BC-5180 Operational Manual PDF
BC-5180 Operational Manual PDF
Operator’s Manual
© 2008-2011 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issued Date is 2011-05.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
Mindray is strictly forbidden.
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only
if:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel.
the electrical installation of the relevant room complies with the applicable national and
local requirements.
I
It is important for the hospital or organization that employs this equipment
to carry out a reasonable service/maintenance plan. Neglect of this may
result in machine breakdown or injury of human health.
Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will
be unreliable, which would damage the analyzer components and cause
personal injury.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the
improper use or application of the product or the use of parts or accessories not approved by
Mindray or repairs by people other than Mindray authorized personnel.
Malfunction of the instrument or part whose serial number is not legible enough.
III
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
E-mail Address: service@mindray.com
Tel: 0049-40-2513175
Fax: 0049-40-255726
IV
Table of Contents
1
Table of Contents
2
Table of Contents
3
1 Using This Manual
1.1 Introduction
This chapter explains how to use your BC-5180 operator’s manual, which is shipped with
your BC-5180 AUTO HEMATOLOGY ANALYZER and contains reference information about
the BC-5180 and procedures for operating, troubleshooting and maintaining the analyzer.
Read this manual carefully before operating your analyzer and operate your analyzer strictly
as instructed in this manual.
1-1
Using This Manual
This manual contains information written for clinical laboratory professionals to:
1-2
Using This Manual
This operator’s manual comprises 12 chapters and 3 appendices. Refer to the table below to
find the information you need.
1-3
Using This Manual
all capital letters enclosed in [ ] indicate a key name on the analyzer or external keyboard,
such as [ENTER].
bold letters included in “ “ indicate text you can find on the screen, such as “Clean”.
bold letters indicate chapter titles, such as Chapter 1 Using This Manual.
All illustrations in this manual are provided as examples only. They may not necessarily
reflect your analyzer setup or data displayed.
1-4
Using This Manual
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
Please check the firmness of all the doors and covers before running the
analyzer.
Make sure all the safety measurements are adopted. Disable any safety
device or sensor is prohibited.
1-5
Using This Manual
Please install the original edition software to prevent the computer from
being infected by virus.
1-6
Using This Manual
1.6 Symbols
You will find the following symbols in this manual:
BIOLOGICAL RISK
HIGH VOLTAGE
1-7
Using This Manual
EARTH (GROUND)
ALTERNATING CURRENT
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
MANUFACTURER
TEMPERATURE LIMITATION
IRRITATING SUBSTANCE
1-8
Using This Manual
1-9
Using This Manual
(1)
To avoid electric shock, disconnect power cord prior to removing or replacing fuse.
1-10
Using This Manual
(1)
To avoid injury, do not put your hands around the guide channel of the syringe board.
(2)
The probe is sharp and may contain biohazardous material. Exercise caution when
working around the probe!
1-11
Using This Manual
(1)
To avoid injury, do not put your hands around the guide channel of the syringe board.
1-12
Using This Manual
(1)
1-13
2 Understanding Your Analyzer
2.1 Introduction
2-1
Understanding Your Analyzer
The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
The BC-5180 is a quantitative, automated hematology analyzer and 5-part differential counter
used in clinical laboratories.
It provides the following 23 basic parameters, 4 parameters for research use, 3 histograms
and 1 scattergram of blood samples. It supports 2 measurement modes: CBC and
CBC+DIFF.
Parameter Name Abbr. CBC CBC + DIFF
White Blood Cell count WBC * *
Neutrophils percentage Neu% / *
Lymphocytes percentage Lym% / *
Monocytes percentage Mon% / *
Eosinophils percentage Eos% / *
Basophils percentage Bas% / *
Neutrophils number Neu# / *
Lymphocytes number Lym# / *
Monocytes number Mon# / *
Eosinophils number Eos# / *
Basophils number Bas# / *
Abnormal Lymphocytes percentage ALY% (RUO) / *
Large Immature Cells percentage LIC% (RUO) / *
Abnormal Lymphocytes number ALY# (RUO) / *
Large Immature Cells number LIC# (RUO) / *
RBC RBC * *
Hemoglobin Concentration HGB * *
Mean Corpuscular Volume MCV * *
Mean Corpuscular Hemoglobin MCH * *
Mean Corpuscular Hemoglobin MCHC * *
Concentration
Red Blood Cell Distribution Width RDW-CV * *
Coefficient of Variation
Red Blood Cell Distribution Width RDW-SD * *
Standard Deviation
Hematocrit HCT * *
Platelet count PLT * *
2-2
Understanding Your Analyzer
“*” means the parameter is provided in the mode. “/” means the parameter is
not provided.
ALY%, LIC%, ALY# and LIC# are parameters for research use only, not for
diagnostic use. For more details about the RUOs, please see 5.3.1 General
Setup Research use only parameter (RUO parameter).
2-3
Understanding Your Analyzer
Please check the firmness of all the doors, covers and boards before
running the analyzer.
2-4
Understanding Your Analyzer
2-5
Understanding Your Analyzer
2-6
Understanding Your Analyzer
2-7
Understanding Your Analyzer
2-8
Understanding Your Analyzer
2-9
Understanding Your Analyzer
2-10
Understanding Your Analyzer
To avoid damage, do not turn on/off the power of the analyzer continually in
a short time.
2-11
Understanding Your Analyzer
The interface can be divided into several areas as follows according to their functions:
2. Analysis status
It indicates the current analysis status and displays in the same way as the Power/Status
indicator on the main unit.
Green icon: it means you can proceed to analyze the sample.
Flickering green icon: it means the analyzer is not ready for analyze yet.
Red icon: it means you can not proceed to analyze the sample but not due to an error
happened. (e.g. the analyzer is in sleep status)
Flickering red icon: it means you can not proceed to analyze the sample and it is due to an
error happened.
Yellow icon: it means you can proceed to analyze the sample under the emergency mode.
2-12
Understanding Your Analyzer
Flickering yellow icon: it means the analyzer is not ready for analyze yet under the
emergency mode.
4. Status area
The area is on the top right of the screen. There are three items from left to right, namely:
LIS/HIS status
Print status:
5. Minimize button
You can click the button to minimize the interface to the taskbar of the operation system.
Click the interface icon displayed on the taskbar, you can get the display of
the interface back after minimizing it.
7. System time
It displays the time of the operation system. When you move the mouse to this area, the tips
will pop up to display the current system time. See Chapter 5 Customizing the Analyzer
2-13
Understanding Your Analyzer
2-14
Understanding Your Analyzer
2-15
Understanding Your Analyzer
2.6.2 Click
Move the pointer to the desired content; left click the mouse then release.
Repeat the operation if failed to choose the content; check the connection
of the mouse if necessary. If the problem still exists, please contact Mindray
customer service department or your local distributor immediately.
Repeat the operation if failed to choose the content; check the connection
of the mouse if necessary. If the problem still exists, please contact Mindray
customer service department or your local distributor immediately.
Repeat the operation if failed to choose the content; check the connection
of the mouse if necessary. If the problem still exists, please contact Mindray
customer service department or your local distributor immediately.
2-16
Understanding Your Analyzer
Move the pointer to the slide bar, left click the mouse and hold, then scroll the bar at will.
2.6.7 Tab
Tab displays one page of the multi-page information. E.g. you can enter the “Settings” tab of
the “L-J” interface to view and set up the information. The “Settings” tab is shown below.
2-17
Understanding Your Analyzer
2.6.8 Buttons
Common buttons
The system will perform the function after you clicking the certain button. E.g. the system will
print after you clicking the “Print” button as shown below.
When a combo list is open, you can choose the desired item in the list by using the [↑]
2-18
Understanding Your Analyzer
When a combo list is open, you can hide it by pressing the [Enter] key on the keyboard
or clicking a certain option. Then, the original content in the combo box will be replaced
by the current chosen one.
When a combo list is open, you can hide it by pressing the [Esc] key on the keyboard.
Then, no change to the combo box.
The scroll bar will appear if the content of the list can not be displayed in
one sight of a screen. You can scroll or use the [PgUp] and [PgDn] keys on
the keyboard to view the information fully.
After clicking the arrow button on the date control, a date box will pop up.
Select the year: click the displayed year, arrows will appear on its right side. Then,
choose the desired year by clicking the arrow button.
Method 1: click the arrow button on the both sides of the date box to switch and choose the
desired month.
Method 2: click the displayed month, then click the desired month from the list appeared as
shown below.
2-19
Understanding Your Analyzer
Select the day: click the desired day, then the date box will hide. The chosen date
(including year, month and day) will replace the original date of the date control.
When the data box popping up, you can hide it by pressing the [Esc] on the keyboard.
Then, no change to the date control.
Radio button
Click the single choice button, a mark appears in the circle, indicating the option is chosen.
E.g. the following figure shows that the “Autoloader-Whole Blood” is chosen whereas the
“Closed-tube-Whole blood” and “Closed-tube-Prediluted Blood” is not chosen.
Only one single choice button can be chosen in one setting option.
Click the “built-in barcode scan” again, the “√” disappeared, it means the option is not
chosen as shown below:
2-20
Understanding Your Analyzer
More than one check box can be chosen at the same time in one setting
option.
You can also proceed to the following operations in the edit box:
Move the cursor to the left or right by using the [←] and [→] keys on the keyboard.
Move the cursor to the left of the initial character or the right or the end character by
pressing the [Home] and [End] key on the keyboard.
Delete the character on the right of the cursor by using the [Delete] on the keyboard.
Delete the character on the left of the cursor by using the [Backspace] key on the
keyboard.
Switch to other edit box by using the [Tab] key on the keyboard.
You don’t have to enter the separators in the date box and the IP box.
The scroll bar (horizontal/vertical) will appear if the content of the edit box
can not be displayed in one sight of a screen. You can scroll or use the
[PgUp] and [PgDn] keys on the keyboard to view the information fully.
2-21
Understanding Your Analyzer
See “Arrow button of the combo box” section for details to complete selecting. See “Edit
box” section for details to complete editing if the combo box is editable.
2.6.13 Form
The form contains several cells and check box (sometimes).
Click the certain cell, it is chosen as shown below:
2-22
Understanding Your Analyzer
Select the cell by using the [↑], [↓], [←] and [→] keys on the keyboard.
Select the initial or end cell of the current row by using the [Home] and [End] keys on the
keyboard.
Select the next cell of the current line by using the [Enter] key on the keyboard.
Select the next cell of the current row by using the [Tab] key on the keyboard(can not
switch to a new row).
Move the mouse to the boundary line between the rows or lists, then left click the mouse
and hold, drawing the line to change the height/width of the row/list, but the height/width
of the whole form remains.
For an editable cell, a cursor will appear in it if it is double clicked. You can enter the
characters from the location of the cursor and the cursor moves to the right at the time. A
edited form is shown below:
Move the cursor to the left or right in the cell by using the [←] and [→] keys on the
keyboard.
Move the cursor to the left of the initial character or the right or the end character by
pressing the [Home] and [End] key on the keyboard.
Delete the character on the right of the cursor by using the [Delete] on the keyboard.
2-23
Understanding Your Analyzer
Delete the character on the left of the cursor by using the [Backspace] key on the
keyboard.
Hide the cursor and quit editing by using the [Enter] key on the keyboard.
If a check box exists in the form, you can operate according to the
instructions of the check box before. The selected check boxes in the forms
will not be cleared when you switching among the screens unless you exit
the software.
If a check box exists in the form, then when you click the check box, it will
be ticked and the record will also be highlighted.
If you wish to choose several continuous records in a form, click the initial
record and a “√” mark appears, the click the end of the desired record while
pressing and holding the [Shift] key on the keyboard, then a set of records
are chosen conveniently.
If you wish to select several consecutive records, click the first record and
then hold and drag the mouse to the last one, and then release the mouse to
select them all.
The scroll bar (horizontal/vertical) will appear if the content of the form can
not be displayed in one sight of a screen. You can scroll or use the [PgUp]
and [PgDn] keys on the keyboard to browse the information of the form fully.
2-24
Understanding Your Analyzer
Method 2: move to the desired option by using the [↑] and [↓] keys on the keyboard and then
enter it by using the [Enter] key on the keyboard.
Method 1: click the first-level menu to open the submenu, then click the desired option on it.
Method 2: first, move to the submenu option on the first-level menu by using the [↑] and [↓]
keys on the keyboard and open it by using the [→]or [Enter] key, then move to the desired
option by the [↑] and [↓] keys, at last open the interface or message box by pressing the
[Enter] key.
A “+”mark indicates the followed menu option has a submenu. Click the menu option, its
submenu will be fully displayed. At the same time, the “+” mark changes to “-”mark(click
the menu option again, its submenu will hide.), and “-” changes to “+” again as shown
below:
The menu option without “+” or “-” mark has no submenu. Click the menu option, the
2-25
Understanding Your Analyzer
Use the [↑] and [↓] keys on the keyboard to move the highlight bar to the desired menu
option.
Use the [Home] and [End] keys on the keyboard to move the highlight bar to the initial or
ending option.
Use the [Enter] key on the keyboard to display the submenu of a menu option. If the
menu option has no submenu, the relative information will display directly.
If the submenu of a menu option has displayed, you can hide it by using the [Enter] key
on the keyboard.
The scroll bar (horizontal/vertical) will appear if the content of the directory
tree can not be displayed in one sight of a screen. You can scroll or use the
[PgUp] and [PgDn] keys on the keyboard to view the information fully.
After choosing the data you want to delete, click the “Ok” button to close the message
box and completing the deletion; click the “Cancel” button to close the message box
without any operation.
Click the button on the right side of the title area to close the message box without
any operation.
2-26
Understanding Your Analyzer
The “2/79” in the above figure indicates the total number of the records is 79, and the current
record is the second one.
If you wish to switch to the previous or the next record, click or button.
If you wish to switch to the first or the last record, click or button.
Click the edit box of the record switch column, enter the desired number of the record,
and then switch to the relative interface by using the [Enter] key on the keyboard.
2-27
Understanding Your Analyzer
The message box displays the help information of the current screen and the corresponding
menu (highlighted).
If you wish to browse other help information, click the desired menu option then the help
information will display on the right.
Click the button on the top right corner to close the message box.
2-28
Understanding Your Analyzer
Then, click the “Search” tab to enter the key words into the search information box.
2-29
Understanding Your Analyzer
After entering the key words, click the “Search” button or press the [Enter] key on the
keyboard to start searching.
You can stop searching by clicking the “Stop” button. Then, the obtained
result displays.
When the search is finished, all the related menu items will be displayed in the left corner and
the corresponding help information will be displayed at the right side screen.
You can click the menu item to check the corresponding help information.
2.7.3 Print
You can click the “Print” button to print the current displayed help information.
2-30
Understanding Your Analyzer
Store and use the reagents as instructed by instructions for use of the
reagents.
When you have changed the diluent, cleansers or lyses, run a background
to see if the results meet the requirement.
Pay attention to the expiration dates and open-container stability days of all
the reagents. Be sure not to use expired reagents.
After installing a new container of reagent, keep it still for a while before use.
2.8.1 Reagents
M-53D Diluent
It provides a stable environment for counting and sizing blood cells.
M-53LH Lyse
It breaks down red blood cell walls and converts hemoglobin to a hemoglobin complex to
determine the HGB. It 2-differentiates WBCs to Basophils and other WBCs, and determines
WBC amount.
M-53 Cleanser
2-31
Understanding Your Analyzer
It is a cleaning solution formulated to prime and clean the fluidic channels and tubing and get
rid of blood albumin and sediment.
2-32
3 Understanding the System
Principles
3.1 Introduction
The measurement methods used in this analyzer are: the Electrical Impedance method for
determining the WBC/BAS, RBC and PLT data; the colorimetric method for determining the
HGB; flow cytometry by laser for determining the WBC data. During each analysis cycle, the
sample is aspirated, diluted and mixed before the determination for each parameter is
performed.
3-1
Understanding the System Principles
3.2 Aspiration
According to different configurations, the analyzer offers two types of sampling mode:
autoloader mode or closed-tube mode, the closed-tube mode supports two types of blood
samples – whole blood samples and prediluted blood samples, and the autoloader mode
supports whole blood samples.
If you are to analyze a whole blood sample under closed-tube mode, the analyzer will
aspirate 16μL of the sample.
If you are to analyze a capillary blood sample under closed-tube mode, you should first
manually dilute the sample (20μL of capillary sample needs to be diluted by 180μL of diluent)
and then present the pre-diluted sample to the analyzer, which will aspirate 80μL(CBC+DIFF)
or 40μL(CBC) of the sample.
If you are to analyze a whole blood sample under autoloader mode, the analyzer will aspirate
16μL of the sample.
3-2
Understanding the System Principles
3.3 Dilution
Then, the sample will be divided into 2 portions and be diluted and processed by different
reagents. After this, they are ready for analysis.
This analyzer can process two types of blood samples – whole blood samples and prediluted
blood samples.
2.5 mL diluent
52.08μL
About 1:416.7 dilution
3-3
Understanding the System Principles
20 μL capillary blood
180 μL diluent
40μL
2.46 mL diluent
60μL
About 1:625 dilution
20 μL capillary blood
180 μL diluent
40μL
1.1 mL M-53LEO(I) lyse
3-4
Understanding the System Principles
After a predetermined volume of blood is aspirated and diluted by a certain amount of reagent,
it is injected into the flow cell. Surrounded with sheath fluid (diluent), the blood cells pass
through the center of the flow cell in a single column at a faster speed. When the blood cells
suspended in the diluent pass through the flow cell, they are exposed to a laser beam. The
intensity of scatter light reflects the blood cell size and intracellular density. The low-angle
scattered light reflects cell size, and the high-angle scattered light reflects intracellular density
(nucleus size and density). The optical detector receives this scatter light and converts it into
electrical pulses. Pulse data collected can be used to draw a 2-dimensional distribution
(scattergram). As shown in Figure 3-2 , X-axis represents the intracellular density and Y-axis
the blood cell size. Various types of analysis data can then be obtained from the
scattergrams.
3-5
Understanding the System Principles
By analyzing the DIFF channel scattergram, the analyzer presents the Lym%, Mon%, Eos%
and Neu%.
3-6
Understanding the System Principles
Each pulse is amplified and compared to the internal reference voltage channel, which only
accepts the pulses of a certain amplitude. If the pulse generated is above the WBC/BAS
lower threshold, it is counted as a WBC/BAS. The analyzer presents the WBC/BAS
histogram, whose x-coordinate represents the cell volume (fL) and y-coordinate represents
the number of the cells.
Basophils number
Bas# is the number of Basophils measured directly by counting the Basophils passing
through the aperture.
Basophils percentage
Bas#
Bas% 100%
WBC
Lymphocytes percentage
Neutrophils percentage
Monocytes percentage
3-7
Understanding the System Principles
Eosinophils percentage
Lymphocytes number
Neutrophils number
Monocytes number
Eosinophils number
3-8
Understanding the System Principles
3.5.2 HGB
The HGB is calculated per the following equation and expressed in g/L.
Blank Photocurrent
HGB(g/L) Constant Ln
Sample Photocurrent
3-9
Understanding the System Principles
Each pulse is amplified and compared to the internal reference voltage channel, which only
accepts the pulses of a certain amplitude. If the pulse generated is above the RBC/PLT lower
threshold, it is counted as a RBC/PLT. The analyzer presents the RBC/PLT histogram, whose
x-coordinate represents the cell volume(fL) and y-coordinate represents the number of the
cells.
3-10
Understanding the System Principles
3.6.2 RBC
RBC
RBC (1012/L) is the number of erythrocytes measured directly by counting the erythrocytes
passing through the aperture.
Based on the RBC histogram, this analyzer calculates the mean cell volume (MCV) and
expresses the result in fL.
This analyzer calculates the HCT (%), MCH (pg) and MCHC (g/L) as follows, where the RBC
is expressed in 1012/L, MCV in fL and HGB in g/L.
RBC MCV
HCT
10
HGB
MCH
RBC
HGB
MCHC 100
HCT
Based on the RBC histogram, this analyzer calculates the CV (Coefficient of Variation, %) of
the erythrocyte distribution width.
RDW-SD (RBC Distribution Width – Standard Deviation, fL) is obtained by calculating the
standard deviation of the red blood cell size distribution.
3.6.3 PLT
Platelet count
PLT (109/L) is measured directly by counting the platelets passing through the aperture.
Based on the PLT histogram, this analyzer calculates the mean platelet volume (MPV, fL).
Platelet distribution width (PDW) is the geometric standard deviation (GSD) of the platelet
size distribution. Each PDW result is derived from the platelet histogram data and is reported
as 10(GSD).
3-11
Understanding the System Principles
PCT
This analyzer calculates the PCT as follows and express it in %, where the PLT is expressed
in 109/L and the MPV in fL.
PLT MPV
PCT
10000
3-12
Understanding the System Principles
3.7 Wash
After each analysis cycle, each element of the analyzer is washed.
3-13
4 Installing Your Analyzer
4.1 Introduction
Your analyzer is tested before it is shipped from the factory. International symbols and special
handling instructions tell the carrier how to treat this electronic instrument. When you receive
your analyzer, carefully inspect the carton. If you see any signs of mishandling or damage,
contact Mindray customer service department or your local distributor immediately
4-1
Installing Your Analyzer
at least 100 cm on each side, which is the preferred access to perform service
procedures.
enough room on and below the countertop to accommodate the diluent and waste
containers.
Before turning on the analyzer, make sure the input voltage meets the
requirements.
Using pinboard may bring the electrical interference and the analysis results
may be unreliable. Please place the analyzer near the electrical outlet to
avoid using the plug-board.
Please use the original electrical wire shipped with the analyzer. Using other
electrical wire may damage the analyzer or cause unreliable analysis
results.
4-2
Installing Your Analyzer
The environment should be as free as possible from dust, mechanical vibrations, loud
noises, pollution and electrical interference.
Do not use this analyzer in close proximity to sources of strong electromagnetic radiation
(e.g. unshielded intentional RF sources), as these may interfere with the proper
operation.
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close.
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
4-3
Installing Your Analyzer
Please make sure the length of the diluent pipe and the waste pipe is not
longer than 1500mm; the length of the lyse pipe and the cleanser pipe is not
longer than 850mm.
4-4
Installing Your Analyzer
4-5
5 Customizing the Analyzer Software
5.1 Introduction
The BC-5180 is a flexible laboratory instrument that can be tailed to your work environment.
