Manual DC Imaging System Flow Visualization
Manual DC Imaging System Flow Visualization
This users guide may not be copied, photocopied, translated, modified, or reduced to any electronic medium or machine-readable form, in whole, or in part, without the prior written consent of Dantec Dynamics A/S. Publication no.: 9040U4164 Date: July 1, 2004 3nd edition printed in 2004. 2004 by Dantec Dynamics A/S, Tonsbakken 16-18, DK-2740 Skovlunde, Denmark. All rights reserved. All trademarks referred to in this document are registered by their owners.
Table of contents
Table of contents................................................................................................................................................i 1. Warranties & disclaims ........................................................................................................................1-1 1.1 Dantec Dynamics License Agreement .................................................................................................1-1 1.1.1 Grant of License..........................................................................................................................1-1 1.1.2 Upgrades .....................................................................................................................................1-1 1.1.3 Copyright ....................................................................................................................................1-1 1.1.4 Limited Warranty........................................................................................................................1-1 1.1.5 Limitation of Liability ................................................................................................................1-2 1.1.6 Product Liability .........................................................................................................................1-2 1.1.7 Governing Law and Proper Forum .............................................................................................1-2 1.1.8 Questions ....................................................................................................................................1-3 1.2 Laser safety..........................................................................................................................................1-3 1.3 CCD sensor warranty..........................................................................................................................1-3 2. DC Image Acquisition System..............................................................................................................2-1 2.1 Overview..............................................................................................................................................2-1 2.2 Cameras supported by the DC Imaging System ..................................................................................2-1 2.2.1 HiSense MkII camera .................................................................................................................2-1 2.2.2 FlowSense 2M camera................................................................................................................2-2 2.2.3 HiSense 4M camera ....................................................................................................................2-2 2.3 Illumination systems ............................................................................................................................2-2 2.3.1 New wave solo pulsed lasers ......................................................................................................2-2 2.3.2 Shutter device .............................................................................................................................2-2 3. Installation .............................................................................................................................................3-1 3.1 Unpacking the system..........................................................................................................................3-1 3.2 System installation...............................................................................................................................3-1 3.2.1 Frame grabber(s) and synchronisation board Driver installation................................................3-1 3.2.2 Frame grabber(s) and synchronisation board installation ...........................................................3-1 3.2.3 Analog Input software and hardware installation .......................................................................3-2 3.2.4 DC Image Acquisition Software installation ..............................................................................3-2 3.2.5 Uninstall the DC Image Acquisition software ............................................................................3-2 3.3 Connections .........................................................................................................................................3-3 3.3.1 Steps by steps procedure.............................................................................................................3-3 3.3.2 Connections diagram ..................................................................................................................3-4 3.3.3 Analog Input ...............................................................................................................................3-4 4. Using the DC Image Acquisition software ..........................................................................................4-1 4.1 First time use .......................................................................................................................................4-1 4.2 Online acquisition window..................................................................................................................4-3 4.2.1 Tab: Acquisition Control ............................................................................................................4-3 Image acquisition ............................................................................................................................... 4-4 View control ....................................................................................................................................... 4-6 Storage control ................................................................................................................................... 4-7
Installation and Users guide for DC Image Acquisition System. i
4.2.2 Tab: Timing Setup ..................................................................................................................... 4-7 Trigger control related parameters...................................................................................................... 4-7 4.2.3 Frame grabber settings (Tabs: FG1 Info and FG2 Info) ............................................................ 4-8 4.2.4 A/D Setup................................................................................................................................... 4-9 A/D Setup control............................................................................................................................... 4-9 Trigger settings ................................................................................................................................. 4-10 Channel 1 settings............................................................................................................................. 4-10 Channel 2 settings............................................................................................................................. 4-10 4.2.5 Fail-safe function ..................................................................................................................... 4-10 5. 6. Troubleshooting.................................................................................................................................... 5-1 Technical specifications ....................................................................................................................... 6-1
ii
1.1.2 Upgrades
This License Agreement shall apply to any and all upgrades, new releases, bug fixes, etc. of the Software which are supplied to you. Any such upgrade etc. may be used only in conjunction with the version of the Software you have already installed, unless such upgrade etc. replaces that former version in its entirety and such former version is destroyed.
