GC-17 v2 Service Manual GC17
GC-17 v2 Service Manual GC17
GC47A Ver.2
SERVICE MANUAL
221-30108
GC-17A Ver. 2
SERVICE MANUAL
SHIMADZU CORPORATION
ANALYTICAL INSTRUMENTS DIVISION
KYOTO, JAPAN
Copyright 0 Shimadzu Corporation 1995. All rights are reserved, including those to
reproduce this publication or parts there of in any form without permission in writing from
Shimadzu Corporation.
Information in this publication is subject to change without notice and does not represent
a commitment on the part of the vendor.
Any errors or omissions which may have occurred in this publication despite the utmost
care taken in its production will be corrected as soon as possible, but not necessarily im-
mediately upon detection.
Shimadzu Corporation will continue to supply original Shimadzu maintenance parts associat-
ed with a given product for a period up to 10 years from the cessation of production of that
product. Please be informed that after this period of time, Shimadzu Corporation cannot guar-
antee supply of such original maintenance parts. However, following discontinuation of a
product, the period of availability of maintenance parts which have been produced on a sub-
contract basis is up to the discretion of the concerned sub-contracting company.
Chapter 1 GG17A Ver. 2 Series .................................................................... I-1
1.1 Configuration of GC-17A Ver.2 Series .............................................................. l-2
1.2 Accessories of GC-17A Ver.2 ............................................................................ l-3
l-l
GC-17A Ver.2 SERVICE MANUAL m
1 GC-17A Ver. 2 1
GC-17AA Ver.2 without _ GC- 17AA V2 common part GC- 17AAT V2 GC- 17AA V2 common part
DET 221-44087-91 221-43799-91 221-44087-91
221-43825-91 -92 -92 -92
-92 -93 -93 -93
-93 -94 -94 -94
-94 -96 -96 -96
-96
GC- 17A V2 accessory TCD-17 V2
221-43312-91 22 l-42474-9 1
-92
GC- 17A V2 assembly parts -93
221-43693-91
GC- 17A V2 accessory
Flow controller make-up 221-43312-91
221-43329-91
GC-17A V2 assembly parts
22 l-43693-9 1
Flow controller make-uu
221-43329-91 L
GC-17AAF V2 GC- 17AA V2 common uart GC-17AA TF V2 - - GC- 17AA V2 common part
221-43305-91 221-44087-91 221-43810-91 221-44087-91
-92 I -92 -92 -92
-93 -93 -93 -93
-94 -94 -94 -94
-96 -96 -96 -96
FID-17 V2 linear - TCD-17 V2
221-43310-91 22 l-42474-9 1
-92 -92
-93 -93
-94
- Fid- 17 V2 linear
GC-17A V2 acessory ASSY 221-43310-91
221-43312-91 -92
GC-17A V2 assembly parts -93
221-43693-91 -94
GC-17AA FW V2 GC- 17AA V2 common part GC-1’7AATFWV2 - - GC- 17AA V2 common part
221-43304-91 221-44087-91 221-43809-91 T 221-44087-91
-92 -92 -92 -92
-93 -93 -93 -93
-94 -94 -94 -94
-96 -96 -96 -96
GC-17A V2 wide-range controller - TCD-17V2
221-43311-91 221-42474-91
-92 -92
-93 -93
-94 - FID-17 V2 wide-range controller
GC-17A V2 accessory Assy 221-43311-91
221-43312-91 -92
-93
GC-17A V2 assembly psrts -94
221-43693-91
> - GC- 17A V2 accessory Assy
Flow controller make-up 221-43312-91
221-43329-91
-91 1OOV type - GC-17A V2 assembly parts
221-43693-91
-92 115V 0
- Flow controller make-up
-93 220v *
221-43329-91
-94240v ” - Additional piping, TCD-17
-96230V * 221-42713-91
Accessories of GC-17A Ver.2 (P/N 221-43312-91)
221-09688-91 Pipe. AIR IN
221-l 5563-91 Graphite ferrule packed with 4 pcs. 221-43947-91 Holder capillary column, V2
2-1
GC-17A Ver.2 SERVICE MANUAL -
.
,:j’&-’
‘2 :>C+ Configuration of GC-17AA Ver.2
-.,-__ ..,
CC-17AA V2 GC-17A V2 over Column oven Assy Grounding cable Assy oven
221-43050-91 22 l-4305 1-9 1 221-41623-91
221-44087-91 -92 -92
-93 -93 Bottom plate Assy Power switch Assy
-94 -94 221-43052-91 221-43053-91
-96
Oven inner tank Assy Oven welding plate Assy
221-43054-91 221-43055-91
% To be continued
2.1 Configuration of GC-17AA Ver.2
SPL-17 v2
I- 221-42939-91
L
-93
-92 115V fi
-93 220v f/
-94240v *
-96 230V ”
2-3
GC-17A Ver.2 SERVICE MANUAL I
2.2.1 Configuration of GC47AA Ver.2
2-4
- GC-17A Ver.2 SERVICE MANUAL
2.2 Configuration of GC-17AA Ver.2
2-5
GC-17A Ver.2 SERVICE MANUAL -
2.2 Configuration of GC-17AA Ver.2
21 l-42903-01 Rear c
2-6
m GC-17A Ver.2 SERVICE MANUAL
2.2 Configuration of GC-17AA Ver.2
I”, -
- LL Aid.--.I1 I 7 I
2-8
m GC-17A Ver.2 SERVICE MANUAL
2.3 Electric Circuit of GC-17A Ver.2
(
(
i (
<
(
c
(
(
t
/;
<
(
C
.