You can use the “Setup” program to customize the software options as introduced in this
chapter.
The analyzer divides the operators into two levels, common user and administrator. Note that
an administrator can access all the functions open to a common user. This chanter introduces
how to customize your analyzer respectively as a common user level and as an administrator.
5-1
Customizing the Analyzer Software
Date format
Date format can be set at this screen. Note that when the date format setting is changed, all
the displayed and printed date format will be affected, including the draw date, delivery date,
run date, entry date of the work list, reagent expiration date, quality control date, calibration
date, and etc.
At the “General Setup” screen, click the “Date Format” button to enter the “Date Format”
screen.
5-2
Customizing the Analyzer Software
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
5-3
Customizing the Analyzer Software
Reagents
Be sure to set the reagent expiration date before the first use of the analyzer
or after a new container of reagent is installed.
You can set the expiration date of the diluent, LEO (I) lyse, LEO (II) lyse, LH lyse and
cleanser at the “Reagent” screen.
At the “General Setup” screen, click the “Reagent” button to enter the “Reagent” screen.
If you wish to set the expiration date for the reagents, you can click the check box “Exp. Date”
to select it. This option is selected as default.
If the “Exp. Date” check box is not selected, then the “Reagent Expired” will
not be alarmed.
5-4
Customizing the Analyzer Software
After selecting the check box of “Exp. Date”, you can click the arrow button of the edit box to
set the expiration date of each reagent by using the date control.
The range of the expiration date is from the current system date to
2099-12-31.
You can not edit the open-container expiration date for it is calculated
automatically by the software and displayed in the form of text.
If the current system date exceeds the displayed expiration date or the
open-container expiration date whichever is earlier, then the “Reagent
Expired” will be alarmed.
When the reagent is expired, you can check the expiration date of the sealed
container and the open container here to determine which one led to the
expiration alarm.
When the reagent is expired, you can not run any samples.
If an external barcode scanner is connected, you can click the “Use Barcode Scanner”
check box to enable it. Scan the barcode with the external barcode scanner. If the scan is
successful, the expiration date of the reagent will be displayed in the corresponding box.
If you wish to set the lot No. for the reagents, you can click the check box “Lot No.” to select it.
This option is selected as default.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
5-5
Customizing the Analyzer Software
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Auxiliary
Entering the “Auxiliary” screen
At the “General Setup” screen, click the “Auxiliary” button to enter the “Auxiliary” screen.
If you have activated the reminder and selected the predilute mode, a message box will pop
up to ask for confirmation every time you try to analyze a sample in the predilute mode.
5-6
Customizing the Analyzer Software
To activate the reminder, click the “Ask for confirmation” radio button (default). To
deactivate the reminder, click the “Do not ask for confirmation” radio button.
Select “Auto Increase” (default) so that the sample ID can increase automatically; select
“Manual entry (by keyboard or bar scanner)” if you want to enter the sample ID manually.
Even in the “Auto increase” mode, you can still change the sample ID by
re-entering the desired number through keyboard or the bar-code scanner.
The new setting of the prefix will only be applied to the later sample IDs. It
will not affect the IDs run previously and those already entered in the
worklist.
If the prefix is entered, and the sample ID entry method is set as “Auto
increase”, then a revisable prefix will be displayed automatically in the
sample ID box every time when you entering/editing the information.
If the sample ID entry method is set as “Manual entry (by keyboard or bar
scanner)”, then no matter the prefix is set or not, the default sample ID of a
new record in the worklist will be empty.
Other settings
If you want to add a new record automatically after the previous one is entered and saved in
the worklist, you can select “Save and jump to next record”.
If you wish to jump to the next record once the current one is validated, you can select the
check box “Validate and jump to next record”. This option is selected as default.
If you wish to switch between different information fields by [Enter] key, you can select the
5-7
Customizing the Analyzer Software
check box “Switch between different information fields by [Enter] key”. The default
setting of this option is selected and also support switch by [Tab] key. Click the “Information
fields” button next to the option and the following message box will pop up.
There is a check box in front of each information field; the default setting is all selected. It
means information entries of all demographics are requested. You can click a check box or
some boxes to cancel the selected mark “√”. It means you can jump over the unselected
demographics by pressing [Enter] or [Tab] key, leaving them in blank. However, you can also
re-locate the cursor in the information field by the mouse to re-enter the information.
Click “Ok” to save the entered information and close the message box, then back to the
“Auxiliary” screen.
If you wish to delete the completed record from the worklist after each run, you can click the
check box “Automatically delete completed records from the worklist”, selecting it by a
“√” in the box. The default setting of this option is not selected.
If you wish to apply the current system date to the “Draw Time” and “Delivery Time” for new
added sample records, you can select the check box “Automatically generate the draw
date and delivery date”. The default setting of this option is not selected.
Any change made to the option “Automatically generate the draw date and
delivery date” will only be applied to later added sample records. Records
entered previously in the worklist will not be affected.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
5-8
Customizing the Analyzer Software
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Change password
5-9
Customizing the Analyzer Software
1. Highlight the current login user in the list, and then click the “Change password” button,
the following message box will pop up.
2. Enter the current login password in the edit box “Old password”, and then enter the new
password in the “New password” and “Confirm new password” box.
3. Finish entering; click “Ok”, then a message box will pop up.
4. Click “Ok” to close the message box and back to the previous screen.
Exit
Click the “Close” button to exit the “User and Password” screen.
5-10
Customizing the Analyzer Software
Lab Information
When you log in as a user of common level, click the “Menu” button, and then select the
“Setup”, and then select “Lab Info.” from the pop-up menu, and then the lab information box
will pop up. You can only browse the information.
Exit
5-11
Customizing the Analyzer Software
You can click the “Department”, “Deliverer”, “Diagnosis” and “Gender” button to check the
corresponding shortcut code.
Exit
5-12
Customizing the Analyzer Software
5.3 Administrator
Auxiliary
Entering the “Auxiliary” screen
At the “General Setup” screen, click the “Auxiliary” button to enter the “Auxiliary” screen.
5-13
Customizing the Analyzer Software
Authority setup
To allow common users to edit the ID of the sample run already in the review screen, you can
select the check box “Edit ID of sample run already”. The default setting of this option is not
selected.
If you wish to enable users of common level the authority of editing/restoring the sample
result, you can select the check box “Edit sample result”. This option is not selected as
default.
If you wish to enable the users of common level the authority of validating the sample result,
you can click the check box “Validate sample”. This option is not selected as default.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save the changes and close the setup screen.
Cancel
5-14
Customizing the Analyzer Software
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Para. Units
Some references/parameters used by the analyzer could in several different units; you can
select the desired unit.
At the “General Setup” screen, click the “Para. Units” button to enter the “Para. Units”
screen.
Parameters of the same group are displayed together, with the first parameter in black and
5-15
Customizing the Analyzer Software
Unit standard
Click the check box “Select unit system” to select the desired unit standard form the
following seven ones: Customized, China, International (default), USA, Canada, Netherlands
and Britain.
When selecting different unit standard, the corresponding unit list and unit
option will be displayed differently.
If “Customized” is selected, then you can modify the unit of each parameter.
If other option is selected except the “Customized”, then the unit of each
parameter can only be browsed.
When “Customized” is selected, click the desired parameter, and then click the unit options
provided on the right to select a new unit for the parameter.
For parameters in a same group, if the unit of any parameter changes, the
units of the rest parameters change accordingly.
The unit of MCH changes according to MCHC and HGB, the operator can not
modify it.
If the parameters units change, the format of the data displayed in the list
will change accordingly.
Default
When “Customized” is selected, click the “Default” button to have the default units
(International) of all parameters displayed in the corresponding cell.
Click the “Print” button to print all the parameters’ units in the current screen.
If you haven’t saved the settings when clicking the “Print” button, a message box will pop up.
5-16
Customizing the Analyzer Software
Click “Yes” to save the new settings and print them; click “No” to print the content of the
original settings without saving.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Ref. Range
The “Ref. Range” screen is where you view and set the high and low limits for your patients.
The analyzer flags any parameter value above (H) or below (L) these limits.
This analyzer divides patients into 5 demographic groups: General, Man, Woman, Child and
Neonate. You can also customize another 5 groups. The default setting is “General”. The
recommended limits are provided for your reference only. To avoid misleading parameter
flags, be sure to set the patient limits according to the characteristics of your local population.
At the “General Setup” screen, click the “Ref. Range” button to enter the “Ref. Range”
screen.
5-17
Customizing the Analyzer Software
Click the “Set ref. group” button, a message box will pop up.
5-18
Customizing the Analyzer Software
You can not modify the names and corresponding information of the five
fixed reference groups in the list.
You can modify the names, age range (including age value and unit) and
gender of the five customized reference groups.
Double click the “Ref. Group” cell of the five customized reference group to modify the name
of the group.
The group title entered can not use General, Man, Woman, Child and
Neonate. The group title can not be the same as existing ones.
Double click the “Age” cell of the customized reference group to modify the age; double click
the age unit cell to open a combo box with different age units for you to choose from: Year,
Month, Day, and Hour.
Double click the “Gender” cell of the customized reference group to open a combo box with
different options for you to choose from: Not defined, Male, Female, Empty.
5-19
Customizing the Analyzer Software
Click the check box of “Automatically match the customized ref. group according to age
and gender” to select it. The option is not selected as default.
When the customized ref. groups are used to match, the matching will be
performed from top down according to the customized ref. groups displayed
in the screen.
Click one of the reference groups to highlight it, and then click the “Set to be default ref.
group” button to set the group as the default group when entering the worklist.
Click one of the reference groups to highlight it, and then click the “Default” button to display
the default information including group title, age limit, age unit and gender in the
corresponding cell.
Click “Print” to print out the settings in accordance with the age and gender of the reference
group. If the previous settings are not saved when you click the “Print” button, a message
box will pop up.
Click “Yes” to save the new settings and print them; click “No” to print the content of the
original settings without saving.
Click the “Ok” button to save and refresh the settings and close the “Set ref. group” message
box.
1. Click the “Ref. Group” combo box, and then select the desired group from the options:
5-20
Customizing the Analyzer Software
2. Drag the scroll bar, and then click the “Upper limit” or “Lower limit” of the parameter you
want to customize.
When a reference group is selected, the upper and lower limit of the target
will change accordingly.
The default reference ranges of the five customized groups are the same as
that of the “General” group.
The change of the reference range will not affect the previous flagging
setup, but only affect the following analysis.
Default
Click the “Default” button to display the default reference range of the current reference
group in the corresponding form.
Click “Yes” to save the new settings and print them; click “No” to print the content of the
original settings without saving.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
5-21
Customizing the Analyzer Software
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Print
Entering the “Print” screen
At the “General Setup” screen, click the “Print” button to enter the “Print” screen.
Enter the print title in the “Title” box. The default title is “Hematology Analysis Report”.
5-22
Customizing the Analyzer Software
Click the “Paper type” box, select the desired paper type from the five types: A5 (default), A4,
continuous paper, B6 and B5.
If you want multiple copies of the same patient report to be printed, you can enter the desired
number (1 - 100) into the “Copies” field. The default number is 1.
Click the “Format” combo box to display the report format options for you to select from. The
report format in the combo box differs according to the selected paper type. The relationships
are shown in the following list:
Print Preview
5-23
Customizing the Analyzer Software
After editing the print setup, you should preview the report before printing
to make sure the setup is correct.
You can click the “Customize” button to enter the “PrintTemplate” screen, and then
customize the print template. See Chapter 12 Customizing the Print Template for details of
how to customize.
Autoprint
The analyzer can auto-print the report in the set format once the analysis result is obtained.
Click “On” to enable the autoprint function; click “Off” to disable the autoprint function. The
default setting is “Off”.
If “Autoprint” is “On”, then the “Autoprint after validation” check box will be activated.
Select the “Autoprint after validation” to autoprint the report only after the sample is
validated; otherwise, the report will be printed once the running is finished.
Print Flag
If the Flag information is needed in the printed report, you can select the “Print Flag” check
box .The default setting of this option is not selected.
If the default report template that you selected can display the flag
information, then the “Print Flag” box here will be available for you to select.
If the default report template that you selected can not display the flag
information, or the selected template is a customized one, then the “Print
Flag” box here will be unavailable for you to select.
If the suspect flags “?” are needed in the printed report, you should select the “Print suspect
flags “?”” check box. This option is selected as default.
If the reference range is needed in the printed report, you should select the “Print ref. range”
check box. This option is selected as default.
5-24
Customizing the Analyzer Software
If the ref. range flags (“H” or “L”) are needed in the printed report, you should select the “Print
ref. range flags” check box. This option is selected as default.
If “Print ref. range” is selected, then the “Print ref. range flags” option will
also be enabled and unavailable for you to edit. The ref. range and the ref.
range flags (“H” or “L”) will be printed in the report together.
If you want to de-select the “Print ref. range flags” option when “Print ref.
range” is selected (print the ref. range in the report only), please contact
Mindray customer service department or your local distributor.
If “Print ref. range” is not selected, then you can select whether to print the
ref. range flags (“H” or “L”) in the report at will.
If the result edited flags (“E” or “e”) are needed in the printed report, you should select the
“Print result edited flags” check box. This option is selected as default. For details of how to
edit result, please see Edit Result section in 7.2.3 or 7.3.3 Function of the Buttons
If the ambient temp. abnormal flags (“T”) are needed in the printed report, you should select
the “Print ambient temp. abnormal flags” check box. This option is selected as default.
If the test date of each QC point needs to be printed out when printing QC graphs, you can
select the “Print QC graph time” check box.
Click the “Default printer” check box to display the printers available to the current system,
and then you can select one type from them as the default printer to perform all the print
tasks.
For operating system of Windows Vista, the right of using network printer is
restricted by users' access level. To enable the network printer, you should
log on the system as administrator, and then right click the shortcut icon of
5-25
Customizing the Analyzer Software
If you change the default printer here, the default printer of the current
system will also change.
If you change the default printer in the operation system, then the printer
name in this check box will also change.
Apply
Click the “Apply” button to save all the changes without closing the setting screen.
Ok
Click the “Ok” button to save all the settings and close the setting screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Communication
Entering the “Communication” screen
At the “General Setup” screen, click the “Communication” button to enter the
“Communication” screen.
5-26
Customizing the Analyzer Software
The settings here are applied to the communication between the analyzer
and the external (i.e. LIS), but not between the analyzer and the terminal
software.
Setting IP address
Enter the IP address into the empty (default) “IP address” box.
Setting Port
Enter the port number into the empty (default) “Port” box.
Auto-communication
The function is used for automatically transmitting the sample result to the external data
management software or LIS/HIS system.
Click “On” to enable the auto-communication; click “Off” to disable the auto-communication.
The default setting is “Off”.
5-27
Customizing the Analyzer Software
The function is used for automatically obtaining the sample/patient information from the
LIS/HIS system after the sample ID is entered or scanned, and automatically transmitting the
sample result to the LIS/HIS system.
Click “On” to enable the bidirectional LIS/HIS communication; click “Off” to disable it. The
default setting is “Off”.
Communication Acknowledgement
Select “On”: when IPU software is communicating with LIS/HIS, the HL7 protocol must be
strictly followed. After receiving the ACK acknowledgement from LIS/HIS, the communication
can be deemed successful; otherwise the communication fails, a communication failure
notice will pop up.
Select “Off”: when IPU software is communicating with LIS/HIS, even if no ACK
acknowledgement from LIS/HIS is received, the communication can be deemed successful;
no failure notice will pop up. Communication flag “T” is recorded in the communication
column in Table Review screen.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
5-28
Customizing the Analyzer Software
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
The RUO parameters are for research use only, not for diagnostic use.
At the “General Setup” screen, click the “RUO” button to enter the “RUO” screen.
5-29
Customizing the Analyzer Software
Setting display
If you wish to display the RUO parameters, select the “Display RUO parameters” check box.
This option is selected as default.
If “Display RUO parameters” is selected and you also wish to display the “*” mark, you can
select the “Display “*” mark” check box. This option is selected as default.
If “Display RUO parameters” and “Display “*” mark” are selected and you also wish to
display declaration (“*” means “research use only, not for diagnostic use”), you can select the
“Display declaration” check box. This option is selected as default.
Setting print
If you wish to print the RUO parameters, you can select the “Print RUO parameters” check
box. This option is selected as default.
If “Print RUO parameters” is selected and you also wish to print the “*” mark, you can select
the “Print “*” mark” check box. This option is selected as default.
If “Print RUO parameters” and “Print “*” mark” are selected and you also wish to print
declaration (“*” means “research use only, not for diagnostic use”), you can select the “Print
declaration” check box. This option is selected as default.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
5-30
Customizing the Analyzer Software
new changes will only be saved after you click the “Apply” or “Ok” button.
Gain
You can adjust each digital pot at the “Gain” screen. It is not recommended to adjust gains
frequently.
At the “General Setup” screen, click the “Gain” button to enter the “Gain” screen.
5-31
Customizing the Analyzer Software
Click the current value of the “RBC” and enter the new value.
You can adjust the HGB blank voltage by adjusting the HGB gain.
You can enter the value directly in the edit box or click the adjusting button to adjust the HGB
gain.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Auto Maintenance
Entering the “Auto Maintenance” screen
Click the “Auto Maintenance” button to enter the “Auto Maintenance” screen.
5-32
Customizing the Analyzer Software
You can set here when to start the auto-maintain procedure after the relevant fluidic operation
stops. Enter the desired time ranging from 15 to 120 minutes into the “Wait” box.
You can enter the desired time in the “Preset time” or click the adjusting button to set the
preset time for time-based maintenance ranging from [0:00(default) – 23:59].
You can enter the desired time in the “Pre-reminding” or click the adjusting button to set the
pre-reminding time for time-based maintenance. The setting range is [0 minute, 2 hours], the
default time is 1 hour.
You can enter the desired time in the "Soak for" box to set the duration of the time-based
maintenance. The range for the entry is [2 hours, 72 hours], and the default time is 4 hours.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the settings and close the setup screen.
5-33
Customizing the Analyzer Software
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Autoloader
When the autoloader is adopted, you can set the conditions when to stop the autoloader, the
rules to arrange the sample ID and whether to display statistics.
At the “General Setup” screen, click the “Autoloader” button to enter the “Autoloader”
screen.
5-34
Customizing the Analyzer Software
To stop the autoloader when there are no diluent and lyses, you can select the “There are no
diluent and lyses” check box. The default setting of this option is not selected.
To stop the autoloader when waste is full, you can select the “Waste is full” check box. The
default setting of this option is not selected.
To stop the autoloader when there is the sample ID reading error, you can select the “Sample
ID reading error” check box. The default setting of this option is not selected.
To stop the autoloader when there are n (editable; default is 3; upper limit is 100) continuous
clog errors, you can select the “There are n continuous clogging” check box. The default
setting of this option is not selected.
If the condition when to stop the autoloader are not set but the certain error
happens (“sample ID reading error” excluded), the analyzer will only stop
autoloading but the sample analysis will continue.
5-35
Customizing the Analyzer Software
You can click the “Increases sequentially according to tube position” radio button to have
the sample ID increased sequentially even an empty tube position is detected. You can click
the “Increases according to actual tube number” radio button to have the sample ID
increased according to actual tube number (i.e. the empty tube position will be excluded).
The default setting of this option is “increases sequentially according to tube position”.
Statistics
Click “On” to enable the statistics after the autoloading is finished every time. Click “Off” to
disable the function (default).
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the changes and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Barcode
You can set the symbology for the analyzer.
Totally 6 code systems are supported by the analyzer, namely, CODE39, CODE93,
CODEBAR, CODE128, UPC/EAN and ITF (chiasmatic 25 code).
At the “General Setup” screen, click the “Barcode Info.” button to enter the “Barcode Info.”
screen.
5-36
Customizing the Analyzer Software
If any of the 4 code systems (CODE39, CODE93, CODEBAR and CODE128) are selected,
you can set to remove the restriction on the number of digits.
Select the “Unlimited number of digits” check box, and the “Digits” edit boxes of the
selected code systems turn gray; you are not able to edit the number of digits, and barcodes
of any length of the selected code systems can be scanned.
Select the check box of the desired code system and enter the digits of the code into the
5-37
Customizing the Analyzer Software
“Digits” box. For CODE39, CODE93, CODEBAR and CODE128, the digits range is 1-20. For
ITF, the digits range is the evens between 1 to 20.For UPC/EAN, the digits range is controlled
by the analyzer automatically.
For the code system and digits, the settings should comply with the actual
one.
Apply
Click the “Apply” button to save all the changes without closing the setup screen.
Ok
Click the “Ok” button to save all the changes and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
Microscopic Para.
You can set the microscopic parameters for the display of the "Microscopic Exam. And
Others" tab at the "Review" screen and the printout template for reports.
At the “General Setup” screen, click the “Microscopic Para.” button to enter the
“Microscopic Para.” screen.
5-38
Customizing the Analyzer Software
2. Enter the name of the new microscopic parameter in the edit box.
3. Click “Ok” to save the information of the new parameter, refresh the para. list without
closing the message box, and the parameter name entered in the message box will be
5-39
Customizing the Analyzer Software
cleared, and then you can continue to add another new microscopic parameter.
New parameter names can be neither empty nor same as existing ones.
1. Click the desired parameter, and then click the “Edit” button. A message box will pop up.
2. You can enter a new name for the microscopic parameter in the edit box.
3. Click “Ok” button to save the modified parameter name and close the message box, and
then the edited parameter will be highlighted in the list.
Modified parameter names can be neither empty nor same as existing ones.
1. Click the desired cell, and then click the “Delete” button. A message box will pop up.
5-40
Customizing the Analyzer Software
2. Click “Ok” to delete the microscopic parameter and close the message box, and then the
parameter will be deleted from the list.
1. Click the “Adjust Order” button and a message box will pop up, displaying the current
order of the microscopic parameters in the para. list.
The buttons ("Top", "Up", "Down" and "Bottom") to the right of the list are
used to adjust the order of the microscopic parameters.
2. Click on a microscopic parameter to highlight it. Adjust the position of this parameter
using the buttons on the right.
Click the "Top" button to move the highlighted microscopic parameter to the top of the
5-41
Customizing the Analyzer Software
list.
Click the "Up" button to move the highlighted microscopic parameter upward by one
position.
Click the "Down" button to move the highlighted microscopic parameter downward by
one position.
Click the "Bottom" button to move the highlighted microscopic parameter to the bottom
of the list.
