0% found this document useful (0 votes)
17 views2 pages

LS Instruments 9 Optical Path Length Reduction

The document discusses the importance of optical path length reduction in dynamic light scattering (DLS) to minimize multiple scattering events, which can lead to measurement errors. It describes two sample cell positions for different sample concentrations to optimize DLS measurements and highlights the use of advanced technologies like Modulated 3D for reliable results. Additionally, it addresses challenges in angle-dependent DLS and the need for precise sample rotation to maintain alignment during measurements.

Uploaded by

tinyhope
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
17 views2 pages

LS Instruments 9 Optical Path Length Reduction

The document discusses the importance of optical path length reduction in dynamic light scattering (DLS) to minimize multiple scattering events, which can lead to measurement errors. It describes two sample cell positions for different sample concentrations to optimize DLS measurements and highlights the use of advanced technologies like Modulated 3D for reliable results. Additionally, it addresses challenges in angle-dependent DLS and the need for precise sample rotation to maintain alignment during measurements.

Uploaded by

tinyhope
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Online Shop Contact Jobs Support

Products & Services Industries Applications Learning About

Dynamic Light Scattering (DLS) Optical Path Length Reduction

Introduction
Both static and dynamic light scattering rely on the measurement of single scattering
Theory events, meaning that each detected photon must have been scattered only once before
reaching the detector. Multiple scattering is thus a major issue, as it leads to significant
DLS Data Analysis: CUMULANT measurement errors, and prevents the characterization of native, undiluted systems.
Method Advanced light scattering techniques such as the Modulated 3D technology, can
suppress multiple scattering such that the result obtained is reliable. In a typically
DLS Data Analysis: The CORENN
turbid sample, both multiple and single scattering events are taking place and the
Method
“filtered” signal consists solely of single scattering events.
3D Cross-Correlation
However, some systems are so turbid that only multiple scattering is detected. In such
Modulated 3D Cross-Correlation cases, the entirety of the signal is filtered out and no reliable DLS signal can be
measured. One must therefore ensure that a minimum of single scattering events is
DLS Microviscometry recorded.

DLS Microrheology This can be achieved by placing the sample cell such that the optical path length within
the sample is reduced. Figure 1 shows two possible sample cell positions:
Optical Path Length Reduction

Experimental Guidelines - DLS


Sample Preparation

a) Corner position: reduces multiple scattering. Used for concentrated samples. b) Body
position: ideal to provide a clean signal when measuring a dilute system.

Figure 1: Positioning of the sample cell.

In Figure 1a), the cell is placed such that the laser beam enters as close as possible to
its corner, and the optical path length can be lowered down to 100 mm, as allowed by
the reproducibility precision of the cuvette positioning and the dimensions of the
cuvette. Since the probability that a photon is scattered multiple times before reaching
the detector is proportional to the laser beam and detection beam path lengths within
the cell, the optical path length reduction results in a drastic decrease of the number of
multiple scattering events. At the same time, the number of single scattering events is
increased. This means that, even in opaque suspensions, one can record single
scattering events and thus obtain a reliable DLS signal.

In the opposite situation of dealing with extremely dilute samples, a different problem
must be addressed: the vicinity of the cell wall to the scattering volume gives rise to
scattering events, known as stray light scattering events, originating from the cell walls
which alter the desired signal. If the frequency of such events becomes comparable to
the low number of scattering events, given the diluted conditions, occurring in the
sample the final DLS measurement result becomes invalid. In this case, one can move
the sample cell in the so-called body position such that a sufficient distance from the
cell wall is established (see Figure 1b)).

Translation of the sample is possible in both the NanoLab 3D and LS Spectrometer. In


combination with the Modulated 3D technology, one can span the widest range of
concentration possible nowadays.

In systems where the scattering angle can be changed as required by SLS and/or angle-
dependent DLS, the technical implementation of the optical path length reduction is
more challenging: Figure 1a) suggests that, if the detection virtual beam is rotated, its
angle of incidence on the cell wall will be different from the angle of incidence of the
illumination beam; as a result, a scattering angle change will result in an alignment
loss. To overcome this limitation, one has to rotate the cell around the scattering
volume by an angle such that the two angles of incidence are equal. This technique,
illustrated in Figure 2, goes under the name of " technique" and requires a very
precise rotation around the axis orthogonal to the scattering plane and passing through
the scattering volume. Such rotation must be implemented by a rotation stage that has
to be coupled to the two translational stages required to place the scattering volume in
the desired position. In the LS Spectrometer, this is achieved using the optional Sample
Goniometer. This unit allows for both rotation and translation of the sample with high
precision1.

Figure 2: Rotation of the square cell in the corner position according to the
technique.

1
For more information about the Sample Goniometer option of the LS Spectrometer
please visit:

https://lsinstruments.ch/en/products/ls-spectrometer#option-sample-goniometer

Previous Chapter Next Chapter

Related Products

LS Spectrometer NanoLab 3D LSI Detection Unit for


PCS

Join our newsletter and stay in the know!


Subscribe to the Newsletter

We promise not to spam you, we also hate that.

Products & Services Technology and About Us


Support
Contact
LS Instruments AG
Products About
Bluefactory, Bât. A
Services Customer Support Jobs
Passage du Cardinal
Tools for Light Center Press
1
Scattering Light Scattering News
CH-1700 Fribourg
Consumables Theory References
Terms & Conditions Webinars

© 2020 by LS Instruments Impressum Disclaimer

You might also like