7044 203 High-Performance Thin-Layer Chromatography / Chemical Tests
First Supplement to USP 38NF 33
Add the following:
203 HIGH-PERFORMANCE THIN-LAYER CHROMATOGRAPHY
PROCEDURE FOR IDENTIFICATION OF ARTICLES OF BOTANICAL
ORIGIN
INTRODUCTION
This chapter describes a procedure for use in a USP Identification test that relies on the technique of high-performance thinlayer chromatography (HPTLC). It is applicable to the identification of articles of botanical origin in USP's compendia that serve
as a drug substance or drug product, or as an ingredient or a dietary supplement. Careful control of the variables for the
HPTLC technique is briefly described with references to more detailed information provided in equipment manuals. Reproducibility of the results allows comparison of closely related botanical materials that are not official ingredients. The analytical technique uses high-performance plates, appropriate equipment to control variables, and a system suitability test for purposes of
performance qualification. [NOTESee Identification of Articles of Botanical Origin Using High-Performance Thin-Layer Chromatography Procedure 1064 for additional information.]
REQUIRED EQUIPMENT
The equipment used for HPTLC technique consists of the following:
Plates: Unless otherwise specified in the individual monograph, use plates coated with a uniform 200-mm layer of porous
(60- pore size) silica gel having irregular particles of 210 mm and an average particle size of 5 mm, a polymeric binder,
and a fluorescence indicator (F254) of 20 10 cm. [NOTEChromatographic methods using high-performance thin-layer
chromatographic glass plates are preferred over aluminum-backed sheets because of greater mechanical stability.]
A device suitable for the application of specified volumes of samples as bands with specified length at the specified positions
A suitable chromatographic chamber (for example, a twin trough chamber) allowing for control of saturation and developing distance
A device suitable for controlling the activity of the stationary phase via relative humidity
A device suitable for reproducible drying of the developed plate
A device suitable for treatment of the plate with derivatization reagent, if required
A device suitable for heating as part of the derivatization procedure, if required
A system suitable for documentation of chromatograms under UV 254 nm, UV 366 nm, and white light
Each of these devices as well as the system as a whole should pass installation qualification (IQ), operational qualification
(OQ), and performance qualification (PQ) in order to assure that the instruments are working according to their specifications
to control the variables within their intended ranges. [NOTESee Analytical Instrument Qualification 1058 for additional information.] Installation qualification and operational qualification are typically performed by the instrument manufacturer. The
system suitability test is performed by the analyst as proof of performance qualification.
PROCEDURE
Preparation of Test Solution
Unless otherwise stated in an individual monograph, 100 mg of a powdered botanical ingredient, 10 mg of a dry extract or
fraction, or the amount of a dosage form containing the equivalent of the aforementioned quantities of the botanical ingredient is sonicated for 15 min with 1 mL of methanol. After centrifugation the filtrate or supernatant is used as the Sample solution. Unless otherwise stated in an individual monograph, 50 mL of an essential oil is dissolved in 1 mL of toluene and used as
the Sample solution.
Preparation of the Standard Solutions
Unless otherwise stated in an individual monograph, USP Reference Standards of individual marker compounds are dissolved
at a concentration of 1 mg/mL in methanol. The USP Reference Standard extracts are shaken and sonicated in methanol at a
concentration of 10 mg/mL or, for essential oils, the USP Reference Materials are dissolved in toluene at a concentration of 50
mL/mL.
Official from August 1, 2015
Copyright (c) 2015 The United States Pharmacopeial Convention. All rights reserved.
Accessed from [Link] by coe2010 on Fri Jun 19 [Link] EDT 2015
First Supplement to USP 38NF 33
Chemical Tests / 203 High-Performance Thin-Layer Chromatography 7045
Sample Application and Plate Layout
Samples are applied as narrow bands of 8.0 0.5 mm length at a distance of 8.0 0.5 mm from the lower edge of the
plate. The system suitability standards are applied on the lane nearest to the edge at NLT 20 mm from the edge of the plate.
The distance between tracks (center to center) is NLT 11.0 0.5 mm. All application volumes are specified in the individual
monograph. Application volumes usually range from 2.0 to 10.0 mL. The developing distance is marked with a pencil close to
one of the edges of the plate before the development, although an electronic solvent front detection device may be substituted.
Preconditioning of the Plate
Following sample application and unless otherwise stated in an individual monograph, the plate is conditioned at a relative
humidity of 33% for a minimum of 10 min (for example, by standing in a closed chamber containing a saturated solution of
magnesium chloride).
Preparation of the Developing Chamber and Development of the Plate
Where a twin trough chamber is used, the rear trough is fitted with filter paper. The chamber is charged with a sufficient
volume of developing solvent to wet the filter paper completely and achieve a level of developing solvent of exactly 5 mm in
both troughs. With the lid closed, the chamber is left 20 min for saturation. The plate is introduced in a vertical position into
the front trough of the chamber so that the coating layer faces the filter paper. When the mobile phase has reached a distance
corresponding to a development path of 6 cm, the plate is removed from the chamber and dried in a vertical position in a
stream of cold air that does not affect the integrity of the separated zones. Other chamber configurations and developing distances may be specified in an individual monograph. [NOTEOther development chambers may be employed if the results
obtained fulfill all of the system suitability criteria.]
