Proline Promag 10 Modbus RS485 Guide
Proline Promag 10 Modbus RS485 Guide
Table of contents
4 "Application" menu . . . . . . . . . . . . . . . . 49
4.1 "Measured values" submenu . . . . . . . . . . . . . . 49
4.1.1 "Totalizer" submenu . . . . . . . . . . . . . . 51
4.2 "System units" submenu . . . . . . . . . . . . . . . . . 53
4.3 "Totalizers" submenu . . . . . . . . . . . . . . . . . . . 58
4.3.1 "Totalizer handling" submenu . . . . . . . 58
4.3.2 "Totalizer 1 to n" submenu . . . . . . . . . 58
4.4 "Sensor" submenu . . . . . . . . . . . . . . . . . . . . . . 63
4.4.1 "Process parameters" submenu . . . . . . 63
4.4.2 "Low flow cut off" submenu . . . . . . . . 66
4.4.3 "Partially filled pipe detection"
submenu . . . . . . . . . . . . . . . . . . . . . . 67
4.4.4 "Sensor adjustment" submenu . . . . . . . 69
4.4.5 "Calibration" submenu . . . . . . . . . . . . 70
4.4.6 "Electrode cleaning cycle" submenu . . . 72
4.5 "Current output" submenu . . . . . . . . . . . . . . . . 74
4.6 " Communication" submenu . . . . . . . . . . . . . . 85
4.6.1 "Modbus configuration" submenu . . . . 85
4.6.2 "Modbus data map" submenu . . . . . . . 88
4.6.3 "Modbus information" submenu . . . . . 89
5 "System" menu . . . . . . . . . . . . . . . . . . . . 90
5.1 "Device management" submenu . . . . . . . . . . . . 91
Endress+Hauser 3
About this document Proline Promag 10 Modbus RS485
1.3.1 Symbols
Types of information
4 Endress+Hauser
Proline Promag 10 Modbus RS485 About this document
Endress+Hauser 5
"Guidance" menu Proline Promag 10 Modbus RS485
2 "Guidance" menu
Main functions for use – from fast and safe commissioning to guided support during
operation.
Navigation Guidance
Guidance
‣ Commissioning → 6
‣ Commissioning
Device tag → 8
Serial number → 8
Firmware version → 8
Device name → 8
Volume unit → 10
Mass unit → 11
Density unit → 12
Temperature unit → 12
Conductivity unit → 12
Unit totalizer 1 to n → 13
6 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Failure mode → 15
Flow damping → 15
Fixed current → 23
Failure current → 24
Value 1 display → 25
Value 2 display → 25
Value 3 display → 26
Value 4 display → 26
Display damping → 26
Time format → 27
Time zone → 27
Set date/time → 28
Endress+Hauser 7
"Guidance" menu Proline Promag 10 Modbus RS485
Device tag
Description Enter a unique designation for the measuring point to be able to easily identify it within
the plant.
User entry Character string comprising numbers, letters and special characters (32)
Serial number
Description Displays the serial number of the measuring device. The serial number can be used to
identify the measuring device and to retrieve further information on the measuring device,
such as the related documentation, via the Device Viewer or Operations app.
Additional information:
The serial number can also be found on the nameplate of the sensor and transmitter.
User interface Character string comprising numbers, letters and special characters
Firmware version
User interface Character string comprising numbers, letters and special characters
Device name
8 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
User interface Character string comprising numbers, letters and special characters
Endress+Hauser 9
"Guidance" menu Proline Promag 10 Modbus RS485
Volume unit
10 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Mass unit
Endress+Hauser 11
"Guidance" menu Proline Promag 10 Modbus RS485
Density unit
Temperature unit
Prerequisite Temperature measurement is only optionally available for Promag H 10 (5HBB): Under
order code for "Functionality", option D (extended transmitter) and order code for "Sensor
option", option CI (medium temperature measurement)
Conductivity unit
12 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Selection SI units
• nS/cm
• µS/cm
• µS/m
• µS/mm
• mS/m
• mS/cm
• S/cm
• S/m
• kS/m
• MS/m
2.1.3 Totalizer 1 to n
Selection • Off
• Volume flow
• Mass flow
Unit totalizer 1 to n
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
or
Endress+Hauser 13
"Guidance" menu Proline Promag 10 Modbus RS485
or
Other units
None *
Selection
The selection is dependent on the process variable selected in the Assign process variable
parameter (→ 13).
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
14 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Failure mode
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
Description Specify how the totalizer should behave in the event of a device alarm.
Additional information:
The failsafe mode that applies to any other totalizers or outputs is specified separately in
other parameters and is not impacted by this setting.
Selection • Stop
• Actual value
• Last valid value
Flow damping
Description Enter value for damping of the flow measured value in order to reduce the variability of
the flow measured value when exposed to interference.
Additional information:
The depth of the flow filter is determined by this setting. As the filter depth increases, so
does the reaction time of the device.
- Value = 0: No damping. Damping of 0 is not recommended, as the measuring signal is
then so noisy that it is almost impossible to perform a measurement.
- Value > 0: Damping increases
Optimal damping depends on the measuring period.
Damping impacts the following measuring device variables:
- Outputs
- Low flow cut off
- Totalizers
Endress+Hauser 15
"Guidance" menu Proline Promag 10 Modbus RS485
User entry 0 to 15
Description Enter time frame for signal suppression (= pressure shock suppression active), for example
to prevent the device from registering flow movements in the pipe when a valve is closed.
Additional information:
Pressure schock suppression is activated when the flow rate drops below the on value for
low flow cutoff. Output values when pressure shock suppression is active:
- Flow: 0
- Totalizer: Last valid value
Pressure shock suppression is deactivated when:
- The time frame specified has elapsed and
- Flow exceeds the off value for low flow cutoff
16 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Q D C D
1 2 3 4
6
5
t
A B
A0012888
Q Flow
t Time
A After run
B Pressure shock
C Pressure shock suppression active as specified by the time entered
D Pressure shock suppression inactive
1 Valve closes
2 Flow falls below the on-value of the low flow cut off: pressure shock suppression is activated
3 The time entered has elapsed: pressure shock suppression is deactivated
4 The actual flow value is now displayed and output
5 On-value for low flow cut off
6 Off-value for low flow cut off
Description Select process variable for low flow cut off to activate low flow cut off.
Selection • Off
• Volume flow
• Mass flow
Endress+Hauser 17
"Guidance" menu Proline Promag 10 Modbus RS485
Q 2 2
4
H
3
t
1 A 1 A
A0012887
Q Flow
t Time
H Hysteresis
A Low flow cut off active
1 Low flow cut off is activated
2 Low flow cut off is deactivated
3 On-value entered
4 Off-value entered
Description Enter off value to switch off low flow cut off. The off value is entered as a positive
hysteresis with respect to the on value.
18 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Description Switch empty pipe detection on or off. Switch on empty pipe detection to detect a partially
filled or empty measuring tube.
Selection • Off
• On
Endress+Hauser 19
"Guidance" menu Proline Promag 10 Modbus RS485
Selection • Off *
• Mass flow
• Conductivity *
• Corrected conductivity *
• Volume flow
• Temperature *
• Noise *
• Coil current shot time *
Description Select the current range for the measured value output and the upper and lower fault
condition signal level.
Additional information:
- The measured value range is specified in the "Lower range value output " parameter and
the "Upper range value output " parameter.
- If the measured value lies outside the scaled measured value range, diagnostic message
"441 Current output faulty" is generated.
- In the event of a device alarm, the current output adopts the behavior specified in the
"Failure behavior current output " parameter.
20 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
The graphic shows the relationship between the current range for the output of the
process value and the lower and upper alarm levels:
2 1 3
I [mA]
A0034351
Selection (current range for process value) Lower level for signal on Upper level for signal on
alarm alarm
Prerequisite In the Current range output parameter, one of the following options is selected:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
Description Enter lower range value for the measured value range.
Additional information:
- Depending on the setting selected for the "Measuring mode current output " parameter,
the value specified for this parameter and the "Upper range value output " parameter must
have the same algebraic sign or not.
- As a rule, the lower range value is scaled to be lower than the upper range value. As a
result, the behavior of the current output is proportional to the process variable assigned.
If the lower range value is scaled to be higher than the upper range value, then the
behavior of the current output will be inversely proportional to the process variable
assigend.
Additional information Examples of the behavior, depending on the option selected in the Measuring mode
current output parameter (→ 75).
