IPS-4 Manual
IPS-4 Manual
IPS-4 Analyzer
Full Spectrum Option
General Purpose/Division 2
User Manual
Process Instruments
455 Corporate Blvd.
PN 403469902, Rev B Newark, DE 19702
Offices
USA - Oklahoma
2001 N. Indianwood Ave., Broken Arrow OK 74012 • Tel: 918-250-7200, Fax: 918-459-0165
USA - Pennsylvania
150 Freeport Road, Pittsburgh PA 15238 • Tel: 412-828-9040, Fax: 412-826-0399
CANADA - Alberta
2876 Sunridge Way N.E., Calgary, AB T1Y 7H9 • Tel: +1-403-235-8400, Fax: +1-403-248-3550
CHINA
Beijing / Tel: 86 10 8526 2111, Fax: 86 10 8526 2141
Chengdu / Tel: 86 28 8675 8111, Fax: 86 28 8675 8141
Shanghai / Tel: 86 21 6426 8111, Fax: 86 21 6426 7818
FRANCE
Tel: 33 1 30 68 89 20, Fax: 33 1 30 68 89 29
GERMANY
Tel: 49 21 59 91 36 0, Fax: 49 21 59 91 3639
SINGAPORE
Tel: 65 6484 2388, Fax: 65 6481 6588
[Link]
Overview 1
OPTICAL BENCH CONFIGURATIONS........................................................... 1
Basic Analytical Theory (Optical Bench) .................................................... 1
Flow Diagram .................................................................................................. 1
Calculation Flow ............................................................................................. 2
Dispersive Ultraviolet/Visible (DUVV) Spectrometer ..................................... 4
Basic Analytical Theory ............................................................................... 4
Optical Bench Design ................................................................................... 4
UV Specifications ............................................................................................ 5
Wavelength Ranges ........................................................................................ 5
Wavelength Accuracy+/- 0.5 nm ..............................................................5
Wavelength Resolution3 nm ...................................................................... 5
Analyzer Rangeppm to 100 %, depending on the application. .......... 5
Photometric Range...................................................................................... 5
Photometric Noise ......................................................................................5
Non-Dispersive Infrared (NDIR) PhotometerBasic Analytical Theory . 6
Optical Bench Design .................................................................................... 6
NDIR Specifications ..........................................................................7
Wavelength Ranges ...................................................................................7
Photometric Noise .....................................................................................7
Sample System ................................................................................................ 7
Electronics ........................................................................................................ 7
Classified, Hazardous Area Protection Components and Functions ........... 9
Specifications 2
Analyzer Specifications ........................................................................................ 1
Analytes ............................................................................................................ 1
Response Linearity ......................................................................................... 1
Measurement Accuracy ................................................................................. 1
Contents | iii
Repeatability .................................................................................................... 1
Linearity ........................................................................................................... 1
Stability ............................................................................................................. 1
24-Hour Zero Drift ......................................................................................... 1
Inputs................................................................................................................ 2
Outputs ............................................................................................................ 2
Sample System Limits .................................................................................... 2
Sample Pressure .........................................................................................2
Oven/Sample System Enclosure Temperature............................................2
Sample Transport ............................................................................................ 3
Instrument Air Requirements....................................................................... 3
Sample Fluid Flow Rate ................................................................................. 3
Electrical Requirements ................................................................................. 3
Power Consumption ...................................................................................3
Heated Sample Line ...................................................................................4
Supply Voltage ............................................................................................4
Cell Construction ............................................................................................ 4
Pressure Input Signals ................................................................................... 4
Environmental ................................................................................................ 4
Ambient Temperature .................................................................................4
Humidity ....................................................................................................4
Pollution Degree .........................................................................................4
Maximum Altitude.....................................................................................4
Installation Category ..................................................................................5
Enclosure Material .....................................................................................5
Ingress Protection .......................................................................................5
On-Board Temperature Sensor..................................................................... 5
Physical Dimensions ...................................................................................... 5
Approvals and Certifications ........................................................................ 6
Contents | v
Modbus Screen .........................................................................................39
APPENDIX – DRAWINGS a
Analyzer Light Path Schematic ........................................................................... 2
UV Schematic ......................................................................................................... 2
NDIR Schematic .................................................................................................... 2
Analyzer Mounting Details ................................................................................. 3
Analyzer Overall Component Layout ............................................................... 4
Detector Interface Board (403441901) ................................................................ 5
Detector Board (100-1955) .................................................................................... 6
Detector board (700030901).................................................................................. 6
Display Interface Board (100-1978)..................................................................... 7
Display Interface board (700043901) .................................................................. 7
Customer Connection Board (100-1971) ............................................................ 8
Relay Board (100-1942) ......................................................................................... 9
Analog Board (100-1922) .................................................................................... 10
MCU Board (100-2045) ....................................................................................... 11
Xenon Lamp Board (100-2061) .......................................................................... 12
Electronics Enclosure Wiring Diagram, Sheet 1 of 5 (403443001) ................ 13
Electronics Enclosure Wiring Diagram, Sheet 2 of 5 (403443001) ................ 14
Electronics Enclosure Wiring Diagram, Sheet 3 of 5 (403443001) ................ 15
Supplemental Information S
Contents | vii
Safety Notes
WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize critical instructions as fol-
lows:
An operating procedure which, if not strictly observed, may result in personal injury or envi-
! ronmental contamination.
WARNING
! An operating procedure which, if not strictly observed, may result in damage to the equip-
CAUTION ment.
Electrical Safety
High voltages are present in the analyzer housings. Always shut down power source(s) before perform-
ing maintenance or troubleshooting. Only a qualified electrician should make electrical connections and
ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection
originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are void if instru-
ment is operated from an improperly grounded power source.
All personnel involved with the installation, start-up, operation, maintenance, service, or troubleshooting
of the IPS-4 Analyzer must review and follow these Warnings and Cautions.
Warnings
Review and follow these Warnings to avoid personal injury or environmental contamination.
Always disconnect main AC power and/or external power sources to the analyzer before
opening any doors on the analyzer to check or perform maintenance on any components
! within the enclosures.
WARNING
If it is necessary to open the analyzer doors while the circuits are live, first monitor the area
for flammable gases and proceed only when the area is safe.
All electrical connections, adjustments, or servicing of the analyzer should be performed only
! by properly trained and qualified personnel.
WARNING
All electrical connections, materials, and methods (plus all safety policies and procedures)
must be made in compliance with local wiring regulations and electrical code for the hazard-
ous area.
! Follow appropriate regulatory and/or company procedures to lock out the analyzer while
WARNING working on the analyzer electronics.
! Before working on the analyzer, test the area for flammable gases. If a flammable gas atmo-
sphere is present, do not open any doors without first disconnecting power and/or alternate
WARNING
power sources to the analyzer.
The glass Xenon Flash Lamp is under high internal pressure which could result in flying glass
! fragments if ruptured. Handle the tube with care to avoid dropping it, subjecting it to impacts,
WARNING applying excessive force to it, or scratching it.
While handling a flash lamp, always wear protective devices (face mask, clothing) to prevent
possible injury, especially to hand and face areas.
!
WARNING
Before working on the sample system, confirm that the system has been purged with Zero
fluid and is isolated (blocked in) from the process.
Contents | ix
Because ultraviolet radiation can harm your eyes, never look directly into the operating lamp.
! If the Xenon Flash Lamp must be viewed while energized, wear safety glasses that block
WARNING ultraviolet radiation.
If the Oven/Sample System Enclosure is heated, the components within it will be hot; take
! precautions to avoid burning yourself.
WARNING
Cautions
The electronic circuit boards and other static-sensitive components should be stored and
! transported in static-shielding carriers or packages.
CAUTION
For electrical-shock protection, the analyzer must be operated from a grounded power source
! that has a securely connected protective-ground contact.
CAUTION
If it becomes necessary to handle any of the electronic circuit boards, do not subject the
boards to static discharge. The ideal solution is a static-safe work area. Since such areas typi-
! cally are not available at analyzer installation sites, the use of a wrist strap connected directly
CAUTION to a ground is recommended. If a wrist strap is not available, you should at the very least
touch the metal chassis (to ground yourself) before handling or touching the boards.
When handling the Xenon Flash Lamp, it is very important not to touch the lamp windows
! because residual oils from the fingers will absorb ultraviolet light. The window is the flat sur-
CAUTION face at the end of the narrow glass tube. The lamp assembly is fragile and should be handled
with care.
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a municipal waste
system (residential trash). The Wheelie Bin marking on this product is a reminder to dispose of the
product properly after it has completed its useful life and been removed from service. Metals, plastics and
other components are recyclable and you can do your part by one of the following these steps:
• When the equipment is ready to be disposed of, take it to your local or regional
waste collection administration for recycling.
• In some cases, your “end-of-life” product may be traded in for credit towards the
purchase of new AMETEK instruments. Contact your dealer to see if this program
is available in your area.
• If you need further assistance in recycling your AMETEK product, contact our of-
fice listed in the front of the instruction manual.
Contents | xi
Electromagnetic Compatibility (EMC)
! Read and follow the recommendations in this section to avoid performance variations or dam-
CAUTION age to the internal circuits of this equipment when installed in harsh electrical environments.
The various configurations of the IPS-4 Analyzer should not produce, or fall victim to, electromagnetic
disturbances as specified in the European Union’s EMC Directive (if applicable to your application).
Strict compliance to the EMC Directive requires that certain installation techniques and wiring practices
are used to prevent or minimize erratic behavior of the Analyzer or its electronic neighbors. Below are
examples of the techniques and wiring practices to be followed.
In meeting the EMC requirements, the various analyzer configurations described in this manual rely
heavily on the use of metallic shielded cables used to connect to the customer’s equipment and power.
Foil and braid shielded I/O and DC power cables are recommended for use in otherwise unprotected situ-
ations. In addition, hard conduit, flexible conduit, and armor around non-shielded wiring also provides
excellent control of radio frequency disturbances. However, use of these shielding techniques is effective
only when the shielding element is connected to the equipment chassis/earth ground at both ends of the
cable run. This may cause ground loop problems in some cases. These should be treated on a case-by-case
basis. Disconnecting one shield ground may not provide sufficient protection depending on the electronic
environment. Connecting one shield ground via a 0.1 microfarad ceramic capacitor is a technique allow-
ing high frequency shield bonding while avoiding the AC-ground metal connection. In the case of shield-
ed cables the drain wire or braid connection must be kept short. A minimal connection distance between
the shield’s end and the nearest grounded chassis point, ground bar or terminal is highly recommended.
An even greater degree of shield performance can be achieved by using metallic glands for shielded cable
entry into metal enclosures. Expose enough of the braid/foil/drain where it passes through the gland so
that the shield materials can be wrapped backwards onto the cable jacket and captured inside the gland,
and tightened up against the metal interior.
Inductive loads connected to the low voltage “Alarm Contacts” are not recommended. However, if this
becomes a necessity, adhere to proper techniques and wiring practices. Install an appropriate transient
voltage suppression device (low voltage MOV, “Transzorb,” or R/C) as close as possible to the inductive
device to reduce the generation of transients. Do not run this type of signal wiring along with other I/O or
DC in the same shielded cable. Inductive load wiring must be separated from other circuits in conduit by
using an additional cable shield on the offending cable.
In general, for optimum protection against high frequency transients and other disturbances, do not allow
installation of this Analyzer where its unshielded I/O and DC circuits are physically mixed with AC mains
or any other circuit that could induce transients into the Analyzer or the overall system. Examples of elec-
trical events and devices known for the generation of harmful electromagnetic disturbances include mo-
tors, capacitor bank switching, storm related transients, RF welding equipment, static, and walkie-talkies.
Refer to Chapter 2 – Specifications for details about the suitability of this equipment in hazard-
ous areas.
! Explosion Hazard – Substitution of Components May Impair Suitability for hazardous loca-
tions.
WARNING
! Explosion Hazard – Do Not Disconnect Equipment Unless Power Has Been Switched Off or
the Area is Known to be Non-Hazardous.
WARNING
For installations using North American wiring practice, all input and output wiring must be in
accordance with Class I, Division 2 wiring methods (NEC Sec 501.10 and 505.15(b) or (CEC-18
106 and in accordance with the authority having jurisdiction.
Contents | xiii
WARRANTY AND CLAIMS
We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which
shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship under normal
use and service will be repaired or replaced (at our option) by us free of charge, provided that written notice of such
defect is received by us within twelve (12) months from date of shipment of portable analyzers or within eighteen (18)
months from date of shipment or twelve (12) months from date of installation of permanent equipment, whichever
period is shorter. All equipment requiring repair or replacement under the warranty shall be returned to us at our factory,
or at such other location as we may designate, transportation prepaid. Such returned equipment shall be examined by
us and if it is found to be defective as a result of defective materials or workmanship, it shall be repaired or replaced
as aforesaid. Our obligation does not include the cost of furnishing any labor in connection with the installation of
such repaired or replaced equipment or parts thereof, nor does it include the responsibility or cost of transportation. In
addition, instead of repairing or replacing the equipment returned to us as aforesaid, we may, at our option, take back
the defective equipment, and refund in full settlement the purchase price thereof paid by Buyer.
Process photometric analyzers, process moisture analyzers, and sample systems are warranted to perform the intended
measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid sample stream com-
position data, process conditions, and electrical area classification prior to order acknowledgment. The photometric light
sources are warranted for ninety (90) days from date of shipment. Resale items warranty is limited to the transferable
portion of the original equipment manufacturer’s warranty to AMETEK. If you are returning equipment from outside
the North America, a statement should appear on the documentation accompanying the equipment being returned
declaring that the goods being returned for repair are North American goods, the name of the firm who purchased the
goods, and the shipment date.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after leaving
our control or which has been replaced by anyone except us, or which has been subject to misuse, neglect, abuse or im-
proper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to include, but shall not be limited
to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications shall be construed
to include improper or inadequate protection against shock, vibration, high or low temperature, overpressure, excess
voltage and the like, or operating the equipment with or in a corrosive, explosive or combustible medium, unless the
equipment is specifically designed for such service, or exposure to any other service or environment of greater severity
than that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original purchaser
from us. Should the Buyer’s technical staff require the on-site assistance of AMETEK’s agents or employees for service
calls covered by this warranty clause, the Buyer shall pay travel time plus actual travel and living expenses.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS
OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANT-
ABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER OBLIGATIONS OR LI-
ABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER
PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES,
CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL
PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR OR
OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY OR OTHERWISE.
WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRAN-
TIES OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT,
WHETHER SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE
DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY
IN CONNECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.
This chapter discusses the features of the IPS-4 Analyzer and a summary of how the
entire system operates. This chapter also provides an overview of the AMETEK Electron-
ics, Full Spectrum options, sample system, and the various sub-systems that make up the
entire analyzer system.
Flow Diagram
Both the DUVV and NDIR optical benches operate in parallel to measure transmittance.
Figure 1.1 illustrates the flow of information from the DUVV and NDIR optical benches
to the IPS-4 analyzer platform. A single PCB acts as the interface between the optical
benches and the MCU board where the signal processing occurs.
Overview | 1-1
DUVV
Detector Array
Detector PCB
100-2046
Dual-Bench
Interface PCB MCU
NDIR 403441901 100-2045
Detector
Preamp
700043901
Figure 1-1.
Signal Flow Diagram
Calculation Flow
The transmittance measurements contain chemical information that is used to determine
the analyzer output (e.g. concentration of HೣS). The following describes the flow of calcu-
lations relating transmittance to the desired analyzer output.
(EQUATION 1.2)
A(λ) = -log10(T(λ))
In this step, the factory established calibration converts either transmittance or absor-
bance information into a value proportional to the analyte of interest. This calibration
will be unique for each application. Because two optical benches are operating in parallel,
transmittance information from one or both of the covered spectral ranges can be used
A median filter is applied to the current analyte value reading V 1 incorporating the pre-
vious analyte value reading V0 . The size of the median filter is set in parameter MEDSZ.
The median filter is follow by an IIR filter (EQUATION 1.3) weighted by a non-linear
function defined by ALPHA and BETA parameters. The weight, wt, is calculated using
the following equation, where Vavg is the average of the last AVERAGE readings (EQUA-
TION 1.4). The values for MEDSZ, ALPHA, BETA, and AVERAGE dictate the response
lag due to filtering. There is a trade-off between response time and signal noise. Decreas-
ing the amount of filtering (e.g. decreasing MEDSZ) will speed up the response time of
the analyzer, but will also increase the noise on the analyte reading. Typically, the filter-
ing is adjusted at the factory to provide a T response of less than 30 seconds (from the
time the sample reaches the cell(s)). T90 is defined as the time required to see 90% of a
step change in the analyte value.
