SPM 9700HT Hardware
SPM 9700HT Hardware
Oct. 2017
SPM-9700HT
Instruction Manual
Hardware
Read the
Read the instruction
instruction manual
manual thoroughly
thoroughly before
before you
you use
use the
the product.
Keep this
product. instruction manual for future reference.
Keep this instruction manual for future reference.
This page is intentionally left blank.
Introduction
Introduction
Read this Instruction Manual thoroughly before using the product.
IMPORTANT
If the user or usage location changes, ensure that this Instruction Manual is always kept together
with the product.
If this manual or a product warning label is lost or damaged, immediately contact your Shimadzu
representative to request a replacement.
To ensure safe operation, read all Safety Instructions before using the product.
To ensure safe operation, contact your Shimadzu representative if product installation, adjustment,
or re-installation (after the product is moved) is required.
Notice
All rights are reserved, including those to reproduce this manual or parts thereof in any form
without permission in writing from Shimadzu Corporation.
Information in this manual is subject to change without notice and does not represent a
commitment on the part of the vendor.
Any errors or omissions which may have occurred in this manual despite the utmost care taken in
its production will be corrected as soon as possible, although not necessarily immediately after
detection.
Shimadzu takes no responsibility for the results obtained through the use of this instrument.
Unforeseen accidents may cause data loss on the computer's hard disk drives. Always regularly
backup important data to avoid data loss due to unforeseen accidents.
Microsoft and Windows are registered trademarks of Microsoft Corporation in the United States
and/or other countries. Other company names and product names mentioned in this manual are
trademarks or registered trademarks of their respective companies.
The TM and ® symbols are omitted in this manual.
© 2017 Shimadzu Corporation. All rights reserved.
Original version is approved in English.
Safety Instructions
To ensure safe product operation, read these important safety instructions carefully before use and
follow all WARNING and CAUTION instructions given in this section.
Installation Site
! CAUTION
Do NOT install this equipment in locations subject to direct sunlight.
Doing so may damage the equipment.
Installation
To ensure safe operation, contact your Shimadzu representative if product installation, adjustment, or
re-installation (after the product is moved) is required.
! WARNING
Take steps to prevent the instrument from falling over due to earthquakes or other
causes.
Not doing so may cause an injury.
Do NOT insert the adapter's ground wire into, or allow it to come in contact with, the
power outlet.
Doing so may result in fire or electric shock.
Do NOT place heavy objects on top of power cables. Also, do NOT run power cables
close to heat-generating equipment.
Doing so may result in damage to the cables and cause a fire, electric shock, or
breakdown. If cables are damaged, contact your Shimadzu representative immediately.
Do NOT modify power cables, or bend or pull power cables with excessive force.
Doing so may damage the cables and cause a fire or electric shock. If cables are
damaged, contact your Shimadzu representative immediately.
Operation
! WARNING
NEVER look directly into the laser beam.
The SPM head uses a Class 2 semiconductor laser with a maximum output of 0.6 mW
that is capable of damaging eyesight.
If a sample with a mirror-like surface or a pair of tweezers is held close to the path of the
laser beam inside the SPM head, high-intensity scattered laser light may be emitted from
the aperture due to reflection. In this case, ensure that high-intensity scattered laser light
does not enter your eyes.
Do NOT raise the SPM head from the SPM main unit while the laser is ON.
Otherwise, you may inadvertently look directly into laser light emitted from the bottom of
the SPM head. Turn power to the instrument OFF before raising the SPM head from the
SPM main unit.
Apert
ure
Do NOT touch the upper part of scanner, when the voltage is applied to a sample. There
is risk of electric shock.
Do NOT install an article in the gap of the upper scanner. High voltage is applied to a
scanner. Doing so may result in it and cause electric shock, or breakdown.
Do NOT perform any control or adjustment procedures other than those stipulated in the
instruction manual.
Doing so may result in dangerous laser exposure.
Do NOT use flammable spray (e.g., hairspray or insecticide) close to the instrument. If it
is used, the spray may ignite and cause a fire.
! CAUTION
Do NOT insert your fingers or foreign objects into the instrument's ventilation slots or in
spaces between the units.
Doing so may cause a breakdown or injury.
Do NOT remove the covers from the SPM unit or the control unit.
Doing so may cause a breakdown or injury. The unit covers do not need to be removed
for everyday inspections and maintenance. Contact your Shimadzu representative for
repairs that require the removal of unit covers.
Do NOT use organic solvents, acids, or alkaline solutions near this instrument.
The instrument may be damaged if it comes into contact with any of these liquids.
Do NOT insert the tweezers to a clearance gap between a guide and a presser bar.
Doing so may cause a contaminating the standard sample.
Do NOT insert the tweezers to a clearance gap between a guide and a presser bar.
It may damage a piezoelectric device or wire inside the cantilever tune.