3. Click the “Ok” button to save the adjusted order, close the message box and go back to
the setup screen. Then the order of parameters will be refreshed.
Ok
Click the “Ok” button to save all the changes and close the setup screen.
Cancel
Click the “Cancel” button to close the setup screen without saving the changes.
You will not lose the new changes by switching to another screen. But the
new changes will only be saved after you click the “Apply” or “Ok” button.
5-42
Customizing the Analyzer Software
Creating a user
2. Enter the information in each edit box, and then click the authority combo box to select
5-43
Customizing the Analyzer Software
3. Click “Ok” to save the information of the new user, refresh the user list without closing the
message box, and all the information fields in the message box will be cleared, and then
you can continue to add another new user.
New user names can be neither empty nor same as existing ones.
1. Click the desired cell, then click the “Edit” button, then a message box will pop up.
2. You can change the content of each item in the edit box or change the users’ authority
level by clicking the authority combo box.
3. Click “Ok” button to save the modified users’ information and close the message box,
then the edited record will be highlighted in the list of users.
Modified user names can be neither empty nor same as existing ones.
If the current login administrator changes the user level into “common
user”, it takes effect only after logging out and then re-login.
Deleting a user
5-44
Customizing the Analyzer Software
1. Click the desired cell, then click the “Delete” button, a message box will pop up.
2. Click “Ok” to delete the user and close the message box, then the user will be deleted
from the list of users.
Reset password
You can reset a new password for users who forgot the password to log into the system:
1. Click the desired cell, and then click the “Reset password” button, a message box will
pop up.
2. Enter the new password in the “New password” and “Confirm new password” box.
3. Click “Ok” button to save the new password and close the message box.
You can not reset the password for the current login user.
Change password
5-45
Customizing the Analyzer Software
2. Enter the current login password in the edit box “Old password”, and then enter the new
password in the “New password” and “Confirm password” box.
3. Finish entering; click “Ok”, a message box will pop up.
4. Click “Ok” to close the message box and back to the previous screen.
Click the “Close” button to exit the message box of “User and Password”.
Lab Information
When you login as a user of administrator level, click the “Menu” button, and then select
“Setup”, and then select “Lab Info.” from the pop-up menu, and then the lab information box
will pop up. All the information fields in the box are activated for you to enter/edit.
5-46
Customizing the Analyzer Software
Enter the contact information (telephone number or E-mail) into the “Contact information”
box.
Entering postalcode
5-47
Customizing the Analyzer Software
Enter the installation date into the “Installation date” box. The installation date must be
entered and it can not be later than the current system date.
Enter the contact information of service department (telephone number or E-mail) into the
“Contact information of service department” box.
Entering remark
Ok
Click the “Ok” button to save the entered/edited information and close the lab information
box.
Cancel
Click the “Cancel” button to close the box without saving the changes.
You can set the shortcut code for the following items: “Department”, “Deliverer”, “Gender”
and “Diagnosis”.
The shortcut code is used to facilitate the entry of the foregoing items. You can enter the
shortcut code and press the [Enter] key instead of entering the whole item.
Department
When you log in as a user of common level, click the “Menu” button, and then select the
“Shortcut Code” to enter the shortcut code screen.
5-48
Customizing the Analyzer Software
Adding department
New added department name must be entered and it can not be the same as
existing ones.
5-49
Customizing the Analyzer Software
Editing department
1. Click the desired form cell and click the “Edit” button, and then the box shown below will
pop up.
3. Click “Ok” to save the information and close the message box, and then the edited
record will be highlighted in the list of department.
New added department name must be entered and it can not be the same as
existing ones.
The shortcut code is not necessary to be entered, but once you set them,
each of the code must be unique.
Deleting department
2. Click “Ok” to delete the department and close the message box, and then it will also be
deleted from the list of department.
5-50
Customizing the Analyzer Software
Exiting
Click the “Close” button to exit the “Shortcut Code” message box.
Deliverer
At the “Shortcut Code” message box, click the “Deliverer” button to enter its shortcut code
settings.
Add, edit and delete the deliverer as instructed in the Department section.
Diagnosis
At the “Shortcut Code” message box, click the “Diagnosis” button to enter its shortcut code
settings.
5-51
Customizing the Analyzer Software
Add, edit and delete the clinical diagnosis as instructed in the Department section.
Gender
At the “Shortcut Code” message box, click the “Gender” button to enter its shortcut code
settings.
Add, edit and delete the Gender as instructed in the “Department” section.
The existed shortcut code settings for gender can not be modified.
5-52
6 Operating Your Analyzer
6.1 Introduction
This chapter provides step-by-step procedures for operating your analyzer on a daily basis.
A flow chart indicating the common daily operating process is presented below.
6-1
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
You should only use the Mindray-specified reagents. Store and use the
reagents as instructed by instructions for use of the reagents.
Check if the reagents are connected correctly before using the analyzer.
After installing a new container of reagent, keep it still for a while before use.
Check and make sure the reagents and waste tubing are properly connected and not bent.
Check and make sure the power cord of the analyzer is properly plugged into the power
outlet.
6-2
Operating Your Analyzer
Check and make sure enough printer paper is installed. Check and make sure the power cord
of the printer is properly plugged into power outlet. Check and make sure the printer is
properly connected to the external computer.
Check and make sure the network cable of the external computer is connected to the
analyzer properly.
Check and make sure the keyboard and the mouse are well connected to the external
computer.
6-3
Operating Your Analyzer
1. Place the power switch at the left side of the analyzer in the ON position (I). The power
indicator light will be on.
3. After entering the operation system, double click the “BC-5180 Auto Hematology
Analyzer” icon to run the software.
4. After starting the software, the message box will pop up.
5. Enter the correct user name and password in the “Login” message box.
6-4
Operating Your Analyzer
Before running the software, make sure the network cable of the external
computer is connected to the analyzer properly. The analyzer starts to
initialize only when the connection are detected.
After startup, please make sure the data/time of the computer is correct.
The initial user name and password of administrator are “ADMIN”, which
was set by service engineer.
1 to 12 numeric digits can be entered for the user name and the password.
No Chinese character is allowed. There could be no password.
7. During the Initialization, the startup information will be displayed in the operation/status
information area at the bottom of the interface.
8. The whole process lasts 4 to 12 minutes. The time needed for initializing the system
depends on how the analyzer was shut down previously.
9. After the initialization process, you can enter the “Graph” screen to check the
background result.
6-5
Operating Your Analyzer
10. After initialization, if the sample records that are not hidden in the worklist are detected, a
message box will pop up.
Click “Yes” to set the first record not hidden in the worklist as the next sample ready to be run.
Click “No” to hide all the records in the worklist.
If the background results exceed the Ref. Range for the first time during
fluidics initialization, then the analyzer will run the background test one
more time.
If error happens during initialization (e.g. the background results exceed the
Ref. Range), the analyzer will alarm. See Chapter 11 Troubleshooting Your
Analyzer for solutions.
For the background Ref. Range of each parameter, please see Appendix B
Specifications.
You can click “Logout” button to switch to another user. Enter the new user
name and password in the log-in message box, then click “Ok” to re-login
as a new user.
During the starting procedure, to startup the analyzer first or to run the
software first are both acceptable.
6-6
Operating Your Analyzer
6-7
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
If you click the “Diluent” button under the “AL-WB” mode, the analyzer will
first switch to the “CT-WB” mode and then dispense the diluent.
1. Use clean K2EDTA (1.5 - 2.2mg/mL) anticoagulant collection tubes to collect venous
blood samples.
6-8
Operating Your Analyzer
For the whole blood samples to be used for WBC differential or PLT count,
you shall store them at the room temperature and run them within 8 hours
after collection.
If you do not need the PLT, MCV and WBC differential results, you can store
the samples in a refrigerator (2℃ - 8℃) for 24 hours. You need to warm the
refrigerated samples at room temperature for at least 30 minutes before
running them.
Be sure to mix any sample that has been prepared for a while before running
it.
Be sure to use the Ф11×40 (mm) adapter when collecting and handling the
prediluted samples. For details of how to install the adapter, please see 6.6.4
Replacing the adapter.
1. At the shortcut button area, click the “Diluent” button, then the compartment door will
open automatically and a message box will pop up.
2. Present a clean centrifugal tube into the sample compartment with the cap opened.
Press the [RUN] key to dispense 180μL of diluent into the tube. During dispensing the
6-9
Operating Your Analyzer
If the current mode is autoloader, the analyzer will switch to the close-tube
mode automatically and then dispense the diluent. Then, the mode will be
switched back to autoloader when exiting.
3. When the dispensing is finished, the compartment door will open automatically and you
can remove the centrifugal tube. Then, the following message box will display.
4. Add 20μL of capillary blood to the diluent, close the tube cap and shake the tube to mix
the sample.
5. After the prediluted sample is prepared, click the “Cancel” button to exit dispensing the
diluent.
6-10
Operating Your Analyzer
You can also dispense 180μL of diluent by pipette into the tube.
After mixing the capillary sample with the diluent, be sure to wait 3 minutes
before running the sample.
Be sure to run the prediluted samples within 30 minutes after the mixing.
Be sure to mix any sample that has been prepared for a while before running
it.
6-11
Operating Your Analyzer
If the analyzer is shut down abnormally, you will lose the worklist
information of the samples that have not been saved yet.
If you want to complete the worklist information after the analysis, see
Chapter 7 Reviewing Sample Results for details.
Click the “Worklist” button on the shortcut area or click “Menu”, then select “Worklist” to
enter the “Worklist” screen.
6-12
Operating Your Analyzer
Click the “New” button, then a new record will be added at the bottom of the worklist and this
blank record is highlighted. All the fields in the information entry area are displayed in defaults
and are activated.
You can switch between options in the Sample Info./Patients Info area by the
[Tab] key. You can also use the [Enter] key to switch after setting, see details
in General Setup section of Chapter 5 Customizing the Analyzer Software.
You can skip the options that do not need the entry when switching by [Tab]
or [Enter], see the setup details in General Setup section of Chapter 5
Customizing the Analyzer Software.
The sample ID must be entered and its acceptable length is [1, 20].
Select the loading mode as “closed -tube”, sample mode (“Whole Blood” or “Predilute”)
and the measurement mode (“CBC” or “CBC+DIFF”) from the three pull-down lists
respectively.
In the “CBC” measurement mode, the analyzer only counts the blood cells
without further differentiating the white blood cells. 13 parameters and
histograms of WBC, RBC and PLT are provided in this mode. In the
“CBC+DIFF” mode, the analyzer counts the blood cells and further
differentiates the white blood cells into 5 sub-populations. 23 basic
parameters, 4 RUO parameters, scattergrams and histograms of
WBC/BASO, RBC and PLT are provided in this mode.
6-13
Operating Your Analyzer
Select the reference group for the sample from the “Ref. Group” pull-down list. The analyzer
will judge the test results according to the reference range of the Ref. group. When the results
exceed the reference range, the analyzer will flag.
If you have entered the gender and age of the patient, then the system will
provide a matching Ref. Group automatically.
If the auto-matching Ref. Group is different from the one that you selected
before (excluding the 5 customized Ref. Groups), then the system will adopt
the auto-matching Ref. Group.
Select the draw date from the date control; enter the draw time into the time edit box.
Select the delivery date from the date control and then enter the delivery time into the time
edit box.
The delivery date/time can not be earlier than the draw date/time.
The draw and delivery date/time can not be later than the current system
date/time.
In the Uni-directional LIS/HIS mode, after you entering the patient ID and
pressing the [Enter] key, the matched patient information (including “Last
Name”, “First Name”, “Gender”, “Age”, “Birthday”, “Department” and “Bed
No.”) will be displayed in the screen automatically if there is any. You can
also proceed to edit the information.
In the Bi-directional LIS/HIS mode, the patient information will adopt those
downloaded from the LIS/HIS as default.
6-14
Operating Your Analyzer
Enter the patient name into the “Last Name” and “First Name” boxes.
Enter the gender of the patient into the “Gender” box or select it from the “Gender” pull-down
list.
The analyzer provides four ways for you to enter the patient age – in years, in months, in
days and in hours. The first way is designed for the adult or pediatric patients older than one
year; the second for the infant patients one month to one year; the third for the neonatal
patients no older than one month and the fourth for the neonatal no older than 24 hours. You
can choose only one of the four ways to enter the patient age.
The “Age” pull-down list provides four ways for you to enter the patient age– in years, in
months, in days and in hours, and you can enter the patient age in the box followed by the
age unit.
After entering the birthday, the age field will calculate automatically
according to the difference between the current system date and the
“Birthday”, and then a new result of age and the corresponding unit will be
displayed in the age edit box and the unit combo box respectively. Then, the
age box will be unavailable to edit unless the “Birthday” is cleared.
If the entered birthday is later than the current system, then it is considered
invalid.
Enter the name of the department, from which the sample came, into the “Department” box
or select it from the “Department” pull-down list.
Enter the bed No. of the patient into the “Bed No.” box.
Enter the name of the deliverer into the “Deliverer” box or select it from the “Deliverer”
6-15
Operating Your Analyzer
pull-down list (if there are previously saved deliverers’ names in the list).
Enter the suspect information of diagnosis into the “Clinical Diagnosis” box.
Save
When finish entering the work list information, you can click the “Save” button or the shortcut
key [F2] to save all the information.
The “Sample ID + Mode” of the current record can not be the same as the
unhidden records in the following status: “To Be Run”, “Running” and
“Error”.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
6-16
Operating Your Analyzer
Repeat piercing the evacuated blood collection tube may break the rubber
tube cap. The fragments produced may lead to incorrect analysis result. It is
recommended that do not pierce each tube for more than three times.
If the sample mode is switched from the “Predilute” to “Whole Blood”, the
analyzer will perform the switching sequence automatically and a progress
bar will be displayed on the screen.
When running as per the worklist, then the next sample ID will always be the
first unhide (or error) sample to be run in the worklist (in closed-tube mode)
till there is no unhide sample left or the worklist is empty; the ID of the latter
sample will increase by 1 automatically. If you set the method of entry for the
sample ID as “Auto Increase”, the ID of the latter sample will increase by 1
automatically.
If you do not run as per the worklist, then no actions will be taken to the
records in the worklist.
1. When it is ready to run samples (i.e. the analysis status icon and analyzer indicator is
green), shake the whole blood sample as shown below to well mix it.
6-17
Operating Your Analyzer
2. Press the [OPEN] key to open the compartment door, and then present the whole blood
sample into the compartment.
4. Click the shortcut button “Run”, the “Run” message box will pop up.
5. Click the “CT-WB” button, and then select the “Run as per the worklist” option.
6. After finishing setting, press [Enter] from keyboard or click the "Run" or “Ok” from the
dialog box, and then press the [Aspirate] key to start the analysis.
7. The sample probe will automatically aspirate the sample. Then, the analyzer will
automatically run the sample while the compartment door will open automatically and the
6-18
Operating Your Analyzer
8. When the analysis is finished, the analysis status icon and analyzer indicator will return
to lasting green.
The sample ID and the measurement mode are determined by the first
unhidden (or error) sample to be run in CT-WB mode in the worklist, and it is
displayed in gray.
If there is no record of CT-WB mode in the worklist, then the “Run as per the
worklist” button in the “Run” message box will be displayed in gray.
If you switch to the review screen from other screens, the latest record
information together with its result and graph will be refreshed and then
displayed.
1. When it is ready to run samples (i.e. the analysis status icon and analyzer indicator is
green), shake the whole blood sample as shown below to well mix it.
2. Press the [OPEN] key to open the compartment door, and then present the whole blood
sample into the compartment.
4. Click the shortcut button “Run”, the “Run” message box will pop up.
6-19
Operating Your Analyzer
5. Click the “CT-WB” button but do not select the “Run as per the worklist” option.
6. You can enter the starting sample ID of the sample to be run and the analysis mode in
the “Run” message box.
7. Click the "Run" button or press the [Aspirate] key after clicking “Ok” to start the analysis.
8. The sample probe will automatically aspirate the sample. Then, the analyzer will
automatically run the sample while the compartment door will open automatically and the
analysis status icon and analyzer indicator is flickering in green.
9. When the analysis is finished, the analysis status icon and analyzer indicator will return
to lasting green.
If you switch to the review screen from other screens, the latest record
information together with its result and graph will be refreshed and then
displayed.
6-20
Operating Your Analyzer
When running sample under the CT-PD mode, be sure to use the Ф11×40
(mm) adapter. For details of how to install the adapter, please see 6.6.4
Replacing the adapter.
1. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green),
press the [OPEN] key to open the compartment door, then present the prediluted sample
(uncapped) into the compartment.
3. Click the shortcut button “Run”, the “Run” message box will pop up.
4. Click the “CT-PD” button and select the “Run as per the worklist” option.
5. After finishing setting, press [Enter] from keyboard or click the "Run" or “Ok” from the
dialog box, and then press the [Aspirate] key to start the analysis.
6-21
Operating Your Analyzer
You can disable the pop-up message box before the predilute run, see
General Setup section in chapter 5 Customizing the Analyzer Software for
details.
7. The sample probe will automatically aspirate the sample. Then, the analyzer will
automatically run the sample while the compartment door will open automatically and the
analysis status icon and analyzer indicator is flickering in green.
8. When the analysis is finished, the analysis status icon and analyzer indicator will return to
lasting green.
The sample ID and the measurement mode are determined by the first
unhidden (or error) sample to run in CT-PD mode in the worklist, and it is
displayed in gray.
If there is no record of CT-PD mode in the worklist, then the “Run as per the
worklist” button in the “Run” screen will be displayed in gray.
If you switch to the review screen from other screens, the latest record
information together with its result and graph will be refreshed and then
displayed.
1. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green),
6-22
Operating Your Analyzer
press the [OPEN] key to open the compartment door, then present the prediluted sample
(uncapped) into the compartment.
3. Click the shortcut button “Run”, the “Run” message box will pop up.
4. Click the “CT-PD” button but do not select the “Run as per the worklist” option.
5. You can enter the starting sample ID of the sample to be run and the analysis mode in the
“Run” message box.
6. Click the "Run" button or press the [Aspirate] key after clicking “Ok”, then a message box
will pop up.
6-23
Operating Your Analyzer
You can disable the pop-up message box before the predilute run, see
General Setup section in chapter 5 Customizing the Analyzer Software for
details.
8. The sample probe will automatically aspirate the sample. Then, the analyzer will
automatically run the sample while the compartment door will open automatically and the
analysis status icon and analyzer indicator is flickering in green.
9. When the analysis is finished, the analysis status icon and analyzer indicator will return to
lasting green.
When the analyzer is running, you can perform any operation (including
new, edit and cancel, etc.) to other “To Be Run” or “Error” samples in the
work list.
When the analyzer is running, you can switch to Graph/Table Review Screen
to perform operations including data browsing, validating, sample
information editing and printing, etc., and you can also switch to other
screens.
When the analyzer is running, all the functions related to the fluidics
sequence are not available.
If you switch to the review screen from other screens, the latest record
information together with its result and graph will be refreshed and then
displayed.
6-24
Operating Your Analyzer
Parameter flags
If parameter follows a “H” or “L”, it means the analysis result has exceeded the upper or
lower limit of the reference range but still within the display range.
If you see *** as opposed to the result, it means the result is either invalid or out of the
display range.
For the background test, the flags of parameter or flags of abnormal blood
cell differential or morphology are not available.
Left Shift?
6-25
Operating Your Analyzer
Macrocytosis
Microcytosis
Dimorphologic
Anemia
Hypochromia
The analyzer will flag abnormal or suspect WBC, RBC and PLT according to the scattergrams
and histograms. The following table shows how the flags affect parameter results:
Left Shift? × √ × ×
Abnormal/Atypical Lymphocyte? × √ × ×
Leucocytosis √ √ √ √
Leucopenia √ √ √ √
Neutrophilia × √ × ×
Neutropenia × √ × ×
Lymphocytosis × √ × ×
Lymphopenia × √ × ×
6-26
Operating Your Analyzer
Monocytosis × √ × ×
Eosinophilia × √ × ×
Basophilia × √ × ×
Dimorphologic √ √ × ×
HGB Abn/Interfere? √ √ × ×
Anisocytosis √ √ × ×
Microcytosis √ √ √ √
RBC/HGB Macrocytosis √ √ √ √
Erythrocytosis √ √ √ √
Anemia √ √ √ √
Hypochromia √ √ √ √
PLT Clump? √ √ × ×
Thrombocytosis √ √ √ √
PLT
Thrombopenia √ √ √ √
When the PLT value is less than 100 109 / L, a manual count by the
microscope is recommended.
Ф12×75(mm)(without the cap) evacuated blood collection tube, used for Ф13×75 (mm)
adapter.
Ф13×75(mm)(without the cap) evacuated blood collection tube, used for Ф13×75 (mm)
adapter.
Ф14×75(mm)(without the cap) evacuated blood collection tube, used for Ф15×75 (mm)
adapter.
Ф15×75(mm)(without the cap) evacuated blood collection tube, used for Ф15×75 (mm)
adapter.
6-27
Operating Your Analyzer
6-28
Operating Your Analyzer
If the analyzer is shut down abnormally, you will lose the work list
information of the samples that have not been saved yet.
If you want to complete the work list information after the analysis, see
Chapter 7 Reviewing Sample Results for details.
Click the “Worklist” button on the shortcut area or click “Menu”, then select “Worklist” to
enter the “Worklist” screen.
6-29
Operating Your Analyzer
Click the “New” button, then a new record will be added at the bottom of the worklist and this
blank record is highlighted. All the fields in the information entry area are displayed in defaults
and are activated.
You can switch between options in the Sample Info./Patients Info area by the
[Tab] key. You can also use the [Enter] key to switch after setting, see the
setup details in General Setup section of Chapter 5 Customizing the
Analyzer Software.
You can skip the options that do not need the entry when switching by [Tab]
or [Enter], see the setup details in General Setup section of Chapter 5
Customizing the Analyzer Software.
6-30
Operating Your Analyzer
The sample ID must be entered and its acceptable length is [1, 20].
Select the sampling mode as “AL” and the measurement mode as “CBC” or “CBC+DIFF”
from the pull-down lists respectively.
If the autoloader mode is selected, then the blood sample mode can only be
“whole blood”.