Endress+Hauser 21
"Guidance" menu Proline Promag 10 Modbus RS485
I [mA]
20
8
4
Q
-250 0 +750
1 1
A0013757
Q Flow
I Current
1 Measuring range is exceeded or undershot
The linear operational range of the measuring device is defined by the values entered for
the Lower range value output parameter (→ 21) and Upper range value output
parameter (→ 22), and by the selected current range.
I [mA]
20
4
0 Q
0
3 1 2
A0013758
Q Flow
I Current
1 Value assigned to 0/4 mA current
2 Forward flow
3 Reverse flow
The current output signal is independent of the direction of flow (absolute amount of the
measured variable). The values for the Lower range value output parameter (→ 21)
and Upper range value output parameter (→ 22) must have the same algebraic sign.
The value for the Upper range value output parameter (→ 22) (e.g. reverse flow)
corresponds to the mirrored value for the Upper range value output parameter
(→ 22) (e.g. forward flow).
Prerequisite In the Current range output parameter, one of the following options is selected:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
22 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
Description Enter upper range value for the measured value range.
Additional information Examples of the behavior, depending on the option selected in the Measuring mode
current output parameter: Lower range value output parameter (→ 21):
Prerequisite A process variable is selected in the Process variable current output parameter and one
of the following options is selected in the Current range output parameter:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
Description Enter time constant (PT1 element) to set the reaction time of the output signal to
fluctuations in the measured value caused by process conditions.
Additional information:
- The smaller the time constant the faster the output reacts to fluctuations in the
measured value.
- If the time constant is 0, damping is deactivated.
Fixed current
Prerequisite In the Current range output parameter in the Current output 1 submenu, the Fixed
value option is selected.
Prerequisite A process variable is selected in the Process variable current output parameter and one
of the following options is selected in the Current range output parameter:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
Endress+Hauser 23
"Guidance" menu Proline Promag 10 Modbus RS485
Description Specify how the output should behave in the event of a device alarm.
Selection • Min.
• Max.
• Last valid value
• Actual value
• Fixed value
Failure current
Prerequisite In the Failure behavior current output parameter in the Current output 1 submenu, the
Fixed value option is selected.
Description Enter the value for the "Fixed value" option in the "Failure behavior current output "
parameter.
24 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
2.1.7 Display
Value 1 display
Description Select the measured value that is displayed first on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Value 2 display
Description Select the measured value that is shown second on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • None
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
• Totalizer 1
• Totalizer 2
• Totalizer 3
Endress+Hauser 25
"Guidance" menu Proline Promag 10 Modbus RS485
Value 3 display
Description Select the measured value that is shown third on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • None
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
• Totalizer 1
• Totalizer 2
• Totalizer 3
Value 4 display
Description Select the measured value that is shown fourth on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • None
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
• Totalizer 1
• Totalizer 2
• Totalizer 3
Display damping
Description Enter time constant (PT1 element) to set reaction time of the display to fluctuations in the
measured value.
Additional information:
- The smaller the time constant the faster the display reacts to fluctuations in the
measured value.
- If the time constant is set to 0, damping is deactivated.
26 Endress+Hauser
Proline Promag 10 Modbus RS485 "Guidance" menu
2.1.8 Date/time
Time format
Selection • 24 h
• 12 h AM/PM
Time zone
Description Select the time zone. Every time the time zone is changed, a logbook entry is created.
Endress+Hauser 27
"Guidance" menu Proline Promag 10 Modbus RS485
Selection
Other units
• UTC-12:00
• UTC-11:00
• UTC-10:00
• UTC-09:30
• UTC-09:00
• UTC-08:00
• UTC-07:00
• UTC-06:00
• UTC-05:00
• UTC-04:00
• UTC-03:30
• UTC-03:00
• UTC-02:00
• UTC-01:00
• UTC 00:00
• UTC+01:00
• UTC+02:00
• UTC+03:00
• UTC+03:30
• UTC+04:00
• UTC+04:30
• UTC+05:00
• UTC+05:30
• UTC+05:45
• UTC+06:00
• UTC+06:30
• UTC+07:00
• UTC+08:00
• UTC+08:45
• UTC+09:00
• UTC+09:30
• UTC+10:00
• UTC+10:30
• UTC+11:00
• UTC+12:00
• UTC+12:45
• UTC+13:00
• UTC+14:00
Set date/time
Description Set the date and local time. Every time the date or time is changed, a logbook entry is
created.
28 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
3 "Diagnostics" menu
Troubleshooting and preventive maintenance – settings for device behavior during process
and device events as well as assistance and measures for diagnostic purposes.
Navigation Diagnostics
Diagnostics
‣ Active diagnostics → 30
‣ Diagnostic list → 32
‣ Simulation → 35
‣ Heartbeat Technology → 38
‣ Diagnostic settings → 39
Endress+Hauser 29
"Diagnostics" menu Proline Promag 10 Modbus RS485
‣ Active diagnostics
Actual diagnostics → 30
Timestamp → 30
Previous diagnostics → 30
Timestamp → 31
Operating time → 31
Actual diagnostics
Timestamp
Description Displays the timestamp for the currently active diagnostic message.
User interface Days (d), hours (h), minutes (m), seconds (s)
Previous diagnostics
30 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
Description Displays the diagnostic message for the last diagnostic event that has ended.
Timestamp
Description Displays the timestamp of the diagnostic message generated for the last diagnostic event
that has ended.
User interface Days (d), hours (h), minutes (m), seconds (s)
Description Indicates how long the device has been in operation since the last time the device was
restarted.
User interface Days (d), hours (h), minutes (m), seconds (s)
Operating time
User interface Days (d), hours (h), minutes (m), seconds (s)
Endress+Hauser 31
"Diagnostics" menu Proline Promag 10 Modbus RS485
‣ Diagnostic list
Diagnostics 1 → 32
Timestamp → 32
Diagnostics 2 → 33
Timestamp → 33
Diagnostics 3 → 33
Timestamp → 33
Diagnostics 4 → 33
Timestamp → 34
Diagnostics 5 → 34
Timestamp → 34
Diagnostics 1
Description Displays the currently active diagnostic message with the highest priority.
Timestamp
Description Displays the timestamp for the diagnostic message with the highest priority.
User interface Days (d), hours (h), minutes (m), seconds (s)
32 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
Diagnostics 2
Description Displays the currently active diagnostic message with the second highest priority.
Timestamp
Description Displays the timestamp for the diagnostic message with the second highest priority.
User interface Days (d), hours (h), minutes (m), seconds (s)
Diagnostics 3
Description Displays the currently active diagnostic message with the third highest priority.
Timestamp
Description Displays the timestamp for the diagnostic message with the third highest priority.
User interface Days (d), hours (h), minutes (m), seconds (s)
Diagnostics 4
Description Displays the currently active diagnostic message with the fourth highest priority.
Endress+Hauser 33
"Diagnostics" menu Proline Promag 10 Modbus RS485
Timestamp
Description Displays the timestamp for the diagnostic message with the fourth highest priority.
User interface Days (d), hours (h), minutes (m), seconds (s)
Diagnostics 5
Description Displays the currently active diagnostic message with the fifth-highest priority.
Timestamp
Description Displays the timestamp for the diagnostic message with the fifth highest priority.
User interface Days (d), hours (h), minutes (m), seconds (s)
34 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
‣ Simulation
Description Select a process variable for the simulation, thereby activating it.
Selection • Off
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
Endress+Hauser 35
"Diagnostics" menu Proline Promag 10 Modbus RS485
Description Enter the simulation value for the selected process variable. Processing of measured values
downstream as well as the signal ouput follow this value. In this way, it is possibe to verify
whether the measuring device has been configured correctly.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • Off
• On
Description Enter the current value for the simulation. In this manner, it is possible to verify the
current output is configured correctly and downstream processing units are functioning
properly.
Additional information:
The valid input range is determined by the "Current range output " parameter.
36 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
Selection • Off
• On
Description Select the category of diagnostic events that should be available for selection in the
"Diagnostic event simulation" parameter.
Selection • Sensor
• Electronics
• Configuration
• Process
Selection Off
Endress+Hauser 37
"Diagnostics" menu Proline Promag 10 Modbus RS485
‣ Heartbeat Technology
38 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
‣ Diagnostic settings
‣ Properties → 39
‣ Diagnostic configuration → 39
‣ Properties
Alarm delay → 39
Alarm delay
Description Enter a duration for the alarm delay. When a diagnostic event of the "Alarm" category
occurs, the diagnostic message is not generated until the delay has elapsed.