Store the standard samples in their specialized case when not using them.
If dust has adhered to the pins or pin-side surface of any power plugs, disconnect the
affected plug and wipe it clean with a dry cloth.
Failure to keep power plugs clean may cause fire to break out.
Only replace parts with those specified in "6.6.1 Maintenance Parts List".
The use of unspecified parts may cause component damage and prevent normal
operation.
In an Emergency
! WARNING
If the instrument emits a strange noise or questionable odor, immediately turn OFF the
power switch, disconnect the power plug from the power outlet, and contact your
Shimadzu service representative.
If the instrument does not start up after turning ON the power switch when recovering
from a power failure, contact your Shimadzu service representative.
Warning Labels
Warning labels are attached at the necessary locations on the instrument to ensure safe usage.
If any warning labels are missing or lost, immediately order replacements from your Shimadzu
representative and attached them at the correct locations. Information for each warning label and part
numbers (P/N) corresponding to each label are shown below.
P/N 037-72421-01
P/N 037-72601-05
SPM Unit
P/N 037-72601-05
P/N 037-72604-03
Identification Label
P/N 228-55336-01
SHIMADZU CORPORATION
1, Nishinokyo-Kuwabaracho,
Nakagyo-ku, Kyoto 604-8511, Japan
Warranty
Shimadzu provides the following warranty for this product.
Failures caused by the following are excluded from the warranty, even if
4. Exceptions: they occur during the warranty period.
1. Improper product handling
2. Repairs or modifications performed by parties other than Shimadzu
or Shimadzu designated companies
3. Product use in combination with hardware or software other than that
designated by Shimadzu
4. Computer viruses leading to device failures and damage to data and
software, including the product's basic software
5. Power failures, including power outages and sudden voltage drops,
leading to device failures and damage to data and software, including
the product's basic software
6. Turning OFF the product without following the proper shutdown
procedure leading to device failures and damage to data and
software, including the product's basic software
7. Reasons unrelated to the product itself
8. Product use in harsh environments, such as those subject to high
temperatures or humidity levels, corrosive gases, or strong vibrations
9. Fires, earthquakes, or any other act of nature, contamination by
radioactive or hazardous substances, or any other force majeure
event, including wars, riots, and crimes
10. Product movement or transportation after installation
11. Consumable items
Note: Recording media such as CD-ROMs are considered
consumable items.
After-Sales Service For information on periodic inspection and maintenance of this product,
refer to "Chapter 6 Maintenance".
If any problem occurs with this product, perform an inspection and take
appropriate corrective action as described in this manual's
troubleshooting section.
If the problem persists, or the symptoms are not covered in the
troubleshooting section, contact your Shimadzu representative.
Replacement Parts Replacement parts for this product will be available for a period of
seven (7) years after the product is discontinued. Thereafter, such
Availability parts may cease to be available.
Note, however, that the availability of parts not manufactured by
Shimadzu shall be determined by the relevant manufacturers.
For daily maintenance, inspection, and replacement parts, see "5. Maintenance".
Replacement cycles described for periodic replacement parts are rough estimate.
Replacement may be required earlier than the described replacement cycles depending on usage
environment and frequency.
Disposal Precautions
This product uses beryllium oxide for one of the electronic parts, and you cannot dispose of the
product as it is.
When disposing the product, be sure to contact your Shimadzu representative.
Contents
Introduction ..................................................................................... iii
About the Instruction Manual ................................................................................... iv
Safety Instructions .................................................................................................... v
Warning Labels ........................................................................................................ ix
Identification and FDA Certification Labels ............................................................... x
Warranty .................................................................................................................. xi
After-Sales Service and Availability of Replacement Parts ..................................... xii
Inspection and Maintenance ................................................................................... xii
Disposal Precautions .............................................................................................. xii
Action for Environment (WEEE) ..............................................................................xiii
1. Overview ................................................................................1
2. Observation Mode ..................................................................2
2.1 Contact Mode ................................................................................................... 2
2.2 Dynamic Mode ................................................................................................. 2
2.3 Phase Mode ..................................................................................................... 3
2.4 Lateral Force (LFM) Mode ................................................................................ 4
2.5 Force Modulation Mode .................................................................................... 5
2.6 Magnetic Force (MFM) Mode ........................................................................... 6
2.7 Current Mode ................................................................................................... 7
2.8 Surface Potential (KFM) Mode ......................................................................... 7
3. Configuration ........................................................................ 10
3.1 Appearance .................................................................................................... 10
3.2 System Configuration ..................................................................................... 10
3.2.1 SPM Unit ..................................................................................................................10
3.2.2 Control Unit ..............................................................................................................11
3.2.3 Data Processing System ..........................................................................................11
3.3 Details of Parts ............................................................................................... 11
4. Details of Instrument Units .................................................... 14
4.1 SPM Unit ........................................................................................................ 14
4.2 SPM Head ...................................................................................................... 14
4.3 Control Unit .................................................................................................... 16
4.4 Data Processing System ................................................................................ 17
4.5 Optical Microscope Unit.................................................................................. 17
4.5.1 Preparation ..............................................................................................................18
4.5.2 Optical Microscope Image Display ............................................................................18
4.5.3 Adjusting Focus and Lighting ....................................................................................19
4.5.4 Magnification ............................................................................................................19
9. Troubleshooting .................................................................... 44
10. Appendix .............................................................................. 47
10.1 Operating Principle of Surface Potential (KFM) Mode .................................... 47
1. Overview
Scanning probe microscope (SPM) is the general term used to describe a microscope that allows the
high-magnification observation of 3D topographic (height) images, force modulation images,
magnetic force images, friction force images, electric-current images, and electric-potential images
by scanning a sample surface with a microscopic probe.
