Operator Manual BT4500-00ING PDF
Operator Manual BT4500-00ING PDF
P/N: MO6313-00ING
This product conforms to the safety requirements of the Council Directives 98/79/EEC of 27
October 1998 (European Parliament) regarding the In-Vitro Diagnostic Medical Devices. This
directive is in accordance with the Article 2, Paragraph 2 of the Directive 89/336/EEC, which
ceases to apply to the products complying with the present directive. Refer to Paragraph 7,
Article No.1 of the IEC Official Gazette No. L331 of Dec. 1998.
It also conforms to Italian Regulations CEI EN 61010-01 and CEI EN 61326-1 (EMC).
The conformity is attested when the equipment is installed in accordance with the conditions
outlined in the manual
CHAPTER 1
GENERAL INFORMATION & SYSTEM DESCRIPTION Page
1. MANUFACTURER........................................................................................ 2
2. ANALYZER IDENTIFICATION AND INTENDED USE..................................2
3. STORAGE AND HANDLING........................................................................ 3
4. WARNINGS AND PRECAUTIONS............................................................... 3
5. ANALYZER INTRODUCTION....................................................................... 5
6. BASIC OPERATING PRINCIPLES OF THE ANALYZER.............................6
7. SYMBOLS: EXPLANATION OF THE USED OR APPLIED SYMBOLS......7
8. BRIEF DESCRIPTION OF THE SYSTEM..................................................10
8.1. Front view of the analyzer........................................................................................ 10
8.2. Modules.......................................................................................................................11
IMPORTANT NOTICE
The introduction of access passwords has been
rendered mandatory since 2004 for safeguarding
sensitive data.
1. MANUFACTURER
The BT4500 analyzer is produced by Biotecnica Instruments S.p.A., in its factory located in Rome,
Italy.
Biotecnica was founded in 1972, based on a lot of technical expertise and manufacturing know-how
in the design, production, sales and technical assistance of standard and OEM clinical chemistry
equipment and relevant reagent products destined for private and state clinical laboratories
(hospitals, clinics, day hospitals, and universities etc.). Since that time the company has developed
and produced the instruments of uncompromising performance characteristics, which has enabled
Biotecnica to become one of Europe’s leading manufacturers of clinical chemistry instruments.
Biotecnica is headquartered in Rome (Italy) and markets products both domestically and
internationally through a network of distributors.
Our actual range of instruments is composed basically of:
- Automatic analyzers for clinical chemistry.
- Flame photometers
- I.S.E. analyzers
Biotecnica has over 3000 square meters of manufacturing facility, which houses the latest in
engineering, manufacturing, testing and quality control equipment and is staffed by highly
experienced personnel.
The company supplies more than just products – it provides highly efficient technical and
manufacturing support based on the requirements of its customers supported by ongoing training
programs to keep you current on advances in technology. With a global network of distributors and
sales engineers, you can be assured of the most comprehensive customer support wherever you
are in the world.
For more information about Biotecnica, visit the Biotecnica web site at w w w. b i o t e c n i c a . i t
Please do not hesitate to contact us and we will be only too pleased to assist with any inquiries you
may have.
manipulated (before, during and after the use) by qualified personnel familiar with their
characteristics in order to safeguard the user as well as the quality of the reagent itself.
Operating Principles
Generally, adding a sample to its reagent determines a chemical reaction (involving enzymes and/or
substrates) whose effect is to increase (or decrease) the solution color and thus the optical density
in the cuvette. As the reaction proceeds, it is "read" by the analyzer in terms of "absorbance" ("A" or
"Abs" for absorbance).
As every analyte has its own reagent with its proper characteristics, it becomes necessary to use
different methodologies (preparation and reading) based upon different wavelengths for each test.
Many tests are based on similar principles, hence they will have in common the method and the
wavelength, but not necessarily the incubation and reading times.
To obtain the concentration of an analyte in a sample, the analyzer multiplies the absorbance (or the
absorbance delta ∆A = absorbance variation) developed by that sample reaction with a
multiplication factor.
Besides some analyses for which a theoretical factor is used, usually the factor is calculated by
means of a calibration. During the calibration the analyzer reads the reaction obtained with a known
concentration sample called "standard". The factor is calculated by dividing the known concentration
value by the absorbance read for the standard.
For the non-linear analyses (e.g. immunoturbidimetric tests) it is necessary to create an interpolation
curve by means of several standards at different concentrations.
12. be sure that the power switch on the side panel of system is off when plugging in, or removing
the power cord from a wall outlet.
13. turn off the mains power switch on the side panel whenever the system is not in use. this
prevents damages due to surge in the mains power.
14. do not attempt to alter the shape of any part of the system.
15. if the system is not operating properly and the trouble-shooting section does not provide a
satisfactory solution to the problem, then do not use the system until the defects are remedied.
16. inspect all accessories and system cords. Do not use if damage can be seen such as cut
insulation or outer covering, frayed or broken wires, corroded or broken connectors etc.
17. to reduce the risk of fire or electric shock, do not allow fluids or any foreign object to enter the
system. Wipe off spills immediately.
18. do not use benzene, thinner, any kind of solvents, or abrasive detergents to clean the case.
Clean with soft dusting cloth dampened with distilled water. If necessary use only neutral
detergent.
19. do not stick objects of any kind into the system through back panel or case slots as they may
touch dangerous voltage points or short out parts that could result in fire or electric shock.
20. install the system in such a way that adequate ventilation is provided all around to properly
dissipate the heat.
21. use only original Biotecnica’s replacements. Do not use conventional parts/replacements. This
will cause malfunction of the system.
22. make sure all fluid lines are free of kinks, nicks, sharp bends, punctures, or occlusions before
installing on system.
23. the halogen lamp must be replaced some minutes after the instrument has been turned off and
power cord unplugged.
24. always allow the burnt out lamp to cool down before handling or attempting replacement.
25. never touch the lamp or the reflector with bare fingers. use a rag when changing.
26. if the lamp is touched inadvertently during installation, clean the lamp or reflector with alcohol
and dry with a clean, soft cloth before burning. Contamination of the lamp or reflector may
reduce lamp performance.
27. this lamp (when lit) emits UV (ultraviolet) radiation. prolonged exposure to this lamp may cause
skin and eye irritation.
28. the analyzer system must not be dismantled or repaired by anyone who has not been qualified
by the manufacturer. Incorrect work may cause fire or irreparable damage to the system.
29. do not overload accessories power outlets and extension cords as this can result in fire or
electric shock.
30. use only secure power source to protect the analyzer system against power surges.
31. do not oil any part of the system.
32. empty waste containers whenever they are full. Ensure that the container lids are screwed on
tightly to prevent leakage or dispersion into the environment.
33. the safe disposal of the analyzer waste material with minimal environmental impact is the
responsibility of the user and will have to meet the local laws and dispositions.
34. do not attempt to remove any panels or coverings of the analyzer system while the system is in
operation.
35. after operation/servicing, cover the system with a protective plastic or cloth sheet.
36. do not use software disks of unknown origin in the analyzer computer as they may introduce
viruses.
37. do not use the computer of the analyzer for any other purpose than the one for which it is
designed for.
38. be particularly cautious that no parts of your body (e.g. fingers hair, etc.) or loose objects (e.g.
cables, tubing, etc.) can be trapped by any moving or rotating parts (e.g. sampling arm, plates,
washer module, pump rollers etc.) of the analyzer system.
NOTE:
a) the careful observation of the proceeding warnings should result in a long and satisfactory
performance. If the above-mentioned notices are not fully observed, then any form of warranty is no
longer valid and Biotecnica Instruments S.p.A. will not be responsible for any subsequent damage
or loss. (see warranty conditions).
b) the information in this manual is based upon the hardware and software currently in use.
Biotecnica Instruments S.p.A. reserves the right to make software and hardware changes or
improvements for product enhancement without notice and without imposing any obligation upon
itself to install these changes or improvements on its products previously manufactured.
5. ANALYZER INTRODUCTION
This instrument is an automatic analyzer for Clinical Chemistry, I.S.E. and Immunoturbidimetry
manufactured by Biotecnica Instruments S.p.A. Rome, Italy.
The analyzer software is based on Windows 7 operative system. The external PC allows easy
management of the operative system and of the stand alone operative program.
The software is easy to learn and offers the operator the maximum flexibility.
MAIN FEATURES
- throughput of 450 tests/h and up to 690test/h including I.S.E.
- sequential access
- routine, urgent samples (STAT – Single Test in Actual Time) and batch programming
- calibrations and quality controls, immediate or time-programmable
- 18 normality classes: 3 fixed (male, female and children) and 15 editable
- dedicated auto-diagnostic function to help the operator in the maintenance of the analyzer
- operative 24h/24
- equipped with the I.S.E. (Ion Selective Electrodes) module
- analytical parameters are flexible and open
- can work with up to four reagents, ready to use or concentrated, dilution can be performed with
water or dedicated diluent
- every reagent can have its duplicate on board, for backup purposes
- refrigerated reagents chamber (40+40 positions) and standard tray (26 positions) ensures a longer
stability of the products in use.
- positive barcode identification of reagents and samples position reduces the possible errors
- repetitions (Re-run) upon operator's request or automatically (pathological and hyperactive results)
- sample pre- and post-treatment
- connection to the Host Computer
- internal software for managing quality control (statistics of control sera and population) and
patients’ archive with data display and printouts
- 100 positions sample tray
- auxiliary solutions in dedicated containers (three, for sample dilution or pre-teatment)
- reading tray of 84 circular 6mm special glass cuvettes
- one single cycle allows two dispensations per arm (two of the first reagent, two sample
dispensations, two dispensations of the second reagent), a reading of all the cuvettes and the
related washings
- washing is done in stages over several cuvettes simultaneously, each of which is subject to a
particular stage of washing
- sampling arms are equipped with a new mechanic for the detection of a crash and new devices for
quick movement of the needle for efficient mixing of reagents and solutions
- the arms have clot sensors and vacuum sensors which can monitor the operations of each cycle of
the machine indicating hydraulic malfunctions
- the control PC is connected to the analyzer via USB port. This allows to use external computers as
laptop or desktop with the specific settings of each country
The methods programming is fully open. The operator can program the mostly used methodologies,
which are already in memory, or brand new methodologies. The most common methods are: End
Point, Fixed Time, Kinetic, Sample Blank. It is possible to store up to 500 different test codes, plus
Relation Tests with no limit. In the stored analyses list the operator can generate customized test
codes sequence for the reagent tray in use, including the relation tests.
The BT 4500 is a compact instrument, based on three levels: in the upper level there are the
devices accessible by the user, protected by a lid that provides state of cleanliness and safety. At
the central level there is a front drawer that allows easier access and maintenance of dilutes and
and tanks and a rear drawer to access the tray of the unified waste pumps. At the lowest level there
are two doors that permit easy access to distilled water tanks, detergent container and waste
containers. A set of supplied fittings allows the use of containers of various sizes.
The I.S.E. (Ion Selective Electrodes) module is a device dedicated to the determination in the
samples of the electrolytes (see chapter L). This device is defined as ion selective as the used
sensors react with the corresponding ions in accordance with the following Nernst law:
E = E0 + RT/nF log aM+
+ +
aM = M ion activity
E = potential in Volt
E0 = constant (H+ electrode redox semi-reaction std potential)
R = gas constant
F = Faraday’s constant
T = temperature expressed in Kelvin degrees
n = ion charge
The sensors life is dependent upon the number of sample runs and the routine maintenance
procedures outlined in this manual.
5
4
6
ICONS BAR
GENERAL ICONS
It is used to delete
As far as the analyzer program runs under Windows O.S., almost all Windows short-cuts are
available. For instance, in the editable fields it is possible to right click to copy and paste the text, or
it is possible to use CTRL+C to copy and CTRL+V to paste.
Catalog number
Manufacturer
Storage temperature
Expiry date
Biological hazard
Risk symbols
* Not all symbols are listed here. These are only some examples.
Direct current
Alternating current
Equipotentiality
ON (Main supply)
The BT 4500 is a compact instrument, based on three levels: in the upper level there are the
devices accessible by the user, protected by a lid that provides state of cleanliness and safety. At
the central level there is a front drawer that allows easier access and maintenance of dilutes and
and tanks and a rear drawer to access the tray of the unified waste pumps. At the lowest level there
are two doors that permit easy access to distilled water tanks, detergent container and waste
containers. A set of supplied fittings allows the use of containers of various sizes.
8.2. Modules
Modules definition
• READING STATION MODULE: comprises cuvette plate, photometer, reading unit,
washing station and electronics.
• POWER SUPPLY MODULE: houses the main power supply of the analyzer.
• REAGENT TRAY MODULE: is composed of the rotating reagent's tray, the refrigeration
chamber, the bar-code reader and the electronics.
• SAMPLE TRAY MODULE: is composed of the rotating samples tray, the bar-code
reader, the sample tube sensors and the control electronics.
• ISE MODULE: consists of the sensors panel, hydraulic path and the electronics.
• SAMPLING ARM: is composed of a two-axes based mechanical system
accommodating sampling needle head with built-in electronics including correct position
sensor (Encoder).
• REAGENT 1 ARM: is composed of a two-axes based mechanical system
accommodating sampling needle head with built-in electronics including correct position
sensor (Encoder).
• REAGENT 2 ARM: is composed of a two-axes based mechanical system
accommodating sampling needle head with built-in electronics including correct position
sensor (Encoder).
CHAPTER 2
INSTALLATION Page
1. UNPACKING INSTRUCTIONS
The crates can be easily opened by applying the lever action, with a large screwdriver, to remove all
the spring clips at the base of the crate as shown in the figure below. Carefully remove the upper
covering. Remove the analyzer and place it on a stable vibration-free surface. Carefully unpack all
the accessories and place them in a protected place. Store the empty wooden crate in a safe place
for future use.
Arrow Pointing
Upwards
Base
Spring Clip
2. INSTALLATION
The analyzer should be easily accessible to allow the operator to load samples, consumables,
reagents, etc. Avoid exposure to direct light, heat, air streams and draught. The instrument’s left,
right and rear sides must be left free (min. 20 cm from the wall) to ensure the produced heat
dispersion and easy tubes and cables connection. Room temperature must not exceed 32°C. It is
very important to place the analyzer away from strong electromagnetic fields, such as centrifuges,
electric motors, big refrigerators, X-ray instruments, etc.
The analyzer refrigerator produces water condensation in the reagent chamber. This is important for
cooling the reagents in the bottles.
NOTE:
All the components, when present, shown in the following figures may undergo
modifications over the time. Therefore, it is recommended to verify them accurately prior to
any repair or installation (refer to eventual specific manuals included).
Before turning the analyzer on, it is necessary to put in seat some items that are shipped separately:
1. I.S.E. sensors and pumps
2. diluent bottles
I.S.E. SENSORS
I.S.E.
slide
After these operations, run an I.S.E. prime to fill the I.S.E. hydraulic circuit.
DILUENT BOTTLES
The analyzer is equipped with three diluent bottles for diluting the samples. Two of these bottles are
located between the Samples arm and the R1 arm. The other one is located next to the I.S.E.
funnel.