User entry 0 to 60 s
‣ Diagnostic configuration
‣ Sensor → 40
‣ Electronics → 40
‣ Configuration → 43
‣ Process → 43
Endress+Hauser 39
"Diagnostics" menu Proline Promag 10 Modbus RS485
"Sensor" submenu
‣ Sensor
Navigation Diagnostics → Diag. settings → Diag. config. → Sensor → Diagnostic no. 043
Description Select behavior for diagnostic event "043 Sensor short circuit detected".
Selection • Off
• Alarm
• Warning
• Logbook entry only
"Electronics" submenu
‣ Electronics
40 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
Navigation Diagnostics → Diag. settings → Diag. config. → Electronics → Diagnostic no. 230
Selection • Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Electronics → Diagnostic no. 231
Description Select behavior for diagnostic event "231 Date/time not available".
Selection • Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Electronics → Diagnostic no. 302
Description Select behavior for diagnostic event "302 Device verification active".
Selection • Off
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Electronics → Diagnostic no. 376
Description Select behavior for diagnostic event "376 Main electronics faulty".
Endress+Hauser 41
"Diagnostics" menu Proline Promag 10 Modbus RS485
Selection • Off
• Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Electronics → Diagnostic no. 377
Description Select behavior for diagnostic event "377 Sensor electronics (ISEM) faulty".
Selection • Off
• Alarm
• Warning
• Logbook entry only
42 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
"Configuration" submenu
‣ Configuration
Navigation Diagnostics → Diag. settings → Diag. config. → Configuration → Diagnostic no. 441
Description Select behavior for diagnostic event "441 Current output faulty".
Selection • Off
• Alarm
• Warning
• Logbook entry only
"Process" submenu
‣ Process
Endress+Hauser 43
"Diagnostics" menu Proline Promag 10 Modbus RS485
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 832
Description Select behavior for diagnostic event "832 Electronics temperature too high".
Selection • Off
• Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 833
Description Select behavior for diagnostic event "833 Electronics temperature too low".
Selection • Off
• Alarm
• Warning
• Logbook entry only
44 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 834
Description Select event category (status signal) for diagnostic event "834 Process temperature too
high".
Selection • Off
• Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 835
Description Select behavior for diagnostic event "835 Process temperature too low".
Selection • Off
• Alarm
• Warning
• Logbook entry only
Endress+Hauser 45
"Diagnostics" menu Proline Promag 10 Modbus RS485
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 842
Description Select behavior for diagnostic event "842 Process value above limit".
Selection • Off
• Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 937
Selection • Off
• Alarm
• Warning
• Logbook entry only
46 Endress+Hauser
Proline Promag 10 Modbus RS485 "Diagnostics" menu
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 938
Selection • Off
• Alarm
• Warning
• Logbook entry only
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 961
Description Select behavior for diagnostic event "961 Electrode potential out of specification".
Selection • Off
• Alarm
• Warning
• Logbook entry only
Endress+Hauser 47
"Diagnostics" menu Proline Promag 10 Modbus RS485
Navigation Diagnostics → Diag. settings → Diag. config. → Process → Diagnostic no. 962
Selection • Off
• Alarm
• Warning
• Logbook entry only
48 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
4 "Application" menu
Targeted optimization to the application – comprehensive device settings from sensor
technology to system integration for optimum application adaptation.
Navigation Application
Application
‣ Measured values → 49
‣ System units → 53
‣ Totalizers → 58
‣ Sensor → 63
‣ Current output 1 → 74
‣ Communication → 85
‣ Measured values
Mass flow → 49
Volume flow → 50
Conductivity → 50
Corrected conductivity → 50
Temperature → 51
‣ Totalizer → 51
Mass flow
Endress+Hauser 49
"Application" menu Proline Promag 10 Modbus RS485
Volume flow
Conductivity
Corrected conductivity
50 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Temperature
Prerequisite Temperature measurement is only optionally available for Promag H 10 (5HBB): Under
order code for "Functionality", option D (extended transmitter) and order code for "Sensor
option", option CI (medium temperature measurement)
‣ Totalizer
Totalizer value 1 to n → 51
Totalizer overflow 1 to n → 52
Totalizer value 1 to n
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
Endress+Hauser 51
"Application" menu Proline Promag 10 Modbus RS485
Totalizer overflow 1 to n
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
52 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
‣ System units
Volume unit → 54
Mass unit → 55
Density unit → 56
Temperature unit → 56
Conductivity unit → 56
Endress+Hauser 53
"Application" menu Proline Promag 10 Modbus RS485
Volume unit
54 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Mass unit
Endress+Hauser 55
"Application" menu Proline Promag 10 Modbus RS485
Density unit
Temperature unit
Prerequisite Temperature measurement is only optionally available for Promag H 10 (5HBB): Under
order code for "Functionality", option D (extended transmitter) and order code for "Sensor
option", option CI (medium temperature measurement)
Conductivity unit
56 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Selection SI units
• nS/cm
• µS/cm
• µS/m
• µS/mm
• mS/m
• mS/cm
• S/cm
• S/m
• kS/m
• MS/m
Endress+Hauser 57
"Application" menu Proline Promag 10 Modbus RS485
‣ Totalizers
‣ Totalizer handling → 58
‣ Totalizer 1 to n → 58
‣ Totalizer handling
Description Reset all totalizers to "0" and restart the totaling process. All flow quantities thus far
totalized are thereby deleted.
Selection • Cancel
• Reset + totalize
‣ Totalizer 1 to n
Unit totalizer 1 to n → 59
Control Totalizer 1 to n → 60
58 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Preset value 1 to n → 61
Failure mode → 61
Selection • Off
• Volume flow
• Mass flow
Unit totalizer 1 to n
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
or
Endress+Hauser 59
"Application" menu Proline Promag 10 Modbus RS485
or
Other units
None *
Selection
The selection is dependent on the process variable selected in the Assign process variable
parameter (→ 13).
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
Control Totalizer 1 to n
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
60 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Selection • Totalize
• Reset + hold
• Preset + hold
• Reset + totalize
• Hold
Preset value 1 to n
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
Example
This configuration is suitable for applications such as iterative filling processes with a fixed
batch quantity.
Failure mode
Prerequisite A process variable has been selected in the Assign process variable parameter in the
Totalizer 1 to n submenu.
Description Specify how the totalizer should behave in the event of a device alarm.
Additional information:
The failsafe mode that applies to any other totalizers or outputs is specified separately in
other parameters and is not impacted by this setting.
Endress+Hauser 61
"Application" menu Proline Promag 10 Modbus RS485
Selection • Stop
• Actual value
• Last valid value
62 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
‣ Sensor
‣ Process parameters → 63
‣ Sensor adjustment → 69
‣ Calibration → 70
‣ Process parameters
Flow damping → 64
Flow override → 64
Conductivity measurement → 64
Fixed density → 66
Endress+Hauser 63
"Application" menu Proline Promag 10 Modbus RS485
Flow damping
Description Enter value for damping of the flow measured value in order to reduce the variability of
the flow measured value when exposed to interference.
Additional information:
The depth of the flow filter is determined by this setting. As the filter depth increases, so
does the reaction time of the device.
- Value = 0: No damping. Damping of 0 is not recommended, as the measuring signal is
then so noisy that it is almost impossible to perform a measurement.
- Value > 0: Damping increases
Optimal damping depends on the measuring period.
Damping impacts the following measuring device variables:
- Outputs
- Low flow cut off
- Totalizers
User entry 0 to 15
Flow override
Description Stops the measuring process. Can be used for example when cleaning the pipeline.
Selection • Off
• On
Conductivity measurement
Prerequisite Conductivity measurement is only optionally available: Under order code for
"Functionality", option D (extended transmitter) and order code for "Sensor option", option
CX (conductivity measurement)
64 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Selection • Off
• On
Endress+Hauser 65
"Application" menu Proline Promag 10 Modbus RS485
Prerequisite Temperature measurement is only optionally available for Promag H 10 (5HBB): Under
order code for "Functionality", option D (extended transmitter) and order code for "Sensor
option", option CI (medium temperature measurement)
Fixed density
Navigation Application → Sensor → Low flow cut off → Low flow cut off
Description Select process variable for low flow cut off to activate low flow cut off.
Selection • Off
• Volume flow
• Mass flow
66 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Q 2 2
4
H
3
t
1 A 1 A
A0012887
Q Flow
t Time
H Hysteresis
A Low flow cut off active
1 Low flow cut off is activated
2 Low flow cut off is deactivated
3 On-value entered
4 Off-value entered
Description Enter off value to switch off low flow cut off. The off value is entered as a positive
hysteresis with respect to the on value.
