Models - Bfc.sofc Unit Cell
Models - Bfc.sofc Unit Cell
2a
Model Definition
An SOFC is constructed with two porous gas diffusion electrodes (GDEs) with an electrolyte sandwiched in the middle; see Figure 1.
CATHODE SIDE Cathode inlet
Cathode outlet
Electrolyte
Porous gas diffusion electrode Anode inlet Flow channel ANODE SIDE
Figure 1: Geometry of the unit cell, with anode at the bottom and cathode at the top. The fuel feed in the cathode and anode is counterflow, with hydrogen-rich anode gas entering from the left.
2011 COMSOL
1 |
The electrochemical reactions in the cell are given below. Anode H2 + O Cathode: 1 -- O + 2 e - O 22 2 The model includes the following processes: Electronic charge balance (Ohms law) Ionic charge balance (Ohms law) Butler-Volmer charge transfer kinetics Flow distribution in gas channels (Navier-Stokes) Flow in the porous GDEs (Brinkman equations) Mass balances in gas phase in both gas channels and porous electrodes (Maxwell-Stefan Diffusion and Convection)
CHARGE BALANCES
2-
H 2 O + 2e
The electronic and ionic charge balance in the anode and cathode current feeders, the electrolyte and GDEs are solved for using a Secondary Current Distribution interface. Assume that Butler-Volmer charge transfer kinetics describe the charge transfer current density. At the anode, hydrogen is reduced to form water, and assuming the first electron transfer to be the rate determining step, the following charge transfer kinetics equation applies: c h2 c h2o F 1.5 F - exp 0.5 ----------- ----------------- exp --------------- i a, ct = i 0, a --------------- c h2, ref RT c h2o, ref RT Here i0,a is the anode exchange current density (A/m2), ch2 is the molar concentration of hydrogen, ch2o is the molar concentration of water, ct the total concentration of species (mol/m3). ch2,ref and ch2,ref is the reference concentrations (mol/m3). Furthermore, F is Faradays constant (C/mol), R the gas constant (J/ (molK)), T the temperature (K), and the overvoltage (V). For the cathode, use the relation
2 |
2011 COMSOL
ct 3.5 F 0.5 F - ---------------------------- i c, ct = i 0,c exp ---------- RT x o2 c o2 ,ref exp RT where i0,c is the cathode exchange current density (A/m2), and xo2 is the molar fraction of oxygen. The overvoltage is defined as = electronic ionic eq where eq is the equilibrium potential difference (V). At the anodes inlet boundary, the potential is fixed at a reference potential of zero. At the cathodes inlet boundary, set the potential to the cell voltage, Vcell. The latter is given by V cell = eq,c eq,a V pol (1)
where Vpol is the polarization. In this model, eq,a = 0 V and eq,c = 1 V , and you simulate the fuel cell over the range 0,2 V V cell 0,95 V by using Vpol in the range 0.05 V through 0.8 V as the parameter for the parametric solver. For the ionic charge balance equations, apply insulating boundary conditions at all external boundaries. At the interior boundaries, continuity in current and potential applies by default.
M U L T I C O M P O N E N T TR A N S P O R T
SOFCs can be operated on many different fuels. This model describes a unit running on hydrogen and air. At the anode, a humidified hydrogen gas is supplied as fuel, meaning that the gas consists of two components: hydrogen and water vapor. In the cathode, humidified air is supplied, consisting of three components: oxygen, water vapor, and nitrogen. The material transport is described by the Maxwell-Stefans diffusion and convection equations, solved for by using a Transport of Concentrated Species physics interface for each electrode flow compartment. The boundary conditions at the walls of the gas channel and GDE are zero mass flux (insulating condition). At the inlet, the composition is specified, while the outlet condition is convective flux. This assumption means that the convective term dominates the transport perpendicular to this boundary.
2011 COMSOL
3 |
Continuity in composition and flux apply for all mass balances at the interfaces between the GDEs and the channels.
GAS-FLOW EQUATIONS
A Free and Porous Media Flow physics interface is used for solving for the velocity field and pressure. The weakly compressible Navier-Stokes equations govern the flow in the open channels and the Brinkman equations describe the flow velocity in the porous GDEs. At the inlet and outlet, you set the pressure, specifying a slight overpressure at the inlet to drive the flow (2 Pa at the anode, and 6 Pa at the cathode).
