Lecture 03
Performance Characteristics for
Measurement and
Instrumentation System
Lecture03 SME3242 Instrumentation 1
3 classifications to define the performance
of measurement system: operational, static,
dynamic
Operational characteristic
Range
Span
Sensitivity
Resolution
Dead band/threshold
Lecture03 SME3242 Instrumentation 2
Range
- will give the minimum and maximum range
Span
- the difference between the maximum and
minimum range
Eg: A specification of a thermometer reads as
follows:
Range and subdivision oC - -0.5 to +40.5 0.1
Min is –0.50C and max is 40.50C
Span = 410C
Lecture03 SME3242 Instrumentation 3
Sensitivity
- ratio of a change in output to the change in
input which causes it at steady-states condition
Eg. : A galvanometer has a sensitivity of 17mm/A
for a 1A input display, a light spot moving
across the scale shows a movement of an
index of 17 mm
Lecture03 SME3242 Instrumentation 4
Resolution
The least incremental value of input or output
that can be detected
Dead band / Threshold
The largest range of values of a measured
variable to which the instrument does not
respond
Lecture03 SME3242 Instrumentation 5
Static characteristic:
Error
Accuracy
Precision
Repeatability
Linearity
Hysteresis
Calibration
Lecture03 SME3242 Instrumentation 6
Error
- Error is the difference between the true value Yn and
instrument reading Xn
e = Yn – Xn
- Types of errors: systematic error (bias error) and
random error (precision error)
Bias (systematic) error : consistent and repeatable
error
Bias error = average readings – true value
Lecture03 SME3242 Instrumentation 7
Random error : the lack of repeatability in the
output of the measuring system
Random error = reading – average reading
Lecture03 SME3242 Instrumentation 8
Accuracy
- Ability of the system to respond to a true value
- Limit of error of a measuring device under certain operating
conditions and can appear is several forms:
1. Measured variable: (error limit)
Ex.: If the accuracy is 0.20C, for a temperature reading of
30.10C the actual temperature lies between 29.90C and
30.30C
2. Percentage of full scale (FS),
Ex.: If the full scale is 10 A and the accuracy 1% f.s., the
accuracy is then 0.1 A
Lecture03 SME3242 Instrumentation 9
3. Percentage of instrument span,
Ex.: If the accuracy 3% span and the span for pressure
measurement is 20-50 psi, the accuracy is =0.03 (50-
20) = 0.9 psi
4. Given as percentage of actual reading,
Ex.: If the true value of the voltmeter is 2 V and the
voltmeter accuracy is 2%, the actual voltage lies
between 2.04 and 1.960C. The accuracy at this instance
is (2V 0.02) = 0.04 V
Lecture03 SME3242 Instrumentation 10
A temperature sensor has a span of 200C –2500C.
A measurement results in a value of 550C for
temperature. Specify the error if the accuracy is:
a) 0.5%FS
b) 0.75% of span
c) 0.8% of reading
What is the possible temperature in each case
Lecture03 SME3242 Instrumentation 11
Full scale = 2500C
0.5% of full scale = [(0.5/100)](2500C)
= 1.250C
Therefore, for a temperature reading of 550C the
actual temperature lies between 53.750C and
56.250C
Lecture03 SME3242 Instrumentation 12
Span= (250-20)0C = 2300C
0.75% of span= [(0.75/100)](2300C)
= 1.7250C
Therefore, for a temperature reading of 550C the
actual temperature lies between 53.2750C and
56.7250C
Lecture03 SME3242 Instrumentation 13
Read temperature = 550C
0.8% of reading = [(0.8/100)](550C)
= 0.440C
Therefore, for a temperature reading of 550C the
actual temperature lies between 54.560C and
55.440C
Lecture03 SME3242 Instrumentation 14
A temperature sensor has a transfer function of 5
mV/0C with an accuracy of 1%. Find the possible
range of the transfer function.
Lecture03 SME3242 Instrumentation 15
Transfer function error = [(1.0/100)](5 mV/0C)
= 0.05 mV/0C
Therefore, the possible range of the transfer
function is 4.95 mV/0C to 5.05 mV/0C
Lecture03 SME3242 Instrumentation 16
Suppose a reading of 27.5 mV results from the
sensor used in previous example (Example 1.9).
Find the temperature that could provide this
reading.