The three diluent bottles, for convenience, have been defined in the software as follows:
Saline solution
Diluent
Sample treatment
The operator can fill these bottles with any necessary working solution and change the definitions as
necessary (see Setup, chapt. 9 par. 2.4)
Before turning the analyzer on, verify also that all necessary parts are in place (covers, trays, etc…)
The start-up process includes the loading of the operating system (bootstrap) into the memory. At
the end of the boot (loading of the operating system), the instrument activates all the devices and
performs mechanical, hydraulic and electronic checks.
Once turning on procedure has completed (lasting few minutes), wait for the system to warm up.
In the meanwhile, it will be necessary to perform a series of primes to make sure all the hydraulic
circuits are filled and there are no bubbles left. From the menu Analyzer, select Analyzer Utilities.
Alternatively, open the Special functions menu on the side bar and select Analyzer Utilities. The
following window will appear.
During warm-up phase the temperature indicator flashes on the bottom right of the display until the
appropriate temperature is reached. The instrument reaches the steady state after approximately 30
minutes.
DESCRIPTIVE TABLE
Note: Check that the waste container is empty or that there is sufficient capacity for at least containing
washing solution corresponding to the daily waste liquid volume.
CHAPTER 3
FUNCTIONS Page
1. BASIC THEORY........................................................................................... 2
1.1. Methods description ..................................................................................................4
2. DESCRIPTION OF THE PROGRAM MENU................................................ 6
2.1. Printing interface........................................................................................................9
3. APPLIED MATHEMATICAL FUNCTIONS ................................................ 11
3.1. Initial Computation...................................................................................................13
4. HOW TO PROGRAM THE ANALYSES .................................................... 14
4.1. Creating a new code ................................................................................................14
4.2. Relation tests............................................................................................................14
4.3. Introduction to the clinical chemistry tests ...........................................................16
5. ANALYSES PROGRAMMING ................................................................... 17
5.1. GENERAL ............................................................................................... 17
5.2. REACTION.................................................................................................................18
5.2.1. REAGENTS .........................................................................................................18
5.2.2. TIMES ..................................................................................................................20
5.2.3. REACTION...........................................................................................................20
5.2.4. BLANK.................................................................................................................23
5.3. SAMPLE ....................................................................................................................24
5.4. CHECK.......................................................................................................................28
5.5. SUPPLEMENTARY ...................................................................................................32
6. CALIBRATIONS......................................................................................... 34
7. CONTROLS................................................................................................ 38
8. PROFILES .................................................................................................. 39
9. APPENDIX ................................................................................................. 41
1. BASIC THEORY
The basic operating principles of the analyzer are described in chapter 1, par. 6.
The BT4500 operates with fixed cycles of 16 seconds. Each cycle allows two dispensations, reading
and washing of the cuvettes. The cuvettes are automatically washed with water and a special acid
solution to avoid possible cross contamination. In every cycle, three arms can operate
simultaneously.
Every test can be freely programmed, with the following limitations:
1. the first reagent is always inserted at cycle 1
2. the sample is always inserted at cycle 2
3. the minimum total cycles is of 7
4. the maximum total cycles is of 35
In the software, the programming of some of the basic test methods is already available (end point,
fixed time, kinetic and sample blank).
As far as every method can be freely programmed (with the above limitations), it is very easy to
create all possible tests using the already available formulas.
For instance the end point test formula is:
Point[Final]-Blank, i.e. the analyzer will use the last cycle absorbance subtracted of the blank
absorbance. But the same formula can be written as Point[30]-Blank, if the last cycle is 30 or if the
last cycle is 34, but the method requires the reading to be performed before the last cycle. Again,
the same formula can be written as Point[Final-4]-Blank.
The same versatility applies to all formulas.
For what concerns the reagent blank, it can be read on the same cuvette used for the reaction. In
this case every sample will have its own blank. On the other hand, it is possible to perform the
reagent blank on a separated cuvette. In this case, the reagent blank will be calculated on the basis
of a specific formula, that usually is Point[Serum-1], i.e. the analyzer will use as blank the cycle
before the sample insertion.
For all those tests that start with the sample insertion (R1, then R2, then Sample), the blank can be
performed on the same cuvette as the test. For the tests that start with the second reagent (R1, then
Sample, then R2), it will be necessary to perform the blank in a second cuvette.
The analyzer works with couples of cuvettes, one in the internal ring and one in the external ring. At
the beginning of every working session, the analyzer will wash the first series of cuvettes, then it will
go on washing until the last sampling. If in the run there are only I.S.E. tests, the analyzer will
anyway start a the first washing series. The cuvettes wash is performed in a sequence that starts
with the aspiration of the cuvettes content and the dispensing of acid solution, followed by the
rinsing with water and drying of the cuvettes.
Once the cuvettes are washed, the sampling starts, using the cuvettes in couples. After the
reagents and sample dispensing, the analyzer will wait the end of the incubation and reading times,
will acquire the absorbances for the results calculation and will then wash again the cuvettes so that
the cycle can continue. The analyzer can work 24h/24, therefore the samples can be loaded
continuously.
The test results are then displayed in the real time results pages. In the Results RT page, the single
results are displayed as soon as ready, while in the Results page, the results are displayed as soon
as the whole patient is completed. Then results can be printed or saved into the Archive.
All results, archives pages as well as reaction graphs can be also saved as *.pdf files. Archives can
also be exported in *.csv or *.rtf formats.
The calibrations, as well as the Quality Controls, have dedicated archives. The QC has also a real
time archive where the controls results can be verified in real time.
In the analyses programming page, there are the analytical parameters, the standards and the
controls. In the same page it is possible to display only the tests on line or all the tests.
The position of the reagents, as well as the positions of the standards or controls on the tray are
programmed in different dedicated windows.
Samples can be programmed in Routine mode (which includes running of STATS or controls) or in
batch mode.
In Routine it is possible to select the type of sample (serum, urine, CSF or other) and eventually the
automatic pre-dilution. Batch runs can be performed on serum or on urine.
Before every run, it is possible to perform a reagents volume check (the analyzer will present the
question) to allow the analyzer to use the backup reagents.
Insert R2
Incubation time
Fixed Time
In this type of reaction, there is an increase (or decrease) of the absorbance during both
incubation’s and reading’s phases. However, the slope of the line may not be the same during the
two phases. The reaction graph displayed to the user is not always linear, but can also appear as
piecewise linear. During reading time, the absorbance delta (ΔA) is computed, which is used for
calculating the final concentration for the analyte in the sample.
Concentration is calculated by multiplying the absorbance delta by the factor obtained from the
calibration:
Conc. in Sample = Factor x (Abs delta)
Abs delta
Kinetics
This kind of reaction is very similar to the previous one, with the difference that the reaction and the
graph derive both from the computation of regression lines. The regression line for the reading
phase is then scaled to minutes to compute absorbance delta (ΔA/min.). This value is then
multiplied by the factor to compute the concentration of the analyte in the sample:
Abs delta
Reading time
Insert sample
Sample Blank
This method is used whenever it is required to eliminate the photometric interference of the sample
(for example turbid sera) from the reaction. These are double-reagent End Point reactions. The
reaction and the computation are performed during two distinct phases: in the first phase (sample
blank) the reaction between the first reagent and the sample (R1+S) takes place, while in the
second phase the second reagent is added to R1+S (R1+S+R2). The final absorbance used for
computing the concentration of the analyte is obtained from the difference in absorbance between
the two phases:
Conc. in sample = Factor x {Abs(R1+S+R2) – [Abs(R1+S) x k]} where k is the volumetric factor
There are five main menus: Patients, Analyses, Analyzer, Utility and Archives.
Patients
Routine/STAT Management & Batch Management: these commands
are used to enter samples in Routine/STAT and Batch mode.
Delete Work list: deletes the entire memorized patients list. The analyzer
will request confirmation before deleting.
Analyses
Parameters: with this command the analyzer will open the analyses page
where it is possible to enter and program new analyses or to modify the
existing ones.
Manage Reagents: this icon gives access to the reagents map, where it is
possible to create the on-line map and to have general information on reagents,
such as residual volume, position, lot number and expiry date.
Profiles: is used to create the analyses lists (profiles) which can be used for
programming patients.
Manage Standard: gives access to the standard page, where the standard
positions and lot number can be viewed. From this page it is possible to run
calibrations.
Manage Controls: gives access to the controls page, where the assigned
positions and lot number can be viewed. From this page it is possible to run the controls.
Export: copies in any desired location, the analytical parameters, optionally with the profiles and the
analyses tray. There are two available options: Back-up (for exporting all the analyses) and Single
Test (exports a selection of single tests).
Import: copies the above-mentioned parameters from any location into the analyzer memory. There
are two available options: Restore (will import all parameters) and Single test (imports the exported
single tests).
NOTE: when a back-up is imported, all data will be overwritten. When a single parameter is
imported, it will be placed in the list of available analyses, not in the "in tray" list. The
operator will have then to correctly place it in the analyses tray.
Analyzer
Mechanical Calibrations: gives access to the page dedicated to the
mechanical devices adjustment (arms and trays).
Analyzer Utilities: this command opens a page where the operator can
select to perform one or more of the available utilities, such as washing the
cuvettes or zeroing the photometer.
Diagnostic: this line gives access to further options as viewing the FCC or the optical transmission.
It also gives access to the general diagnostic page, which is normally used only by the well-qualified
technical assistance personnel.
Utility
Setup: it is used to define the system parameters, such as the language,
the communication, the bar-code, etc. This command is disabled during
analyzer operation.
View Reaction Graph: this command displays the reaction’s curves on a
graph, with all reaction information and print capability.
Log-Off: this command is used for logging off and on the different users.
When logging off, a window will appear where the user must enter the
user name and password
Switch-off analyzer + PC: with this command it is possible to turn off the
analyzer program together with the PC operative program. The analyzer
itself will remain turned on. Press the ON/OFF button on the side of the
analyzer to turn also the analyzer off off.
Note: The analyzer’s program will guide the operator through screen messages in the analyzer
turning off procedures. He will be invited to place the solutions when needed and to ensure the
correct execution of washing procedures. Bear in mind that it is not possible to ensure data
precision and accuracy if the normal washing and maintenance procedures are not observed.
Serial: this option allows the analyzer to send data to the host computer, upon operator’s request.
The type of data transportation can be selected in the Setup.
Log Files: this read-only area is very important for the Technical Assistance/service personnel.
Under this option there are the following:
- View Logs: gives access to the files where the operations performed by the analyzer are stored,
together with the logged users.
- View Test Counter: in this page it is possible to view the total of the tests performed by the
analyzer, divided into blanks, controls, standards, etc.
- Lot's management: shows all the reagents' lots with the expiry dates. The positions on the tray,
as well as if the reagent is the 1st or 2nd are not displayed.
Archives
This menu gives access to all the archives: patients archive, population
archive, quality control archive, calibrations' archive and real time QC
archive. The latter gives information in real time on the performed
controls.
In this bar are presented the most common and most useful icons.
Opens the routine programming page
Gives access to the reagent tray pages (volumes information, make tray function)
It is used to delete
Click on the printer icon to print the document. The printer interface will allow the operator to select
the printer, the copies, the number of pages, etc….
Click on the ""paper" icon to export the document in different type of file formats. The available file
formats are:
*.pdf => will save the document as Adobe document.
*.rtf => will save the document in rich text format. This can be opened
with Word, Notepad, Wordpad, etc…
*.csv => will save the document in comma separated values format.
This can be opened with Excel. The older versions of Excel will
correctly open (in columns ) these files only following this procedure.
Open Excel, the click on "Open" or "Open a file". In the "Open" window select "All files (*.*)" at the
line "File type". Then browse for the *.csv file and open it.
Text (matrix printer) *.prn => will save the document in a text file for dot-matrix printers.
where:
n : Number of readings
i : Number of reading (i)
T : Times
L : Readings
♦ Linear Regression
where:
M : Angular coefficient for the line
Q : Final point for the line
n : Number of readings
i : Number of reading (i)
T : Times
L : Readings
♦ Distance point-line
where:
M : Angular coefficient for the line
Q : Final point for the line
X : Point Abscissa
Y : Point Ordinate
where:
X : X axis
Y : Y axis
x0 : First Point X Axis
x1 : Second Point X Axis
y0 : First Point Y Axis
y1 : Second Point Y Axis
where:
K : Volumetric factor
vS : Serum volume
vR1 : First reagent volume
vR2 : Second reagent volume
♦ Clinical Chemistry
where:
Z : Zeroing with water
F1 : First Filter’s Value
F2 : Second Filter’s Value
Fz1 : First Filter’s Zero-Value
Fz2 : Second Filter’s Zero-Value
Op : Optical path
Of : F.C.C.
The analyzer can store virtually endless analysis codes (with parameters). It is possible to view only
the analyses which are on the tray (Only in tray), only the analyses which are not on the tray (Only
not in tray) or all programmed codes (All Codes). To create the working tray it is possible to use
the Make Tray command in the Reagents Management window (Tray icon).
Secondary: write here the secondary unit, if necessary. In the Factor field, write the
conversion factor from the primary unit to the secondary one.
Limits: insert the min. and max. values of the normal range with reference to the selected group
(male, female, etc).
Decimals: it is possible to choose the number of decimals after the point. Leaving the Automatic
option the analyzer will follow this principle of the floating point:
for values like 0.XXX three decimals
for values up to 9.XX two decimals
for values up to 99.X one decimal
for values over 100 no decimals
Otherwise it is possible to decide how many digits must be shown after the point, with a maximum of
6 digits.
Instrumental Factor: this function introduces a constant correction of the final data of the
calculated test. It may be used for making adjustments to test data obtained from analytical methods
or different type of instruments. Calculation: final result = value x instrumental factor
Shift: this function introduces a constant quantitative correction of the final data. It may be used for
making adjustments to test data obtained through analytical methods or instruments of different
types. Calculation: final result = value + shift.
To enter the mathematical function select Formula
A window divided in two parts will be displayed: one for the calculator and one for the analyses list.
The mathematical function can be composed of simple values and operations or can recall sample
results acquired by the analyzer on other tests (complex function). To enter a simple mathematical
function avail yourself of the displayed calculator. To enter a complex function, select the code of
the test to be inserted into the function. A small field will appear, where it is possible to select the
type of sample (among serum, urine, CSF and Other) result for that test. Then complete the function
with the needed operations.
To create more complex functions (involving more than one test’s result) it is advisable to use the
parenthesis as for all normal mathematical functions. Ex. For the creatinine clearance with urine/24h
= 900ml [(urine CRE x urine ml 24/h)/(serum CRE x 1440)] the formula would be: ( [CRE&U] * 900) /
([CRE&S] * 1440).
NOTE: use only the numbers and symbols on the screen, by clicking on each and remember that
the same operators on the keyboard are disabled.
There are some commands which are in common for all the analyses and these commands are
available in the menu bar.
Print menu. From this menu it is possible to print the analytical parameters, the calibrations and the
controls. It gives access to other choices: it is possible to print the actual test, the selected tests (if
more than one test is checked) or the whole area (only in tray, only not in tray or all codes).