An atomic force microscope (AFM) is a device that allows the observation of a sample's surface
topography by holding an extremely small cantilever close to the sample surface and detecting the
force (atomic force) acting between the cantilever and the sample.
In addition to height images, switching between observation modes allows the observation of force
modulation images, magnetic force images, friction force images, current images, and potential
images.
The SPM-9700HT is a microscope and not a measuring device. Data values acquired using the
SPM-9700HT should only be used as a guide. Shimadzu does not guarantee the accuracy of such values.
2. Observation Mode
2.1 Contact Mode
In this mode, the repulsive force that acts between the cantilever and the sample is detected and
feedback* control of the scanner is performed along the Z-axis so that the vertical deflection of the
cantilever remains constant while the surface of the sample is scanned.
Contact mode is used to observe height images, which represent the sample's surface topography,
and deviation images, which indicate the difference from the target feedback value.
4-element photodiode
Cantilever
Signal
processing X/Y scan
controller
Monitor
Tube piezo
scanner
Feedback PC
controller
* The SPM detects the force acting between the cantilever and the sample as a displacement or
variation in amplitude of the cantilever. Feedback in the SPM is defined as using the scanner to
control the distance between the sample and the cantilever so that the detected displacement or
amplitude of the cantilever equals the operating point, which is the target feedback value.
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
カンチレバーの位置信号
Cantilever position signal
加振信号 signal
Excitation
Asinδ
ロックイン
Lock-in
検出器
detector Reference
参照信号
signal
振幅(A) (A)
Amplitude
位相(δ)
Phase ()
感度
Sensitivity 位相オフセット
Phase offset
x1, x2, x4, 8
1, 2, 4, x8 -100~100度,
-100° to 100°,
ロックイン
Lock-in +180度
+180°
検出器
detector
Acosδ
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
A signal output from the lock-in detector.
AsinD This signal and the Acos signal are used to calculate the phase () and
amplitude (A).
A signal output from the lock-in detector.
AcosD This signal and the Asin signal are used to calculate the phase () and amplitude
(A).
This signal contains topographical information and differences in properties
Phase
across the sample surface, such as viscoelasticity.
As for deflection, this signal indicates the difference from the target feedback
Amplitude
value.
走査角度:0deg
Angle: 0 deg 走査角度:90deg
Angle: 90 deg
Clockwise
時計回りに回転rotation
Cantilever
カンチレバー
試料
Sample
トレース
Trace
リトレース
Retrace
Trace
トレース
Retrace
リトレース
Fig. 2.4.1 Cantilever Scan Direction as Displayed on Screen
検出器
Detector
レーザー
Laser
Cantilever
カンチレバー
トレース
Trace
リトレース
Retrace トレース リトレース
Trace Retrace
(A)
Fig. 2.4.2 How the Deflection (Horizontal) Signal Is Output
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
The cantilever torsion signal.
Def.(Hori.) Indicates the lateral force (frictional force) when scanning vertically with respect to
the longitudinal direction of the cantilever.
P
D
Cantilever
Asin Lock-in
A,
Acos Amplifier
Sample
Oscillator
Tube scanner
Large
Small
カンチレバーの位置信号
Cantilever position signal
加振信号 signal
Excitation
Lock-in
Asinδ
ロックイン
検出器
detector Reference
参照信号
signal
振幅(A)
Amplitude (A)
位相(δ)()
Phase
感度
Sensitivity 位相オフセット
Phase offset
x1, x2, x4, 8
1, 2, 4, x8 -100~100度,
-100° to 100°,
ロックイン
Lock-in +180度
+180°
検出器
detector
Acosδ
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
A signal output from the lock-in detector.
AsinD
This signal and the Acos signal are used to calculate the phase () and amplitude (A).
A signal output from the lock-in detector.
AcosD
This signal and the Asin signal are used to calculate the phase () and amplitude (A).