Endress+Hauser 67
"Application" menu Proline Promag 10 Modbus RS485
New adjustment → 68
Description Switch empty pipe detection on or off. Switch on empty pipe detection to detect a partially
filled or empty measuring tube.
Selection • Off
• On
New adjustment
Description Select empty pipe or full pipe adjustment to perform a new adjustment. To adjust empty
pipe detection, perform the empty pipe adjustment first and then the full pipe adjustment.
Additional information:
The measuring device is pre-adjusted at production using water (approx. 300 μS/cm). For
liquids that deviate from this conductivity, a new empty pipe and full pipe adjustment must
be performed on site.
Selection • Cancel
• Empty pipe adjust
• Full pipe adjust
68 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Description Displays the value currently measured for empty pipe detection.
‣ Sensor adjustment
Installation direction → 69
Integration time → 70
Measuring period → 70
Installation direction
Endress+Hauser 69
"Application" menu Proline Promag 10 Modbus RS485
Integration time
User interface 1 to 65 ms
Measuring period
‣ Calibration
Nominal diameter → 70
Calibration factor → 71
Zero point → 71
Nominal diameter
70 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
User interface Character string comprising numbers, letters and special characters
Calibration factor
Description Displays the current calibration factor for the flow rate measuring sensor.
Additional information:
The factory setting for the calibration factor can be found on the sensor's nameplate.
Zero point
Description Displays the zero point correction value for the sensor.
NOTE
Users logged on in the Service role have write access!
Endress+Hauser 71
"Application" menu Proline Promag 10 Modbus RS485
ECC duration → 72
ECC interval → 73
ECC polarity → 73
Selection • Off
• On
ECC duration
Description Specify a timespan for recovery after a cleaning cycle has completed to prevent
interference with the signal outputs. The output signal values will be frozen for the
duration of the recovery.
72 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
ECC interval
Description Specify the duration of the interval between one cleaning cycle and the next.
ECC polarity
Endress+Hauser 73
"Application" menu Proline Promag 10 Modbus RS485
‣ Current output 1
Fixed current → 80
Failure current → 83
Output current 1 → 84
Selection • Off *
• Mass flow
• Conductivity *
• Corrected conductivity *
• Volume flow
• Temperature *
• Noise *
• Coil current shot time *
74 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Prerequisite A process variable is selected in the Process variable current output parameter
(→ 20).
Endress+Hauser 75
"Application" menu Proline Promag 10 Modbus RS485
I [mA]
20
4
0 Q
1 2
A0028084
I Current
Q Flow
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)
76 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
A0028091
Q Flow
t Time
I [mA]
t
A0028092
I Current
t Time
I [mA]
A0028093
I Current
t Time
Endress+Hauser 77
"Application" menu Proline Promag 10 Modbus RS485
I [mA]
A
t
S
S=A
A0028094
I Current
t Time
S Flow components saved
A Balancing of saved flow components
I [mA]
20
4
0 Q
0
1 2
A0028095
I Current
Q Flow
1 Lower range value (value assigned to 0/4 mA current)
2 Upper range value (value assigned to 20 mA current)
Q
1
+
0 t
-
2
a=
b=
A0028098
Q Flow
t Time
a Lower range value (value assigned to 0/4 mA current)
b Upper range value (value assigned to 20 mA current)
78 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
I [mA]
t
a=
b=
A0028100
I Current
t Time
I [mA]
t
S
S=A
A0028101
I Current
t Time
S Flow components saved
A Balancing of saved flow components
Description Select the current range for the measured value output and the upper and lower fault
condition signal level.
Additional information:
- The measured value range is specified in the "Lower range value output " parameter and
the "Upper range value output " parameter.
- If the measured value lies outside the scaled measured value range, diagnostic message
"441 Current output faulty" is generated.
- In the event of a device alarm, the current output adopts the behavior specified in the
"Failure behavior current output " parameter.
Endress+Hauser 79
"Application" menu Proline Promag 10 Modbus RS485
2 1 3
I [mA]
A0034351
Selection (current range for process value) Lower level for signal on Upper level for signal on
alarm alarm
Fixed current
Prerequisite In the Current range output parameter in the Current output 1 submenu, the Fixed
value option is selected.
Prerequisite In the Current range output parameter, one of the following options is selected:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
80 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Description Enter lower range value for the measured value range.
Additional information:
- Depending on the setting selected for the "Measuring mode current output " parameter,
the value specified for this parameter and the "Upper range value output " parameter must
have the same algebraic sign or not.
- As a rule, the lower range value is scaled to be lower than the upper range value. As a
result, the behavior of the current output is proportional to the process variable assigned.
If the lower range value is scaled to be higher than the upper range value, then the
behavior of the current output will be inversely proportional to the process variable
assigend.
Additional information Examples of the behavior, depending on the option selected in the Measuring mode
current output parameter (→ 75).
I [mA]
20
8
4
Q
-250 0 +750
1 1
A0013757
Q Flow
I Current
1 Measuring range is exceeded or undershot
The linear operational range of the measuring device is defined by the values entered for
the Lower range value output parameter (→ 21) and Upper range value output
parameter (→ 22), and by the selected current range.
I [mA]
20
4
0 Q
0
3 1 2
A0013758
Q Flow
I Current
1 Value assigned to 0/4 mA current
2 Forward flow
3 Reverse flow
Endress+Hauser 81
"Application" menu Proline Promag 10 Modbus RS485
The current output signal is independent of the direction of flow (absolute amount of the
measured variable). The values for the Lower range value output parameter (→ 21)
and Upper range value output parameter (→ 22) must have the same algebraic sign.
The value for the Upper range value output parameter (→ 22) (e.g. reverse flow)
corresponds to the mirrored value for the Upper range value output parameter (→ 22)
(e.g. forward flow).
Prerequisite In the Current range output parameter, one of the following options is selected:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
Description Enter upper range value for the measured value range.
Additional information Examples of the behavior, depending on the option selected in the Measuring mode
current output parameter: Lower range value output parameter (→ 21):
Prerequisite A process variable is selected in the Process variable current output parameter and one
of the following options is selected in the Current range output parameter:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
Description Enter time constant (PT1 element) to set the reaction time of the output signal to
fluctuations in the measured value caused by process conditions.
Additional information:
- The smaller the time constant the faster the output reacts to fluctuations in the
measured value.
- If the time constant is 0, damping is deactivated.
82 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Prerequisite A process variable is selected in the Process variable current output parameter and one
of the following options is selected in the Current range output parameter:
• 4...20 mA NE (3.8...20.5 mA)
• 4...20 mA US (3.9...20.8 mA)
• 4...20 mA (4... 20.5 mA)
Description Specify how the output should behave in the event of a device alarm.
Selection • Min.
• Max.
• Last valid value
• Actual value
• Fixed value
Failure current
Prerequisite In the Failure behavior current output parameter in the Current output 1 submenu, the
Fixed value option is selected.
Description Enter the value for the "Fixed value" option in the "Failure behavior current output "
parameter.
Endress+Hauser 83
"Application" menu Proline Promag 10 Modbus RS485
Output current
84 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
‣ Communication
‣ Modbus configuration → 85
‣ Modbus information → 89
‣ Modbus configuration
Bus address → 85
Baudrate → 86
Parity → 86
Byte order → 87
Telegram delay → 87
Failure mode → 87
Bus address
Endress+Hauser 85
"Application" menu Proline Promag 10 Modbus RS485
Baudrate
Parity
Selection • Odd
• Even
• None / 1 stop bit
• None / 2 stop bits
86 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
Byte order
Description Select byte transmission sequence. The transmission sequence must be coordinated with
the Modbus master.
Additional information:
The byte sequence is not standardized by the Modbus protocol. However, if the host
system and the measuring device do not use the same byte sequence, correct data
exchange is not possible.
Changing the byte sequence in the host system often requires extensive knowledge and
significant programming effort. This parameter can be used to keep the standard settings
of the host system, while adjusting the byte sequence on the measuring device by means
of trial and error. If it is not possible to achieve correct data exchange in this manner, then
it is the host system's settings for the byte sequence that must be modified.
Selection • 0-1-2-3
• 3-2-1-0
• 1-0-3-2
• 2-3-0-1
Telegram delay
Description Enter a delay time after which the measuring device replies to the request telegram of the
Modbus master. This is to enable communication with slow Modbus RS485 masters.