In the “CBC” measurement mode, the analyzer only counts the blood cells
without further differentiating the white blood cells. 13 parameters and
histograms of WBC, RBC and PLT are provided in this mode. In the
“CBC+DIFF” mode, the analyzer counts the blood cells and further
differentiates the white blood cells into 5 sub-populations. 23 basic
parameters, 4 RUO parameters, scattergrams and histograms of
WBC/BASO, RBC and PLT are provided in this mode.
Select the reference group for the sample from the “Ref. Group” pull-down list. The analyzer
will judge the test results according to the reference range of the Ref. group. When the results
exceed the reference range, the analyzer will flag.
If you have entered the gender and age of the patient, then the system will
provide a matching Ref. Group automatically.
If the auto-matching Ref. Group is different from the one that you selected
before (excluding the 5 customized Ref. Groups), then the auto-matching
Ref. Group will be adopted by the system.
Select the draw date from the date control; enter the draw time into the time edit box.
6-31
Operating Your Analyzer
Select the delivery date from the date control and then enter the delivery time into the time
edit box.
The delivery date/time can not be earlier than the draw date/time.
The draw and delivery date/time can not be later than the current system
date/time.
Enter the rack No. and tube No. into the “Rack No.” and “Tube No.” box respectively.
The entry range of the rack No. is [1, 100]; the entry range of the tube No. is
[1, 10].The rack No. and tube No. can not be empty.
In the Uni-directional LIS/HIS mode, after you entering the patient ID and
pressing the [Enter] key, the matched patient information (including “Last
Name”, “First Name”, “Gender”, “Age”, “Birthday”, “Department” and “Bed
No.”) will be displayed in the screen automatically if there is any. You can
also proceed to edit the information.
In the Bi-directional LIS/HIS mode, the patient information will adopt those
downloaded from the LIS/HIS as default.
Enter the patient name into the “Last Name” and “First Name” boxes.
Enter the gender of the patient into the “Gender” box or select it from the “Gender” pull-down
list.
6-32
Operating Your Analyzer
The analyzer provides four ways for you to enter the patient age – in years, in months, in
days and in hours. The first way is designed for the adult or pediatric patients older than one
year; the second for the infant patients one month to one year; the third for the neonatal
patients no older than one month and the fourth for the neonatal no older than 24 hours. You
can choose only one of the four ways to enter the patient age.
The “Age” pull-down list provides four ways for you to enter the patient age– in years, in
months, in days and in hours, and you can enter the patient age in the box followed by the
age unit.
After entering the birthday, the age field will calculate automatically
according to the difference between the “current system date” and the
“birthday”, and then a new result of age and the corresponding unit will be
displayed in the age edit box and the unit combo box respectively.
If the entered birthday is later than the current system, then it is considered
invalid.
Enter the name of the department, from which the sample came, into the “Department” box
or select it from the “Department” pull-down list.
Enter the bed No. of the patient into the “Bed No.” box.
Enter the name of the deliverer into the “Deliverer” box or select it from the “Deliverer”
pull-down list (if there are previously saved deliverers’ names in the list).
Enter the suspect information of diagnosis into the “Clinical Diagnosis” box.
6-33
Operating Your Analyzer
Save
When finish entering the work list information, you can click the “Save” button to save all the
information.
The “Sample ID + Mode” or “Rack No. +Tube No.” of the current record can
not be the same as the unhide records in the following status: “To Be Run”,
“Running” and “Error”.
If the Bi-directional LIS/HIS is selected, the terminal software will obtain the
corresponding information from LIS/HIS after you clicking the “Save” button
and display them in the right field.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
Repeat piercing the evacuated blood collection tube may break the rubber
tube cap. The fragments produced may lead to incorrect analysis result. It is
recommended that do not pierce each tube for more than three times.
Be sure that the entered sample ID, rack No., tube No. and the analysis
mode are strictly in accordance with the sample to be run.
6-34
Operating Your Analyzer
If the sample mode is switched from the “Predilute” to “Whole Blood”, the
analyzer will perform the switching sequence automatically and a progress
bar will be displayed on the screen.
Do not scan the sample ID or rack No. and do not run as per the worklist
1. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green), click
the shortcut button “Run”, the “Run” message box will pop up.
2. Select the “AL-WB” mode in the message box, but do not select the “Run as per the
worklist”, "Automatically scan sample ID" or "Automatically scan rack No." option.
3. Select the desired analysis mode, enter the starting sample ID, rack No. and tube No.
4. Place the tubes into the corresponding tube positions according to the starting rack No.
and tube No. entered.
6-35
Operating Your Analyzer
5. Place racks loading tubes in ascending order on the level of the right tray of the
autoloader, with the back of “MINDRAY” mark on the carrier facing the analyzer.
6. After finishing setting, press [Enter] from keyboard or click the "Run" or “Ok” from the
dialog box, and then press the [Aspirate] key to start the analysis. The analyzer will start
the analysis from the starting position as you set. The analysis status icon and analyzer
indicator is flickering in green at the time.
7. After every analysis cycle, the results will be saved to the Review Screen.
8. When the analysis is finished, the following statistics message box will pop up.
At the Setup screen, you can set whether to display the statistics after the
autoloading is finished every time.
If there are statistics displayed in the message box, you can click
“Details…” to check the detailed sample ID, run date and sample position
accordingly.
9. Click “Ok” to close the box, and then the analysis status icon and analyzer indicator will
return to lasting green.
10. When finish running, all the racks come to the left tray of the autoloader. Remove them
safely.
6-36
Operating Your Analyzer
If more racks are needed when running the samples, you should load the
rack from the right of the autoloader while removing the completed rack
from the left of the autoloader in time.
Do not scan the sample ID or rack No. but run as per the worklist
1. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green), click
the shortcut button “Run”, the “Run” message box will pop up.
2. Select the “AL-WB” mode in the message box, then select the “Run as per the
worklist” but do not select the "Automatically scan sample ID" or "Automatically scan
rack No." option.
3. Mark the prepared tubes according to the record in the list. Place the tubes into the
corresponding tube positions.
4. Place racks loading tubes in ascending order on the level of the right tray of the
autoloader, with the back of “MINDRAY” mark on the carrier facing the analyzer.
5. After finishing setting, press [Enter] from keyboard or click the "Run" or “Ok” from the
6-37
Operating Your Analyzer
dialog box, and then press the [Aspirate] key to start the analysis. The analyzer will start
the analysis according to the tube positions sequentially. The analysis status icon and
analyzer indicator is flickering in green at the time.
6. After every analysis cycle, the results will be saved to the Review Screen.
7. When the analysis is finished, the following statistics message box will pop up.
At the Setup screen, you can set whether to display the statistics after the
autoloading is finished every time.
If there are statistics displayed in the message box, you can click
“Details…” to check the detailed sample ID, run date and sample position
accordingly.
8. Click “Ok” to close the box, and then the analysis status icon and analyzer indicator will
return to lasting green.
9. When finish running, all the racks come to the left tray of the autoloader. Remove them
safely.
6-38
Operating Your Analyzer
For the unhidden (or error) samples to run in the autoloader mode in the
worklist, the analyzer will run the samples according to the sample position.
In the “Run” message box, the gray-displayed sample ID, rack No., tube No.
and measurement mode are determined by the record in the worklist with
the following features: the earliest, in autoloader mode, unhidden, to be run
(or error).
If there is no record of AL-WB mode in the worklist, then the “Run as per the
worklist” button in the “Run” screen will be displayed in gray.
When “Run as per the worklist”, if the current tube is excluded in the
worklist, the analyzer will still skip it without running.
If more racks are needed when running the samples, you should load the
rack from the right of the autoloader while removing the completed rack
from the left of the autoloader in time.
Automatically scan the sample ID and/or rack No. but do not run as per
the worklist
1. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green), click
the shortcut button “Run”, the “Run” message box will pop up.
6-39
Operating Your Analyzer
2. Select the “AL-WB” mode in the message box, then select "Automatically scan
sample ID" and/or "Automatically scan rack No." but do not select the “Run as per the
worklist” option.
3. Then select the desired analysis mode. If "Automatically scan sample ID" is not
selected, enter the sample ID; if "Automatically scan rack No." is not selected, enter the
starting rack No. and tube No; if both of the two check box are selected, you can not enter
in these edit boxes.
4. If "Automatically scan sample ID" is selected, place labels on the tubes, and then
place them into the corresponding tube positions.
5. Place racks loading tubes in ascending order on the level of the right tray of the
autoloader, with the back of “MINDRAY” mark on the carrier facing the analyzer.
6. After finishing setting, press [Enter] from keyboard or click the "Run" or “Ok” from the
dialog box, and then press the [Aspirate] key to start the analysis. The analyzer will scan
the sample ID and/or rack No. from the starting position as you set, then start the analysis.
The analysis status icon and analyzer indicator is flickering in green at the time.
7. After every analysis cycle, the results will be saved to the Review Screen.
8. If "Automatically scan rack No." is selected, when invalid rack No. is scanned during
counting, the following message box will pop up. Click “Ok” to close it, and the analyzer
will skip the tubes on this rack and continue with other tubes.
9. When the analysis is finished, the following statistics message box will pop up.
6-40
Operating Your Analyzer
At the Setup screen, you can set whether to display the statistics after the
autoloading is finished every time.
If there are statistics displayed in the message box, you can click
“Details…” to check the detailed sample ID, run date and sample position
accordingly.
10. Click “Ok” to close the box, and then the analysis status icon and analyzer indicator will
return to lasting green.
11. When finish running, all the racks come to the left tray of the autoloader. Remove them
safely.
If “Run as per the worklist” is not selected but “Built-in barcode scanner” is
selected, then no actions will be taken to the records in the worklist during
running in autoloader mode.
If the “Sample ID” box displays “Invalid”, it means the scanning is invalid.
You can re-enter the sample ID when reviewing the sample results, see
Chapter 7 Reviewing Sample Results for details.
If more racks are needed when running the samples, you should load the
rack from the right of the autoloader while removing the completed rack
from the left of the autoloader in time.
6-41
Operating Your Analyzer
Automatically scan the sample ID and/or rack No. and run as per the
worklist
1. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green), click
the shortcut button “Run”, the “Run” message box will pop up.
2. Select the “AL-WB” mode in the message box, then select "Automatically scan
sample ID" and/or "Automatically scan rack No.", as well as the “Run as per the
worklist” option.
3. If "Automatically scan sample ID" is selected, place labels on the tubes, then place
them into the corresponding tube positions.
4. Place racks loading tubes in ascending order on the level of the right tray of the
autoloader, with the back of “MINDRAY” mark on the carrier facing the analyzer.
5. After finishing setting, press [Enter] from keyboard or click the "Run" or “Ok” from the
6-42
Operating Your Analyzer
dialog box, and then press the [Aspirate] key to start the analysis. The analyzer will start
scanning sample IDs and/or rack No., and running samples according to the matched
sample in the worklist sequentially. The analysis status icon and analyzer indicator is
flickering in green at the time.
6. After every analysis cycle, the results will be saved to the Review Screen.
7. If "Automatically scan rack No." is selected, when invalid rack No. is scanned during
counting, the following message box will pop up. Click “Ok” to close it, and the analyzer
will skip the tubes on this rack and continue with other tubes.
8. When the analysis is finished, the following statistics message box will pop up.
At the Setup screen, you can set whether to display the statistics after the
autoloading is finished every time.
If there are statistics displayed in the message box, you can click
“Details…” to check the detailed sample ID, run date and sample position
accordingly.
9. Click “Ok” to close the box, and then the analysis status icon and analyzer indicator will
return to lasting green.
6-43
Operating Your Analyzer
10. When finish running, all the racks come to the left tray of the autoloader. Remove them
safely.
If there is no record of AL-WB mode in the worklist, then the “Run as per the
worklist” button in the “Run” screen will be displayed in gray.
If "Automatically scan rack No." is not selected, the system will search for
the matching record in the worklist according to the scanned sample ID, and
then the rack No. and tube No. will be overwritten. The run status of the
record will be changed from “To Be Run” to “Running”.
If more racks are needed when running the samples, you should load the
rack from the right of the autoloader while removing the completed rack
from the left of the autoloader in time.
Stop
During running, the “Run” button at the shortcut buttons area will be replaced by “Stop”. .
Click the “Stop” button, after the previously analyzing cycle is finished, the system stops the
analysis and the current sample carrier will be pushed out.
STAT
If there is STAT sample requires running first during the autoloader sampling analysis.
Insert STAT
1. Click the shortcut button “STAT”, the message box will pop up.
2. Click “Yes”. After the previously analyzing of the pierced sample is finished, the analysis
stops and the system switch the autoloader mode to the CT-WB sampling mode and
6-44
Operating Your Analyzer
3. Present the whole blood sample into the compartment, push the compartment door to
close it.
4. When the STAT run is ready (i.e. the analysis status icon and analyzer indicator is in
yellow), you can run the STAT sample in ways as the closed tube sampling mode, see
details in Closed Tube Sampling Analysis.
When the current sample is being run, the massage box “Please finish the
current sample then perform STAT operation. Please wait!” will pop up. You
can click the “Ok” button to close it.
When a STAT is inserted, the system will switch to the worklist screen
automatically and add a new STAT record at the first line of the worklist and
it is displayed in yellow background.
You can start the STAT analysis only by pressing the [RUN] key on the
analyzer.
After running a STAT sample, you can switch to the graph or table review
screen to edit the corresponding information.
Cancel
After the running is finished, you can click the “Cancel” button to cancel the SATA, and then
the analyzer will judge whether to switch to the analysis mode before the inserted STAT
accordingly, and then proceed with autoloading.
If you click the “Cancel” button when there are unfinished samples left in
the worklist, a message box will pop up and then you can select whether to
delete the unfinished STAT sample.
After exiting the SATA mode, the background of all the SATA samples will
return to normal color.
6-45
Operating Your Analyzer
Parameter flags
If the parameter follows a “H” or “L”, it means the analysis result has exceeded the upper
or lower limit of the reference range, but still within the display range
If the parameter follows a “?”, it means the analysis result obtained is suspect.
If you see *** as opposed to the result, it means the result is either invalid or out of the
display range.
For the background test, the flags of parameter or flags of abnormal blood
cell differential or morphology are not available.
6-46
Operating Your Analyzer
Left Shift?
Erythrocytosis
RBC abnormal distribution
Anisocytosis
Macrocytosis
Abnormal
RBC/HGB Microcytosis
Dimorphologic
Anemia
Hypochromia
The analyzer will flag abnormal or suspect WBC, RBC and PLT according to the scattergrams
and histograms. The following table shows how the flags affect parameter results:
Left Shift? × √ × ×
Abnormal/Atypical Lymphocyte? × √ × ×
6-47
Operating Your Analyzer
Leucocytosis √ √ √ √
Leucopenia √ √ √ √
Neutrophilia × √ × ×
Neutropenia × √ × ×
Lymphocytosis × √ × ×
Lymphopenia × √ × ×
Monocytosis × √ × ×
Eosinophilia × √ × ×
Basophilia × √ × ×
Dimorphologic √ √ × ×
HGB Abn/Interfere? √ √ × ×
Anisocytosis √ √ × ×
Microcytosis √ √ √ √
RBC/HGB Macrocytosis √ √ √ √
Erythrocytosis √ √ √ √
Anemia √ √ √ √
Hypochromia √ √ √ √
PLT Clump? √ √ × ×
Thrombocytosis √ √ √ √
PLT
Thrombopenia √ √ √ √
When the PLT value is less than 100 109 / L, a manual count by the
microscope is recommended.
6-48
Operating Your Analyzer
Blood, scratch and powders of the gloves will affect the accurate barcode
scanning. Be sure there are no blood, scratch and powders on the barcode
labels.
The label must be pasted properly to ensure the accurate barcode scanning. You should
paste the label to the area marked with “X” as shown below.
6-49
Operating Your Analyzer
6-50
Operating Your Analyzer
6.8 Worklist
Click the “Worklist” button on the shortcut area or click “Menu”, then select “Worklist” to
enter the “Worklist” screen.
The upside of the interface is the worklist; the downside is information entry area including
Sample Info. and Patient Info. The bottom of the screen is the function button area.
All the information fields in the worklist are entered through the information
entry area except the “No.”, “Run Status” and “Entry Time”.
If the worklist is empty, all the information fields in the information entry
area are blank and displayed in gray.
In the “Worklist” screen, you can perform the following operations to the worklist in the
upside of the screen.
6-51
Operating Your Analyzer
Click and hold the title of the column then drag the column to the desired position to adjust
the display order.
Click and hold the boundary line between the two columns, then drag the line to adjust the
width of each column.
1. Right click the highlighted record; then the following shortcut menu will pop up:
2. Click “Top” to set the highlighted record as the first record in the worklist.
5. Click “Bottom” to set the highlighted record as the last record in the worklist.
If you click a record in the worklist to highlight it, the corresponding information of the record
will display in the information entry area. You can edit each information field in the information
entry area.
For the record whose “Run Status” is “Running”, you can not edit its
“Sample ID”, “Mode”, “Rack No.” and “Tube No.”(autoloader mode).
The information entry area of the records whose “Run Status” are
“Finished” will be displayed in gray and unavailable to edit and modify. You
can switch to the Graph review or Table review screen to edit and modify the
corresponding information.
When the mouse is moved just on the function buttons, the name of the button and the
corresponding shortcut key will appear. For example, when moving the mouse on the “Save”
button, the tips will pop up:
The function buttons at the bottom of the “Worklist” screen and their shortcut keys are shown
in the table.
6-52
Operating Your Analyzer
New
You can click the “New” to add a new sample record, see Entering Work List Information
section of this chapter for details.
Insert
2. Click the “Insert” button to insert a new record before the highlighted record and then the
new added blank row will be highlighted. All the fields in the information entry area are
displayed in defaults and are activated.
3. You can enter the sample/patient information in the information entry area, see Entering
Work List Information section of this chapter for details.
Save
After performing the “Edit”, “New” or “Insert” operation, you can click the “Save” button to
save all the information.
Delete
1. Click the ”Delete” button, then the “Delete” message box will pop up.
6-53
Operating Your Analyzer
2. Click the radio button “Selected Samples”, “All finished records” or “All records” to
select the records you want to delete.” Selected Samples” are those selected with “√”
marks in the worklist.
The records whose “Run Status” are “Running” can not be deleted.
Search
1. Click the ”Search” button, then the “Search” message box will pop up.
2. Click one or more check boxes to define the desired search condition(s).
3. Enter the search content in the edit box of the desired search condition.
6-54
Operating Your Analyzer
4. If you wish to perform the precise search, you can select the “Whole Words Only”
check box; if you wish to perform the fuzzy search (means to search the related records
which contain the content that you entered), you should leave the check box in blank.
If the first/last record is reached, then the searching circle will start again
from the last/first record upwards/downwards.
6. A searching circle will be completed when backing to the initial record. If there is no
matching record found, the prompt message box “No record found!” will pop up at the
screen; otherwise, the prompt message box “Search finished!” will pop up.
8. You can repeat procedure 2 to 6 to search for other content; or click the “Close” button to
finish searching and close the message box.
Copy
2. Click the “Copy” button to add a new record in the worklist and highlight it. The sample ID
of this new added record is empty or will automatically increase by 1 based on the last
sample ID in the worklist, the other information remains the same as the record copied
from. For the autoloader samples, the “Rack No.-Tube No.” of this new copied record will
increase by 1 based on the greatest “Rack No.-Tube No.” existed in the worklist.
Hide
2. Click the “Hide” button to hide the selected record and display it in gray.
6-55
Operating Your Analyzer
If the selected records include both hidden and unhidden records, when you
click the “Hide” button, all of them will be hidden.
The records whose “Run Status” are “Running” or “Finished” can not be
hidden.
Cancel
2. Click the “Cancel” button to cancel the hide and gray display status of the record.
If the selected records are all hidden records, the “Hide” button will
replaced by the “Cancel” button.
2. Click the “Print” button, and then a message box will pop up.
6-56
Operating Your Analyzer
6.9 Shutdown
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The shutdown procedure includes closing the analyzer and exiting the software. The
following content will introduce the two procedures respectively.
6-57
Operating Your Analyzer
3. During the shutdown procedure, the shutdown information will be displayed in the
information indicating area at the bottom of the interface.
5. Place the power switch at the left side of the analyzer in the OFF position (O). The
message box will be closed automatically.
If the analyzer disconnects with the computer, you can not perform the
shutdown procedure.
If error happens during shutdown procedure, the analyzer will return to the
status before the shutdown procedure is performed, and then alarm. See
Chapter 11 Troubleshooting Your Analyzer for details to remove the error.
You can click the “Restart” button to restart the analyzer and perform
startup initialization.
You will not exit the software after the shutdown of the analyzer, and you can
still perform operations that are available without the cooperation of the
analyzer.
1. Click the shortcut button “Exit”, or select “Menu”“Exit”“Exit” option, the following
message box will pop up.
6-58
Operating Your Analyzer
You must shut down the analyzer before exiting the software.
1. Close the external computer according to the shutdown procedures of the operation
system.
You should exit the terminal software first and then turn off the external
computer according to standard procedures. Otherwise, the database of the
terminal software might be lost!
6-59
7 Reviewing Sample Results
7.1 Introduction
The analyzer automatically saves analysis results. Totally 40,000 results can be saved,
including sample information, parameters, flag prompts, scattergrams and histograms.
You can browse sample results either in the table or graph mode.
7-1
Reviewing Sample Results
The “Graph” screen consists of three parts. The upside displays the Sample/Patient Info. The
downside displays the Results, Scattergrams, Histograms, Flags, DIFF Graphs, Microscopic
Exam Results and Blood Type/ESR results in accordance with the Sample/Patient Info. in the
form of tabs including “Data/Graph”, “DIFF”, “Microscopic Exam. and Others” and
“Research”. The bottom displays the functional buttons available in the current screen.
7-2
Reviewing Sample Results
You can enable users of common level to edit the sample ID by setting in the
“Setup” screen, see chapter 5 Customizing the Analyzer Software for
details.
You can edit all the information of the sample except the Mode and Run
Time.
7.2.2 Tabs
After selecting a sample record, you can click the tab at the downside of the screen to see the
corresponding information.
Data/Graph
Click the “Data/Graph” tab to see the data/graph information of the record.
You can select whether to display the four RUOs, the “*” mark and the
corresponding declarations (“*” means “research use only, not for
diagnostic use”) in “Setup” screen, see Chapter 5 Customizing the Analyzer
Software for details.
When the results of Bas% and Bas# are expressed in “*”, the second
histogram discriminator will not be displayed.
