Model E1005 Indicator

E1005

User Instructions
ENGLISH

43091-0018 Issue AC

August 30, 2007

*43091-0018*

E1005_rev2006_u.book

Avery Weigh-Tronix LLC, 2006. All rights reserved.


The information contained herein is the property of Avery WeighTronix, Inc and is supplied without liability for errors or omissions.
No part may be reproduced or used except as authorized by
contract or other written permission. The copyright and the
foregoing restriction on reproduction and use extend to all media
in which the information may be embodied.
Avery Weigh-Tronix is a trading name of Weigh-Tronix, Inc
Trademarks and acknowledgements
Avery Weigh-Tronix, Avery Berkel, Dillon, NCI and Salter
Brecknell are registered trademarks in certain jurisdictions.
All third party brands and product names used within this
document are trademarks or registered trademarks of their
respective holders.
IMPORTANT
When programming or configuring the equipment you must ensure
that you comply with all relevant standards and legislation. The
example settings given in this book may not be legal for trade with
the public.

Declarations of compliance
United States
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment
in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.

Canada
This digital apparatus does not exceed the Class A limits for radio noise
emissions from digital apparatus set out in the Radio Interference
Regulations of the Canadian Department of Communications.
Le prsent appareil numrique nmet pas de bruits radiolectriques
dpassant les limites applicables aux appareils numriques de la Classe A
prescrites dans le Rglement sur le brouillage radiolectrique edict par le
ministre des Communications du Canada.

European Countries
WARNING
This is a Class A product. In a domestic environment this product may cause
radio interference in which the user may be required to take adequate
measures.

Model E1005 Indicator User Instructions

CAUTION: DANGER OF EXPLOSION IF BATTERY IS INCORRECTLY


REPLACED. REPLACE ONLY WITH THE SAME OR EQUIVALENT TYPE
RECOMMENDED BY THE MANUFACTURER. DISPOSE OF USED
BATTERIES ACCORDING TO THE MANUFACTURER'S INSTRUCTIONS.
ATTENTION: IL Y A DANGER D'EXPLOSION S'IL Y A REMPLACEMENT
INCORRECT DE LA BATTERIE, REMPLACER UNIQUEMENT AVEC UNE
BATTERIE DU MME TYPE OU D'UN TYPE QUIVALENT
RECOMMAND PAR LE CONSTRUCTEUR. METTRE AU REBUT LES
BATTERIES USAGES CONFORMMENT AUX INSTRUCTIONS DU
FABRICANT.
CAUTION: THE POWER SUPPLY CORD IS USED AS THE MAIN
DISCONNECT DEVICE, ENSURE THAT THE SOCKET-OUTLET IS
LOCATED/INSTALLED NEAR THE EQUIPMENT AND IS EASILY
ACCESSIBLE
ATTENTION: LE CORDON D'ALIMENTATION EST UTILIS COMME
INTERRUPTEUR GNRAL. LA PRISE DE COURANT DOIT TRE
SITUE OU INSTALLE PROXIMIT DE L'QUIPEMENT ET TRE
FACILE D'ACCS".

Model E1005 Indicator User Instructions

Declaration of Conformity

Manufacturer

Avery Weigh-Tronix

Type

E1005, E1010

No. of EC type approval certificate

UK 2723

corresponds to the requirements of the following EC directives:


NonAutomatic Weighing Instruments
Directive

90/384/EEC 1

EMC Directive

89/336/EEC

Low Voltage Directive

73/23/EEC

The applicable harmonised standards


are:

EN 45501, EN 61000-4, EN 55022, EN


60950

Note 1 : This declaration is only valid if the nonautomatic weighing


instrument was verified by the manufacturer or with a certificate of
conformity issued by a notified body.

A copy of the original signed declaration for this instrument is available


from:
Avery Weigh-Tronix, Foundry Lane, Smethwick, West Midlands B66 2LP England

Verklaring van Overeenstemming

Fabrikant

Avery Weigh-Tronix

Type

E1005, E1010

Nummer van de Verklaring van EGtypegoedkeuring

UK2723

is in overeenstemming met de voorschriften van de volgende EG-richtlijnen:


Richtlijn Niet-automatische
Weegwerktuigen

90/384/EEG 1

Richtlijn EMC

89/336/EEG

Richtlijn Laagspanning

73/23/EEG

Toegepaste geharmoniseerde normen:

EN 45501, EN 61000-4
EN 55022, EN60950

Noot 1 : Deze verklaring is alleen geldig indien het weegwerktuig door


de fabrikant is geverifieerd of met een Verklaring van
Overeenstemming, afgegeven door een bevoegde instantie

Een kopie van de originele ondertekende verklaring is verkrijgbaar


door contact op te nemen met:
Avery Weigh-Tronix, Foundry Lane, Smethwick, West Midlands B66 2LP England

Dclaration de Conformit

Fabricant

Avery Weigh-Tronix

Type

E1005, E1010

No. de certificat dapprobation de type


CE

UK2723

correspond aux exigences des directives CE suivantes :


Directive pour les instruments de pesage
fonctionnement non automatique

90/384/CEE 1

Directive CEM

89/336/CEE

Directive Basse Tension

73/23/CEE

Les normes harmonises applicables


sont :

EN 45501, EN 61000-4
EN 55022, EN 60950

Nota 1 : Cette dclaration est valide seulement si linstrument de


pesage fonctionnement non automatique a t vrifi par le fabricant
ou avec une attestation de conformit dlivre par un organisme
notifi.

Une copie originale de la dclaration signe pour cet appareil est


disponible ladresse suivante :
Avery Weigh-Tronix, Foundry Lane, Smethwick, West Midlands B66 2LP England

Konformittserklrung

Hersteller

Avery Weigh-Tronix

Typ

E1005, E1010

Nr. der EU-Bauartzulassung

UK2723

entspricht den Anforderungen folgender EU-Richtlinien:


Waagenrictlinie fr nichtselbstttige
Waagen

90/384/EWG1

EMV-Richtlinie

89/336/EWG

Niederspannungsrichtlinie

73/23/EWG

Die angewendeten harmonisierten


Normen sind:

EN 45501, EN 61000-4
EN 55022, EN 60950

Anmerkung 1: Diese Erklrung gilt nur, wenn die nichtselbstttige


Waage vom Hersteller geeicht wurde oder in Verbindung mit einer
Konformittsbescheinigung einer benannten Stelle.

Eine Abschrift der original unterschriebenen Konformittserklrung ist


unter untengenannter Adresse erhltlich von:
Avery Weigh-Tronix, Foundry Lane, Smethwick, West Midlands B66 2LP England

Dichiarazione di conformit

Produttore

Avery Weigh-Tronix

Tipo

E1005, E1010

N. - certificato di approvazione
CE

UK2723

conforme alle caratteristiche previste dalle seguenti direttive CE:


Normativa per satrumenti di
pesatura non automatici

90/384/CEE 1

Normativa EMC

89/336/CEE

Normativa per la bassa tensione

73/23/CEE

Le norme standard armonizzate


applicate sono:

EN 45501, EN 61000-4
EN 55022, EN 60950

Nota 1 : Questa dichiarazione valida solamente se lo strumento di


pesatura non automatico stato verificato dal produttore o
provvisto di un certificato di conformit rilasciato da un ente
riconosciuto.

Se richiesta, disponibile una copia del certificato originale firmato


presso:
Avery Weigh-Tronix, Foundry Lane, Smethwick, West Midlands B66 2LP England

Declaracin de Conformidad

Fabricante

Avery Weigh-Tronix

Tipo

E1005, E1010

Nmero del certificado de


homologacin CE

UK2723

conforme a las exigencias de las siguientes directivas CE:


Directiva para equipos de pesaje
no automticos

90/384/EEC 1

Directiva CEM

89/336/EEC

Directiva de baja tensin

73/23/EEC

Las normas armonizadas en


vigor son:

EN 45501, EN 61000-4
EN 55022, EN60950

Nota 1 : Esta declaracin es vlida solamente si el equipo de pesaje


no automtico ha sido verificado por el fabricante o con
certificado de conformidad emitido por un organismo
notificado.

Una copia del certificado de homologacin que corresponde a este


equipo se puede obtener en:
Avery Weigh-Tronix, Foundry Lane, Smethwick, West Midlands B66 2LP England

10

Table of Contents

page
Table of Contents ................................................................................................11
Safety and Warnings ...........................................................................................13
Installation and service (English) ...................................................................13
Installatie en service (Dutch) .........................................................................15
Installation et Service (French) ......................................................................17
Installation und Wartung (German) ...............................................................19
Installazione e manutenzione (Italian) ...........................................................21
Instalacin y Mantenimiento (Spanish) ..........................................................23
Introduction .........................................................................................................25
About This Manual .........................................................................................25
Unpacking and Setup ....................................................................................26
Front Panel ..........................................................................................................27
Keys ...............................................................................................................27
Numeric Entry Procedure ..............................................................................29
Battery Information ..............................................................................................30
Menu Mode .........................................................................................................31
User Menu .....................................................................................................31
Supervisor Menu ...........................................................................................32
DATE submenu (Set date) .......................................................................33
HOUR submenu (Set time) ......................................................................33
SETUP submenu (Setup menu) ...............................................................34
TEST submenu (Test menu) ....................................................................39
AUDIT (Audit counters) ............................................................................40
SLEEP (Sleep mode) menu .....................................................................41
Indicator Operations ............................................................................................42
General Weighing ..........................................................................................42
Using Outputs ..........................................................................................43
Accumulator Weighing ...................................................................................43
Using Outputs ..........................................................................................44
Checkweighing ..............................................................................................45
Limit Type: Entering Upper and Lower Limits ..........................................46

Sample Type: Using Product to Set Target Weight .................................46


Performing a Checkweighing Weighment ................................................46
Using Outputs ..........................................................................................46
Counting ........................................................................................................47
Using Outputs ..........................................................................................47
Batch Weighing .............................................................................................48
Peak Weighing ..............................................................................................49
Using Outputs ..........................................................................................49
Remote Display .............................................................................................50
Communications .................................................................................................51
Error Messages ...................................................................................................52
Indicator Diagnostics ...........................................................................................53
Specifications ......................................................................................................55

Safety and Warnings

1.1

Installation and service (English)


THE EQUIPMENT CONTAINS NO USER SERVICEABLE COMPONENTS.
Installation and maintenance of the equipment must only be carried out by
trained and authorized personnel.