(EQUATION 1.3)
VIIR = V0 + (V1 − V0 ) • wt
(EQUATION 1.4)
wt =1 − ALPHA • e ( (
− BETA • abs Vavg −V0
))
STEP 5: Apply linearization
For some applications, linearization is then applied to the resulting filtered value. If
linearization is turned on, the value is checked against linearization breakpoints. The
response between any two consecutive breakpoints is assumed linear. The slope and
intercept are calculated for each segment and used to calculate the linearized value.
Finally, calibration span and offset corrections are applied to the resulting filtered (and
potentially linearized) value, v (see equation 1.5).
(EQUATION 1.5)
Overview | 1-3
Dispersive Ultraviolet/Visible (DUVV) Spectrometer
The DUVV optical bench operates in the near ultraviolet (200 to 400 nm) and visible (400
to 800 nm) wavelength ranges of the electromagnetic spectrum.
A xenon flash lamp is utilized as the source for the ultraviolet and visible regions. The
collimated beam from the xenon lamp travels through the measuring cell. There are sev-
eral styles of measuring cells for the IPS-4 (see Figure 5-2). When an application requires
sensitivity at shorter wavelengths, UV-grade fused silica windows are used.
Figure 1-2.
Analyzer light path
UV-schematic.
UV Specifications
Wavelength Ranges
Wavelength Range Detector Grating
Wavelength Accuracy
+/- 0.5 nm
Wavelength Resolution
3 nm
Analyzer Range
ppm to 100 %, depending on the application.
Photometric Range
The photometric absorbance range is dependent on wavelength.
Photometric Noise
The photometric noise is dependent on wavelength.
Typically +/- 0.001 AU
Overview | 1-5
Non-Dispersive Infrared (NDIR) Photometer
The NDIR optical bench operates in the infrared (1-5 microns) wavelength range of the
electromagnetic spectrum.
A coiled filament, blackbody radiation steady-state emitter is utilized as the source for the
infrared spectral region. The near-collimated beam from the emitter assembly (containing a
parabolic reflector and CaFೣ lens) travels through the optical bandpass filter (up to six filters
located in the filter wheel assembly) and then through the measuring cell. There are several
styles of measuring cells for the IPS-4 (see Figure 5-2). Typically, sapphire windows and gold
mirrors are used in the cell assembly. Next, a condensing lens focuses the beam onto a two-
stage thermoelectrically cooled lead selenide (PbSe) detector.
CONDENSING
Figure 1- 3
Analyzer light path
NDIR-schematic.
Wavelength Ranges
1-5 microns
Photometric Range
Photometric Noise
Sample System
The sample system is contained within an attached but separate, insulated enclosure
(Oven/Sample System Enclosure). If this enclosure is heated, components within this
enclosure are specified to operate at temperatures of up to 150 °C.
The system components are configurable for both gas and liquid applications. A typi-
cal configuration for the sample system contains the connections for the inlet and outlet
sample lines, the Measuring Cell, an aspirator (gas applications only) to drive the sample
fluid through the system, a Zero Fluid inlet, and a Span Fluid inlet. The system may also
contain pressure sensors/indicators, a filter, and other application-dependent compo-
nents. Information about specific (or optional) components that make up your sample
system – if applicable – are included in a separate Manual Supplement document, located
in the “Supplemental Information” section of this manual or the Documentation Package
shipped with the analyzer.
Electronics
The 512 or 1024 element channel detector is mounted on a board which performs some
signal conditioning. The signal passes to the MCU board which contains the main ana-
lyzer processor. The analyzer internal and external analog and digital inputs/outputs,
along with relays, etc., are handled by the Analog and Relay boards. The main user
interface consists of a keypad/display, also known as the User Interface Panel. Other
boards include the Customer Connection board, a Display Interface board to run the
display, and a Xenon Lamp board. Smaller processors on some of the ancillary boards –
all tied into the main MCU – handle local tasks.
Overview | 1-7
The main functions of the processors are to:
For external communication, RS232, RS485, and Ethernet ports are available. Analog
signals are also available to the customer through configurable 0–20 mA or 4–20 mA out-
puts.
Component selection, redundant heater control, and PCB design assure that under nor-
mal operating conditions within stated environmental specifications and some abnormal
conditions, the system will not produce any sparks or hot spots that could ignite sur-
rounding gases. For this reason, any repairs must not bypass these safety considerations
and must use only recommended components.
Overview | 1-9
This page intentionally left blank.
Analyzer Specifications
Analytes
Up to eight analytes, depending on the application.
Response Linearity
+/- 2% of full scale range
Measurement Accuracy
+/- 2% of full scale (typical – application specific). Accuracy is determined by comparing
analyzer response to a known standard sample after a calibration has been performed.
Repeatability
< 1.0 % of full-scale range (application specific).
Linearity
< 2.0 % of full-scale range.
Stability
+/- 0.5% of full scale range
Specifications | 2-1
Inputs
Two (2) non-isolated analog inputs, configurable as 0–20 mA, 4–20 mA, or 0–5 V.
Two (2) optically isolated discrete DC inputs (minimum voltage: 11.5 V; maximum volt-
age: 24.5 V).
Outputs
256 x 64 pixel vacuum-fluorescent display with multilingual capability.
Up to eight (8) relays (NO, contact rating 100 VA, 240 V maximum), which indicate the
operational status of the analyzer. Each relay provides one SPST (Form A) dry (potential
free) contact.
Sample Pressure
Up to 10,000 kPag (1,450 psig / 100 barg) can be accommodated, depending on the ap-
plication See product markings or customer-specific drawings in the “Supplemental
Information” section of this manual.
For applications with a heated sample system, the Oven will operate at temperatures up
to 150 °C (302 °F).
Liquid Applications:
Using sample stream pressure.
Minimum 5psi differential pressure inlet to outlet.
Minimum pressure:
490–700 kPag (70–100 psig / 4.8–6.9 barg)
Air Quality:
As per ANSI/ISA-S7.0.01 (1996) Quality Standard for Instrument Air.
Nitrogen Requirements:
If using Nೣ :
Use only ultrahigh purity (UHP) Nitrogen.
Electrical Requirements
Power Consumption
Specifications | 2-3
Heated Sample Line
Depending on voltage and length - see customer wiring diagram. Max rating 240Vac
20A.
Supply Voltage
Cell Construction
Gas and Liquid Cells available, in a variety of materials (body can be 316 stainless steel,
Hastelloy C-276, or Monel with Sapphire or UV-grade Quartz windows), to suit a variety
of applications.
Environmental
Ambient Temperature
Humidity
Pollution Degree
Pollution Degree 2
Maximum Altitude
2000 metres
Installation Category II
Enclosure Material
Ingress Protection
Physical Dimensions
Height: 660 mm (26”), analyzer only
Specifications | 2-5
Approvals and Certifications
The IPS-4 Analyzer is certified with regard to electrical safety for Pollution Degree 2
(normally nonconductive environmental pollution occurs with occasional condensa-
tion) and Installation Category II (local level transients, less than those found at power
distribution level). Mains supply voltage fluctuations are not to exceed 10 percent of the
nominal supply voltage.
The Installation & Start-Up chapter contains information about the following topics:
• Where to find safety information in this guide before working with the ana-
lyzer.
• How to uncrate and inspect the analyzer for damage before installing it.
• How to install the mechanical components and sample lines, and perform a
leak check on all of the fittings and connections before powering up the ana-
lyzer.
The installation of the analyzer must be in accordance with all of the customer (end
! user) and local regulatory standards and procedures. There are no operator-serviceable
WARNING components inside the analyzer. Refer service requirements to qualified personnel.
The IPS-4 Analyzer is certified with regard to electrical safety for Pollution Degree
! 2 (normally nonconductive environmental pollution occurs with occasional conden-
WARNING sation) and Installation Category II (local level transients, less than those found at
power distribution level). Mains supply voltage fluctuations are not to exceed 10
percent of the nominal supply voltage.
Upon receiving the analyzer system, remove the shipping crate and check the exterior
of the shelter and/or analyzer for damage. Remove the Optical Bench Assembly from its
box and check it for any physical damage. Replace the Optical Bench Assembly in its box
until it is ready to be installed.
Open the shelter and/or analyzer and check to ensure that all of the components within
are secure and that there is no physical damage. Check that all of the components and
wiring within the Electronics Enclosure and Oven/Sample System Enclosure are secure
and that there is no physical damage.
After the inspection, close and secure all doors with at least one latch. This will keep
! the electronics equipment secure and will prevent damage to the doors and electronic
CAUTION components during installation.
! The analyzer (on backpanel) weighs approximately 129 kg (285 lb)depending on the
WARNING system. Use caution when lifting it from its crate.
• Door latchkeys (two keys are supplied with the analyzer), attached to the
exterior of the analyzer.
• Set of open-end wrenches for fittings and set of metric ball drivers.
• Wire Cutters, strippers, and crimpers.
• Flat blade instrument screwdriver.
• Soft, non-abrasive cloth.
• Wrist Strap (for grounding).
• Detergent-based leak detector (Snoop® or another suitable leak detection
agent is permissible).
• Ethernet cable (supplied by customer), for systems using a permanent Ether-
net connection. Length will vary, depending on application.
Minimum requirements: CAT 5e (maximum 100 m).
Special requirements may be necessary for outdoor analyzer installations,
hazardous locations, and/or lengths greater than 100 m.
Systems using Ethernet also require a ferrite core (supplied by AMETEK).
For systems using a temporary Ethernet connection, a crossover cable must be
used.
• One power-disconnect switch (breaker), rated for at least 250 VAC, 20 A. For
safety reasons during maintenance, this switch allows the main AC power to
be disconnected from the analyzer. Supplied by customer.
• Tubing for the Instrument Air (or N2) and Span fluid lines. Length will vary,
depending on the distance between the analyzer and the supply source. Sup-
plied by the customer.
! Do not mount the Optical Bench Assembly in the Electronics Enclosure until after the
analyzer has been mounted in its designated location.
CAUTION
Figure 3-2 illustrates the mounting hole locations and dimensions, plus necessary clear-
ances on all sides.
The entire analyzer system can be mounted directly on a wall, in a specially designed
cabinet, or in a custom-built shelter. With its IP65 / NEMA 4x rating, the analyzer system
is also suitable for installation in outdoor locations.
The minimum allowable clearances are (if possible, provide extra room to allow sufficient
access for servicing the analyzer). See Figure 3-2a:
Figure 3-1.
BENCH ASSEMBLY BENCH ASSEMBLY
CELLS (APPLICATION SPECIFIC)
XENON FLASH
LAMP ASSEMBLY
OVEN ENCLOSURE
RELAY PWB
MCU PWB
POWER SUPPLY
CAL (SPAN) FLUID IN
DC TERMINAL ASSEMBLY
ANALOG PWB
SAMPLE IN
AC TERMINAL/
MAIN FUSES
SAMPLE OUT
P2, SAMPLE LINE POWER C3, SAMPLE LINE HEAT TRACER POWER
Outdoor Installations:
In most cases, the analyzer must be shielded from harsh environmental elements such as
wind, rain, and sun. In these cases a rain shelter or sun shade is required.
Also, proper water drainage must be considered in the mounting location to allow for
suitable and safe working conditions for service personnel.
Ensure there is no power being supplied to the analyzer while installing the Optical
! Bench Assembly.
WARNING
Refer to Figure 3-3a for the locations of the Optical Bench Assembly, boards, cable con-
nectors, and other components discussed in this procedure.
1. Using the locator pins on the UV Optical Bench Assembly, position the Bench
Assembly to the upper Window Housing Assembly (Figure 3-3a). Mount the
clamp on the interface and secure the assembly in place.
3. Connect the RTD cable connector from the Optical Bench to the in-line con-
nector (labelled “CON2”) that leads to J402 on the Analog board.
4. Connect the DC cable connector from the Xenon Flash Lamp Assembly to J2
on the Xenon Lamp board, located near the top of the Electronics Enclosure.
[ 12.7] 6 PLS
NOTE
46.25
Analyzer mounting
[1174.8]
22.00
HEATED SAMPLE LINE
[558.8] SLEEVE SHOWN
(OPTIONAL)
SPECIFIC
27.81 [706.3]
BOTTOM VIEW
58.74 [1491.9]
Figure 3-2b.
Analyzer mounting,
door clearance.
6. If the lamp assembly is provided with a purge fitting, connect the tube from
the air inlet on the bottom of the enclosure to the fitting on the lamp.
Clamp the ribbon cable using the clamp provided, leaving enough length on
each end to avoid pulling or binding.
Do not attempt to to adjust the optical bench. The optical bench has been factory
! aligned to the mounted cell. See additional information on the dual bench application
WARNING in the “Overview section of this manual.
Ensure there is no power being supplied to the analyzer while installing the Optical
! Bench Assembly.
WARNING
Refer to Figure 3-3b for the locations of the Optical Bench Assembly, boards, cable con-
nectors, and other components discussed in this procedure.
1. Using the locator pins on the NDIR Optical Bench Assembly, position the
Bench Assembly to the lower Window Housing Assembly (Figure 3-3b).
Mount the clamp on the interface and secure the assembly in place.
2. Connect the three pin DC cable connector from DC supply to the mating con-
nector on the underside of the Optical Bench Assembly.
3. Connect the ribbon cable from the Optical Bench to connector P2 on the De-
tector Interface board, mounted to the MCU board.
4. Clamp the ribbon cable using the clamp provided, leaving enough length on
each end to avoid pulling or binding.
5. Connect the purge supply tubing to the purge fitting on the Optical Bench
Assembly.
XENON FLASH
LAMB ASSEMBLY
CLAMP TO SECURE
OPTICAL BENCH
OPTICAL BENCH TO ANALYZER
ASSEMBLY
Figure 3-3b.
Optical Bench-IR
Assembly.
CLAMP, TO SECURE
RIBBON CABLE OPTICAL BENCH
CONNECTS TO P2 TO ANALYZER
OF DETECTOR INTERFACE
BOARD
PURGE FITTING
If your application uses any of these or other special-requirement sample system com-
ponents, the customer-specific drawings and Manual Supplements – that describe and
illustrate installation, operation (if applicable), and maintenance (if applicable) – will be
included in the “Supplemental Information” section of this manual and the Documenta-
tion Package shipped with the analyzer.
If an aspirator is used to drive the sample fluid (typical in gas applications), the
! Sample and Vent Lines must be minimum 3/8” O.D. to avoid back pressure. This is of
particular importance for the Vent Lines.
NOTE
For Vent Lines longer than 15 metres (50 feet), 1/2” O.D. tubing is recommended.
This procedure applies to both unheated and heated Sample and Vent Lines. Refer to ap-
propriate customer-specific Sample/Vent Line Installation drawings (if lines supplied by
AMETEK) for additional information. These drawings are located in the “Supplemental
Information” section of this manual and the Documentation Package shipped with the
analyzer.
! To avoid damaging the Sample or Vent Lines, use two wrenches to ensure that the fit-
ting body does not turn on the line.
CAUTION
• When laying out the lines, ensure the ends with the electrical leads are at the
location of the external power supply.
• Install heat shrink around the Sample and Vent Line wiring and terminate the
Sample and Vent Line wiring to the external power source.
Refer to power termination instructions supplied with the Sample and Vent
Lines for details.
Ensure there is no power being supplied to the Sample and Vent Line temperature-con-
! trol zone circuits or external power source. Do not apply power until the entire system
has been installed, leak checked, and the analyzer is ready for power-up.
WARNING
Route the Sample Line from the analyzer to the sample probe (or process tap)
and the Vent Line from the analyzer to the return/vent location.
Ensure there are no loops, sags, or other traps in the Sample and Vent Lines.
Provide support where needed.
4. Connect the Vent Line from the analyzer to the return/vent location.
Pressure at the analyzer inlet connector must not exceed the ratings marked on the
! inlet and on the system drawings.
NOTE
Instrument Air or Nitrogen can be used as the aspirator/pneumatic valve drive fluid and
to zero the analyzer. This procedure assumes Instrument Air is being used for these func-
tions. The procedure is the same, regardless of which medium is being used.
The customer supplies the Instrument Air/Zero fluid line. The tubing must be 1/4” O.D.
316 stainless steel tubing with appropriate 1/4” tube connectors.
• The Instrument Air must meet the ANSI/ISA-7.0.01 (1996) specifications at all
times. If the Instrument Air cannot meet these requirements, an inlet filter
system must be installed.
If using N2:
1. Route the line from the Instrument Air/Zero fluid supply to the analyzer.
2. Before connecting it to the analyzer, blow down (purge) the line to remove
any liquids or particulate that may be present in it.
3. Connect the line to the Instrument Air/N2 supply and to the analyzer. The
plumbing for purge systems and the Instrument Air/N2 feed through the ana-
lyzer manifold is already made by AMETEK.