May result in a signal that displays predominant viscosity when observing comparatively
Phase
soft samples, such as polymer materials and biological samples.
May result in a signal that displays predominant elasticity when observing comparatively
Amplitude
soft samples, such as polymer materials and biological samples.
Reference:
M. Radmacher, R. W. Tillmann, and H. E. Gaub. 1993. Imaging viscoelasticity by force modulation
with the atomic force microscope. Biophys. J. 64:735-742
Do NOT use the force modulation mode in conjunction with the petri-dish type solution cell (option).
Cantilever
S S
Scanning direction
N N
Repulsive force Attractive force
Sample N S N S
Fig. 2.6.1 Magnetic Force (MFM) Mode
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
AsinD An intermediate signal used to calculate the phase () and amplitude (A).
AcosD An intermediate signal used to calculate the phase () and amplitude (A).
Indicates the delay in phase with respect to the excitation signal applied to the cantilever.
Phase
This signal also indicates magnetic information.
Amplitude Indicates magnetic information, although with lower sensitivity when compared to the phase.
Photodetector
Bias voltage
Sample
Sample holder
Sample stage
Scanner
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target feedback value.
Current Indicates the electric current flowing from the sample to the cantilever.
Laser beam
Lock-in amplifier
Photodetector
Cantilever
(conductive)
AC
~ +
Sample
DC
++++ ----
See "10.1 Operating Principle of Surface Potential (KFM) Mode" for details on the operating principle
of surface potential (KFM) mode.
カンチレバーの位置信号
Cantilever position signal
加振信号 signal
Excitation カンチレバー振動
Cantilever vibration
Potential signal
電位信号(AC+DC)
(AC+ DC) 電位フィードバック(DC)
Potential feedback (DC)
(Potential deflection)
Asinδ(電位偏差)
電位
Potential
ロックイン
Lock-in Reference
検出器
detector 参照信号
signal
感度
Sensitivity カンチレバー電位(AC)
Cantilever potential (AC)
1, 2, x4,
x1, x2, 4,x8
8 位相オフセット
Phase offset
-100~100度,
-100° to 100°,
+180度
+180°
Fig. 2.8.2 Signal Processing in Surface Potential (KFM) Mode
Signal Description
Height Indicates the surface topography.
Deflection Indicates the difference from the target height feedback value.
AsinD Indicates the difference from the target potential feedback value.
Potential Indicates the electric potential on the sample surface.
! CAUTION
Do NOT allow the cantilever or cantilever hold block to make contact with other metal parts.
The cantilever holders for current mode (option) and surface potential (KFM) mode (option)
communicate electric signals with the cantilever via the hold block. The cantilever is held
down by an insulated presser bar.
Cantilever
Hold block
Presser bar
In order to obtain suitable image data, use a cantilever appropriate to the observation mode.
Note that inappropriate parameter settings may cause:
Noise to appear in image data
Trace and retrace in the line profile to be out of synchronization or unstable
For details on recommended parameter values when starting observation and the adjustment of parameters,
refer to the "SPM-9700HT Observation Guide" (P/N 305-26223).
3. Configuration
3.1 Appearance
Fig. 3.1.1 SPM-9700HT Scanning Probe Microscope shows the SPM-9700HT scanning probe
microscope system.
SHIMADZU
SPM Control Unit
SPM-9700HT Accessories
Sample holders 10
Cantilever/probe
for contact mode 1 (includes 34) *1
for dynamic and phase modes 1 (includes 10) *1
Standard samples
Niobium-coated plate 1
Precision grid plate 1
Instruction manuals
Hardware instruction manual 1
Software instruction manual 1
Observation guide 1
Backup media 1
Maintenance accessories (contained in accessory case)
Tweezers, double curve 1
Tweezers, precision 1
Hexagon wrench, 3 mm 1
Grease 1
Flat-bladed screwdriver, 1.6 mm (contained in scanner box) 1
Accessory case 1
The following items are included only in the model with CCD.
Microscope TV adapter 1
CCD camera 1
AC adapter for CCD camera 1
Coaxial cable with BNC connector, 5 m 1
BNC-RCA conversion connector 1
Image capture cable 1
Installation media for image capture cable 1
! CAUTION
If the sample observation lever is set to position B
and then returned to position A, check that the laser
spot position has not been displaced. If it is
displaced, perform readjustment.
The laser spot position with respect to the cantilever
may be displaced when the sample observation lever
is set to position B.
This window is used to observe the cantilever and sample from directly
above with an optical microscope.
Sample Observation
4 Window
When observing the cantilever and sample with the optical microscope
unit (option) or high-magnification microscope unit (option), set the
objective lens on top of the sample observation window.
This knob is used to move the position of the cantilever with respect to
the sample along the X (lateral) direction and can also be used to adjust
5 X Stage-Positioning Knob the position for sample observation.
The cantilever can be moved ±3 mm if the sample size does not exceed
18 mm along the X direction.