1.2

Electrical installation
The mains lead must be connected to a supply outlet with a protective earth
contact. The electrical supply at the socket outlet must provide over current
protection of an appropriate rating.
For your protection all mains (110V or 230V) equipment used out of doors
or in wet or damp conditions should be supplied from a correctly fused
source and protected by an approved ground fault protection device (RCD,
GFCI etc.)
IF IN DOUBT SEEK ADVICE FROM A QUALIFIED ELECTRICIAN.

Pluggable equipment

Pluggable equipment must be installed near an easily accessible socket


outlet.

Permanently wired equipment Isolator requirements

Permanently connected equipment must have a readily accessible


disconnect device incorporated in the fixed wiring. (An isolator or circuit
breaker with at least 3mm contact separation).
The isolator MUST NOT be installed into the flexible mains cable supplied
with the unit.

Electrical installation (supplementary notes for wet


conditions)

Under wet conditions the plug must be connected to the final branch circuit
via an appropriate socket / receptacle designed for wash down use.
Installations within the USA should use a Rain tight while in use
specification grade cover, such as those manufactured by Leviton.
These allow the unit to be plugged in with a rain tight cover fitted over the
plug / receptacle allowing it to meet a NEMA 3R rating, as required by the
national electrical code under section 410-57. Note that the particular cover
mentioned is listed in UL file #E13397.
Installations within Europe must use a socket which provides a minimum
of IP56 protection to the plug / cable assembly. Care must be taken to
ensure that the degree of protection provided by the socket is suitable for the
environment.

Model E1005 Indicator User Instructions

13

Safety and Warnings

1.3

Routine maintenance
To avoid the possibility of electric shock or damage to the machine,
always switch off the machine and isolate from the power supply
before carrying out any routine maintenance.
To avoid the risk of the machine falling, where applicable, ensure that it is
placed securely on a flat and level surface.

1.4

Cleaning the machine


The outside of standard products may be wiped down with a clean cloth,
moistened with water containing a small amount of mild detergent.
Harsh abrasives, solvents, scouring cleaners and alkaline cleaning
solutions, such as washing soda, should not be used especially on
the display windows. Under no circumstances should you attempt to
wipe the inside of the machine.
Do not spray any liquid directly onto the display windows. If you are
using a proprietary cleaning fluid ensure you spray the cloth and not
the display.

1.5

Training
Do not attempt to operate or carry out any procedure on a machine unless
you have received the appropriate training or read the Instruction Books.
To avoid the risk of RSI (Repetitive Strain Injury) it is important to ensure
that the machine is placed on a surface which is ergonomically satisfactory
to the user. It is recommended that frequent breaks are taken during
prolonged usage.

1.6

Sharp objects
Do not use sharp objects (screwdrivers, long fingernails etc.) to operate
the keys.

1.7

EMC compliance
The following warning may be applicable to your machine.
WARNING: This is a class A product. In a domestic environment this
product may cause radio interference in which case the user may be
required to take adequate measures.

14

Model E1005 Indicator User Instructions

1.8 Installatie en service (Dutch)

1.8

Installatie en service (Dutch)


DE APPARATUUR BEVAT GEEN VOOR EEN SERVICEBEURT IN
AANMERKING KOMENDE ONDERDELEN.
Installatie en onderhoud van de apparatuur mag alleen worden uitgevoerd
door getraind en bevoegd personeel.

1.9

Elektrische installatie
De netstroomkabel moet worden aangesloten op een geaarde contactdoos.
De stroomvoorziening op de contactdoos moet overstroombeveiliging van
een toelaatbaar vermogen bieden.
Voor uw veiligheid moet alle in open lucht of onder natte of vochtige
omstandigheden gebruikte netstroomapparatuur (110 V of 230 V) zijn
voorzien van een goedgekeurde aansluitingsbescherming (RCD, GFCI
enz.)
BIJ TWIJFEL DIENT ADVIES TE WORDEN GEVRAAGD AAN EEN
GEKWALIFICEERDE ELEKTRICIEN.

Op een contactdoos aan te


sluiten apparatuur

Op een contactdoos aan te sluiten apparatuur moet worden genstalleerd in


de buurt van een goed te bereiken wandcontactdoos.

Permanent bedrade apparatuur


- Vereisten voor
isolatieschakelaar

Apparatuur die permanent op een contactdoos is aangesloten, moet in de


vaste bedrading zijn voorzien van een goed te bereiken
scheidingsschakelaar. (Een isolatieschakelaar of stroomonderbreker met
een minimale contactscheiding van 3 mm).
De isolatieschakelaar MAG NIET worden genstalleerd in de met de
installatie meegeleverde flexibele netstroomkabel.

Elektrische installatie (aanvullende opmerkingen voor


natte omstandigheden)

Onder natte omstandigheden moet de stekker op het eindtakcircuit worden


aangesloten met een contactdoos dat geschikt is voor contact met water.
Installaties in de VS moeten gebruik maken van een Rain tight while in use
specification grade cover, zoals die worden gemaakt door Leviton.
Hierdoor kan de apparatuur worden aangesloten met een waterdicht deksel
dat over de stekker / contactdoos wordt geplaatst. Zo voldoet de installatie
aan NEMA 3R zoals vereist in de "national electrical code", sectie 410-57.
Het betreffende deksel is vermeld in UL reg #E13397.
Installaties in Europa moeten gebruik maken van een stopcontact dat
minimaal IP56 bescherming biedt aan de stekker / kabelassemblage. Men
dient ervoor te zorgen dat de door de contactdoos geboden mate van
bescherming geschikt is voor de omgeving.

1.10

Periodiek onderhoud
Schakel de apparatuur altijd uit en haal deze uit de contactdoos
voordat u periodiek onderhoud pleegt om het risico van een
elektrische schok of beschadiging van de apparatuur te voorkomen.
Zorg dat de apparatuur zich altijd op een vlak en effen oppervlak bevindt om
te voorkomen dat deze op de grond valt.

Model E1005 Indicator User Instructions

15

Safety and Warnings

1.11

Het apparaat reinigen


De buitenkant van de standaardproducten kan met een schone vochtige
doek met een klein beetje mild reinigingsmiddel worden afgenomen.
Gebruik geen schuurmiddelen, oplosmiddelen, schuurpoeder of
alkalische reinigingsmiddelen (zoals soda) om het apparaat te
reinigen. Gebruik deze middelen in elk geval nooit op de
displayvensters. De binnenkant van het apparaat mag nooit worden
gereinigd.
Spuit geen vloeistof direct op het displayvenster. Als u een
schoonmaakmiddel in een spuitbus gebruikt, spuit het middel op de
doek en niet op het displayvenster.

1.12

Training
Tracht geen procedure op de machine uit te voeren, behalve wanneer u de
relevante training gehad, of de instructieboeken gelezen heeft.
Om RSI (Repetitive Strain Injury = letsel door repetitieve werkzaamheden)
te voorkomen, is het belangrijk ervoor te zorgen dat het apparaat op een
werkoppervlak wordt geplaatst dat ergonomisch voldoet voor de gebruiker.
In gevallen van lang gebruik adviseren wij regelmatig een pauze in te
lassen.

1.13

Scherpe voorwerpen
Gebruik geen scherpe voorwerpen (schroevendraaiers, lange
vingernagels etc.) om de toetsen te bedienen.

1.14

EMC-naleving
De volgende waarschuwing kan op uw apparaat van toepassing zijn.
WAARSCHUWING: Dit is een klasse A-product. In huiselijke omgeving
kan dit product radiostoring veroorzaken in welk geval de gebruiker de
noodzakelijke maatregelen dient te nemen.

16

Model E1005 Indicator User Instructions

1.15 Installation et Service (French)

1.15

Installation et Service (French)


CET EQUIPEMENT NE CONTIENT AUCUN COMPOSANT PERMETTANT
L'INTERVENTION TECHNIQUE DE L'UTILISATEUR.
L'installation et la maintenance de cet quipement doivent tre excutes
uniquement par du personnel spcifiquement form et habilit intervenir.

1.16

Installation lectrique
Le cble d'alimentation doit tre branch sur une prise quipe d'un
dispositif de masse - mise la terre. L'alimentation lectrique au niveau de
la prise doit assurer une protection d'une valeur approprie contre une
surcharge de courant.
Par mesure de scurit, tous les quipements fonctionnant sous 110 ou 230
V utiliss en extrieur ou dans des conditions d'humidit doivent tre
raccords une source avec un coupe-circuit et protg par un appareil
homologu de protection contre les dfauts de masse (normes RCD, GFCI,
etc.)
EN CAS DE DOUTE, DEMANDER L'AVIS D'UN TECHNICIEN QUALIFIE.

Equipement avec prise secteur

Ces quipements lectriques doivent tre installs prs d'une prise


facilement accessible.

Equipement connect en
permanence - Consignes
d'isolement

Les quipements qui doivent tre aliments en permanence doivent tre


dots d'un systme de dconnexion facilement accessible et intgr au
cblage (un circuit isolant ou coupe-circuit avec une sparation de contact
d'au moins 3 mm).
Le circuit isolant ne doit pas tre install dans le cble flexible fourni avec
l'appareil.