For applications that require spanning, the Span fluid line is supplied by the customer.
Refer to the customer-specific Flow Diagram in the “Supplemental Information” section
of this manual or the Documentation Package shipped with the analyzer.
1. Route the line from the calibration fluid supply to the analyzer.
2. Before connecting it to the analyzer, blow down (purge) the line to remove
any liquids or particulate that may be present in it.
3. Connect the line to the calibration fluid supply and to the analyzer. The
plumbing for the calibration fluid feed through the analyzer manifold is al-
ready made by AMETEK.
For AC electrical supply cable and conduit requirements, refer to – and comply with –
! local wiring regulations and electrical codes for the area.
WARNING
Electrical Connections
Termination points for power and all signals are located in the Electronics enclosures.
Four metric M25 straight thread conduit entries are provided on the bottom of the enclo-
sure for Analyzer power, analog signal, discete signal, and optional HAG probe power
and HAG probe RTD. NPT conduit adapters are available, or the installer may resize the
holes as needed.
All end-user input/output signal and alarm relay contact connections are made at the
Customer Connection board. AC power connections are made at the Terminal Block 1
and the Ground connection is made at the ground stud on the bottom of the Electronics
enclosure, near the power entry gland.
Four metric M25 straight thread conduit entries are provided on the Sample Line enclo-
sure. (Two on the right, two on the bottom and one on the left side of the enclosure. The
conduit entries are for sample line power, sample line trace power, and sample line RTD
#1, #2 wiring. Note that unused holes must be plugged with M25 or M32 hole plugs ap-
plicable to the enclosure. Do not use NPT hole plugs with thread adapters. Only prop-
erly certified conduit connectors must be used.
Figure 3-5 illustrates the general locations of all customer termination points, while
the Customer Connections drawing specifically details the signal, relay contact,
and AC power connection points. This customer-specific drawing is located in the
! “Supplemental Information” section of this manual and in the Documentation Pack-
NOTE age shipped with the analyzer.
1. Open the Electronics Enclosure and open all of the main fuse (F1, F2, F3, F4)
at TB1 11, 12, 13, 14.
2. Route the analog input/output signal wiring and alarm relay contact con-
ductors through one of the Signal cable entry ports and into the Electronics
Enclosure.
Terminate the AC power wires wires at TB1 as detailed in the Customer Con-
nections drawing.
3. Route the AC power conductors through the Power cable entry port and into
the Electronics Enclosure.
! Do not apply power to the system until after all of the wiring has been installed, con-
nected, and verified, and only if the purging system (if used) is ready for operation.
NOTE
For systems using a permanent Ethernet connection, a ferrite core (Part Number
! 301-1044, supplied by AMETEK) must be installed on the Ethernet cable (supplied by
NOTE customer), inside the Electronics Enclosure.
For details about setting up Ethernet parameters, refer to “Ethernet Network Screen” in
Chapter 4.
1. Route the Ethernet cable through one of the Signal cable entry ports and into
the Electronics Enclosure. Pull through enough cable to loop around the fer-
rite.
2. Insert the Ethernet cable through the ferrite core and loop the cable once
around the ferrite, leaving enough cable length to reach the Ethernet port at
J18 on the Customer Connection board. Adjust the cable and ferrite so the
ferrite is as close as possible to the Customer Connection board.
3. Route the cable to the network connection in the plant and connect it.
Before powering up the analyzer, test the area for flammable gases. If a flammable
! gas atmosphere is present, do not apply power to the analyzer or any alternate power
sources that supply power to the analyzer components. Proceed only when the area is
CAUTION
found to be safe. DO NOT LEAVE POWER CONNECTED TO THE ANALYZER
IF THE DOORS ARE OPEN UNLESS YOU ARE CONTINUOUSLY SAMPLING
THE AREA FOR FLAMMABLE GASES AND ATTENDING TO THE ANALYZ-
ER.
When the analyzer doors are open, take appropriate precautions to avoid electrical
shock. Hazardous voltages are present inside.
Do not apply main AC power to the analyzer if the Xenon Flash Lamp Assembly elec-
! trical connections have not been made.
NOTE
The analyzer is shipped from the factory without a password (security is disabled).
Upon powering up the analyzer for the first time, all configurable parameters can be
! changed until a password is entered in the Change Password function (Login/Set
NOTE
Password screen, SYSTEM SETTINGS menu).
Setting up security for the analyzer is necessary to ensure analyzer settings are not
1. Close the Main AC Power and DC Power fuses. Refer to the Fuse Legend
(at the back of the Electronics Enclosure) for the location of these fuses. If a
power-disconnect switch (breaker) was not installed, main AC power will be
applied to the analyzer.
Open the Bench Heater and Oven Heater fuses. Refer to the Fuse Legend
(at the back of the Electronics Enclosure) for the location of these fuses. If a
power-disconnect switch (breaker) was not installed, main AC power will be
applied to the analyzer.
Do not power up the temperature-control zone circuitry (Bench Heater fuse and
! Oven Heater fuse, if Oven/Sample System Enclosure is heated) until after the sample
NOTE system leak check has been performed and verified, as described later in this procedure.
3. After applying AC main power to the analyzer, the message “System Start-
ing” is displayed momentarily on the top line of the User Interface Panel.
Following this, the AMETEK Process Instruments start-up screen will appear.
This screen also displays the software version.
After the start-up screen, the system defaults to the HOME screen which
allows you to view the current status of up to nine factory-configured param-
eters. These parameters are displayed on the left side of this screen. The right
side of this screen displays the current operational status of the analyzer. De-
scriptions of these parameters and states are described under “HOME Menu
Screen” and “System Status Messages (HOME Screen),” in Chapter 4.
! If these screens do not appear, check the AC wiring to ensure it is properly connected.
WARNING
To check the AC wiring, take all necessary safety precautions to power down
the analyzer and open the Main AC Power and DC Power fuses.
> Heater 19 °C
Bench 24 °C
Cell 23 °C
Electronics 24 °C
Figure 3-6. Manifold Pressure 85 psia
STATUS menu Aspirator Pressure 3 psia
(example screen). HOME ALARMS STATUS SETUP
a. Press the STATUS soft key to view the STATUS screen (Figure 3-6).
Check and record the current operating temperature readings. This will
help you determine if the signal wiring is properly terminated and if all of
the circuitry is operating properly.
Pay close attention to the temperatures of the (Oven) Heater (if used), (Opti-
cal) Bench, and Cell (if used) readings. If these temperatures do not increase,
check the wiring for proper connections. If the wiring is properly connected,
it may indicate a shorted RTD for that temperature zone.
! The Oven Heater temperature parameter is available only for applications with a
heated Oven.
NOTE
b. Check the pressure of the pilot valves – Manifold block and Zero fluid (if
used) pressure. Adjust the Actuator/Zero pressure regulator to obtain a
minimum pressure of 490 kPag (70 psig / 4.8 barg).
c. Check the Aspirator Pressure (gas applications only) from the STATUS
screen; the pressure should be adjusted to the value stated on the system
drawings, or enough to see a flow on the Flow Indicator (or meter), if used,
or to achieve an acceptable response time. Typically, a good response time
is less than 30 seconds to T90 when the sample system is clean and operat-
ing at peak efficiency – response time may vary due to Sample Line length.
The response is application-specific and will vary, depending on the sample
system.
a. Press the SETUP soft key to view the SETUP menu screen. Scroll down to
select Current Outputs and press ENTER to display the Current Outputs
1 thru 4 screen. Scroll down to select Current Outputs 1 screen (Figure
3-7).
Review the Customer Connections drawing and check the wiring termina-
tions from the main AC power supply source to the analyzer.
Take all necessary safety precautions and reapply power to the analyzer by
closing the Main AC Power and DC Power fuses. Check the User Interface
Panel again to see if these screens appear. If so, continue with the next step.
4. Perform a leak (pressure) check to ensure there are no leaks in the system,
then return to this procedure and complete the remaining start-up and verifi-
cation tests.
To perform the leak check, follow the procedure under “Sample System Leak
Check,” later in this chapter.
5. After the leak check has been performed and has passed, close the Bench
Heater fuse and Oven Heater fuse (if Oven/Sample System Enclosure is
heated) to apply power to the Measuring Cell and Oven Heater temperature-
control zone circuitry. Ensure that the sample system door has been closed
and latched.
No alarms will be displayed on the HOME screen or entered into the Alarm Log
! until AFTER the system has left the Not Ready state for the first time after power-
NOTE up, which occurs after the entire sample system has stabilized (2–3 hours) and reached
normal operating temperature.
6. While performing the remaining start-up tests, close the analyzer doors to
allow the sample system to warm up to normal operating temperature (neces-
sary only if Oven/Sample System Enclosure is heated).
The Oven and the rest of the sample system will reach normal operating temperature
in approximately 2 hours, however (application-specific) analyzers may take up to
! another hour to stabilize.
NOTE
To minimize the time it takes for the entire sample system to warm up to normal op-
erating temperature, be sure to close the doors after performing the checks that require
them to be open.
7. While working from the User Interface Panel, perform the following checks
to ensure the analyzer temperature-control zone circuitry is operating prop-
erly and the system pressures are set appropriately. Record the temperature
value for each zone and the system pressures to use as a reference later while
comparing the values.
Current Out 1
Source Analyte 1
Full 1000.0
Zero 0.0
Figure 3-7. Type 4-20 mA
Current Output Setup Over Range Disabled
screen (Current Out 1). HOME ALARMS STATUS SETUP
a. Press the SETUP soft key to view the SETUP menu screen. Scroll down to
select Relays and press ENTER to display the Relay screen. Scroll down to
select Relay 1 and press ENTER to display the Relay 1 screen (Figure 3-8).
Relay 1
Function Disable
Source None
Value 0
Figure 3-8. Delay 0 Sec
Relay Setup screen
(Relay 1). HOME ALARMS STATUS SETUP
Check the Relay 1 Function (operational status setting for this relay), Source
(species), Value, and Delay (zero point) settings/values and compare them
to the settings/values listed on the analyzer documentation shipped with
the analyzer to ensure they correspond.
10. If using an Ethernet connection to communicate with the analyzer, set up the
Ethernet parameters as described in the section “Ethernet Network Screen” in
Chapter 4.
Temporarily open the Electronics Enclosure door and listen for an intermit-
tent clicking sound, which indicates the Xenon Flash Lamp is on (flashing).
If there is no clicking sound, and the analyzer is in a state other than the Not
Ready state, this can indicate problems with the wiring. If necessary, power
down the analyzer and check all associated wiring.
12. If the analyzer sample system cools down to below its normal operating tem-
perature and returns to the Not Ready state, wait until it warms up enough
and the analyzer leaves the Not Ready state.
Check the User Interface Panel for current alarms. To view alarms that have
been detected, press the ALARMS soft key to view the ALARMS screen. From
here, use the UP/DOWN arrow key to select Current Alarms and press ENTER
again to view any active alarms (on the Current Alarms screen, Figure 3-9).
If any alarms exist, take appropriate action to correct the alarms before con-
tinuing. For more information about alarms and corrective action to take to
correct and reset alarms, refer to “Alarm Conditions and Corrective Action” in
Chapter 5.
If no alarms exist (Figure 3-10), or after all alarm conditions have been cleared
(reset), recheck the temperature of each zone from the
STATUS screen.
Compare the values to the values recorded earlier to ensure they have in-
creased, have reached their normal operating temperatures, and have stabi-
lized.
13. Open the Vent Line isolation valve (on the Sample Probe, process tap, or
inside the Oven/Sample System Enclosure). Open the Sample Line isolation
valve (on the Sample Probe, process tap, or inside the Oven/Sample System
Enclosure) to allow sample fluid into the sample system.
Cell Temperature
Heater Temperature
Figure 3-9.
Active alarms displayed
(example), Current
Alarms screen. HOME ALARMS STATUS SETUP
Figure 3-10.
No active alarms,
Current Alarms screen. HOME ALARMS STATUS SETUP
14. Zero the analyzer as described under “Manually Zeroing the Analyzer Sam-
ple System,” later in this chapter. From the HOME screen (Figure 3-11), check
the response time of the analytes. Typically, a good response time is less than
30 seconds to T90 when the sample system is clean and operating at peak ef-
ficiency. Response time may vary due to Sample Line length.
Record the initial sample response time in a log book, as described under “Re-
cording Initial Readings,” later in this chapter.
Heater 21 °C Measure
Figure 3-11.
Bench 20 °C
HOME screen Cell 23 °C
(example). HOME ALARMS STATUS SETUP
The start-up and verification checks are complete and the analyzer is ready for normal
operation.
Problem Encountered
Status/error code conditions for current alarms are displayed on the
User Interface Panel.
Corrective Action
To view current alarms, press the ALARMS soft key to view the
ALARMs screen. For information about viewing current alarms, refer
to “Alarm Notification (HOME Screen)” and “ALARMS Menu Screen”
in Chapter 4. For detailed information about these alarms, refer to the
alarm conditions and corrective action in the “Maintenance & Trouble-
shooting” chapter of this manual.
Problem Encountered
The Xenon Flash Lamp is not operating (intermittent clicking sound is
not heard).
Corrective Action
Check the system state on the right side of the HOME screen. Note
that the Xenon Flash Lamp will not operate when the system is in the
Not Ready state.
Problem Encountered
The LEDs on the Relay, Customer Connection, or MCU boards are
not on.
Corrective Action
Take appropriate safety precautions and check for proper I/O wiring
connections.
Problem Encountered
Temperature zone values do not increase.
Corrective Action
The over-temperature circuits may have been tripped; press the Reset
Switch (S501) on the Relay board or the Reset Switch on the Optical
Bench. (Refer to Figure 3-3 of this manual.
However, fittings can loosen during transport. Also, the Sample, Vent, Instrument Air (or
N2), and Calibration (Span, if used) Fluid lines are installed on-site. Therefore, it is impor-
tant that a leak check is performed on the entire sample system prior to commissioning
the analyzer for operation for the first time, or following the replacement of any lines/fit-
tings or Measuring Cell parts.
If the analyzer is configured for a heated sample system, perform the sample system leak
check prior to applying power to the Oven/Sample System Enclosure circuitry (must not
be hot).
Preventing leaks in the sample handling system is critical to ensure proper analyzer
! operation. If sample fluid migrates into the Optical Bench Assembly or Reflector Block
due to a leak in the Measuring Cell Assembly, the optics will become damaged and will
CAUTION
likely require replacement. Most leaks are preventable with the regular replacement of
the Measuring Cell o-rings.
Do not use a leak detection fluid on hot fittings. If the analyzer system is at operating
temperature, the temperature-control zone circuitry fuses must be opened to allow the
temperature-control zones to cool before using a leak detection fluid on the fittings.
Refer to the system Analyzer flow diagram in the Ametek supplied drawing package
(403900XXX-3).
Identify the location of the pressure sources that can be applied for leak testing.
! When pressurizing the system, do not exceed the rated pressure for that branch of the
sample system.
NOTE
If a pressure sensor and isolation valves are not available, pressurize the system and bub-
ble-test all joints with a soap solution. (eg. Snoop or the fluid specific to the Analyzer.)
For heated systems, all temperature zones must be at their operating set points before
performing a Zero. If necessary, view the temperature zones to verify this.
Pressing CANCEL at any time during this procedure will abort the procedure and
! return the software to the HOME screen.
CAUTION
1. Ensure that the Zero fluid cylinder source is connected and turned on.
2. Press the SETUP soft key to display the SETUP menu. Select the Calibration
Control sub-menu and press enter. Scroll down to the Calibration sub-menu
and press enter to display the Calibration screen.
3. Use the DOWN arrow key to select Start zero calibration and press ENTER to
purge the sample system with Zero fluid. The message “Please wait ...” will
appear while the analyzer performs the calibration function.
Typically, the IPS-4 Analyzer controls the zero flow rate; no adjustments are required.
!
NOTE If the customer provides the sample system zero, run the manual zero at a flow rate of
2.5 L/minute (0.08 scfm).
4. From the Calibration screen, select Start measure and press ENTER to return
the analyzer to Measure mode.
Typically, the IPS-4 Analyzer controls the zero (and span, if used) flow rate; no ad-
! justments are required.
NOTE
If the customer provides the sample system zero, run the manual zero at a flow rate of
2.5 L/minute (0.08 scfm). Since excess air flow into the analyzer sample system will
cool the Measuring Cell and Oven/Sample System Enclosure, the optimum Zero fluid
flow rate is the lowest consistent with a good zero.
A good zero can be defined as a state where the concentration outputs are stable and fur-
ther increases in Zero fluid flow rate do not reduce the concentrations observed on the
analyzer User Interface Panel. A mid-scale flow indicator (or meter) should be sufficient
for this purpose.