This knob is used to move the position of the cantilever with respect to
the sample along the Y (vertical) direction and can also be used to adjust
6 Y Stage-Positioning Knob the position for sample observation.
The cantilever can be moved ±3 mm if the sample size does not exceed
18 mm along the Y direction.
This knob is used to change the angle of the mirror inside the SPM head.
7 Mirror Adjustment Tab The mirror is used to direct laser light reflected from the cantilever to the
detector.
Fig. 4.2.4 Moving the Detector Vertically (Left) and Laterally (Right)
No. Name
1 Tool microscope
2 Objective lens, 2x
3 Eyepiece lens, 20x
4 Stand
5 Microscope TV adapter
6 CCD camera
7 AC adapter for CCD camera
8 Coaxial cable with BNC connector, 5 m
9 BNC-RCA conversion connector
0 Image capture cable
a CD-ROM containing Image capture cable software
* If there is no CCD, the configuration consists of items 1, 2, 3, and 4 only.
4.5.1 Preparation
1. Adjust the stand height and the arm length, and position the tool microscope above the SPM
head's observation window.
2. Position the end of the fiber light's (option) light guide in front of the SPM head so that the
sample can be illuminated.
Be sure to connect the USB cable to the USB 2.0 port ( ). If the cable is connected to the USB 3.0 port
( ), the microscope cannot operate correctly and thus no image is displayed.
Do NOT enable to put the computer to sleep (Default setting is “Never”). It may cause the CCD camera to fail
to work properly.
Click (Live) in the [OM Image] window to display the optical microscope image.
This window is enabled when a CCD-equipped optical microscope is provided with the instrument.
Do NOT insert or extract other USB device while the optical microscope is live. In this case, there is a
possibility that the optical microscope is undetected.
4.5.4 Magnification
Images sent to the host computer from the CCD camera via the image capture cable are displayed on
the monitor. The display magnification of the displayed image can be changed using the size and
zoom settings (digital zoom) in the [OM Image] window.
The display magnification for the initial window size without zoom is given below.
The magnification for observation using an eyepiece lens instead of the CCD camera is 40x.
4.5.5 Options
The options given below are available for Optical Microscope Unit.
No. Name
1 Tool microscope
2 TV rear capacitor lens
3 CCD camera
4 AC adapter for CCD camera
5 Lamp house
6 Lamp house controller
7 Image capture cable
8 Universal stand
9 Bonder mount
0 Coaxial cable with BNC connector
a BNC-RCA conversion connector
b Installation media for image capture cable
4.6.1 Preparation
1. Adjust the universal stand's height and the arm length, and position the zoom microscope
above the SPM head's observation window.
2. Position the end of the fiber light's (option) light guide in front of the SPM head so that the
sample can be illuminated.
Be sure to connect the USB cable to the USB 2.0 port ( ). If the cable is connected to the USB 3.0 port
( ), the microscope cannot operate correctly and thus no image is displayed.
Do NOT enable to put the computer to sleep (Default setting is “Never”). It may cause the CCD camera to fail
to work properly.
4.6.5 Magnification
If the zoom magnification of the zoom microscope is changed (on a scale of 0.8 to 5), the optical
zoom operates and the display magnification changes.
The display magnification of the displayed image can be changed using the size and zoom settings
(digital zoom) in the [OM Image] window.
The optical magnification changes by a factor of 3 depending on whether or not the TV rear capacitor
lens is attached.
The display magnification with and without the TV rear converter lens for the initial window size
without zoom is given below.
Display Magnification
Observation Range Display Range Zoom Microscope Scale
0.8 to 5
Approx. 0.9 0.7 mm to
Approx. 92 69 mm Approx. 16x to 100x
approx. 5.7 4.3 mm
Display Magnification
Observation Range Display Range Zoom Microscope Scale
0.8 to 5
Approx. 0.3 0.2 mm to
Approx. 92 69 mm Approx. 48x to 300x
approx. 1.9 1.4 mm
If the zoom magnification of the zoom microscope is between 0.8 and 1.5 (on a scale of 0.8 to 5),
coaxial incident-light illumination does not reach the area surrounding the viewing field, causing it to
become darker.
! CAUTION
Always store standard samples according to the following conditions.
Do NOT touch the surface of standard samples.
Do NOT wet the surface of standard samples with any type of liquid.
Store standard samples in their specialized case when not in use.
Do NOT store standard samples in environments subject to high temperatures and
humidity or significant fluctuations in temperature and humidity.
Temperature: 23 °C ± 5 °C
Humidity: 30 % max.
3. Turn ON the power to the standard optical microscope unit's (option) or the high-magnification
microscope unit's (option) CCD camera.