Installation lectrique Information supplmentaire


pour conditions d'humidit

Dans des environnements humides, la fiche d'alimentation doit tre


connecte une prise ou botier spcialement conu et protg contre les
jets d'eau.
Les installations effectues aux tats-Unis doivent utiliser un couvercle
impermable, tel que ceux fabriqus par Leviton. Ce couvercle, class
NEMA 3R, permet de protger l'appareil et la prise conformment au code
national lectrique, section 410-57. Remarque : le couvercle mentionn cidessus est rpertori dans le fichier UL #E13397.
Pour les installations en Europe il est ncessaire d'utiliser au minimum
une connexion aux normes IP56 pour la prise, la fiche secteur et le cble.
Veiller ce que la protection fournie par le point de connexion soit adapte
l'environnement.

1.17

Maintenance de routine
Avant toute maintenance de routine, pour viter toute possibilit de
choc lectrique ou de dommage l'appareil, teindre toujours la
machine et la dconnecter de l'alimentation secteur.
Pour viter tout risque ventuel de chute de la machine, vrifier qu'elle soit
place sur une surface plane et niveau.

Model E1005 Indicator User Instructions

17

Safety and Warnings

1.18

Nettoyage de l'appareil
L'extrieur des produits standard peut tre nettoy avec un chiffon propre,
humidifi avec de l'eau et une petite quantit de dtergent.
Ne pas utiliser de dtergents abrasifs ou de solutions alcalines telle
que la soude mnagre, surtout sur les crans. Ne jamais essayer de
nettoyer l'intrieur de la machine.
Ne jamais vaporiser de liquide directement sur l'cran. En cas
d'utilisation d'un liquide de nettoyant mnager, asperger le chiffon et
non l'cran.

1.19

Formation
Ne pas essayer d'utiliser la machine ou suivre une procdure quelconque
sans avoir reu la formation approprie ou avoir lu les manuels
d'utilisation.
Pour viter le risque de microtraumatismes rpts, il est important de
s'assurer que la machine soit place sur une surface ergonomiquement
satisfaisante pour l'utilisateur. Il est recommand que l'utilisateur fasse
des pauses frquentes en cas d'une utilisation prolonge.

1.20

Objets tranchants
Ne pas utiliser d'objets tranchants (tournevis, ongles trop longs) pour
actionner les touches.

1.21

EMC
L'avertissement suivant pourrait concerner votre appareil.
Attention. Ceci est un appareil de classe A. Dans un environnement
domestique, cet appareil peut tre la cause d'interfrences radio. Dans ce
cas, l'utilisateur peut tre oblig de prendre les mesures adquates.

18

Model E1005 Indicator User Instructions

1.22 Installation und Wartung (German)

1.22

Installation und Wartung (German)


DAS GERT ENTHLT KEINE TEILE, DIE VOM BENUTZER GEWARTET
WERDEN KNNEN.
Die Installation und Wartung des Gertes darf nur von geschultem und
autorisiertem Personal durchgefhrt werden.

1.23

Elektrische Installation
Die verwendete Steckdose muss unbedingt mit einem Schutzleiter
ausgestattet sein. Die Stromversorgung ber die Steckdose muss durch
einen vorschriftsmigen Fehlerstrom-Schutzschalter geschtzt sein.

Zu Ihrer eigenen Sicherheit sollten


alle Netzstrom betriebenen (110V
oder 230V) Gerte, die im Freien
oder unter feuchten Bedingungen
verwendet werden, von einer
gesicherten Stromquelle, die ber
eine zugelassene FehlerstromSchutzeinrichtung (FI-Schalter)
geschtzt ist, versorgt werden.
IN ZWEIFELSFLLEN WENDEN
SIE SICH BITTE AN EINEN
QUALIFIZIERTEN
ELEKTRIKER.Angeschlossene
Gerte
Elektrische Installation (Zusatz-Hinweise bei nassem
Umfeld)

Ans Netz angeschlossene Gerte mssen leicht vom Netz trennbar sein.
Stndig angeschlossene Gerte - TrennschalterAnforderungenDauerhaft angeschlossene Gerte mssen ber einen
leicht zugnglichen, in den festen Stromkreis integrierten Notschalter
verfgen. (Einen Trenn-schalter oder Sicherungsautomaten mit mindestens
3mm Kontakt-Trennung).
Der Trennschalter DARF NICHT in das mitgelieferte flexible Netzkabel
installiert werden.

Im Freien oder unter feuchten Bedingungen muss der Stecker direkt ber
eine geeignete abwaschbare Steckdose an die Stromversorgung
angeschlossen werden.
Anschlsse in den USA sollten mit einer Abdeckung mit dem
Spezifikationsgrad Wasserundurchlssig whrend in Betrieb versehen
sein, wie beispielsweise die von Leviton.
Dadurch kann das Gert an eine Steckdose mit einer wasserdichten
Abdeckung laut NEMA 3R angesteckt werden, wie dies nach dem National
Electrical Code Abschnitt 410-57 erforderlich ist. Hinweis: die genannte
Abdeckung ist unter UL E13397 aufgelistet.
Anschlsse in Europa mssen ber eine Steckdose erfolgen, die die
Mindest-Anforderungen der IP56-Schutzvorkehrung fr Stecker / Kabelanschlsse erfllt. Es muss darauf geachtet werden, dass der von der
Steckdose gewhrleistete Schutz den Anforderungen des entsprechenden
Umfelds entspricht.

1.24

Regelmige Wartung
Zum Schutz vor Stromschlgen und um Beschdigungen des Gertes
vorzubeugen, muss dieses vor der Durchfhrung von Wartungsarbeiten ausgeschaltet und vom Netz getrennt werden.
Um das Herunterfallen des Gertes zu vermeiden, sollte dieses sicher auf
einer ebenen und waagerechten Oberflche platziert werden.

Model E1005 Indicator User Instructions

19

Safety and Warnings

1.25

Reinigung des Gertes


Das Gehuse kann von auen mit einem feuchten Tuch gereinigt werden.
Dem Wasser kann eine kleine Menge eines milden Reinigungsmittels
beigefgt werden.
Zu Reinigungszwecken, besonders im Fall von Monitoren, drfen
keine Scheuermittel, Lsungsmittel oder andere alkalische
Reinigungsmittel verwendet werden. Die Innenseite des Gertes darf
unter keinen Umstnden nass gereinigt werden.
Sprhen Sie keine Flssigkeiten direkt auf das Display. Wenn Sie
eine Reinigungsflssigkeit verwenden, sprhen Sie diese auf ein
Tuch und nicht auf das Display.

1.26

Schulung
Ohne Durchlesen der Bedienungsanleitung bzw. einer Einweisung durch
das Fachpersonal sollte das Gert nicht bedient oder ein Verbundsystem in
Betrieb genommen werden.
Um Fehlbelastungen zu vermeiden, sollte die Waage so platziert werden,
dass sie den ergonomischen Ansprchen des Bedieners gerecht wird. Es
wird empfohlen, bei lang andauernder Bedienung des Gertes regelmige Pausen einzulegen.

1.27

Spitze Gegenstnde
Verwenden Sie keine spitzen Gegenstnde (Schraubenzieher, lange
Fingerngel usw.), um die Tasten zu bedienen.

1.28

EMV-gerecht
Der folgende Warnhinweis trifft unter Umstnden auf Ihr Gert zu.
WARNUNG: Bei diesem Gert handelt es sich um ein Gert der Klasse A.
Dieses Gert kann bei einem Betrieb in Wohngegenden Funkstrungen
verursachen. Ist dies der Fall muss der Benutzer eventuell entsprechende
Manahmen treffen.

20

Model E1005 Indicator User Instructions

1.29 Installazione e manutenzione (Italian)

1.29

Installazione e manutenzione (Italian)


L'APPARECCHIATURA NON CONTIENE COMPONENTI RIPARABILI
DALL'UTENTE.
L'installazione e la manutenzione dell'apparecchiatura devono essere
eseguite esclusivamente da personale addestrato e autorizzato.

1.30

Impianto elettrico
Il cavo di alimentazione deve essere collegato ad una presa di corrente con
un contatto di messa a terra protettiva. La fornitura elettrica alla presa di
corrente deve prevedere una protezione contro le sovracorrenti di un indice
appropriato.
Per sicurezza l'apparecchiatura di rete (110 V o 230 V) utilizzata all'aperto
oppure in condizioni di bagnato o umidit deve essere alimentata da una
sorgente opportunamente equipaggiata con fusibili e protetta da un
dispositivo approvato di protezione contro i guasti di terra, come ad esempio
un dispositivo di corrente residua (RCI), un interruttore del circuito di
rilevamento di fughe a terra (GFCI), ecc.
IN CASO DI DUBBI RIVOLGERSI AD UN ELETTRICISTA QUALIFICATO.

Apparecchiatura collegabile

L'apparecchiatura collegabile deve essere installata in prossimit di una


presa elettrica facilmente accessibile.

Apparecchiatura cablata
permanentemente - requisiti
dell'isolatore

L'apparecchiatura collegata in modo permanente deve disporre di un


dispositivo di disconnessione facilmente accessibile, incorporato nel
cablaggio fisso un isolatore o un interruttore salvacircuito con una
separazione dei contatti di almeno 3 mm.
L'isolatore NON DEVE essere installato nel cavo di rete flessibile in
dotazione con l'unit.