While the sample system is warming up, or when the analyzer is in standby mode (Not
Ready state), zero flow may be desired. During these conditions, it is important that the
sample system is kept free of contamination by the sample fluid.
For additional information, refer to the Sample System Manual Supplement in the “Sup-
plemental Information” section of this manual or in the Documentation Package shipped
with the analyzer.
To avoid damaging the pressure transducer, do not set the pressure higher than 105
! kPag (15 psig).
CAUTION
The analyzer pressure – measured by the pressure transducer – can also be used to set
the sample flow rate. This is accomplished by first measuring the sample stream pressure
with no aspiration (Drive Air Isolation Valve closed) and then adjusting the Aspirator
Pressure Regulator with the Drive Air Isolation Valve open to obtain a pressure of 5–10
Sample response time can also be used as a parameter for setting the Aspirator Pressure.
Response time is determined by switching the analyzer from Zero Gas Flush mode to
Measure mode and recording the time it takes the analyzer to display the first reading
after the switch.
2. Press the SETUP soft key to display the SETUP menu and then press the
DOWN arrow key until the Calibration sub-menu item is displayed and se-
lected.
3. Manually zero the analyzer. To do this, press ENTER to display the Calibration
screen. Use the DOWN arrow key to select Start zero calibration and press
ENTER. The message “Please wait ...” will appear while the analyzer initializes
and performs the calibration function.
Run the manual zero for 10 minutes. For more information about setting the
Zero fluid flow rate, refer to “Adjusting the Zero/Span Fluid Flow Rate” in this
chapter.
4. From the Calibration screen, select Start measure and press ENTER to return
the analyzer to Measure mode.
5. Press the HOME soft key to return to the HOME screen and observe the ana-
lyte output concentration(s) again.
If the sample response time is adequate (first response after a Zero is less than
30 seconds to T90), no further adjustment is necessary.
! The sample response time may vary, depending on Sample Line length.
NOTE
AMETEK recommends that you observe and record this information every month to ob-
tain a history. From these recordings, you will be able to determine if there are problems
with the analyzer, such as increasing sample response. These problems typically indicate
that maintenance is required.
Keep the log book in a safe location until you need to review it for patterns of a longer
sample response time.
The analyzer sample response time can be used for two functions. One function is to
help you set the sample fluid flow rate, as discussed earlier in this chapter, under “Set-
ting Sample Response Time.” The other function is to help you determine if the typical
response time is increasing, which can indicate a plug in the sample system which, in
turn, can indicate maintenance is required.
2. After the zero is complete, view the HOME screen and observe the analyte
output concentration changes. Observe and record the time it takes the
analyzer to display the first reading after the analyzer System State switches
from “Zero” to “Measure”.
! It is important to record the response time only after performing a manual Zero func-
tion to simulate the same conditions during the original recordings.
NOTE
3. If the sample response time is adequate (first reading after a Zero function
and after the analyzer System State switches from “Zero” to “Measure” is less
! The sample response time may vary, depending on Sample Line length.
NOTE
Analyzer Configuration
The Parameter List (Factory Settings) file is shipped with each analyzer. These configu-
ration sheets list all of the settings applicable to the factory analyzer (to meet customer
requirements). The files are provided on a CD and located in the “Supplemental In-
formation” section of this manual and in the Documentation Package shipped with the
analyzer.
If any changes are made to the original configuration, be sure to record the changes for
later reference. If the MCU board is ever replaced, this data must be re-entered in order
to override the factory defaults.
If necessary, refer to the Parameter List (Factory Settings) to revert the analyzer back to
its original settings.
The Controller / User Interface chapter contains information about the following topics:
• What the various keys that makes up the keypad do, and how to use them
to work from the analyzer User Interface Panel.
Before working from the User Interface Panel, read the section titled “Introduction to
! the User Interface Panel,” in the following pages. This section discusses the functions
NOTE that can be performed by pressing the keypad keys.
This section also discusses how to enter Edit mode, and how to quickly return to the
HOME screen from anywhere in the software.
! Pressing ENTER after making any changes will save the configuration data in non-
volatile memory – the old values will be overwritten.
NOTE
Main Keypad
Figure 4-1.
User Interface Panel
layout.
• The Vacuum Fluorescent Display consists of a 256 by 64 dot graphic display. The
information displayed depends on the mode of operation you are working in.
- The bottom line displays the main menu items which can be accessed by
pressing the corresponding blank keypad square (on the Soft Keys Keypad)
directly below it.
For information about the main menu items, and the sub-menu items, refer
to “Working in the Main Menus” in this chapter.
- The other lines can display the output results (concentrations, temperatures,
pressures), alarms (current and historical), system status, system information,
and various configurable parameters that allow you to set up the analyzer
to perform various functions (calibration, communication, etc.).
• The Main Keypad allows you to change information from any of the editable
parameters. The keypad consists of:
0–9 Numeric keys 0–9 are used to enter numerical data or as part
of a password.
X The CANCEL key allows you to discard any changes you make
to operating parameters while in Edit mode, or to back out
of previous menu levels one level at a time until the HOME
screen is displayed. This key will also allow you to answer “No”
to software prompts (e.g., “Restore settings. Are you sure?”).
8 The ENTER key allows you to view sub-menu items, enter Edit
mode, select parameters and/or their options, initiate analyzer
functions (e.g., calibration), and to exit Edit mode and save any
changes that have been made. This key will also allow you to
answer “Yes” to software prompts (e.g., “Restore settings.
Are you sure?”).
ñ ò The UP and DOWN arrow keys are used to scroll up or down to
view sub-menu items or parameters that are not currently on
the screen, as indicated by a DOWN arrow icon on the lower-
left side of the screen. These keys are also used to change
a selected character – to the next or previous alphanumeric
character – while working in Edit mode.
Character Description
u This character (cursor) beside a menu or sub-menu item indi-
cates the item is selected, and that it has additional sub-menus
or parameters to view. This character beside a parameter
indicates the parameter can be modified.
When both of these characters (located in the lower-left corner)
are displayed, there are other options to be viewed above and
below the items currently being displayed. To scroll through
other options, press the UP or DOWN arrow key.
The analyzer is shipped from the factory without a password (security is disabled).
! Upon powering up the analyzer for the first time, all configurable parameters can be
changed until a password is entered in the Change Password function (Login/Set
NOTE
Password screen, SYSTEM SETTINGS menu).
Setting up security for the analyzer is necessary to ensure analyzer settings are not
inadvertently changed while working from the User Interface Panel. If changing the
password, be sure to record and store the password in a safe place for reference.
• If security is enabled (if a password has been entered in the Change Password
function – Login/Set Password screen, SYSTEM SETTINGS menu), the software will
revert to Operator (read-only) mode after 5 minutes of no keypad activity.
• If security is enabled but you are not logged in, the message “Password required”
is displayed when you press ENTER to initiate analyzer functions (such as calibration
functions) or go into Edit mode to change the configuration settings.
• On all screens, a cursor (‘>’ or ‘u’ ) is displayed beside the current item; to view
other related items that are not currently displayed (if available), press the UP or DOWN
arrow keys.
The cursor ‘>’ indicates the item is read-only. The ‘u’ cursor indicates the item is
editable.
• When the ‘u’ cursor is next to a sub-menu item press ENTER to view parameters
under that item.
• When in Edit mode (for numeric entries only) the cursor ( ‘_’ ) is displayed on or
to the right of the selected character.
• You can view all menus and sub-menus – and the current settings or values of all
parameters or functions on each screen – without a password.
• The display will revert to the HOME screen if no keys are pressed within 60 sec-
onds.
• Active alarm names will flash on the HOME screen only. To stop the flashing,
press any key once. The alarm will begin flashing again after 20 seconds of no keypad
activity, unless the alarm is reset.
• To return to the HOME screen from any other screen (except when in Edit mode),
press the HOME soft key.
This list includes variables to set up Communications, Units of Measures, Language, Tim-
ers for analyzer operations, Inputs, Outputs, Relays and Solenoids, Alarm Limits, Lamp
Control, Heater Setpoints and Control, Filtering, and various other parameters. It is also
continuously updated by the main processor with (read-only) variables such as Relay/
Solenoid/Valve Status, Current Outputs, and other parameters.
A subset of these parameters may be changed from the User Interface Panel. The se-
quence of analyzer operations is defined by a file called “[Link].” The proces-
sor continuously communicates with the “[Link]” and “[Link]” files to
define what function the analyzer should be performing at any given time. These files
also control the setting of valves and other devices.
Examples of “states” include Zero, Span, and Measure. These and other states are de-
scribed in detail under “System Status Messages (HOME Screen),” later in this chapter.
Generally, the analyzer will sequence through a series of factory-set, predefined states.
However, through settings that can be changed from the User Interface Panel, you can
vary the interval between analyzer states, manually change the analyzer to another state,
or enable or disable certain states.
The analyzer will leave its normal sequence of states if there is an alarm condition, such
as a temperature or pressure out of range, and will return to the first state in the se-
quence once the offending condition has been cleared (reset).
The bottom line displays the main menu items which can be accessed by pressing the
corresponding keypad square (soft key) directly below each item.
The right side of this screen displays the current operational status of the analyzer. De-
scriptions of these states are described under “System Status Messages (HOME Screen),”
in the following pages. Also, new alarms and warnings that have been set are displayed
on this screen. Refer to “Alarm and Warning Notification (HOME Screen),” later in this
section, for more information.
When you press ENTER, the software goes into Edit mode, allowing you to set up the
parameters that are displayed. Items that can be displayed here include:
Heater (optional)
Displays the real-time temperature (°C) of the Oven Heater, as measured by the Oven
Heater RTD (sealed inside the Oven Heater). This parameter is available only for applica-
tions with a heated Oven.
Bench
Displays the real-time temperature (°C) of the Optical Bench, as measured by the Optical
Bench RTD.
Measure
u Heater 21 °C
Figure 4-2.
Bench 20 °C
HOME screen Cell 23 °C
(example). HOME ALARMS STATUS SETUP
Manifold Pressure
Displays the real-time pressure (psia, or other unit) measured at the Manifold Block. The
manifold pressure is used for valve actuation and for zero flush (if applicable).
Aspirator Pressure
Displays the real-time pressure (psia, or other unit) being used to drive the Aspirator
system (gas applications only).
Cell Pressure
Displays the real-time pressure (psia, or other unit) being measured in the Measuring
Cell.
Probe Temperature
Displays temperature @ sample probe.
Not Ready
During this state, the analyzer is not measuring sample fluid due to alarm conditions
that must be resolved before it will resume normal sampling operation.
Sample Flush
Indicates the analyzer is allowing sample fluid into the sample system. There will be a
short delay before it reaches the Measuring Cell, where the analyzer begins measuring
the sample.
Measure
Indicates the analyzer is measuring analyte values.
Span
This occurs following the Span Fluid Flush and indicates the analyzer is being spanned.
When complete, the span fluid valve is closed.
Zero
Indicates the analyzer is being zeroed. During this function, the zero fluid valve remains
open until the Zero is complete. The Zero consists of an average reading of a factory-set
number of readings (usually 10 to 90; default = 60). The valve has to be closed by chang-
ing the system state to Measure or Span mode.
Span Hold
Indicates the analyzer is measuring concentration with the span fluid valve open. To
leave this state, the analyzer must be manually changed to another state from the Cali-
bration screen.
Zero Hold
Indicates the analyzer is measuring concentration with the zero fluid valve open. To
leave this state, the analyzer must be manually changed to another state from the Cali-
bration screen.
When an alarm or warning is set by the analyzer, the message will flash on the HOME
screen only. The information on the HOME screen will be replaced by a reverse (white)
screen with the alarm name in black text. It will then switch to a black screen with white
text and will continue to flash until the alarm conditions that triggered the alarm(s) are
reset.
For more information about active and historical alarms and warnings, refer to “Viewing
Current Alarms and Warnings” and “Viewing Logged Alarms or Warnings” in the fol-
lowing pages.
Alarms and warnings that have been reset (no longer active) are stored in the Alarm Log
and Warning Log, accessed from the ALARMS AND WARNINGS screen. Logged (or
historical) alarms and warnings – which occupy two lines – include the time and date
each alarm was set and reset. The alarm or warning at the top of the list is most recent;
the alarm or warning at the bottom is the oldest.
Figure 4-3.
ALARMS and
WARNINGS screen. HOME ALARMS STATUS SETUP
All active alarms and warnings are displayed on the Current Alarms and Warnings
screen. Each line will occupy one alarm or warning description and up to six active
alarms or warnings can be displayed on the screen at once. If more than six are active,
use the DOWN arrow key to scroll down the list. The alarm or warning at the top of the
list is most recent; the alarm or warning at the bottom is the oldest.
To view active conditions, select Current Alarms and Warnings and press ENTER. If there
are alarms active, the alarm description is displayed (Figure 4-4). If no alarms are active,
“No alarms” is displayed (Figure 4-5).
For a complete list of alarms and warnings generated by the analyzer, refer to the
! “Maintenance & Troubleshooting” chapter.
NOTE
Cell Temperature
Heater Temperature
Figure 4-4.
Active alarms displayed,
Current Alarms (example
screen). HOME ALARMS STATUS SETUP
No alarms
Figure 4-5.
No active alarms,
Current Alarms screen. HOME ALARMS STATUS SETUP
Alarms that have been reset (no longer active) are stored in the Alarm Log, accessed
from the ALARMS screen. Each logged (or historical) alarm – which occupies two lines
– includes the time and date it was set and reset. The logged alarm at the top of the list is
most recently reset alarm; the alarm at the bottom is the oldest. The Alarm Log will be
cleared when the analyzer is powered off.
To view the history of alarms or warnings that have been reset, select Alarm Log or
Warnings Log and press ENTER. If there is history of alarms or warnings that have been
reset, the descriptions are displayed (Figure 4-6 and 7). If there is no history, “Log empty”
is displayed (Figure 4-8 and 9).
Alarm Log
05/13/2005 [Link] 05/13/2005 [Link]
Cell Temperature
Figure 4-6. 05/13/2005 [Link] 05/13/2005 [Link]
History of alarms Heater Temperature
displayed, Alarm Log
(example screen). HOME ALARMS STATUS SETUP
Warnings Log
05/13/2005 [Link] 05/13/2005 [Link]
Intensity to low during zero calibration
Figure 4-7. 05/13/2005 [Link] 05/13/2005 [Link]
History of warnings Zero calibration failed
displayed, Warning Log
(example screen). HOME ALARMS STATUS SETUP
Alarm Log
Log empty
Figure 4-8.
No history of alarms,
Alarm Log screen. HOME ALARMS STATUS SETUP
Warnings Log
Log empty
Figure 4-9.
No history of
warnings, Warnings
Log screen. HOME ALARMS STATUS SETUP
> Heater 19 °C
Bench 24 °C
Cell 23 °C
Electronics 24 °C
Figure 4-10. Manifold Pressure 85 psia
STATUS menu screen Aspirator Pressure 3 psia
(example). HOME ALARMS STATUS SETUP
Heater (optional)
Displays the real-time temperature (°C) of the Oven Heater, as measured by the Oven
Heater RTD (sealed inside the Oven Heater). This parameter is available only for applica-
tions with a heated Oven.
Bench
Displays the real-time temperature (°C) of the Optical Bench, as measured by the Optical
Bench RTD.
Electronics
Displays the real-time temperature (°C) inside the Electronics Enclosure. This tempera-
ture is measured by a sensor on the MCU board.
Manifold Pressure
Displays the real-time pressure (psia, or other unit) being measured in the Manifold
Block, for the Manifold and Zero fluid (if used).
Aspirator Pressure
Displays the real-time pressure (psia, or other unit) being measured in the Aspirator sys-
tem (gas applications only).
Cell Pressure
Displays the real-time pressure (psia, or other unit) being measured in the Measuring
Cell.
Next Cal 2
Displays the time remaining until the next Auto-Calibration 2 function. The duration
and units between Auto-Calibrations is determined by the duration entered in the timer
interval and the unit entered in the unit on the Automatic Calibration 2 screen. The
Calibration Trigger (TIMER 2) must be selected for the countdown time to be displayed.
Refer to the section “Calibration Control” in this chapter for descriptions about setting up
all Auto-Calibration parameters.
The settings for your analyzer have been configured at the factory to meet specified
! customer requirements. Changing the factory-set configuration could cause the ana-
CAUTION lyzer to operate incorrectly.
To view other sub-menu items under this screen, use the UP/DOWN arrow keys to scroll
through the list of analyzer functions to select the appropriate sub-menu item.
u Calibration Control
Relays
Current Outputs
Setpoints
Diagnostics
Figure 4-11. System Settings
SETUP menu screen. HOME ALARMS STATUS SETUP
Detailed descriptions of each sub-menu item – and the parameters under each – are
listed in the following pages (each listed as a heading in this manual). The list of available
sub-menu items includes:
• Calibration Control
• Relays
• Current Outputs
• Setpoints
• Diagnostics
• System Settings
• Modbus
The Calibration Control screen displays the available analyzer calibration controls and
allows the user to initialize the calibration functions.
u Calibration Settings
Calibration Valves
Calibration Actions
Figure 4-12.