1.1.3 Copyright
The Software (including text, illustrations and images incorporated into the Software) and all proprietary rights therein are owned by Dantec or Dantecs suppliers, and are protected by the Danish Copyright Act and applicable international law. You may not reverse assemble, decompile, or otherwise modify the Software except to the extent specifically permitted by applicable law without the possibility of contractual waiver. You are not entitled to copy the Software or any part thereof. However you may either (a) make a copy of the Software solely for backup or archival purposes, or (b) transfer the Software to a single hard disk, provided you keep the original solely for backup purposes. You may not copy the Users Guide accompanying the Software, nor print copies of any user documentation provided in on-line or electronic form, without Dantecs prior written permission.
1-1
A defect in the Software shall be regarded as material only if it has an effect on the proper functioning of the Software as a whole, or if it prevents operation of the Software. If until 90 days after the delivery of the Software, it is established that there is a material defect in the Software, Dantec shall, at Dantec's discretion, either deliver a new version of the Software without the material defect, or remedy the defect free of charge or terminate this License Agreement and repay the License fee received against the return of the Software. In any of these events the parties shall have no further claims against each other. Dantec shall be entitled to remedy any defect by indicating procedures, methods or uses ("work-arounds") which result in the defect not having a significant effect on the use of the Software. The Software was tested prior to delivery to you. However, software is inherently complex and the possibility remains that the Software contains bugs, defects and inexpediences which are not covered by the warranty set out immediately above. Such bugs etc. shall not constitute due ground for termination and shall not entitle you to any remedial action. Dantec will endeavour to correct all bugs etc. in subsequent releases of the Software. The Software is licensed "as is" and without any warranty, obligation to take remedial action or the like thereof in the event of breach other than as stipulated above. It is therefore not warranted that the operation of the Software will be without interruptions, free of defects, or that defects can or will be remedied.
1-2
1.1.8 Questions
Should you have any questions concerning this License Agreement, or should you have any questions relating to the installation or operation of the Software, please contact the authorised Dantec distributor serving your country. You can find a list of current Dantec distributors on our web site: www.dantecdynamics.com Dantec Dynamics A/S Tonsbakken 16 18 DK - 2740 Skovlunde Denmark Tel: +45 44 57 80 00 Fax: +45 44 57 80 01 Web site: http://www.dantecdynamics.com
1-3
Precautions 1. Cap the lens whenever the camera is not in use. 2. Cap the lens during set-up and alignment of the light sheet. Before removing the cap, make sure that reflections off the surface of any objects inside the light sheet do not hit the lens by observing where reflections go. 3. As general precautions to avoid eye damage, carefully shield any reflections so that they do not exit from the measurement area. You must wear appropriate laser safety goggles during laser alignment and operation.
1-4
The high dynamic range is a valuable flexibility in the practical performance of the PIV and LIF experiments (although most PIV experiments do not require 12 bit resolution, LIF does in terms of scalar resolution, precision and accuracy in some cases). Also, there is less need to adjust the laser power and the seeding, simply
Installation and Users guide for DC Image Acquisition System. 2-1
because a wider range of input intensity levels provide successful results. Likewise, if problematic windows, dirt or other produces uneven illumination over the area, there is less loss of local information, because the signal is still received at the CCD due to the higher dynamic range.
2-2
3. Installation
3.1 Unpacking the system
The DC Image Acquisition System comprises:
1 frame grabber for each camera package delivered with the DC System (also
FlowManager data analysis software Note: The frame grabber(s) and the synchronisation board must be installed in the same computer as the FlowManager software. If the computer is part of the delivery the installation has already been done and do not require any further software installation. In this case you may go directly to section 4.4, otherwise please read carefully all Section 3 of this document. To use Dantec Dynamics DC Imaging System you need an illumination system, like a double cavity laser with lightsheet and optics.