Be sure to connect the USB cable to the USB 2.0 port ( ). If the cable is connected to the USB 3.0 port
( ), the microscope cannot operate correctly and thus no image is displayed.
6. Start the online, browser, or offline software according to the operation you wish to perform.
If higher quality image data is required, start the online software and wait several hours after illuminating the
laser before performing observation.
5. Shut down Windows and turn OFF the power to the host computer.
! CAUTION
Do NOT turn OFF power to the host computer, before the host computer completely stops.
Turning OFF power in operating conditions, especially terminating the host computer, may
cause serious damage to the SPM software, the windows, or the host computer.
6. Maintenance
6.1 Handling the Scanner
6.1.1 Magnet Type Stub
A magnet type stub is attached in each type of scanner for the SPM-9700HT. (Fig. 6.1.1.)
The magnet type stub is fixed with a stab mounting screw, and can be removed.
The stab mounting screw may become loose during transportation or use. Before attaching the
sample holder, confirm using the dedicated screwdriver attached that the mounting screw is not loose.
If it is loose, slightly tighten it.
! CAUTION
When turning the stub mounting screw in the scanner, use the attached screwdriver, and do
not apply an excessive force or impact force.
Scanner cap
! CAUTION
Do NOT allow the scanner to suffer physical impacts while the scanner cap is removed.
! CAUTION
The scanner is a precision component and must be handled with care.
2. Turn OFF the power to the control unit and the host computer.
5. Lift the SPM head unit slightly and move it to the right.
1. Insert the new scanner so that the expanding slot is located on the near side.
At this time, pull out the cable through the notch on the left side so that it is not pinched.
3. Return the SPM head and slide it a short distance horizontally until it sets in the home position.
The SPM head will not move horizontally while in the home position.
3. If the scanner cannot be selected in the manager, copy the scanner file with the file extension
".PZT" from the media provided with the scanner to the Env folder
(C:\Users\Public\SPM-9700\Env).
4. Start the manager, select the scanner, and then start the online software.
6.4 Storage
6.4.1 Main Unit
When not using the instrument for a long period, or when storing it, maintain the surrounding
environment at conditions conforming as closely as possible to the installation specifications, cover
the instrument with a vinyl sheet, and store it together with a desiccant. It is recommended that, if
possible, the instrument is stored with the cables connected (except power cables).
Do not store the equipment in the following locations:
Locations subject to high temperatures and high humidity levels
Locations subject to direct sunlight, wind, or rain
Locations close to highly corrosive chemicals
Unstable locations from which the instrument may fall
Screws
Fig. 6.6.1 Z-axis Drive Unit Screws and Grease Applied to Prevent Rust
Replacing Fans
Fans must be replaced if they emit strange noises, do not rotate properly, or exhibit some other
irregularity. The service life of a fan is approx. 2 years. It is recommended that fans are replaced
regularly even if there are no irregularities. Contact your Shimadzu representative to request
replacement.
Replacing Fuses
Before replacing a fuse, remove the cause of the overcurrent.
1. Turn OFF the power to the control unit and disconnect the power cable.
2. Remove the control unit's fuse cover ((5) in Fig. 4.3.1) with a flat-bladed screwdriver.
7. Specifications
7.1 Main Units
7.1.1 SPM Unit
External dimensions Approx. 180 (W) 255 (D) 260 (H) mm
Weight Approx. 5.5 kg
Contact mode
Dynamic mode
Phase mode
Lateral force (LFM) mode
Observation modes
Force modulation mode
Magnetic force (MFM) mode (option)
Current mode (option)
Surface potential (KFM) mode (option)
XY: 0.2 nm
Resolution
Z: 0.01 nm
Displacement detection Light source, optical lever, detector
SPM head Light source Laser diode (650 nm, 0.6 mWmax, ON/OFF possible)
Detector Photodetector
Drive element Tube-type piezo element
10 m 10 m 1 m
30 m 30 m 5 m (option)
Scanner Max. scanning size
125 m 125 m 7 m (option)
(X Y Z)
55 m 55 m 13 m (option)
2.5 m 2.5 m 0.3 m (option)
24 (dia.) 8 mm
Maximum sample size
(20 (dia.) 8 mm in current mode)
Stage Sample replacement method Head-slide mechanism
Sample securing method Secured with magnet
SPM Head drive range 6 6 mm (for sample diameter of 18 mm)
Type Fully automated mechanism based on stepping motor
Z-axis drive unit
Maximum stroke 10 mm
Signal display panel Display volume Total light intensity incident to the detector (digital display)
Vibration isolation
Vibration isolator Built into SPM unit
mechanism
7.1.4 Software
Interface Language English
Online Observation mode, scanner selection, startup
Offline Function selection, startup
Manager
Specifying and saving of data paths for each user and
User settings
temporary paths
Signal Select from a maximum of 6 signals
Scanning direction Trace, retrace (simultaneous measurement possible)
Scanning angle -90 to 90 deg
Scanning size 0.1 nm to the scanner's max. scanning size, offset settings
Scanning rate 0.015 to 100 Hz
2048 2048*3, 1024 1024*3, 512 512,
256 256, 128 128, 64 64*4, 32 32*4
(*3 Performs displaying as compressed 512 512.