Calibration Control
screen. HOME ALARMS STATUS SETUP
The Calibration Settings screen (Figure 4-13) displays all of the available analyzer cali-
bration functions and allows the user to set up the calibration functions.
u Automatic Calibration 1
Automatic Calibration 2
Flush Times
Figure 4-13.
Calibration Settings
screen. HOME ALARMS STATUS SETUP
Use these to set up the type of Auto-Calibration function that will be performed during
normal analyzer operation. The two setup screens allow multiple timers to be set.
Calibration Trigger
Use this to select how auto-calibration is triggered.
Automatic Calibration 1
uCalibration Trigger Timer
Mode Zero + Span
Timer Interval 0 day
Figure 4-14. Unit day
Calibration Settings Start Time [Link]
screen. HOME ALARMS STATUS SETUP
• Off
Disables Auto-Calibration. The system will not automatically Zero or Span
the analyzer regardless of what the automatic calibration mode function
is set to. The analyzer can still be manually calibrated using the Start zero
calibration, Start span calibration or Start zero+span calibration functions
on the Calibration screen.
Mode
Use this to set up the type of Auto-Calibration function that will be performed during
normal analyzer operation.
Span only
For its regular sequence of operation, the analyzer will perform a Span and then return
to Measure mode.
Zero only
For its regular sequence of operation, the analyzer will perform a Zero and then return to
Measure mode.
This function is not available for applications with multi-species span capability (i.e.,
when “Select span fluid” function is available on the Calibration screen). In this
! case, Auto cal – if enabled (ON) – will perform a zero calibration only. Span calibra-
NOTE tion must be done manually as described under “Manually Starting a Span Function”
on page 4-21.
Timer interval
Defines the time between each Auto-Calibration cycle. The unit is determined by the
Unit setting.
Start Time
Use this to set the start Auto-Calibration in military time.
2. Use the UP/DOWN arrow keys to view the options available for each parameter.
3. Change the timer durations (Auto cal timer interval, Zero flush time, Span flush
time):
• If a value exists, use the LEFT arrow key as a backspace to remove (erase) the
characters to the left of the cursor ( ‘_’ ) if applicable, then use the numeric
keys to enter the new value. This works best for drastic changes.
Flush Times
The Flush Times screen (Figure 4-15) displays all of the available flush times.
Flush Times
uZero 10 sec
Span 30 sec
Sample 20 sec
Figure 4-15.
Flush Times Settings
screen. HOME ALARMS STATUS SETUP
The Calibration Valves screen (Figure 4-16) allows the user to enter the calibration fluid
concentration value used for each analyte (Analyte cal conc). Note that the actual species
name will be displayed on the screen, not “Analyte” (e.g., SO2).
1. Select the analyte you wish to enter a calibration fluid concentration for and press
ENTER to go into Edit mode.
• When no value exists, use the numeric and decimal keys to enter a value.
• When a value exists, use the LEFT arrow key as a backspace to erase the
characters to the left of the cursor ( ‘_’ ), then use the numeric and decimal
keys to enter the new value. This works best for drastic changes.
3. For multi-species spanning, repeat Steps 1 and 2 for each species that will be
spanned.
The Calibration action screen displays all of the available analyzer calibration functions
and allows the user to initialize the calibration functions. Figures 4-17a and 4-17b illus-
trate the differences between analyzers that use single-species spanning and those that
use multi-species spanning.
Figure 4-16.
Calibration Valves
screen. HOME ALARMS STATUS SETUP
To select a span fluid, press ENTER to go into edit mode. Use the UP/DOWN arrow
key to select the fluid you wish to span and then press ENTER to select the span fluid.
Next, select Start span calibration and press ENTER to start the span.
Start measure
This function returns the analyzer to normal Measure mode from any of the Inject Zero
fluid, Inject Span fluid, Zero Gas Flush, or Span Gas Flush functions. To return the
Refer to “Manually Zeroing the Analyzer Sample System” in Chapter 3 for information
about how to perform a manual zero on the analyzer.
! Pressing CANCEL at any time during this procedure will abort the procedure and
return the software to the HOME screen.
NOTE
2. Ensure the Span fluid cylinder is connected to the Span Fluid Inlet on the right
side of the analyzer and is turned on.
3. From the SETUP menu, select the Calibration Control sub-menu and press enter.
Scroll down to the Calibration Fluid Concentration sub-menu and press enter to display
the Calibration Fluid Concentration screen (Figure 4-16).
Ensure that the concentration value for the Analyte being spanned corresponds
with the concentration of the span fluid. If not, edit the value as described under “Cali-
bration Fluids Screen,” later in this section.
5. From the Calibration screen, select Start span calibration and press ENTER to start
the span. The analyzer will span the selected fluid and then return to normal operation
(Measure mode).
The Relay Setup screens (Figure 4-18a and 4-18b) allow the user to view or set up the
parameters for up to eight relays. For example, Concentration Alarms can be set up from
this screen (Relay Function option).
Relay 1
u Function Disable
Source None
Value 0
Figure 4-18b. Delay 0 Sec N
Relay Setup screen Normal Oper Normally Open
(Relay 1). HOME ALARMS STATUS SETUP
u Relays
Relay 1
Relay 2
Relay 3
Figure 4-18a. Relay 4
Relay Setup screen. Relay 5
HOME ALARMS STATUS SETUP
Disable
Select Disable to turn off the functionality of a relay.
Calibrating
The relay will be de-energized if the analyzer is in a calibration state (Span Flush, Span,
Zero Flush, or Zero).
Data Valid
The relay will be de-energized when the analyzer is in any state other than the Measure
state.
Fault
The relay will be de-energized if any system fault alarm condition is triggered within the
analyzer diagnostic system. The analyzer requires service.
The relay will reset automatically upon correction of the fault [Link] Limit
Select this to enable an analyte concentration Low Limit [Link] Limit
Select this to enable an analyte concentration High Limit Alarm.
Heartbeat
Relay periodicaly will turn on and off to indicate that the
Analyzer is “ON”.
Check Request
Relay will be “ON” if any warnings are active.
User
Relay action controlled by analyzer internal task.
NONE
Select NONE if you do not wish to assign a relay to an output channel.
Analyte 1–8
Defines the analyte (species) of interest for the relay output.
You can assign a different relay for each channel. To assign a relay to a channel, select an
analyte and press ENTER. Note that the actual species name (e.g., SO2) will be displayed
on the screen, not “Analyte.”
For example, if Relay Delay is set to 5 seconds, the analyte concentration can exceed the
High Limit for up to 5 seconds at a time without de-energizing the relay. If the concen-
tration stays high for more than 5 seconds, the relay will be de-energized. The range
is 0–3600 seconds. This parameter only applies if Relay Function is set to Low Limit or
High Limit.
2. Select one of the Relay (1–8) sub-menus. Press ENTER to view the parameters
for that relay.
3. Select the required parameter (Relay Function, Source, Value, or Delay), change
the settings, and press ENTER.
When entering a value for the Relay Value and Delay parameters:
• If a value exists, use the LEFT arrow key as a backspace to erase the characters
to the left of the cursor ( ‘_’ ) if applicable, then use the numeric keys to enter
the new value. This works best for drastic changes.
The Current Output screens (Figure 4-19a and 4-19b) display current output assignments
and allows you to set up the parameters for each of the four current outputs.
Source
Defines the analyte assigned to each analog output. Current Out Source options include:
NONE
Select NONE if you do not wish to assign an output to a [Link] 1–8
This defines the analyte (species) of interest for the current output.
You can assign a different species for each channel. To assign an analyte to a channel,
select an analyte (species) and press ENTER. Note that the species name will be displayed
on the screen, not “Analyte” (e.g., SO2).
Full
This is the output full-scale value, where the user enters the current full-scale set point
corresponding to 20 mA DC for the output. For example, if Current Out 1 Full = 100
ppm, Output 1 would read 20 mA at 100 ppm.
Current Outs
u Current Out 1
Current Out 2
Current Out 3
Figure 4-19a. Current Out 4
Current Output Setup
screen. HOME ALARMS STATUS SETUP
Current Out 1
uSource Analyte 1
Full 1000.0
Zero 0.0
Figure 4-19b. Type 4-20 mA
Current Output Setup Over Range Disabled
screen (Current Out 1). HOME ALARMS STATUS SETUP
Zero
This is the output low-scale value, where the user enters the current low-scale set point
corresponding to 4 mA DC for the output. For example, if Current Out 1 Zero = 0 ppm,
Output 1 would read 4 mA at 0 ppm.
Type
This allows you to select the output value range, either 4–20 mA DC or 0–20 mA DC for
the output. Select ‘0’ for 4–20 mA DC or ‘1’ for 0–20 mA DC.
Over Range
When enabled, output readings of 3.8 and 20.5 mA will occur if the analyzer exceeds 4-20
mA.
2. Select one of the Current Output (1–8) sub-menus. Press ENTER to view the
Current Output screen.
3. Select the required parameter ( Source, Full, Zero, Type, or Over Range), change
the settings, and press ENTER.
• If a value exists, use the LEFT arrow key as a backspace to erase the characters
to the left of the cursor ( ‘_’ ) if applicable, then use the numeric keys to enter
the new value. This works best for drastic changes.
The Setpoints screens (Figure 4-20 thru 4-23) allow you to view configured normal oper-
ating (factory set) parameters for span, pressure, and temperature low/high limits.
Setpoints
> Span Limits
Pressure Limits
Temperature Limits
Figure 4-20.
Setpoints Screen. HOME ALARMS STATUS SETUP
Span Limits
If Span limits are exceeded, calibration is required.
Pressure Limits
If pressure limits are exceeded, the system alarm will activate and the analyzer will go
into “Not Ready” state. Limits are disabled if set to 0.
Temperature Limits
If temperature limits are exceeded, and in some heated applications, the system alarm
will activate and the analyzer will go into “Not Ready” state.
Span Limits
uSpan Limit Low 1 0.90
Span Limit High 1 1.10
Figure 4-21.
Span Limits
Setpoints Screen. HOME ALARMS STATUS SETUP
Pressure Limits
uCell Pres High Limit 142-15.2 PSI
Cell Pres Low Limit 142-15.2 PSI
Pres X 2 High Limit 0 PSI
Figure 4-22. Pres X 2 Low Limit 0 PSI
Pressure Limits Manifold Pres High Limit 70-100 PSI
Setpoints Screen. HOME ALARMS STATUS SETUP
Temperature Limits
uBench Deviation Limit 2º C
Bench Set Point 50º C
Heater Deviation Limit 0º C
Figure 4-23. Heater Set Point 190º C
Temperature Limits
Setpoints Screen. HOME ALARMS STATUS SETUP
The diagnostics screen (Figure 4-24) allows the user to disable normal operations while
troubleshooting or setting up external control systems.
Diagnostics
> Current Output Diagnostics
Relay Diagnostics
Valve Diagnostics
Lamp Diagnostics
UV Diagnostic
IR Diagnostic
Figure 4-24.
Diagnostics Screen.
HOME ALARMS STATUS SETUP
Relay Diagnostics
Force relay contacts on or off to test the hardware.
Relay Diagnostics
uDiagnostics State Off
Relay 1 On
Relay 2 On
Figure 4-27. Relay 3 On
Relay Diagnostics Relay 4 On
Diagnostics Screen. HOME ALARMS STATUS SETUP
Valve Diagnostics
uDiagnostics State Off
Valve 1 On
Valve 2 Off
Figure 4-28. Valve 3 Off
Valve Diagnostics Valve 4 Off
Diagnostics Screen. HOME ALARMS STATUS SETUP
Lamp Diagnostics
Monitor detector operation for a given wavelength.
Lamp Diagnostics
uWavelength 0.0
Measurement 16804
Dark Read 16718
Figure 4-29. Absorbance 0.1
Lamp Diagnostics Pixel 0
Diagnostics Screen. HOME ALARMS STATUS SETUP
The System Settings screen (Figure 4-30) allows the user to view and/or setup analyzer
system parameters.
The Login/Set Password screen (Figure 4-31) allows you to set up password protec-
tion for users who need access to modify analyzer configuration parameter settings, or
restrict users to read-only function. The analyzer is shipped from the factory without a
password. Upon powering up the analyzer for the first time, all configurable parameters
can be changed until a password is entered in the Change Password function.
When you first power up the analyzer, or following a system reset, the software secu-
u Enter Password
Change Password
Lock
Figure 4-31.
Login/Set Password
screen. HOME ALARMS STATUS SETUP
rity level defaults to read-only mode – you will have to re-enter the password to allow
changes. Password protection can also be removed, as described under “Removing the
Password.”
When a password has been entered, the software will remain in Edit (read-write) mode
as log as the user is working from the User Interface Panel. If no keys are pressed within
10 minutes, the software will revert to read-only mode. The password will have to be re-
entered before further changes can be made.
Enter Password
Use this function to enter a password that will allow you to make changes to many of the
analyzer configuration settings, or to perform calibration and other functions. Because
the analyzer is shipped from the factory without a password, you must first enter a pass-
word in the Change Password function. Record and store the password in a safe place
for reference.
Change Password
Use this function to enter a password for the first time after installing the analyzer or
whenever you are changing the password. The password entered here is the password
that must be entered in the Enter Password function that will allow you to change the
analyzer configuration settings.
Because the analyzer is shipped from the factory without a password, you must enter a
password in the Change Password function before you can change parameter settings.
If you attempt to change the password without first entering the password, the message
“System Locked” appears.
Keys that can be used to enter a password include any of the numeric keys (0–9), deci-
mal key ( . ), and the minus ( – ) key. A maximum of nine characters can be used for the
password.
Lock
Use this functions to clear the password from memory and “lock” all parameters, set-
tings, and functions as read-only. When you select Lock and press ENTER, the param-
eters are locked out and the message “System locked” is displayed. When the system is
locked, a padlock icon ( ) appears in the top-right corner of the HOME screen.
Occasionally, you should change the password that allows you to make changes to the
analyzer configuration settings. If password protection already exists, you must enter the
existing password before you can change it.
1. Select Enter Password and press ENTER. Key in the password and press EN-
TER again. The software will now allow you to change the password.
2. Select Change Password and press ENTER. Key in the new password and
press ENTER. Record and store the password in a safe place for reference.
Use the Lock function to disable the password from the current session. This reverts all
analyzer functions to read-only. Changes to the analyzer settings cannot be made until
the password is entered again.
Select Lock and press ENTER. The message “System Locked” appears. A padlock icon ap-
pears on the top-right side of the HOME screen to indicate the system is currently locked
from making any changes.
1. Select Enter Password and press ENTER. Key in the password and press EN-
TER. The software will now allow you to remove the password.
2. Select Change Password and press ENTER. Leave the password blank and
press ENTER again. The password is removed.
The Backup/Restore screen (Figure 4-32) allows you to back up the current analyzer con-
figuration settings, restore previously saved analyzer configuration settings, or restore
analyzer configuration parameters to their original factory-shipped settings.
u Backup settings
Restore settings
Restore factory settings
Figure 4-32.
Backup/Restore
Settings screen. HOME ALARMS STATUS SETUP
Restore settings
Retrieves and restores the most recently saved analyzer configuration settings from the
“[Link]” file. See “Restoring Analyzer Configuration Settings” for details about how to
restore configuration settings.
1. Press the SETUP soft key and then use the DOWN arrow to scroll down until
the Backup/Restore sub-menu is selected. Press ENTER to view the Backup/
Restore screen.
2. Select Backup settings and press ENTER. The software will prompt: “Backup
settings. Are you sure?” (Figure 4-33).
1. Press the SETUP soft key and then use the DOWN arrow to scroll down until the
Backup/Restore sub-menu is selected. Press ENTER to view the Backup/Restore screen.
• Select Restore settings if you wish to restore settings you last saved using
the Backup settings command.
• Select Restore factory settings if you wish to restore settings to the factory
default settings.
3. Press ENTER to initiate the command to restore the settings. The software will
prompt: “Restore settings. Are you sure?” (Figure 4-34) or “Restore factory
settings. Are you sure?” (Figure 4-35).
- If you selected Restore settings, the analyzer begins restoring its most
recently saved configuration settings (when the Backup settings com-
mand was last used) from a file called “[Link].”
After the settings have been restored, the analyzer enables the settings.
The Ethernet Network screen (Figure 4-36) allows the user to enter Ethernet settings.
This is used to enable your analyzer system to communicate over an Ethernet network
for viewing analyzer and stream composition data, viewing analyzer diagnostics, or
backing up/restoring analyzer configuration settings. For addresses, contact your net-
work administrator.