Failure mode
Description Select the preferred Modbus communication output mode for the measured value, when a
diagnostic event of the category specified in the "Assign diagnostic behavior" parameter
occurs.
Endress+Hauser 87
"Application" menu Proline Promag 10 Modbus RS485
88 Endress+Hauser
Proline Promag 10 Modbus RS485 "Application" menu
‣ Modbus information
Device ID → 89
Device revision → 89
Device ID
Device revision
Endress+Hauser 89
"System" menu Proline Promag 10 Modbus RS485
5 "System" menu
Overall device management and security settings – management of system settings and
adaption to operational requirements.
Navigation System
System
‣ Device management → 91
‣ User management → 93
‣ Connectivity → 96
‣ Date/time → 97
‣ Information → 99
‣ Display → 104
90 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
‣ Device management
Device tag → 91
Locking status → 91
Configuration counter → 92
Device reset → 92
Device tag
Description Enter a unique designation for the measuring point to be able to easily identify it within
the plant.
User entry Character string comprising numbers, letters and special characters (32)
Locking status
Description Indicates the write protection with the highest priority that is currently active.
Endress+Hauser 91
"System" menu Proline Promag 10 Modbus RS485
Configuration counter
Device reset
Description Reset the device configuration - either entirely or in part - to a defined state.
Selection • Cancel
• To delivery settings
• Restart device
• Restore S-DAT backup *
• Create T-DAT backup
• Restore T-DAT backup *
92 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
‣ User management
User role → 93
User role
Description Displays the role the user is currently logged on in. The role determines the user's access
rights for the parameters.
Additional information:
- Until a Maintenance code has been set in the "Define Maintenance code" parameter, all
users are automatically logged on in the Maintenance role. Once the Maintenance code
has been set, all users are automatically logged on in the Operator role.
- The access rights can be changed via the "Enter access code" parameter.
Endress+Hauser 93
"System" menu Proline Promag 10 Modbus RS485
Description For users logged on in the Operator role, enter the Maintenance code to change the access
status to Maintenance and disable write protection of parameters. For users logged on in
the Maintenance role, enter the Service code to change the access status to Service and
enable read and write access to Service parameters.
Description Enter the code provided by Endress+Hauser Technical Support to reset the Maintenance
code.
User entry Character string comprising numbers, letters and special characters (32)
94 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
Navigation System → User manag. → Def. Maint. code → Def. Maint. code
Description Specify an access code that is required to obtain the access rights for the Maintenance role.
Navigation System → User manag. → Def. Maint. code → Conf. Maint code
Description Confirm the access code entered for the Maintenance role.
Endress+Hauser 95
"System" menu Proline Promag 10 Modbus RS485
‣ Connectivity
‣ Bluetooth configuration → 96
‣ Bluetooth configuration
Bluetooth → 96
Communication established → 96
Bluetooth
Selection • Enable
• Disable
• Not available *
Communication established
User interface • No
• Yes
96 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
‣ Date/time
Set date/time → 97
Time format → 97
Time zone → 97
Set date/time
Description Set the date and local time. Every time the date or time is changed, a logbook entry is
created.
Time format
Selection • 24 h
• 12 h AM/PM
Time zone
Description Select the time zone. Every time the time zone is changed, a logbook entry is created.
Endress+Hauser 97
"System" menu Proline Promag 10 Modbus RS485
Selection
Other units
• UTC-12:00
• UTC-11:00
• UTC-10:00
• UTC-09:30
• UTC-09:00
• UTC-08:00
• UTC-07:00
• UTC-06:00
• UTC-05:00
• UTC-04:00
• UTC-03:30
• UTC-03:00
• UTC-02:00
• UTC-01:00
• UTC 00:00
• UTC+01:00
• UTC+02:00
• UTC+03:00
• UTC+03:30
• UTC+04:00
• UTC+04:30
• UTC+05:00
• UTC+05:30
• UTC+05:45
• UTC+06:00
• UTC+06:30
• UTC+07:00
• UTC+08:00
• UTC+08:45
• UTC+09:00
• UTC+09:30
• UTC+10:00
• UTC+10:30
• UTC+11:00
• UTC+12:00
• UTC+12:45
• UTC+13:00
• UTC+14:00
98 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
‣ Information
‣ Device → 99
‣ Device
Device name → 99
Manufacturer → 102
Device name
Endress+Hauser 99
"System" menu Proline Promag 10 Modbus RS485
User interface Character string comprising numbers, letters and special characters
Device tag
User interface Character string comprising numbers, letters and special characters
Serial number
Description Displays the serial number of the measuring device. The serial number can be used to
identify the measuring device and to retrieve further information on the measuring device,
such as the related documentation, via the Device Viewer or Operations app.
Additional information:
The serial number can also be found on the nameplate of the sensor and transmitter.
User interface Character string comprising numbers, letters and special characters
Order code
User interface Character string comprising numbers, letters and special characters
Firmware version
User interface Character string comprising numbers, letters and special characters
100 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
Description Displays the first, second and/or third part of the extended order code. Due to character
length restrictions, the extended order code is split into a maximum of 3 parameters. The
extended order code indicates for each feature in the product structure the selected option,
thereby uniquely identifying the device model.
Additional information:
The extended order code can also be found on the nameplate.
User interface Character string comprising numbers, letters and special characters
Description Displays the first, second and/or third part of the extended order code. Due to character
length restrictions, the extended order code is split into a maximum of 3 parameters. The
extended order code indicates for each feature in the product structure the selected option,
thereby uniquely identifying the device model.
Additional information:
The extended order code can also be found on the nameplate.
User interface Character string comprising numbers, letters and special characters
Description Displays the first, second and/or third part of the extended order code. Due to character
length restrictions, the extended order code is split into a maximum of 3 parameters. The
extended order code indicates for each feature in the product structure the selected option,
thereby uniquely identifying the device model.
Additional information:
The extended order code can also be found on the nameplate.
User interface Character string comprising numbers, letters and special characters
ENP version
Endress+Hauser 101
"System" menu Proline Promag 10 Modbus RS485
User interface Character string comprising numbers, letters and special characters
Manufacturer
User interface Character string comprising numbers, letters and special characters
Firmware version
‣ Display module
102 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
Firmware version
Endress+Hauser 103
"System" menu Proline Promag 10 Modbus RS485
‣ Display
Language → 104
Brightness → 107
Language
Selection • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык (Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• ( الْعَرَبيّةArabic) *
• Bahasa Indonesia
104 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
• ภาษาไทย (Thai) *
• tiếng Việt (Vietnamese)
• čeština (Czech)
Value 1 display
Description Select the measured value that is displayed first on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Value 2 display
Description Select the measured value that is shown second on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • None
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
• Totalizer 1
• Totalizer 2
• Totalizer 3
Endress+Hauser 105
"System" menu Proline Promag 10 Modbus RS485
Value 3 display
Description Select the measured value that is shown third on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • None
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
• Totalizer 1
• Totalizer 2
• Totalizer 3
Value 4 display
Description Select the measured value that is shown fourth on the local display.
Additional information:
The applicable unit of measure is specified in the "System units" submenu.
Selection • None
• Volume flow
• Mass flow
• Conductivity *
• Corrected conductivity *
• Temperature *
• Totalizer 1
• Totalizer 2
• Totalizer 3
Display damping
Description Enter time constant (PT1 element) to set reaction time of the display to fluctuations in the
measured value.
Additional information:
- The smaller the time constant the faster the display reacts to fluctuations in the
measured value.
- If the time constant is set to 0, damping is deactivated.
106 Endress+Hauser
Proline Promag 10 Modbus RS485 "System" menu
Rotation display
Description Select rotation angle of the display text to optimize local display readability.
Selection • Auto
• 0 degree
• 90 degree
• 180 degree
• 270 degree
Brightness
Color scheme
Selection • Light
• Dark
Endress+Hauser 107
"System" menu Proline Promag 10 Modbus RS485
‣ Software configuration
Activate SW option
Description Enter application package code or code of the functionality ordered separately to activate
it.
Additional information:
- If a measuring device was ordered with an add-on software option, the activation code is
programmed into the measuring device ex factory.
- After entering the activation code: Check whether the new software option is displayed
in the "Software option overview" parameter and therefore active.
NOTE
If an an invalid code is entered the software options that have already been activated are
invalidated!
Before entering a new activation code: Create a record of the existing activation code.
Description Displays all software options included in the order ex factory or ordered at a later date that
have been enabled via the operating interface.
