Understanding Infrared Spectroscopy
Understanding Infrared Spectroscopy
1
Infrared Absorption Spectroscopy (IR)
2
Common Spectroscopic Methods based on Electromagnetic Radiation
Type Spectroscopy Usual wavelength range Usual wave number range, cm-1 Type of quantum
transition
Gamma-ray emission 0.005-1.4 Å - Nuclear
X-ray absorption, emission, 0.1-100 Å - Inner electron
fluorescence,and diffraction
Vacuum uv absorption 10-180 nm 1x106 to 5x104 Bonding electrons
uv-vis absorption, emission, 180-780 nm 5x104 to 1.3x104 Bonding electrons
and fluorescence
IR absorption and Raman 0.78-300 m 1.3x104 to 3.3x101 Rotation/
scattering vibration of
molecules
Microwave absorption 0.75-3.75 mm 13-27 Rotation of molecules
Electron spin resonance 3 cm 0.33 Spin of electrons in a
magnetic field
Nuclear magnetic resonance 0.6-10 m 1.7x10-2 to 1x103 Spin of nuclei in a
magnetic field
1 Å = 10-10 m = 10-8 cm
1nm = 10-9 m = 10-7 cm
1mm = 10-6 m = 10 –4 cm
3
Infrared Spectral Regions
The infrared spectrum is conveniently divided into;
❖ Near
❖ Mid
❖ Far
radiation
Region Wavelength Range Wavenumber Range Frequency Range
(µm) (cm-1) (Hz)
Near 0.78 to 2.5 12,800 to 4000 3.8x1014 to 1.2x1014
Middle 2.5 to 50 4000 to 200 1.2x1014 to 6.0x1012
Far 50 to 1000 200 to 10 6.0x1012 to 3.0x1011
Most Used 2.5 to 15 4000 to 670 1.2x1014 to 2x1013
4
Representation of an Electromagnetic Wave
Electric Field
Direction of Propagation
Frequency :
the number of oscillations of the
field that occur per second and is
equal to 1/p (p: period)
Wavelength :
linear distance between any two
equivalent points on successive
waves (for example successive
maxima or minima)
6
ELECTROMAGNETIC RADIATION
Velocity of Propagation (vi):
Multiplification of the frequency () in cycles per second by the
wavelength () in meters per cycle gives the velocity of propagation
vi in meters per second:
vi = i
c =
The relation between frequency, , of the light and energy, E, is defined as;
E=h
h is the Planck constant. 7
ELECTROMAGNETIC RADIATION
Whenever a molecule interacts with radiation, a quantum of energy
(or photon) is either emitted or absorbed. In each case, the energy of
the quantum of radiation must exactly fit the energy gap such as E1-
E0.
E3
E2
E1
Eo 8
ELECTROMAGNETIC RADIATION
The wavenumber :
Reciprocal of the wavelength in cm, and the unit of wavenumber is cm-1.
Wavenumber is widely used in IR spectroscopy. The wavenumber is a useful unit
because, in contrast to wavelength, it is directly proportional to the frequency.
= 1/=/c
9
Infrared absorption spectrum of dichloromethane
Wavenumber (cm-1)
15
Types of Molecular Transitions
There are three types of molecular transitions that occur in IR:
a) Rotational transitions
• When an asymmetric molecule rotates about its center of mass, the
dipole moment seems to fluctuate.
• E for these transitions correspond to < 100 cm-1
b) Vibrational-rotational transitions
c) Vibrational transitions
• The most important transitions observed in qualitative mid-IR
spectroscopy.
= 13,000 – 675 cm-1 16
Types of Molecular Vibrations
1. Stretching - A stretching vibration involves a continuous change in the
interatomic distance along the axis of the bond between two
atoms.
17
The atoms in a CH2X2 group, commonly found in organic compounds and where
X can represent any other atom. Six of these vibrations involve only
the CH2 portion:
Symmetry
Symmetric Antisymmetric
Direction
Radial Streching
Antisymmetric
Symmetric stretching (νs)
stretching (νas)
Latitudinal Bending
Scissoring (δ) Rocking (ρ)
19
Mechanical Model of Stretching Vibrations
Simple harmonic oscillator:
where: F = Force
k = Force Constant
(depends on stiffness
of spring)
y = Displacement
16𝑥10−3 𝑘𝑔/𝑚𝑜𝑙
𝑀𝑦 = 23 𝑥1 𝑎𝑡𝑜𝑚 = 2.7𝑥10−26 𝑘𝑔
6.0𝑥10 𝑎𝑡𝑜𝑚𝑠/𝑚𝑜𝑙
1𝑥103 𝑁/𝑚
𝜈ҧ = 5.3𝑥10−12 𝑠/𝑐𝑚 = 1.6𝑥103 𝑐𝑚 −1
1.1𝑥10−26 𝑘𝑔 22
Anharmonic Oscillators
In reality, bonds act as anharmonic oscillators because as atoms get close, they
repel one another, and at some point, a stretched bond will break, thus, the
potential energy can be expected to rise more rapidly than the harmonic
approximation preditcs.
26
Fingerprint Region (1200-700 cm-1):
Small differences in
the structure and
constitution of a
molecule result in
significant changes in
the distribution of
absorption peaks in
this region of the
spectrum.
As a consequence a
close match between
two spectra in this
fingerprint region
constitutes strong
evidence for the
identity of the
compounds.
FTIR Spectra of Toluene
The output from the instrument is referred as a spectrum. Most commercial
instruments present a spectrum with the wavenumber decreasing from left to
right.