5. ANALYSES PROGRAMMING
The first page shown is the Parameters page. In this page there are five buttons: General,
Reaction, Sample, Check and Supplementary. Each of these choices opens a different dedicated
page.
5.1. GENERAL
In this page the operator must give some general information on the test.
Putting the check in the bar-code checkbox, the bar-code field will be displayed. By means of the
Bar-Code it is possible to assign a numerical code for positive identification of the reagents. It
enables, if activated in the Setup, the bar-code scanning to correctly identify the bottle during
reagent’s insertion phase.
Method: write in this field the reaction principle used for the test (for example: Jaffè, IFCC, etc.).
This option is useful when recalling the tests from the Quality Control archive, in accordance with
different principles.
Filters: the operator can select the desired filter value for the Filter 1 and the Filter 2 (normally
used as reference filter) from the available filter wavelengths: 340, 380, 405, 436, 480, 510, 546,
578, 630 and.
For the tests in dichromatic, click on the Filter 1 and select the desired filter from the cascading
window and then go to Filter 2 and select the desired filter from its cascading window.
For the tests in monochromatic, select the desired filter in the Filter 1 cascading window. Since the
second filter is not used in this type of test, therefore select the last position, Ref, in the cascading
window of Filter 2. In this case the analyzer will use a reference filter only to stabilize the readings.
Process: defines the kind of test’s calibration: with factor, linear, or with curve. The following
choices are available: cubic spline, polynomial, best fit, multi points, log-logit 4, log-logit 5 and semi-
log.
linear: this function is used for linear reactions, it requires analytical test calibration to process the
computing factor;
with factor: it is used for linear reactions whenever the computing factor used is theoretical;
with curve: non-linear tests, distinguished in:
- cubic spline: the typical cubic interpolation is used. There is a good connection between
subsequent point couples. Approximation is perfect on single points, but is not flex point free;
- polynomial: a third degree polynomial function is used. It has a very good approximation, but
sometimes it may show anomalies;
- best fit: a second or third degree polynomial interpolation is used. It gives the best
approximation to the minimum squares. It is a very simple curve, but sometimes may be distant
from the standard points;
- Multi-point: Linear interpolating function for several standard concentrations (max 6);
- log-Logit 4 and log-Logit 5. It is a logarithmic approximation on four or five points, used for non-
linear tests;
- Semi-log: a second degree curve is used, which is function of the concentration logarithm, and
gives the best approximation to the standard points minimum squares. It is asymptotical on the
last point, has no flex points and well approximates the standard points. It may be used for
highly non linear processes.
Always bear in mind that the serum parameters will be used for the calibrations, regardless of the
sample that is going to be used.
The used calibrators or standards must be placed in the positions assigned on the
standard tray in the page Manage Standard
In linear analysis it is possible to use up to 3 different standard concentrations or execute up to a
maximum of 3 repetitions on a single point. It is possible to program the same positions for various
analyses if a multi-calibrator is used.
In analyses with curve, the standard positions available are:
- cubic, polynomial, best fit, multi point and semi-log 3 to 6 positions.
- log logit4: 4 positions
- log logit5: 5 positions
5.2. REACTION
In the Reaction page the operator must give those parameters that are necessary to define the
reaction, from the number reagents, to the type of calculation to be used for the test itself.
5.2.1. REAGENTS
Number of reagents: enter the number of reagents the methodology requires, max. 4. Use
up/down arrow keys "v" or move the cursor directly on the box and enter the value. Remember that
the number of reagents is the only parameter that must be defined before moving the test into the
on line tray.
The selected number of reagents will appear at the page base with one tag for each reagent. For
every reagent the operator must enter the following information.
Reagent Quantity (µl): insert the volume of each reagent (in micro liters) by selecting the
corresponding table (Rgt 1, Rgt 2, Rgt 3 & Rgt 4) in accordance with number of reagents
programmed. Always bear in mind that there are some limits for the final solution (reagent +
sample) as well as for sample or reagents (for limitations see table 1 at the end of par. 5.3).
Dilution: this field refers to concentrated reagents. If ready-to-use reagents are used the Type
field should be set to None. If the reagent has to be diluted, in the Type field it is necessary to
select the diluent, whether to use a dedicated Diluent or distilled Water. Then insert in the
Diluent Quantity (µl) text box the volume of the diluent to be added to the concentrated reagent.
For example: for a dilution ratio of 1:3 write 100µl for the concentrated reagent and 200µl for the
diluent. The dedicated diluent is considered by the analyzer as a further reagent and will
therefore take a position of its own in the reagents’ tray. If the diluent is the distilled water, the
analyzer will take it from the main reservoir.
In case the test is already in the tray, selecting the dedicated diluent will cause the test to be
removed from the tray in order to program the position for the dedicated diluent.
5.2.2. TIMES
The times are expressed in cycles. Every cycle is 16 seconds.
The max number of programmable cycles is 35, corresponding to 560 seconds.
As far as the reagent 1 and sample insertion cycles are fixed (respectively to cycle 1 and cycle 6),
the only programmable times are for the total reaction time and the other reagents insertion.
One single cycle allows two dispensations per arm (two of the first reagent, two sample
dispensations, two dispensations of the second reagent), a reading of all the cuvettes and the
related washings.
End Cycle: use the arrows to enter the final cycle of the reaction.
Reagent 1 & Sample: these fields are fixed as the first reagent's and the sample's cycles are fixed
by the program, respectively at the first and sixth cycle.
Depending on the number of reagents, the following fields will appear.
Reagent 2, Reagent 3 and Reagent 4. In these fields it is necessary to enter the number of cycle at
which the corresponding reagent must be introduced in the reaction cuvette. Using the mouse
cursor, click on the arrows until it reaches the correct value. In case of cycles overlapping with other
reagents, the cycle number will become red, and in case of overlapping with sample cycle, the
number will become red on yellow background.
NOTE: in every reagent cycle it is possible to insert two reagents. If one of the reagents is diluted,
the reagents become three, so it will be necessary to move the insertion of one reagent to the next
semi-cycle. The sample cycle allows the insertion of the sample and of two reagents.
5.2.3. REACTION
Calculation type: the type of calculation may be quantitative, qualitative or a combination of these
two. Select the desired calculation and, except for the quantitative calculation, enter the cut-off
value.
Formula: the final absorbance, to be used in the result concentration calculation, will be calculated
on the basis of the given formula. It is possible to select one of the already present calculations, or it
is possible to introduce brand new calculations. Use only round parenthesis when entering the
formula, as the square parenthesis are used by the program for the already available formulas.
For instance, an End point test formula will be: Point[Final]-Blank, thus meaning that the analyzer
will use the last read point absorbance subtracted of the reagent blank. A Fixed time test formula
will be: Point[Final]-Point[LastInsert], thus meaning that the analyzer will calculate the delta between
the absorbance read at the moment of the last insertion cycle and the absorbance read at the final
point.
When using the cycles, bear in mind that the analyzer will, in the same cycle, perform the insertion
first, followed by the reading phase. The reading phase will always be the last thing performed in a
cycle.
Clicking on the f(x) icon, it is possible to display the available Functions or the
formula parameters (Help).
Point[X]
[X] definition "X" meaning
[LastInsert] absorbance read at the last insertion cycle
[Final] absorbance read at the final cycle
[n] Number of the corresponding cycle. Where min. n=1
Eg:
Point[LastInsert]: uses the absorbance read when the last item (reagent or sample)
was introduced in the reaction
Point[LastInsert-1]: uses the absorbance read one cycle before the last item (reagent
or sample) was introduced in the reaction
Point[Final]: uses the absorbance read at the final cycle of the reaction
Point[6]: uses the absorbance read at cycle number 6
Regress[1,2], with the same definitions as above for X: is the linear regression calculated from
point 1 to point 2.
RegrxMinute[1,2], with the same definitions as above for X: is the linear regression calculated per
minute, from point 1 to point 2.
Mean[1,6]: with the same definitions as above for X: is the mean value calculated from point 1 to
point 6.
Sample: uses the absorbance read in the same cycle when introducing the sample in the reaction
Reagent[n]: uses the absorbance read in the same cycle when introducing reagent number "n" in
the reaction
The "0" in the formula must be changed by the operator with the necessary number of cycles.
These formulas are only examples and should be used only as guidelines to start the reaction
programming. The above mentioned formulas may also be different from the presented ones.
When programming formulas, bear in mind the absorbance sign that is expected for the calculation.
This and other info will be very important when programming the checks page.
5.2.4. BLANK
This page is dedicated to the reagent blank calculation. As for the final absorbance in the test, also
the blank must be calculated by means of a specific formula.
The reagent blank, when performed in the same cuvette where the reaction will also take place, will
be automatically read for every single sample. The absorbance value of the blank is then displayed
in the reaction graph. In the standard page will only be displayed the last blank read.
Use another cuvette: click on this checkbox to let the analyzer use a second cuvette for performing
the reagent blank. This option is useful every time the blank must calculated with a dynamic
different from the sample, or when the reagent blank must be calculated with reagents that are
inserted after the sample. In this case it is necessary to introduce the formula for calculating the
absorbance. The formula is based on the same criteria as the one for the reaction.
End cycle: use the arrow to select the number of the last cycle to be used.
Reagents used: use this option when more reagents are used for the test, considering that not
all of them are necessary to calculate the reagent blank. Just select by means of the checkbox
the reagents that participate to the blank calculation, then move the corresponding arrows until
the correct insertion cycle is displayed. In case of cycles overlapping with other reagents, the
cycle number will become red.
Repetition: when the reagent blank is performed in the same cuvette as the test, it does not
need to be repeated as it is read for every sample, but when it is performed in another cuvette,
the operator should decide when to perform the reagent blank reading.
The reagent blank, depending on the reagent itself, may need to be performed once a week, as
well as every four hours, for instance. This option gives the possibility to select when to perform
the blank.
Every run: the absorbance determination for the reagent will be performed at every work
start-up for the analyses that are going to be run.
Every day at: the absorbance determination for the reagent will be performed every day at
the specified hour. For example, setting "03"
"03" "15" determination will be performed every
every day at 03:15 in the morning.
Days + hours interval: the absorbance determination for the reagent will be performed at the
time intervals as programmed into the Days +
Days + Hours boxes. For example, setting
setting "02 + 8" determination will be
performed every two days and eight hours.
hours. Setting "00 + 8" the blank determination will be performed every eight hours.
Every "n" samples: the absorbance determination for the reagent will be performed every
"n" samples. For example, setting 15, the
analyzer will perform the reagent blank every
every 15 samples.
5.3. SAMPLE
This page is dedicated to the sample parameters, necessary for the reaction. At the bottom left of
the page there are the four tags with the various type of samples: serum, urine, CSF and Other.
The "Other" tag is left free for any other type of sample that the operator will use. For every type of
sample, it will be necessary to fill the fields described below, in order to allow the analyzer to
correctly perform the necessary calculations and to use the correct references.
Name: write here the full name of the test. The name written here will be used in the real time pages
and in the final report.
Quantity (µl): write the sample volume, expressed in µl.
Limits: the limits that have to be selected here are the normal range and the panic range, each
defined for every group: male, female, children and the other 15 available groups that have been
defined in the Setup (see chapter 9, par. 2.1 ).
Group: select the desired group for assigning the ranges.
Normal values: enter the minimum and maximum values for the normal range of the selected
group. All results falling within this range will be considered as normal and will not be flagged.
Results below the minimum or above the maximum, will be considered as pathological and
will be flagged with "-" and "+" respectively.
Panic values: enter the minimum and maximum values of the panic range for the selected
group. All results falling within this range will be simply considered as pathological and will not
be repeated when pathological repetition is active. Results below the minimum or above the
maximum, will be repeated if the pathological repetition is active. In this case, samples can be
automatically repeated with or without dilution. Attention: the pathological repetition criteria
are selected in the sample page, but the check and the automatic repetition are selected in
the Check page.
The type of sample treatment is set in the Sample management page, explained hereafter.
Sample management: the sample can be pre-diluted as well as post-treated. The sample handling
performed after the first result of the test (post-treatment) can be a sample dilution or a sample
"concentration". The sample pre-treatment, which is the only automatic pre-dilution, is applied to all
samples performing the test. It is also possible to program the pre-dilution of a single sample, but
this can be done in the Routine programming of the sample.
First it is necessary to select the type of treatment, then to set the correct treatment parameters and
to activate the sample management by clicking the Active checkbox. This procedure will only
activate the sample management parameters. Then, to allow the analyzer to perform the
corresponding check and eventually the repetition, it will be necessary to separately activate also
these other options (see par. 5.4).
Example
Sample Parameters:
Check Parameters:
Pre-treatment: this parameter is set when all samples need a pre-dilution and this is the only type
of pre-treatment, as all the other options are treatments performed after the first sampling result and
are activated only in certain conditions.
Pathological High: when a result is higher than the max value of the panic range, it may be useful
to repeat it with dilution.
Hyperactive: when a result is hyperactive (out of Test Limit, Max Abs Delta and/or Final Abs), it is
necessary to repeat the diluted sample in order to bring the absorbance back into the linearity
range.
Out of curve Above: in analyses with non linear calibration, the absorbances falling out of the
calibration curve may be misinterpreted. For this reason, it is necessary to dilute the sample and
repeat the test.
Dilution: the preparation of the diluted sample may be performed with water (from the reservoir)
or any other diluent as necessary. The type
of diluents can be edited in the Setup page
(see chapt. 9, par. 2.4) as every lab may
need different and specific diluents. The
default diluents are defined as follows:
saline solution, diluent and sample treatment
solution. Select the Type of solution to be
used, then write in the corresponding fields
the Sample volume and the Diluent volume. These two volumes will be mixed in a dedicated
cuvette from which the analyzer will withdraw the sample volume as set in the sample Quantity
field. This diluted sample will then be used for the reaction. The minimum volume for
Sample+Diluent is 120µl.
For instance, setting the parameters as follows:
Type: saline
Sample volume: 30µl
Diluent volume: 90µl
Sample Quantity
The analyzer will dilute 30µl of original sample with 90µl of saline solution (1:4), then will
withdraw 16µl of the diluted sample for performing the reaction. The read absorbance
will be then multiplied by 4 to give the final result already included of the performed
dilution.
The minimum volume for Sample+Diluent is 120µl if the Sample Quantity is less than 120µl. If the
Sample Quantity is more than 120µl, the minimum volume for Sample+Diluent would be "Sample
Quantity" + 5µl.
Pathological Low & Out of curve Below: in case of a result below the minimum panic range or
below the minimum point of the calibration curve, it may be useful to increase the sample volume to
bring the sample absorbance back into the range or the curve. Pay attention that in this case the
reaction stoichiometric ratio will be changed.
Enter the Sample volume to be withdrawn in place of the volume set in
the Quantity field.
The analyzer will calculate the multiplication factor (which will be <1) in
order to correctly calculate the result.
Min Max
Sample volume 1.0 150
Reagent volume 1.0 320
Reagent diluent volume 1.0 320
Volume in cuvette (reagents and sample) 150 420
SAMPLE MANAGEMENT
Sample 1 150
Diluent 1 320
Sample + diluent 120 320
The minimum volume for sample + diluent in sample management is "Sample
quantity" + 5µl when the Sample quantity is lower than 120µl
5.4. CHECK
The check parameters page contains a list of checks that are necessary and useful to verify the
correct reaction evolution and end.