Figure 2: Oxygen mole fraction in the gas channel and in the gas diffusion cathode while operating at a cell voltage of 0.5 V.
4 |
2011 COMSOL
The mole fraction of hydrogen in the anode also decreases along the channel. Figure 3 below shows the distribution of hydrogen. It shows that the depletion is not as pronounced as for the cathode.
Figure 3: Hydrogen distribution in the anode at 0.5 V cell voltage. A consequence of the concentration distribution is that the current density will be nonuniform in the GDEs. Figure 4 depicts the current density distribution at the cathode side of the ionic conductor.
2011 COMSOL
5 |
Figure 4: The current density in the unit cell operating at 0.5 V. The cathode inlet is to the right. As a consequence of oxygen depletion, the current density distribution is poor, with most of the current produced close to the cathode inlet. One way to improve the operating conditions is to increase the cathode flow rate, thus improving the oxygen mass transport.
6 |
2011 COMSOL
Figure 5 shows the voltage as a function of the total current (polarization curve).
2011 COMSOL
7 |
Figure 6 shows the power output as a function of the cell voltage. The model predicts a maximum power-output of 940 W/m2 for the unit cell.
References
1. J. Hartvigsen, S. Elangovan, and A. Khandkar, Science and Technology of Zirconia V, S.P.S. Badwal, M.J. Bannister, and R.H.J. Hannink (eds), p. 682, Technomic Publishing Company Inc., Lancaster, PA, 1993. 2. R. Herbin, J.M. Fiard, and J.R. Ferguson, First European Solid Oxide Fuel Cell Forum Proceedings , U. Bossel (ed.), p. 317, Lucerne, Switzerland, 1994.
8 |
2011 COMSOL
Modeling Instructions
MODEL WIZARD
1 Go to the Model Wizard window. 2 Click Next. 3 In the Add physics tree, select Electrochemistry>Secondary Current Distribution (siec). 4 Click Add Selected. 5 In the Add physics tree, select Chemical Species Transport>Transport of Concentrated Species (chcs). 6 Click Add Selected. 7 Click Add Mass Fraction. 8 In the Mass fractions table, enter the following settings:
wO2 wH2Oc wN2
9 In the Add physics tree, select Chemical Species Transport>Transport of Concentrated Species (chcs). 1 0 Click Add Selected. 1 1 In the Mass fractions table, enter the following settings:
wH2 wH2Oa
1 2 In the Add physics tree, select Fluid Flow>Porous Media and Subsurface Flow>Free and Porous Media Flow (fp). 1 3 Click Add Selected. 1 4 Click Next. 1 5 Find the Studies subsection. In the tree, select Preset Studies for Selected Physics>Stationary. 1 6 Click Finish.
GLOBAL DEFINITIONS
2011 COMSOL
9 |
Parameters
1 In the Model Builder window, right-click Global Definitions and choose Parameters. 2 Go to the Settings window for Parameters. 3 Locate the Parameters section. Click Load from File. 4 Browse to the models Model Library folder and double-click the file
sofc_unit_cell_parameters.txt.
Here, R_const and F_const are predefined constants representing the gas constant and the Faraday constant, respectively.
GEOMETRY 1
In this model you will first create the 2D cross section of the device. Then you will extrude it to create your 3D model geometry. Then you creat a mapped mesh on the faces at one end of the model. This mapped mesh will then be swept along the length. This technique is particularly useful if you have a stretched-out object where you want to spend a few mesh layers along the object length.
Work Plane 1
1 In the Model Builder window, right-click Model 1>Geometry 1 and choose Work Plane. 2 Go to the Settings window for Work Plane. 3 Locate the Work Plane section. From the Plane list, choose yz-plane.
Rectangle 1
1 In the Model Builder window, right-click Geometry and choose Rectangle. 2 Go to the Settings window for Rectangle. 3 Locate the Size section. In the Width edit field, type W_channel+W_rib. 4 In the Height edit field, type H_gde. 5 Click the Build Selected button. 6 Click the Zoom Extents button on the Graphics toolbar.
Rectangle 2
1 In the Model Builder window, right-click Work Plane 1>Geometry and choose Rectangle. 2 Go to the Settings window for Rectangle. 3 Locate the Size section. In the Width edit field, type W_channel+W_rib. 4 In the Height edit field, type H_electrolyte.