When changing the IP Address, IP Subnet Mask, and Gateway IP address, use the
numeric keys to enter a new value. When a period ( . ) character is necessary, after every
third digit of an address, the system will only accept input from the decimal key ( ‘.’ ).
Use the LEFT arrow key as a back space and to delete numerical characters to the left of
the cursor, or the RIGHT arrow key to advance to the right of the cursor.
Ethernet enable
Allows the user to enable or disable the Ethernet interface, to allow communication be-
tween the analyzer and network. This should be set disabled when the Ethernet is not in
use to prevent unauthorized network access.
To change this, press ENTER to enter Edit mode and then use the UP/DOWN arrow
keys to toggle between Enabled or Disabled.
DHCP enable
Allows the user to enable or disable the DHCP (Dynamic Host Configuration Protocol)
client. This function should be disabled to ensure the network uses a static address. The
DHCP server must be on the same Subnet as the analyzer.
IP Address
Defines the network of the analyzer. This is a required parameter. To change this, press
ENTER to enter Edit mode and change this address.
IP Subnet Mask
A TCP/IP number used to determine to which the TCP/IP Subnet device belongs. Devices
in the same Subnet can be communicated with locally without going through a router.
This is a required parameter.
Gateway (optional)
A Gateway (or “Router”) is a device which is used to forward IP packets to a remote des-
tination. The definition of “remote,” in this case, is a device that is not directly attached to
the same network segment as the sending device (e.g., the same Ethernet segment). This
function defines the Gateway IP address for the network, which allows local network
traffic to be sent to another network.
Setup of this parameter is required only if communicating from outside the internal net-
work. To change this, press ENTER to enter Edit mode and change this address.
MAC address
Defines the MAC (Media Access Control) address (unique byte number) of the analyzer
(e.g., 00-0F-88-00-00-79). The MAC address is factory-set. This address is read-only.
3. Press the SETUP soft key and then scroll down and select Ethernet Network.
Press ENTER to view the Ethernet Network screen.
4. Enter the IP Address, IP Subnet Mask address, and Gateway IP address (if
required), pressing ENTER after entering each address.
The System screen (Figure 4-37) allows the user to view and/or set up analyzer system
parameters.
Model Name, Serial Number, and Software Version are not editable, as indicated by the
‘>’ cursor.
System Time and System Date are editable, as indicated by the ‘u’ cursor. To enter Edit
mode for either of these parameters, press ENTER and then press the appropriate nu-
meric character. You can also use the RIGHT/LEFT arrow key to jump over a character. The
cursor will jump over the colon ( : ) and slash ( / ) characters when entering the time and
date.
> Model Name IPS-4
Serial Number 12345678
Software Version 1.10
System Time [Link]
System Date 07/11/2005
Figure 4-37.
System screen. HOME ALARMS STATUS SETUP
Model Name
This displays the Analyzer Model name.
Serial Number
This displays the analyzer serial number.
Software Version
This displays the analyzer software version.
System Time
The System Time clock is used for time-stamping various analyzer functions
(such as alarms).
To set the time, use the RIGHT/LEFT arrow key to select hours, minutes, and sec
onds and then either the UP/DOWN arrow key or numeric keys to enter the time.
When using the numeric keys to enter a value (while setting the time), the
cursor (the active, editable character) will move one character to the right
and skip over the colon ( : ) character.
System Date
The System Date is used for date-stamping various analyzer functions (such as
alarms). The format is defined by the Date format setting on the Customization
screen.
To set the date, use the RIGHT/LEFT arrow keys to toggle between day (DD), month
(MM), and year (YY) and the UP/DOWN arrow key to set the day, month, and year. When
using the numeric keys to enter a value, the cursor (the active, editable character) will
move one character to the right after each input.
The Customization screen (Figure 4-38) allows the user to view how information is dis-
played.
To customize how the analyzer displays information, use the UP /DOWN arrow key to
select an item and then press ENTER to go into Edit mode. Then, press the UP or DOWN
arrow key to scroll through the list of options and press ENTER to select the option.
Date Format
Defines the format for displaying the date, which is displayed on the System screen.
Format options include:
- MM/DD/YYYY - DD/MM/YYYY
- DD-MM-YYYY - MM-DD-YYYY
Decimal point
Defines the format for displaying the decimal point in a value:
- Period (example, 3.14)
- Comma (example, 3,14)
Language
Defines the language in which the software descriptions will be displayed. To view the
software in a language other than the default English, select Language, press ENTER,
then use the UP/DOWN arrow keys to scroll through the list until you locate the language
of choice. Press ENTER again to select the language.
Screen Brightness
This displays and allows you to change the screen brightness.
Modbus Screen
The Modbus screen (Figure 4-39) allows the user to view and/or set up Modbus param-
eters to enable the analyzer to communicate with the Modbus master.
Baud
This displays and allows you to change the current Baud Rate at which data will be
transferred using Modbus communications. The typical default is 9600. Other options
include 115200, 57600, 38400, 19200.
Parity
This displays and allows you to change the current Parity setting used for Modbus com-
munications. The typical default is even. Other options include odd or none.
u Baud: 9600
Parity even
Stop Bits 0
Modbus Address 1
Modbus Timeout 3000 msec
Figure 4-39. Modbus Port Off
Modbus screen. HOME ALARMS STATUS SETUP
Stop Bits
This displays and allows you to change the current Stop Bits setting used for Modbus
communications. The typical default is 0. It can also be set to 2.
Modbus Address
This displays and allows you to change the address assigned for Modbus communica-
tions. Valid addresses are 1–247. Setting the address to zero (‘0’) disables Modbus.
Modbus Timeout
This displays and allows you to change the time out value (duration) that the software
will use to attempt to establish communications with the analyzer when using Modbus
communications. AMETEK recommends a value of 1000 ms.
Modbus Port
This displays and allows you to change the Modbus communications with the analyzer
when using Modbus communications Port format. The default options include Modbus
RTU and Modbus TCP.
1. Select the parameter you wish to change and press ENTER to go into Edit
mode.
2. For Baud, Parity, and Stop Bits, use the UP/DOWN arrow key to scroll through
the available options. To select an option when it is displayed, press EN-
TER.
For Modbus Address and Modbus Timeout, enter a value using the numeric
keys.
Safety Considerations
Before performing any maintenance, service, or troubleshooting on
the analyzer, review and follow all personnel and equipment safety in-
! formation under “Personnel and Equipment Safety Information” fol-
NOTE lowing the Table of Contents near the beginning of this manual. This
information describes procedures to follow to avoid personal injury
and/or damage to the equipment. All regulatory agency and person-
nel safety procedures for your jurisdiction must be followed.
Preventive Maintenance
To reduce the occurrence of problems with the analyzer, AMETEK recom-
mends that you follow the “Analyzer Preventive Maintenance Schedule,”
as outlined in the following pages. Since most analyzer problems originate
within the sample handling system, the primary objective of the preven-
tive maintenance schedule is proper care of the sample system.
Frequency Task
As required Clean the Analyzer
Cleaning the exterior of the analyzer is required occasionally to re-
move dust and other debris. Use a damp cloth to clean the analyz-
er’s exterior, including the User Interface Panel screen.
Daily Check for Alarms
To view alarms, press the ALARMS soft key to view the Alarms
screen.
To view active alarms, select Current Alarms and press ENTER. For
detailed information about these alarms, refer to the alarm condi-
tions and corrective action in the “Troubleshooting and Diagnostics”
section of this chapter.
Check the history buffers to see if alarms are recurring. The history
buffers contain alarms that have been reset. To view a history of
alarms that were set by the analyzer but have since been reset,
select Alarm Log and press ENTER. If there is a history of alarms
that have been reset, the alarm descriptions are displayed. If there
is no history of alarms, “Log empty” is displayed. To scroll through
the list, use the UP/DOWN arrow key.
Monthly Zero the Analyzer / Check Analyzer Response Time
Manually Zero the analyzer (from the Calibration screen, select
Start zero calibration and press ENTER).
Observe the readings and record the time it takes the analyzer to
display the first reading after the switch. Typically, a good response
time is less than 30 seconds to T90. Check your log book to verify
response times. A response time that is slower than normal may
suggest plugging problems in the analyzer sample system.
Aspirator pressure should be set higher than 70 kPa (10 psi) above
the sample stream pressure.
CAUTION:
To avoid damaging the pressure transducer, do not set the pressure
higher than the Cell pressure transducer rating. See drawing pack-
age for maxium pressure ratings.
The secondary gaskets (with the wire mesh) on the door are used
for electrical contact between each door and its enclosure. Ensure
these gaskets are intact.
Frequency Task
At the same time, replace the o-rings (always) and windows and/or
window/mirror combinations (if chipped, cracked, or scratched).
Observe the Cell Pressure reading from the STATUS screen, and if the
reading approaches the Zero fluid supply pressure, a plug is present
somewhere in the sample inlet.
After confirming that the plug is somewhere in the sample inlet, take
! all appropriate safety precautions and dismantle the sample system,
NOTE starting at the Sample Probe or process tap.
Plugs in the vent side of the sample system are rare. Plugs in the vent sys-
tem cause the aspirator drive air to flow back into the Measuring Cell and,
essentially, simulate a Zero condition.
This is done by supplying and maintaining the necessary heat and insula-
tion to the sample system.
Check current operating temperatures and compare them to the set point
temperatures for each temperature zone.
1. While working from the User Interface Panel, manually force the
analyzer to continuously backflush the sample system with Zero fluid
(Zero Gas Flush mode).
2. The flow indicator should be in the centre of the meter for an optimal
flow rate.
3. Close the Sample Valve on the Sample Probe. Observe the flow indica-
tor. If the flow reduces to a stop, a large plug has occurred somewhere
in the sample system.
5. After locating the cause of the restriction or plug (and after cleaning
or replacing the part or line), leak test the fittings or connections that
were loosened or replaced. To do this, follow the procedures under
“Sample System Leak Check” in Chapter 3.
6. Return the Vent Line Shut Off valve to the open position. Both isola-
tion valves should be open.
While working on the Cell the following tools, equipment, and supplies
are required (supplied by AMETEK only where indicated):
• Measuring Cell spare parts. Refer to either the Measuring Cell Manual
Supplement or the Custom Spare Parts List (if applicable), located in the
“Supplemental Information” section of this manual and the Documen-
tation Package shipped with the analyzer.
• Set of open-end wrenches for fittings and set of metric ball drivers.
• Cotton swabs, to clean the interior of the Cell Body and other metal
parts.
For example:
If suitable for the species, an acceptable material is Chem-Thane – or a
nonabrasive detergent and water solution. Reagent-grade acetone (if
suitable), can then be used to rinse the Cell Body, followed by a rinse
with pharmaceutical-grade distilled water.
The o-rings in the Measuring Cell should be replaced every year to avoid
a leak in the sample system. If the sample system operates above atmo-
spheric pressure, sample fluid can leak from the Measuring Cell into the
Oven/Sample System Enclosure. While the Measuring Cell is disassem-
bled, clean its chambers and parts.
Replace the o-rings in the Measuring Cell every time this assembly is
apart.
2. After the zero is complete, close the Sample Line and then the Vent
Line Shut Off valves (in the Oven/Sample System Enclosure) to isolate
the analyzer from the sample stream.
Before proceeding, test the area around the analyzer for flammable
gases. If a flammable gas atmosphere is present, do not power down
! the analyzer or any alternate power sources that supply power to the
WARNING analyzer components. Proceed only when the area is found to be safe.
REFLECTOR BLOCK
SCREW
300-8430
(3 PLS)
O-RING
MIRROR (2 PLS) FITING APLLICATION SPECIFIC
272285001 (IR) 203336018 (4 PLS)
403007001 (UV) (2 PLS)
O-RING
300-2361 (2 PLS)
MIRROR CAP
300-1261 (2 PLS)
Figure 5-2.
Measuring Cell Assembly
(40 cm Measuring Cell).
Figure 5-2 illustrates a 40 cm Measuring Cell. If your application uses
a different Cell, refer to the “Supplemental Information” tab of this
! manual for a Measuring Cell Manual Supplement and for application-
specific spare parts, either in the Measuring Cell Manual Supplement
NOTE or in a Custom Spare Parts List.
a. Disconnect the Sample Line and Vent Line tubes from the Cell.
Loosen and remove the Cell RTD (if used) from the RTD Clamp
(not shown) on the Cell.
b. Note the position of the Cell and its Sample/Vent fittings and Re-
flector Block.
Remove the three screws that secure the Cell to the Window
Housing Assembly (not shown). Carefully pull the Cell straight out
from the Window Housing and place it on a work bench or other
suitable location to work on the removed cell.
! In the next steps use a soft, non-abrasive cloth to place the Cell parts
on.
NOTE
Use an o-ring removal tool to remove the outer o-ring from the
Cell. Take care not to scratch or damage the o-ring groove.
d. While holding the Cell vertically so that the Reflector Block end is
up, remove the three M4 x 20 mm screws that secure the Reflector
Block to the Cell. Remove the Reflector Block carefully to ensure
the Cell Window behind it does not drop out.
e. With the Reflector Block end up, carefully remove the o-rings and
Cell Window from the Cell.
g. Inspect the interior of the Cell Body for foreign material. If neces-
sary, use a cotton swab to clean the Cell Body.
Clean the fittings using the same method as the Cell Body.
For example:
If suitable for the species, an acceptable material is Chem-Thane –
or a nonabrasive detergent and water solution. The Cell Body can
then be rinsed with reagent-grade acetone (if suitable), followed
by a rinse with pharmaceutical-grade distilled water.
a. Insert a new o-ring in the o-ring groove on the flat surface of the
Reflector Block.
With the Cell held vertically so that the Reflector Block end of the
Cell is up, install a new o-ring by using your thumbs to push it
into the o-ring groove until it is completely seated in the groove.
b. With the Cell held vertically so that the Reflector Block end is
down, install a new o-ring by using your thumbs to push it into
the o-ring groove until it is completely seated in the groove.
8. Reconnect the Sample Line and Vent Line tubes to the Cell.
Reconnect all other sample system parts/tubing (use Teflon tape on fit-
tings that originally were taped).
Do not close the Oven Heater fuse (if used) until after the Leak Check
has been performed.
10. Perform a leak check on all of the fittings that were reconnected. If
necessary, follow the procedure under “Sample System Leak Check,”
in Chapter 3.
After the leak check has been performed and has passed, return to this
procedure and continue with the remaining steps.
Close the Oven Heater fuse to apply power to the Oven Heater
temperature-control zone circuitry.
Close the analyzer enclosure doors and secure them with their latches
to allow the analyzer to warm up to operating temperature and stabi-
lize. This could take up to 2 hours if this sample system is allowed to
cool completely.
Open the Vent Line Shut Off valve (in the Oven/Sample System En-
closure). Open the Sample Line Shut Off valve to allow sample fluid
into the sample system.
13. Switch the analyzer back to Measure mode (from the Calibration
screen, select Start measure and press ENTER). Press the HOME soft key
to return to the HOME screen.
15. Close and secure the analyzer enclosure doors with their latches.
Figure 5-3.
Infrared Source
Assembly (700046901)
2. After the zero is complete, close the Sample Line and then the Vent
Line Shut Off valves (in the Oven/Sample System Enclosure) to isolate
the analyzer from the sample stream.
Before proceeding, test the area around the analyzer for flammable
! gases. If an explosive gas atmosphere is present, do not power down
the analyzer or any alternate power sources that supply power to the
WARNING
analyzer components. Proceed only when the area is found to be safe.
5. Loosen the locking clamp on the circular mount and carefully remove
the optical bench. Remove the bench cover.
7. Insert the new assembly into the optical bench. Coat the three contact
faces with a thin coat of heat-sink compound.
Insert the two screws removed earlier and attach the wires to TB1 pins
three (red) and 4 (black).
8. Replace the cover and slide the bench back onto its mounting stud.
Retighten the locking clamp.
Before proceeding, test the area around the analyzer for flammable
! gases. If an explosive gas atmosphere is present, do not apply power to
CAUTION the analyzer or any alternate power sources that supply power to the
analyzer components. Proceed only when the area is found to be safe.
Close the analyzer enclosure doors and secure them with their latches
to allow the analyzer to warm up to operating temperature and stabi-
lize. This could take up to 2 hours if this sample system is allowed to
cool completely.
11. Switch the analyzer back to Measure mode (from the Calibration
screen, select Start measure and press ENTER). Press the HOME soft key
to return to the HOME screen.
13. Close and secure the analyzer enclosure doors with their latches.
The glass Xenon Flash Lamp is under high internal pressure which
could result in flying glass fragments if ruptured. Do not subject the
lamp to drop impact, vibration, or shock.