Additional information:
If a new software option is not displayed after entering the activation code, the code
entered was inaccurate or invalid. In this case, contact the appropriate Endress+Hauser
sales organization to activate the software option.
108 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
6.1 Notes
Name of parameter Indicated in decimal • Float length = Possible type of access Selection Page number information
numerical format 4 byte to parameter: List of the individual and cross-reference to the
• Integer length = • Read access via options for the standard parameter
2 byte function codes 03, parameter description
• String length, 04 or 23 • Option 1
depending on • Write access via • Option 2
parameter function codes 06, • Option 3 (+)
16 or 23
• Factory
setting
highlighted in
bold
• (+) = Factory
setting
depends on
country, order
options or
device
settings
Input
Input range for the
parameter
NOTICE
If non-volatile device parameters are modified via the MODBUS RS485 function codes
06, 16 or 23, the change is saved in the EEPROM of the measuring device.
The number of writes to the EEPROM is technically restricted to a maximum of 1 million.
‣ Make sure to comply with this limit since, if it is exceeded, data loss and measuring
device failure will result.
‣ Avoid constantly writing non-volatile device parameters via the MODBUS RS485.
Endress+Hauser 109
Modbus RS485 register information Proline Promag 10 Modbus RS485
Function code Access type Register in accordance with "Modbus Register in accordance with
Applications Protocol Specification" "Modicon Modbus Protocol
Reference Guide"
Guidance → 119
‣ Commissioning → 119
110 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Diagnostics → 125
Endress+Hauser 111
Modbus RS485 register information Proline Promag 10 Modbus RS485
Timestamp → 125
Timestamp → 125
‣ Simulation → 126
‣ Properties → 126
‣ Sensor → 126
‣ Electronics → 127
112 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
‣ Configuration → 127
‣ Process → 127
Application → 128
Conductivity → 128
Temperature → 128
‣ Totalizer → 128
Endress+Hauser 113
Modbus RS485 register information Proline Promag 10 Modbus RS485
‣ Totalizers → 132
‣ Totalizer 1 to n → 132
‣ Sensor → 133
114 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
‣ Calibration → 134
Endress+Hauser 115
Modbus RS485 register information Proline Promag 10 Modbus RS485
‣ Communication → 135
Baudrate → 135
Parity → 135
116 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Device ID → 136
System → 136
‣ Connectivity → 137
Bluetooth → 137
‣ Date/time → 137
Endress+Hauser 117
Modbus RS485 register information Proline Promag 10 Modbus RS485
‣ Information → 138
‣ Device → 138
Manufacturer → 138
‣ Display → 139
Language → 139
118 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Brightness → 140
"Commissioning" wizard
Parameter Register Data type Access Selection / User entry / User interface →
Device tag 29243 to 29258 String Read / Write Character string comprising numbers, 8
letters and special characters (32)
Serial number 7003 to 7008 String Read Character string comprising numbers, 8
letters and special characters
Firmware version 7277 to 7280 String Read Character string comprising numbers, 8
letters and special characters
Device name 7263 to 7270 String Read Character string comprising numbers, 8
letters and special characters
Endress+Hauser 119
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
120 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Parameter Register Data type Access Selection / User entry / User interface →
71 = gal/d (imp)
72 = Mgal/s (imp)
73 = Mgal/min (imp)
74 = Mgal/h (imp)
75 = Mgal/d (imp)
76 = bbl/s (imp;beer)
77 = bbl/min (imp;beer)
78 = bbl/h (imp;beer)
79 = bbl/d (imp;beer)
80 = bbl/s (imp;oil)
81 = bbl/min (imp;oil)
82 = bbl/h (imp;oil)
83 = bbl/d (imp;oil)
88 = kgal/s (us)
89 = kgal/min (us)
90 = kgal/h (us)
91 = kgal/d (us)
92 = MMft³/s
93 = MMft³/min
94 = MMft³/h
96 = Mft³/d
Endress+Hauser 121
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
122 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Parameter Register Data type Access Selection / User entry / User interface →
Totalizer operation mode 2605 Integer Read / Write 0 = Net flow total 14
1 = Forward flow total
2 = Reverse flow total
On value low flow cutoff 5138 to 5139 Float Read / Write Positive floating-point number 18
Off value low flow cutoff 5104 to 5105 Float Read / Write 0 to 100.0 % 18
Empty pipe adjust value 2181 to 2182 Float Read Positive floating-point number 19
Full pipe adjust value 2832 to 2833 Float Read Positive floating-point number 19
Endress+Hauser 123
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
Current range output 5923 Integer Read / Write 0 = 4...20 mA (4... 20.5 mA) 20
1 = 4...20 mA US (3.9...20.8 mA)
2 = 4...20 mA NE (3.8...20.5 mA)
4 = Fixed value
Lower range value output 6195 to 6196 Float Read / Write Signed floating-point number 21
Upper range value output 5915 to 5916 Float Read / Write Signed floating-point number 22
Damping current output 5903 to 5904 Float Read / Write 0.0 to 999.9 s 23
124 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
Timestamp 29726 String Read Days (d), hours (h), minutes (m), seconds 30
(s)
Timestamp 29715 String Read Days (d), hours (h), minutes (m), seconds 31
(s)
Endress+Hauser 125
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
Operating time from restart 2624 String Read Days (d), hours (h), minutes (m), seconds 31
(s)
Operating time 2631 String Read Days (d), hours (h), minutes (m), seconds 31
(s)
"Simulation" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Process variable value 6814 to 6815 Float Read / Write Signed floating-point number 36
Current output value 5995 to 5996 Float Read / Write 3.59 to 21.5 mA 36
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
126 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
"Electronics" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Assign behavior of diagnostic no. 27632 Integer Read / Write 1 = Logbook entry only 41
230 2 = Warning
3 = Alarm
Assign behavior of diagnostic no. 28535 Integer Read / Write 1 = Logbook entry only 41
231 2 = Warning
3 = Alarm
"Configuration" submenu
Parameter Register Data type Access Selection / User entry / User interface →
"Process" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Endress+Hauser 127
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
"Totalizer" submenu
Parameter Register Data type Access Selection / User entry / User interface →
128 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Parameter Register Data type Access Selection / User entry / User interface →
Endress+Hauser 129
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
69 = gal/min (imp)
70 = gal/h (imp)
71 = gal/d (imp)
72 = Mgal/s (imp)
73 = Mgal/min (imp)
74 = Mgal/h (imp)
75 = Mgal/d (imp)
76 = bbl/s (imp;beer)
77 = bbl/min (imp;beer)
78 = bbl/h (imp;beer)
79 = bbl/d (imp;beer)
80 = bbl/s (imp;oil)
81 = bbl/min (imp;oil)
82 = bbl/h (imp;oil)
83 = bbl/d (imp;oil)
88 = kgal/s (us)
89 = kgal/min (us)
90 = kgal/h (us)
91 = kgal/d (us)
92 = MMft³/s
93 = MMft³/min
94 = MMft³/h
96 = Mft³/d
130 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Parameter Register Data type Access Selection / User entry / User interface →
Endress+Hauser 131
Modbus RS485 register information Proline Promag 10 Modbus RS485
"Totalizers" submenu
"Totalizer handling" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
Totalizer operation mode 1: 2605 Integer Read / Write 0 = Net flow total 60
2: 2805 1 = Forward flow total
3: 3005 2 = Reverse flow total
Preset value 1 to n 1: 2590 to 2591 Float Read / Write Signed floating-point number 61
2: 2592 to 2593
3: 2594 to 2595
132 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
"Sensor" submenu
"Process parameters" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Conductivity temperature coefficient 2886 to 2887 Float Read / Write Signed floating-point number 65
Fixed density 2830 to 2831 Float Read / Write Positive floating-point number 66
Parameter Register Data type Access Selection / User entry / User interface →
On value low flow cutoff 5138 to 5139 Float Read / Write Positive floating-point number 67
Off value low flow cutoff 5104 to 5105 Float Read / Write 0 to 100.0 % 67
Parameter Register Data type Access Selection / User entry / User interface →
Empty pipe adjust value 2181 to 2182 Float Read Positive floating-point number 68
Full pipe adjust value 2832 to 2833 Float Read Positive floating-point number 69
Measured value EPD 2298 to 2299 Float Read Positive floating-point number 69
Parameter Register Data type Access Selection / User entry / User interface →