Characteristic Infrared Absorption of groups
Group Frequency Range (cm-1) Intensity* (m-s)
A. Alkyl
C-H stretching 2853-2962 (s)
Isopropyl, - CH(CH3)2 1380-1385 and 1365-1370 (s)
Tert-Butyl, - C(CH3)3 1385-1395 (m)
~1365 (s)
B.Alkenyl (out of plane C-H bending)
C-H (stretching) 3010-3095 (m)
C=C (stretching) 1620-1680 (v)
R-CH=CH2 985-1000 (s)
R2-C=CH2 880-900 (s)
cis-RCH=CHR 675-730 (s)
trans- RCH=CHR 960-975 (s)
C. Alkynyl
≡CH (stretching) ~3300 (s)
C≡C (stretching) 2100-2260 (v)
D. Aromatic
Ar-H (stretching) ~3030 (v)
Aromatic substitution type
(C-H out-of-plane bendings)
monosubstituted 690-710 (very s)
730-770 (very s)
730-770
o-Disubstituted 735-770 (s)
m-Disubstituted 680-725 (s)
750-810 (very s)
p-Disubstituted 800-860 (very s)
E. Alcohols, Phenols, and Carboxylic Acids
O-H (stretching)
Alcohols, phenols (dilute solutions) 3590-3650 (sharp, v)
Alcohols, phenols (hydrogen bonded) 3200-3550 (broad, s)
Carboxylic acids (hydrogen bonded) 2500-3000 (broad, v)
F. Aldehydes, Ketones, Esters, Carboxylic Acids, and Amides
C=O (stretching) 1630-1780 (s)
Aldehydes 1690-1740 (s)
Ketones 1680-1750 (s)
Esters 1735-1750 (s)
Carboxylic Acids 1710-1780 (s)
Amides 1630-1690 (s)
29
*Abbreviations: s = strong, m= medium, w = weak, v = variable, ~ = approximately.
Vibrational Coupling
The energy of a vibration and thus the wavelength of its absorption peak may be
influenced (or coupled) by other vibrations in the molecule.
As an example of coupling effects, let us consider the infrared spectrum of
carbon dioxide:
If no coupling occured between two C=O bonds, an absorption peak would be
expected at the same wavenumber as the peak for the C=O stretching vibration in an
aliphatic ketone (about 1700 cm-1). However, experimentally, carbon dioxide exhibits
two absorption peaks, the one at 2330 cm-1 and the other 667 cm-1.
in plane bending
30
Infrared Instruments
Four types of instruments for infrared absorption measurements are
available for commercial sources:
32
Fourier Transform Infrared (FT-IR)
Spectrometers
33
Fourier Transform Infrared Spectrometers
❖ This type of instrument employs an interferometer. FTIR has greatly
improved the quality of infrared spectra and minimized the time required
to obtain data.
❖ The majority of commercially avialable FTIR instruments are based upon the
Michelson interferometer.
36
The radiation striking the detector after passing through the interferometer will generally
be much lower in frequency than the source frequency. The relationship between the two
frequencies is readily derived by reference to P(t) versus plot. One cycle of signal occurs
when the mirror moves a distance that corresponds to one half a wavelength (/2). If the
mirror is moving at a constant velocity of vM and is the time required for the mirror to
move /2 cm, it can be expressed as below;
𝜆
𝑣𝑀 𝜏 =
2
1 2𝑣𝑀
𝑓= =
𝜏 𝜆
37
The relationship between the optical frequency of the radiation and
the frequency of the interferogram
• The frequency of the
radiation () is directly
related to the frequency
of the interferogram (f).
2 m
f =
c
= frequency of radiation
f = frequency of inteferogram
m = velocity of the mirror
c = speed of light (3.00 x 1010 cm/s)
2. Globar - Silicon carbide rod with water cooled contacts to prevent arcing.
Others are; mercury arc, tungsten filament lamp, carbon dioxide laser source
40
Detectors – measure minute changes in temperature.
1. Thermal detectors
• Have a slow response time, so they are not well suited to FT-IR.
2. Pyroelectric detectors
• Constructed of crystalline wafers of pyroelectric materials (triglycine
sulfate) which are insulators (dielectric materials) with very special
thermal and electrical properties.
• Have a fast response time and are well suited for FT-IR.
3. Photoconducting detectors
• Constructed of a semiconducting material (lead sulfide,
mercury/cadmium telluride, or indium antimonide) deposited on a
nonconducting glass surface and sealed in an evacuated envelope to
protect the semiconducting material from the environment.
• Fast response time comparing to pyroelectric detectors and are well suited
for FT-IR. 41
Advantages of FTIR
42
Infrared Absorption Spectroscopy (IR)
❖ Liquids ❖ Powders
❖ Solutions ❖ Films
❖ Pastes ❖ Fibres
❖ Gases
43
Sample Handling Techniques
Gas Samples:
The spectrum of gas sample can be obtained by
permitting the sample to expand into an evacuated cell.
Solutions:
45
Sample Handling Techniques
Solid:
46
Sample Handling Techniques
Solid:
Mull Method: A finely powdered
sample (2 to 5 mg, <2 µm) is
combined with Nujol (a brand of
mineral oil by Plough Inc) in a mortar
and pestle to make a mull (a very
thick suspension). This mull is
sandwiched between sodium chloride
plates before being placed in the
spectrometer. The disk is placed in
the light path of IR radiation and its IR
spectrum recorded.