c
C
I
.
J
- J3 - 320 - 34
I
2-15
GC-17A Ver.2 SERVICE MANUAL m
GC-17AA Ver.2
R4
---VA
IOKSI
BASE 2w
UNIT TLP633 332K
AC TAB 1 35 SW REGULATOR
ACC TAB 2 J6
INJ C J7
J&l
GND TAB 3
VALVE SPLl
VALVE SPL2
VALVE AIR
512
4 I
!h--?j 0 0 0
wz,i-fER FLOW I l=LOU L
Ii
i+biALyS~ S
R-V’ 0
2-17
GC-17A Ver.2 SERVICE MANUAL -
2.3 Electric Circuit of GC-17A Ver.2
/-
[
’ i
ii -
j /
1 !I
2-18
m GC-I-IA Ver.2 SERVICE MANUAL
2.3 Electric Circuit of G&l-/A Ver.2
AUX temperature control unit Installation manual for AUX temperature control unit
22 I-43345-91
2-19
GC-17A Ver.2 SERVICE MANUAL m
CNS (AUX4.5 LAMP)
R5 /
AUXPC 6
AUX3C 7
AUX4C 8 OVER IiEAT
.*ToA
AUX5C 9
AUXPS 11
AUXBS 12
I
AUX4S 13
AUX5S 14
AGND 15
‘CA
H2
,,/ 11, ,,,
I ! I ! I-!-- I
CN3 (AUX2) CN4 (AUXJ) CN5 (AUX4) CN6 (AUX5)
+4oooPPwc
R3 .
38.2 ’
3-,l
GC-17A Ver.2 SERVICE MANUAL m
3.1 .I Configuration of AFC Ver.2
AFC 17A V2 Assy
AFC 17A V2 Assy Front Assy Solenoid valve Assy Coil Assy
221-43234-91 221-43227-91------I 221-43221-92 221-43216-91
-92 -92
-93 -93 Needle valve
-94 -94 22 l-26995
-95 -95
Small solenoid valve ASS)- Relay Assy
AFC-SPL pipe Assy 22 I -43690-g 1 221-43996-91 (115V)
221-43233-91 -92 -92 (24OV)
-93 -93 (230V)
= PCB FLOW CONT 17A Assy -94 Not necessary for
g 22 1-43096-g 1 -95 1OOV and 220V.
s
0 TFC Assy Solenoil valve Assy
g 221-43228-91 221-43221-91 Coil Assy
G - Capillary Assy
i4 221-43220-91
(3
- Pressure gauge sensor Assy
221-43642-91
3-2
m GC-17AVer.2 SERVICE MANUAL
3.1 AFC (Advanced Flow Controller) Ver.2
Vt
TFC + PURGE VENT
vt=vc+vp+vs
Split mode, splitless mode (Not ready, after elapsing sampling time)
TFC
Pressure
control
vt=vc+vp+vs
vt=vc+vp
Completely closed
vt=vc+vp
3-4
I GC-17A Ver.2 SERVICE MANUAL
3.1 AFC (Advanced Flow Controller) Ver.2
C&l @
fU.2 -0
Ml I---- ----.-,
0.91 7c
3-5
GC-17A Ver.2 SERVICE MANUAL I,
670-l 8558-i 1 Pressure gauge 2OOkPa
AUX IN
-
221-43073 Make-up fixture 221-14775-25 Sticking nameplate AUX
3-6
- GC-17A Ver.2 SERVICE MANUAL
Configuration of manual flow controller Ver.2 (CFC-17A Ver.2)
CFC- 17 Ver. 2 Instruction manual, Manual flow controller,
221-43833-91 supplement J
-92 221-40159
-93 --I-
-94
-96
( Instruction manual, Manual flow controller,
I supplement E
22140160
1
I
- Manual flow controller main body V2 - - Small solenoid valve Assy
22 1-43999-9 1 22 I-43690-9 1
-92 -92
-93 -93
-94 -94
-96 -95
- Needle valve S 1
221-26995
3-7
GC-17A Ver.2 SERVICE MANUAL m
3.3 Additional manual flow controller Ver.2
221-31745 Housing
-94 24Ov ”
-95 230v ”
PURGE >
Fig. 3.3 Flow line diagram of manual flow controller main body 17A V2
3-8
- GC-I-IA Ver.2 SERVICE MANUAL
Configuration of APC-17 V2
konfiguration (quantity) of APC mounting parts Assy (for 2ch, 4ch, and 6ch)
221-44015 -91, -92, -93
3-9
GC-17A Ver.2 SERVICE MANUAL -
Chapter 4 Injection Unit
/CONTENTS1
4.1 SPL-17 Ver. 2.. ...............................................................................4-2
4.1.1 SPL- 17 Ver.2 configuration ................................................. 4-2
4.1.2 SPL-17 Ver.2 configuration ................................................. 4-3