Number of pixels
Performs processing and analysis after conversion to 512
512 pixels or below)
(*4 Pixels cannot be saved)
Data size 8.03 MB, 2.03 MB, 544 kB, 160 kB, 64 kB
Operating point, P gain, I gain
Feedback settings
Switch between constant force or constant height
Online Scanner operating status display
Scanner operation settings Z range settings
Offset settings (automatic adjustment possible)
Approach (automatic, fast, slow)
Z-axis course adjustment
Release (10 μm, 100 μm, release)
Stop
Switch XY scanning ON/OFF
Scanning Switch Y scanning ON/OFF, restart
Start position of Y scanning can be changed (top, center,
bottom)
Number of screens (single screen display, dual screen
display (upper/lower, left/right)), quad screen display)
View method (gray, solid, mix)
Display range settings (Z display range, offset)
Image data display
Color palette settings (400 types)
Ruler display
Tilt correction settings (none, Y, XY)
Image history display (list display, single image display)
7.2 Cantilever
Table 7.2.1 Recommended Cantilevers
Probe Reflective k f L
Part Number Part Name Application Material Lever Shape Quantity
Coating Coating [N/m] [kHz] [µm]
Cantilever,
0.57 73 100
SSOT-AK0546 OMCL-TR800PS Contact mode *5 SiN None Au Triangle 34
0.15 24 200
A-1
Cantilever, Dynamic and 330
SSOT-AK0556 NCHR-20 phase modes
Single-crystal Si None Al Rectangle 20 42
(260-410)
125
Cantilever, 0.2 13
SSOT-AK0551 CONTPT-10
Current mode Single-crystal Si PtIr5 Al Rectangle 10
(0.07-0.4) (10-17)
450
*5 Also usable in lateral force (LFM) mode and force modulation mode.
H/2 H
8. Installation
In principle, installation of the SPM-9700HT is performed by Shimadzu service personnel. Facilities
that satisfy the installation specifications described below are required for installation and must be
prepared by the customer.
For details on installation specifications, refer to the installation guide.
Switchboard
Max.10m
Stepdown
3m Transformer
3m
Monitor
800
Control unit
600
250 420 454
Approx. 18.5 kg
Optical microscope Fiber light
unit (OP) (OP)
1200 800
! CAUTION
Always install the control unit on a table. Installation on anything other than a table may cause
the following issues.
Legs, feet, or chairs may come into contact with the power switch, turning OFF the control
unit during operation and resulting in the loss of important data.
Legs, feet, or chairs may come into contact with and damage the power switch.
Personnel may injure themselves due to incorrect posture when operating the power switch
or checking the illumination status of lamps.
8.1.1 Tables
A large table with dimensions of approx. 1200 800 mm and another large table with dimensions of
approx. 900 800 mm are required for installation.
If higher quality image data is required, instead of using a table with approximate dimensions of
900 mm 800 mm, Shimadzu recommends installing the SPM unit on an air-spring vibration isolator
table of the same size.
Installing the instrument in an environment that does not satisfy the installation specifications may cause
noise to appear in images.
Capacity/Power
Unit Power Supply Grounding
Consumption
100 to 120 VAC, 50/60 Hz 100 Ω
Control unit 100 W
200 to 240 VAC, 50/60 Hz or less
SPM-9700HT
100 to 120 VAC 10%, 50/60 Hz 100 Ω
Monitor 45 W
200 to 240 VAC 10%, 50 Hz or less
Optical Microscope AC adapter for Not
100 to 120 VAC, 50/60 Hz 100 W
Unit (Option) CCD required
AC adapter for Not
High-Magnification 100 to 120 VAC, 50/60 Hz 100 W
CCD required
Optical Microscope
Not
Unit (Option) Light controller 100 VAC, 50/60 Hz 35 W
required
100 to 120 VAC/200 to 240 VAC, 100 Ω
Fiber Light (Option) 150 W
50/60 Hz or less
Two extension cables (each equipped with a switch) of the following specifications are provided with
this system.
Number of outlets: 6
Cable length: 3m
Ratings: 15 A, 125 V (less than 1.5 kW)
Use the extension cable at the specified ratings.
! CAUTION
Be sure to connect the grounding wire.
