the RTC and secure them.
19. Check all boards for secure seating and good contact.
E 20. .
21. Check whether all connectors are correctly engaged and, if applicable, threaded in
place.
22. Swing back the RTC and close it.
23. Switch on the system and let it boot.
24. Click on Options > Service > Local Service to open the ’Service Software’ window.
25. Enter the ’Password’ and confirm with OK.
26. Click on Configuration in the “Service SW” window.
27. Click on SWM RTC for “Preparing download ...)
28. Click on “>” for “RTC download without disk format”
29. Click on Go for "Download Mode".
30. Select the SIRC/ECG Download checkbox in the 'Artis Software Manager RTC' win-
dow.
Fig. 34: Artis RTC Software Manager (Download Mode): SIRC/ECG Download
32. If the nominal and the actual firmware version are the same, exit the window with Fin-
ish.
Continue with Point 37.
If not, select all checkboxes, in which the nominal and actual firmware versions are
not the same.
Continue with Point 32.
33. Click on Go to start the download.
¹ There must be an ok in the “Status” column for all firmware components.
34. Click on Finish.
35. Wait until the following window appears.
Fig. 35: Message box: Switch the RTC Off/On after the SIRC/ECG download
36. Switch the RTC(s) off and back on.
37. Click on OK to close the window.
38. Wait until the 7-segment display on RTC A shows “2A” and “2b” on RTC B and the fol-
lowing window appears:
Fig. 36: Artis RTC SW Manager (Download Mode), after the SIRC/ECG download
39. Click on Exit to close the “Artis Software Manager (Download Mode)” window.
40. Click on Next or on Config Check.
¹ The Config check starts. No errors may occur.
41. Click on Exit to close the Config Check window.
42. Click on Next or on Version Check.
¹ The version check starts.
If SW conflicts occur, click on Refresh.
43. Click on Exit to close the “Version check result” window.
44. Click on Finish to close the “Configuration” window.
45. Click on Home.
46. No adjustments are required.
47. Check of the Bypass Mode:
a) Click on Configuration in the “Service SW” window.
b) Click on System Config
X c) Release fluoroscopy
¹ The bypass fluoroscopy image must appear on the Live TFT monitor in the
examination room in less than 2 sec.
d) Click on Exit >Finish> Home.
48. Perform a function test.
X Trigger a fluoro and an ECG-triggered acquisition at the highest selectable frame
rate and matrix size. Check all radiation displays.
With biplane systems, perform in Planes A and B.
A
9Index
G General Information . . . . . . . . . . . . . . . .7
General Safety Information. . . . . . . . . . .5
H hm_Serv_CheckBypassImage . . . . . . . .6
hm_Serv_FixCablesAfterService . . . . . .6
hm_serv_LabelPowerPresent. . . . . . . . .6
hm_serv_NoteProtectEyes. . . . . . . . . . .6
hm_Serv_Training . . . . . . . . . . . . . . . . .5
hm_serv_WarnSetUPSPowerMayBePresent5
N Notes . . . . . . . . . . . . . . . . . . . . . . . . . . .5