4.2 WBI-17 Ver. 2.. ..............................................................................4-5
4.2.1 WBI-17 Ver.2 configuration.. ............................................... 4-5
4.2.2 WBI-17 Ver.2 configuration ................................................. 4-5
4.2.3 WBI-17 Ver.2 flow line.. ...................................................... 4-6
4.3 OCI-17 Ver.2 ..................................................................................4-7
4.3.1 OCI- 17 Ver.2 configuration .................................................. 4-7
4.3.2 OCI-17 Ver.2 configuration .................................................. 4-8
4.4 SPWWBI flow line selection unit 17A Ver.2.. ............................... 4-9
4.4.1 Configuration of SPL/WBI selection unit 17A Ver.2.. ......... 4-9
4.4.2 Flow line of SPL/WBI selection unit 17A Ver.2.. ................ 4-9
4-l
GC-17A Ver.2 SERVICE MANUAL -
4.1 .I SPL-17 Ver.2 configuration
SPL-17 v2 SPL Injection port Assy 17A V2
221-42939-g 1 221-41270-92
L
-93
-L Heater block i7A V2
221-43011
4-2
- GC-17A Ver.2 SERVICE MANUAL
4.1 SPL-17 Ver. 2
221-43886
___- Heat insulator inj
020-46539 SUS S
Aluminum foil
4-5
GC-17A Ver.2 SERVICE MANUAL -
221-43869-91 Dividing pipe Assy
-L
----- ---- ----l
I M 1
w ’ SPLIT VENT
L----------I
Inside of L.:
Y -’ indicates the flow line of AFC V2.
Injection, OCI- 17 Ver.2 Injection unit, OCI-17 Ver.2 without heater OCI injection port Assy,17 Ver.2
221-42218-92
L
Themocouple Assy
221-41569-91
4-7
GC-17A Ver.2 SERVICE MANUAL M
4.3 OCI-17 Ver. 2
221-39276 Sleeve
_;--------I--
-- 221 -43777
221-42221-91
Cover, lower OCI-17
Capillary adaptor
V2
OCI-17
4-8
m GC-17A Ver.2 SERVICE MANUAL
4.4.1 Configuration of SPL/WBI selection unit 17A Ver.2
SPUWBI selection unit 17A V2 Instruction manual, SPUWBI selection unit, Japanese version
221-43829-91 22140141
-92 Instruction manual, SPUWBI selection unit, English version
221-40142
(
INJ U2 selection unit
221-43866-91
-4-l- I I c2?Is5% --
Needle valve / 1 r GX-SLN
Fig. 4.4.2 Flow line diagram of SPL/VVBl selection unit 17A Ver.2
Chapter 5 Detector Unit
5-l
GC-17A Ver.2 SERVICE MANUAL m
5.1 .I WFID-17 Ver.2 configuration
/ FID-47 wide-range controllerVer.2 (
-mDc
FID- Nozzle base Assy, FID- 17
Ver.2 v2 V2 Assy 221-39692-92
221-43311-91 221-42951-91 221-39631-92
-92 -93 Nozzle Assy Nozzle with cap 0.3mm dia
-93 221-37304-03 22 l-33265-02
-94
High-voltage unit Assy, FID-17 Ver.2
221-42940-91
5-2
- GC-17A Ver.2 SERVICE MANUAL
5.1 WFID-17 VeCYFID-17 Ver.2 unit
221-43289-91 T 221-43268-91
-9424ov ”
5.1 WFID-17 Ver.2/FID-17 Ver.2 unit
\.., 221-42947_______
BotIom-plate heater block, 17 V2-
5-4
m GC-17A Ver.2 SERVICE MANUAL
5.1 WFID-17 Ver.2/FID-17 Ver.2 unit
/
221-43683 Top heat insulator, DET-17 V2
I
26mm
6 >
221-31745 Housing
Hz IN
5-6
- GC-17A Ver.2 SERVICE MANUAL
5.1 WFID-17 Ver.2/FID-17 Ver.2 unit
5-7
GC-17A Ver.2 SERVICE MANUAL m
Detector Unit
ab
A0 (UlEl-
ovy
Al
A’ YI
43 vt
CZZ
22OP I ,,
4 Fig. 5.1.4-2 Circuit diagram of WFID-17 V2 (112)
5.1 WFID-17 Ver.2/FID-17 Ver.2 unit
Q’ v
.F
IA
5-9
GC-17A Ver.2 SERVICE MANUAL m
5.1 WFID-17 Ver.2/FID-17 Ver.2 unit
.