Endress+Hauser 133
Modbus RS485 register information Proline Promag 10 Modbus RS485
"Calibration" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Nominal diameter 2048 to 2057 String Read Character string comprising numbers, 70
letters and special characters
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
*
Process variable current output 5927 Integer Read / Write 0 = Off 74
1 = Volume flow
4 = Conductivity *
7 = Temperature *
9 = Corrected conductivity *
11 = Mass flow
40 = Noise *
41 = Coil current shot time *
Measuring mode current output 5899 Integer Read / Write 0 = Forward flow 75
2 = Reverse flow compensation
13 = Forward/Reverse flow *
Current range output 5923 Integer Read / Write 0 = 4...20 mA (4... 20.5 mA) 79
1 = 4...20 mA US (3.9...20.8 mA)
2 = 4...20 mA NE (3.8...20.5 mA)
4 = Fixed value
Lower range value output 6195 to 6196 Float Read / Write Signed floating-point number 80
Upper range value output 5915 to 5916 Float Read / Write Signed floating-point number 82
Damping current output 5903 to 5904 Float Read / Write 0.0 to 999.9 s 82
134 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
Parameter Register Data type Access Selection / User entry / User interface →
"Communication" submenu
"Modbus configuration" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
Endress+Hauser 135
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
Device tag 29243 to 29258 String Read / Write Character string comprising numbers, 91
letters and special characters (32)
Parameter Register Data type Access Selection / User entry / User interface →
Reset Maintenance code 8880 to 8895 String Read / Write Character string comprising numbers, 94
letters and special characters (32)
Parameter Register Data type Access Selection / User entry / User interface →
136 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
"Connectivity" submenu
"Bluetooth configuration" submenu
Parameter Register Data type Access Selection / User entry / User interface →
"Date/time" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Endress+Hauser 137
Modbus RS485 register information Proline Promag 10 Modbus RS485
"Information" submenu
"Device" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Device name 7263 to 7270 String Read Character string comprising numbers, 99
letters and special characters
Device tag 2026 to 2041 String Read Character string comprising numbers, 100
letters and special characters
Serial number 7003 to 7008 String Read Character string comprising numbers, 100
letters and special characters
Order code 2058 to 2067 String Read Character string comprising numbers, 100
letters and special characters
Firmware version 7277 to 7280 String Read Character string comprising numbers, 100
letters and special characters
Extended order code 1 2212 to 2221 String Read Character string comprising numbers, 101
letters and special characters
Extended order code 2 2222 to 2231 String Read Character string comprising numbers, 101
letters and special characters
Extended order code 3 2232 to 2241 String Read Character string comprising numbers, 101
letters and special characters
ENP version 4003 to 4010 String Read Character string comprising numbers, 101
letters and special characters
Manufacturer 8001 to 8016 String Read Character string comprising numbers, 102
letters and special characters
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
138 Endress+Hauser
Proline Promag 10 Modbus RS485 Modbus RS485 register information
"Display" submenu
Parameter Register Data type Access Selection / User entry / User interface →
Display damping 27602 to 27603 Float Read / Write 0.0 to 999.9 s 106
Endress+Hauser 139
Modbus RS485 register information Proline Promag 10 Modbus RS485
Parameter Register Data type Access Selection / User entry / User interface →
Parameter Register Data type Access Selection / User entry / User interface →
Software option overview 2902 Integer Read 16384 = Heartbeat Monitoring 108
32768 = Heartbeat Verification
140 Endress+Hauser
Proline Promag 10 Modbus RS485 Index
Index
A Date / Time (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 97
Activate SW option (Parameter) . . . . . . . . . . . . . . . 108 Define access code (Wizard) . . . . . . . . . . . . . . . . . . . 95
Active diagnostics (Submenu) . . . . . . . . . . . . . . . . . . 30 Density unit (Parameter) . . . . . . . . . . . . . . . . . . . 12, 56
Actual diagnostics (Parameter) . . . . . . . . . . . . . . . . . 30 Device (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Alarm delay (Parameter) . . . . . . . . . . . . . . . . . . . . . . 39 Device alarm simulation (Parameter) . . . . . . . . . . . . . 37
Application (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Device management (Submenu) . . . . . . . . . . . . . . . . . 91
Assign behavior of diagnostic no. 043 (Parameter) . . . 40 Device name (Parameter) . . . . . . . . . . . . . . . . . . . . 8, 99
Assign behavior of diagnostic no. 230 (Parameter) . . . 41 Device reset (Parameter) . . . . . . . . . . . . . . . . . . . . . . 92
Assign behavior of diagnostic no. 231 (Parameter) . . . 41 Device tag (Parameter) . . . . . . . . . . . . . . . . . 8, 91, 100
Assign behavior of diagnostic no. 302 (Parameter) . . . 41 Diagnostic configuration (Submenu) . . . . . . . . . . . . . 39
Assign behavior of diagnostic no. 376 (Parameter) . . . 41 Diagnostic event category (Parameter) . . . . . . . . . . . . 37
Assign behavior of diagnostic no. 377 (Parameter) . . . 42 Diagnostic event simulation (Parameter) . . . . . . . . . . 37
Assign behavior of diagnostic no. 441 (Parameter) . . . 43 Diagnostic list (Submenu) . . . . . . . . . . . . . . . . . . . . . 32
Assign behavior of diagnostic no. 832 (Parameter) . . . 44 Diagnostic settings (Submenu) . . . . . . . . . . . . . . . . . . 39
Assign behavior of diagnostic no. 833 (Parameter) . . . 44 Diagnostics (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Assign behavior of diagnostic no. 834 (Parameter) . . . 45 Diagnostics 1 (Parameter) . . . . . . . . . . . . . . . . . . . . . 32
Assign behavior of diagnostic no. 835 (Parameter) . . . 45 Diagnostics 2 (Parameter) . . . . . . . . . . . . . . . . . . . . . 33
Assign behavior of diagnostic no. 842 (Parameter) . . . 46 Diagnostics 3 (Parameter) . . . . . . . . . . . . . . . . . . . . . 33
Assign behavior of diagnostic no. 937 (Parameter) . . . 46 Diagnostics 4 (Parameter) . . . . . . . . . . . . . . . . . . . . . 33
Assign behavior of diagnostic no. 938 (Parameter) . . . 47 Diagnostics 5 (Parameter) . . . . . . . . . . . . . . . . . . . . . 34
Assign behavior of diagnostic no. 961 (Parameter) . . . 47 Display (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . 104
Assign behavior of diagnostic no. 962 (Parameter) . . . 48 Display damping (Parameter) . . . . . . . . . . . . . . . 26, 106
Assign process variable (Parameter) . . . . . . . . . . . 13, 59 Display module (Submenu) . . . . . . . . . . . . . . . . . . . 102
Assign simulation process variable (Parameter) . . . . . 35 Document
Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
B
Bluetooth (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 96 E
Bluetooth configuration (Submenu) . . . . . . . . . . . . . . 96 ECC duration (Parameter) . . . . . . . . . . . . . . . . . . . . . 72
Brightness (Parameter) . . . . . . . . . . . . . . . . . . . . . . 107 ECC interval (Parameter) . . . . . . . . . . . . . . . . . . . . . . 73
ECC polarity (Parameter) . . . . . . . . . . . . . . . . . . . . . . 73
C ECC recovery time (Parameter) . . . . . . . . . . . . . . . . . 72
Calibration (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 70 Electrode cleaning cycle (Parameter) . . . . . . . . . . . . . 72
Calibration factor (Parameter) . . . . . . . . . . . . . . . . . . 71 Electrode cleaning cycle (Submenu) . . . . . . . . . . . . . . 72
Color scheme (Parameter) . . . . . . . . . . . . . . . . . . . . 107 Electronics (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 40
Commissioning (Wizard) Empty pipe adjust value (Parameter) . . . . . . . . . . . . . 19
. . . . . . . . . . . . . . . 6, 8, 9, 13, 15, 16, 17, 19, 20, 25, 27 Empty pipe detection (Parameter) . . . . . . . . . . . . . . . 19
Communication (Submenu) . . . . . . . . . . . . . . . . . . . . 85 ENP version (Parameter) . . . . . . . . . . . . . . . . . . . . . 101
Communication established (Parameter) . . . . . . . . . . 96 Enter access code (Parameter) . . . . . . . . . . . . . . . . . . 94
Conductivity (Parameter) . . . . . . . . . . . . . . . . . . . . . . 50 Extended order code 1 (Parameter) . . . . . . . . . . . . . 101
Conductivity calibration factor (Parameter) . . . . . . . . 71 Extended order code 2 (Parameter) . . . . . . . . . . . . . 101