Lecture03 SME3242 Instrumentation 17
Note: output
transfer function =
input
output
i.e. input =
transfer function
Lecture03 SME3242 Instrumentation 18
Since the transfer function range is 4.95 mV/0C to
5.05 mV/0C, the possible input values range for an
output of 27.5 mV can be formulated as follows:
For transfer function = 4.95 mV/0C,
27.5 mV
input =
4.95 mV/0C
= 5.56 0C
Lecture03 SME3242 Instrumentation 19
For transfer function = 5.05 mV/0C,
27.5 mV
input =
5.05 mV/0C
= 5.45 0C
Therefore, for a output reading of 27.5 mV the input
lies between 5.450C and 5.560C
Lecture03 SME3242 Instrumentation 20
System accuracy is the consideration of overall accuracy of
many elements in a process-control loop to represent a
process variable.
Consider, a system with two transfer function that act on
dynamic variable to produce an output voltage shown in the
Figure below:
C
K K G G V V
Lecture03 SME3242 Instrumentation 21
The output can be described as:
V V = (K K) (G G)C
or
V V = KGC GCK KCG KGC
where,
V = output voltage
V = uncertainty in output voltage
K,G = nominal transfer functions
K, G = uncertainty in transfer functions
C = dynamic variable
Lecture03 SME3242 Instrumentation 22
Knowing that,
V = KGC
Therefore,
V = GCK KCG KGC
In term of fractional uncertainties (error per output),
V GCK KCG KGC
=
V KGC
V GCK KCG KGC
V
= KGC
KGC
KGC
Lecture03 SME3242 Instrumentation 23
V GCK KCG KGC
V
= KGC KGC KGC
The numerator of the last term, KG, is a product of two
small numbers and thus is really small. Therefore,
V K G
=
V K G
Hence, the percentage of output error,
V K G
x100% = x 100%
V K G
Lecture03 SME3242 Instrumentation 24
i.e.
V K G
x100% = x100% x 100%
V K G
or in general term,
System accuracy or percentage of output error is equals
to summation of all individual transfer function errors
(percentage)
Lecture03 SME3242 Instrumentation 25
The overall system accuracy as the root-mean-
square (rms):
V K G
2 2
=± +
V rms K G
Lecture03 SME3242 Instrumentation 26
Find the system accuracy of a flow process if the
transducer transfer function is 10 mV/(m3/s)
±1.5% and the signal-conditioning system-
transfer function is 1mA/mV ±0.05%
Lecture03 SME3242 Instrumentation 27
Using the equation,
V K G
x100% = x100% x 100%
V K G
Total of system error = 1.5 % + 0.05 %
= 1.55 %
= 0.0155 mA
Lecture03 SME3242 Instrumentation 28
Using the rms equation,
V K G
2 2
=± +
V rms K G
= ( 0.015) 2 + (0.005) 2
= ±0.0158 1.6%
Lecture03 SME3242 Instrumentation 29
Precision
The degree of exactness of which an
instrument is designed or intended to
performed
Significant figures convey actual information
regarding the magnitude and the
measurement precision of a quantity
The more the significant the figure, the
greater the precision of measurement
Lecture03 SME3242 Instrumentation 30
Lecture03 SME3242 Instrumentation 31
Lecture03 SME3242 Instrumentation 32
Repeatability
- The ability of the system to display the same
output for a series of applications of the same
input signal, under the same operating
conditions
Linearity
- The output reading of the measurement is
linearly proportional to the quantity being
measured
- when linear relationship occurs, straight line
eq. is used cm = mc + c0
Lecture03 SME3242 Instrumentation 33
FIGURE 1.24 Comparison of an actual curve and its best-fit
straight line, where the maximum deviation is 5% f.s.
Copyright ©2006 by Pearson
Education, Inc.
Curtis Johnson Upper Saddle River, New Jersey
Process Control Instrumentation Technology, 8e] 07458
All rights reserved.
Lecture03 SME3242 Instrumentation 34
A sensor resistance changes linearly from 100 to
180 as temperature changes from 200 to 1200C.
Find a linear equation relating resistance and
temperature
Lecture03 SME3242 Instrumentation 35
Hysteresis
- different reading may be obtained if the
variable was increasing prior to taking the
reading and if the variable was decreasing
- causes: friction, mechanical flexure of
internal part, electrical capacitance
Lecture03 SME3242 Instrumentation 36
FIGURE 1.23 Hysteresis is a predictable error resulting from
differences in the transfer function as the input variable increases or
decreases.
Copyright ©2006 by Pearson
Education, Inc.
Curtis Johnson Upper Saddle River, New
Process Control Instrumentation Technology, 8e] Jersey 07458
All rights reserved.
Lecture03 SME3242 Instrumentation 37
Lecture03 SME3242 Instrumentation 38
Calibration
- process of checking a measuring system
against a standard reading
- purposes:
To ensure readings from an instrument are
consistent with other measurements
To determine the accuracy of the instrument
readings
To establish the reliability of the instrument
i.e. that it can be trusted
Lecture03 SME3242 Instrumentation 39
Lecture03 SME3242 Instrumentation 40