?
-4
30
i P 0
t
-
-6
5-10
- GC-17A Ver.2 SERVICE MANUAL
5.1 V\iFID-17 Ver.2/FID-17 Ver.2 unit
5-11
GC-17A Ver.2 SERVICE MANUAL -
5.1 WFID-17 Ver.2/FID-17 Ver.2 unit
--
t
I/ ----I I
5-12
m GC-17A Ver.2 SERVICE MANUAL
,:,,’ :g-. : 2, ‘.
TCD-17 Ver.2 unit
xI
TCD-17 Assy - - TCD unit Assy - Heater, F’T, CA Assy, TCD Heater Assy, 1OOV/ I SOW
22 l-42474-91 221-41900-91 221-41902-91 221-41903-91
-92 -93 -93 -93
-93
- Sample inlet section Assy Temperature sensor Assy
221-41897-92 221-41568-91
Thermocouple Assy
221-41569-91
5-13
GC-17A Ver.2 SERVICE MANUAL -
5.2 TCD-17 Ver.2 unit
221-41955-02 Bottom sheet, high-temperature side 221-41960 Heat insulator, bottom, inner tank
r-
I- 221-41934 Cell fixing block 22141961 Heat insulator, side, inner tank
020-05157 SUS SEMUS pan head screw, M4x45 221-41963 Heat insulator, lid, inner tank
023-77040 SW spring washer 2-M4 221-41931-02 Heat insulator, center, inner tank
221-41936-02 Bottom sheet 221-41964 Heat insulator, inside bottom, outer tank
2 Spacer, SBE-306
5-15
GC-17A Ver.2 SERVICE MANUAL m
5.2 TCD-17 Ver.2 unit
7-38855-29 SBA-320
5-16
- GC-17A Ver.2 SERVICE MANUAL
5.2 TCD-17 Ver.2 unit
5-17
GC-l-/A Ver.2 SERVICE MANUAL B
R39 hDo 7l.m
+lq2+
Fig. 5.2.3-2 Circuit diagram of TCB-17 V2 (112)
5.2 TCD-17 Ver.2 unit
5-19
GC-17A Ver.2 SERVICE MANUAL m
II ,,:
.~ I. :,
/.’ _.:s*.:,
.. 3 FPD-17 Ver.2 unit
t- Thermocouple, 17A V2
221-43349-91
- Photo-multiplier Assy, FPD- 17 Photo-multiplier R268
221-41823-91 670-14404
- Head section
Heater, PT, CA, Assy FPD
221-42168-91
-93 Temperature sensor Assy
221-41568-91
/-- Thermocouple Assy
FPD linear controller control Assy Cable, ITG, LINEAR 221-41569-91
221-41834-91 221-41123-91
-r
Accessory, FPD-17 Ver.2 Signal cord DET Assy
221-43758-91 221-41085-91
t PCB, linear amp. FPD Assy
22 1-399.59-9 1
Flow controller, Hz-Air FPD ____ Small solenoid valve, 17A Ver.2
221-43757-91 221-43309-91
-92 -92
-93 -93
-94 -94 -91 1OOV type
-II
----.._ “L _ 221-42965 Cover, lower FPD17V2
221-41777 EIod
____
221-41772-91 Nozzle base Assy FPDlP
l- IN OUT
OUT
221-31745 Housing
5-22
m GC-17A Ver.2 SERVICE MANUAL
5.3 FPD-17 Ver.2 unit
0
E-f/.?
5-23
GC-17A Ver.2 SERVICE MANUAL I
+R< Ml0
+12v
OV(12)
+15v
-15v
AD7248
A GND
+5v
D GND
Ii:
EE
-
Ei
Ii;
A0
2:
A3
1
t15v 2
4 AG 5
-15v 3.4
DC1 7
i DC24 8
DC4L B’3 +24V
+I--+
OUT#
I 1 I
m H4
--I LSi5 RN1 01 I X2
,R4,,q ’ -4iSl588 1
3>,4
c40.022p 2, ,l 7 132qr$ RN2227Xr
5""S I
K2 I nr
1OOK
Q I
I
I- T-
ITJIA12EZX41 I I I
Detector Unit
DC24V T101
(CN1.613)
181832 X4
4 D102 IK 150K TPI-\
ov (12) <- 1 1
(CNI .B6)
H.V.