For details of how to edit and restore the result, please see the following Edit Result and
7-3
Reviewing Sample Results
When finish browsing, you can click “X” on the top right of the box to close it.
DIFF
You can click the “DIFF” tab to check the WBC differential information of the record.
You can select whether to display the four RUOs, the “*” mark and the
corresponding declarations (“*” means “research use only, not for
diagnostic use”) in “Setup”, see Chapter 5 Customizing the Analyzer
Software for details.
When the results of Bas% and Bas# are expressed in “*”, the second
histogram discriminator will not be displayed.
7-4
Reviewing Sample Results
For details of how to edit and restore the result, please see the following Edit Result and
Restore Result section of this chapter.
When moving the mouse to any graph of the scattergram/histogram, an icon of magnifier will
appear. Click the icon, a box of enlarged graph will pop up and you can drag the box at will.
When finish browsing, you can click “X” on the top right of the box to close it.
Click the “Sample Type” combo box; select the sample type as “Venous blood” (default) or
“Capillary Blood”.
7-5
Reviewing Sample Results
Click the “Microscopic exam. time” edit box, enter the date and time of microscopic exam.
The Microscopic exam. time can not exceed the current system time.
You can enter the morphology information for WBC, RBC and PLT respectively into the
multi-line edit box.
You can enter the percentage or other form of differential result of each cell differential into
the edit box next to the cell differential name respectively.
You can enter a value within the range [0.0-100.0] and the unit is “%”.
You can select the blood type of the patient in the “Blood Type/ESR” column. Click the first
combo box next to the blood type, you can select from “Blank”, “A”, “B”, “O” and “AB”; click
the second combo box, you can select from “Blank”, “RH+” and “RH-”.
You can enter the blood ESR value into the edit box follows the “ESR”. If the value exceeds
the Ref. Range, the flags “H” or “L” will appear to indicate the value exceeds the upper limit or
lower limit.
You can modify the reference range of Blood ESR by the following steps:
1. Click “Set Reference Range” button, a message box will pop up. Enter the upper limit
and lower limit of the blood ESR in the edit box “Upper limit” and “Lower limit”
respectively.
7-6
Reviewing Sample Results
2. Click the “Ok” button to save the settings and refresh the information.
You can enter the value up to 3 numeric characters within the range [0,999].
The upper limit can not be smaller than the lower limit.
The entered reference range of the Blood ESR is only applied to the current
record, and the default range is [0, 20].
Research
Click “Research” tab on the screen, the specific value of each parameter will be displayed.
The specific values of the parameter results that are out of the display range
or without data collected cannot be provided.
7-7
Reviewing Sample Results
Edit of the results in the “Data/Graph” tab will not affect the display of
parameters in the “Research” tab.
The content of this tab can only be viewed and used for research; it cannot
be edited or printed.
The shortcut keys of the function buttons in the graph review screen are shown in the
following table.
Validate F3
Print F4
Edit Result F5
Restore Result F6
Auto-refresh
At the graph review screen, when browsing the results, you can select to activate or
deactivate the auto-refresh function to display the latest results.
Click the button at the lower right of the graph review screen to make it raised. Later
on, the graph review screen will refresh automatically to display the latest results and graphs
if any. At the mean time, the graph will be enlarged as shown below.
When the button is raised, all the information fields and buttons at the graph review
screen will be unavailable (displayed in gray) except the records switching column.
Click the button at the lower right of the graph review screen to make it sunk. Later
on, the graph review screen will not refresh even the new results are obtained, but still display
the current sample information, results and graphs that you are now browsing. At the mean
7-8
Reviewing Sample Results
When the button is sunk, all the operations are available to the current displayed
records.
The default status of the button at the graph review screen is sunk.
The status of the button keeps still when you return to the graph
review screen after exiting.
When you browsing records at the graph review screen by using the
Save
Click the “Save” button to save the modified information on all tabs of the current result.
7-9
Reviewing Sample Results
Print
Click the “Print” button to print the information, result, histogram and scattergram of the
current sample.
You can set the amount of copies for the printed report in the “Setup”
screen.
At the “Setup” screen, you can select whether to print the Flag information
in the report.
Delete
2. Click “Ok” to delete the current displayed sample record in the “Graph” screen.
The “Delete” button and the corresponding deleting operation are not
available to users of common-level.
Validate
Click the “Validate” button to perform the validating operation.
You can enable the users of common level to validate by setting in the
“Setup” screen. Otherwise, the user name and the pass word of
administrator level are required.
After validating, you can not edit the Sample/Patient Info. and result.
7-10
Reviewing Sample Results
Cancel (validate)
Click the “Cancel” button to cancel the validating operation.
After canceling, you can edit the Sample/Patient Info. and result.
Edit Result
1. Click the “Edit Result” button, then the result of each parameter and WBC DIFF results
will be displayed in an edit box for you to edit.
2. After editing, click the “Save” button to save the change and the edit box disappears.
You can enable the users of common level to edit result by setting in the
“Setup” screen. Otherwise, the user name and the pass word of
administrator level are required.
If the result of one parameter is modified, then the result of other related
parameter(s) will be changed accordingly and the high or low/suspect flags
will also be refreshed.
Only the result of the measurement parameters (WBC, RBC, HGB, HCT and
PLT) and WBC DIFF results can be modified.
After editing and then saving WBC DIFF results, the absolute value of each
DIFF result will be re-calculated and then refreshed.
If the sum of the DIFF results does not equal to 100.00% after being edited,
then the message box “The sum of the DIFF results is not 100.00%!” will pop
up when you clicking the “Save” button.
7-11
Reviewing Sample Results
Restore Result
1. Click the “Restore Result” button, and then the following message box will pop up.
2. Click “Ok” to restore the result to the original measurement value and remove the result
edited flags (“E” or “e”).
7-12
Reviewing Sample Results
Click the shortcut button “Table” or Click the “Menu” button, and then select the “Review”
“Table” to enter the following table review screen.
The “Table” screen consists of three parts. The upside of the screen displays the sample
records in tables. The downside of the screen displays the Result, Sample/Patient Info.,
Microscopic Exam Result and Blood Type/ESR Result of the current sample record in the
form of tabs. The top and bottom of the screen displays the functional buttons available in the
current screen.
You can browse each sample record and its Sample/Patient information in the “Table” screen.
7-13
Reviewing Sample Results
For the error sample record, the content of each information field is
displayed in red.
For the printed sample, its cell in the “Print” column displays “P”. For the
unprinted sample, its cell in the “Print” column is blank.
For the validated sample, its cell in the “Validate” column displays “V”; for
the sample not validated, its cell is blank.
For the transmitted sample, its cell in the “Transmission” column displays
“T”; for the sample not transmitted, its cell is blank.
In the sample records table area, you can perform the following operations:
Click the “Sample List” combo box, then you can select “Samples within today” (default),
“All Samples” and ”Samples found”.
The Review List will display different records according to the different options:
Samples within today Display only the sample records within today
Samples found Display all the sample records met the search requirements.
Click and hold the title of the column then drag the column to the desired position to adjust
the display order.
1. Right click the “Run Date” title, and then a shortcut menu will pop up:
2. Click “Ascending” to rank the records in ascending order of “Run Date + Time” (i.e. the
latest testing sample ranks last in the list).
3. Click “Descending” to rank the records in descending order of “Run Date + Time” (i.e.
the latest testing sample ranks the first in the list).
7-14
Reviewing Sample Results
The records in the graph screen will be ranked the same way as the table
screen, in ascending/descending order of “Run Date + Time”.
Switching
Double click a record in the list; the screen will switch to the graph review screen of the record
automatically.
7.3.2 Tabs
Result
You can click the “Result” tab to see all the results of the highlighted record.
You can select whether to display the four RUOs, the “*” mark and the
corresponding declarations (“*” means “research use only, not for
diagnostic use”) in “Setup”, see Chapter 5 Customizing the Analyzer
Software for details.
7-15
Reviewing Sample Results
For details of how to edit and restore the result, please see the following Edit Result and
Restore Result section of this chapter.
Sample/Patient Information
You can click the “Sample/Patient Info.” tab to see the sample information and patient
information of the highlighted record in the list.
For details to edit information, see Editing work list information section in Chapter 6
Operating Your Analyzer.
7-16
Reviewing Sample Results
Click the “Sample Type” combo box, and then select the sample type as “Venous blood”
(default) or “Capillary Blood”.
Click the “Microscopic exam. time” edit box, enter the date and time of microscopic exam.
The Microscopic exam. time can not exceed the current system time.
You can enter the morphology information for WBC, RBC and PLT respectively into the
multi-line edit box.
You can enter the percentage or other form of differential result of each cell differential into
the edit box next to the cell differential name respectively.
You can enter a value within the range [0.0-100.0] and the unit is “%”.
You can select the blood type of the patient in the “Blood Type/ESR” column. Click the first
7-17
Reviewing Sample Results
combo box next to the blood type, you can select from “Blank”, “A”, “B”, “O” and “AB”; click
the second combo box, you can select from “Blank”, “RH+” and “RH-”.
Enter the blood ESR value in the edit box follows the “ESR”. If the value exceeds the Ref.
Range, the mark “H” or ”L” will appear to indicate the value exceeds the upper limit or the
lower limit.
You can modify the reference range of Blood ESR by the following steps:
1. Click “Set Reference Range” button, and then a message box will pop up. Enter the
upper limit and lower limit of the blood ESR into the edit box “Lower limit” and “Upper
limit” respectively.
2. Click the “Ok” button to save all the settings and refresh the information.
You can enter the value up to 3 numeric characters within the range [0,999].
The upper limit can not be smaller than the lower limit.
The entered reference range of the Blood ESR is only applied to the current
record, and the default range is [0, 20].
Research
Click “Research” tab on the screen, the specific value of each parameter will be displayed.
7-18
Reviewing Sample Results
The specific values of the parameter results that are out of the display range
or without data collected cannot be provided.
Edit of the results in the “Data/Graph” tab will not affect the display of
parameters in the “Research” tab.
The content of this tab can only be viewed and used for research; it cannot
be edited or printed.
The shortcut keys of the function buttons in the table review screen are shown in the following
table:
Validate F3
Print F4
Edit Result F5
Restore Result F6
Batch Validate F7
7-19
Reviewing Sample Results
Search F8
Communication F9
Delete Alt+Delete
Deselect F10
CV F12
Save
Click the “Save” button to save the modified information on all tabs of the current result.
Search
You can search for the specified sample record from all records in the current list as default.
1. Click the “Search” button, and then a “Search” message box will pop up.
Select the check box of “Sample ID”, and then enter the desired sample ID into the “Sample
ID” edit box.
7-20
Reviewing Sample Results
Select the check boxes of “Last Name” and “First Name”, and then enter the desired patient
name into the boxes.
Select the check box of “Run Date”, and then select the limits of the run date.
Select the check box of “Gender”, and then click the radio button “Male”, “Female” or
“Empty” to select the patient gender.
Select the check box of “Patient No.”, and then enter the desired patient No. into the “Patient
No.” edit box.
Select the check box of “Department”, and then enter the desired department name in the
“Department” edit box.
Select the check box of “Bed No.”, and then enter the desired bed No. into the “Bed No.” edit
box.
Select the check box of “Deliverer”, and then enter the desired deliverer into the “Deliverer”
edit box.
Select the check box of the “Validate Status”, and then click the radio button “Validated” or
“Not Validated” to select the validate status.
Select the check box of “Print Status”, and then click the radio button “Printed” or “Not
Printed” to select the print status.
Select the check box of the “Communication Status”, and then click the radio button
7-21
Reviewing Sample Results
Select the check box of “Whole Words Only”, and then the precise search will be performed;
otherwise, the fuzzy search (means to search the related records which contain the content
that you entered) will be preformed.
Select the check box of “Case Sensitive”, and then the capital letters and small letters in the
edit box will be distinguished when searching; otherwise, the search will be insensitive to the
form the letters (i.e. the capita letters and small letters will not be distinguished).
3. Click “Ok” to perform the search and switch to the “Samples found” list of the “Table”
screen, and the searching results will display..
The desired record is searched from all the sample records as default.
Validate
Click the “Validate” button to validate the current highlighted record in the list.
You can enable the users of common level to validate by setting in the
“Setup” screen. Otherwise, the user name and the password of
administrator level are required.
After validating, you can not edit the sample/patient information and the
result
Batch Validate
1. Click the “Batch Validate” button, and then the following message box will pop up.
7-22
Reviewing Sample Results
2. Click the radio button “Selected Samples” or “Specified Samples” to select the records
you want to validate.” Selected Samples” are those selected with “√” marks in the
review list.
3. Click the “Specified Samples” radio button to specify the starting and finishing time of
the Run Date for the record to be validated.
For the validated record, you can not edit the sample/patient information
and the result.
Cancel (validate)
Click the “Cancel” button to cancel the validating operation.
7-23
Reviewing Sample Results
The “Cancel” button is only available to the highlighted record in the list.
After canceling, you can edit the sample/patient Information and the result.
1. Click the “Print” button, and then a message box will pop up.
2. Click the radio button “Selected Samples” or “Specified Samples” to select the record
you want to export.” Selected Samples” are those selected with “√” marks in the review
list.
3. Click the “Specified Samples” radio button to set the starting and finishing run date of
the records to be printed.
4. Click the “Report” or “List” radio button to select the print format.
5. When “List” is selected, you can continue to select one of the four print templates, the
7-24
Reviewing Sample Results
You can set the amount of copies of the printed report in the “Setup” screen.
At the “Setup” screen, you can choose whether to print the Flag prompts in
the report or not.
Communication
You can do as follows to transmit the sample record to the LIS/HIS system.
1. Click “Transmit” button, the following message box will pop up.
2. Click the radio button “Selected Samples” or “Specified Samples” to select the record
you want to transmit.” Selected Samples” are those selected with “√” marks in the
review list.
3. Click the “Specified Samples” radio button to specify the starting and finishing time of
the Run Date for the record to be transmitted.
Once the transmission starts, if you click the “Com.” button again, then the
foregoing message box will also pop up but the “Start” button will be
7-25
Reviewing Sample Results
replaced by the “Stop” button. You can click the “Stop” button to stop
transmitting once the transmission of the current sample record is done.
CV
You can check the reproducibility of the selected sample record.
2. Click the “CV” button to start calculating the reproducibility, and then the result message
box will pop up:
3. Click the “Calculate Deviation” button, and then a message box will pop up. You can
check the absolute deviation of the 5 WBC related parameters of percent-style.
7-26
Reviewing Sample Results
4. When finish browsing, you can click the “Close” button to exit.
Trend Graph
You can check the trend graph of all the parameters of the selected sample record. Do as
follows:
2. Click the “Trend Graph” button, and then a message box with the trend graph of all
parameters of the selected record will pop up.
7-27
Reviewing Sample Results
At least 3 records or at most all records in the review list can be selected.
Checking data
Method 1: click the certain data group to move the green line to the place, and then you can
check the data of this group.
Method 2: click the arrow buttons on the “Pos./Total” control to move the green line and check
the data of each group.
Method 3: when the green line is located, you can press the [←] and [→] key on the keyboard
to move the green line and check the data of each group.
Method four: when the green line is located, you can press the [Home] or [End] key on the
keyboard to check the first or the last group of data on the graph.
Method 1: click the arrow buttons next to the range to adjust it. The trend graph will refresh
immediately once the range is changed.
Method 2: click the edit box of the range and enter the value into it. After entering, press the
[Enter] key or switch to other focus to refresh the trend graph.
The modified range will also be saved after closing the trend graph.
7-28
Reviewing Sample Results
Delete
The “Delete” button and the corresponding deleting operation are not
available for users of common-level.
2. Click the “Delete” button, and then a message box will pop up.
Edit Result
1. Click the “Edit Result” button, and then in the “Result” tab, you can edit the result of
each parameter and WBC DIFF results in the activated edit box.
You can enable the users of common level to edit result by setting in the
“Setup” screen. Otherwise, the user name and the pass word of
administrator level are required.
If the result of one parameter is modified, then the result of other related
parameter(s) will be changed accordingly and the high or low/suspect flags
will also be refreshed.
Only the result of the measurement parameters (WBC, RBC, HGB, HCT and
PLT) and WBC DIFF results can be modified.
After editing and then saving WBC DIFF results, the absolute value of each
DIFF result will be re-calculated and then refreshed.
If the sum of the DIFF results does not equal to 100.00% after being edited,
then the message box “The sum of the DIFF results is not 100.00%!” will pop
7-29
Reviewing Sample Results
The scattergram of the sample will not be changed even when the
differential result of the WBC is modified.
Restore Result
1. Click the “Restore Result” button, and then the following message box will pop up.
2. Click “Ok” to restore the result to the original measurement value and remove the result
edited flags (“E” or “e”).
The users of common level are enabled the authority of “Restore Result”
together with “Edit Result” when setting in the “Setup” screen. Otherwise,
the user name and the pass word of administrator level are required.
7-30
Reviewing Sample Results
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the data backup function, you
should log on the system as administrator, and then right click the shortcut
icon of the terminal software and run it as administrator.
You can back up the data of the sample base in the external computer. Do as follows:
1. At the table review screen, click the “Menu” button, then select “Review” “Data
Backup” on the pop-up menu, and then a message box will pop up.
2. Click the radio button “Selected Samples” or “Specified Samples” to select the record
you want to backup.” Selected Samples” are those selected with “√” marks in the review
list.
3. Click the “Specified Samples” radio button to specify the starting and finishing time of
the Run Date for the record to be backed up.
4. Click “Ok” button, the following dialog box will pop up.
7-31
Reviewing Sample Results
6. Click the “Save” button, and then a progress bar will pop up.
7-32
Reviewing Sample Results
You can not choose the file format when backing up.
The backup sample record can not be edited and can only be reviewed in
“History”.
7-33
Reviewing Sample Results
1. At the “Table” screen, click the “Menu” button, then select “Review” “Data Export” in
the pop-up menu, and then a “Export” message box will pop up.
2. Click the radio button “Selected Samples” or “Specified Samples” to select the records
you want to export.” Selected Samples” are those selected with “√” marks in the review
list.
3. Click the “Specified Samples” radio button to set the starting and finishing run date of
the record to be printed.
4. Click the “Ok” button, and then a message box will pop up.
7-34
Reviewing Sample Results
5. Select the directory and format and enter name for the exported file.
6. Click the “Save” button, and then a process bar will pop up.
7-35
Reviewing Sample Results
The default format of the exported files is “.csv” and you can also choose
the “.txt” format.
The exported sample record will keep the same order as displayed in the
table review screen (run time ascending/descending).
You can not review the exported files in “History”, but you can perform the
operations including editing and deleting, etc. to the exported files on an
external computer.
7-36
Reviewing Sample Results
7.6 Auto-Backup
When the terminal software connects the analyzer for 4 hours, the backup will be performed
automatically. A progress bar will pop up.
You can not perform any operation during the process of backup.
The record in the sample base will also be backed up automatically when
you exiting the application software.
7-37
Reviewing Sample Results
7.7 Auto-Restore
If the data is damaged in the sample base, but the corresponding auto-backed up data is fine,
then a message box will pop up.
Click “Yes” to close the box and display the auto-backed up data into the sample list.
If you choose not to restore the auto-backed up data and running no new
samples, then the foregoing message box will still pop up when you run the
program again.
7-38
Reviewing Sample Results
7.8 Compare
Click the “Menu” button, and then select “Review” “Compare” from the pop-up menu to
enter the following “Compare” screen.
The “Compare” screen consists of three parts. The top of the screen displays the search
conditions: “Patient ID”, “Last Name”, “First Name” and “Run Date”. The middle of the
screen displays the tab of the patient’s test results and the trend graphs. The available
function buttons are also displayed in the screen.
7.8.1 Tabs
Comparison Summary
Click the “Comparison Summary” tab in the screen to check the test results of the patient.
7-39
Reviewing Sample Results
If “Display RUO parameters” is not selected in the “Setup” screen, then the
ALY%, LIC%, ALY# and LIC# will not be displayed in the list.
The red flags (“?”, “H” or “L”) indicate the result is either out of limit or
suspect.
Result Trend
Click the “Result Trend” tab in the screen to check the trend graphs of the patient.
7-40
Reviewing Sample Results
You can click the “Select Parameter” combo box at top-left of the screen to select the desired
the trend graph.
If “Display RUO parameters” is not selected in the “Setup” screen, then the
ALY%, LIC%, ALY# and LIC# will not be displayed in the pull-down list of
the combo box.
The x-axis of the “Parameter Trend” shows the No. of the test and the y-axis shows the result
of each test. Up to 30 data points can be displayed in the trend graph. If the matched data are
more than 30 groups, then only the latest 30 ones will be displayed. The data points in the
trend graph are displayed from left to right in the ascending order of “Run Date /Time”. Thus,
the latest data point places the last.
The scale on the x-axis of the trend graph changes according to the number
of the result.
The scale on the y-axis of the trend graph changes according to the value of
the result.
7-41
Reviewing Sample Results
The down side of the tab displays the result of the single parameter in the form of list. The
corresponding No., run date/time and result are displayed in the list.
The data in the list corresponds with the data point in the trend graph one by
one.
The red flags (“?”, “H” or “L”) indicate the result is either out of limit or
suspect.
Search
You can search for the certain result of the patient by entering the search condition in the top
of the screen.
Enter the patient name into the “Last Name” and “First Name” boxes.
Use the date control to specify the range for the run date.
An empty entry of run date indicates the whole database will be searched for
the certain result and there is no restriction to the run date.
2. Click the “Search” button to start searching the specified record and the result will be
displayed in the two tabs. You can switch between the two tags to check the result.
7-42
Reviewing Sample Results
Re-fill
You can click the “Re-fill” button to clear all the entered search conditions and then re-fill.
The display order of the parameter in the “Comparison Summary” tab and the
parameter order when printing
The display order of the parameter in the combo list of the “Result Trend” tab
1. Click the “Adjust Parameter Order” button, then the following box with the list of the
parameter order will pop up
The right of the parameter list displays the order-adjusting buttons: “Top”,
“Up”, “Down” and “Bottom”.
7-43
Reviewing Sample Results
Click the “Top” button to move the parameter to the top of the list.
Click the “Up” button to move the parameter upward by one position.
Click the “Down” button to move the parameter downward by one position.
Click the “Bottom” button to move the parameter to the bottom of the list
Print
When the current tab is “Comparison Summary”, you can click the “Print” button to print all
the contents of the summary.