CCD
Optical Host PC
microscope
Image
capture
cable
AC adapter
To switchboard
Extension cable
Fig. 8.2.1 Cable Connection Diagram for Standard Optical Microscope Unit
CCD
Host PC
High-magnification
microscope
Light controller
Image
capture
cable
AC adapter
To switchboard
Extension cable
Fig. 8.2.2 Cable Connection Diagram for High-Magnification Optical Microscope Unit
Video SVideo
Video Composite(Yellow)
Be sure to connect the USB cable to the USB 2.0 port ( ). If the cable is connected to the USB 3.0 port
( ), the microscope cannot operate correctly and thus no image is displayed.
Fiber light
To switchboard
Extension cable
9. Troubleshooting
Problem Main Cause Corrective Action
Did the control unit start up Check that the control unit's Ready
normally? lamp is lit.
Check that the host computer and
Are the host computer and control
control unit are connected with a LAN
unit connected with a LAN cable?
cross cable.
A connection is not Check that the TCP/IP protocol
established with the control settings for the host computer's LAN
unit when the online software adapter are configured as follows:
is started. IP address: 192.9.200.2
Are the communication settings on Subnet mask: 255.255.255.0
the host computer correct? Display a [Command Prompt] window
from the [Start] menu on the host
computer, enter the command "ping
192.9.200.1", and check that a
response is received.
Select [Windows Firewall] in the
The online software is
[Control Panel] window on the host
interrupted (i.e. scanning Is Windows Firewall enabled?
computer and then disable the
stops).
firewall.
Has the optical axis been adjusted
Readjust the optical axis.
correctly?
Replace the cantilever and readjust
The laser beam is not Is the cantilever bent or deformed?
the optical axis.
detected.
Check the wiring of each component
Are there any broken wires? (especially around the SPM head) for
broken or disconnected wires.
The laser power has Has the optical axis been adjusted
Readjust the optical axis.
dropped. correctly?
Has the optical axis been adjusted
Readjust the optical axis.
correctly?
The approach (coarse Are all the connectors inserted Check that each connector is inserted
adjustment) is not completed correctly? securely.
normally. Check the wiring of each component
Are there any broken wires? (especially around the SPM head) for
broken or disconnected wires.
Is the appropriate cantilever Use a cantilever suitable for the
selected? observation mode.
Is the cantilever correctly attached
Correctly attach the cantilever.
to the cantilever holder?
Is the cantilever holder correctly
Image data cannot be Correctly attach the cantilever holder.
attached to the SPM head?
obtained.
The resolution of the images Is the cantilever bent or deformed? Replace the cantilever and readjust
data is poor (exhibits noise). Is the cantilever worn out? the optical axis.
Trace and retrace in the line
profile is poorly synchronized Is the laser beam spot reflected
or unstable. correctly around the end of the Readjust the optical axis.
cantilever?
Is each parameter in the online Refer to the observation guide and
software set appropriately? adjust each parameter.
Are all the connectors inserted Check that each connector is inserted
correctly? securely.
10. Appendix
10.1 Operating Principle of Surface Potential (KFM)
Mode
Surface potential (KFM) mode generates images based on the electric potential of the sample
surface.
Applying an AC voltage to a conductive cantilever and detecting the resulting electric force makes it
possible to observe the surface profile and, at the same time, the surface potential distribution.
Here, we consider a cantilever and sample with uniform surfaces. We assume that they are both
conductors and that their respective surface potentials are Vt and Vs respectively. In this case, the
electrostatic free energy, G, in a system kept at a constant potential can be expressed by formula (1)
below, where d is the distance between the cantilever and the sample surface and C is the
capacitance between the cantilever and the sample.
1
G C (Vt Vs ) 2 (1)
2
The electric force, Fze(d), that acts on the cantilever can be expressed by formula (2).
G 1 C
Fze (d ) (Vt Vs ) 2 (2)
d 2 d
Because the capacitance, C, decreases as d increases, the electric force between the cantilever and
the sample is always attractive, and the size is proportional to the capacitance gradient. The voltage
applied externally to the cantilever can be expressed by formula (3).
1 C 1 2 1 2
F (d , t ) Fzm (d ) (VDC Vs ) 2 VAC 2(VDC Vs )VAC cos t VAC cos 2t (4)
2 d 2 2
From this, the amplitude of the vibration electric force corresponding to angular frequency can be
expressed by formula (5).
C
Fe (VDC Vs )VAC (5)
d
In SPM profile measurement, the distance between the cantilever and the sample is constant and so
it can be assumed here that the capacitance gradient and VAC are constant. Therefore, if feedback
control is performed for VDC, which is applied to the cantilever, so that Fωe is always equal to zero, VDC
will be equal to Vs and can be obtained as the potential of the sample surface.
Bibliography:
1) M. Nonnenmacher, M. P. O'Boyle and H. K. Wickramasinghe, Appl. Phys. Lett. 58 (1991) 2921
2) S. Morita, Atomic/Molecular Nanomechanics (2003) 33