10K
---O(GND)/
HV.OFF - II
t
Clot
0.11
i-l
AGND
(CNl.A2) <
5-27
GC-17A Ver.2 SERVICE MANUAL I
5.4.1 ECD-17 Ver.2 configuration
1ECD-17 V2 1
ECD cell unit with ECD-17 Ver.2 without cell Base heater block, ECD Ver.2
indication, IOOV 22 I-42960-9 1 221-42954
221-4375 l-50
-----L- ECD cell with [email protected]) Heater, PT. CA DET Ver.2 Heater, DET Ver.2
221-25055-50 221-43346-91 221-43347-91
T
Temperaturesensor,DET Ver.2
221-43348-91
t Termbcouple, DET Ver.2
22 1-43349-P 1
ECD cell unit ECD- 17 Ver.2 without cell Base heater block, ECD Ver.2
221-43751-91 22 l-42960-9 1 221-42954
-93 -93
Heater, Pt. CA DET V2 Heater, DET Ver.2
---I ECD cell (ECD-9) 221-43346-91 22 1-43347-9 1
221-25055-91 -93 -93
Temperaturesensor,DET Ver.2
221-4334X-91
Termocoupl.e, DET Ver.2
221-43349-91
5-28
m GC-17A Ver.2 SERVICE MANUAL
5.4 ECD-17 Ver.2 unit
5-29
GC-17A Ver.2 SERVICE MANUAL 7
5.4 ECD-17 Ver.2 unit
CNl SC -I
- IMADZU MADE IN JAPAN
5-30
m GC-17A Ver.2 SERVICE MANUAL
CN2
0
R32 475K
_- vvx
I’ l-
R21
10K
+15
10K
R18 1K
Li+---- +15
cm
(SH : 1)
Detector Unit
CNl w1a tiw
C
+15v
-15v C
A GND C
M7
( Ml3 AD7248
+5v -15v +15v
D GND (
DO (
Dl (
D2 (
D3 (
(
D4
(
D5
(
D6
(
D7
Ml5
A4
A0 , ‘A
Al
A2
A3
G2B __
4#
(SH.2)
OUT#
- B9 IN#
IN#
5-33
GC-17A Ver.2 SERVICE MANUAL -
5.5.1 FTD-17 Ver.2 configuration
1 FTD-17V2 ]
221-41871-93
Power controller FTD- 17 23 >
k;stt;~;p9iy,FrD-17 / ;;;gggysy
Branch pipe Assy
221-25288-92
! FTD resistance tube Assy, Hz
221-18713-95
i- Stop valve
201-35039-00
5-34
- GC-17A Ver.2 SERVICE MANUAL
5.5 FTD-17 Ver.2 unit
ndentation, M3 x 6
FiD nozzle
, High-pressure cable
AIR - J!i “1
~7304-03
221-39692-92
Nozzle
Nozzle
Assy-
base portion FID17 V2
H2+ -
Make-up gas
Capillary adapter
H2 IN
LI-------
5-36
0 GC-17A Ver.2 SERVICE MANUAL
5.5 FTD-17 Ver.2 unit
5-37
GC-17A Ver.2 SERVICE MANUAL -
04
Dl
D2 24V
13
D3 c21 24V
D4 RSl 0
D5
D6
D7
A4
Al
A2
A3 READ
6# LOW 2sc
2661
OUl
IN
CN2
R3
5.11K
OUT
Y4t
(64)
--lIp,,l’” “-“‘P
5-39
GC-17A Ver.2 SERVICE MANUAL -
5.5 FTD-17 Ver.2 unit
5-40
- GC-17A Ver.2 SERVICE MANUAL
Chapter 6 Appendix
6-1
GC-17A Ver.2 SERVICE MANUAL -
Maintenance parts list
“~.,:.....j, ,,__ I
22 l-43 122-9 1 PCB, POWER CONT 17AV2 I- 22 l-43227-9 1 Front panel Assy ICOV
-92 Front panel Assy 115v
is--- 065-68207-03 SSR, S26MDOl
-93 Front panel Assy 220v
072-01665-36 Fuse, 326 015, ISA 1OOV system -94 Front panel Assy 240v
072-01665-34 Fuse, 326 010, 10A 1OOV system -95 Front panel Assy 230V
072-01664-33 Fuse, 313 005, 5A lOOV, 2COV system 22 l-4322 l-92 Solenoid valve Assy ESC
072-01665-33 Fuse, 326 008, 8A 200V system t 22 l-43690-9 1 Small solenoid valve Assy 1OOV
072-01664-30 Fuse 3A 200V system
-92 Small solenoid valve Assy 115V
074-80662-o 1 SW power supply 5V, ?c 15 1OOV system -93 Small solenoid valve Assy 220V
074-80667-01 SW power supply 5V. t 15 200V system -94 Small solenoid valve Assy 240V
-95 Small solenoid valve Assy 230V
074-80654-3 1 SW power supply 24V 1OOV system
374-80666-01 SW power supply 24V 250V system 221-43096-91 PCB, FLOW CONT 17A V2
670-18558-12 Pressuregauge300kPa
6-2
- GC-l-/A Ver.2 SERVICE MANUAL
6.1 Maintenance parts list
6.2.2 Mode selection By pressing the (Eg key ten times, a different mode from the
normal one is selected. Pressing the [SC] key returns the mode
to normal.