When the current tab is “Result Trend”, you can click the “Print” button to print the displayed
result trend and the result list.
7-44
Reviewing Sample Results
7.9 Statistics
You can select or enter the “Statistical Condition” to realize the statistics of the workload.
The top of the screen displays the “Statistical Item”, namely, “Department”, “Deliverer”,
“Operator” and “Run Date”. Below the “Statistical Item”, it displays the corresponding
“Statistical Condition”. The corresponding records are displayed in the workload summary
together with the workload statistics. The available function buttons are also displayed in the
screen.
Workload Summary
All the records that match the statistical conditions will be displayed in the Workload
Summary and be included to calculate the total workload. The records of the same category
(i.e. those with the same field in the first column) will be taken to calculate the subtotal
workload. The default information fields in the workload summary are “Department”,
“Deliverer”, “Operator”, “Run Date” and “Sample Load”.
7-45
Reviewing Sample Results
For several results of the same item, they will be displayed in ascending order by run date. Be
aware that not all the statistics results in the summary are displayed in ascending order by
run date.
In the workload summary, the subtotal workload will be displayed below the records of the
same category, and the total workload will be displayed in the last.
Statistics
You can select the desired statistical item and enter the statistical condition to check the
statistics of the workload of the matched record.
1. You can select the desired statistical item by clicking the check box.
2. After selecting the statistical item, the corresponding statistical condition will be activated
for you to enter.
Enter the department into the “Department” box or select it from the “Department” pull-down
list.
Enter the name of the deliverer into the “Deliverer” box or select it from the “Deliverer”
pull-down list.
Enter the name of the operator into the “Operator” box or select it from the “Operator”
pull-down list.
Use the date control to specify the range for the run date.
When selecting “All”, it means all the available options of this field will be
calculated.
7-46
Reviewing Sample Results
3. Click the “Statistics” button to start calculating the workload for the matched records,
and then displays the result in the workload summary.
Re-fill
You can click the “Re-fill” button to clear all the entered statistical conditions and then re-fill.
You can use the “Adjust order” function to adjust the column order (including “Department”,
“Deliverer”, “Operator” and “Run Date”) in the workload summary. Do as follows:
1. Click the “Adjust Order” button, then the following box with the list of the column (field)
order will pop up.
The right of the list displays the order-adjusting buttons, namely, “Top”,
“Up”, “Down” and “Bottom”.
Click the “Top” button to move the field to the top of the list.
Click the “Up” button to move the field upward by one position.
Click the “Down” button to move the field downward by one position.
Click the “Bottom” button to move the field to the bottom of the list
7-47
Reviewing Sample Results
3. Click “Ok” to save the changes, close the box and back to the “Calculate Workload”
screen. Then, the column order in the workload summary refreshes.
The statistical result will also be refreshed together with the column order.
1. Click the “Print” button, and then a message box will pop up.
2. Click the “Print Summary” or the “Print statistics only” radio button to select the
content you want to print.
“Print Summary” means to print out all the records and statistical results in
the workload summary.
“Print statistics only” means to print out the statistical results only
(including subtotal and total).
7-48
Reviewing Sample Results
7.10 History
You can review the backed up sample data in “History”. Do as follows:
1. At the Table review screen, click the “Menu” button, and then select “Review”
“History”, and then a dialog box will pop up.
2. Select the desired directory and the file name, and then click “Open” to go to the
"Sample Data History" screen shown as follows.
7-49
Reviewing Sample Results
If you select another backed up file, then the history data displayed in the
list will be refreshed.
7-50
8 Using the QC Programs
8.1 Introduction
Quality Control (QC) consists of strategies and procedures that measure the precision and
stability of the analyzer. The results imply the reliability of the sample results. QC involves
measuring materials with known, stable characteristics at frequent intervals.
Analysis of the results with statistical methods allows the inference that sample results are
reliable. Mindray recommends you run the QC program daily with low, normal and high level
controls. A new lot of controls should be analyzed in parallel with the current lot prior to their
expiration dates. This may be accomplished by running the new lot of controls twice a day for
five days using any empty QC file. The QC files calculate the mean, standard deviation and
coefficient of variation for each selected parameter. The instrument-calculated means of
these ten runs should be within the expected ranges published by the manufacturer.
The analyzer provides 4 QC programs: L-J QC, X mean QC, X mean R QC and X-B QC.
You should only use the Mindray-specified controls and reagents. Store and
use the controls and reagents as instructed by instructions for use of the
controls and reagents.
8-1
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow, and the change will
be recorded into the system log.
Before analyzing the new lot of controls, you should set a QC file for each lot of controls and
you can edit the QC settings in the QC file by one of the following ways:
Manual entry
1. You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-2
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-3
Using the QC Programs
You can select the file No. within the range [1, 60].
Different QC files can not have the same lot No. and QC mode.
In the "AL-WB" QC mode, you can perform analysis to more than one tube
of control by selecting "Build-in Barcode Scanner".
5. Set QC ID: if you like to run control in the interval of running blood samples, you can set
a specific ID for the control. If such ID is recognized when the analyzer is running blood
samples, it will identify the sample as control automatically. When the running ends, the
result will be stored in the QC file of the corresponding ID.
8-4
Using the QC Programs
Letters, numbers and all characters that can be entered through the
keyboard (including special characters) are allowed for the QC ID, but the
number must end with a nonzero number. Chinese and other languages
(such as Japanese, Korean, etc) are not supported.
6. Click the “Read File” button, a message box will pop up for you to select the directory.
8. Click the “Ok” button to close the message box shown above, and the following
message box will pop up.
8-5
Using the QC Programs
The QC files for selection are displayed in the form of "Lot No. (Level)".
9. Click the “Ok” button to close the message box and go back to the "Read File" box. The
selected directory is displayed in the "Read From:" edit box.
10. In the "Read File" message box, select the "Read Target/Limits" check box, and click
“Ok” to read the selected QC information to the current QC file.
8-6
Using the QC Programs
If the “Read Target/Limits” is not selected, you have to enter the target and
limits manually.
The expiration date can not be earlier than the current system date.
Different QC files can not have the same lot No. and QC mode.
Manual entry
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-7
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-8
Using the QC Programs
You can select the file No. within the range [1, 60].
3. You can enter the lot No. of the controls by one of the following ways:
Manual entry
The lot No. can not be empty and up to 16 digits can be entered. You can
enter characters, numbers, letters and special characters, but no Chinese
characters are allowed.
Different QC files can not have the same lot No. and QC mode.
8-9
Using the QC Programs
You must enter the expiration date, and the entry range is [current system
date, 2099-12-31].
Different QC files can not have the same lot No. and QC mode.
7. Set QC ID: if you like to run control in the interval of running blood samples, you can set
a specific ID for the control. If such ID is recognized when the analyzer is running blood
samples, it will identify the sample as control automatically. When the running ends, the
result will be stored in the QC file of the corresponding ID.
Letters, numbers and all characters that can be entered through the
keyboard (including special characters) are allowed for the QC ID, but the
number must end with a nonzero number. Chinese and other languages
(such as Japanese, Korean, etc) are not supported.
8. According to the target list of the corresponding lot No., enter the target and limits into
the edit boxes of the parameters to be included in the QC run.
9. Click the “Save” button to save all the settings of the QC.
8-10
Using the QC Programs
If there are the saved preset values (Target and Limits) for the current level,
you can read the preset values into the current QC file. For details of
calculating and saving the preset values, see Section 8.2.3 Reviewing QC
results.
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-11
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-12
Using the QC Programs
You can select the file No. within the range [1, 60].
3. You can enter the lot No. of the controls by one of the following ways:
Manual entry
The lot No. can not be empty and up to 16 digits can be entered. You can
enter characters, numbers, letters and special characters, but no Chinese
characters allowed.
Different QC files can not have the same lot No. and QC mode.
8-13
Using the QC Programs
You must enter the expiration date, and the entry range is [current system
date, 2099-12-31].
Different QC files can not have the same lot No. and QC mode.
7. Set QC ID: if you like to run control in the interval of running blood samples, you can set
a specific ID for the control. If such ID is recognized when the analyzer is running blood
samples, it will identify the sample as control automatically. When the running ends, the
result will be stored in the QC file of the corresponding ID.
Letters, numbers and all characters that can be entered through the
keyboard (including special characters) are allowed for the QC ID, but the
number must end with a nonzero number. Chinese and other languages
(such as Japanese, Korean, etc) are not supported.
8. Click the “Have Preset Values” button to read the saved preset target and limits
(correspond to the current level) into the current QC file.
8-14
Using the QC Programs
9. Click the “Save” button to save all the settings of the QC.
Setting Limits
You can take the following steps to adjust the display format of the limits and the calculation
method of the preset limits.
1. Click the “Set Limits” button, and then the following message box will pop up.
2. Click “By SD” to display the limits in the form of absolute value; click “By CV” to display
the limits in the form of percentage.
3. If “By SD” is selected, click the “2SD” or “3SD” to select either double or triple standard
deviation to be the limits; if “By CV” is selected, click the “2CV” or “3CV” to select either
double or triple coefficient of variation to be the limits.
4. Click the “Ok” button to save all the settings for the limits.
Print
Click the “Print” button to print the setting information of the current QC file.
8-15
Using the QC Programs
Run controls in the interval of running blood samples under sample count screen
CT-WB
CT-PD
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
8-16
Using the QC Programs
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill
into your eyes, wash them off with plenty of water and immediately go see a
doctor.
If the sample mode is switched from the “Predilute” to “Whole Blood”, the
analyzer will perform the switching sequence automatically and a progress
bar will be displayed on the screen.
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-17
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
Be sure to use the adapter of Ф13x75 (mm) model when running the QC in
CT-WB mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
8-18
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “CT-WB” and the analysis status icon and analyzer indicator is
green.
2) Shake the prepared control as shown below to well mix it.
8-19
Using the QC Programs
7. When finish running, the QC results will be displayed in the current screen and be saved
in the QC file automatically.
8-20
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-21
Using the QC Programs
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-22
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
Be sure to use the adapter of Ф11x40 (mm) model when running the QC in
CT-PD mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
8-23
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “CT-PD” and the analysis status icon and analyzer indicator is
green.
2) Click the shortcut button “Diluent”, the compartment door will open automatically. A
message box shown below will pop up.
8-24
Using the QC Programs
3)Present a clean centrifugal tube into the sample compartment with the cap opened. Press
the [RUN] key to dispense 180μL of diluent into the tube. During dispensing the diluent, a
progress bar will display.
If the current mode is autoloader, the analyzer will switch to the close-tube
mode automatically and then dispense the diluent. Then, the mode will be
switched back to autoloader when exiting.
4)When the dispensing is finished, the compartment door will open automatically and you can
remove the centrifugal tube. Then, the following message box will display.
5) Add 20μL of control to the diluent, close the tube cap and shake the tube to mix the
sample.
6) Click the “Cancel” button to exit the “Diluent” message box.
8-25
Using the QC Programs
9) Present the centrifugal tube into the tube holder with the cap opened and then close the
compartment door.
10) Press the [Run] key or click the “Run” button on the message box to start running.
11) When the running is finished, the compartment door will open automatically and you can
remove the centrifugal tube.
7. When finish running, the QC results will be displayed in the current screen and be saved
in the QC file automatically.
You can also dispense 180μL of diluent by pipette into the tube.
After mixing the control with the diluent, be sure to wait 3 minutes before
running.
Be sure to run the prediluted samples within 30 minutes after the mixing.
Be sure to mix any sample that has been prepared for a while before running
it.
8-26
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
The pusher will push the rack inside the autoloader. Be sure you hand is
away from the rack before starting the autoloader.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-27
Using the QC Programs
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-28
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-29
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “AL-WB” and the analysis status icon and analyzer indicator is
green.
2) Click the “Start” button.
8-30
Using the QC Programs
4) Place racks loading controls on the level of the right tray of the autoloader, with the back of
“MINDRAY” mark on the carrier facing the analyzer.
5) Press the [Run] key or click the “Run” button on the message box to start running.
6) When finish running, you can remove the racks from the left of the autoloader.
Without built-in barcode scanner, only one control can be run in single
running.
7. When finish running, the QC results will be displayed in the current screen and be saved
in the QC file automatically.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
8-31
Using the QC Programs
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
The pusher will push the rack inside the autoloader. Be sure you hand is
away from the rack before starting the autoloader.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-32
Using the QC Programs
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-33
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
2. Select the QC file No. to be run; the screen displays the corresponding file information.
As long as the QC mode of the select QC file is “AL-WB” and “With Built-in
Barcode Scanner” is also selected, even the lot No. of the current QC file
does not match the lot No. of the control, you can still save each QC result
into the corresponding QC file by the scanned barcode.
3. Be sure that the level of the control to be run is the same with the current QC file.
8-34
Using the QC Programs
For details to paste the barcode, see Chapter 6 Operating Your Analyzer.
Be sure that the barcode of the lot No. of the controls is readable by the
built-in barcode scanner, for details to set the symbology, see Chapter 5
Customizing the Analyzer Software; for details to test the built-in barcode
scanner, see Chapter 10 Maintenance.
1) Make sure the QC mode is “AL-WB” and the analysis status icon and analyzer indicator is
green.
2) Be sure “Built-in Barcode Scanner” is selected for all the corresponding QC files, and the
barcode labels are pasted to the containers of the controls.
3) Click the “Start” button.
With built-in barcode scanner, more than one control can be run together in
a single running.
When the barcode of the control is scanned by the built-in barcode scanner,
the screen will switch to the corresponding QC file automatically.
If nothing is scanned or no matching lot No. is found, the analyzer will jump
over the certain control without analyzing, and proceed to the later barcode
scanning and QC analyzing.
8-35
Using the QC Programs
8. When finish running, the newest QC results will be displayed in the current screen and
be saved in the QC file automatically.
If any control is jumped over without being analyzed, a prompt will pop up
when the analysis cycle is completed. Click the “Ok” button to close the
prompt.
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to switch to the pervious or the next QC result.
You can click the button or button to switch to the earliest or the latest QC result
saved in the QC file.
The running result of the expired control will begin with an “O” mark.
The flag prompt “H” or “L” will appear in front of the result that out of the
limits.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
Click the “Print” button to print the results of the current QC Run screen.
8-36
Using the QC Programs
After setting special “QC ID” for controls under the “Settings” screen, you can run controls in
the intervals of running blood samples under sample count screen.
Before running blood samples, when you are editing worklist or entering information for the
next sample in the “Run” message box, be sure to enter the special “QC ID” as “sample ID”.
(Refer to 6.6.1 or 6.7.1 Entering Work List Information for the way to enter sample IDs)
You can start one of the following QC analyses according to the selected QC mode.
CT-WB
CT-PD
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
8-37
Using the QC Programs
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill
into your eyes, wash them off with plenty of water and immediately go see a
doctor.
If the sample mode is switched from the “Predilute” to “Whole Blood”, the
analyzer will perform the switching sequence automatically and a progress
bar will be displayed on the screen.
2. When it is ready to run (i.e. the analysis status icon and analyzer indicator is green),
shake the prepared control as shown below to well mix it.
8-38
Using the QC Programs
4. Present the prepared control into the tube holder and then close the compartment door.
5. Press the [Run] key or click the “Run” button on the message box to start running.
6. When the running is finished, the compartment door will open automatically and you can
remove the control.
7. When finish running, the QC results will be automatically saved in the corresponding QC
file of the QC ID
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
8-39
Using the QC Programs
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
2. Make sure the analysis status icon and analyzer indicator is green, and then click the
shortcut button “Diluent”, the compartment door will open automatically. A message box
shown below will pop up.
8-40
Using the QC Programs
3. Present a clean centrifugal tube into the sample compartment with the cap opened.
Press the [RUN] key to dispense 180μL of diluent into the tube. During dispensing the
diluent, a progress bar will display.
If the current mode is autoloader, the analyzer will switch to the close-tube
mode automatically and then dispense the diluent. Then, the mode will be
switched back to autoloader when exiting.
4. 4)When the dispensing is finished, the compartment door will open automatically and
you can remove the centrifugal tube. Then, the following message box will display.
5. Add 20μL of control to the diluent, close the tube cap and shake the tube to mix the
sample.
8-41
Using the QC Programs
8. Present the centrifugal tube into the tube holder with the cap opened and then close the
compartment door.
9. 1Press the [Run] key or click the “Run” button on the message box to start running.
10. 1When the running is finished, the compartment door will open automatically and you
can remove the centrifugal tube.
11. When finish running, the QC results will be automatically saved in the corresponding QC
file of the QC ID.
You can also dispense 180μL of diluent by pipette into the tube.
After mixing the control with the diluent, be sure to wait 3 minutes before
running.
Be sure to run the prediluted samples within 30 minutes after the mixing.
Be sure to mix any sample that has been prepared for a while before running
it.
8-42
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
The pusher will push the rack inside the autoloader. Be sure you hand is
away from the rack before starting the autoloader.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-43
Using the QC Programs
2. Make sure the analysis mode is “AL-WB” and the analysis status icon and analyzer
indicator is green.
4. Place racks loading controls on the level of the right tray of the autoloader, with the back
of “MINDRAY” mark on the carrier facing the analyzer.
5. Press the [Run] key or click the “Run” button on the message box to start running.
6. When finish running, you can remove the racks from the left of the autoloader.
Without built-in barcode scanner, only one control can be run in single
running.
7. When finish running, the QC results will be automatically saved in the corresponding QC
file of the QC ID.
8-44
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
The pusher will push the rack inside the autoloader. Be sure you hand is
away from the rack before starting the autoloader.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-45
Using the QC Programs
For details to paste the barcode, see Chapter 6 Operating Your Analyzer.
Be sure that the barcode of the lot No. of the controls is readable by the
built-in barcode scanner, for details to set the symbology, see Chapter 5
Customizing the Analyzer Software; for details to test the built-in barcode
scanner, see Chapter 10 Maintenance.
3. Make sure the analysis mode is “AL-WB” and the analysis status icon and analyzer
indicator is green.
4. Be sure “Built-in Barcode Scanner” is selected for all the corresponding QC files, and the
barcode labels are pasted to the containers of the controls.
6. Place racks loading controls on the level of the right tray of the autoloader, with the back
of “MINDRAY” mark on the carrier facing the analyzer.
7. Press the [Run] key or click the “Run” button on the message box to start running.
8. When finish running, you can remove the racks from the left of the autoloader.
8-46
Using the QC Programs
With built-in barcode scanner, more than one control can be run together in
a single running.
When the barcode of the control is scanned by the built-in barcode scanner,
the screen will switch to the corresponding QC file automatically.
If nothing is scanned or no matching lot No. is found, the analyzer will jump
over the certain control without analyzing, and proceed to the later barcode
scanning and QC analyzing.
9. When finish running, the QC results will be automatically saved in the corresponding QC
file of the QC ID.
If any control is jumped over without being analyzed, a prompt will pop up
when the analysis cycle is completed. Click the “Ok” button to close the
prompt.
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to switch to the pervious or the next QC result.
You can click the button or button to switch to the earliest or the latest QC result
saved in the QC file.
8-47
Using the QC Programs
The running result of the expired control will begin with an “O” mark.
The flag prompt “H” or “L” will appear in front of the result that out of the
limits.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
Click the “Print” button to print the results of the current QC Run screen.
8-48
Using the QC Programs
Graph
Table
Graph Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-49
Using the QC Programs
1. Select the QC file No. you want to review, and then the screen will display the
corresponding information and the graph.
2. You can drag the scroll bar on the right of the graph to browse the desired graph of the
parameter. You can drag the scroll bar down to the graph horizontally to browse all the
QC results.
8-50
Using the QC Programs
1- The Mean, SD and CV% of all the QC results of each parameter in the current graph.
2- The saving date and time of the QC points located on the green line.
3- The operator who run the QC analysis and obtained the QC points located on the green
line.
4- The QC results of the parameters that correspond to the QC points located on the green
line.
5- The QC points in each graph are displayed from left to right according to the sequence
from the earliest to the latest. The QC points are connected by a line to illustrate the
distribution trend.
6- The QC point corresponds to each QC result. Only the selected QC point displays its value
under the parameter. The black QC point indicates the value is within the limit; the red QC
point indicates the value is out of the limit.
7- When you clicking a QC point in the graph, the QC points of other parameters that saved
together with this one will be marked by a green line.
8- The relative position of the QC point located on the green line and the total QC points
saved currently.
The outliers are excluded from the calculation of Mean, SD and CV%.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
8-51
Using the QC Programs
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to move the green line to the pervious or the
next QC point; you can click the button or button to move the green line to the first
or the last QC point in the graph. When the location of the green line is selected, you can
check the QC results of the QC points located on the green line under each parameter.
New Vial
If the reviewed QC results are obtained by analyzing a new vial of control of the same batch,
you should mark the QC points of the new vial to distinguish the QC results from the old.
1. Move the green line to the first QC point of the new vial.
2. Click the “New Vial” button, and then a blue line appears at the QC point of the new vial.
3. After another new vial of control (of the same batch) is run and its QC results are saved,
you can continue to mark the current QC point of the new vial according to step 1 and 2.
4. If the current QC point is marked with the blue line, the “New Vial” button will turn into
“Cancel”; you can click the button to remove the blue line, and then the “Cancel” button
will turn back to “New Vial”.
Data Compare
If you wish to compare the graphs of the certain parameter obtained by running controls of
different lot No., do as follows:
1. Click the “Data Compare” button to start selecting the desired graph.
8-52
Using the QC Programs
2. Select the desired QC file No. into the “File No.” box (3 files can be selected at most).
Then, the graph of the selected QC file will be displayed below together with its lot No.,
QC mode and level.
Browse the graph here the same way as instructed in the “Graph” screen.
Be aware that, for controls of different level, their graphs will be
distinguished by the color of orange, black and blue.
Display Order
1. Click the “Display Order” button to check the current display order of the graphs.
8-53
Using the QC Programs
3. You can click the button or button to move parameter upward or downward;
you can click the button or button to move the parameter to the first or the last
position.
4. Click the “Ok” button to refresh the display order of the graphs.
If there are 3 or more than 3 QC results within the limits obtained for the parameters, you can
take the following steps to calculate and save the preset value for each parameter:
1. Click the “Calculate Preset Values” button, and then the screen displays two lines for
you to select the range for calculating the preset values.
2. Click and drag the two lines respectively to locate them at the beginning and the ending
of the range for calculating the preset values.