Derivatization Procedure
Where derivatization reagents are used, defined volumes of reagents in solution (typically 12 mL) are homogeneously
sprayed onto the plate or the plate is immersed into the reagent solution at a defined speed and for a defined dwell time.
[NOTEImmersion speed of 50 mm/s and dwell time of 1 s works for most nonaqueous reagents.]
Visualization
Chromatograms on the plate are visualized as stated in an individual monograph. Observation and evaluation may be performed under UV 254 nm, UV 366 nm, or white light prior to and after derivatization.
System Suitability
To check the suitability of the system for resolution, position, and color of the bands, unless otherwise stated in an individual
monograph, two or more reference substances are selected that have similar but just separable RF values under the chromatographic conditions to be used; for example, chlorogenic acid (blue) and hyperoside (yellow-orange) in chromatographic systems used for flavonoids. USP Reference Standard mixtures for system suitability may be provided, or the substances designated to check the system suitability for resolution, position, and colors of the bands may be included in the USP Reference
Standard extracts. Description of the resolution, position, and colors for the key bands of the reference material fingerprint
should match the description in the monograph within a specified tolerance range. The system suitability requirements in an
individual monograph are satisfied when the results obtained comply with those specified in the monograph.
Evaluation and Acceptance Criteria
Chromatograms of the Sample solution and Standard solution are compared against the description in the Acceptance criteria
section of the monograph with respect to zone position, zone separation, color, and relative intensity.
Documentation
Documentation is necessary to record the results in an auditable manner to comply with current good manufacturing practices. Proper documentation tools should be employed; for example, a camera suitable for taking digital pictures under UV and
white light and an imaging software suitable for archiving, retrieving, and analyzing the results makes it easy to maintain electronic records.
1S (USP38)
Official from August 1, 2015
Copyright (c) 2015 The United States Pharmacopeial Convention. All rights reserved.
Accessed from [Link] by coe2010 on Fri Jun 19 [Link] EDT 2015
7046 361 Barbiturate Assay / Chemical Tests
First Supplement to USP 38NF 33
OTHER TESTS AND ASSAYS
Delete the following:
361 BARBITURATE ASSAY
Internal Standard, Internal Standard Solution, Standard Preparation, and Assay PreparationPrepare as directed in
the individual monograph.
Chromatographic SystemUnder typical conditions, the gas chromatograph is equipped with a flame-ionization detector
and contains a 4-mm 0.9-m glass column packed with 3% liquid phase G10 on support 80- to 100-mesh S1A. The column is
maintained at a temperature of 200 10, and the injection port and detector are maintained at about 225, the column temperature being varied within the designated tolerance, as necessary, to meet System Suitability specifications and provide suitable retention times. Use a suitable carrier gas, such as dry nitrogen, at an appropriate flow rate, such as 60 to 80 mL per minute. Use on-column injection. [NOTEIf the instrument is not equipped for on-column injection, use an injection port lined
with glass that has been washed successively with chromic acid cleansing solution, water, methanol, chloroform, a 1 in 10
solution of trimethylchlorosilane in chloroform, and chloroform.]
System Suitability (see Chromatography 621)Chromatograph five replicate injections of the Standard Preparation, and
record peak responses as directed for Procedure: the relative standard deviation for the ratio RS is not more than 1.5%. In a
suitable chromatogram, the resolution, R, between the barbituric acid and the Internal Standard is not less than the value given
in the individual monograph, and the tailing factor, T, for each of the two peaks is not more than 2.0.
ProcedureInject a suitable portion (about 5 mL) of the Standard Preparation into a suitable gas chromatograph, and record the chromatogram. Similarly inject a suitable portion of the Assay Preparation, and record the chromatogram. Calculate
the content of the barbiturate or barbituric acid in the assay specimen by the formula given in the individual monograph, in
which RU is the ratio of the peak response of the barbituric acid to that of the Internal Standard obtained for the Assay Preparation; QS is the ratio of the weight of the barbituric acid to that of the Internal Standard in the Standard Preparation; Ci is the
concentration, in mg per mL, of Internal Standard in the Internal Standard Solution; and RS is the ratio of the peak response of
the barbituric acid to that of the Internal Standard in the Standard Preparation. 1S (USP38)
561 ARTICLES OF BOTANICAL ORIGIN
SAMPLING
In order to reduce the effect of sampling bias in qualitative and quantitative results, it is necessary to ensure that the composition of the sample used be representative of the batch of drugs being examined. The following sampling procedures are the
minimum considered applicable to vegetable drugs. Some articles, or some tests, may require more rigorous procedures involving more containers being sampled or more samples per container.
Gross Sample
Where external examination of containers, markings, and labels indicates that the batch can be considered to be homogeneous, take individual samples from the number of randomly selected containers indicated below. Where the batch cannot be
considered to be homogeneous, divide it into sub-batches that are as homogeneous as possible, then sample each one as a
homogeneous batch. It is recommended to include samples from the first, middle, and last containers where the No. of Containers in Batch (N) is 11 or more and each container in the batch is numbered or lettered in order.
No. of Containers
in Batch (N)
No. of Containers
to Be Sampled (n)
110
All
1119
11
>19
n = 10 + (N/10)
(Round calculated n to next highest whole number.)
Official from August 1, 2015