1. Monitor function
The following functions are available with the (m] , and
m or @ key.
0 The monitor value for each temperature block is displayed to
the first decimal place. [In this case, the decimal point itself
is not displayed.)
@ The monitor displays the pressure values for FLOW 1, 2,
and 3 and the flow rate value to the first decimal place.
@ AFC and the room temperature are displayed.
6-4
- GC-17A Ver.2 SERVICE MANUAL
6.2 Special key operations
6-5
GC-17A Ver.2 SERVICE MANUAL B
6.2 Special key operations
CONTROL No. Initial values Contents CONTROL No. Initial values Contents
1 10 FLOW 1 (BP) Time constant 26 IO FLOW 4 (F) Proportionalconstan
2 10 FLOW I (BP) Proportionalconstant 27 20 FLOW 4 (F) Integral constan
3 8 FLOW 1 (BP) Integral constant 28 0 FLOW 4 (F) Differentialconstan
4 0 FLOW 1 (BP) Differentialconstant 29 10 FLOW 4 (P) Time constant
5 1 FLOW 2 (F) Time constant 30 200 FLOW 4 (P) Proportionalconstanl
6 10 FLOW 2 (F) Proportionalconstant 31 5 FLOW 4 (P) Integral constant
7 20 FLOW 2 (F) Integral constant 32 0 FLOW 4 (P) Differentialconstant
8 0 FLOW 2 (F) Differentialconstant
9 10 FLOW 2 (P) Time constant 37 I FLOW 3 (P) Time constant
IO 200 FLOW 2 (P Proportionalconstant 38 20 FLOW 3 (P) Proportionalconstant
11 5 FLOW 2 (F Integral constant 39 5 FLOW 3 (P) Integral constant
12 0 FLOW 2 (P Differentialconstant 40 0 FLOW 3 (P) Differentialconstant
@ ADJUST value
The ADJUST values of the AFC and APC can be checked
with the [E], B , and @ keys.
The meaning of each parameter is as follows.
6-6
m GC-17A Ver.2 SERVICE MANUAL
6.2 Special key operations
6-7
GC-17A Ver.2 SERVICE MANUAL m
6.3.1 Maintenance precautions
1. Make sure to turn off both the power switch of the GC-l7A
and main power supply to the equipment. (The same applies
to peripheral devices.)
2. Always make sure to close the hydrogen gas valve on the sup-
ply side before working on any equipment which use the gas.
3. Make sure to turn off the main valve on the supply side
before inspecting any flow system and never remove any
joint until the gas pressure completely decreases.
h!L
II\ If the gas flows in the reverse direction due to loosening the
joint before completely lowering the gas pressure, it may
result in a pressure control malfunction caused by dust and
external particles.
(A filter is installed in the primary side of the pressure con-
troller but not in the secondary side.)
4. When connecting a pipe, always hold the pipe in one hand
and manually tighten the joint with your other hand, making
sure it is correctly aligned. Following this, grip two wrench-
es with one hand, squeezing them together until the joint is
fully tightened. (Refer to Fig. 6.3.1- 1)
-\!!L
‘1, While only one wrench can be used to connect a pipe to a
firmly secured joint, it is recommended to use two wrenches
at all times.
6-8
- GC-17A Ver.2 SERVICE MANUAL
6.3 GC general
Cut here
6-9
GC-17A Ver.2 SERVICE MANUAL I
6.3 GC general
N:h%ote2
Fig 6.3.2 Procedure for restrictor replacement
<<Note2+> 1. Be careful not to bend the restrictor when inserting it into the
joint.
2, Make sure to secure the designated length of the joint so that
it does not touch the end of the restrictor.
6-10
m GC-17A Ver.2 SERVICE MANUAL
6.3 GC general
stop the gas flow before removing the molecular sieve filter.)
@ Short-circuit the joints where the molecular sieve filter was
installed.
@ Remove the column from the column oven.
@ Install the molecular sieve filter to the INJ outlet using a
wrench, and firmly tighten it to prevent leakage.
@ Allow the carrier gas to flow at 40-50 ml/min. (N2 or He gas
can also be used.)
@ Heat the column oven to 34OC and maintain the tempera-
ture for more than 5 hours. (This procedure is called aging.)
<<Note>> Make sure that the carrier gas is flowing through the molecular fil-
ter during heat aging.
6-11
GC-17A Ver.2 SERVICE MANUAL m
6.3 GC general
<<Note>> If the carrier gas is stopped while the column oven is still hot, it can
damage the column. Pay special attention when setting the FLOW
OFF TIME.