10 |
2011 COMSOL
5 Locate the Position section. In the y edit field, type -H_electrolyte. 6 Click the Build Selected button.
Rectangle 3
1 In the Model Builder window, right-click Work Plane 1>Geometry and choose Rectangle. 2 Go to the Settings window for Rectangle. 3 Locate the Size section. In the Width edit field, type W_channel+W_rib. 4 In the Height edit field, type H_gde. 5 Locate the Position section. In the y edit field, type -H_electrolyte-H_gde. 6 Click the Build Selected button.
Rectangle 4
1 In the Model Builder window, right-click Work Plane 1>Geometry and choose Rectangle. 2 Go to the Settings window for Rectangle. 3 Locate the Size section. In the Width edit field, type W_channel. 4 In the Height edit field, type H_channel. 5 Locate the Position section. In the x edit field, type W_rib/2. 6 In the y edit field, type H_gde. 7 Click the Build Selected button.
Rectangle 5
1 In the Model Builder window, right-click Work Plane 1>Geometry and choose Rectangle. 2 Go to the Settings window for Rectangle. 3 Locate the Size section. In the Width edit field, type W_channel. 4 In the Height edit field, type H_channel. 5 Locate the Position section. In the x edit field, type W_rib/2. 6 In the y edit field, type -H_gde-H_electrolyte-H_channel. 7 Click the Build Selected button. 8 Click the Zoom Extents button on the Graphics toolbar.
Extrude 1
1 In the Model Builder window, right-click Work Plane 1 and choose Extrude. 2 Go to the Settings window for Extrude.
2011 COMSOL
11 |
3 Locate the Distances from Work Plane section. In the table, enter the following
settings:
DISTANCES (M)
4 Click the Build All button. 5 Click the Zoom Extents button on the Graphics toolbar.
The model geometry is now complete, and it should look like that in Figure 1.
DEFINITIONS
Explicit 1
1 In the Model Builder window, right-click Model 1>Definitions and choose Selections>Explicit. 2 Select Domain 4 only. 3 Right-click Explicit 1 and choose Rename. 4 Go to the Rename Explicit dialog box and type Anode Flow Channel in the New name edit field. 5 Click OK.
Explicit 2
1 Right-click Definitions and choose Explicit. 2 Select Domain 1 only. 3 Right-click Explicit 2 and choose Rename. 4 Go to the Rename Explicit dialog box and type Anode Electrode in the New name
edit field.
5 Click OK.
Explicit 3
1 Right-click Definitions and choose Explicit. 2 Select Domain 3 only. 3 Right-click Explicit 3 and choose Rename. 4 Go to the Rename Explicit dialog box and type Cathode Electrode in the New name
edit field.
5 Click OK.
12 |
2011 COMSOL
Explicit 4
1 Right-click Definitions and choose Explicit. 2 Select Domain 5 only. 3 Right-click Explicit 4 and choose Rename. 4 Go to the Rename Explicit dialog box and type Cathode Flow Channel in the New name edit field. 5 Click OK.
Variables 1
1 Right-click Definitions and choose Variables. 2 Go to the Settings window for Variables. 3 Locate the Geometric Entity Selection section. From the Geometric entity level list,
choose Domain.
4 From the Selection list, choose Anode Electrode. 5 Locate the Variables section. In the Variables table, enter the following settings:
NAME EXPRESSION DESCRIPTION
eta_a ict_a
Variables 2
1 In the Model Builder window, right-click Definitions and choose Variables. 2 Go to the Settings window for Variables. 3 Locate the Geometric Entity Selection section. From the Geometric entity level list,
choose Domain.
4 From the Selection list, choose Cathode Electrode.
2011 COMSOL
13 |
5 Locate the Variables section. In the Variables table, enter the following settings:
NAME EXPRESSION DESCRIPTION
eta_c ict_c
6 In the Model Builder window, right-click Definitions and choose Probes>Boundary Probe. 7 Go to the Settings window for Boundary Probe. 8 Locate the Probe Settings section. From the Type list, choose Average. 9 Locate the Source Selection section. Click Clear Selection. 10 Select Boundary 6 only. 11 In the upper-right corner of the Expression section, click Replace Expression. 12 From the menu, choose Secondary Current Distribution>Electrolyte current density vector>Electrolyte current density vector, z component (siec.Ilz).