Note:
To make sure the boards are properly installed, start the MAX utility: Select Start Programs National Instruments Measurement & Automation (from Windows environment) If you are prompted about User preferences, just click OK to proceed In the Configuration part of this software (View Configuration), select Devices and Interfaces. Among many PCI devices, the PCI-6601 device and 1 (or 2) IMAQ PCI-1428 devices should be present Click on PCI-6601: Device number should have the value 1 Click on Tools IMAQ Max number of buffer and set the value to 1000 Press the Ok button Exit MAX utility software
Note:
It is important that the version number of the FlowManager software matches the version number of the DC Imaging System software. If not, the installation will not succeed. Note that when installing new software on your PC, you need to have administrative rights.
Note:
Select Start Settings Control panel and select Add/Remove programs Select DC Image Acquisition and click on the Remove button. Select then FlowManager and click on Change Next Repair Next Install to repair FlowManager software
Installation and Users guide for DC Image Acquisition System.
3-2
Now the DC part has been removed and FlowManager will operate as in the original installation.
3.3 Connections
3.3.1 Steps by steps procedure
Before utilisation of the DC system, all parts must be properly connected. Always turn off the power of the PC while hardware connections are made. On all cameras there is a DIP-switch. For the HiSense MKII/4M camera, verify that all four pins are in the lower position (i.e. off position). For the FlowSense camera, all pins must be in the left position. Connect the camera Digital I/O connector to the frame grabber Camera link connector using the 10 meter data cable. The other connector on the frame grabber is not used. Connect the synchronisation board to the Timer Box using the 1m interface cable. Timer / laser box / camera(s) connections Connect the connector marked Out1 on the Timer Box to the Camera Timing I/O connector (HiSense MKII or 4M) or DC In/Trig (FlowSense) using the 10 m trigger cable. If two cameras should be connected, use a BNC T-piece to connect both cameras to the Timer Box Out1 connector. Using the 10 m long BNC cables, link the laser Flashlamp and Q-switch connectors to the timer box as follows: Connection on Timer box Out2 Out 3 Out 4 Out 5 Out 6 Out 7 Out 8 In 1 In 2 Laser cables Fire flashlamp 1 Fire flashlamp 2 (For shutter option) Fire Q-switch 1 Fire Q-switch 2 Not used Not used Trigger input Not used
Figure 3-1: Connecting the interface cables to the Timer box Connect the camera power supply(ies) to the camera power connector(s) and to the 220/110V supply. Turn on the power for the camera(s) and the PC
3-3
3-4
Figure 4-1: Utility software for initialisation of each frame grabber board installed in the computer. Start FlowManager software and select the correct device library. - This is found by selecting FileLibrariesDevice Library. - Press FileOpen and select DevLib440.fdl which contains the available devices. (When the default installation path for FlowManager is used, this file is found in the path C:\Program Files\Dantec Dynamics\FlowManager\Libs)
Figure 4-2: Selecting the device library. Create a new database (FileDatabaseNew) and a project with a new setup and add a HiSense MkII, HiSense 4M or FlowSense 2M camera and the laser source from the library. If the setup contains two cameras, add a second camera of the same type.
Note:
The DC Image Acquisition System does not support mix of camera types; e.g. 1 FlowSense + 1 HiSense MkII. For multiple camera settings, the user shall select and use 2 identical cameras; i.e. either 2 FlowSense 2M, 2 HiSense MkII or 2 HiSense 4M.
4-1
Figure 4-3: Selecting the appropriate camera settings using FlowManager set-up functionalities.. Select OptionsSystem Unit Type and Direct Computing Imaging
Figure 4-4: Selecting the DC System among the many acquisition systems available from Dantec Dynamics A/S. Press then the green Run online icon ( acquisition part. ) on the tool bar to enter the online
Figure 4-5: FlowManager main window with iconised shortcuts to direct set-up and image acquisition start.
4-2
Next to this dialog box a window displays images acquired during preview or while viewing images from temporary storage. If two cameras are installed (and set in the acquisition settings), images from each camera are displayed in separate windows. Note: If only one camera is included in the setup but two frame grabbers/cameras are installed in the system, the software always use the camera connected to the first frame grabber detected (and this, even if it has been specified as camera number 2 in the setup).