Conductivity damping time (Parameter) . . . . . . . . . . . 65 Extended order code 3 (Parameter) . . . . . . . . . . . . . 101
Conductivity measurement (Parameter) . . . . . . . . . . . 64
Conductivity temperature coefficient (Parameter) . . . . 65 F
Conductivity unit (Parameter) . . . . . . . . . . . . . . . 12, 56 Failure behavior current output (Parameter) . . . . . 23, 83
Configuration (Submenu) . . . . . . . . . . . . . . . . . . . . . 43 Failure current (Parameter) . . . . . . . . . . . . . . . . . 24, 83
Configuration counter (Parameter) . . . . . . . . . . . . . . 92 Failure mode (Parameter) . . . . . . . . . . . . . . . . . . 15, 61
Connectivity (Submenu) . . . . . . . . . . . . . . . . . . . . . . 96 Firmware version (Parameter) . . . . . . . 8, 100, 102, 103
Control Totalizer 1 to n (Parameter) . . . . . . . . . . . . . . 60 Fixed current (Parameter) . . . . . . . . . . . . . . . . . . 23, 80
Corrected conductivity (Parameter) . . . . . . . . . . . . . . 50 Fixed density (Parameter) . . . . . . . . . . . . . . . . . . . . . 66
Current output 1 (Submenu) . . . . . . . . . . . . . . . . . . . 74 Flow damping (Parameter) . . . . . . . . . . . . . . . . . 15, 64
Current output 1 simulation (Parameter) . . . . . . . . . . 36 Flow override (Parameter) . . . . . . . . . . . . . . . . . . . . . 64
Current output value (Parameter) . . . . . . . . . . . . . . . 36 Full pipe adjust value (Parameter) . . . . . . . . . . . . . . . 19
Current range output (Parameter) . . . . . . . . . . . . 20, 79
G
D Guidance (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Damping current output (Parameter) . . . . . . . . . . 23, 82
Endress+Hauser 141
Index Proline Promag 10 Modbus RS485
H S
Heartbeat Technology (Submenu) . . . . . . . . . . . . . . . 38 Sensor (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 40, 63
Sensor adjustment (Submenu) . . . . . . . . . . . . . . . . . . 69
I Sensor electronic module (ISEM) (Submenu) . . . . . . 102
Information (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 99 Serial number (Parameter) . . . . . . . . . . . . . . . . . . 8, 100
Installation direction (Parameter) . . . . . . . . . . . . . . . 69 Set date/time (Parameter) . . . . . . . . . . . . . . . . . . . . . 28
Integration time (Parameter) . . . . . . . . . . . . . . . . . . . 70 Simulation (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 35
Software configuration (Submenu) . . . . . . . . . . . . . . 108
L Software option overview (Parameter) . . . . . . . . . . . 108
Language (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 104 Submenu
Locking status (Parameter) . . . . . . . . . . . . . . . . . . . . 91 Active diagnostics . . . . . . . . . . . . . . . . . . . . . . . . 30
Low flow cut off (Parameter) . . . . . . . . . . . . . . . . 17, 66 Bluetooth configuration . . . . . . . . . . . . . . . . . . . . 96
Low flow cut off (Submenu) . . . . . . . . . . . . . . . . . . . . 66 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Lower range value output (Parameter) . . . . . . . . . 21, 80 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
M Connectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Manufacturer (Parameter) . . . . . . . . . . . . . . . . . . . . 102 Current output 1 . . . . . . . . . . . . . . . . . . . . . . . . . 74
Mass flow (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 49 Date / Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Mass flow unit (Parameter) . . . . . . . . . . . . . . . . . 11, 55 Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Mass unit (Parameter) . . . . . . . . . . . . . . . . . . . . . 11, 55 Device management . . . . . . . . . . . . . . . . . . . . . . . 91
Measured values (Submenu) . . . . . . . . . . . . . . . . . . . 49 Diagnostic configuration . . . . . . . . . . . . . . . . . . . . 39
Measuring mode current output (Parameter) . . . . . . . 75 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Measuring period (Parameter) . . . . . . . . . . . . . . . . . . 70 Diagnostic settings . . . . . . . . . . . . . . . . . . . . . . . . 39
Menu Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Display module . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Electrode cleaning cycle . . . . . . . . . . . . . . . . . . . . 72
Guidance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Heartbeat Technology . . . . . . . . . . . . . . . . . . . . . 38
Modbus configuration (Submenu) . . . . . . . . . . . . . . . 85 Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Modbus data map (Submenu) . . . . . . . . . . . . . . . . . . 88 Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Modbus information (Submenu) . . . . . . . . . . . . . . . . 89 Measured values . . . . . . . . . . . . . . . . . . . . . . . . . 49
Modbus configuration . . . . . . . . . . . . . . . . . . . . . 85
N
Modbus data map . . . . . . . . . . . . . . . . . . . . . . . . . 88
Nominal diameter (Parameter) . . . . . . . . . . . . . . . . . 70
Modbus information . . . . . . . . . . . . . . . . . . . . . . . 89
O Partially filled pipe detection . . . . . . . . . . . . . . . . . 67
Off value low flow cutoff (Parameter) . . . . . . . . . . 18, 67 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
On value low flow cutoff (Parameter) . . . . . . . . . . 18, 67 Process parameters . . . . . . . . . . . . . . . . . . . . . . . 63
Operating time (Parameter) . . . . . . . . . . . . . . . . . . . . 31 Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Operating time from restart (Parameter) . . . . . . . . . . 31 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40, 63
Order code (Parameter) . . . . . . . . . . . . . . . . . . . . . . 100 Sensor adjustment . . . . . . . . . . . . . . . . . . . . . . . . 69
Output current 1 (Parameter) . . . . . . . . . . . . . . . . . . 84 Sensor electronic module (ISEM) . . . . . . . . . . . . . 102
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
P Software configuration . . . . . . . . . . . . . . . . . . . . 108
Partially filled pipe detection (Submenu) . . . . . . . . . . 67 System units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Preset value 1 to n (Parameter) . . . . . . . . . . . . . . . . . 61 Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Pressure shock suppression (Parameter) . . . . . . . . . . 16 Totalizer 1 to n . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Previous diagnostics (Parameter) . . . . . . . . . . . . . . . . 30 Totalizer handling . . . . . . . . . . . . . . . . . . . . . . . . 58
Process (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Process parameters (Submenu) . . . . . . . . . . . . . . . . . 63 User management . . . . . . . . . . . . . . . . . . . . . . . . 93
Process variable current output (Parameter) . . . . . 20, 74 System (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Process variable value (Parameter) . . . . . . . . . . . . . . . 36 System units (Submenu) . . . . . . . . . . . . . . . . . . . . . . 53
Properties (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 39
T
R Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Reset all totalizers (Parameter) . . . . . . . . . . . . . . . . . 58 Temperature (Parameter) . . . . . . . . . . . . . . . . . . . . . 51
Reset Maintenance code (Parameter) . . . . . . . . . . . . . 94 Temperature damping time (Parameter) . . . . . . . . . . 66
Rotation display (Parameter) . . . . . . . . . . . . . . . . . . 107 Temperature unit (Parameter) . . . . . . . . . . . . . . . 12, 56
Time format (Parameter) . . . . . . . . . . . . . . . . . . . . . . 27
Time zone (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 27
142 Endress+Hauser
Proline Promag 10 Modbus RS485 Index
U
Unit totalizer 1 to n (Parameter) . . . . . . . . . . . . . 13, 59
Upper range value output (Parameter) . . . . . . . . . 22, 82
User management (Submenu) . . . . . . . . . . . . . . . . . . 93
User role (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 93
V
Value 1 display (Parameter) . . . . . . . . . . . . . . . . 25, 105
Value 2 display (Parameter) . . . . . . . . . . . . . . . . 25, 105
Value 3 display (Parameter) . . . . . . . . . . . . . . . . 26, 106
Value 4 display (Parameter) . . . . . . . . . . . . . . . . 26, 106
Volume flow (Parameter) . . . . . . . . . . . . . . . . . . . . . 50
Volume flow unit (Parameter) . . . . . . . . . . . . . . . . 9, 53
Volume unit (Parameter) . . . . . . . . . . . . . . . . . . . 10, 54
W
Wizard
Commissioning 6, 8, 9, 13, 15, 16, 17, 19, 20, 25, 27
Define access code . . . . . . . . . . . . . . . . . . . . . . . . 95
Z
Zero point (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 71
Endress+Hauser 143
*71575137*
71575137
[Link]