SECONDARY CURRENT DISTRIBUTION
1 In the Model Builder window, click Model 1>Secondary Current Distribution. 2 Select Domains 13 only.
Porous Electrode 1
1 Right-click Model 1>Secondary Current Distribution and choose Porous Electrode. 2 Go to the Settings window for Porous Electrode. 3 Locate the Domain Selection section. From the Selection list, choose Anode Electrode. 4 Locate the Electrolyte Current Conduction section. From the l list, choose User defined. In the associated edit field, type kleff_a. 5 From the Effective conductivity correction list, choose No correction. 6 Locate the Electrode Current Conduction section. From the s list, choose User defined. In the associated edit field, type kseff_a. 7 From the Effective conductivity correction list, choose No correction.
14 |
2011 COMSOL
Porous Electrode 2
1 In the Model Builder window, right-click Secondary Current Distribution and choose Porous Electrode. 2 Go to the Settings window for Porous Electrode. 3 Locate the Domain Selection section. From the Selection list, choose Cathode Electrode. 4 Locate the Electrolyte Current Conduction section. From the l list, choose User defined. In the associated edit field, type kleff_c. 5 From the Effective conductivity correction list, choose No correction. 6 Locate the Electrode Current Conduction section. From the s list, choose User defined. In the associated edit field, type kseff_c. 7 From the Effective conductivity correction list, choose No correction.
Electrolyte 1
1 In the Model Builder window, click Electrolyte 1. 2 Go to the Settings window for Electrolyte. 3 Locate the Electrolyte section. From the l list, choose User defined. In the associated
2011 COMSOL
15 |
Electric Ground 1
1 In the Model Builder window, right-click Secondary Current Distribution and choose Electrode>Electric Ground. 2 Select Boundaries 3 and 20 only.
Electric Potential 1
1 In the Model Builder window, right-click Secondary Current Distribution and choose Electrode>Electric Potential. 2 Select Boundaries 10 and 22 only. 3 Go to the Settings window for Electric Potential. 4 Locate the Electric Potential section. In the s,bnd edit field, type V_cell.
Initial Values 2
1 In the Model Builder window, right-click Secondary Current Distribution and choose Initial Values. 2 Select Domain 3 only. 3 Go to the Settings window for Initial Values. 4 Locate the Initial Values section. In the phis edit field, type V_cell.
TR A N S P O R T O F C O N C E N T R A T E D S P E C I E S
3 Click OK. 4 Select Domains 3 and 5 only. 5 Go to the Settings window for Transport of Concentrated Species. 6 Locate the Transport Mechanisms section. From the Diffusion model list, choose Maxwell-Stefan. 7 Locate the Species section. From the From mass constraint list, choose wN2.
TR A N S P O R T O F C O N C E N T R A T E D S P E C I E S - C A T H O D E
16 |
2011 COMSOL
3 Locate the Density section. From the Mixture density list, choose Ideal gas. 4 In the MwO2 edit field, type Mo2. 5 In the MwH2Oc edit field, type Mh2o. 6 In the MwN2 edit field, type Mn2. 7 Locate the Diffusion section. In the Dik table, enter the following settings:
1 Do2h2oeff Do2n2eff
Do2h2oeff
Do2n2eff Dn2h2oeff
1 Dn2h2oeff
8 Locate the Model Inputs section. From the u list, choose Velocity field (fp/fp1). 9 In the T edit field, type T. 1 0 From the p list, choose Pressure (fp/fp1). 1 1 In the pref edit field, type p_atm.
Initial Values 1
1 In the Model Builder window, click Initial Values 1. 2 Go to the Settings window for Initial Values. 3 Locate the Initial Values section. In the wO2 edit field, type w_o2ref. 4 In the wH2Oc edit field, type w_h2oref.
Do2h2o
Do2n2
2011 COMSOL
17 |
Do2h2o Do2n2
1 Dn2h2o
Dn2h2o
9 Locate the Model Inputs section. From the u list, choose Velocity field (fp/fp1). 10 In the T edit field, type T. 11 From the p list, choose Pressure (fp/fp1). 12 In the pref edit field, type p_atm.
Reaction Coefficients 1
1 In the Model Builder window, expand the Porous Electrode Coupling 1 node, then click Reaction Coefficients 1. 2 Go to the Settings window for Reaction Coefficients. 3 Locate the Model Inputs section. From the iv list, choose Local current source (siec/ per1). 4 Locate the Stoichiometric Coefficients section. In the nm edit field, type 4. 5 In the wO2 edit field, type -1.