Impianto elettrico - (note


aggiuntive in caso di condizioni
di bagnato)

In condizioni di bagnato, la spina deve essere collegata al circuito terminale


di derivazione tramite un'apposita presa concepita per applicazioni di
lavaggio.
Per gli impianti negli Stati Uniti occorre utilizzare un "coperchio conforme
alla specifica e un dispositivo di tenuta a prova di pioggia", come quelli
prodotti da Leviton.
Questi consentono il collegamento dell'unit con un coperchio a prova di
pioggia montato su presa che consente di soddisfare la classificazione
NEMA 3R, secondo quanto richiesto dal codice nazionale per il materiale
elettrico nell'articolo 410-57. Si noti che il coperchio menzionato elencato
nel file UL #E13397.
Per gli impianti in Europa occorre utilizzare una presa che fornisca almeno
una protezione IP56 per il complesso spina/cavo. Assicurarsi che il grado di
protezione fornito dalla presa sia adatto per l'ambiente.

1.31

Manutenzione di routine
Per evitare la possibilit di scosse elettriche o danni alla macchina,
spegnere sempre la macchina e isolarla dall'alimentazione elettrica
prima di eseguire gli interventi di manutenzione di routine.
Per evitare il rischio di caduta della macchina, laddove applicabile,
assicurarsi che sia posizionata in modo sicuro su una superficie piana e
orizzontale.

Model E1005 Indicator User Instructions

21

Safety and Warnings

1.32

Pulizia della macchina


La parte esterna dei prodotti standard pu essere pulita con un panno
pulito, inumidito con acqua contenente una piccola quantit di detergente
neutro.
Non utilizzare sostanze abrasive corrosive, solventi, detergenti
sgrassanti e soluzioni detergenti a base alcalina, come ad esempio
carbonato di sodio, particolarmente sui display. Non tentare mai di
pulire l'interno della macchina.
Non spruzzare alcun liquido direttamente sui display. Se si utilizza un
apposito liquido detergente assicurarsi di spruzzare il liquido su un
panno e non sul display.

1.33

Formazione
Non tentare di azionare la macchina o non eseguire alcun intervento su di
questa a meno che non si riceva una formazione appropriata o senza aver
letto il manuale di istruzioni.
Per evitare lesioni derivanti da sollecitazioni ripetitive (RSI, Repetitive
Strain Injury), importante assicurarsi che la macchina venga posta su
una superficie che sia posizionata in modo soddisfacente da un punto di
vista ergonomico per l'utente. Si consiglia di effettuare frequenti
interruzioni durante l'uso prolungato.

1.34

Oggetti appuntiti
Non utilizzare oggetti appuntiti (cacciaviti, unghie lunghe, ecc.) per
azionare le chiavi.

1.35

Conformit alla direttiva sulla compatibilit elettromagnetica (EMC)


Per la macchina possibile che sia applicabile la seguente avvertenza.
AVVERTENZA: questo un prodotto di classe A. In un ambiente
domestico, questo prodotto pu causare interferenze radio e in tal caso
possibile che l'utente debba intraprendere misure adeguate.

22

Model E1005 Indicator User Instructions

1.36 Instalacin y Mantenimiento (Spanish)

1.36

Instalacin y Mantenimiento (Spanish)


EL EQUIPO NO TIENE PARTES QUE PERMITEN UN MANTENIMIENTO
POR EL USUARIO.
La instalacin y el mantenimiento del equipo slo pueden ser realizados por
el personal calificado y autorizado.

1.37

Instalacin elctrica
El cable de electricidad debe ser conectado a un enchufe con toma de tierra.
La alimentacin elctrica en el enchufe debe estar provista de una
proteccin adecuada de sobrecorriente.

Para garantizar su seguridad, todo


el equipo conectado a la red (110V
o 230V) que se utiliza en el exterior
o bajo condiciones de humedad
debe ser alimentado por una fuente
con fusibles adecuados y protegida
por un dispositivo de proteccin
contra la prdida de tierra
aprobado (RCD, GFCI, etc.).
EN CASO DE DUDA, CONSULTE
UN TCNICO CALIFICADO.
Equipo conectado

Todo el equipo que se puede conectar a la red elctrica se debe instalar


cerca de una toma de corriente fcilmente accesible.

Equipo conectado
permanentemente Requisitos
de aislador

Si el equipo est permanentemente conectado, debe existir un mecanismo


de desconexin accesible incorporado en el cableado de la instalacin
elctrica. (Un aislador o interruptor del circuito con una separacin mnima
del contacto de 3mm).
NO SE DEBE instalar el aislador dentro del cable de red flexible facilitado
con la mquina.

Instalacin elctrica - (avisos


adicionales para condiciones de
humedad)

Bajo condiciones de humedad se debe conectar el enchufe directamente al


suministro de corriente por medio de una toma de corriente adecuada e
impermeable.
Instalaciones en los EEUU debern proveer de una cubierta con un grado
de especificacin Impermeable mientras en funcionamiento como aquellas
fabricadas por Leviton.
Esto permite que la unidad sea enchufada con una cubierta impermeable
colocada sobre el enchufe permitiendo cumplir con el grado NEMA 3R,
segn es requerido por el cdigo nacional elctrico bajo la seccin 410-57.
Aviso: La cubierta mencionada est listada en el archivo UL nmero
E13397.
Instalaciones en Europa deben usar una toma de corriente que provea un
mnimo de proteccin de IP56 a la conexin de enchufe / cable. Hay que
asegurarse de que el grado de proteccin realizado por el enchufe
corresponde a las exigencias del entorno.

Model E1005 Indicator User Instructions

23

Safety and Warnings

1.38

Mantenimiento rutinario
Para evitar la posibilidad de un choque elctrico o daos a la mquina,
siempre apague la mquina y desconctela del suministro elctrico
antes de realizar cualquier operacin rutinaria de mantenimiento.
Para evitar el riesgo de cada de la mquina, asegrese de colocarla
adecuadamente sobre una superficie lisa y nivelada.

1.39

Limpieza de mquina
El exterior de los productos estndares se puede limpiar con un pao
limpio humedecido con agua conteniendo una pequea cantidad de
detergente suave.
No se deben usar abrasivos duros, disolventes, estropajos o
soluciones de limpieza alcalinas, tales como soda custica,
especialmente en las pantallas de visualizacin. En ningn caso
debe intentar limpiar el interior de la mquina.
No roce ningn lquido directamente a las pantallas de visualizacin.
Si utiliza un fluido de limpieza especfico asegrese que lo roca al
pao y no a la pantalla.

1.40

Formacin
No intente operar o llevar a cabo ningn procedimiento en la mquina si no
ha recibido la capacitacin apropiada o ledo los Libros de Instruccin.
Para evitar el riesgo de LTR (lesin de la tensin repetida) es importante
asegurarse que la mquina sea colocada en una superficie
ergonmicamente satisfactoria para el usuario. En casos de un uso
prolongado se les recomienda descansos frecuentes a los operarios.

1.41

Objetos afilados
No utilice objetos afilados (destornilladores, uas largas, etc.) para
teclear.

1.42

Conformidad de ESD
La siguiente advertencia se puede aplicar a su mquina.
ADVERTENCIA: Esto es un producto de clase A. En un ambiente
domstico este producto puede causar interferencias radiofnicas, en
cuyo caso podra ser necesario que el usuario adopte medidas
adecuadas.

24

Model E1005 Indicator User Instructions

Introduction

The applications are available one


at a time. They must be switched
on through a password protected
menu. Contact your local supplier
or Avery Weigh-Tronix distributor
for assistance.

2
The Model E1005 is an easy to use, uncomplicated indicator for general
weighing applications. It is ideal for bench scales, floor scales and tank
weighing applications. The display includes a multi-segment fan graph for
fast visual awareness for checkweighing. Also, the indicator can perform
counting functions, peak weight functions, act as a remote display and
operate on battery power.
Communication port allows connection to a printer, remote display or
computer. The indicator also has three setpoint controls and can
accommodate a footswitch for zero, print or tare function.
All this in an IP54 rated enclosure.

2.1

About This Manual


Major sections of this manual are numbered and headed by titles in a black
bar like Introduction above. Subheadings appear in the left column.
Instructions and text appear on the right side of the page. Occasionally
notes, tips, and special instructions appear in the left column.

Model E1005 Indicator User Instructions

25

Introduction

2.2

Unpacking and Setup


Unpack your indicator and check for any shipping damage. If shipping
damage is found, save all packing materials and contact the shipping
company immediately.
1.

Use the included hardware and attach the indicator to the bracket.

2.

Connect all necessary cables to the appropriate connector on the


back of the indicator. See Figure 2.1.

The Model E1005 is battery


powered with an AC wall mount
transformer for charging.
Charge the battery overnight
before using the indicator on
battery power only.

Figure 2.1 Connection panel on the E1005


(Caution: connections may be reversed based on
viewpoint)

The unit can be run on the


transformer if the battery is drained
or absent.

3.

Plug the power transformer into an appropriate wall outlet. See note
at left.

4.

Press and hold the ON/OFF key for one second to turn the indicator
on.

5.

The indicator powers up in normal operation mode.

See Battery Information for more


information.

It is important that the battery be kept fully charged, either by the preferred
method of float charging (i.e. keeping the unit permanently plugged into AC
power supply) or by charging for 24 hours every 40-50 days. If a longer
period of storage is required and float charging is not practical, the battery
should be disconnected.
If the indicator is sealed for Weights and Measures purposes, disconnecting
the battery may not be possible. In this case the only option is to ensure that
the indicator is connected to the AC supply to keep the battery topped off.
The battery should never be stored in a discharged state as permanent
damage will occur.

26

Model E1005 Indicator User Instructions

Front Panel

3
The front panel, shown in Figure 3.1, consists of the keys and display

net

Figure 3.1 E1005 front panel

3.1

Keys
There are seven keys on the front panel. Their functions are listed below.

Never press a key with anything


but your finger. Damage to the
overlay may result if sharp or rough
objects are used.

Model E1005 Indicator User Instructions

27

Front Panel

Press the TARE key to perform a tare function. Also acts as a left
arrow key when in the menu structure.