When the FLOW OFF TIME is set at 9999 (initial setting), press
the [%mJ (E] and [w] keys to stop the carri-
er gas after the temperature of the column oven lowers below 50 “C
(2) Make sure that the temperature of the injection port has suf-
ficiently lowered (below 100 “C ) before removing the sep-
tum securing nut by fingers as shown in Fig, 6.3.5-l.
After removing the nut, take out the needle guide and sep-
tum. (Do not turn the septum securing nut while it is hot, as
this can cause scorching.)
(3) Insert a new septum.
(4) Attach the needle guide and secure it with the nut. Turn the
nut all the way with fingers.
6-12
m GC-17A Ver.2 SERVICE MANUAL
6.3 GC general
Needle guide
Septum
Fig. 6.3.6-l
Aluminum foil
6-13
GC-17A Ver.2 SERVICE MANUAL II
6.3 GC general
START
Turn the heater ON. When the READY lamp lights up, the
column oven starts increasing its inside temperature.
After the temperature has reached the specified setting, turn
OFF the power of the Gas Chromatograph. Then, when the
column oven has cooled down to room temperature, remove
the septums wrapped in the aluminum foil. (It will take about
1 hour for the oven to sufficiently cool down.)
Place the aged injection port septums in a clean bottle and
seal it tight for storage.
<<Note>> Be sure to lower the column inlet pressure before removing the
septum. if the septum is removed when the column inlet pressure is
still high, a sudden drop in pressure may cause silica wool to squirt
or oxygen to enter the column and cause oxidation.
6-14
m GC-l-/A Ver.2 SERVICE MANUAL
Relationship between temperature and temperature sensor resistance
i-----l
'C 1 Ohm 1 Ohm/‘; “C Ohm Ohm /“C ‘C Ohm Ohm /“C
120 51.32 80 131.42 280 206.66
0.415 0.388 0.364
110 55.47 90 135.30 290 210.30
0.412 0.386 0.363
-100 59.59 + 100 139.16 + 300 213.93
0.411 0.385 0.361
90 63.70 110 143.01 310 217.54
0.409 0.384 0.361
80 67.79 120 146.85 320 221.15
0.408 0.382 0.359
70 71.87 130 150.67 330 224.74
0.406 0.382 0.358
60 75.93 140 154.49 340 228.32
0.404 0.380 0.357
50 79.97 150 158.29 350 231.89
0.403 0.379 0.356
40 84.00 160 162.08 360 235.45
0.402 0.378 0.354
30 88.02 170 165.86 370 238.99
0.401 0.377 0.354
20 92.03 180 169.63 380 242.53
0.399 0.375 0.352
10 96.02 190 173.38 390 246.05
0.398 0.375 0.35 1
i0 100.00 + 200 177.13 +400 249.56
0.397 0.373 0.350
10 103.97 210 180.86 410 253.06
0.396 0.372 0.349
20 107.93 220 184.58 420
0.395 0.371
30 111.88 230 188.29 430
0.393 0.370
40 115.81 240 191.99 440
0.392 0.368
50 119.73 250 195.67 450
0.391 0.368
60 123.64 260 199.35 460
0.390 0.366
70 127.54 270 203.03
0.388 0.365
___--
6-15
GC-17A Ver.2 SERVICE MANUAL m
.I I CA Thermocouple
“C 0 1 2 3 4 5 6 7 8 9 10 “C
1
t millivolts
- - 1
(+)o 0.00 0.04 0.08 0.12 0.16 0.20 0.24 0.28 0.32 0.36 0.40 (-t)C
10 0.40 0.44 0.48 0.52 0.56 0.60 0.64 0.68 0.72 0.76 0.80 10
20 0.80 0.84 0.88 0.92 0.96 1.oo I .04 1.08 1.12 1.16 I .20 20
30 1.20 1.24 1.28 1.32 1.36 1.40 1.44 1.49 1.53 1.57 1.61 30
40 1.61 1.65 1.69 1.73 1.77 1.81 1.85 1 .vo 1 .v4 1.98 2.02 40
50 2.02 2.06 2.10 2.14 2.17 2.23 2.27 2.3 1 2.35 2.39 2.43 50
60 2.43 2.47 2.5 1 2.56 2.60 2.64 2.68 2.72 2.76 2.80 2.85 60
70 2.85 2.89 2.93 2.97 3.01 3.05 3.10 3.14 3.18 3.22 3.26 70