Inflow 1
1 In the Model Builder window, right-click Transport of Concentrated Species - Cathode
Outflow 1
1 In the Model Builder window, right-click Transport of Concentrated Species - Cathode
18 |
2011 COMSOL
TR A N S P O R T O F C O N C E N T R A T E D S P E C I E S 2
1 In the Model Builder window, right-click Model 1>Transport of Concentrated Species 2 and choose Rename. 2 Go to the Rename Transport of Concentrated Species dialog box and type Transport
of Concentrated Species 2 - Anode in the New name edit field.
3 Click OK. 4 Select Domains 1 and 4 only. 5 Go to the Settings window for Transport of Concentrated Species. 6 Locate the Transport Mechanisms section. From the Diffusion model list, choose Maxwell-Stefan. 7 Locate the Species section. From the From mass constraint list, choose wH2Oa.
TR A N S P O R T O F C O N C E N T R A T E D S P E C I E S 2 - A N O D E
1 Dh2h2oeff
Dh2h2oeff
7 Locate the Model Inputs section. From the u list, choose Velocity field (fp/fp1). 8 In the T edit field, type T. 9 From the p list, choose Pressure (fp/fp1). 1 0 In the pref edit field, type p_atm.
Initial Values 1
1 In the Model Builder window, click Initial Values 1. 2 Go to the Settings window for Initial Values. 3 Locate the Initial Values section. In the wH2 edit field, type w_h2ref.
2011 COMSOL
19 |
1 Dh2h2o
Dh2h2o
12 Locate the Model Inputs section. From the u list, choose Velocity field (fp/fp1). 13 In the T edit field, type T. 14 From the p list, choose Pressure (fp/fp1). 15 In the pref edit field, type p_atm.
Reaction Coefficients 1
1 In the Model Builder window, click Reaction Coefficients 1. 2 Go to the Settings window for Reaction Coefficients. 3 Locate the Model Inputs section. From the iv list, choose Local current source (siec/ per1). 4 Locate the Stoichiometric Coefficients section. In the nm edit field, type 2.
20 |
2011 COMSOL
5 In the wH2 edit field, type 1. 6 In the wH2Oa edit field, type -1.
Inflow 1
1 In the Model Builder window, right-click Transport of Concentrated Species 2 - Anode
Outflow 1
1 In the Model Builder window, right-click Transport of Concentrated Species 2 - Anode
1 In the Model Builder window, click Model 1>Free and Porous Media Flow. 2 Select Domains 1 and 35 only. 3 Go to the Settings window for Free and Porous Media Flow. 4 Locate the Physical Model section. From the Compressibility list, choose Compressible flow (Ma<0.3).
Fluid Properties 1
1 In the Model Builder window, expand the Free and Porous Media Flow node, then click Fluid Properties 1. 2 Go to the Settings window for Fluid Properties. 3 Locate the Fluid Properties section. From the list, choose Density (chcs2/chcs2). 4 From the list, choose User defined. In the associated edit field, type mu.
Wall 2
1 In the Model Builder window, right-click Free and Porous Media Flow and choose Wall. 2 Select Boundaries 2, 8, 23, and 25 only. 3 Go to the Settings window for Wall. 4 Locate the Boundary Condition section. From the Boundary condition list, choose Slip.
2011 COMSOL
21 |
Fluid Properties 2
1 In the Model Builder window, right-click Free and Porous Media Flow and choose Fluid Properties. 2 Select Domains 3 and 5 only. 3 Go to the Settings window for Fluid Properties. 4 Locate the Fluid Properties section. From the list, choose Density (chcs/chcs). 5 From the list, choose User defined. In the associated edit field, type mu.
Mh2 Mh2o
Reaction Coefficients 1
1 In the Model Builder window, expand the Porous Electrode Coupling 1 node, then click Reaction Coefficients 1. 2 Go to the Settings window for Reaction Coefficients. 3 Locate the Model Inputs section. From the iv list, choose Local current source (siec/ per1).
22 |
2011 COMSOL
4 Locate the Stoichiometric Coefficients section. In the nm edit field, type 2. 5 In the 1 edit field, type 1. 6 In the 2 edit field, type -1.
Reaction Coefficients 1
1 In the Model Builder window, expand the Porous Electrode Coupling 2 node, then click Reaction Coefficients 1. 2 Go to the Settings window for Reaction Coefficients. 3 Locate the Model Inputs section. From the iv list, choose Local current source (siec/ per1).