Press the SELECT key to toggle between Gross, Tare, Net, Count,
Gross Accumulator, Net Accumulator, Transaction Counter, Piece
Weight, and Peak. Dependent on the current application. Also acts
as an up arrow key when in the menu structure.

Press the ZERO key to zero the display. Also acts as an Escape
key when in the menu structure.

Press the PRINT key to send information to a peripheral device


through the Comm port. Also acts as a down arrow key when in
the menu structure.

Press the UNITS key to scroll through the available units of


measure while in normal operating mode. Also acts as a right
arrow key when in the menu structure.

Press the F1 key to select application specific choices. Also acts


as an ENTER key in the menu structure.

Press and hold the ON/OFF switch for one second to turn the
indicator on. Press and hold the key until the unit turns off. The unit
must be in normal weigh mode to turn off.

When the indicator goes to sleep


you must press the ON/OFF switch
to restart the indicator.

28

Model E1005 Indicator User Instructions

3.2 Numeric Entry Procedure

3.2

Numeric Entry Procedure


Some keys have alternate functions when you need to enter numbers. See
Figure 3.2.

Figure 3.2 Alternate key functions


Press the ZERO key to terminate a
value entry and leave the previous
value, if any, active.

In screens where numeric entry is possible, choose the first digit using the
UP or DOWN keys. Use the LEFT and RIGHT keys to advance or
backspace through the entry. Press the F1 key to accept an entry. Below is
an example:
Example: To key in the number 507Press the SELECT or PRINT key until 5 appears on the display.
Press the UNITS key once to move cursor one space to the right.
Press the SELECT or PRINT key until 0 appears on the display.
Press the UNITS key once to move cursor one space to the right.
Press the SELECT or PRINT key until 7 appears on the display.
Press the F1 key to enter the value.
You can move the entry function one digit to the left with a press of the TARE
key. This effectively deletes the current value in that position and allows you
to enter a new value in that position.

Model E1005 Indicator User Instructions

29

Battery Information

Battery Information

4
This unit contains a sealed rechargeable 6 volt, 3.0Ah, lead-acid battery. Life
expectancy of this battery is 3-5 years in standby use or:

When the indicator goes to sleep


you must press the ON/OFF switch
to restart the indicator.

180 charging cycles (approx.) if discharged 100%

400 charging cycles (approx.) if discharged 50%

1200 charging cycles (approx.) if discharged 30%

Battery life is 25 hours with one 350 ohm weight sensor and 15 hours with
four 350 ohm weight sensors. Recharge time from complete discharge is 14
hours while powered up and in service (single loadcell). The AC adapter/
charger will charge the battery as it powers the indicator.
The charger has a 12 VDC 800 mA output with center positive connection.
It is important that the battery be kept fully charged, either by the preferred
method of float charging (i.e. keeping the unit permanently plugged into AC
power supply) or by charging for 24 hours every 40-50 days. If a longer
period of storage is required and float charging is not practical, the battery
should be disconnected.
If the indicator is sealed for Weights and Measures purposes, disconnecting
the battery may not be possible. In this case the only option is to ensure that
the indicator is connected to the AC supply to keep the battery topped off.
The battery should never be stored in a discharged state as permanent
damage will occur.

30

Model E1005 Indicator User Instructions

Menu Mode

5
The E1005 has two menus you can use to set up and test the indicator: the
User menu and the Supervisor menu. Both are explained below.

5.1

User Menu
The E1005 has a User menu which you can use to do the following:
Audit the number of configurations and calibrations performed
See software and revision information
Display test
Button test
Serial port test
1.

Access the User menu by pressing and holding the ZERO key for 3-5
seconds.
PASS_ is displayed.

User menu password is 111. You


must key in the password within 10
seconds or the display returns to
normal operation mode.

2.

Use the numeric entry procedure to scroll in the User menu password,
which is 111. See note at left. Figure 5.1 shows a flowchart of the
User menu items. Use the keys shown in the dotted box in Figure 5.1
to navigate through the menu and choose the items you want.

Choose the first digit using the UP


or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry.

Figure 5.1 User menu flowchart


Specific instructions on the User menu appear in the section Indicator
Diagnostics later in the manual.

Model E1005 Indicator User Instructions

31

Menu Mode

5.2

Supervisor Menu

Password for the Supervisor menu


is 1793.

The Supervisor menu is shown in Figure 5.2. Use this menu to set time and
date, print and clear reports, perform diagnostic tests and view audit
counters.
WARNING: Entering this menu and changing settings may affect
operation of the indicator and may require a service call to correct. Be
sure you want to change settings before doing so.

Figure 5.2 Supervisor menu

32

Model E1005 Indicator User Instructions

5.2 Supervisor Menu

Password for the Supervisor menu


is 1793.

1.

Access the Supervisor menu by pressing and holding the ZERO key
for 3-5 seconds
PASS_ is displayed.

DATE submenu
(Set date)

2.

Scroll in the password, 1793, by using the numeric entry procedure


and press F1
DATE is displayed. Use this to set the current date.

3.

Press the PRINT key


TYPEX is displayed. X is one of the four values shown below.
These are date styles along with number you enter to create that
style:
1=MM/DD/YY
2=MM/DD/YYYY
3=DD/MM/YY
4=DD/MM/YYYY

4.

Use the TARE or UNITS key to scroll to the style number you want to
use in all dated reports. When it is displayed, press the F1 key
M is displayed. This stands for month.

5.

Key in the month number (1 for Jan., 2 for Feb., etc.) using the
numeric entry procedure and press the F1 key
DD XX is displayed. DD stands for day and XX represents the
current value.

6.

Key in the date value and press the F1 key


YY XX is displayed. YY stands for year and XX represents the
current value.

7.

Key in the year (04=2004, etc.) using the numeric entry procedure
and press the F1 key
DATE is displayed.

HOUR submenu
(Set time)

1.

Press the UNITS key


HOUR is displayed. Set the time in this item.

2.

Press the PRINT key


TYPEX is displayed. X is one of the two values shown below:
1 is 24 hr. military format
2 is 12 hour, AM/PM format

3.

Use the TARE or UNITS key to scroll to the number for the style you
want to use in all time stamped reports and press the F1 key
HH XX is displayed. This stands for hour and its current value. See
note at left.

Time must be entered in 24 hour


(military) time.
1 = 1AM
13 = 1 PM

4.

M XX is displayed. M stands for minute and XX represents the


current value.
5.

Model E1005 Indicator User Instructions

Key in the hour in military time using the numeric entry procedure and
press the F1 key

Key in the minutes using the numeric entry procedure and press the
F1 key
33

Menu Mode

SS XX is displayed. SS stands for seconds and XX represents the


current value.
6.

Key in the seconds using the numeric entry procedure and press the
F1 key
HOUR is displayed.

SETUP submenu
(Setup menu)

1.

APP
(Applications)

2.

Press the UNITS key


SETUP is displayed. Use this submenu to print and/or clear
application reports and choose the operation modes or values for
applications which have choices. Each is explained in the following
steps.
Press the PRINT key
APP is displayed. Each application is listed below this menu item.
Applications are enabled and disabled in a password protected
menu, but you do each applications setup in this area of the
Supervisor menu. Contact your local supplier or Avery Weigh-Tronix
distributor for assistance.

PLU
(Product Look Up)
3. Press the PRINT key
PLU is displayed. This stands for Product Look Up. This memory
channel contains all the parameter values and accumulator totals
associated with all the different applications. This menu item prints
all the information for each application and/or clears all the
information.
4.

Press the PRINT key


PRINT is displayed. Use this item to print out a complete report of
all application parameters and totals.

5.

Press the PRINT key to print the PLU report.

6.

Press the UNITS key


CLEAR is displayed. Use this item to clear all the information
stored for each application. WARNING - Only do this if you are
sure you want the information permanently removed!
You may want to print out the reports before clearing all the
information.

7.

Press the PRINT key


SURE_ is displayed.

8.

Press the ZERO key to abort the Clear process or press the F1 key to
clear all the information
CLEAR is displayed.

9.

Press the SELECT key


PLU is displayed.

ACC
(Accumulator Application)
10. Press the UNITS key
ACC. is displayed. This stands for the accumulator application.
11. Press the PRINT key
34

Model E1005 Indicator User Instructions

5.2 Supervisor Menu

PRINT is displayed. Use this item to print out a complete report of


accumulator totals.
12. Press the PRINT key to print the accumulator report.
13. Press the UNITS key
CLEAR is displayed. Use this item to clear all the information
stored for this application. WARNING - Only do this if you are
sure you want the information permanently removed!
You may want to print out the report before clearing all the
information.
14. Press the PRINT key
SURE_ is displayed.
15. Press the ZERO key to abort the Clear process or press the F1 key to
clear all the information
CLEAR is displayed.
16. Press the SELECT key
ACC. is displayed.
BATCH
(Batching Application)
17. Press the UNITS key
BATCH is displayed.
18. Press the PRINT key
PRINT is displayed. Use this item to print out a complete report of
batch information.
19. Press the PRINT key to print the batch report.
20. Press the UNITS key
TYPE is displayed. Use this item to set the type of batching to be
done; Automatic or Manual
AUTOAs weight is added to the scale and the first output is
reached, OP1 annunciator lights and Output #1 is
activated. When weight reaches the second output value,
OP2 annunciator lights and Output #2 activates. When
weight reaches the third output value, OP3 annunciator
lights and Output #3 activates.
MAN.After the user begins the batching process, the user must
press the F1 key to activate each subsequent output after
each output weight is reached.
21. Press the PRINT key
The current setting is displayed.
22. Toggle between the choices by pressing the TARE or UNITS key.
Press the F1 key when your choice is displayed
TYPE is displayed.
23. Press the UNITS key
CLEAR is displayed. Use this item to clear all the information
stored for this application. WARNING - Only do this if you are
sure you want the information permanently removed!