Figure 4-6: The online acquisition window with Acquisition control tab, Timing setup tab and camera Information tab(s)
4-3
Note that the system memory in the PC is used as temporary storage for the images acquired. Therefore to make the acquisition process more efficient in terms of maximum number of images, the system memory must be prepared for the data before the acquisition actually takes place. Also, depending on the amount of images to acquire this initialisation process will take some time. When the acquisition is started (by pressing the Preview or Initialise Acquire button), a message is displayed informing about memory initialisation. Once done, the button Begin is enabled and when pressed, the acquisition begins with the settings described in the sections below. Note: At the time of preparation, max. 90% of the available memory can be reserved for temporary storage, while the remaining 10% are reserved for the operating system.
Image acquisition
To acquire a serie of images, specify the number of records to acquire or check the option Use full range to record the maximum of records possible with the system used. (Note that 1 record refers to a double image from each camera.) Also, this maximum number of images is limited by the available memory in the PC. Below the Initialize Acquire button, a status line shows the storage capacity in terms of maximum records (and MBytes space). If the user enter say 250 records when the system only can store 150 images, the software automatically correct the value typed in the Records to acquired dialog and replaces it by the maximum available. Additionally, the total time the acquisition will take (in seconds) is given. The latter depends on the trigger frequency when the Internal master clock triggering mode is used, and it is set to zero when the External sync. mode is used (since the trigger frequency is not known.)
Pressing the Start laser button immediately starts the laser, which flashes at the Trigger frequency specified in the Tab: Timing setup. (Note that this is also a way to get the laser stabilised before record acquisition, which is a critical issue when e.g. running LIF experiments.) When the trigger mode is set to External sync., the laser flashes each time the external trigger signal is received by the timer box.
4-4 Installation and Users guide for DC Image Acquisition System.
Note:
In any synchronisation case, no images are recorded until the Show preview or the Initialize Acquire buttons are pressed.
The Show preview button offers the possobility to acquire images without saving the data in the memory. This is useful to ensure the system is set up properly for the wanted acquisition or for the focusing of camera(s) prior to image acquisition. Press the Stop Acquisition button to stop the continuous preview. Note: When the Show preview button is pressed, the system memory is initialised and the synchronisation board runs on triggering the camera at the rate specified in the timing setup (i.e. fixed rate or externally triggered). The software transfers each double image from the camera(s) and display it in the image map window. At high trigger frequencies (i.e. approx. 2-3Hz for 1 camera system and approx. 1-2Hz for 2 cameras systems), overlap of images during updating the window may occur. This cosmetic display problem depends on the processing power of the PC and does not indicate at all technical problems with the image acquisition itself.
Once the number of images (and timing setup see next section) are specified, acquisition can be done. Press the Start laser button and then the Initialize Acquire button: the PC memory is initialised and when the system is ready the Begin message is displayed (Figure 4.8b). Click on OK to start the acquisition process (or on Cancel to stop the acquisition process). Note: If the Start laser button has not been pressed before the acquisition, a warning message is displayed asking to start the laser or cancel the operation (see Figure a below).
Figure 4-8: Dialog boxes when starting measurements Important tip: It can take 30 seconds and more to initialise the memory, depending on the number of images to be recorded. During that period of time, if the Start laser has been pressed, the laser flashes. For the safety of the operator, it may be adequate not to press the Start laser button. When the Initialize Acquire button is pressed without the laser flashing, a dialog box pups-up to inform the user to start the laser. That way, the laser flashes only during acquisition and not during the initialisation period.
4-5
The Acquire images button now changes text and functionality to a Stop Acquisition button. Acquisition will run until the specified number of images to acquire is reached or when the user presses the Stop Acquisition button. Note that during the acquisition, a status line shows the progress and time left to finish acquistion process. When the acquisition has ended, the records are stored temporarily in to the PC memory until the Clear memory button is pressed or on return to the main FlowManager software for the processing of the images.
