Signet 2839 To 2842 Conductivity Sensors: Warning
Signet 2839 To 2842 Conductivity Sensors: Warning
WARNING
SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Conrm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and
appropriate eye, face, hand, body, and/or respiratory protection should be used.
Table of Contents
1. Specications
2. Cell constant
selection
3. In-line Installation
4. Submersible or Tank
installation
5. Wiring
6. Maintenance
7. Ordering information
1. Specications
Dimensions
38.9 mm/
1.530 in.
38.9 mm/
1.53 in.
)
38.9 mm/
1.53 in.
144 mm/
5.66 in.
140.7 mm/
5.54 in.
109 mm/
4.29 in.
3-2840-1
3-2840-1D
0CR
905
890
886
0-X
885
0
590
0CR
580
285
Shipping Weight:
2839:
2840, 2841, 2842:
Process connection:
-1 and -2 versions:
-1D and -2D versions:
Temperature measurement:
Temperature response time ():
2839: 5 s
2841: 20 s
3-2841-2
3-2841-2D
3-2842-2
3-2842-2D
75
50
25
C
-20 0 20
60
100
140
-4
140
212
284
32 68
Operating temperature/pressure:
(with thread engagement per ANSI B1.20.1)
/4 in. NPT
ISO 7/1-R3/4
41.9 mm/
1.65 in.
BAR PSI
7 100
CR Electrodes
with DryLoc (-2X)
Quality Standard:
CE
26.8 mm/
1.13 in.
3-2841-1
3-2841-1D
3-2842-1
3-2842-1D
3-2840-2
3-2840-2D
General
CR Electrodes
with Cable (-1X)
35.8 mm/
1.41 in.
15.9 mm/
0.625 in.
21.5 mm/
0.85 in.
15.9 mm/
0.625 in.
15.9 mm/
0.625 in.
35.8 mm/
1.41 in.
15.9 mm/
0.625 in.
41.9 mm/
1.65 in.
73.7 mm/
2.90 in.
3-2839-2
3-2839-2D
Compatibility:
112 mm/
4.41 in.
105.9 mm
4.17 in.
3-2839-1
3-2839-1D
59.3 mm/
2.33 in.
15.9 mm/
0.625 in.
73.7 mm/
2.90 in.
15.9 mm/
0.625 in.
102.7 mm/
4.05 in.
2% of cell value
PT1000
2840:
2842:
10 s
30 s
-1X versions:
-10C to 100C @ 6.9 bar (14F to 212F @ 100 psi)
-10C to 131C @ 2.76 bar (14F to 268F @ 40 psi)
-2X versions:
-10C to 85C @ 6.9 bar (14F to 185F @ 100 psi)
Storage temperature:
Wetted materials:
DryLoc connector (-2 versions only):
CPVC
Threaded tting:
PEEK
Insulator:
PEEK
Insulator O-ring (2841, 2842):
FPM
Electrode contacts:
316L stainless steel
PEEK is a trademark of Victrex plc
Inspect threads to ensure integrity. Do not install an electrode with damaged threads.
Apply sealant or PTFE tape to threads.
Wetted materials include 316L stainless steel, PEEK and FPM (FPM O-ring inside 2841, 2842).
Check for chemical compatibility before installing electrode.
The -1X electrodes are supplied with 5 m (15 ft.) of cable. It may be extended to a maximum 30 m (100 ft.)
For resistivity measurements above 10 M or below 20C, maximum cable length is 25 ft. (7.6 m)
Wetted materials include 316L stainless steel, PEEK (FPM O-ring inside 2841, 2842).
Check for chemical compatibility before installing electrode.
The -1X electrodes are supplied with 15 ft. of cable. It may be extended to a maximum 30 m (100 ft.)
For resistivity measurements above 10 M or below 20C, maximum cable length is 25 ft. (7.6 m)
1.
2.
3.
4.
Product name
Do not route electrode cable in conduit containing AC power wiring. Electrical noise may interfere with electrode signal.
Routing electrode cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
Seal cable entry points to prevent moisture damage.
For resistivity measurements above 10 M or below 20C, maximum cable length is 25 ft. (7.6 m)
8850
Conductivity Transmitter
locking ring
conduit base
integral
adapter
electrode
ProcessPro
To 8850-1
ProPoint
10
Sensr Gnd
(SHIELD)
Iso. Gnd
(BLACK)
SILVER (SHLD)
Temp. IN
(WHITE)
WHITE
RED
Signal IN
(RED)
BLACK
14
Sensr Gnd
(SHIELD)
13
Iso. Gnd
(BLACK)
12
Temp. IN
(WHITE)
11
Signal IN
(RED)
Shld
Temp. IN
Iso. Gnd
Signal IN
To 5800CR
Intelek-Pro
To 9050CR
RLY3
ANL2
RLY4
NO
NO
R+
SERIAL
CH2
CH1
T1
SG
9 10 11 12 13
14
R-
Tx GND Rx
S2
T2
S1
To 8860
BLACK
WHITE
RED
SHLD
CH 2
ISO GND
TEMP 1
Ch 1
SHIELD (Gnd)
RED
WHITE
BLACK
CLEAR (SHLD)
SGNL 1
RED
WHITE
BLACK
WHITE
RED
BLACK
SHLD
CH 1
ISO GND
TEMP 1
CLEAR (SHLD)
Ch 2
SHIELD (Gnd)
SGNL 1
product name
6. Maintenance
Conductivity electrodes require little maintenance except for periodic cleaning in installations where contaminants are present.
Electrode tip
(side view)
Electrode tip
(end view)
9 mm
0.34 in.
Electrode
14 mm
0.55 in.
11 mm
0.44 in.
Cotton swab
C
Insulator
B
New Insulator
D
electrode tip
(side view)
7. Ordering Information
Mfr. Part No.
3-2839-1
3-2839-1D
3-2840-1
3-2840-1D
3-2841-1
3-2841-1D
3-2842-1
3-2842-1D
Code
159 000 921
159 000 923
159 000 786
159 000 788
159 000 790
159 000 792
159 000 794
159 000796
Description
Cell 0.01, 15 ft cable, NPT
Cell 0.01, 15 ft cable, ISO
Cell 0.1, 15 ft cable, NPT
Cell 0.1, 15 ft cable, ISO
Cell 1.0, 15 ft cable, NPT
Cell 1.0, 15 ft cable, ISO
Cell 10.0, 15 ft cable, NPT
Cell 10.0, 15 ft cable, ISO
3-2839-2
3-2839-2D
3-2840-2
3-2840-2D
3-2841-2
3-2841-2D
3-2842-2
3-2842-2D
3-2850-1
3-2850-2
3-2850-3
3-2850-4
George Fischer Signet Inc., 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. Tel. (626) 571-2770 Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com Or call (in the U.S.): (800) 854-4090