3. The Mean, SD and CV% (on the right of the graph) will change into the new results
which obtained by calculating the selected range.
4. If you wish to save the new results, you can click the “Save Preset Values” button to
save the current Mean, SD and CV% as the preset values for the corresponding level
8-54
Using the QC Programs
(high/normal/low). Then, the two selecting lines disappear and the Mean, SD and CV%
return to the calculated results of all QC results.
The calculation and display of the preset values are only available to the
parameter (within the calculation range) which has 3 or more than 3 results
within the limit. Otherwise, the display of the preset values will be empty.
According to the high, normal and low level of the controls, three sets of
preset values can be saved respectively.
1. After moving the green line to the desired QC point, you can click the “Outliers” button to
display the QC results, targets and limits of all the parameters located on the green line
(the QC results exceed the limit will be displayed in red) and enter the reasons for the
outliers.
2. You can select the reason form the given ones or enter the reasons (up to 200
characters) into the edit box manually after selecting “Others”.
3. Click the “Ok” button to save the reasons for the outliers and exit.
8-55
Using the QC Programs
If you enter the reason for the group of QC points whose results are actually
within the limits, then their corresponding QC data both in the QC Graph and
QC Table will be displayed in the color of red. And the data will return in the
color of black if you cancel the reason and then save the changes.
Delete
1. If you wish to delete a single QC result, move the green line to the desired QC result; if
you wish to delete all the data, perform step 2 directly.
4. Click the “Ok” button and then confirm to delete the selected data.
Click the “Print” button to print all the file information and graphs of the parameters of the
current QC file.
The green line and the corresponding values of the QC points will not be
printed.
8-56
Using the QC Programs
Table Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
You can enter the Table screen in one of the following ways:
Click the “Menu” button on the screen; then select “QC””L-J” on the pop-up menu.
8-57
Using the QC Programs
8-58
Using the QC Programs
2. Select the QC file No. you want to review, and then the screen will display the
corresponding information and the table.
3. You can drag the scroll bar on the right of the table vertically to browse the desired table
of the parameter. You can drag the scroll bar down to the table horizontally to browse all
the QC results.
1- The No. of the QC result saved in the QC file (arranged from left to right in the order that
from the earliest to the latest)
2- QC Result
3- QC parameters (displayed in the same order as the Graph screen)
4- QC flag: The flag “H” or “L” will be used to prompt the result that out of the limits
5- The relative position of the highlighted QC point and the total QC points saved currently.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
8-59
Using the QC Programs
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to highlight the pervious or the next QC result;
you can click the button or button to highlight the first or the last QC result in the
table.
Delete
1. If you wish to delete a single QC result, click the column contains the desired QC result;
if you wish to delete all the data, perform step 2 directly.
4. Click the “Ok” button and then confirm to delete the selected data.
1. Click the “Print” button, and then you can select “All Data” or “Specified Data” to be
printed.
8-60
Using the QC Programs
2. Click the “All Data” button and then click the “Ok” button to print all the file information
and tables of the parameters of the current QC file; after clicking the “Specified Data”
button, you can select the starting and finishing date (the saved date of the QC results)
for printing, then click the “Ok” button to print the specified data.
Communication
If you wish to transmit the QC data to the external data management software or LIS/HIS, do
as follows:
1. Click the “Communication” button, and then you can select “All Data” or “Specified
Data” to be transmitted.
2. Click the “All Data” button and then click the “Start” button to transmit the information of
the current QC file and QC data. After clicking the “Specified Data” button, you can
specify a date range (the date when the QC result was saved), then click the “Ok” button
to transmit the specified data. The “LIS/HIS connected” icon on the status bar of the
8-61
Using the QC Programs
3. While transmitting, the “Start” button in the pop-up message box will be replaced by
Data Backup
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the data backup function, you
should log on the system as administrator, and then right click the shortcut
icon of the terminal software and run it as administrator.
If you wish to backup the information and the result of the current QC file, do as follows:
1. Click the “Data Backup” button, a message box will pop up.
2. Select the backup directory (the default directory is the folder of “QC Data” under the
8-62
Using the QC Programs
3. Enter the name for the backup data (the default name is [L-J_QC_date saved_time
saved]).
5. When the backup is finished, a message box will pop up, and then click “Ok” to exit.
The backup data can not be modified. You can review the data in “History”
screen.
You can click the “Cancel” button to cancel backup when it is in process.
Data Export
If you wish to export the information and the result of the current QC file, do as follows:
1. Click the “Data Export” button, and then a message box will pop up.
8-63
Using the QC Programs
2. Select the export directory (the default directory is the folder of “QC Data” under the
installation location of the terminal software).
3. Enter the name for the export data (the default name is [L-J_QC_date saved_time
saved]).
6. When the export is finished, a message box will pop up, and then click “Ok” to exit.
You can modify the exported data but can not review the data in the
“History” screen.
You can click the “Cancel” button to cancel export when it is in process.
8-64
Using the QC Programs
History
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the history function, you should
log on the system as administrator, and then right click the shortcut icon of
the terminal software and run it as administrator.
1. Click the “History” button, and then a message box will pop up.
8-65
Using the QC Programs
4. The display format and the function button are the same as the QC Table review screen.
Browse, print, transmit and export the data as instructed in the section of QC Table
Review.
8-66
Using the QC Programs
The single para. QC graph screen consists of the file information area, QC graph area and
the comments area.
You can select among all valid QC file No. from the “File No.” pull-down list. The default file
No. is the L-J QC file used by the current operator, and the file No. cannot be null.
The lot No, level, exp. date, QC mode, editor and QC sample ID will be displayed
automatically after you select the QC file No..
QC graph area
You can select the parameter to plot QC graph from the “Parameter” pull-down list, and
select the plotting mode from the “Plotting” pull down list: “By QC Date”, “1/Day”, “All”.
The default values of the “Mean” and “SD” boxes are calculated from all the valid QC points
on the graph of the selected parameter in the current QC file by the default plotting mode. You
can edit the values manually.
8-67
Using the QC Programs
If there are less than 3 valid QC points of the current parameter, the mean
and SD will be null.
If you select “By QC Date” or “1/Day” plotting mode, and there are more
than 1 QC point in one day, the points will all be displayed in the graph, but
only the latest one will be the connecting point of the plot.
Drag the scroll bar after setting up all the information, and the QC graph will be plotted
automatically and the analysis results and operator information will be displayed under the
graph.
Comments area
This area displays the conclusion drawn based on the preset rule of outliers, which can be
edited by operators.
Click the arrow buttons under the screen to browse all the QC points.
You can click the button or button to highlight the previous or the next QC point;
you can click the button or button to highlight the first or the last QC point in the
graph.
New Vial
If the reviewed QC results are obtained by analyzing a new vial of control within the same
batch, you should mark the QC points of the new vial to distinguish the QC results from the
old.
1. Move the green line to the first QC point of the new vial.
2. Click the “New Vial” button, and then a blue line appears at the QC point of the new vial.
3. After another new vial of control (within the same batch) is run and its QC results are
saved, you can continue to mark the current QC point of the new vial according to step 1
and 2.
4. If the current QC point is marked with the blue line, the “New Vial” button will turn into
“Cancel”; you can click the button to remove the blue line, and then the “Cancel” button
8-68
Using the QC Programs
Calculate Target
Click the “Calculate” button, the following screen will pop up.
After selecting the date range, the mean, SD and CV will be calculated and displayed in to the
corresponding boxes automatically.
Click “Apply” to enter the calculated target into the single para. QC graph screen, and then
click “OK” to save the target and close the dialog box. Click “Cancel” to exit the “Calculate”
screen without saving the calculated target.
If there are no data in the selected date range, a prompt will be displayed:
“Not enough data! Reselect a date range and make sure there are more than
5 groups of valid data.”
Save
Click “Save” to save the mean and SD (including the manually edited values and the values
obtained by calculating target) and display them on the bottom left of the QC graph.
Monthly QC Graph
Monthly QC graph is the QC graph of one parameter of the high, normal and low levels within
one month.
Click “Menu” “QC” ”L-J QC”, and then click the “Monthly QC Graph” tab to enter the
following screen.
8-69
Using the QC Programs
The monthly QC graph screen consists of the file information area and QC graph area.
You can select the parameter to plot QC graph from the “Parameter” pull-down list; select the
level(s) of the control you want to check from the “Level” pull down list; select the plotting
mode from the “Plotting” pull down list: “1/Day”, “All”; and select the date range of all QC
data in the QC graph from the “Date Range” pull-down list.
After you set the above conditions, the matching QC files will be acquired and the Lot No.,
exp. Date, X mean and SD will be displayed in the boxes of the corresponding QC levels. If
no data is acquired, the boxes will be null.
If the “X Mean” and “SD” boxes are both null, their values cannot be edited, otherwise you
can edit the values of X mean and SD, then click “Save” to save the change.
Two QC files of each QC level can be displayed in one month; if there are
more than two QC files, the latest two will be displayed.
QC graph area
8-70
Using the QC Programs
Drag the scroll bar after setting up all the information, the QC graph will be plotted
automatically and the QC point No., analysis results and operator information will be
displayed under the QC graph. If there are QC points in the graph, the Lot No., X mean, SD
and CV% will be displayed above the graph.
Each screen of the QC graph can display 31 groups of QC data. To view more QC data, drag
the horizontal scroll bar.
Click the arrow button under the screen to browse all the QC points.
You can click the button or button to highlight the previous or the next QC point;
you can click the button or button to highlight the first or the last QC point in the
graph.
8-71
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
For the QC files with saved QC results, if changing and then saving the
reference range or limits caused the change of target and limits, then the
changed data will be displayed in yellow background and recorded in the
system log.
Before analyzing a new batch of controls, you should set a QC file for each lot of controls and
edit the QC settings in the QC file by one of the following ways:
Manual entry
1. You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen; then select “QC””X mean” on the pop-up menu.
8-72
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-73
Using the QC Programs
You can select the file No. within the range [1, 60].
Different QC files can not have the same lot No. and QC mode.
5. Click the “Read File” button, a message box will pop up for you to select the directory.
8-74
Using the QC Programs
7. Click the “Ok” button to close the message box shown above, and the following
message box will pop up.
8-75
Using the QC Programs
The QC files for selection are displayed in the form of "Lot No. (Level)".
8. Click the “Ok” button to close the message box and go back to the "Read File" box. The
selected directory is displayed in the "Read From:" edit box.
9. In the "Read File" message box, select the "Read Target/Limits" check box, and click
“Ok” to read the selected QC information to the current QC file.
8-76
Using the QC Programs
If the “Read Target/Limits” is not selected, you have to enter the target and
limits manually.
The expiration date can not be earlier than the current system date.
Different QC files can not have the same lot No. and QC mode.
Manual Entry
You can enter the graph screen by one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean” on the pop-up
menu.
8-77
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-78
Using the QC Programs
You can select the file No. within the range [1, 60].
3. You can enter the lot No. of the controls in one of the following ways:
Manual entry
The lot No. can not be empty and up to 16 digits can be entered. You can
enter characters, numbers, letters and special characters, but no Chinese
characters allowed.
Different QC files can not have the same lot No. and QC mode.
8-79
Using the QC Programs
You must enter the expiration date, and the entry range is [current system
date, 2099-12-31].
Different QC files can not have the same lot No. and QC mode.
7. According to the target list of the corresponding lot No., enter the target and limits into
the edit boxes of the parameters to be included in the QC run.
8. Click the “Save” button to save all the settings of the QC.
8-80
Using the QC Programs
If there are the saved preset values (Target and Limits) for the current level,
you can read-in the preset values into the current QC file. For details of
calculating and saving the preset values, see Section 8.3.3 Reviewing QC
results.
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean” on the pop-up
menu.
8-81
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-82
Using the QC Programs
You can select the file No. within the range [1, 60].
3. You can enter the lot No. of the controls in one of the following ways:
Manual entry
The lot No. can not be empty and up to 16 digits can be entered. You can
enter characters, numbers, letters and special characters, but no Chinese
characters allowed.
Different QC files can not have the same lot No. and QC mode.
8-83
Using the QC Programs
You must enter the expiration date, and the entry range is [current system
date, 2099-12-31].
Different QC files can not have the same lot No. and QC mode.
7. Click the “Have Preset Values” button to read-in the saved preset target and limits
(correspond to the current level) into the current QC file.
8. Click the “Save” button to save all the settings of the QC.
Setting Limits
Do as followings to adjust the display format of the limits and the calculation method of the
preset limits.
1. Click the “Set Limits” button, and then the following message box will pop up.
8-84
Using the QC Programs
2. Click “By SD” to display the limits in the form of absolute value; click “By CV” to display
the limits in the form of percentage.
3. If “By SD” is selected, click the “2SD” or “3SD” to select either double or triple standard
deviation to be the limits; if “By CV” is selected, click the “2CV” or “3CV” to select either
double or triple coefficient of variation to be the limits.
4. Click the “Ok” button to save all the settings for the limits.
Print
Click the “Print” button to print the setting information of the current QC file.
8-85
Using the QC Programs
CT-WB
CT-PD
AL-WB
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-86
Using the QC Programs
If the sample mode is switched from the “Predilute” to “Whole Blood”, the
analyzer will perform the switching sequence automatically and a progress
bar will be displayed on the screen.
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean” on the pop-up
menu.
8-87
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
Be sure to use the adapter of Ф13x75 (mm) model when running the QC in
CT-WB mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
8-88
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “CT-WB” and the analysis status icon and analyzer indicator is
green.
2) Shake the prepared control as shown below to well mix it.
8-89
Using the QC Programs
9) Mix the control well again, and then to run the control for the second time according to the
prompt.
8-90
Using the QC Programs
You can click the “Cancel” button in the message box to cancel the second
run and the results obtained in the first run will not be saved as well.
10) When the running is finished, the compartment door will open automatically and you can
remove the control.
7. When finish running, the QC results (values of the two QC runs and the X mean) will be
displayed in the current screen and be saved in the QC file automatically.
When the QC result of the second QC run is obtained, the screen will refresh
the displayed histograms and scattergrams according to the second QC
run.
8-91
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-92
Using the QC Programs
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean” on the pop-up
menu.
8-93
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
Be sure to use the adapter of Ф11x40 (mm) model when running the QC in
CT-PD mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
8-94
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “CT-PD” and the analysis status icon and analyzer indicator is
green.
2) Click the shortcut button “Diluent”, the compartment door will open automatically. A
message box shown below will pop up.
8-95
Using the QC Programs
3) Present a clean centrifugal tube into the sample compartment with the cap opened. Press
the [RUN] key to dispense 180μL of diluent into the tube. During dispensing the diluent, a
progress bar will display.
If the current mode is autoloader, the analyzer will switch to the close-tube
mode automatically and then dispense the diluent. Then, the mode will be
switched back to autoloader when exiting.
4) When the dispensing is finished, the compartment door will open automatically and you
can remove the centrifugal tube. Then, the following message box will display.
8-96
Using the QC Programs
10) Present the centrifugal tube into the tube holder with the cap opened and then close the
compartment door.
11) Press the [Run] key or click the “Run” button on the message box to start running.
12) When the running is finished, the compartment door will open automatically and you can
remove the centrifugal tube.
13) After the analyzing is finished, the QC result of the first analysis will be displayed on the
screen.
14) Add 20μL of control to the other centrifugal tube with diluent, close the tube cap and
shake the tube to mix the sample, and then present the centrifugal tube into the tube holder
with the cap opened to start the second running.
You can click the “Cancel” button in the message box to cancel the second
run and the results obtained in the first run will not be saved as well.
15) When the running is finished, the compartment door will open automatically and you can
remove the centrifugal tube.
7. When finish running, the QC results (values of the two QC runs and the X mean) will be
displayed in the current screen and be saved in the QC file automatically.
8-97
Using the QC Programs
You can also dispense 180μL of diluent by pipette into the tube.
After mixing the control with the diluent, be sure to wait 3 minutes before
running.
Be sure to run the prediluted samples within 30 minutes after the mixing.
Be sure to mix any sample that has been prepared for a while before running
it.
When the QC result of the second QC run is obtained, the screen will refresh
the displayed histograms and scattergrams according to the second QC
run.
8-98
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
The pusher will push the rack inside the autoloader. Be sure your hand is
away from the rack before starting the autoloader.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-99
Using the QC Programs
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean” on the pop-up
menu.
8-100
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-101
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “AL-WB” and the analysis status icon and analyzer indicator is
green.
2) Click the “Start” button.
8-102
Using the QC Programs
4) Place racks loading controls on the level of the right tray of the autoloader, with the back of
“MINDRAY” mark on the carrier facing the analyzer.
5) Press the [Run] key or click the “Run” button on the message box to start running.
6) When the racks are pushed out, you can remove the racks from the left of the autoloader.
7) After the analyzing is finished, the QC result of the first analysis will be displayed on the
screen.
8) Present the prepared control into the rack according to the prompt, and then place racks
loading controls on the level of the right tray of the autoloader, with the back of “MINDRAY”
mark on the carrier facing the analyzer, and then start the second running.
You can click the “Cancel” button in the message box to cancel the second
run and the results obtained in the first run will not be saved as well.
9) When the racks are pushed out, you can remove the racks from the left of the autoloader.
7. When finish running, the QC results (values of the two QC runs and the X mean) will be
displayed in the current screen and be saved in the QC file automatically.
When the QC result of the second QC run is obtained, the screen will refresh
the displayed histograms and scattergrams according to the second QC
run.
8-103
Using the QC Programs
You can click the button or button to switch to the pervious or the next QC result.
You can click the button or button to switch to the earliest or the latest QC result
saved in the QC file.
The running result of the expired control will begin with an “O” mark.
The flag prompt “H” or “L” will appear in front of the X mean that out of the
limits.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
Print
Click the “Print” button to print the results of the current QC Run screen.
Graph
Table
Graph Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
8-104
Using the QC Programs
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC”” X mean” on the pop-up
menu.
1. Select the QC file No. you want to review, and then the screen will display the
corresponding information and the graph.
8-105
Using the QC Programs
2. You can drag the scroll bar on the right of the graph to browse the desired graph of the
parameter. You can drag the scroll bar down to the graph horizontally to browse all the
QC results.
1- The Mean, SD and CV% of all the QC results of each parameter in the current graph.
2- The saving date and time of the QC points located on the green line.
3- The operator who run the QC analysis and obtained the QC points located on the green
line.
4- The QC results of the parameters that correspond to the QC points located on the green
line.
5- The QC points in each graph are displayed from left to right according to the sequence
from the earliest to the latest. The QC points are connected by a line to illustrate the
distribution trend.
6- The QC point corresponds to each QC result. Only the selected QC point displays its value
under the parameter. The black QC point indicates the value is within the limit; the red QC
point indicates the value is out of the limit.
7- When you clicking a QC point in the graph, the QC points of other parameters that saved
together with this one will be marked by a green line.
8- The relative position of the QC point located on the green line and the total QC points
saved currently.
8-106
Using the QC Programs
The outliers are excluded from the calculation of Mean, SD and CV%.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to move the green line to the pervious or the
next QC point; you can click the button or button to move the green line to the first
or the last QC point in the graph. When the location of the green line is selected, you can
check the QC results of the QC points located on the green line under each parameter.
New Vial
If the reviewed QC results are obtained by analyzing a new vial of control of the same batch,
you should mark the QC points of the new vial to distinguish the QC results from the old.
1. Move the green line to the first QC point of the new vial.
2. Click the “New Vial” button, and then a blue line appears at the QC point of the new vial.
3. After another new vial of control (of the same batch) is run and its QC results are saved,
you can continue to mark the current QC points of the new vial according to step 1 and 2.
4. If the current QC point is marked with the blue line, the “New Vial” button turns into
“Cancel”, you can click the button to remove the blue line, then the “Cancel” button will
turn back to “New Vial”.
Data Compare
If you wish to compare the graphs of the certain parameter obtained by running controls of
different lot No., do as follows:
1. Click the “Data Compare” button to start selecting the desired graph.
8-107
Using the QC Programs
2. Select the desired QC file No. into the “File No.” box (3 files can be selected at most).
Then, the graph of the selected QC file will be displayed below together with its lot No.,
QC mode and level.
Browse the graph here the same way as instructed in the “Graph” screen.
Be aware that, for controls of different level, their graphs will be
distinguished by the color of orange, black and blue.
Display Order
1. Click the “Display Order” button to check the current display order of the graphs.
8-108
Using the QC Programs
3. You can click the button or button to move parameter upward or downward;
you can click the button or button to move the parameter to the first or the last
position.
4. Click the “Ok” button to refresh the display order of the graphs.
If there are 3 or more than 3 QC results within the limits obtained for the parameters, you can
take the following steps to calculate and save the preset value for each parameter:
1. Click the “Calculate Preset Values” button, and then the screen displays two lines for
you to select the range for calculating the preset values.
2. Click and drag the two lines respectively to locate them at the beginning and the ending
of the range for calculating the preset values.
3. The Mean, SD and CV% (on the right of the graph) will change into the new results
which obtained by calculating the selected range.
4. If you wish to save the new results, you can click the “Save Preset Values” button to
save the current Mean, SD and CV% as the preset values for the corresponding level
8-109
Using the QC Programs
(high/normal/low).Then, the two selecting lines disappear and the Mean, SD and CV%
return to the calculated results of all QC results.
The calculation and display of the preset values are only available to the
parameter (within the calculation range) which has 3 or more than 3 results
within the limit. Otherwise, the display of the preset values will be empty.
According to the high, normal and low level of the controls, three sets of
preset values can be saved respectively.
1. After moving the green line to the desired QC point, you can click the “Outliers” button to
display the QC results, targets and limits of all the parameters located on the green
line(the QC results exceed the limit will be displayed in red) and enter the reasons for the
outliers.
2. You can select the reason form the given ones or enter the reasons (up to 200
characters) into the edit box manually after selecting “Others”.
3. Click the “Ok” button to save the reasons for the outliers and exit.
8-110
Using the QC Programs
If you enter the reason for the group of QC points whose results are actually
within the limits, then their corresponding QC data both in the QC Graph and
QC Table will be displayed in the color of red. And the data will return in the
color of black if you cancel the reason and then save the changes.
Delete
1. If you wish to delete a single QC result, move the green line to the desired QC result; if
you wish to delete all the data, perform step 2 directly.
4. Click the “Ok” button and then confirm to delete the selected data.
Click the “Print” button to print all the file information and graphs of the parameters of the
current QC file.