! While handling a flash lamp, always wear protective devices (face
WARNING
mask, clothing) to prevent possible injury, especially to hand and face
areas.
When handling the flash lamp, never touch the glass bulb of the lamp
with bare hands. Dust or fingerprints on the glass bulb may greatly
! reduce transmittance in the ultraviolet range.
CAUTION
If the glass bulb must be cleaned, wipe it using a soft, lint-free cloth
moistened with high-quality alcohol.
2. After the zero is complete, close the Sample Line and then the Vent
Line Shut Off valves (in the Oven/Sample System Enclosure) to isolate
the analyzer from the sample stream.
Figure 5-3.
Xenon Flash Lamp
Assembly 403225901
Before proceeding, test the area around the analyzer for flammable
gases. If a flammable gas atmosphere is present, do not power down
! the analyzer or any alternate power sources that supply power to the
WARNING analyzer components. Proceed only when the area is found to be safe.
5. While holding the Lamp Assembly with one hand, loosen (by hand)
the blue knurled nut that secures it to the Optical Bench Assembly
and remove the Lamp Assembly. Dispose of the Lamp Assembly fol-
lowing company policy, or local regulations.
! Do not use any tool to loosen the blue nut on the Optical Bench when
removing or replacing the Lamp Assembly.
WARNING
6. Take the new Xenon Flash Lamp Assembly 403225901 and, using the
alignment pin on the Optical Bench Assembly as a guide, gently push
the Lamp Assembly straight into place.
Tighten, by hand, the blue knurled nut to secure the Lamp Assembly
to the Optical Bench Assembly.
! Do not apply main AC power to the analyzer if the Xenon Flash Lamp
Assembly electrical connections have not been made.
NOTE
The battery on the MCU board powers the analyzer clock when the main
analyzer AC power is not connected. The clock’s primary function is to
timestamp the Alarm Log entries.
The battery’s expected life-span is greater than five years; therefore, it is
important to order and stock a battery when your analyzer is approaching
five years from the original manufacture date. It is also possible that the
battery can become damaged by a power spike. Ordering information is
located in Chapter 6.
2. After the zero is complete, close the Sample Line and then the Vent
Line Shut Off valves (in the Oven/Sample System Enclosure) to isolate
the analyzer from the sample stream.
Before proceeding, test the area around the analyzer for flammable
! gases. If a flammable gas atmosphere is present, do not power down
WARNING the analyzer or any alternate power sources that supply power to the
analyzer components. Proceed only when the area is found to be safe.
6. Check the System Time and System Date (System screen, SETUP
menu) and set them to the current time and date.
Open the Vent Line Shut Off valve (in the Oven/Sample System En-
closure). Open the Sample Line Shut Off valve to allow sample fluid
into the sample system.
10. Close and secure the analyzer enclosure doors with their latches.
Alarm Notification –
When an alarm is set by the analyzer, the alarm message flashes on the
HOME screen to alert the user. The information on the HOME screen will
be replaced by a reverse (white) screen with the alarm name in black text.
It will then switch to a black screen with white text and will continue to
flash until the alarm conditions that triggered the alarm are reset.
All active alarms are displayed on the Current Alarms screen. Each line
will occupy one alarm description. Up to six active alarms can be dis-
played on the screen at once. If more than six alarms are active, use the
UP/DOWN arrow keys to scroll through the list. The alarm at the top of the
list is most recent alarm; the alarm at the bottom is the oldest.
Alarms that have been reset (no longer active) are stored in the Alarm
Log, accessed from the ALARMS screen. Logged (or historical) alarms
– which occupy two lines – include the time and date each alarm was
set and reset. The logged alarm at the top of the list is most recently re-
set alarm; the alarm at the bottom is the oldest. The Alarm Log will be
cleared when the analyzer is powered down.
Use the UP/DOWN arrow keys to scroll through the list (if there are
more than three logged alarms).
Note: The system alarm relay contact opens on alarm. A system alarm
! can result from any of the alarm conditions listed below. Some alarms
NOTE may not apply based on the system hardware configuration. For ex-
ample, systems without a heated sample system will not have a heater
RTD alarm.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
and the Oven/Sample System Enclosure doors, and:
• Using an Ohm Meter, measure the resistance of the Oven Heater
Plate RTD and test it for an open circuit. If the RTD is faulty, contact
AMETEK for assistance.
Cell Temperature
Bench Temperature
Oven Temperature Note:
The Cell and Over Temperature alarms apply only to systems with a
heated Oven/Sample System Enclosure.
Corrective Action:
For alarms caused by Low / High Temperatures:
• From the User Interface Panel:
- Check current operating temperatures for each temperature zone
and compare them to their set point temperatures.
- Check each of the associated configuration parameters for
temperature control. These parameters may have been lost as a
result of a power interruption or a power spike.
Corrective Action:
• The parameters have exceeded the normal operating range. From the
User Interface Panel, check that the full-scale range is correct for the
current sample. If Pressure Transducer alarms persist during normal
operation, contact the factory. A range change and recalibration may
be required.
Condition/Alarm Name Description & Suggested Corrective Action
Manifold pressure
Aspirator pressure These alarm conditions indicate that the pressure within these
components exceeds its full-scale range by more than 10 psia (typical;
set point and pressure unit may vary).
Corrective Action:
• The parameters have exceeded the normal operating range. From
the User Interface Panel, check that the full-scale range is correct for
the current sample. If Manifold or Aspirator pressure alarms persist
during normal operation, contact the factory. A range change and
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
door and:
• Check all of the ribbon cables between boards for proper connections
and inspect them for damage (cuts, nicks, burn marks, etc.). For
cable connection locations, refer to the Electronics Enclosure Wiring
Diagram in the Appendix.
• Contact AMETEK for assistance.
Failed communications to the Analog board
This alarm condition indicates that the MCU board cannot establish
reliable communications with the Analog board.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
door and:
• Check the ribbon cable (Cable 3) between the MCU board (P3) and
Analog board (J101) – both boards mounted on the Electronics
Enclosure door – for proper connections and inspect it for damage
(cuts, nicks, burn marks, etc.). For cable connection locations, refer to
the Electronics Enclosure Wiring Diagram in the Appendix.
• One or both boards may need to be replaced. Contact AMETEK for
assistance.
Failed communications to the Relay board
This alarm condition indicates that the MCU board cannot establish
reliable communications with the Relay board.
Corrective Action:
Take appropriate safety precautions, open the Electronics Enclosure
door and:
• Check the flat cable (Cable 3) between the MCU board (P3, on
the Electronics Enclosure door) and Relay board (J101, on the
Electronics Enclosure backpan) for proper connections and inspect
it for damage (cuts, nicks, burn marks, etc.). For cable connection
locations, refer to the Electronics Enclosure Wiring Diagram in the
Appendix.
• One or both boards may need to be replaced. Contact AMETEK for
assistance.
Condition/Alarm Name Description & Suggested Corrective Action
Relay board firmware not compatible
This alarm condition occurs if the Relay board cannot communicate with
the MCU board due to incompatible firmware versions.
Corrective Action:
The Relay board and/or MCU board firmware will have to be replaced:
• Contact AMETEK for assistance.
Corrective Action:
The Display board and/or MCU board firmware will have to be replaced:
• Contact AMETEK for assistance.
File system error This alarm condition occurs if the Detector data file (also known as the
DSP, or digital signal processing, file) is missing or has become corrupt.
Corrective Action:
• Contact AMETEK for assistance.
Corrective Action:
Perform a successful span calibration:
• Contact AMETEK for assistance.
Zero Cal Required
This warning is generated after a zero calibration failure or if a zero has
not been performed.
Corrective Action:
Perform a successful zero calibration:
• Contact AMETEK for assistance.
Zero Cal Failure
This warning is generated after a zero calibration failure.
Corrective Action:
Perform a successful zero calibration:
• Contact AMETEK for assistance.
Span Cal Failure
This warning is generated after a span calibration failure.
Corrective Action:
Perform a successful span calibration:
Analyzer Reset
If a situation arises where the analyzer must be reset, take appropriate
safety precautions power down the analyzer using the power-disconnect
switch (breaker). Reapply power.
The Service & Parts chapter discusses what to do if you need technical support from
AMETEK, or if you are returning parts for service. This chapter also lists the recom-
mended spare parts to have on hand to ensure all consumable and replaceable parts
are replaced according to the “Analyzer Preventive Maintenance Schedule” included in
Chapter 5.
Technical Support
AMETEK Western Research is committed to providing you the best technical support in
the industry. If you need service or application assistance, contact your local or nearest
AMETEK Service Centre or the AMETEK factory AMETEK at (403) 235-8400 or 1-800-661-
9198), or contact your local AMETEK Western Research representative.
Before contacting AMETEK with questions regarding the installation, operation, or main-
tenance/troubleshooting of your analyzer system, carefully review the contents of this
manual. If you are unable to find an explanation for your problem in this manual, please
gather the following information prior to contacting AMETEK:
• AMETEK part number for the specific component you are enquiring about, if
known.
Before ordering spare parts for the Measuring Cell, refer to the “Supplemental Informa-
tion” section of this manual or the Documentation Package shipped with the analyzer for
a Measuring Cell Manual Supplement or a Custom Spare Parts List. If the Cell maintenance
procedure is described in a separate Manual Supplement, check that document first for a
Spare Parts list. If included, use those part numbers; if not, use the part numbers listed in
the Custom Spare Parts List.
Choose the appropriate fuses (120 or 240 VAC) for the application.
! These main analyzer fuses do not need to be changed out at regular intervals, but
NOTE AMETEK recommends these fuses are available in the event that a replacement fuse is
required.
The User Web Interface chapter contains information about the following topics:
• An introduction to the User Web Interface, and how to access, monitor, and
control your analyzer through a web interface
• What the various screens that make up the interface do and how to use them
to work from the analyzer User Web Interface.
Before working from the Web Interface, read the section titled “Introduction to the
! Web Interface,” in the following pages. This section discusses the functions that can be
NOTE performed.
The analyzer IP address is listed on the Ethernet Network Screen on the user interface
panel. To use the web interface the analyzer ethernet must be enabled. Refer to Chapter
4 for ethernet network information.
Analyzer IP Address
Figure 7-1.
Web Interface with
Analyzer IP Address.
The information displayed on the right-hand side varies depending on the selected sec-
tion.
• The web interface header displays the “state” of the analyzer and the concen-
tration of the analytes being measured. This section is read only.
- Examples of “states” include Zero, Span, and Measure. These and other states
are described in detail under “System Status Messages (HOME Screen),” in
Chapter 4.
• Alarm and warning notifications are displayed in the banner located below
the header on every screen. The most current notification is displayed with 1
of x if there are multiple active notifications.
The Web Interface is shipped from the factory without password protection. All con-
figurable parameters can be changed through the web interface.
!
NOTE
Trend and Spectral data are stored in temporary memory. For security reasons, tempo-
rary data can not be stored directly onto the computer hard drives. Any intentional or
unintentional recycling of power will cause a loss of saved data.
Data cannot be saved directly to your computer. To make a record of diplayed informa-
tion use your PC screen capture function and paste the image into another program.
Navigation
Buttons
All active alarms and warnings are displayed on the Active Alarms screen. Each line will
occupy one alarm or warning description and up to 15 active alarms or warnings can be
displayed on the screen at once. If more than 15 are active, use the arrow keys to scroll
through the list. The alarm or warning at the top of the list is most recent.
To view active conditions, click the Show Active button. If there are alarms active, the
alarm description is displayed (Figure 7-3). If no alarms are active, the display will be
empty.
! For a complete list of alarms and warnings generated by the analyzer, refer to the
NOTE “Maintenance & Troubleshooting” chapter.
Alarms / Warnings
Notification Banner
Indicates Number
of Active Alarms
Figure 7-3.
Active Alarms Screen.
Alarms that have been reset (no longer active) are stored in the Alarm Log, accessed
from the ALARMS screen. Each logged (or historical) alarm includes the time and date
it was set. The logged alarm at the top of the list is most recently reset alarm. The log file
could be up to 800 lines long. After 800 lines it will overwrite the oldest messages.
When an alarm or warning is set by the analyzer, the alarm is displayed in the banner
accross the top of the screen.
Use Check
Boxes to Filter
Alarms/Warnings
Displayed
To refresh the display with the most recent data click the “Get Trends” button.
Select Parameter
to Display
Select Trending
Interval
Click to
Refresh Display
Figure 7-5.
Trends Menu Screen.
Trends for the following system parameters are available for display:
• Bench Temperature
• Heater Temperature
• Lamp Temperature
• Cell Temperature
• Cell Pressure
• Manifold Pressure
• Aspirator Pressure
• Electronics Temperature
• Analytes Concentration
Intervals
Trend intervals for selected parameters are available in one second, fifteen seconds or
one hour data points. The analyzer stores 3600 parameter data points in temporary
memory prior to overwriting data.
• 1 hour interval will contain 3,600 hours or 150 days before overwriting.
Select data
source to display
Click to
Update Chart
Click to View
Spectra
Click to Clear
User Data
Click to Save
User Data
Figure 7-6.
Spectra Menu Screen.
Live
View Live Spectra by selecting the data source to display.
User
Compare User Spectra to Live Spectra for trouble shooting.
• Click “Save User” button to save user spectra to the hard drive. The file is
saved as a cookie in the browser.
The settings for your analyzer have been configured at the factory to meet specified
! customer requirements. Changing the factory-set configuration could cause the ana-
NOTE lyzer to operate incorrectly.
Detailed descriptions of each menu item and parameters are described in Chapter 4. The
list of available sub-menu items includes:
• Calibration Control
• Actions
• Current Outputs
• Relays
• Analytes
• Limits
• Diagnostics
The CALIBRATION SETTINGS screen (Figure 7-7) displays the available analyzer
calibration controls and allows the user to set up and initialize the calibration functions.
Descriptions of these parameters are described under “Calibration Settings Screens” in
Chapter 4.
Use to
Set up Calibration
The ACTIONS screen (figure 7-8) allows the user to enter calibration concentrations and
initiate on demand calibration actions.
Initiate
Action
Figure 7-8.
Settings Menu
Actions Screen.
The CURRENT OUTPUT screen (Figure 7-9) displays current output assignments and
allows you to set up the parameters for each of the four current outputs. Descriptions of
these parameters are described under “Current Output Setup Screens” in Chapter 4.
The Relay Setup screens (Figure 7-10) allow the user to view or set up the parameters for
up to eight relays. For example, Concentration Alarms can be set up from this screen (Re-
lay Function option). Descriptions of these parameters are described under “Relay Setup
Screens” in Chapter 4.
The NETWORK & MODBUS screen (Figure 7-11) allows the user to enter ethernet set-
tings and set up modbus parameters.
Ethernet settings
Enable the analyzer system to communicate over an Ethernet network for viewing ana-
lyzer and stream composition data, viewing analyzer diagnostics, or backing up/restoring
analyzer configuration settings. Descriptions of these parameters are described under the
“Ethernet Network Screen” in Chapter 4.
Modbus screen
Set up Modbus parameters to enable the analyzer to communicate with the Modbus
master. Descriptions of these parameters are described under the “Modbus Screen” in
Chapter 4.
Use to Use to
Enter Ethernet Settings Set Up Modbus Parameters
The SYSTEM & CUSTOMIZATION screen (Figure 7-12) allows the user to view and/or
setup analyzer system parameters.
Use to
View System Parameters Customize Display
The ANALYTES screen (figure 7-13) allows you to view and modify analyte names and
span factor min/max.
Figure 7-13.
Settings Menu
Analyte Screen.
The LIMITS screen (figure 7-14) allows you to view and modify pressure and tempera-
ture limits. Descriptions of these parameters are described under “Setpoints Screens” in
Chapter 4.
Figure 7-14.
Settings Menu
Limits Screen.
The DIAGNOSTICS screen (Figure 7-15) allows the user to disable normal operations
while troubleshooting or setting up external control systems. Descriptions of these pa-
rameters are described under “Diagnostics” in Chapter 4.
The analyzer should be in normal operations for lamp diagnostics. The lamp does not
! flash in diagnostics mode. For the diagnostics to operate, click the ‘Turn Diagnostics
NOTE On button’.
Click to put
Analyzer in Diagnostics
Mode Use to Check
Current outputs
Use to Check
Solenoid
Valves
Use to Test
Hardware
Enter Wavelength
to Test During Normal Operation
Figure 7-15.
Settings Menu
Diagnostics Screen.
• Save IPS-4 software, user configuration settings, and individual files (matrix.
xml for example) to your PC.
Click to Save or
Transfer Data
Figure 7-16.
Settings Menu
Save/Restore Files Screen.