As for the other parameters, there is the possibility to set the checks in different ways for serum as
well as for urines, CSF and other liquids.
In the Check list there are two columns: on the left side there are
the checkboxes to activate the automatic repetitions, on the right
side there are the checkboxes to activate the checks themselves.
On the right side of the page there are the check tags, one for each
check. Click on the tags to set the parameters for every control that
the analyzer will have to perform. Once the limits have been set for
one of the sample types (for instance "serum"), it is possible to copy
the same information to the other sample types by clicking on the
Copy these data to other types button.
Reagent Limit
The reagent limit verifies the good quality of the reagent.
For every test it is possible to set an Upper limit and a Lower
limit. Usually for an increasing reaction it may be expected that the
reagent blank increases, while it may decrease for a decreasing
Example for an increasing test
reaction. In some cases it may be useful to set both upper and
lower limits. If the reagent’s absorbance is beyond these limits, the
analyzer will check the results with O flag (Reagent out of limit).
This parameter allows monitoring reagents quality, as well as
checking any variation from the specific techniques.
Example for a decreasing test
Note:
This control has the highest priority on all the other check flags and inhibits the automatic
repetition functions. In case of reagent out of limit, the blank will be repeated every time the test
is performed in a run.
Form
This check is performed at three levels: reaction form,
reaction direction and tolerance within the previously
selected options.
n o p q t
r s
m
...
In the above example, the delta are always increasing up to point o, then the two couples p-q and
q-r have a decreasing delta. The delta calculated for the two couples must remain within the given
tolerance.
Test Limit
This parameter is used in all methodologies and it allows verification
of the final concentration of the analyses. It represents a threshold
value beyond which the analyzer detects an out-of-linearity condition
(hyperactivity). The concentration should be expressed with the same units used for the standard.
When the test result exceeds this value, the analyzer will flag the result as hyperactive (I flag =
Hyperactive Sample – out of Test Limit). If the Repeat check is enabled, the analyzer will also
automatically repeat the sample with the criteria expressed in the Sample Management page for
hyperactive situations.
Sample Interference
The sample interference is used to verify whether there is any
interference in the absorbance reading due to the sole sample (ex.
lipemic sample). The reading is performed after sample insertion,
therefore at cycle 7. The absorbance expected for this cycle is supposed to be a sum of the reagent
absorbance + a normal sample absorbance. If the absorbance read exceeds this value, the sample
will be flagged with i flag.
Hook
In this check the analyzer verifies the whole reaction with relation to
the reaction direction. The analyzer will verify the delta read between
every two readings (two cycles). The calculated delta should be
consistent: always>0 or always<0 (depending on the reaction direction). In case the delta
calculation results in an inverted sign (delta<0 in an increasing reaction or delta >0 in a decreasing
reaction), the analyzer will keep in memory the absorbance of the first inverted absorbance delta
point. Then it will go on calculating the delta between every couple of readings until the delta is
inverted again. The analyzer will then calculate the total delta between the first inverted absorbance
delta point and the last one, when the absorbance delta is again normal.
u
o p
...
q t
r
s
In the above example, the Hook will be calculated between points o and r. If this delta exceeds the
given value, the result will be flagged with H flag. If the Repeat check is enabled, the analyzer will
also automatically repeat the sample with the criteria expressed in the Sample Management page
for hyperactive situations.
Final Abs
This check verifies the final reaction absorbance. It indicates the limit
(upper limit for increasing analyses, lower limit for decreasing ones). If
the set absorbance is exceeded, the analyzer will detect an out-of-
linearity situation and will flag the result with A flag (= Hyperactive Sample – out of Final ABS).
This check can be selected only if also the Form check is selected, this to allow the analyzer to
know which is the reaction direction. If the Repeat check is enabled, the analyzer will also
automatically repeat the sample with the criteria expressed in the Sample Management page for
hyperactive situations.
Pathological
The results falling outside the normal range are flagged with + if above the range and with – if below
the range. The pathological check will verify if the results fall within the panic range.
If the Repeat check is enabled, the analyzer will also automatically repeat
the sample with the criteria expressed in the Sample Management page for
Pathological situations.
NOTE 1: the available checkboxes must be enabled to allow the analyzer to correctly flag the
results. If programmed but not checked, the analyzer will use the internal default values, thus giving
wrong flags (ex. a decreasing fixed time test should have the Final Abs Sign selected to Negative, if
the check is not enabled, the analyzer will use the default setting which is Positive and the result
would be <NC>).
NOTE 2: when the Repeat check is enabled, but the corresponding Check is not, the automatic
repetition does not take place.
5.5. SUPPLEMENTARY
The supplementary parameters are used to define the measurement units, the decimals, the
instrumental factor and shift.
UNITS: the measurement units will be used to report the sample's final result in the correct
concentration unit. Enter for the Primary unit the Dimension. The units may be the same as used
for the standard or may be different. In this case it is necessary to enter the Std Conv. Factor,
which is a factor by which the concentration, calculated with the standard units, will be converted
into the Primary unit.
If necessary, a second unit can be used. In the Secondary part, check Present, then enter the
secondary unit dimension and a multiplication factor to convert the primary unit into the secondary.
In the Decimals field it is possible to decide whether to use the floating point algorithm for the
decimals in the results, or to have them fixed. In this last case, disable the Automatic check and set
the number of fixed decimals desired (max 6).
The Supplementary parameters must be programmed for every type of sample independently.
Example 1. units are mg/dl both for serum and for its calibrator. For urines the dimension is mg/24h.
This value is calculated as follows:
=> mg/24H
therefore if the diuresis is given (patient entry – see chapter 5 par. 4.2.1.) in dl, the dimensional
equation will turn out correctly. But if the diuresis is given in a different unit, such as liters, the
dimensional equation must be corrected and the same is for the Std Conv. Factor.
(dl=l*0.1)
=> 0.1 * l/24H the correction factor is 0.1.
Example 2. units are mg/dl both for serum and for its calibrator. For urines the dimension is g/24h.
This value is calculated as follows:
(mg=gr*0.001)
=> 0.001 * g/24H => the correction factor is 0.001.
To print the analytical parameters, go to the menu bar, select print and then select what to print.
6. CALIBRATIONS
The calibration parameters are accessible from the same page of the analytical parameters,
selecting the Standard tag.
The standard parameters must be programmed for all type of processes: with factor, linear, with
curve, to allow the analyzer to correctly calculate the results on the basis of the read absorbance.
LINEAR PROCESS
The analyses with linear process will display a standard page as the following.
Factor
In the Factor section there are three fields that must be programmed.
Value: in this field the analyzer will automatically update the factor value calculated in the
calibration.
Min: enter the minimum acceptable value of the calculated factor.
Max: enter the maximum acceptable value of the calculated factor.
The Min and Max factor fields are necessary to verify the validity of the calibration. In case of out of
range factor value, a warning is given and the calibration is not saved, keeping the previously
memorized factor.
Standard: in this section it is possible to program the standard parameters, but not
the standard positions on the standard tray. This setting is possible only from the
Manage Standard page.
Diluent type: if the serial dilution has been programmed, select also the type of diluent to be used
for the standard dilution. Bear in mind that the diluent solutions are user-defined (see Setup, chapt.
9, par. 2.4).
Dilution 1: : if the standard is diluted, but not serially, it is possible to assign the specific dilution
ratio in this field, together with the corresponding concentration in the Conc fields.
Conc: the standard concentration that the analyzer will use to calculate the factor must be typed in
these fields.
ABS: the absorbance read by the analyzer, for every standard will be updated in these fields.
AUTOMATIC SERIAL DILUTION: the automatic serial dilution is always performed directly in
cuvette, therefore there is no need to put three cups on the standard tray. Only the highest standard
must be placed on the standard tray. The automatic serial dilution is automatically activated when
the standard positions assigned in the Manage Standard page are all the same. Only the position
corresponding to the higher standard must be programmed, repeated for the number of standard
used. For instance, if the calibration will be performed as in the example below, the standard
positions should be programmed as in part A of the table.
ATTENTION: the dilution ratio (both Serial dilution 1: and Dilution 1: ) allows the
use of the floating point. It is therefore possible to type a dilution such as Dilution
1: 3.22. The arrows will increment or decrement by 0.5.
MULTIPLE STANDARDS: when more than one standard is used, its position on the standard tray
must be defined in the Manage Standard page. In the following example it is possible to see that
the analyzer can take the standard three times from the same cup or one time each from three
different cups.
Use sample pre-dilution: enable this check if you want the analyzer to pre-dilute the standard in
the same as it pre-dilutes the sample (see par. 5.3 Sample – Sample management – Pre-treatment)
Variance%: this parameter allows the good quality of the calibration verification. It represents the
acceptable difference (in percentage) among the factors calculated if various standards are used or
if replicates are executed on a single point. After a variation above the programmed limit, a warning
is given and the previously memorized factor is not changed.
Last Calibration: in this field it is updated the date for the last positive
calibration performed for the test.
The calibrations have a dedicated archive (menu Archives – Calibrations' Archive) where all the
performed calibrations are stored. Here it is possible to view whether the calibration had a positive
result or not and the type of error. The calibrations that had problems of no reagent or no standard
are not stored.
Calibrations can also be time-programmed in order to let the analyzer remember the operator when
to run them. This feature is in the Manage Standard page (see chapt. 5 par. 3.1).
Chapter 3 Functions Page 36 of 43
BT4500 OPERATOR MANUAL
The analyzer will automatically calculate and update the scalar concentrations starting from the
highest concentration value already present in the fields. In the Dilution fields the dilution ratio will
be automatically updated for each point.
ATTENTION: the dilution ratio (both Serial dilution 1: and Dilution 1: ) allows the
use of the floating point. It is therefore possible to type a dilution such as Dilution
1: 3.22. The arrows will increment or decrement by 0.5.
Zero: used to automatically reset the concentration of the lowest point of the curve to zero. During
the predilution phase of the standards the analyzer will dispense the diluent in this position.
Use sample pre-dilution: some analyses are programmed with the automatic sample pre-dilution.
It is possible to use the same pre-dilution also for the standards. The standards will be diluted as set
in the Sample Management page at the Sample pre-treatment option.
Clicking on the button Graph the analyzer will display the calibration graph for the test.
For all the tests with calibration curve, the analyzer will warn the operator if samples (patients) have
results above or below the calibration curve. The associated flag, next to the result, will be: below
curve "<" and above curve ">". It is possible to select automatic re-running of samples outside the
calibration curve in the Check parameters (see par. 5.4) considering also that the repetition
parameters are set in the Sample Management page.
To print the calibrations, go to the menu bar, select print and then select what to print.
Calibrations as well as controls can be programmed in order to let the analyzer remember when a
timed repetition is necessary. The programming is performed in the same pages where the physical
positions on the standard & control tray are assigned: Manage Standard pages (see chapter 5, par.
3.1).
7. CONTROLS
The controls are divided into Known and Unknown,. For every type there are three levels.where
Known: Level 1, 2, 3
Unknown: Level 1, 2, 3
Enter the Lot ID# both for known and unknown controls, then complete the fields for the known
control, assigning the minimum, theoretical and maximum values of the lot in use.
To print the controls, go to the menu bar, select print and then select what to print.
The physical positions for the controls as well as the timed repetitions are programmable in the
Manage Controls page (see chapter 5, par. 3.2).
.
8. PROFILES
Analyses can be grouped in coherent lists which are called profiles. Having these
lists already programmed can make the
patients programming easier. Click on
the Profiles icon on the main icon bar. The following
page will be displayed.
The list of available profiles is displayed in the left column,
while the tests belonging to each profile are listed in the
right column.
Click on the button New Entry to enter a new profile. The following window will appear.
In the Name filed, write the name of the profile, then select a number to be assigned to this profile
(for instance for bar-coded programming). All possible tests, available or not on the tray, can be
added to a profile.
Creating a profile: click over New button: the Profile Name textbox will appear. Enter the name for
the profile and click on Analyses: this will open a window where it will be possible to select the
analyses for the profile. Click "Save" to store the selected analyses (Fig. 29). It is possible to enter
in the Code for Bar-Code field a numerical code. The analyzer, when reading the bar-code labels
of Patients, will recognize it. This code is used for completely computerized acquisition of the
analyses assigned to the patient. The bar-code settings are explained in chapter 9, par. 1.4.
To update an existing profile, select it from the list and click on Modify, and then proceed as
outlined previously. To delete an existing profile, select its name and press Erase, confirmation is
requested prior to erasing.
Attention: during the patient acquisition, only the analyses present on the tray are displayed for
each profile.
9. APPENDIX
In the following pages are presented some examples on how to program the times of an end point,
a fixed time and a kinetic test, starting from the BT3500 parameters.
End Point:
GLY
BT3500
Method End Point
Kind Of Process Linear
Filters 510 / 700
Reaction Direction Increasing
Reagent #1 (µl) 200
Reagent # 2 (µl) 0
Sample Starter Inactive
Delay Time (sec) 0
Incubation Time (sec) 360
Reading Time (sec) 10
Insert R1 Insert S
In a mono-reagent test, the blank is read automatically in the same cuvette where the reaction will
take place, therefore every single sample will have its own blank value.
BT4500
R2 starter
Insert R1 Insert S Insert R2
BT4500
Sample starter
Insert R2
Insert R1 Insert S
BT4500
R2 starter
BT4500
Sample starter
Insert R2
Insert R1 Insert S
CHAPTER 4
TECHNICAL SPECIFICATIONS Page
GENERAL INFO
Operating Mode: Sequential access
Methods: Clinical Chemistry, I.S.E., Immune-Chemistry, DOA
Test Mode: Routine, Batch, Emergencies (STAT), Profiles
Tests in memory: 500 single or double- reagent + unlimited Relation Tests
Test Re-runs: automatic or on demand
Calibrations and Controls: automatic or on demand
Quality Control: 3 known levels and 3 unknown levels
Maintenance: Automatic program
PERFORMANCES
Tests on line: 80 refrigerated reagents + relation test (40 large + 40 small bottles)
Measurement: direct reading of 84 cuvettes in optical glass
Samples Tray: 100 samples positions
Standard & Controls Tray: 26 refrigerated positions
1 arm for Samples
Sampling arms:
2 arms for Reagents
2 distinct barcode scanners for positive identification of samples and
Bar code scanner:
reagents
Direct on serum (K, Na, Cl, Li), pre-diluted urine sample (K, Na, Cl),
I.S.E. Module:
Microsample
Analytical Throughput: Up to 450 test/hour clinical chemistry; up to 690 test/hour with I.S.E.