80 3.26 3.30 2.35 3.39 3.43 3.47 3.5 1 3.56 3.60 3.64 3.68 80
90 3.68 3.72 3.76 3.81 3.85 3.89 3.93 3.97 4.01 4.06 4.10 90
--
100 4.10 4.14 4.18 4.22 4.26 4.3 1 4.39 4.43 4.47 4.5 1 100
110 4.5 1 4.55 4.60 4.64 4.68 4.72 4.76 4.80 4.84 4.88 4.92 110
120 4.92 4.96 5.01 5.05 5.09 5.13 5.17 5.21 5.25 5.29 5.33 120
130 5.33 5.37 5.41 5.45 5.49 5.53 5.57 5.61 5.65 5.69 5.73 130
140 5.73 5.77 5.81 5.85 5.89 5.93 5.97 6.01 6.05 6.09 6.13 140
1.50 6.13 6.17 6.2 1 6.25 6.29 6.33 6.37 6.41 6.45 6.49 6.53 150
160 6.53 6.57 6.61 6.65 6.69 6.73 6.77 6.81 6.85 6.89 6.93 160
.-x
170 6.93 6.97 7.01 7.05 7.09 7.13 7.17 7.21 7.25 7.29 7.33 170
z 180 7.33 7.37 7.41 7.45 7.49 7.53 7.57 7.61 7.65 7.69 7.73 180
x 190 7.73 7.77 7.81 7.85 7.89 7.93 7.97 8.01 8.05 8.09 8.13 190
--
a” 8.50 8.54
200 8.13 8.17 8.21 8.25 8.29 8.33 8.37 8.41 8.46 200
--
210 8.54 8.58 8.62 8.66 8.70 8.74 8.78 8.82 8.86 8.90 8.94 210
220 8.94 8.98 9.02 9.06 9.10 9.14 9.18 9.22 9.26 9.30 9.34 220
230 9.34 9.38 9.42 9.46 9.50 9.54 9.59 9.63 9.67 9.7 1 9.75 230
240 9.75 9.79 9.83 9.87 9.9 i 9.95 9.99 10.03 10.07 10.11 10.16 240
250 10.16 10.20 10.24 10.28 i0.32 10.36 10.40 10.44 10.48 10.52 10.57 250
260 10.57 10.61 10.65 10.69 10.73 10.77 10.81 10.85 10.89 10.93 10.98 260
270 10.98 11.02 11.06 11.10 11.14 11.18 11.22 il.26 11.30 11.34 Il.39 270
280 11.39 11.43 11.47 11.51 11.55 li.59 11.63 11.67 11.72 11.76 11.80 280
290 11.80 11.84 11.88 il.92 11.96 12.01 12.05 12.09 12.13 12.17 i2.21 290
-- --
300 12.21 12.25 12.29 12.34 12.38 12.42 12.46 12.50 12.54 12.58 12.63 3oa
310 12.63 12.67 12.71 12.75 12.79 12.83 12.88 12.92 12.96 13.00 13.04 310
320 13.04 13.08 13.12 13.17 13.21 13.25 13.29 13.33 13.37 13.42 13.46 320
330 13.46 13.50 13.54 13.58 13.62 13.67 13.71 13.75 13.79 13.83 13.88 330
340 13.88 13.92 13.96 14.00 14.04 14.09 14.13 14.17 14.21 14.25 14.29 340
350 14.29 14.34 14.38 14.42 14.46 14.50 14.55 14.59 14.63 14.67 14.71 350
360 14.71 14.76 14.80 14.84 14.88 14.92 14.97 15.01 15.05 15.09 15.13 360
370 15.13 15.18 15.22 15.26 15.30 15.34 15.39 15.43 15.47 15.51 15.55 370
380 15.55 15.60 15.64 15.68 15.72 15.76 15.81 15.85 15.89 15.93 15.98 380
390 15.98 16.02 16.06 16.10 16.14 16.19 16.23 16.27 16.31 16.36 16.40 390
400 16.40 16.44 16.48 16.52 16.57 16.61 16.65 16.69 16.74 16.78 16.82 400
410 16.82 16.86 16.91 16.95 16.99 17.03 17.07 17.12 17.16 17.20 17.24 410
420 17.24 17.29 17.33 17.37 17.41 17.46 17.50 17.54 17.58 17.62 17.67 420
430 17.67 17.71 17.75 17.79 17.84 17.88 17.92 17.96 18.01 18.05 18.09 430
“C 0 1 2 3 4 5 6 7 8 9 10 ‘C
6-16
- GC-17A Ver.2 SERVICE MANUAL
: :.-;. I ~
: 6. i ‘6. ” Linear velocity and flow rate
Where:
The column flow rate for the GC-17A V2 indicates the value that
has been converted to the under ambient temperature and atmo-
spheric pressure. It is equivalent to the value measured by a flow
meter connected the column outlet under ambient temperature.
i.D.=0.22mm
P
100Column inlet pressure
VW
x (PfPo)Z---PO2 x 103
Fc= rp$z PO
6-17
GC-17A Ver.2 SERVICE MANUAL I
6.6 Linear velocity and flow rate
<=Sleference>> Method of obtaining the carrier gas flow rate (Fc) in the column
from the retention time (to) of methane.
r2L
Fc= rtoJ Unit of FC 1 ml/min
Therefore:
4-18
- GC-77A Ver.2 SERVICE MANUAL
CB
3 Kanda-Nlshtktcho
SHIMADZU