This button is only enabled when at least 1 record is stored in the memory of the PC. When pressing this Clear memory button, all the data are removed from the temporary storage, which frees the memory to another complete range of records acquisition. This button must therefore be pressed when acquiring a new set of images; i.e. after a set of records is saved to the FlowManager database, or when the user is not content of the quality of the images recorded and thereby wishes to run a new acquisition, keeping the settings as defined. View control Once a number of records are acquired and stored in the memory of the PC, they can be browsed one by one, using the view control buttons. (Note that if only the Show Preview button is pressed there is no record in the system memory and therefore nothing to view.)
Figure 4-9: Records in PC memory can be browsed via the View control feature of the DC Image Acquisition Software
Play buttons
Use the play buttons to view the next image in memory (>) or the previous image in memory (<). This option can use to check the quality of some of the records before they are transferred to the database.
The Fast play button (>>) has a similar functionality as the Play buttons. Position the slider at the first image you want to see, press the Fast play button, and all subsequent records are shown at a rate of approx. 1Hz.
Scrolling
Records can also be viewed by clicking with the mouse cursor on the scrolling bar and moving the mouse cursor towards the right (next record) or the left (previous record).
4-6
Storage control To save the images stored in the PC memory to FlowManager database environment, specify which images to save, using the From and To input fields or check the Save all. Press then the Save in database button, which will start the transfer of the records selected from the system memory to the database and the harddisk. A status line will appear and show the progress and time left to finish the transfer process. If needed, the user may stop this transfer by pressing the Stop transfer button.
Trigger mode
The DC Image Acquisition System allows both internal run (i.e. user-defined acquisition frequency) and external triggering. In the latter case, an external trigger source must be connected to the Timer Box In1 connector (see Figure 3-3). When selecting the trigger mode Internal master clock, the trigger pulses are generated by the synchronisation board at a rate specified in the Trigger frequency field (between 0.25Hz and a camera dependent maximum rate) When setting the trigger mode to External sync., the trigger pulse generated by the external trigger source control the acquisition system. Note that the software makes sure that the camera is not triggered at a rate faster than maximum rate If the trigger pulses arrive faster than this frequency, only every Nth pulse triggers the measurement with N chosen so that the frame rate comes just below the maximum rate. Trigger frequency
The trigger frequency determines the rate (in Hertz, Hz) at which the camera is triggered to deliver double images. In External sync. trigger mode, this field is disabled as the trigger frequency source determines this acquisition rate.
Installation and Users guide for DC Image Acquisition System. 4-7
Trigger delay
Note that this option is only available when the External sync. trigger mode is selected. The Trigger delay specifies the time necessary for the external trigger pulse to arrive so the laser fires (or the shutter is opened). Delays for the camera and laser/shutter are also taken into consideration. Due to these delays in camera and laser, the trigger delay must be above a minimum value. In case the specified value is below the minimum value, the software will automatically update this value and display the adequate message. Note: With the present DC Image Acquisition System, the maximum trigger delay is 4 seconds.
dt
The dt is the time between the first and the second laser pulse. The first laser pulse is exposed in the first image (frame A) while the second laser pulse is exposed in the second image (frame B). The value of dt can vary in the range 5sec to 80 msec. Naturally, the adequate value of dt depends on the flow measured and camera set-up. Prior knowledge of expected velocities is typically very helpful.
4.2.3 Frame grabber settings (Tabs: FG1 Info and FG2 Info)
There are no settings to be applied for the frame grabber and the camera that are part of the system. The only information displayed is the grabber type, the serial number and the memory on the board. (Figure 4.7) with as many dialog windows as there are frame grabbers installed in the PC. The frame grabber is automatically detected when installed and its parameters are matched with the settings of the camera.
Note: Note:
If the onboard memory is displayed as N/A, there may have been problems during the detection of the system. Rebooting the PC very often fix this problem. The camera(s) is(are) always set to double frame mode, full pixel resolution and 12 or 10 bit for the HiSense MkII/4M and FlowSense 2M, respectively. Note that no camera settings can be changed from the software.