VOLUMES
WORKING SOLUTIONS
Min / Max Sample Vol. µl 1.5 / 100
Min / Max Reagent Vol. µl 1.5 / 300
Min / Max reaction (reag +
180 / 400
sample) µl
BT3500 TIME USED WASHING SOL. (approx) USED DEDICATED SOL. (approx)
Wash with water 5’ 200 ml
Wash cuvettes 7’ 250 ml 15 ml
Extra wash cuvettes 11’ 500 ml 15 ml
Zeroing on water 6’ 200 ml
Sleep mode wash 5’ 200 ml
FCC Function 15’ 300 ml 15 ml
Cuvette single wash 2.5 ml
Needle single wash 2 ml
Consumption per test 4.5 ml
BT1000 & BT2000Plus TIME USED WASHING SOL. (approx) USED DEDICATED SOL. (approx)
Wash with water 5’ 200 ml
Wash cuvettes 7.5’ 250 ml 15 ml
Extra wash cuvettes 11’ 500 ml 15 ml
Zeroing on water 6’ 200 ml
Sleep mode wash 5’ 200 ml
FCC Function 15’ 300 ml 15 ml
Cuvette single wash 6 ml
Needle single wash 2 ml
Consumption per test 8 ml
BT1500 TIME USED WASHING SOL. (approx) USED DEDICATED SOL. (approx)
Wash with water 6’ 140 ml
Wash cuvettes 6’ 230 ml 6.6 ml
Extra wash cuvettes 12’ 350 ml 6.6 ml
Zeroing on water 8’ 140 ml
Sleep mode wash 6’ 140 ml
FCC Function 15’ 300 ml 10 ml
Cuvette single wash 2.5 ml
Needle single wash 2 ml
Consumption per test 4.5 ml
NOTE: stated times and liquid consumptions should be considered only as indicative as they
may vary in different conditions.
Operating Limits
The instrument cannot guarantee the preceding performance specs in the following conditions:
1) Ambient conditions beyond the specified range.
2) Use of non-conformant clinical chemistry products such as washing solution, distilled water, and etc.
3) Maintenance schedule and expiry date ignored.
4) Use of non-original spare parts and consumables.
The manufacturer does not guarantee the correct instrument performances in case of implementation of
unanticipated methodologies. Consult your nearest sales/service office or factory for the use of different
methodologies.
CHAPTER 5
OPERATIVE PROCEDURE Page
1. TURNING ON PROCEDURE
Turn on the analyzer by pressing the green on/off switch on the left side (refer to Chapter 2,
paragraph 3.1. Turning on the instrument for the first time).
analyzer itself and does not load the program. To properly turn on all the analyzer system, turn on
the external PC and load the analyzer program. Once the analyzer program is fully loaded, the
following main screen will appear.
On the lower bar right side there will be a number blinking. This number is the cuvettes temperature
read by the analyzer. This number will blink until the correct cuvettes temperature is reached.
On the lower bar left side there is the messages area. Double click on this area to view the
messages displayed by the analyzer.
At the center of the lower bar there is the software version of the main program and the authorized
protection level.
After the turning on procedure, allow the analyzer to warm-up. The instrument is ready for use after
approx 20 minutes, when the temperature in the cuvettes tray has reached the proper value and the
photometric lamp is stable. At this point the analyzer will require a zeroing of the photometer
2. REAGENTS MANAGEMENT
This function is accessed either by pressing F10 key or by clicking on the specific icon.
In the management pages it is possible to view the reagent volumes, insert and remove reagents,
create the current analyses tray, assign the reagent lot and expiry date and duplicate the reagents.
To create the on line tray, select from the Available list the code or codes to move. Keep pressed
the CTRL key while clicking on the codes, to select codes randomly. Keep pressed the SHIFT key
while clicking on the first and the last code of a series to select them all.
Click here to move only the selected code or codes to the on line tray.
Click here to remove only the selected code or codes from the on line tray.
When moving the codes, the analyzer will use the following criteria to assign the position on the
tray:
- first will fill the empty positions
- then will fill the positions with lower number of reagents already present.
For the type of bottle, the analyzer will first place the reagents in the large /small order, then large /
large, then small / small.
REAGENT INFO
Once all the codes have been moved, it is possible to customize the tray.
Click on this icon to enter the reagent info. The following page will be displayed.
Position: select the physical position on the tray for the corresponding reagent. Max 20 reagents
can be programmed in a single position.
Type: set the type of bottle, among those presented by the analyzer. The reagent bottles may be of
80ml, 50ml, 20ml, 10ml.
Lot ID#: type the reagent lot number.
Expiry: type the reagent expiry date.
ATTENTION
Remember that the lot number as well as the expiry date are linked to the position, not to the code
in that position, therefore, all reagents in a single physical position will have same lot and expiry
date. It is advisable to set the lot and expiry date only after all reagents in a single physical position
have been programmed to that position. The lot and expiry will be automatically copied to all the
codes in that position. Adding another reagent in the same position will automatically delete the
already present lot and expiry.
REAGENT DUPLICATION
It is possible to duplicate a reagent in order to have a backup bottle on board, already available. A
reagent can be duplicated a max of two times, for a total of three bottles on board. The analyzer will
automatically switch from the first to the second and then to the third bottle as soon as the reagent
is finished. Note that the analyzer will not switch from the third to the first bottle. Before every run
the analyzer will ask if the operator would like to perform a volumes check. This will update the
reagent volume in the bottles and enable the analyzer to use a backup bottle if necessary. The
samples performed with the backup bottle will be marked with the K flag, meaning that a new
calibration is necessary for the test.
Click on the + (plus sign) to duplicate a code or on the – (minus sign) to delete the
duplicates.
The duplicated reagents will be added on the list after the master and will be
identified as BAK and colored in light green.
The duplicated reagents must be programmed in higher positions respect to the master. To set lot
number and expiry date, proceed as for the other reagents.
Once all necessary settings have been made, click on the Save button, or click on Cancel to abort
all modifications made. Click on the Print button to print the reagents map.
The reagent tray is divided into sectors, identified by the letters A, B, C, D and E. Each sector
has 8 positions. The screen displays the representation of 8+8 bottles. The analysis codes of
large bottles are displayed on the lower positions, while the upper positions indicate the
codes used for small bottles. The diluents and I.S.E. reagents are displayed below the
chemistry reagents.
The symbols (+) or 1.XXX and 2.XXX can be displayed inside the bottles. The symbol (+) indicates
that the position is used for more than one code, while the symbols 1.XXX, 2.XXX, 3.XXX and
4.XXX indicate the position of the first, second, third and fourth bottle pertaining to a multi-reagent
code.
SECTORS
In the Sector area, the five sectors are represented on a rotating wheel. Click
on the letter corresponding to the sector and this will be presented in the
reagents page. The letter indicating the present sector is in capital letters and it
is placed at the bottom center of the wheel.
To insert a sector, select it first as explained above, then click on the Insert
Sector button. Open the reagents cover and introduce the sector, then answer
to the analyzer's question. Click on the Check sector volumes button. The
analyzer will now verify the volumes present in the sector's bottles and will
update the graphical representation. When using the bar-code, it is possible to identify the reagent
bottles by scanning the sector. Click on the Scan sector button. The bar-code settings are
explained in chapter 9, par. 1.4.
The analyzer will read the bar-code labels and assign the reagent positions automatically. In case of
errors or misreading a window will appear listing the occurred errors.
BOTTLES
To insert a single bottle, right click on it and the analyzer will display the following options.
Insert bottle: allows the operator to enter a reagent in the correct position. The
requested bottle will be presented right in front of the reagents chamber. Open
the cover and opportunely answer the analyzer's question. Then the analyzer will
verify the volume present in the bottle and will update the graphical
representation.
Present position: the analyzer will only present the required position.
Scan position: the analyzer will read the bar-code label and assign the correct code and analytical
parameters to the selected position. The bar-code settings are explained in chapter 9, par. 1.4.
VOLUMES
For every reagent and diluent bottle is graphically indicated the remaining volume in a green to red
fading scale. The I.S.E. Pack and I.S.E. Sample diluent display also the approx number of tests still
available with the remaining volume.
To see the remaining volume and tests available in the
chemistry bottles, left-click on the corresponding
position.
The upper bottle corresponds to a 10ml or a 20ml
bottle, while the lower bottle corresponds to a 50ml or
80ml bottle.
When clicking on the desired bottle, a small window
will appear showing the test name, the remaining
reagent volume and the number of tests which can be
performed with the remaining volume, according to the
set parameters.
When more than one test in the same position, the analyzer will show the list
of codes, the remaining volume and, for each code, the available number of
tests, according to the set parameters.
The positions on the standard & controls tray are programmed by clicking on the dedicated icons on
the main icons bar.
3.1. Calibrations
In the Manage standard page it is possible to select all the codes stored in memory, as well as only
those that are not in the tray or only those that are on line. Click on the corresponding button on the
top icon bar.
The selected code type list will be displayed in the page. In this page there are two tags: one is for
the Immediate standards, the other is for the Timed standards. The only difference between the
two pages is in the timed standards programming, which appears in the last column on the right and
will be explained afterward.
If the Lot ID# is not assigned, the analyzer will not run the calibration.
The buttons on the right side of the window are in common for both tags.
Select/deselect: allows simultaneous selection/deselection of all listed codes.
Modify lot name: allows the modification of the lot name for multiple codes simultaneously.
Print: opens the printing interface (see chapt. 3, par. 2.1). From this page it is possible to print the
calibrations, to save the file in pdf format or to export it in other formats.
The printout will list the codes, the positions on the tray and the Lot ID. The codes selected to run
the calibration will be marked with an X.
Run: click on this button to run the calibrations. See also par. 3.1.4. Running calibrations.
Exit: click on this button to close the Manage standard window.
The following columns are identified as #1, #2, #3, #4, #5 and #6. These numbers correspond to the
same numbers in the standard analytical parameters page.
Immediate standard
It is necessary to respect the above described relationship between numbers, in order to correctly
assign the positions on the tray.
In the above picture there are three analyses programmed. The ALBC test has a calibration curve
on positions from 4 to 9 of the standard & controls tray. The tests CRE and GLY are linear tests
both programmed to repeat three times the calibrator in position 1. It is possible to notice that the
positions assigned concomitantly to more than one test are highlighted in bold red. The same
character format is used also if the same test is repeated three times in the same position: again the
position will be in bold red.
To change the position assigned to a standard, click on the number and use the arrows or type a
different number. The field in editing will be the only white one in a line with blue background.
NOTE: the standard & controls tray positions are now in share between standards and
controls. When the analyzer is already running calibrations (or controls) and the operator
runs controls (or calibrations) which are in the same physical positions on the tray, the
analyzer will not run all of those controls which are in the same positions of the already in
run calibrations.
In the calibrations page, the three standard must be in the same position on the tray.
With this programming the analyzer will take the calibrator three times from the same cup.
The positions on the tray are in bold red, thus meaning that the same position is used for all the
three standards.
In the calibrations page, the three standard must be in different positions on the tray.
With this combination, the analyzer will take the standards from the three different cups, without
performing any automatic dilution and will calculate the factor as a mean of the three factors
calculated each on every single point.
LINEAR CALIBRATION WITH THREE SAMPLINGS FROM THE SAME POSITION AND WITH
AUTOMATIC DILUTION
In the standard analytical parameters page, all
three standards must have their dilution, manually
entered or automatically calculated by means of the
serial dilution.
The three standards must also have three different
concentrations, according to the respective dilution.
The lower concentration standard is automatically
placed as first standard.
In the calibrations page, the three standard must be in the same position on the tray.
With this combination, the analyzer will take the standard from the cup in position 1 of the tray for
three times. For the standard #1 and #2, the analyzer will also perform an automatic dilution.
It is also possible to program the analyzer to take the standards from three different cups,
performing the programmed dilution for each, but usually this is only a waste of calibrator.
The automatic dilution is performed as follows. The analyzer will take the standard from the cup and
will dilute it into a first cuvette, then it will take the diluted standard from the first cuvette and use it
for the reaction in a second cuvette. The separated preparation of every standard allows the
analyzer to dilute the standards (or the samples) with the maximum flexibility.
In the standard analytical parameters page, all six standards must have their dilution, manually
entered or automatically calculated by means of the serial dilution.
The six standards must also have different concentrations, according to the respective dilution.
The lower concentration standard is automatically placed as first standard.
CALIBRATION WITHOUT DILUTION
In the calibrations page, every of the six standards must be in a different position.
The analyzer will simply take the standard from each cup to perform the reaction.
In the calibrations page, the six standards must be programmed in the same position.
The analyzer will take the standard from the cup six times, performing every time the necessary
dilution in a separated cuvette.
It is also possible to mix the two methods, if necessary.
Verify that the selected codes have their Lot ID#. To assign a Lot name or number, double click on
the Lot ID# field of the desired code. It is also possible to select multiple codes to assign
simultaneously the lot name. Click on scattered codes keeping the CTRL key pressed or click on the
first and last code in a sequence keeping the SHIFT key pressed. Then click
on the Modify lot name button. Enter the lot name and press Save.
Once all desired codes have been selected, press on the Run button. The
analyzer will display the following question.
In case of positive answer, the analyzer will start the sampling. In case of negative answer, the
analyzer will display the standard positions map.
Clicking on every position number, the analyzer will rotate the standard tray to present the correct
position where the operator can insert the standard cup. The already inserted standards will turn
background color from white to green.
Once finished, click on the button Accept to run the calibrations. Press Exit to abort the procedure.
When there are calibrations in run, these will have a different background color in the Manage
standard page.
Attention: when a code is in run in the Immediate standards, it will be of the same color also in the
Timed standards page.
The timed calibrations programming window will be opened by double clicking on the Program
column field corresponding to every code. The programming will only activate a warning reminder
message: the calibrations will not run automatically. It will be on the user responsibility to verify the
presence of the cups and reagents and then to run the determination.
Every day at: the reminder message will be presented every day at the specified hour. For
example, setting "08" "30" the message will be
displayed every day at 08:30 in the morning.
Days + hours interval: the reminder message will be displayed at the time intervals as
programmed into the Days + Hours boxes. For
example, setting "02 + 8" the reminder will be
presented every two days and eight hours.
Setting "00 + 8" the message will be displayed
every eight hours. Setting "1 + 0" the reminder
will be presented 24 hours after the last determination.
In the Timed Standard page the programmed timers will be displayed synthetically in the Program
column as follows.
3.2. Controls
All the Controls pages are very similar to the same page dedicated to the controls.
In the Manage controls page it is possible to select all the codes stored in memory, only those that
are not in the tray or only those that are on line. Click on the corresponding button on the top icon
bar.
The selected code type list will be displayed in the page. In this page there are two tags: one is for
the Immediate controls, the other is for the Timed controls. The only difference between the two
pages is in the timed controls programming, which appears in the last column on the right and will
be explained afterward.
If the Lot ID# is not assigned, the analyzer will not run the control.
The buttons on the right side of the window are in common for both tags.
Select/deselect: allows simultaneous selection/deselection of all listed codes.
Modify lot name: allows the modification of the lot name for multiple codes simultaneously.
Print: opens the printing interface (see chapt. 3, par. 2.1). From this page it is possible to print the
controls, to save the file in pdf format or to export it in other formats.
The printout will list the codes, the positions on the tray and the Lot ID. The codes selected to run
the control will be marked with an X.
Run: click on this button to run the controls. See also par. 3.1.4. Running controls.
Exit: click on this button to close the Manage control window.
To allow the analyzer to correctly elaborate the results, it is necessary to respect the above
described relationship between numbers, in order to correctly assign the positions on the tray.
