HF Analyzer HF-400
Electrosurgical Analyzer for function tests of
HF Surgical Equipment
menu guided cursor operation or PC-operation
HF - power measurement
HF - voltage measurement
HF - current measurement
HF - leakage current measurement
test load resistances 10 Ohm, 25 – 6375 Ohm
in steps of 25 Ohm
6 selectable pre resistance
user specific language setting
option - suit case
Test and measurement technic
for medicine and industry
Technical Data
Line voltage: 83 – 264 V ac, 50 /60 Hz Measurement range error
Power consumption: 50 VA HF output power: 0 – 500 W ± 1 W or
Class: 1 ± 2,5 % of value
Environmental temperature: + 5 - + 40°C HF leakage current:: 0 – 250 mA ± 2 mA
Storage temperature: - 10 - + 50°C ± 5 % of value
Load resistors: 10 Ohm,
Measuring ranges: 25 – 6375 Ohm ±5%
HF-current RMS: 0 - 5000 mA
HF-current Peak: 0 - 5000 mA Keyboard: 6 key foil keyboard
Discrimination: 0,1 mA Display: 4 x 20 char LCD B/W display
HF- output power RMS: 1 – 500 Watt Interfaces: 1 x USB for PC interface
(in dependence of RL 1 x RS-232 for PC interface
Crest Faktor: (V2) 1 – 10 (bei > 1000 mA) 1 x RS-232
HF-leakage current: 0 – 250 mA for additional test devices
Discrimination: 0,1 mA
Bandwidth 0,3 – 10 MHz Testing plugs: 2 x safety plugs 4 mm for
Measuring principle: thermal electric converter HF power
Load resistors: 10 Ohm 2 x safety plugs 4 mm for
25 Ohm – 6375 Ohm HF leakage current
In steps of 25 Ohm 1 x safety plug 4 mm for PE
Swing in time: < 3 sec 1 x potential balance
Output power: 500 W: 1 min on, 5 min off
permanent: max. 200 W at 25°C Accessories: 1 x potential balance cable
environmental temperature 1 x RS-232 interface cable
Mechanical data: light way metal case IP20
Dimensions: 340 x 87 x 290 mm (W x H x D)
Weight: approx. 3,8 kg
Selectable languages: german, english, french, polish
spanish, italian, portuguese, turkish
Description of functions: HF output power:
HF-400 serves to test the function of HF During the measurement of power, firstly the
Surgical Equipment. In accordance to the software sets the prescribed load resistance
instructions of the manufacturer of such to 10 Ohm or from 25 Ohm to 6375 Ohm in
surgical devices, the user can measure the 25 Ohm steps. Than the HF output power
HF output power and the HF leakage current can be send to the HF-400 and is measured.
given on a load resistor. The load resistor is An automatic range switcher takes care of
adjustable to 10 Ohm and from 25 – 6375 the optimal control of the RMS-converter.
Ohm in steps of 25 Ohm. The test The RMS converter based on a thermal
parameters for testing can laid down in a test conversion principle and, together with the
instruction and can be tested automatically driver module, is designed for frequencies up
with a PC. This makes it possible to reduce to 10 MHz.
the time for testing. In the use as multi
functional test device the measured values HF leakage current:
were displayed directly. For example:
The high-frequency leakage current is
HF output power measured through a 200 Ohm load resistor.
HF leakage current During this test load resistor is adjustable.
HF current, RMS
HF voltage, RMS