4-8
Figure 4-13: Frame grabber information tabs A/D Setup control Use analog input When the analog input is not needed, disable the input by removing the check mark from the box Use analog input'. Max number of analog samples Below the check box Use analog input, the maximum number of analogue samples appears. This number, which depends on the onboard memory size and the Sample delay/range setting, shows how many waveforms the board can hold.
4-9
When the user specifies a number of records that exceeds the max number of analog samples, only the first max number of analog samples will contain analog data. Sample delay/range (in the bottom of the dialog) This input has two functions: 1. The user specifies the delay from the trigger to the analog values that are saved with image data. 2. This setting specifies the length of the waveform in time. (The range is from 0.1 s up to 1 ms.) Trigger settings Trigger select The trigger input of the CompuScope board is edge sensitive. It is possible to specify if it is a positive edge or a negative edge that triggers the acquisition. Level Together with specifying the trigger polarity a voltage must be specified. (The range is from 5.0V to +5.0V.) Channel 1 settings Range Specify here the input voltage range for channel A. Coupling This input is used to specify if the input coupling is DC or AC. Like on an ordinary oscilloscope, the user has the possibility to select or to remove any DC level offset by choosing AC. Attenuation Attenuation is related to the probe that is used. If the operator uses a x10 probe then attenuation x10 must be used, likewise for a x100 probe the attenuation is set to x100. When no probe are used, just connect the BNC cable from the signal source set the Attenuation to x1. Channel 2 settings Channel Enable For some boards it is possible to disable channel B. Range See Channel 1 settings. Coupling See Channel 1 settings. Attenuation See Channel 1 settings.
To enable this safety system, the Timer box (which contains a watchdog circuit) must be connected to the laser interlock. The watchdog functions as a relay that establishes a connection between the two pins in the cable labelled Interlock when it is active. This circuit is always held active by the DC Image Acquisition Software and if/when the software crashes or when leaving the online acquisition section of the software it breaks the connection Important note: If the software crash ends by sending an error report, the safety connection will not open until the user sends the report. Also if a blue screen crash ends with memory dump, the connection will not open until the end of the dump. The cable from the Laser connector on the Timer box, fits into the Interlock connector of e.g. the New Wave Solo PIV laser provided with the DC Image Acquisition System. You can check that by entering the online acquisition part. The interlock then opens and it stays open until you return to the FlowManager main part again. If this is not the case, check that the connection between the Timer box and the laser interlock control is established.
Note:
Note:
During the memory preparation, occasional closing and opening of the interlock may occur.
4-11
5. Troubleshooting
Problem: The very first time images are acquired after the camera is powered on, the frames A and B may be switched. Solution: Stop the acquisition/preview, and the image sequence will be correct in the following acquisitions. Problem: A timeout error is received Solution: Turn off the power for the camera, and on again. Problem: When using trigger mode External sync., the first image received may be exposed at a wrong moment (before or after the laser pulse) Solution: If possible, begin the acquisition before the external trigger source is activated. Problem: If you cannot select the DC System in the the emnu OptionsSystem Unit Type, it may be due to conflicts between the installed drivers and old drivers of the NI-IMAQ or NI-DAQ type. Solution: Uninstall the DC software, the NI-DAQ driver and the NI-IMAQ driver. Repair the FlowManager software and install the DC Image Acquisition software and the NI_DAQ driver again. Problem: You cannot change settings in the timing setup even if no data belong to the current setup. Solution: Exit the online acquisition part and enter again. It can be because you have deleted images, which belong to the currrent setup while you were in the online acquisition part. Problem: No image on either frame A or frame B Solution: Check the cable connecting the Timer Box and the synchronisation board in the PC.
5-1
6. Technical specifications
Number of cameras Image acquisition rate Camera supported 1 or 2 Camera dependent HiSense MkII HiSense 4M FlowSense 2M Trigger mode Laser/shutter trigger Records to acquire System memory Operating system Internal timer or external trigger TTL level Depending on number of cameras and available system memory Up to 2 GBytes (limited by the Windows operating system) Windows 2000/XP
6-1