For all controls the different levels are the selected by means of the following checks.
In case the operator does not want this type of simultaneous running, it is possible to program the
analyzer in order to skip on certain levels, as shown in the picture below.
In this case, instead of a numerical position on the tray, the Known #1, Known #2 and Unknown #3
levels show an empty field. This is obtained by typing 0 (zero) instead of a number in the position
field. In this condition the analyzer will not consider the empty positions.
When the test is highlighted, it is possible to see the "0" in white on blue background.
AAA 10 11 0 0 0 0
BBB 10 11 0 0 0 0
CCC 10 11 12 0 0 0
DDD 10 11 0 14 15 0
EEE 10 11 0 14 0 0
FFF 10 0 12 14 15 16
If usually the known and unknown third level are not used, it is simply possible to program the
analyzer as in this picture.
The analyzer will ignore the positions programmed for these two levels and will not perform them.
NOTE: the standard & controls tray positions are now in share standard and controls. When
the analyzer is already running controls (or standards) and the operator runs standards (or
controls) which are in the same physical positions on the tray, the analyzer will not run all of
those standards which are in the same positions of the already in run controls.
Verify that the selected codes have their Lot ID# programmed for all selected levels. To assign a Lot
name or number, double click on the Info field of the desired code.
It is also possible to select multiple codes to assign simultaneously the lot name. Click on scattered
codes keeping the CTRL key pressed or click on the first and last code in a sequence keeping the
SHIFT key pressed. Then click on the Modify lot name button. Enter the lot name and press Save.
Once all desired codes have been selected, press on the Run button. The
analyzer will display the following question.
In case of negative answer, the analyzer will display the control positions map.
Clicking on every position number, the analyzer will rotate the control tray to present the correct
position where the operator can insert the control cup. The already inserted controls will turn
background color from white to green.
Once finished, click on the button Accept to run the controls. Press Exit to abort the procedure.
When there are controls in run, these will have a different background color in the Manage control
page.
Attention: when a code is in run in the Immediate controls, it will be of the same color also in the
Timed controls page.
The timed controls programming window will be opened by double clicking on the Program column
field corresponding to every code. The programming will only activate a warning reminder message:
the controls will not run automatically. It will be on the user responsibility to verify the presence of
the cups and reagents and then to run the determination.
Every day at: the reminder message will be presented every day at the specified hour. For
example, setting "08" "30" the message will be
displayed every day at 08:30 in the morning.
Days + hours interval: the reminder message will be displayed at the time intervals as
programmed into the Days + Hours boxes.
For example, setting "01 + 6" the reminder will
be presented every one day and six hours.
Setting "00 + 8" the message will be displayed
every eight hours. Setting "1 + 0" the reminder
will be presented 24 hours after the last determination.
Every "n" samples: the controls determination will be performed every "n" samples. For example,
setting 20, the analyzer will remind to run the
controls every 20 samples.
In the Timed Controls page the programmed timers will be displayed synthetically in the Program
column as follows.
4. SAMPLES
The samples, i.e. the patients can be programmed easily in batch mode or in a better detailed way
using the routine mode (Patients menu).
It is possible to program STAT samples at any moment in the Patients work-list.
As in other parts of the program it is possible to display all test codes, only codes in the current tray
or only codes not in the current tray. Select the desired list by clicking on the button.
The list of the codes is displayed. Click on a code to program the batch. The samples tray has 100
positions available. To make the programming easier, use the Select and Deselect buttons after
having entered a numerical range to operate with.
For instance: select as range "38" and "48" and click on Select: all positions in this range will be
selected. On the opposite, if all 100 positions (or less) are selected and it is necessary to free other
positions, select the range, again "38" and "48" and click on Deselect: all positions in this range will
be deselected.
After having selected the positions on the tray, it is necessary to confirm which tests
will have to be performed. This is done by clicking on the checkbox near each
analysis code. When the check is present, the test will be performed.
The type of sample used in the batch is by default the serum. It is possible
to select a whole batch to be on urine by clicking on the checkbox Batch
on urine.
In this case all samples will be dealt with as urines.
When the batch has been selected (positions, codes and sample type) click on the Run button to
perform the tests selected.
The analyzer will display the following question.
The patients already programmed in the current work-list will not be lost, but will be moved to the
extra list. This is an additional list, where the positions on the tray are not assigned. In a second
moment it will be possible to move again the programmed patients to the current list (see par. 4.4).
Answering no, the procedure will be aborted. Answer yes if the current list is empty or if there are no
problems in moving the patients. The analyzer will display the following question.
Answer yes if the cups are already positioned in the tray, or answer no to let the analyzer guide you
in the cups insertion. In this case the analyzer will display the batch list as below.
Click on a single position and the analyzer will rotate the tray in order to insert easily the cup in the
correct position. The inserted cups' positions will turn from red to blue background.
Once all selections are done, click on the Accept button to let the analyzer start the run.
To start the run immediately, skipping on the above mentioned procedure, answer Yes to the
question "Do you have already inserted the samples?.
The analyzer will display the following question.
The volumes check is useful when there are backup reagents on board. With the check the analyzer
will update all reagents volumes and will be therefore able to swap to the backup reagent when
necessary.
NOTE: the swapping procedure is explained at par. 2.3.
The running samples are always loaded into Current list. To view these, click on the Patients icon.
There it will be possible to view all the work-lists: Current list, Extra list and Samples performed.
The samples programmed in batch will be identified as Batch XX.
To program new patients it is possible to click on the New Entry button or to click on the number of
any free position in the list. The Sample's Data window will be displayed.
On the right and bottom sides of this window there are respectively the Profiles and Analyses
fields. Click on these fields to open/close the corresponding lists. When the Sample's Data window
is opened, by default it will have the analyses list open and the profiles list closed as in the picture
below.
Samples can be saved only if a code is assigned. If analyses have not been assigned, the analyzer
will warn the operator with a message, but the sample can be saved and completed later.
In this page the number of the position in programming is reported in the header.
Code: type in this field the ID code for the patient. This ID# can be used for fast search in the
archives. It is the only mandatory information for saving the
sample. The user can also enter the code of a patient saved in the
work list, even if in execution. In this case, a message asks the
user to confirm patient’s data cloning. All the data relevant to the patient are instantly displayed and
are linked to the current position. The cloning of a patient’s code allows the user to obtain one report
in case different samples are used.
The "plus" symbol allows the duplication of the patient. A duplicate can be changed in every part of
its programming. Clicking on this symbol the following window will appear.
Enter the Number of duplicates as desired, considering that the
final number will be "master+duplicates". The ID code can be auto-
assigned (Autobatch XX) or can be assigned by the operator.
Enter in this case the Start code ID#. It is possible to use numbers
and letters. The codes assigned by the analyzer will follow these
criteria:
Numeric code: it is increased by 1 unit
Alphanumeric code: one character is added, starting from 1.
The "0" at the beginning of the numeric codes will be deleted.
Examples:
Start code ID# is 001 – four duplicates
Start code ID# is 0 – five duplicates (codes from 1 to 5 have been skipped as already present
in the work-list)
When entering the Start code ID# and the Number of duplicates in an already existing work-list,
remember that the already present codes will be skipped. For instance if codes 10-11-12 and 13 are
already present and the starting code for duplication is 9, the analyzer will create code 9, followed
by 14-15-16, etc.
PATIENT DATA
Surname and Name: enter the patient's data. Then enter in the Note field the necessary
information to complete the patient's clinical situation. These information will also be saved into the
Patients Archive.
Then it is necessary to select the following.
Element: select the type of processing for the sample. It may be a normal Routine, a STAT or a
Control.
In case the operator wants to run a control among the samples, it is possible to set
here the control parameters. Select the Type if known or unknown and the Level if
Level 1, Level 2, or Level 3.
Group: select one of the available groups. Remember that groups are user-defined (see Setup,
chapt. 9, par. 2.1). The group definition is necessary to correctly assign the
pathological flags and to allow the automatic repetitions (when programmed) for
the samples that fall outside the panic range.
Type: in this field select the type of sample. As every type of sample has separated parameters, it is
necessary to correctly assign this option to allow the analyzer to correctly
process the sample.
Birth date: entering the patient's birth date in this field, the analyzer will automatically calculate and
update the Age field. It is also possible to write the Age, in this case the
analyzer will calculate the Birth date (only the year, day and month will be the
current). The birth date field will be updated when saving or clicking on it.
Draw date: the current date is automatically updated here. Click on the calendar icon to select a
different date or type it in the field.
These two functions are used alternatively, depending on the programming. If the programming is
as in the picture, none of the two functions is enabled.
Inactive
Programming in one of
these two ways, there
is no pre-dilution and
no result recalculation
With Predilution
Pre-dilution is performed with the programmed pre-
dilution factor, using the diluent which is programmed in
the Sample Management (see chapter 3, par. 5.3). The
analyzer will look for the suitable diluent into the Sample
Management items, searching for an active option,
starting from the Pre-treatment. If none of these is active, the analyzer will use the saline
solution.
In these condition the analyzer will also disregard all other sample treatments programmed in the
analytical parameters of the assigned tests. If a repetition of the sample is necessary, the
analyzer will again use only the dilution programmed in the patient data and will skip on the
Sample Management parameters.
With External Factor
In this conditions, the analyzer will manage the sample
for every test as programmed in the analytical
parameters (Sample management), but the final result
will be multiplied by the external dilution factor.
ANALYSES
As already seen for the analytical parameters, it is possible to display, when programming the
patient, the current analyses tray or all the programmed codes.
To assign the analyses to a patient, click on the test code button or on the checkbox.
Selected tests will have the check next to the code, in the check-box. To deselect the
assigned analyses, click again.
To assign the profiles, first open the profiles list by clicking on the right side of the window, on the
Profiles bar. Click once on each profile to see the analyses belonging to it,
or double click to assign the profile to the patient.
Once profiles are assigned it is still possible to change the patient's analyses by clicking singularly
on each.
It is also possible to restart all the patient's analyses programming by clicking
on the Deselect all button. This will deselect all assigned analysis, without
changing anything in the patient data.
To move back and forth inside the work-list it is possible to use the two arrows.
When a patient programming is finished, click on the Save button and the analyzer will automatically
show the next free position.
Press Cancel to abort the programming and to go back to the work-list.
The user can run immediately the sample or save it in the work-list.
If the sample is run immediately, the analyzer will ask if the sample has already been inserted in the
tray, if not, the analyzer will guide the operation in placing the cup in the correct position on the
samples tray (see par. 4.3. Running samples).
After the STAT sample has been run, the analyzer will have to complete the R1+S phases that were
already started. This means a waiting time that may vary from 0 to 6 cycles, for a maximum of 90".
After the STAT sampling is completed, the analyzer will go back to the normal routine work.
Alternatively it is possible to select the samples one by one, clicking on each check-box.
Then click on the Options menu, go to Selected samples and select Run.
As already seen for batches, the analyzer will display the following message.
Answer yes if the cups are already positioned in the tray, or answer no to let the analyzer guide you
in the cups insertion. In this case the analyzer will display the work-list as below.
Click on a single position and the analyzer will rotate the tray in order to insert easily the cup in the
correct position. The inserted cups' positions will turn from red to blue background.
Chapter 5 Operating procedure Page 31 of 37
BT4500 OPERATOR MANUAL
Once all selections are done, click on the Accept button to let the analyzer start the run.
To start the run immediately, skipping on the above mentioned procedure, answer Yes to the
question "Do you have already inserted the samples?.
The analyzer will display the following question.
The volumes check is useful when there are backup reagents on board. With the check the analyzer
will update all reagents volumes and will be therefore able to swap to the backup reagent when
necessary.
NOTE: the swapping procedure is explained at par. 2.3.
The current work-list will appear in this way: the red colored samples are in run, while the blue
colored samples are already programmed, but not in run.
When the bar-code reading is active (see bar-code settings in chapter 9, par. 1.4), the patients
samples can be placed directly on the tray for positive identification.
When the tubes are positioned, click on the Options menu of the Current list, select Bar-code and
then one of the available commands for running samples: see next par. 4.4.
Bar-code: when the bar-code on samples is enabled, clicking on this line will open another menu
with the following functions.
Scan all tray: starts a reading of all bar-coded tubes on the samples
tray.
Scan single position: reads the bar-code in a position selected by
the operator.
Scan all tray and run: makes a bar-code reading and starts the run
of the positively identified samples. Errors, if present, will be
displayed on the screen.
Scan single position and run: makes the bar-code reading of a
selected position and runs the sample.
Extra list Options. In the Extra list options, the functions are similar to those seen for the Current
list. The Run line is not available as the extra is a storage list with
no position assigned.
When using the bar-coded tubes and the Host computer, it is possible to send the work-list from the
Host and to use it immediately or to store it into the Extra list. When the tubes are introduced into
the samples tray they are read with the bar-code reader. The analyzer will match every read code
ID with those present in the Extra list and will move only these samples to the current list for
running.
It is possible also to view all the test results and reaction graphs by double clicking on the sample's
code. In this case, the analyzer will open a window like the following.
On the left side there are the patient's data, while on the right side are listed the performed tests
with their own result. The red background indicates the presence of flags for the test.
Next to each code there is also a small icon representing a graph. Click on this icon to view the test
graph.
The same information that are shown clicking the Graph icon in the main icons bar are shown here,
but these info refer strictly to the selected test belonging to a specific patient. In the upper part of the
window there is the reaction graph. On the left there are the absorbances read, with the cycle they
were read at. On the lower part of the window there is a summary of the reaction the information
and the information used for calculation. The Print button opens the printing interface (see chapt. 3,
par. 2.1).
New Entry: this command will open the Samples Data window for
programming a new patient.
Close: this command will close the whole Patients management window.
Exit from the program: with this command, only the analyzer
program will be closed. The PC will remain turned on.
NOTE: if cuvettes are not properly washed at the shut-down, at the following start up the
analyzer will ask for the cuvettes extra wash. Performing the normal cuvettes wash (with the
cuvettes washing solution) will not prevent the analyzer from warning again that the extra
wash is needed.
Conditions are:
a. If no test has been performed before the shut-down, at the following start-up of the program, no wash
message will appear.
b. If no test has been performed before the shut-down and the shut-down wash has been started but it did not
complete correctly, then at the following start-up of the program, the analyzer will warn the operator that the
preceding wash was not correctly completed.
c. If tests have been performed and the shut-down wash was not performed, at the following start-up of the
program, the analyzer will ask for the Extra wash cuvettes and will warn the operator that the preceding wash
was not correctly completed.
NOTES:
1) Having performed a test: it means having performed either a single test with reagents and sample or any
procedure involving colored solutions, such as the FCC.
2) The analyzer asks for the Extra wash cuvettes as it does not know how long the solutions were left inside
the cuvettes before washing them.
CHAPTER 6
QUALITY CONTROLS Page
1. QUALITY CONTROLS
As all the external archives, the QC has a backup function. It is advisable to perform a periodical
backup of the archive, every 3 or 6 months, depending on the amount of data which are present in
every archive. The backup file format is not readable by a program different from the Analyzer.
It is also possible to export the archives in different file formats as pdf, rtf and csv (see Chapter 3,
par. 2.1 Printing interface) that can be easily read with an appropriate program on the PC.