Model E1005 Indicator User Instructions

35

Menu Mode

You may want to print out the report before clearing all the
information.
24. Press the PRINT key
SURE_ is displayed.
25. Press the ZERO key to abort the Clear process or press the F1 key to
clear all the information
CLEAR is displayed.
26. Press the SELECT key
BATCH is displayed.
TARGET
(Checkweighing application)
1.

Press the UNITS key


TARGET is displayed. Use this item to print and clear reports for
the checkweigher application and to set the type of sampling to be
used, Net or Sample.

2.

Press the PRINT key


PRINT is displayed. Use this item to print out a complete report of
checkweigher information.

3.

Press the PRINT key to print the checkweigher report.

4.

Press the UNITS key


TYPE is displayed. Use this to set the way you set the target weight
for the checkweighing application. You have two choices; LIMIT
and SPL (sample).
LIMITYou enter the upper and lower limits for your item and the
indicator will use those values to run the display.

Figure 5.3 Limit Mode

36

Model E1005 Indicator User Instructions

5.2 Supervisor Menu

SPLYou use a correct weight product on the scale to set the


target weight. The indicator will use this weight to run the
display. Upper and lower limits will automatically be 1
division above and below the target weight respectively.
Each colored display graduation is equal to 1 scale
division. The TARGET light stays lit if weight is 1 division
of the target weight.

Figure 5.4 Sample mode


5.

Toggle between the choices by pressing the TARE or UNITS key.


Press the F1 key when your choice is displayed
TYPE is displayed.

6.

Press the UNITS key


CLEAR is displayed. Use this item to clear all the information
stored for this application. WARNING - Only do this if you are
sure you want the information permanently removed!
You may want to print out the report before clearing all the
information.

7.

Press the PRINT key


SURE_ is displayed.

8.

Press the ZERO key to abort the Clear process or press the F1 key to
clear all the information
CLEAR is displayed.

9.

Press the SELECT key


TARGET is displayed.

Model E1005 Indicator User Instructions

37

Menu Mode

COUNT
(Counting Application)
1.

Press the UNITS key


COUNT is displayed. Use this item to clear and print reports for the
count application.

2.

Press the PRINT key


PRINT is displayed. Use this item to print out a complete report of
count application information.

3.

Press the PRINT key to print a count report.

4.

Press the UNITS key


CLEAR is displayed. Use this item to clear all the information
stored for this application. WARNING - Only do this if you are
sure you want the information permanently removed!
You may want to print out the report before clearing all the
information.

5.

Press the PRINT key


SURE_ is displayed.

6.

Press the ZERO key to abort the Clear process or press the F1 key to
clear all the information
CLEAR is displayed.

7.

Press the SELECT key


COUNT is displayed.

TOP
(Peak Weight Application)
1.

Press the UNITS key


TOP is displayed. Use this item to clear and print reports for the
peak application.

2.

Press the PRINT key


PRINT is displayed. Use this item to print out a complete report of
peak application information.

3.

Press the PRINT key to print a peak weight report.

4.

Press the UNITS key


CLEAR is displayed. Use this item to clear all the information
stored for this application. WARNING - Only do this if you are
sure you want the information permanently removed!
You may want to print out the report before clearing all the
information.

5.

Press the PRINT key


SURE_ is displayed.

6.

Press the ZERO key to abort the Clear process or press the F1 key to
clear all the information
CLEAR is displayed.

7.

Press the SELECT key


TOP is displayed.

38

Model E1005 Indicator User Instructions

5.2 Supervisor Menu

8.

Press the SELECT key


APP is displayed.

9.

Press the SELECT key


SETUP is displayed.

LOGS
(Calibration report)

1.

From App, press the UNITS key


LOGS is displayed. Use this menu item to print a calibration report.

2.

Press the PRINT key


CAL is displayed.

3.

Press the PRINT key to print the calibration report


CAL is displayed.

4.

Press the SELECT key twice


SETUP is displayed.

TEST submenu
(Test menu)

1.

ABOUT
(Indicator information)

2.

Proceed to the next menu by pressing the UNITS key


TEST is displayed. This menu lets you view indicator information
and test the display, keypad, serial port, inputs and outputs.
Press the PRINT key
ABOUT is displayed. Press the PRINT key then the UNITS key to
view the part number and revision level for the software found in
your indicator.
Press SELECT key to return to ABOUT.

ADC
(Analog to Digital converter)

3.

Press the UNITS key


ADC is displayed. This stands for the analog to digital converter
value in mV/Vs. This is a live reading of the system to aid in
diagnostics.

4.

Press the PRINT key


The mV/V value coming into the indicator is displayed.

5.

Press the SELECT


ADC is displayed.

DISP
(Display test)

BUTTON
(Key test)

6.

Press the UNITS key


DISP is displayed. This is the display test item.

7.

Press the PRINT key to perform a dynamic test of the display.

8.

Press the ZERO key to stop the dynamic test.

9.

Press the UNITS key


BUTTON is displayed. This is the button test item.

10. Press the PRINT key to perform a button test. Each key you press will
be reflected on the display screen to confirm the button is functioning
correctly.
11. Press the ZERO key to stop the button test.
BUTTON is displayed.

Model E1005 Indicator User Instructions

39

Menu Mode

SERIAL
(Serial port test)

12. Press the UNITS key


SERIAL is displayed. This is the serial test item. To test the serial
port, jumper the TX and RX lines. See illustration at left.
13. Press the PRINT key to access the serial test.
The display will show PASS if the serial port is working properly. If
there is a problem the display will show FAIL.
14. Press SELECT key to exit the serial test.
SERIAL is displayed.

INPUT
(Input test)

15. Press the UNITS key


INPUT is displayed. This is the input test item.
16. Press the PRINT key to access the test.
1 2 3 is displayed. 1 stands for input 1, etc.
17. Activate any remote switches connected to the indicator to verify
hardware and wiring
The display will show the input number change to 0 if the external
switch is operating properly.
18. Press the SELECT key to exit the test
INPUT is displayed.

OUTPUT
(Output test)

19. Press the UNITS key


OUTPUT is displayed. This is the output test item. See note at left.
20. Press the PRINT key to access the test.
OUT 1 is displayed. This stands for output 1.
21. Press the PRINT key
The display toggles between ON and OFF. This will toggle the
output off and on. Monitor the output to see that it is turning off and
on. Use a Trips Interface Unit (TIU3) or other output device.
22. Stop the test by pressing the SELECT key

Follow all lockout and red tag


procedures. Disconnect all
devices not intended to start
before running this test.

OUT 1 is displayed.
23. Press the UNITS key
OUT 2 is displayed.
24. Repeat steps 21 and 22 for outputs 2 and 3
25. Press the SELECT key
OUTPUT is displayed.
This completes the TEST menu item. Press the UNITS key to go to the
AUDIT submenu or press ZERO to return to normal weighing mode.

AUDIT
(Audit counters)

The next section of the Supervisor menu is the AUDIT submenu. This menu
lets you view configuration and calibration audit counters. These counters
cannot be changed, only viewed.
Follow these steps to access each item in the AUDIT submenu:

40

Model E1005 Indicator User Instructions

5.2 Supervisor Menu

CFG
(Configuration audit counter)

1.

Press the UNITS key


AUDIT is displayed.

2.

Press the PRINT key


CFG is displayed. This stands for the Configuration audit counter.
Use this item to see how many times this indicator has been
configured.

3.

Press the PRINT key


A number is briefly displayed, then CFG is displayed. This is the
number of times this indicator has been configured.

CAL
(Calibration audit counter)

4.

Press the UNITS key


CAL is displayed. This stands for the Calibration audit counter. Use
this item to see how many times this indicator has been calibrated.

5.

Press the PRINT key


A number is briefly displayed, then CAL is displayed. This is the
number of times this indicator has been calibrated.

6.

Press the SELECT key


AUDIT is displayed.

SLEEP (Sleep mode) menu

1.

Press the UNITS key


SLEEP is displayed. This stands for the sleep mode timer.

When the indicator goes to sleep


you must press the ON/OFF switch
to restart the indicator.
Any motion or any key press
restarts the sleep timer and will
prevent the indicator from entering
sleep mode.

2.

Press the PRINT key


Current sleep timer value is shown in hours. Scroll through the
choices (1/2, 2, 4, or OFF) by pressing the TARE or UNITS key.
Choose OFF to disable sleep mode. Choose any of the other
values to enable the sleep mode after the chosen time of keypad
inactivity and no scale motion.

3.

Press F1 when choice is displayed


SLEEP is displayed.

This completes the Sleep item and the Supervisor menu.


4. Press ZERO to return to normal weighing operation.

Model E1005 Indicator User Instructions

41

Indicator Operations

Indicator Operations

The applications are available one


at a time. They must be switched
on through a password protected
menu. Contact your local supplier
or Avery Weigh-Tronix distributor
for assistance.

The E1005 comes equipped with several weighing applications;


Accumulator weighing
Batch weighing
Checkweigher
Counting
Peak capture
Remote display
General (default setting)

When the indicator goes to sleep


you must press the ON/OFF switch
to restart the indicator.

The applications are activated using a password protected Service menu.


Contact your local supplier or Avery Weigh-Tronix distributor for assistance
on changing applications.
The general application comes as the default application. You can do gross
weighments, tare and net weighments. Below are instructions for each.

6.1

General Weighing
To perform gross weighing, power up the unit and follow these steps:
1. Empty the scale and press ZERO key to zero the display
0 is displayed and gross and center of zero annunciators are lit.

To change unit of measure, press


the UNITS key.

2.

Place item to be weighed on the scale


Weight is displayed.

To perform a net weighment, power up the unit and follow these steps:
1.