2011 COMSOL
23 |
4 Locate the Stoichiometric Coefficients section. In the nm edit field, type 4. 5 In the 1 edit field, type -1.
Initial Values 1
1 In the Model Builder window, click Initial Values 1. 2 Go to the Settings window for Initial Values. 3 Locate the Initial Values section. In the p edit field, type 0.
Inlet 1
1 In the Model Builder window, right-click Free and Porous Media Flow and choose Inlet. 2 Select Boundary 30 only. 3 Go to the Settings window for Inlet. 4 Locate the Boundary Condition section. From the Boundary condition list, choose Pressure, no viscous stress. 5 Locate the Pressure, No Viscous Stress section. In the p0 edit field, type dp_c.
Inlet 2
1 In the Model Builder window, right-click Free and Porous Media Flow and choose Inlet. 2 Select Boundary 11 only. 3 Go to the Settings window for Inlet. 4 Locate the Boundary Condition section. From the Boundary condition list, choose Pressure, no viscous stress. 5 Locate the Pressure, No Viscous Stress section. In the p0 edit field, type dp_a.
Outlet 1
1 In the Model Builder window, right-click Free and Porous Media Flow and choose Outlet. 2 Select Boundaries 15 and 29 only. 3 Go to the Settings window for Outlet. 4 Locate the Boundary Condition section. From the Boundary condition list, choose Laminar outflow. 5 Click the Exit pressure button. 6 In the Lexit edit field, type 1e-3. 7 Select the Constrain outer edges to zero check box.
24 |
2011 COMSOL
MESH 1
Mapped 1
1 In the Model Builder window, right-click Model 1>Mesh 1 and choose More Operations>Mapped. 2 Select Boundaries 1, 4, 7, 11, and 15 only.
Distribution 1
1 Right-click Mapped 1 and choose Distribution. 2 Click the Zoom Extents button on the Graphics toolbar. 3 Select Edges 12, 13, 17, 20, 22, and 26 only. 4 Go to the Settings window for Distribution. 5 Locate the Distribution section. In the Number of elements edit field, type 5.
Distribution 2
1 In the Model Builder window, right-click Mapped 1 and choose Distribution. 2 Select Edges 2, 10, 24, and 27 only. 3 Go to the Settings window for Distribution. 4 Locate the Distribution section. In the Number of elements edit field, type 6.
Distribution 3
1 In the Model Builder window, right-click Mapped 1 and choose Distribution. 2 Select Edges 1 and 30 only. 3 Go to the Settings window for Distribution. 4 Locate the Distribution section. In the Number of elements edit field, type 6.
Distribution 4
1 In the Model Builder window, right-click Mapped 1 and choose Distribution. 2 Select Edges 7 and 34 only. 3 Go to the Settings window for Distribution. 4 Locate the Distribution section. In the Number of elements edit field, type 6.
Distribution 5
1 In the Model Builder window, right-click Mapped 1 and choose Distribution. 2 Select Edges 4 and 32 only. 3 Go to the Settings window for Distribution. 4 Locate the Distribution section. In the Number of elements edit field, type 1.
2011 COMSOL
25 |
Mapped 1
In the Model Builder window, right-click Mapped 1 and choose Build Selected.
Swept 1
Right-click Mesh 1 and choose Swept.
Distribution 1
1 In the Model Builder window, right-click Swept 1 and choose Distribution. 2 Go to the Settings window for Distribution. 3 Locate the Distribution section. In the Number of elements edit field, type 10.
Swept 1
In the Model Builder window, right-click Swept 1 and choose Build Selected.
STUDY 1
1 In the Model Builder window, right-click Study 1 and choose Show Default Solver. 2 Expand the Study 1>Solver Configurations node.
Solver 1
1 In the Model Builder window, expand the Study 1>Solver Configurations>Solver 1
node. This rather small problem solves more quickly using a direct solver.
2 In the Model Builder window, expand the Stationary Solver 1 node. 3 Right-click Stationary Solver 1>Direct and choose Enable. 4 Right-click Stationary Solver 1 and choose Parametric. 5 Go to the Settings window for Parametric. 6 Locate the General section. From the Defined by study step list, choose User defined. 7 In the Parameter names edit field, type V_pol. 8 In the Parameter values edit field, type 0.05 range(0.1,0.1,0.8). 9 In the Model Builder window, right-click Study 1 and choose Compute.