From the Data menu it is possible to enter external QC data, view the QC stored
data and also the Youden graph.
In the Utility menu, there are the Backup and Restore functions, together
with the Manage external data and Delete data functions.
When the button Save is pressed, the control range window will appear. Type the control range for
the selected type and level and save with the button Yes.
Make all the selections necessary for processing the data, then press the Search button.
The results will be displayed as a list. The data are ordered by date and time. Out of range controls
are indicated with an asterisk and have red background.
The Westgard graph provides a global vision of a given lot by plotting data on a diagram having the
origin at 0 line representing the lot mean value, calculated as (lot min + lot max)/2, which usually
corresponds to the lot theoretical value. As division, are reported values such as Mean±1S ±4S. All
data within the given range will be displayed with a green circles while the out of range data will be
plotted using red circles.
It is a procedure for classifying values of a Known lot taken into consideration for processing.
Mean: mean between the maximum and minimum lot value
S: (Maximum lot value – Minimum lot value) / 8
The following procedure is observed for classification:
Class C (2-2S): Two consecutive results exceed mean by 2S in the same direction.
Class D (R-4S): Difference between two consecutive results is higher than 4S and at least one
result exceeds mean by +/- 2S.
Class E (4-1S): Four consecutive results exceed mean by more than 1S in the same direction
and at least one result exceeds mean by +/-2S.
Class F (10x): Ten consecutive results are all in the same direction of the mean value and at
least one result exceeds the mean by +/- 2S.
Classes are controlled from F to A. The classes are mutually exclusive, that is, if a given value is
mapped to one class then it can't be part of another class.
Example:
2 E classes means that 5 consecutive results exceed Mean by more than 1S in the same direction
and at least one result exceeds mean by +/-2S; alternatively two groups of 4 consecutive results
exceed the mean by more than 1S in the same direction and at least one result exceeds mean by
+/-2S.
Click on the Print button to print the graph and/or to export it (see printing interface chapt. 3, par.
2.1).
The Daily Chart graph provides a global vision of a given lot, plotting data on a diagram having the
origin represented by the actual mean (not by the theoretical value) of the same lot and as divisions
values such as Mean+Standard Deviation*1, Mean+Standard Deviation*2 to Mean+Standard
Deviation*4. In addition the red dashed lines indicate the upper and lower limits of the lot. All data
within the range will be represented by a green circle, while the out of range data will be plotted
using red circles.
Click on the Print button to print the graph and/or to export it (see printing interface chapt. 3, par.
2.1).
After selecting the general search parameters Analyses, From Date, To Date, choose the specific
Level, Lot ID#, Method for the X and Y axes. Data processing and visualization is performed
pressing the button Search. The controls are ordered by date and in case of known controls the Out
of limit condition is indicated.
This function relates two different levels for the same lot displaying distribution of controls within the
limits of lots.
Click on the Print button to print the graph and/or to export it (see printing interface chapt. 3, par.
2.1).
1.4. Utility
The UTILITY menu contains a few functions: the backup, the restore
Backup: it is used to make a backup copy of the present archive. The
exported archive can be restored in any moment, but it will overwrite the
existing QC archive. The backup function will keep a copy of the archive in
a safe place (the files can be copied in any other location or PC). It is
advisable to make a backup every 3 to 6 months and to delete completely
the QC archive every time a backup is performed. An important alternative
is to use the export functions available in the printing interface chapt. 3, par.
2.1
Restore: the restore function will import and overwrite the present archive with one of the available
backup.
The backup and restore functions are very important as every PC may face memory or hard disk
problems. In this cases it is possible that the archives are lost or damaged. A copy of the archives
stored in another PC or any other device, different from the analyzer's PC may allow the total
restoration of data.
Manage external data: this function allows to view an exported archive (backup) without restoring it
into the analyzer's HD.
Select the location where the archive was saved and open it.
The Manage external data sentence will be displayed on the
main title bar and in the Utility
menu there will be the check
sign indicating that the analyzer
is dealing with the External
data. Once the external archive
has been loaded, just click on
the Manage data button to view
it, exactly as is normally done
with the internal archive. To go
back to the internal archive, just remove the check sign by clicking again on the Manage external
data line.
Delete data: data in the archive can be deleted with the same search criteria already seen in the
backup search. It is possible to delete the whole archive
performing a search leaving in the criteria fields all the
default values. Otherwise it is possible to perform a
search by lot number, by date, etc..
CHAPTER 7
EXTERNAL ARCHIVES Page
1. EXTERNAL ARCHIVES
The External Archives are: Patients Archive, Population Archive, Quality
Controls, Calibrations Archive and Quality Controls in Real Time. The
two Quality Controls archives are explained in Chapter 6.
With the exception of the QC in real time, all the external archives have a
backup function. It is advisable to perform a periodical backup of the
archive, every 3 or 6 months, depending on the amount of data which
are present in every archive. The backup file format is not readable by a
program different from the Analyzer.
The backup and restore functions are very important as every PC may
face memory or hard disk problems. In this cases it is possible that the archives are lost or
damaged. A copy of the archives stored in another PC or any other device, different from the
analyzer's PC may allow the total restoration of data.
It is also possible to export the archives in different file formats as pdf, rtf and csv (see Chapter 3,
par. 2.1 Printing interface) that can be easily read with an appropriate program on the PC.
2. PATIENTS ARCHIVE
The Patients Archive module stores
the patients reports. For every patient
the given personal information are
saved together with the analyses
results, possible flags and notes.
It is possible to perform searches in
the archive, based on different criteria.
The archive can be exported, copied
as internal backup, viewed as external
archive, deleted and printed.
In the archive main page there are two
main menu: Utility and Data. Two
buttons allow: entering in the Manage
page and Exit from the archive.
The patients in the Archive can be searched for, with a certain number of criteria:
From code / To code: indicate the first code and the last code within which the search must be
performed.
Surname / Name: these options are useful when searching for a known patient.
From date / To date: allows to select a range of dates within which the analyzer can search
patients data.
Once the criteria are specified, click on the button Search. To repeat the search with different
criteria, click on Clear search.
Leaving all fields unchanged (default values), and clicking on the Search button, the analyzer will list
all available patients.
When patients are listed, click on one line to view that patient's data and results in the right side of
the window. The tests with red background also have an asterisk to remember there is a flag.
Double click on one of these red lines to read the flag.
Sort order: the patients can be listed in a specific order, that can be
decided by the operator.
The available options are: by code (alphanumeric order), by Draw date
or by Surname. With the None option, patients will be listed by date.
The backup and restore functions are very important as every PC may face memory or hard disk
problems. In this cases it is possible that the archives are lost or damaged. A copy of the archives
stored in another PC or any other device, different from the analyzer's PC may allow the total
restoration of data.
Delete data: data in the archive can be deleted with the same
search criteria already seen in the backup search. It is possible to
delete the whole archive performing a search leaving in the criteria
fields all the default values. Otherwise it is possible to perform a
search by code, by date, etc..
The DATA menu offers, besides the possibility to open the Manage data
window, also the Manage external data option. This function allows to
view an exported archive (backup) without restoring it into the analyzer's
HD.
Select the location where the archive was saved and open it.
The Manage external data sentence will be displayed on the
main title bar and in the Data menu there will be the check sign
indicating that the analyzer is dealing with the External data.
Once the external archive has been loaded, just click on the
Manage data button to view it, exactly as is normally done with
the internal archive. To go back to the internal archive, just
remove the check sign by clicking again on the Manage
external data line.
3. POPULATION
Press the Save button and go ahead with the following result. Close when finished. At this point
statistical processing of the data becomes available.
Data inserted manually will be managed exactly like the data from the analyzer.
Once the criteria are specified, click on the button Search. To repeat the search with different
criteria, click on Clear search.
Leaving all fields unchanged (default values), and clicking on the Search button, the analyzer will list
all available data.
The statistics related to the search will be displayed in the
colored area on the right side of the window. The analyzer will
calculate:
Number of (selected) records
Minimum result
Maximum result
Mean
Standard Deviation
Correlation coefficient
Variation coefficient
Variance
Deviance
Median
The performed search can be printed or exported (see printing interface chapt. 3, par. 2.1), by
clicking on the Print button.
It is possible to print only the statistics or these and the search data, as in the above example.
The Trender graph shows the data sequence together with its related minimum square line in the
lower part of the graph the equation for the line is shown.
The L. Jennings graph shows data plotted with the mean and the standard deviation.
The Statistic Histogram graph shows the histogram for the selected data with respect to the mean
value, used a origin line.
Standard Deviation:
SD: is a quantity that measures the spread of data across its mean value. If data is mostly located
near the mean SD assumes a small value, otherwise a large value indicates large data spread.
n
DS =
Σ (Xi – X)
i=1
2
n-1
Variation coefficient:
CV%: is computed as the ratio between mean square error and arithmetic mean. CV% is a relative
quantity and independent from the measurement unit used.
DS*100
CV=
X
Minimum square line y = ax + b
Where: Σ XY-nΣxΣy
a=
Σ X2 - n(ΣX)2
b= ΣY-aΣx
Correlation Coefficient:
Σ[(Y-Y)*(X-X)]
CC =
Σ(X-X)*(Y-Y)
Variance:
[ X* X]
ΣΧ2 - Σ nΣ
V=
n-1
Deviance:
( ΣX)2
D =Σ X2 - n
Median:
The Median for an ordered ser of data is the central value or the arithmetic mean of the two central
values, depending on the fact that the number of elements in the set be odd or even. In particular
the median for N odd elements, is the X[(N+1)/2]th element.
Example.: given the following 9 elements set
1, 2, 2, 17, 21, 34, 34, 34, 67
The median for N even elements, is the { X(N/2) +X[(N+2)/2] }/2 element.
Ex.: be given the following 8 elements set
1, 2, 12, 24, 26, 45, 45, 46
{X[4]+X[5]}/2 = (24+26)/2 = 25
median is 25.
By comparing the median and the arithmatical mean, one can assume the existence of a
measurement error or an asymmetry in the distribution function, in case these two quantities differ
greatly.
The backup and restore functions are very important as every PC may face memory or hard disk
problems. In this cases it is possible that the archives are lost or damaged. A copy of the archives
stored in another PC or any other device, different from the analyzer's PC may allow the total
restoration of data.
Delete data: data in the archive can be deleted with the same
search criteria already seen in the backup search. It is possible to
delete the whole archive performing a search leaving in the criteria
fields all the default values. Otherwise it is possible to perform a
search analyses, by date, etc..
The DATA menu offers, besides the possibility to open the New Data and
Manage data windows, also the Manage external data option. This
function allows to view an exported archive (backup) without restoring it into
the analyzer's HD.
Select the location where the archive was saved and open it.
The Manage external data sentence will be displayed on the
main title bar and in the Data menu there will be the check
sign indicating that the analyzer is dealing with the External
data.
Once the external archive has been loaded, just click on the
Manage data button to view it, exactly as is normally done
with the internal archive. To go back to the internal archive,
just remove the check sign by clicking again on the Manage
external data line.
CHAPTER 8
DISPLAYING AND PRINTING RESULTS Page
Results: this button will open the pages where the patients results in real time are displayed. This
page is updated every time a patient is completed and is the only RT page that includes other
commands.
Results RT: in this page all results, independently from the patient to whom they belong, are
updated as soon as every test is completed.
Graph: this page shows all the available test graphs.
Calibrations and controls are listed in these pages together with the test results. Calibrations are
written with green text and controls are written with blue text.
In this pages are updated the information related to patients, calibrations, controls and possible
errors that occurred during execution.
If there is no data the page appears blank. A color code makes it possible to quickly identify the
information:
red text: presence of flags in at least one test of the sample
blue text: controls
green text: calibrations
black text: no anomaly in the test or the entire sample
It is a brief representation of data and allows visualization of results of patient in execution as the
tasks for the single patients are completed.
Once the results are archived, the information present in this page will no longer be available.
a) Sample Position (#XX) Physical position of the sample in the samples tray or in the standard
& controls tray.
b) Sample Code Patient ID#, batch or autobatch number, standard or control code..
c) Surname, Name Patient’s personal data.
d) Sample Type Between parenthesis it is indicated the type of sample, whether
Routine, STAT, control or standard type.
e) Date and time Indicates when (date and time) the sample was performed.
f) Results The patients results are represented as follows:
At the top right of the Results page there is a small button Info flags.
Click here to see a summary of the flags that are used in the Results page.
At the bottom of the Results page there are more useful buttons.
Print: click here to print the Results page. This button will open a printing interface as already
described in Chapter 3, par. 2.1 Printing interface.
For Analyses: with this button the analyzer will arrange the results per test instead than per patient.
Adjust: this function allows results recalculation. Note that recalculated results will be flagged with
the additional flag E.
The Adjust button gives the following choices.
With standard: this function will re-calculate the results in case there is a new calibration available.
The analyzer runs the recalculation starting from the absorbance memorized for the test, thus the
various analytical parameters and the last valid
calibration are taken into consideration. In this
case, if the instrumental factor and shift have
been set in the analytical parameters, they will
be used in calculating the new value.
With factor: this correction is performed in percentage and may be increasing as well as
decreasing.
Once the type of correction, with factor, has been selected,
the analyzer will display the analyses that can be re-
calculated.
Click on one code to open the correction window.
With standard (only if performed with backup reagent): this type of correction is performed when
a certain number of tests have been partially performed with the backup reagent and a calibration
with the backup reagent has been performed too, but after the test results. In this case it is possible
to recalculate the results with the new calibration.
Store data: Click this button to save the results into the patients archive. Bear in mind that once the
results are archived, the graphs are not available anymore. Therefore, it is a good idea to print any
pertinent graph before archiving data.
It is possible to save all results or to save data for a selected group of samples.
In the same way, iy is possible to decide whether to archive also the not completed samples.
If the uncompleted samples are not stored, they will remain in the work-list in order to be run and
completed in a second moment.
Clicking on the Select samples line, the analyzer will show the following window. Here it is possible
to select the single patients by
means of the check sign next to the
ID code. There is also a small arrow
near each code. Click on this arrow
to view the list of tests performed for
that sample. With this option it is
possible to select, not only the
sample, but also the single test that
has to be stored.
Clear: click on this button to delete all results in this page. All data will be definitively lost.
Sort: click on this button to arrange the results with the criteria selected in the Setup (see chapter 9,
par. 2.3.).
NOTE: once the results are archived, the information present in the results page will no longer be
available.
The data refer to the results obtained by the analyzer in real time, i.e. as the tests are completed.
For this reason the results of tests with a shorter incubation and reading time may appear first even
if they belong to later patients. The results are not sorted in any way, not per patient, not for type of
test. The data display is synthetic.
Once the results are archived, the information present in this page will no longer be available. The
results of tests associated to flags are shown in red.
a) Sample Position (#XX) Physical position of the sample in the samples tray or in the standard
& controls tray.
b) Sample Code Patient ID#, batch or autobatch number, standard or control code..
c) Sample Type Between parenthesis it is indicated the type of sample, whether
Routine, STAT, control or standard type.
e) Date and time Indicates when (date and time) the sample was performed.
f) Results The patients results are represented as follows:
With the Print button it is possible to print this page, and with Close it is possible to close the page.