Empty the scale and press ZERO key to zero the display
0 is displayed and gross and center of zero annunciators are lit.

2.

Place item to be tared on the scale


Weight is displayed.

3.

Press the TARE key


0 is displayed and net annunciator is lit.

4.

Place material to be weighed on the scale


Net weight of material is displayed and Net annunciator is lit.

5.

42

Press the SELECT key to scroll through gross, tare, and net modes.
Remove the weight from the scale and press TARE to return to gross
mode.

Model E1005 Indicator User Instructions

Using Outputs

The output will trigger when weight goes above the target set for that output
and will remain on until the weight falls below the target. Follow these steps
to configure three outputs:
1.

With the indicator powered up, press and hold the SELECT key until
OP1 is displayed.

2.

Press the PRINT key to set the value for the output
Use the numeric entry procedure to key in a value, see note at left,
and press the F1 key to accept the value. OP1 will be displayed

Choose the first digit using the UP


or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry.

OR
Press the ZERO key to select the next output.
3.

Repeat step 2 for each output. Press ZERO key to select the next
output and then return to normal operation with the ingredients active.

Output operation in general


weighing mode:
Below Configured Value:
Outputs are ON
Annunciators are OFF
TIU3 Relays are ON
Above Configured Value:
Outputs are OFF
Annunciators are ON
TIU3 Relays are OFF

6.2

Accumulator Weighing
The accumulator is memory that collects individual weighments (gross and
net) and stores the totals. These totals can be recalled at any time and the
total number of weighments included in the totals can be displayed. With the
proper password all information can be printed and/or deleted. Contact your
local supplier or Avery Weigh-Tronix distributor for assistance.
The accumulator maximum is 999,999. It does not rollover and start over at
0.
To use the accumulator, power up the unit and follow these steps:
1.

Empty the scale and press ZERO key to zero the display
0 is displayed and Gross and Center of zero annunciators are lit.

You can use tare/net weighing with


the accumulator application. The
accumulator stores both gross and
net totals for later recall.

Model E1005 Indicator User Instructions

2.

Place item on the scale


Weight is displayed.

3.

You can press the PRINT or F1 key to add weight to the accumulator.
If you press PRINT, the weight is accumulated and the information
printed. If you press F1, the weight is accumulated. Scale weight must
return to within the configured Auto-print range before another
weighment can be accumulated.

4.

Repeat 2 and 3 for each weighment you want to accumulate.

43

Indicator Operations

5.

All weight must be removed from


the scale and the display must be
at zero before you can review
these values.

To review the accumulator total and the number of weighments,


remove all weight from the scale and press the SELECT key
repeatedly
1st press = Net weight displayed
2nd press = Tare weight displayed
3rd press = Gross total of all weighments is displayed
4th press = Net total of all weighments is displayed
5th press = Number of weighments is displayed
6th press = Display returns to gross weigh mode

See the Supervisor Menu on page 32 for instructions on clearing the


accumulator.

Using Outputs

You can use the outputs function if so desired while in the Accumulator
application. The output will trigger when weight goes above the target set for
that output and will remain on until the weight falls below the target. Follow
these steps to configure three outputs:

Output operation in
accumulation mode:
Below Configured Value:
Outputs are ON
Annunciators are OFF
TIU3 Relays are ON

1.

Above Configured Value:


Outputs are OFF
Annunciators are ON
TIU3 Relays are OFF

44

With the indicator powered up, press and hold the SELECT key until
OP1 is displayed.

2.

Press the PRINT key to set the value for the output
Use the numeric entry procedure to key in a value and press the F1
key to accept the value. OP1 will be displayed
OR
Press the ZERO key to select the next output.

3.

Repeat step 2 for each output. Press ZERO key to select the next
output and then return to normal operation with the ingredients active.

Model E1005 Indicator User Instructions

6.3 Checkweighing

6.3

Checkweighing
This section applies if your indicator has the checkweighing application
active. Applications are activated through a password protected menu.
Checkweighing allows a quick, visual check of the acceptability or
unacceptability of an items weight.
You can set your target weight in one of two ways. The type is set in a
password protected menu. The two types are explained below:
Limit Type

Enter the upper and lower limits for your item and the
indicator will use those values to run the display. See
Figure 6.1.

The graph is based off of net


weight so if a tare is active only the
net weight is considered for
checkweighing. If there is no tare,
gross weight is used as the basis
for the graph.

Figure 6.1 Limit type checkweighing


Sample Type

Place a correct weight product on the scale and press


the F1 key. The indicator will use this weight to run the
display. Upper and lower limits will automatically be 1
division above and below the target weight respectively.
Figure 6.2 shows how the graphic display works when
using a sample. Each graduation is equal to 1 scale
division. The TARGET light stays lit if weight is 1
division of the target weight.

Figure 6.2 Sample type checkweighing


Directions for using each type follows.

Model E1005 Indicator User Instructions

45

Indicator Operations

Limit Type: Entering Upper


and Lower Limits

Follow these steps to setup and use the checkweigher function if limit type
checkweighing is enabled, where you set upper and lower weight limits:

Choose the first digit using the UP


or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry.

1.

Press the F1 key


Hi is displayed followed by an underscore cursor.

2.

Key in the upper weight limit using the numeric entry procedure. See
note at left. Press the F1 key or press the ZERO key
Lo is displayed.

3.

Key in the lower weight limit using the numeric entry procedure. Press
the F1 key or press the ZERO key
The indicator returns to normal weigh mode.

4.

Sample Type: Using


Product to Set Target
Weight

Place items on the scale and the display will show if the weight is
over, under or acceptable based on the limits you have set.

Follow these steps to setup and use the checkweigher function if sample
type checkweighing is enabled, where you set target weight based on an
actual product:
1.

Place a sample, of the correct weight, on the scale


Weight is displayed.

2.

Press the F1 key.


The target weight is captured, the display reads NET 0 and your
indicator is ready to use as a checkweigher. The target weight will
be the same as your sample item and the target will stay lit
whenever an items weight is within 1 division of the target weight.

Performing a
Checkweighing Weighment

1.

With your target weight set, as described in one of the previous two
sections, place your item on the scale
If the weight equals the acceptable value, the TARGET
annunciator lights. If the weight varies from the target value, upper
or lower segments may be lit and the weight will show a plus or
minus weight reading for the deviation from the target weight.

2.

Repeat step 1 for all products of this weight.

Using Outputs

You can use the output function if so desired while in the checkweigher
application. Follow these steps to set up to three outputs:

Output operation in
checkweighing mode:
Outputs are OFF in gross zero
band
Annunciators are OFF
TIU3 Relays are OFF

1.

Outputs latch on for appropriate


ABOVE, OVER and ACCEPT
Annunciators are ON
TIU3 Relays are ON

46

With the indicator powered up, press and hold the SELECT key until
OP1 is displayed.

2.

Press the PRINT key to set the value for the output
Use the numeric entry procedure to key in a value and press the F1
key to accept the value. OP2 will be displayed
OR

3.

Press the ZERO key to proceed to the next output.

4.

Repeat step 2 and 3 for each output. Press ZERO key to select the
next output and then return to normal operation with the ingredients
active.

Model E1005 Indicator User Instructions

6.4 Counting

6.4

Counting
This section applies if your indicator has the counting application active.
Applications are activated through a password protected menu.
Follow these steps to perform a counting function with the E1005:

Choose the first digit using the UP


or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry
.

1.

In gross weight mode, press the F1 key


PCS is displayed followed by an underscore cursor.

2.

Enter the sample size you desire using the numeric entry procedure
and press F1
Add X is displayed. X is the sample size you keyed in.

3.

Place the correct number of parts on the scale and press the F1 key.
BUSY is briefly displayed, followed by one of two possible displays:
a. If the sample met the minimum sample requirements, the display
will show the correct number of parts on the scale.
b. If the sample size was not large enough, the display will show
ERROR briefly and will tell you how many more parts to add to the
scale (For example: 5 might be displayed.) Add the number
requested, wait for the scale to stabilize, then press F1 again. The
display will read the correct number of parts on the scale.

4.

Place the parts on the scale to be counted.

Using Outputs

You can use the output function if so desired while in the counting
application. Follow these steps to set up to three outputs:

Output operation in counting


mode:
Below Configured Value:
Outputs are ON
Annunciators are OFF
TIU3 Relays are ON

1.

With the indicator powered up, press and hold the SELECT key until
OP1 is displayed.

2.

Press the PRINT key to set the value for the output (go to step 3)
OR
Press the ZERO key to skip to the next output

Above Configured Value:


Outputs are OFF
Annunciators are ON
TIU3 Relays are OFF

OP2 is displayed. Skip to step 5.


3.

Key in a value using the numeric entry procedure and press the F1
key to accept the value
OP1 will be displayed.

Choose the first digit using the UP


or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry
.

Model E1005 Indicator User Instructions

4.

Press the ZERO key


OP2 is displayed.

5.

Repeat steps 2 through 4 for each output. Press ZERO key to return
to normal operation with the ingredients active.

47

Indicator Operations

6.5

Batch Weighing
This section applies if your indicator has the batching application active. The
batching application allows you to set three outputs based on three
progressively larger weights. There are two possible types of batching; Auto
or Manual. Each are explained below.
AUTO
As weight is added to the scale and the first output is reached, OP1
annunciator lights and Output #1 is activated. When weight reaches the
second output, OP2 annunciator lights and Output #2 activates. When
weight reaches the third output, OP3 annunciator lights and Output #3
activates.
MANUAL
In manual batching, after you begin the batching process, you must press
the F1 key to activate each subsequent output after each output weight is
reached.

Creating a Recipe
Choose the first digit using the UP
or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry
.

Follow these steps to create a recipe with three ingredients:


1.