This chapter describes the customer serial communication interface on Model IPS-4
Analyzer. The communication protocol implemented is Modicon Modbus as defined in
“Modicon Modbus Protocol Reference Guide” (PI-MBUS-300 RevJ). The Modbus proto-
col transmission mode implemented is Remote Terminal Unit (RTU) with the analyzer
operating as a slave device.
The IPS-4 Analyzer supports RS485 serial communication standards. The physical com-
munication connection between a Model IPS-4 analyzer and a customer DCS\SCADA\
PLC\DAS or a general-purpose computer is RS485. The analyzer RS485 connection sup-
ports 2-wire multi-drop systems.
Hardware:
The jumpers on the CPU board J9 connector should be installed between 1&2 and 3&4
pins.
The converter terminal marked TD(B) which is same as Tx+ should be connected to the
CCB_J14 connector marked Tx+ (pin3) on the Customer Connection Board, and TD(A)
should be connected to Tx- (pin 4).
! Make sure that you rebooted the analyzer after changing the MBUSPORT parameter
value to 1.
NOTE
When designing a Modbus RS485 multi-drop communication system with the Model
IPS-4 analyzer, the system designer should consider the following:
• The update rate is one time per minute or less. Polling of these registers more
frequently than once a second is not recommended.
• When operating in MODBUS TCP mode. Use TCP Port 502 for modbus com-
munications.
Modbus Address
The analyzer needs to be assigned a Modbus slave address, which can be a number from
1 to 247 .
Modbus Functions
As the Modbus protocol is designed for communication among Programmable Logic
Controllers (PLCs), not all Modbus function codes supported by a slave PLC are appli-
cable to the IPS-4 analyzer. Only the following relevant function codes are implemented:
CODE DESCRIPTION
03 read multiple holding register (4x references)
06 write one holding register (4 references)
16 write multiple holding registers (4x references)
Figure 8-1(B7-1).
Modbus Functions Screen
This Appendix includes many of the drawings that are included in the main body of this
manual, as well as additional analyzer layout and wiring diagram drawings.
Drawings such as “Analyzer Mounting Details, Unistruts” and “Analyzer Overall Compo-
nent Layout” are examples only; customer-specific drawings for your system are located
in the “Supplemental Information” section of this manual and in the Documentation
Package shipped with the analyzer.
!
NOTE
[ 12.7] 6 PLS
46.25
[1174.8]
analyzer.
19.70
[500.5] 28.45
22.00
21.70 [722.7]
[558.8]
[551.3]
48.00
[1219.2]
Analyzer Mounting Details
22.00
HEATED SAMPLE LINE
[558.8] SLEEVE SHOWN
(OPTIONAL)
!
NOTE
XENON FLASH
LAMP ASSEMBLY
OVEN ENCLOSURE
ELECTRONIC ENCLOSURE
analyzer.
DISPLAY INTERFACE PWB
OVEN HEATER
RELAY PWB
MCU PWB
POWER SUPPLY
CAL (SPAN) FLUID IN
DC TERMINAL ASSEMBLY
ANALOG PWB
Analyzer Overall Component Layout
SAMPLE IN
AC TERMINAL/
MAIN FUSES
SAMPLE OUT
P2, SAMPLE LINE POWER C3, SAMPLE LINE HEAT TRACER POWER
Information” section of this manual or the Documentation Package shipped with the
Detector Interface Board (403441901)
C20 C13
+
C7 R6 U6
P2 R9
+
C5 P4
C3
R7
C19 JP1 C14
C9 X1 C18 J3 R8 R13 1
U5
C8
U10 J4
U8 D2
R10 1 3 5
U2 C15
R11
U1
R4 C11
U11 U9 C12 U12 2 4 6
C17
R12
C10
403441901
REV
C2
TP1
U13 J1
U7
C6
GND C21
+
C19 C23 C36
C14
J4 R9 L5
C13 C18 R18
C38 R21
C2 U13
R6 L3 P3
J3 R25
R5 C37 R19 R28
U3 C12 R22
R7
C20 R17 R37
L1
C5 C8 C42 R36 R38
+
D1 U6 R35
R8
C10 C9
U2 R32 J2
C6 J1 700030901 U10 C11
C3 L4 C28 C16
U5 TP4
C4 C31 C32 TP6 R24
C45
C30 D2
U11 R26
C7 R2 R3
R23 R13 RV1
R1 TP8 C44
R33
C39 U9 R15
C27 REV P2
R30 D4 R31
R27
C34
C35
R29
C41
C33 C43
R4 R11 TP3
D5
COM
VI
700043901
C101
R101
C104
R102
R100
R103
C100
U100
C102
REV
C103 C105
WHT
BLK
BLK
BLK
BLK
RED
RED
RED
RED
RTD 301-0860 301-0859
CABLE17
18BLK300
18RED300
XENON FLASH LAMP DIB_P2 4AT 2AT
SSR1-4/A2
SSR1-3/A1
1 18BLK600
1 18BLK600
1 18BLK600
1
2
3 18GRN/YEL600
4
2 18BLK600
3
2 18WHT600
3 18GRN/YEL600
4
2
3
4
7
8
9
13
1
2
1
2
3
4
5
6
10
ASSY. 100-1994 OPTICAL BENCH ASSY. 301-0862 301-0861
CON3A
100-1987 or 100-1999 4B 4A 2AT
CABLE16 BLK 3B 3A WHT 12 301-0861
RB_J402
RB_J401B
RB_J401A
RB_J401C
16GRN/YEL600
AB_J404 CELL RTD
RB_J109 RB_J203 RED 2B 2A WHT
18RED300
18BLK300
G-IN
GOUT
GOUT
N/C
N/C
L-IN
LOUT
L-IN
N-IN
N/C
LOUT
NOUT
LOUT
NOUT
WHT 1B RED 6.3 A T
1A 11
OVEN HTR BENCH & LAMP BENCH HTR LAMP HTR 12 VDC SOLENOID VALVES 301-0863
CABLE3
4 HTR MCU_P3 CON3B
3
2 4B 4A
CT
1
CABLE15 BLK 3B 3A RED
RB_J404
S/V LINE
AB_J402
RED 2B 2A RED
SSR2-4/A2 22BLK300 4 WHT 1B 1A WHT
BENCH RTD
SSR2-3/A1 22RED300 3 TB2-1
2 SHEET3 TB2-8 CUSTOMER CON2B CON2A
1
RELAY BOARD
SR
SHIELD SHIELD
S/V LINE
RB_J403
(PROGRAMMED) 100-2050 CONNECTION PRESSURE
22BLU300 4 BOARD 100-1971 TRANSDUCER
SV-OT-1 BLK 4B 4A BLK
22YEL300
18BLK300
18RED300
SV-OT-0 3 CABLE13 WHT 3B 3A WHT
2 RB_J101 AB_J302
+
CCB_J19 GRN 2B 2A GRN
SHEET2
1
RB_J504
RED 1B 1A RED
OT RELAYS
1C1 AUX. POWER 1 22BLU600 CON4B CON4A
VL SL OVEN HEATER 1C2 2 22RED600
OT RTD OT RTD OT RTD RELAY CONTACTS CONTACT IN FLOW SW IN 2C1 CCB_J1B 3 22GRN600
SHEET4
2C2 4 22YEL600
RB_J503 RB_J502 RB_J501 RB_J301 RB_J201 RB_J202
3C1 5 22VLT600 RED 6
3C2 6 22OR600 5
4C1 7 22GRY600
1
1
1
1
1
4
2
2
2
2
2
4
4
4
4
4
3
3
3
3
3
9
8
7
5
4
3
2
1
11
16
15
14
13
12
10
TB3-13 22RED300 8 22BLK600 CABLE9 SENSOR 2
C4
4C2 3
SHEET2
0.1u
RED
RED
WHT
TB3-11 5C2 10 22BRN600 BLK 1
6C1 11 22BLU600 WHT NOTE: PIN 1
END
6C2 12 22RED600 HAS A NOTCH
RED 6 NC
22YEL600
22GRN600
22RED600
22BLU600
22GRY300
22VLT600
22BLK600
22WHT300
22BLK300
22BRN300
22GRY600
18BLK600 13
22OR600
22GRN600
22WHT600
22BRN600
22RED600
22BLU600
22OR600
22VLT600
22GRN600
22YEL600
SSR1-2/T1 5 NC
18
14 22YEL600
4 NC
18WHT600 15 22VLT600 CABLE8 SENSOR 1
C3
3 VS
17
0.1u
16
WHT
WHT CCB_J2B
18WHT600 1 22GRY300
15
2 22BLK300
18BLK600 CONTACT INPUT 3 22WHT300
14
4 22BRN300
18BLK600 CCB_J2A
13
ETHERNET
CCB_J17
18BLK600
18BLK600
12
PS1 ACL CABLE12
SHEET3
MCU_J5 NDIR OPTICAL BENCH ASSY.
18BLK600
11
18BLK 600 CCB_J18 22VIO-250 BRAID
SSR1-1/L1 CCB_J8
CON-A PS1-GND
18BLK600 700046901
10
22BRN-300 18RED 600
WHT/RED
CON-B IRB_P10 IRB_P20
SHEET3
9
CON-D DIB_P2 CABLE5 CABLE18
PS1 CURRENT OUT 1-2 18BLK600
SHEET3
18GRN/YEL600 PS1-(-)
8
5 4 3 2 1
CCB_J9 CCB_J16
7
CCB_J10
18WHT600 RS-232 8
VAR1
6
CURRENT OUT 3-4
CIRCUIT BOARDS
CCB_J12
5
CABLE6 SYMBOL BOARD NAME PART No. SHEET
VAR2
MCU_P6 AB ANALOG BOARD 100-2047 2
4
CCB_J15 CCB CUSTOMER CONNECTION BOARD 100-1971 1
18BLK600 CCB_J11 CABLE7 DB DETECTOR BOARD 100-2046 1
3
CCB_J3 MCU_P7 DIB DETECTOR INTERFACE BOARD 100-2048 2
CCB_J6 DIO DISPLAY INTERFACE BOARD 100-2049 2
2
SHEET2
VAR3
AB_J103 RB RELAY BOARD 100-2050 1
ANALOG IN 1
1
XLB XENON LAMP BOARD 100-2061 3
4B 4A
END CABLE14 WIRE CODE
BLK 3B 3A RED
AB_J401 RED 2B 2A RED ANALOG XX YYY/YYY ZZZ
TB1 WHT 1B 1A WHT RS-485
14GRN/YEL 600
BRAID
IN2 RTD1 RTD2
CCB_J13
LF1 CON1B CON1A CCB_J7 CCB_J4 CCB_J5 CCB_J14 AWG COLOR VOLTAGE RATING
LOAD
BACK PANEL GROUND
V+
VR
TX+
TX-
RX+
RX-
EXC
CM1
CM2
GND
EXC
OVEN RTD
18GRN/YEL 600 OVEN HEATER OT RTD
18GRN/YEL600 WHT/RED
L
WHT OVEN
N
GROUND STUD H 14GRN/YEL600 GRN/YEL G HEATER TB3-7 22 WHT 300
TB3-8 22 RED 300
(SHEET2) GROUND GROUND 22 RED 300
BRAID OVEN HEATER CONDUIT TB3-9
STUD B STUD C
BACK PANEL
GROUND POINT D
Electronics Enclosure Wiring Diagram, Sheet 1 of 4 (403444001)
DIO_P3
BOARD 100-2049
DIO_P2
DIO_P4
1 16 CABLE NUMBER
4
3
2
CABLE1 100-2029
18RED300
TB2-3 CABLE2 273119003
CABLE3 273119002
CABLE11
18BLK300
SHEET3
TB2-6 CABLE4 301-0733
CABLE5 100-2029
KEY
PAD
CABLE5
IRB_P10 CABLE6 301-0897
CABLE7 301-0897
CABLE3
RB_J101
SHEET1
CABLE8 301-0930
CABLE9 301-0930
MCU_P3
CABLE10 301-0896
CABLE3
MCU_P7
CABLE13 301-0936
DIB_P2 DIB_P4
CABLE14 301-0930
DIB_P1
MCU_P2
DETECTOR INTERFACE
SHEET1
CABLE15 301-0930
CABLE6 (PROGRAMMED) 403441901
CCB_J12 CABLE16 301-0930
AB_J101
CONNECTOR AND COVER
MCU_P6
271229001,280517001
CABLE17 273119001
MCU BOARD 100-2045 1 18RED300 CABLE18 700065901
2
TB2-4
(PROGRAMMED)
3
CABLE10 4
MCU_P1 MCU_J5 5
MCU_P4
SHEET3
MCU_P5
6 18BLK300
TB2-5
CABLE17
DB_J1 (SHEET1)
CABLE12
CCB_J17 (SHEET1)
CABLE4
XLB_J1 (SHEET3)
CABLE16
CON3B
CABLE15
CON2B BRAID GROUND STUD F
SHEET1
MCU_P3
CABLE14 (KEYPAD PLATE)
CON1B
GROUND STUD G
(BEZEL MOUNTING SCREW)
SHIELD
SHIELD
SHIELD
WHT
WHT
WHT
BLK
BLK
BLK
RED
RED
RED
CABLE3
1
4
2
3
1
1
4
1
2
3
4
2
3
2
3
4
14GRN/YEL600 GROUND
SHIELD
STUD B (SHEET 1)
BLK 4
CABLE13 GROUND
WHT 3 AB_J404 AB_J403 AB_J402 AB_J401
CON4B GRN 2
STUD H
AB_J101 CELL RTD LAMP RTD BENCH RTD OVEN HEATER
RED 1 RTD
AB_J302
SENSOR 1
PRESSURE
4
3
2
ANALOG BOARD
AB_J303
SHIELD 1
SENSOR2
PRESSURE
CABLE9 (PROGRAMMED) 100-2047
SENS2 1 18RED300
TB2-2
SHEET1
BLK 6 2 18BLK300
WHT 5 TB2-7
12 VDC
AB_J104
SHEET3
RED 4
AB_J103
SHIELD 3
BLK 2
AB_J304
SENSOR3
WHT 1
CABLE8
Electronics Enclosure Wiring Diagram, Sheet 2 of 4 (403444001)
SENS1 RED
CABLE2 CCB_J3
(SHEET1)
TB1-11
18BLK600 XENON LAMP BOARD 100-2061
18BLK600
TB1-18
XLB_J2 XLB_J1 XLB_J3
18BLK300 J403-2
(SHEET2)
BOARD OFF ON
CON* 22BRN300 CCB_J16_3
22BLK300
CUSTOMER
18BLK300 - CCB_J16_5 CONNECTION
C B A BOARD
22GRY300 CCB_J16_2
+
(SHEET1)
18RED300
F E D
AUX 18BLK300
M L K
TB1 TB4 J H G
IRB_P20
11 I0+ RTN 46
12 RTN OUT3 45 18RED300
13 RTN OUT2 44 18BLK300 22GRY300
14 I1+ RTN 43 22GRY300
15 RTN OUT1 42 END
DIGITAL OUTPUT
ANALOG OUTPUTS
16 RTN OUT0 41 SHIELD
CABLE 18
CCB_J19-1
(SHEET1) 18RED300 18RED300 RB_J109-1 (SHEET1)
1
RS485
RTN D 34 +V
24 I3+ D 33 12V RET 18BLK300 RED
25 RTN +B 18RED300 -V
32 BLK BLK
SHEET2
18BLK300
PWR
ANALOG OUTPUTS
26 RTN A 31 -V
18BLK300 MCU_P4-6
18BLK300 DIO_P4-3
ACN
8 7 6 5 4 3 2
R1
18BLK600 RS-75-12
1M
TB1-12
18WHT600 CCB_J19-2 18BLK300
TB1-16 END
18GRN/YEL600 (SHEET1)
SHEET 1
TB1-8
TB2
Electronics Enclosure Wiring Diagram, Sheet 3 of 4 (403444001)
SHEET 1
J504-4 22 BLU 300
4
12BLK600
SSR2
+3/A1 2/T1
12BLK600
1
12 GRN/YEL600
TB3
GROUND
STUD 1 2 3 4 5 6 7 8 9 10 11 12 13 14
G L N N L G
POWER HEAT TRACER RTD1 RTD2
Disconnect Enclosure Wiring Diagram, Sheet 4 of 5 (403444001)
SUPPLEMENTAL INFORMATION
This chapter consists of information that supports the operation of this analyzer. All
information in this chapter comprises the Documentation Package. A duplicate of the
Documentation Package is also shipped with the analyzer.
• Manual Supplements – that describe and illustrate installation, operation (if ap-
plicable), and maintenance (if applicable) procedures for specific components
or optional equipment that make up your sample system.
Supplements are typically used for Measuring Cells, Sample Systems, or in-
formation that is intended to replace similar information in the User Manual.
• Any OEM products incorporated with this analyzer system, if applicable (ex-
ample, EXPO MiniPurge System manual, if used).