The reaction graphs are divided into parts by dashed red axes, each of which indicates an
absorbance value used in the result calculation. The above example is a fixed time test. In this case
the result is calculated by means of an absorbance delta, which is determined between the two
cycles (17th and 6th) corresponding to the red dashed lines.
The X axe indicates the cycles (time) and the Y axe indicates the absorbance.
The single absorbance per cycle is indicated in the two columns at the right of the graph.
In the Statistic area, there are the most important information concerning the reaction. This area is
divided in two parts. The left part lists the absorbances in relation to the whole reaction, while the
right part refers only to the absorbances used in the result calculation.
The last line in this area gives the final reaction result and the reagent blank value.
When opening the graphs page, the analyzer will show the first available graph.
Use the left/right arrows to see the other reaction graphs of the
same patient.
Use the up/down arrows to select the preceding or the
following sample.
It is also possible to click on the Code line to choose one of
the available sample codes and view its graphs.
O Reagent Limit
! Final Abs Sign
I Test Limit
d Max Abs Delta
A Final Abs
H Hook
F Form
+ Path High - Range Normal Values >Max
- Path Low- Range Normal Values <Min
K Test performed with backup reagent
> Out Of Calibration Curve – Above
< Out Of Calibration Curve – Below
>> Out Of Calibration Curve After Repetition - Above
<< Out Of Calibration Curve After Repetition - Below
i Sample Interference
S No Sample
R No Reagent
-- Negative Concentration
? Not Calculable – Relation test – one or more of the built-in analyses is missing
* Relation Test - Error In Built-In Analyses – there are error flags
X Result Impossible – negative value
P No Data To Calculate – data are not enough
E Sample Recalculated
d No Diluent (saline, diluent, etc)
N Reagent Expired
M Error In Instr Fact/Shift
I.S.E.
Y I.S.E. Hydraulic Malfunctioning – funnel does not empty
B Bubbles In I.S.E.
D Drift On I.S.E.
++ Concentration Too High
a No Sol A (I.S.E.)
b No Sol B (I.S.E.)
c No I.S.E. Sample Diluent
r No Reference (I.S.E.)
U Urine Calibration is missing
w I.S.E. Waste Full
CHAPTER 9
SETUP Page
ANALYZER SETUP
The analyzer setup is used to customize certain aspects of the analyzer program, such as the report
format.
To open the Setup, click on the Utility main menu and select the command Setup.
The Setup window is divided in two parts. On the left there are the two main menus, System and
Patients which can be alternatively selected. On the right there is a window where all the details
dedicated to each setting are available.
1. System
In this menu are available all of those settings generally applied to
the whole system.
1.1. Language
In this window it is possible to select the desired
language among those available. To select a
language it is first necessary to have saved the
external translations.
The external translations are text files that can be
easily translated in other languages, therefore it is
virtually possible to translate the analyzer software
in every language. The only limitation is that the
computer Operative System should support the
same language.
Click on the button Load external translations.
Select the folder where these translations are
stored and click OK. The available translations will
be visible in the list Stored Translations. Double
click on the desired language and save the Setup.
1.2. Printer
The Printer page can be used to customize the
analyzer reports with the name and address of the
Lab and the name of the Lab responsible. There
are two areas where it is possible to add the text:
the header and the footer. The text character can
be formatted as desired.
1.4. Bar-code
This option enables the bar-code scanning. It can
be separately enable on samples or on reagents or
on both.
The bar-code on reagents can simply be enabled
or disabled, with no further settings.
ENTERING PATIENTS
Tests to be run can be acquired from the Host Computer, from those stored after manual input and
through barcode scanning (see chapt 5, par. 4.2.1, 4.3 and 4.4).
Manual programming: the patient is programmed manually (code and analyses), then his position
on the sample tray is checked by barcode reading. In this case the samples can be inserted
randomly on the tray, since the barcode reader will assign the position on the tray to the patient
corresponding to each code read which has the same ID code.
Programming via Host Computer: The patient is programmed (code and analyses) on the Host
Computer and transferred to the analyzer. The position on the sample tray is checked by barcode
reading as described above.
Entering patients with and without bar-code on the same tray
Patients with barcode and those without can be run together. Patients without barcode must be
inserted exactly into the positions assigned during programming. Patients with bar-code can be
freely placed in any of the free position with no need to respect the assigned position number.
In case a bar-coded sample is found in the same position where a manually programmed patient is
stored, the analyzer will move the manually programmed patient to the Extra list.
Supported Symbologies
- Barcode (1D): JAN/UPC/EAN incl. add on, Codabar/ NW-7, Code 11, Code 39, Code 93, Code
128, GS1-128 (EAN-128), GS1 DataBar (RSS), IATA, Industrial
- 2of5, Interleaved 2of5, ISBN-ISMN-ISSN, Matrix 2of5, MSI/Plessey, S-Code, Telepen, Tri-Optic,
UK/Plessey
- Postal code: Chinese Post, Korean Postal Authority code
- 2D code: Composite codes, MicroPDF417, PDF417
EAN13
If the bar-code used is an EAN13, there are some limitations: the first number must not be 0 and the
actual number of digits is 12 as the last digit is the checksum X (automatically created).
1.6. Temperature
There are three devices that are temperature-controlled. The cuvettes, the standards and the
reagents trays. All the three devices can be separately enabled or disabled.
Cuvettes tray: the cuvettes are kept at a
temperature of 37°C. The heating of the
cuvettes can be disabled if necessary. For
instance when the room temperature is
higher than the cuvettes temperature. It
makes no sense to have the system heating
the cuvettes at 37°C when the room
temperature is at 40°C.
Reagents tray: the reagent are normally
stored at controlled temperatures. The
ensure a longer life to the reagents on
board is useful to have the reagents chamber at a relatively low temperature. The choice of the
temperature depends on many factors: the room temperature, the time the reagents will remain on
board, etc… The lower the temperature is, the more will be the analyzer effort to keep it down. If the
reagents will last only a few days on board, there is no need to keep the temperature too low..
Standards tray: the standard and control tray is refrigerated to ensure a longer stability to the
solutions. In any case, once poured into the cups, the standards or controls cannot be re-introduced
in the original vials, to avoid contamination, therefore also here it is not advisable to keep the
temperature too low.
1.7. Administrator
The Administrator has the possibility to use or not the reagents Lots' control. When this option is
disabled, the analyzer will not perform any control on the reagents lot or expiry date. On the other
side, when this option is enabled, it is possible to enable also the following checks.
Check dates: this check will allow the analyzer to warn the user when
a reagent has reached its expiry date. An expired reagent will be
automatically removed from the reagents tray and will not be useable
anymore.
Check volumes: this is an additional check that will allow the analyzer
to verify if the reagent has been added to an existing bottle. The
analyzer will verify the volume increment to avoid bottles re-filling. If
the analyzer detects a volume increment higher than 15%, it will consider that bottle as
counterfeited and will remove it from the tray. In the Reagents Lot Management are listed all the
analyses with the Expiry date and volume.
2. Patients
In this menu are available those settings applied generally to the
patients and patients results.
2.2. Communication
The Communication page allows to set the parameters for the communication between the
analyzer and a host computer.
Once the communication is activated, all the
parameters will be displayed. These parameters
are different for the two type of Transport. This can
be Serial or TCP/IP.
Selecting TCP/IP it will be necessary to select the
Port.
The default communication port is 8888. It can be
set by the technical assistance personnel with their
password. If the communication port is changed, it
will be necessary to turn the external PC off and
then on to activate the new port.
The TCP/IP protocol version used is version 4 (not
version 6).
COM Port: select the appropriate communication port. (for example COM 1, COM 2 etc.)
Baud-Rate: Sets data transmission speed between the analyzer and
the host computer. Click u to view all the possible transmission speeds.
To select, click on the desired value.
Parity: None
Word Length: 8
Stop Bits: 1
Hand-Shake: Enables Data Flow Control to be used during data
transmission. Click u to view the available options. To select, click on
the desired option.
Protocol: it is possible to use the ASTM protocol or a variable protocol. When choosing the variable
protocol, it will be possible to set the protocol options by clicking on the
dedicated button Modify Scripts. Only expert operators should set the
variable protocol.
ONLY WITH ASTM PROTOCOL. In case the analyzer, after having read the
bar-code on the tray, does not find in its programmed work-lists the patient
corresponding to that code, it will require this information to the host computer.
PATIENTS
BY CODE: sorting is done by alphanumerical order of the ID code. The alphanumerical order means
that codes can be ordered as follows: CODE1, CODE10, CODE2, or as follows, depending on the
way numbers are written, CODE01, CODE02 AND CODE10
BY POSITION: patients are ordered by their position on the tray. In case of patients programmed in
the same position, the analyzer will insert them in the new order as it finds them in the Result page.
BY TIME: patients are ordered upon the execution date and time. Within every date, patients are
ordered by time.
ANALYSES
Within every patient, analyses can be sorted with their own criteria.
BY CODE: analyses are sorted by the alphanumerical order of the code. If codes are all
alphabetical, the codes will be ordered in alphabetical order, but if there are alphanumerical codes,
these will be ordered as for patients.
BY POSITION: analyses are ordered upon their position on the tray.
BY TIME: analyses are ordered upon the execution time.
SPECIAL ANALYSES: I.S.E. & Relation tests
Both I.S.E. and Relation tests can be ordered specifically. If set to NONE, the tests will be inserted
in the list in the order as the analyzer finds them in the Result page.
AT TOP: the special tests will be placed in the new order at the beginning of the list.
AT BOTTOM: the special tests will be placed in the new order at the end of the list.
As far as every Lab may have or use special solutions for different needs, these three bottles’
definitions have been left "open", to let the operator name them as necessary.
To assign a definition, simply click on the editable field and type the solution's name.
CHAPTER 10
TECHNICAL FUNCTIONS Page
TECHNICAL FUNCTIONS
In this chapter are first explained the analyzer utilities, the mechanical calibrations and then the
available diagnostic functions.
In the normal routine use, there are only a few adjustments that may become necessary. On the
other side, there are many utility functions that are commonly used. The diagnostic pages are
normally reserved to the technical assistance personnel.
1. ANALYZER UTILITIES
The analyzer utilities can be accessed by the main menu Analyzer or from the side bar, under the
option Special Functions.
Empty Fluidics: This command completely empties the fluidic circuit. It is to be used exclusively as
preliminary operation for maintenance or eventual moving of the instrument.
Zeroing on water: Performs photometer zeroing with distilled water. It is important to remember
that zeroing is necessary to avoid any drift of the photometer zero and has to be performed after the
analyzer has reached the steady state (approx 20 minutes after turning on).
F.C.C. calculation: This procedure is for calculating the optical correction factor for the reading
cuvettes. A screen message invites to insert a bottle with appropriate solution into the reagents tray.
Generally a solution of Potassium Bichromate with absorbance around 0,500 Abs (read at 340 -
700nm) is used. The analyzer guides the user through the procedure. This function is password
protected.
It is performed only when one or more cuvettes are replaced or after a thorough service is
performed on the instrument computer (e.g. hard disk substitution) where correction factors may be
lost.
Lamp Setup: this function is only necessary after the photometric lamp has been changed. It is
used to align the new lamp to preset internal values. It also carries out photometer zeroing at the
same time. It is advisable that only technical assistance personnel use this function.
Washing station prime:
Na Conditioning: The Na sensor requires reconditioning of its proper pH value. The
reconditioning procedure is simple and rapid and a specific solution "Na conditioning solution"
has to be used. When to perform: usually once a week or when one notices a sharp decline of
the Na Slope, beyond 6 mV. In practice the Na Slope descends a couple of mV per day, but it is
also dependent upon the number of tests executed. The procedure consists in placing a serum
cup with at least 500µl of Na conditioning solution in the serum plate in the position indicated on
the displayed message and then press start. This procedure will last for approximately 4
minutes (after the aspiration of the Na solution into the sensors, the analyzer will wait for approx
3 min) and ends with the release of the solution A into the sensors. Then it is suggested to run
a calibration (slope).
2. MECHANICAL CALIBRATIONS
This function allows mechanical centering and adjustments for the different positions of the
sampling arm, cuvettes plate, sample plate and reagents tray. Bear in mind that the analyzers are
already factory-calibrated and a new adjustment is seldom necessary.
The mechanical calibrations are divided into two
main sections: Arms and Trays
Each section, once opened gives access to the
specific and dedicated functions.
To perform calibrations, click on the desired
function: the Left and Right command keys will be
enabled on the screen, which determine the
movement of the selected object either clockwise or
counterclockwise, step by step. The [!] Test key
lowers and lifts the sampling needle or the wash
piston to check centering.
NOTE:
It is highly recommended to calibrate the trays first and then the positions of sampling arm.
ARMS
There are three sampling arms. Two for the reagents and one dedicated to the samples. Each arm
must be calibrated in all positions. Click on the line corresponding to each position, then use the
arrows buttons to adjust the needle, which has to be centered above the selected position. To test
the centering, use the Test button. The Height adjustment is normally performed in factory. Please
read carefully the dedicated paragraph before using this function.
Arm on sample
TRAYS
There are four trays. Each has to be positioned in order to center the needle. Follow the same
instructions as for the arms.
Samples tray
Click on every arm position: the needle
will move to that position to allow the
tray adjustment.
The samples tray must be adjusted in
order to have the needle centered on
the three samples rings. Then it has to
be set the position of the tray respect to
the bar-code readers. The external and
internal samples' tray rings are read
from the external bar-code reader,
while the internal samples ring is read
from the internal bar-code.
Reagents tray
Click on every reagent position: the
needle will move above that position to
allow the tray adjustment.
The reagents tray moves in order to
allow the needle centering on the
reagent bottles (first and second, large
and small). It has also to be centered to
the reagent insertion position and on
the bar-code reader position.
Cuvettes
NOTE: the position presented when adjusting the needle on the tray's rings is always the
first position for that ring, i.e. the position in that ring having the lowest number. For
instance, in the samples tray, the first position of the external ring is #1, of the central ring is
#37 and of the internal ring is #73.
Chapter 10 Technical Functions Page 5 of 7
BT4500 OPERATOR MANUAL
HEIGHT ADJUSTMENT: this button is used to memorize the maximum descent of the needle in the
selected position. For the reagents positions, including the I.S.E. sampling diluent, an empty bottle
must be in place. For samples positions a primary tube must be in place. For standards and controls
a cup must be placed on the tray. The needle will go fully down until the crash sensor is activated. In
this way the analyzer memorizes the maximum possible descent in every position.
3. DIAGNOSTICS
Diagnostic pages are accessible from Analyzer’s main menu. The
Diagnostic functions are partially reserved for approved service personnel
only. Therefore the access to some of the functions is password protected.
It contains the following options: View F.C.C., View Optical Transmission
and General Diagnostics.
Show F.C.C.: It shows the optical correction factor calculated for all the reading cuvettes. For each
cuvette, the mABS and the F.C.C. values are
reported. It also indicates the cuvettes with
correction factor exceeding the preset limits
(± 3%). It is possible to print this page.