3D Plot Group 10
Follow these instructions to reproduce the plot in Figure 2 of the oxygen distribution in the anode at a cell voltage of 0.5 V.
26 |
2011 COMSOL
1 In the Model Builder window, right-click Results and choose 3D Plot Group. 2 Right-click Results>3D Plot Group 10 and choose Rename. 3 Go to the Rename 3D Plot Group dialog box and type Oxygen Mole Fraction in
3D Plot Group 11
The following steps reproduce the corresponding plot for hydrogen (compare with Figure 3).
1 In the Model Builder window, right-click Results and choose 3D Plot Group. 2 Right-click Results>3D Plot Group 11 and choose Rename. 3 Go to the Rename 3D Plot Group dialog box and type Hydrogen Mole Fraction in
2011 COMSOL
27 |
4 From the menu, choose Transport of Concentrated Species 2 - Anode>Species wH2>Mole fraction (chcs2.x_wH2). 5 Locate the Plane Data section. From the Plane list, choose yz-planes. 6 In the Planes edit field, type 20. 7 Click the Plot button.
1D Plot Group 12
The following instructions reproduce the plot of the polarization curve for the SOFC (see Figure 5).
1 In the Model Builder window, right-click Results and choose 1D Plot Group. 2 Right-click Results>1D Plot Group 12 and choose Rename. 3 Go to the Rename 1D Plot Group dialog box and type Polarization curve in the New name edit field. 4 Click OK. 5 Go to the Settings window for 1D Plot Group. 6 Locate the Title section. From the Title type list, choose Manual. 7 In the Title text area, type Polarization curve. 8 Locate the Plot Settings section. Select the x-axis label check box. 9 In the associated edit field, type Average current density (A/m<sup>2</
sup>).
10 Select the y-axis label check box. 11 In the associated edit field, type V<sub>cell</sub> (V).
Polarization curve
1 Right-click Results>1D Plot Group 12 and choose Global. 2 Go to the Settings window for Global. 3 Locate the y-Axis Data section. In the table, enter the following settings:
EXPRESSION DESCRIPTION
V_cell
Cell Voltage
4 Locate the x-Axis Data section. From the Parameter list, choose Expression. 5 In the Expression edit field, type bnd1. 6 Click the Plot button.
28 |
2011 COMSOL
1D Plot Group 13
Next, reproduce a plot showing the power output as a function of the cell voltage (Figure 6.
1 In the Model Builder window, right-click Results and choose 1D Plot Group. 2 Right-click Results>1D Plot Group 13 and choose Rename. 3 Go to the Rename 1D Plot Group dialog box and type Power vs Current in the New name edit field. 4 Click OK. 5 Go to the Settings window for 1D Plot Group. 6 Locate the Title section. From the Title type list, choose Manual. 7 In the Title text area, type Total output power. 8 Locate the Plot Settings section. Select the x-axis label check box. 9 In the associated edit field, type Average current density (A/m<sup>2</
sup>).
1 0 Select the y-axis label check box. 1 1 In the associated edit field, type Average Cell Power (W/m<sup>2</sup>).
Power vs Current
1 Right-click Results>1D Plot Group 13 and choose Global. 2 Go to the Settings window for Global. 3 Locate the y-Axis Data section. In the table, enter the following settings:
EXPRESSION DESCRIPTION
V_cell*bnd1
4 Locate the x-Axis Data section. From the Parameter list, choose Expression. 5 In the Expression edit field, type bnd1. 6 Click the Plot button.
3D Plot Group 14
Next reproduce the plot in Figure 4 showing the current density in the unit cell at 0.5 V.
1 In the Model Builder window, right-click Results and choose 3D Plot Group. 2 Right-click Results>3D Plot Group 14 and choose Rename.
2011 COMSOL
29 |
3 Go to the Rename 3D Plot Group dialog box and type Electrolyte current
density in the New name edit field.
4 Click OK.
Before defining the plot, add the data set for the Boundary 9.
Data Sets
1 In the Model Builder window, right-click Results>Data Sets and choose Solution. 2 Right-click Probe Solution 2 and choose Add Selection. 3 Go to the Settings window for Selection. 4 Locate the Geometric Entity Selection section. From the Geometric entity level list,
choose Boundary.
5 Select Boundary 9 only.
30 |
2011 COMSOL