With the indicator powered up, press and hold the SELECT key until
OP1 is displayed.

2.

Press the PRINT key to set the value for the output (go to step 3)
OR
Press the ZERO key to skip to the next output
OP2 is displayed. Skip to step 5.

If the system cannot accept the


value you key in, ABORT will be
displayed and display will return to
the menu.

3.

Key in a value using the numeric entry procedure and press the F1
key to accept the value
OP1 will be displayed.

4.

Press the ZERO key


OP2 is displayed.

5.

48

Repeat steps 2 through 4 for each output. Press ZERO key to return
to normal operation with the ingredients active.

Model E1005 Indicator User Instructions

6.6 Peak Weighing

6.6

Peak Weighing
This section applies if your indicator has the Peak application active.
Peak weight is defined as the highest stable weight reached by the scale.
Momentary higher weights that do not stabilize are ignored.
Follow these steps to perform a peak weighment:
1.

Empty the scale and press the ZERO key to zero the display
0 is displayed.

2.

Place the item(s) on the scale


Weight is displayed.

3.

After the motion annunciator (~) turns off remove the item(s) from the
scale,
The peak weight is displayed and the center-of-zero annunciator
lights.

4.

To clear the peak value, press the F1 key


0 is displayed.

5.

Repeat steps 2-4 for other weighments.

Using Outputs

You can use the outputs function if so desired while in the peak weighing
application. Follow these steps to set up to three outputs:

Choose the first digit using the UP


or DOWN keys. Use the LEFT and
RIGHT keys to advance or
backspace through the entry
.

1.

With the indicator powered up, press and hold the SELECT key until
OP1 is displayed.

2.

Press the PRINT key to set the value for the output (go to step 3)
OR
Press the ZERO key to skip to the next output
OP2 is displayed. Skip to step 5.

3.
Output operation in peak mode:
Below Configured Value:
Outputs are ON
Annunciators are OFF
TIU3 Relays are ON

OP1 will be displayed.


4.

Model E1005 Indicator User Instructions

Press the ZERO key


OP2 is displayed.

5.
Above Configured Value:
Outputs are OFF
Annunciators are ON
TIU3 Relays are OFF

Key in a value using the numeric entry procedure and press the F1
key to accept the value

Repeat steps 2 through 4 for each output. Press ZERO key to return
to normal operation with the ingredients active.

49

Indicator Operations

6.7

Remote Display
This indicator can be configured as a remote display or as the host to
another Evolution series indicator which is configured as a remote display.
As a user, there is nothing you do to set up these functions but you need to
understand the four modes in which this indicator can work as a remote
display.
The remote display is capable of displaying weight information; gross, tare,
or net weight, sent from the host indicator. In addition to displaying weight,
the remote display can be configured to display the scale annunciators;
Motion, Center of Zero, outputs, or other application specific data.
The remote display can also transmit key presses for basic weight functions
back to the host indicator. Functions include; Print, Zero and Tare. Your
indicator will be configured for these individual functions or a combination of
these functions. These four modes of operation are explained below.
Mode 1 - Remote weight display only:
In this mode the indicator will function as a remote display without
annunciators or key functions. The keypad and annunciators of the remote
indicator are disabled but valid messages received from the serial port are
displayed.
Mode 2 - Remote weight display with annunciators:
In this mode the indicator acts the same as in Mode 1 plus annunciator
information will be displayed.
Mode 3 - Remote weight display with keypad:
In this mode the indicator acts the same as in Mode 1 plus the remote
keypad is enabled.
Mode 4 - Remote weight display with keypad and annunciators:
In this mode the indicator acts the same as in Mode 1 plus annunciator
information will be displayed and the remote keypad is enabled.
If communication is lost between the host and the remote, the remote will
display -, middle dashes, until a signal is acquired again.

50

Model E1005 Indicator User Instructions

Communications

7
The E1005 provides an RS-232 output for data transmission to a peripheral
device. Contact your local supplier or Avery Weigh-Tronix distributor for
assistance with RS-232 interface connections.
If your indicator has a peripheral device connected, from the gross/net
weighing mode press the PRINT key to transmit the selected print formats
or strings.
The PRINT annunciator will illuminate while data is transmitted and the data
configured to be printed will be output to the printer.

The default serial port parameters


are 9600 baud, 8 databits, no parity
and 1 stop bit.

Print Format #1 for weighing applications


G 1234.56 lb<CR><LF>
T 34.56 lb<CR><LF>
N 1200.00 lb<CR><LF>
Print Format #1 for counting application
Count: 12230
Print Format #1 for peak application
12230 lb<CR><LF>

Model E1005 Indicator User Instructions

51

Error Messages

Error Messages

8
The following are displays you may see if problems occur or if invalid
operations are attempted with your indicator:

Display

Description
Over-range weight

Under-range weight

The unit cannot perform a function. Displayed only while


the key is held down.

Displayed while a key is pressed when attempting to


modify a sealed selection without edit privileges.

Appears on the display when the EEprom is not functioning


correctly.

Appears on the display when the Analog to Digital


converter is not functioning correctly.

Appears on the display when trying to change something


and the scale is sealed.

52

Model E1005 Indicator User Instructions

Indicator Diagnostics

The user menu lets you test various functions of the indicator. The user
menu is shown in Figure 9.1. Instructions for using the Test portion of the
menu are found below.

Figure 9.1 User Menu


1.

Access the User menu by pressing and holding the ZERO key for 3-5
seconds.
PASS_ is displayed.

2.

Use the numeric entry procedure to enter the User menu password,
111, and press F1.
TEST is displayed. This menu lets you view indicator information
and test the display, keypad, serial port, inputs and outputs.

3.

Press the PRINT key.


ABOUT is displayed. Press the PRINT key then the UNITS key to
view the part number and revision level for the software found in
your indicator.
Press SELECT key to return to ABOUT.

4.

Press the UNITS key


DISPLAY is displayed. This is the display test item.

5.

Press the PRINT key to perform a dynamic test of the display.

6.

Press ZERO key to stop the dynamic test.

7.

Press the UNITS key


BUTTON is displayed. This is the button test item.

Model E1005 Indicator User Instructions

53

Indicator Diagnostics

8.

Press the PRINT key to perform a button test. Each key you press will
be reflected on the display screen to confirm the button is functioning
correctly.

9.

Press ZERO key to stop the button test.


BUTTON is displayed.

10. Press the UNITS key


SERIAL is displayed. This is the serial test item. To test the serial
port, jumper the TX and RX lines. See illustration at left.
11. Press the PRINT key to access the serial test.
The display will show PASS if the serial port is working properly. If
there is a problem the display will show FAIL.
12. Press ZERO key to exit the serial test.
SERIAL is displayed.
13. Press the ZERO key to return to normal operation mode.

54

Model E1005 Indicator User Instructions

Specifications

10

Power:

120 VAC, 60 Hz, 800 mA


Internal battery- 25 hours of continuous operation with one weight sensor
- 15 hours of continuous operation with four weight sensors

Excitation:

5 volts supports up to four 350 ohm weight sensors

Analog Signal Input Range:

12 mV/V

Analog Signal Sensitivity:

0.5 V/division minimum

Operational Keys:

Tare, Select, Zero, Print, Units, F1, On/Off

Operational Annunciators:

Center of Zero, Motion, Gross, Tare, Net, Battery status, Under/Target/Over,


Unit of measure (LB, KG), Peak, Print, OP1, OP2, OP3

Display:

Seven digits, seven -segment, 0.75 inch high, LCD

Display Rate:

Selectable (1, 2, 5, 10 times per second)

A to D conversion rate:

60 times per second

Unit of Measure:

Three, independently programmable (pounds, kilograms, custom)

Capacity:

999,999 with decimal located from zero to five places.

Incremental Selections:

Multiples and submultiples of 1, 2, 5

Programmable Selections:

Zero range, motion detection, automatic zero tracking, five-point


linearization

Time and Date / RAM:

Battery backed up real time clock and RAM are standard

Internal Resolution:

64,424,509 counts per mV/V per sec.

Standard Inputs:

Three configurable logic level inputs for Zero, Print, Tare, Units, and F1

Standard Outputs:

Three cutoff outputs, open collector design


Serial port - RS-232 or RS-422 or RS-485 or 20 mA current loop

Serial Command Inputs:

Programmable serial response to ASCII character input, SMA protocol

Self Diagnostics:

Display, keys, inputs, outputs, serial port, A to D converter

Circuitry Protection:

RFI, EMI, and ESD protection

Model E1005 Indicator User Instructions

55

Specifications

Options:

Trips Interface Unit (TIU3)

Operating applications

General weighing, simple counting, checkweighing, accumulation, peak


measurement, batching, and remote display. One available at a time.

Operating Temperature:

14 to 104 F (-10 to 40 C)
10 to 90% non-condensing humidity

Enclosure:

Composite, IP 54

Dimensions

9 W x 6.25 H x 4.25 D (229mm x 159mm x 108mm) without bracket


9 W x 8.25 H x 4.25 D (229mm x 210mm x 108mm) with bracket

Weight:

8 lb (3.7 kg)

Agencies:

NTEP
Class III/IIIL, 10,000 divisions - CC# 04-029
CE Marked
OIML
6,000 divisions
Canadian Weights and Measures
UL
CUL
CSA

56

Model E1005 Indicator User Instructions

Avery Weigh-Tronix USA


1000 Armstrong Dr.
Fairmont MN 56031 USA
Tel: 507-238-4461
Fax: 507-238-4195
Email: industrial@weigh-tronix.com
www.wtxweb.com
Avery Weigh-Tronix UK
Foundry Lane,
Smethwick, West Midlands,
England B66 2LP
Tel: +44 (0)870 903 4343
Fax: +44 (0)121 224 8183
Email: info@awtxglobal.com
www.averyweigh-tronix.com