The green line and the corresponding values of the QC points will not be
printed.
8-111
Using the QC Programs
Table Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
You can enter the Table screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC”” X mean” on the pop-up
menu.
8-112
Using the QC Programs
8-113
Using the QC Programs
2. Select the QC file No. you want to review, and then the screen will display the
corresponding information and the table.
3. You can drag the scroll bar on the right of the table vertically to browse the desired table
of the parameter. You can drag the scroll bar down to the table horizontally to browse all
the QC results.
1- The No. of the QC result saved in the QC file (arranged from left to right in the order that
from the earliest to the latest)
2- QC Result
3- QC parameters (displayed in the same order as the Graph screen)
4- QC flag: The flag “H” or “L” will be used to prompt the result (X mean) that out of the limits
5- The relative position of the highlighted QC point and the total QC points saved currently.
For the QC files with saved QC results, if any change is made to the target or
the limits, the changed data will be highlighted in yellow.
8-114
Using the QC Programs
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to highlight the pervious or the next QC result;
you can click the button or button to highlight the first or the last QC result in the
table.
Delete
1. If you wish to delete a single QC result, click the column contains the desired QC result;
if you wish to delete all the data, perform step 2 directly.
4. Click the “Ok” button and then confirm to delete the selected data.
1. Click the “Print” button, and then you can select “All Data” or “Specified Data” to be
printed.
8-115
Using the QC Programs
2. Click the “All Data” button and then click the “Ok” button to print all the file information
and tables of the parameters of the current QC file; after clicking the “Specified Data”
button, you can select the starting and finishing date (the saved date of the QC results)
for printing, then click the “Ok” button to print the specified data.
Communication
If you wish to transmit the QC data to the external data management software or LIS/HIS, do
as follows:
1. Click the “Communication” button, and then you can select “All Data” or
2. Click the “All Data” button and then click the “Start” button to transmit the
information of the current QC file and QC data. After clicking the “Specified Data”
button, you can specify a date range (the date when the QC result was saved), then
click the “Ok” button to transmit the specified data. The “LIS/HIS connected” icon on
8-116
Using the QC Programs
3. While transmitting, the “Start” button in the pop-up message box will be replaced by
Data Backup
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the data backup function, you
should log on the system as administrator, and then right click the shortcut
icon of the terminal software and run it as administrator.
If you wish to backup the information and the result of the current QC file, do as follows:
1. Click the “Data Backup” button, a message box will pop up.
2. Select the backup directory (the default directory is the folder of “QC Data” under the
8-117
Using the QC Programs
3. Enter the name for the backup data (the default name is [X_QC_date saved_time
saved]).
5. When the backup is finished, a message box will pop up, and then click “Ok” to exit.
The backup data can not be modified. You can review the data in “History”
screen.
You can click the “Cancel” button to cancel backup when it is in process.
Data Export
If you wish to export the information and the result of the current QC file, do as follows:
1. Click the “Data Export” button, and then a message box will pop up.
8-118
Using the QC Programs
2. Select the export directory (the default directory is the folder of “QC Data” under the
installation location of the terminal software).
3. Enter the name for the export data (the default name is [X_QC_date saved_time
saved]).
6. When the export is finished, a message box will pop up, and then click “Ok” to exit.
You can modify the exported data but can not review the data in the
“History” screen.
You can click the “Cancel” button to cancel export when it is in process.
8-119
Using the QC Programs
History
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the history function, you should
log on the system as administrator, and then right click the shortcut icon of
the terminal software and run it as administrator.
1. Click the “History” button, and then a message box will pop up.
8-120
Using the QC Programs
4. The display format and the function button are the same as the QC Table review screen.
Browse, print, transmit and export the data as instructed in the section of QC Table
Review.
8-121
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
Before analyzing the new lot of controls, you should set a QC file for each lot of controls and
you can edit the QC settings in the QC file by the following way:
Manual entry
Manual Entry
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean R” on the pop-up
menu.
8-122
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-123
Using the QC Programs
You can select the file No. within the range [1, 60].
3. You can enter the lot No. of the controls by one of the following ways:
Manual entry
The lot No. can not be empty and up to 16 digits can be entered. You can
enter characters, numbers, letters and special characters, but no Chinese
characters allowed.
Different QC files can not have the same lot No. and QC mode.
8-124
Using the QC Programs
You must enter the expiration date, and the entry range is [current system
date, 2099-12-31].
Different QC files can not have the same lot No. and QC mode.
7. Click the “Save” button to save all the settings of the QC.
Print
Click the “Print” button to print the setting information of the current QC file.
8-125
Using the QC Programs
CT-WB
CT-PD
AL-WB
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-126
Using the QC Programs
If the sample mode is switched from the “Predilute” to “Whole Blood”, the
analyzer will perform the switching sequence automatically and a progress
bar will be displayed on the screen.
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC””X mean R” on the pop-up
menu.
8-127
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
Be sure to use the adapter of Ф13x75 (mm) model when running the QC in
CT-WB mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
8-128
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “CT-WB” and the analysis status icon and analyzer indicator is
green.
2) Shake the prepared control as shown below to well mix it.
8-129
Using the QC Programs
9) Mix the control well again, and then to run the controls for the second time according to the
prompt.
8-130
Using the QC Programs
You can click the “Cancel” button in the message box to cancel the second
run and the results obtained in the first run will not be saved as well.
10) When the running is finished, the compartment door will open automatically and you can
remove the control.
7. When finish running, the QC results (values of the two QC runs, X mean and range R)
will be displayed in the current screen and be saved in the QC file automatically.
When the QC result of the second QC run is obtained, the screen will refresh
the displayed histograms and scattergrams according to the second QC
run.
Up to 300 QC results (X mean and range R) can be saved for each QC file.
8-131
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
Be sure to place the collection tubes in the right adapter before running,
otherwise, the collection tubes may be broken and cause biohazard.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-132
Using the QC Programs
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC”” X mean R” on the
pop-up menu.
8-133
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
Be sure to use the adapter of Ф11x40 (mm) model when running the QC in
CT-PD mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
8-134
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “CT-PD” and the analysis status icon and analyzer indicator is
green.
2) Click the shortcut button “Diluent”, the compartment door will open automatically. A
message box shown below will pop up.
8-135
Using the QC Programs
3) Present a clean centrifugal tube into the sample compartment with the cap opened. Press
the [RUN] key to dispense 180μL of diluent into the tube. During dispensing the diluent, a
progress bar will display.
If the current mode is autoloader, the analyzer will switch to the close-tube
mode automatically and then dispense the diluent. Then, the mode will be
switched back to autoloader when exiting.
4) When the dispensing is finished, the compartment door will open automatically and you
can remove the centrifugal tube. Then, the following message box will display.
8-136
Using the QC Programs
10) Present the centrifugal tube into the tube holder with the cap opened and then close the
compartment door.
11) Press the [Run] key or click the “Run” button on the message box to start running.
12) When the running is finished, the compartment door will open automatically and you can
remove the centrifugal tube.
13) After the analyzing is finished, the QC result of the first analysis will be displayed on the
screen.
14) Add 20μL of control to the other centrifugal tube with diluent, close the tube cap and
shake the tube to mix the sample, and then present the centrifugal tube into the tube holder
with the cap opened to start the second running.
You can click the “Cancel” button in the message box to cancel the second
run and the results obtained in the first run will not be saved as well.
15) When the running is finished, the compartment door will open automatically and you can
remove the centrifugal tube.
7. When finish running, the QC results (values of the two QC runs, X mean and range R)
will be displayed in the current screen and be saved in the QC file automatically.
8-137
Using the QC Programs
You can also dispense 180μL of diluent by pipette into the tube.
After mixing the control with the diluent, be sure to wait 3 minutes before
running.
Be sure to run the prediluted samples within 30 minutes after the mixing.
Be sure to mix any sample that has been prepared for a while before running
it.
When the QC result of the second QC run is obtained, the screen will refresh
the displayed histograms and scattergrams according to the second QC
run.
Up to 300 QC results (X mean and range R) can be saved for each QC file.
8-138
Using the QC Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The sample may spill from the unclosed collection tubes and cause
biohazard. Exercise caution to the unclosed collection tubes.
Be sure to avoid reversing the collection tube when loading, otherwise, the
collection tube may be broken and cause biohazard.
The pusher will push the rack inside the autoloader. Be sure your hand is
away from the rack before starting the autoloader.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
8-139
Using the QC Programs
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC”” X mean R” on the
pop-up menu.
8-140
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-141
Using the QC Programs
2. Select the QC file No. to be run; the screen displays the corresponding file information.
3. Be sure that the level of the control to be run is the same with the current QC file.
1) Make sure the QC mode is “AL-WB” and the analysis status icon and analyzer indicator is
green.
2) Click the “Start” button.
8-142
Using the QC Programs
4) Place racks loading controls on the level of the right tray of the autoloader, with the back of
“MINDRAY” mark on the carrier facing the analyzer.
5) Press the [Run] key or click the “Run” button on the message box to start running.
6) When the racks are pushed out, you can remove the racks from the left of the autoloader.
7) After the analyzing is finished, the QC result of the first analysis will be displayed on the
screen.
8) Present the prepared control into the rack according to the prompt, and then place racks
loading controls on the level of the right tray of the autoloader, with the back of “MINDRAY”
mark on the carrier facing the analyzer, and then start the second running.
You can click the “Cancel” button in the message box to cancel the second
run and the results obtained in the first run will not be saved as well.
9) When the racks are pushed out, you can remove the racks from the left of the autoloader.
7. When finish running, the QC results (values of the two QC runs, X mean and range R)
will be displayed in the current screen and be saved in the QC file automatically.
When the QC result of the second QC run is obtained, the screen will refresh
the displayed histograms and scattergrams according to the second QC
run.
Up to 300 QC results (X mean and range R) can be saved for each QC file.
8-143
Using the QC Programs
You can click the button or button to switch to the pervious or the next QC result.
You can click the button or button to switch to the earliest or the latest QC result
saved in the QC file.
The running result of the expired control will begin with an “O” mark.
Print
Click the “Print” button to print the results of the current QC Run screen.
Graph
Table
Graph Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
8-144
Using the QC Programs
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC”” X mean R” on the
pop-up menu.
1. Select the QC file No. you want to review, and then the screen will display the
corresponding information and the graph.
8-145
Using the QC Programs
2. You can drag the scroll bar on the right of the graph to browse the desired graph of the
parameter. You can drag the scroll bar down to the graph horizontally to browse all the
QC results.
1- The Mean, SD and CV% of all the QC results of each parameter in the current graph.
2- The saving date and time of the QC points located on the green line.
3- The operator who run the QC analysis and obtained the QC points located on the green
line
4- The QC results of the parameters that correspond to the QC points located on the green
line.
5- The QC points in each graph are displayed from left to right according to the sequence
from the earliest to the latest. The QC points are connected by a line to illustrate the
distribution trend.
6- The QC point corresponds to each QC result. Only the selected QC point displays its value
under the parameter. The black QC point indicates the value is within the limit; the red QC
point indicates the value is out of the limit.
7- When you clicking a QC point in the graph, the QC points of other parameters that saved
together with this one will be marked by a green line.
8-146
Using the QC Programs
8- The relative position of the QC point located on the green line and the total QC points
saved currently.
The graphs for different parameters will be provided only after 10 batches of
QC results (20 times of QC runs) are obtained.
The values of the QC point are the X mean and range R of each batch of QC
results.
The outliers are excluded from the calculation of Mean, SD and CV%.
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to move the green line to the pervious or the
next QC point; you can click the button or button to move the green line to the first
or the last QC point in the graph. When the location of the green line is selected, you can
check the QC results of the QC points located on the green line under each parameter.
New Vial
If the reviewed QC results are obtained by analyzing a new vial of control of the same batch,
you should mark the QC points of the new vial to distinguish the QC results from the old.
1. Move the green line to the first QC point of the new vial.
2. Click the “New Vial” button, and then a blue line appears at the QC point of the new vial.
3. After another new vial of control (of the same batch) is run and its QC results are saved,
you can continue to mark the current QC points of the new vial according to step 1 and 2.
4. If the current QC point is marked with the blue line, the “New Vial” button turns into
“Cancel”, you can click the button to remove the blue line, then the “Cancel” button will
turn back to “New Vial”.
Data Compare
If you wish to compare the graphs of the certain parameter obtained by running controls of
different lot No., do as follows:
1. Click the “Data Compare” button to start selecting the desired graph.
8-147
Using the QC Programs
2. Select the desired QC file No. into the “File No.” box (3 files can be selected at most).
Then, the graph of the selected QC file will be displayed below together with its lot No.,
QC mode and level.
Browse the graph here the same way as instructed in the “Graph” screen.
Be aware that, for controls of different level, their graphs will be
distinguished by the color of orange, black and blue.
Display Order
1. Click the “Display Order” button to check the current display order of the graphs.
8-148
Using the QC Programs
3. You can click the button or button to move parameter upward or downward;
you can click the button or button to move the parameter to the first or the last
position.
4. Click the “Ok” button to refresh the display order of the graphs.
1. After moving the green line to the desired QC point, you can click the “Outliers” button to
display the QC results, total mean and average range of all the parameters located on
the green line (the QC results exceed the limit will be displayed in red) and enter the
reasons for the outliers.
8-149
Using the QC Programs
2. You can select the reason form the given ones or enter the reasons (up to 200
characters) into the edit box manually after selecting “Others”.
3. Click the “Ok” button to save the reasons for the outliers and exit.
If you enter the reason for the group of QC points whose results are actually
within the limits, then their corresponding QC data both in the QC Graph and
QC Table will be displayed in the color of red. And the data will return in the
color of black if you cancel the reason and then save the changes.
Delete
1. If you wish to delete a single QC result, move the green line to the desired QC result; if
you wish to delete all the data, perform step 2 directly.
8-150
Using the QC Programs
4. Click the “Ok” button and then confirm to delete the selected data.
Click the “Print” button to print all the file information and graphs of the parameters of the
current QC file.
The green line and the corresponding values of the QC points will not be
printed.
8-151
Using the QC Programs
Table Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
You can enter the Table screen in one of the following ways:
Click the “Menu” button on the screen, and then select “QC”” X mean R” on the
pop-up menu.
8-152
Using the QC Programs
8-153
Using the QC Programs
2. Select the QC file No. you want to review, and then the screen will display the
corresponding information and the table.
3. You can drag the scroll bar on the right of the table vertically to browse the desired table
of the parameter. You can drag the scroll bar down to the table horizontally to browse all
the QC results.
1- The No. of the QC result saved in the QC file (arranged from left to right in the order that
from the earliest to the latest)
2- QC Result
3- QC parameters (displayed in the same order as the Graph screen)
4- QC flag: If 10 batches of QC results (20 times of QC runs) are obtained, the flag “H” or “L”
will be used to prompt the result (X mean) that out of the limits
5- The relative position of the highlighted QC point and the total QC points saved currently.
8-154
Using the QC Programs
The total mean, average range and flag prompt for the parameters will be
provided only after 10 batches of QC results (20 times of QC runs) are
obtained.
The values of the QC result are the X mean and the range R of each batch of
QC results.
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to highlight the previous or the next QC result;
you can click the button or button to highlight the first or the last QC result in the
table.
Delete
1. If you wish to delete a single QC result, click the column contains the desired QC result;
if you wish to delete all the data, perform step 2 directly.
4. Click the “Ok” button and then confirm to delete the selected data.
8-155
Using the QC Programs
1. Click the “Print” button, and then you can select “All Data” or “Specified Data” to be
printed.
2. Click the “All Data” button and then click the “Ok” button to print all the file information
and tables of the parameters of the current QC file; after clicking the “Specified Data”
button, you can select the starting and finishing date (the saved date of the QC results)
for printing, then click the “Ok” button to print the specified data.
Communication
If you wish to transmit the QC data to the external data management software or LIS/HIS, do
as follows:
1. Click the “Communication” button, and then you can select “All Data” or
8-156
Using the QC Programs
2. Click the “All Data” button and then click the “Start” button to transmit the
information of the current QC file and QC data. After clicking the “Specified Data”
button, you can specify a date range (the date when the QC result was saved), then
click the “Ok” button to transmit the specified data. The “LIS/HIS connected” icon on
3. While transmitting, the “Start” button in the pop-up message box will be replaced by
Data Backup
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the data backup function, you
should log on the system as administrator, and then right click the shortcut
icon of the terminal software and run it as administrator.
If you wish to backup the information and the result of the current QC file, do as follows:
1. Click the “Data Backup” button, a message box will pop up.
8-157
Using the QC Programs
2. Select the backup directory (the default directory is the folder of “QC Data” under the
installation location of the terminal software).
3. Enter the name for the backup data (the default name is [X-R_QC_date saved_time
saved]).
5. When the backup is finished, a message box will pop up, and then click “Ok” to exit.
8-158
Using the QC Programs
The backup data can not be modified. You can review the data in “History”
screen.
You can click the “Cancel” button to cancel backup when it is in process.
Data Export
If you wish to export the information and the result of the current QC file, do as follows:
1. Click the “Data Export” button, and then a message box will pop up.
2. Select the export directory (the default directory is the folder of “QC Data” under the
installation location of the terminal software).
3. Enter the name for the export data (the default name is [X-R_QC_date saved_time
saved]).
8-159
Using the QC Programs
6. When the export is finished, a message box will pop up, and then click “Ok” to exit.
You can modify the exported data but can not review the data in the
“History” screen.
You can click the “Cancel” button to cancel export when it is in process.
History
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the history function, you should
log on the system as administrator, and then right click the shortcut icon of
the terminal software and run it as administrator.
1. Click the “History” button, and then a message box will pop up.
8-160
Using the QC Programs
8-161
Using the QC Programs
4. The display format and the function button are the same as the QC Table review screen.
Browse, print, transmit and export the data as instructed in the section of QC Table
Review.
8-162
Using the QC Programs
It is recommended the X-B analysis be activated when the sample volume of your laboratory
is greater then 100 samples per day. The analyzer can save maximum 500 X-B QC results.
When the saved QC results have reached the maximum number, the newest result will
overwrite the oldest.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
At the X-B QC setting screen, you can edit the QC information and configure the sample
validity setup.
Manual entry
Manual Entry
You can enter the graph screen in one of the following ways:
8-163
Using the QC Programs
Click the “Menu” button on the screen; then select “QC””X-B” on the pop-up menu.
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-164
Using the QC Programs
2. In the “Samples/Batch” edit box, you can enter the amount of samples [within the range
20(recommended) to 200] to be included in calculating for an X-B QC point.
3. Click the “Open” button of “X-B” to open the X-B QC, and from the time on, all the
samples results will be included to calculate the X-B.
All the targets and limits for the QC parameters shall be entered without
empty.
When first use, the default setting will provide the Initial values for the
targets and limits of the three QC parameters.
If the QC data have existed in the QC file, you are not allowed to edit the
target and limits.
5. Click the “Save” button to save all the settings of the QC.
8-165
Using the QC Programs
If there are the saved preset values (Target and Limits) for the X-B QC, you
can read-in the preset values into the X-B QC file. For details of calculating
and saving the preset values, see Section 8.5.3 Reviewing QC results.
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen; then select “QC””X-B” on the pop-up menu.
8-166
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-167
Using the QC Programs
2. In the “Samples/Batch” edit box, you can enter the amount of samples [within the range
20(recommended) to 200] to be included in calculating for an X-B QC point.
3. Click the “Open” button of “X-B” to open the X-B QC, and from the time on, all the
samples results will be included to calculate the X-B.
4. Click the “Have Preset Values” button to read-in the saved preset target and limits into
the X-B QC file.
All the targets and limits for the QC parameters shall be entered without
empty.
If some QC parameters have no preset values, you should enter the target
and limits for them manually.
If the QC data have existed in the QC file, you are not allowed to have the
preset values.
5. Click the “Save” button to save all the settings of the QC.
8-168
Using the QC Programs
Setting Limits
Do as followings to adjust the display format of the limits and the calculation method of the
preset limits.
1. Click the “Set Limits” button, and then the following message box will pop up.
2. Click “By SD” to display the limits in the form of absolute value; click “By CV” to display
the limits in the form of percentage.
3. If “By SD” is selected, click the “2SD” or “3SD” to select either double or triple standard
deviation to be the limits; if “By CV” is selected, click the “2CV” or “3CV” to select either
double or triple coefficient of variation to be the limits.
4. Click the “Ok” button to save all the settings for the limits.
Restoring Defaults
When editing the QC settings, if you wish to restore the target and limits to the defaults, you
can click the “Restore Default” button to read-in the defaults to the X-B QC file.
MCV: 89.5fL
MCH: 30.5pg
MCHC: 340g/L
MCV: 2.7 fL
MCH: 0.9 pg
MCHC: 10 g/L
8-169
Using the QC Programs
If the QC data have existed in the QC file, you are not allowed to restore
defaults.
2. Background results;
5. Calibration data;
6. Results generated while there are errors which could affect the accuracy of the results
(insufficient aspiration volume or clogging for example).
"Sample Validity Setup" is to set up the ranges of valid RBC, MCV, MCH and MCHC results.
Only when the results of all these four parameters are within the specified ranges, the sample
results can be used for X-B QC calculation. Do as follows to set the sample validity:
Click the “Menu” button on the screen; then select “QC””X-B” on the pop-up menu.
8-170
Using the QC Programs
For details to edit the name of the login user, see chapter 5 Customizing the
Analyzer Software.
8-171
Using the QC Programs
3. Set the upper and lower limits of the four parameters in the "Sample Validity Setup"
area.
5. If any value you entered is out of range or any upper limit entered is less than the
corresponding lower limit, the following message box will pop up when you click the
"Save" button.
6. Click "Ok" to go to the QC setting screen and modify the invalid values.
8-172
Using the QC Programs
1.0×1012/L≤RBC≤8.0×1012/L
50fL≤MCV≤150fL
20pg≤MCH≤40pg
240g/L≤MCHC≤440g/L
The validity entry range for RBC is its linearity range, and the validity entry
ranges for other three parameters are those of their display ranges.
All the entries should be numbers with only one decimal point, and the
entries should be restricted to the length of the edit boxes.
Once the validity range is changed, the previous results will not be used in
the QC calculation as valid results, for example, if 20 valid samples are
needed for the X-B QC calculation, when you change the validity range after
10 groups of valid sample results have been acquired, these 10 groups of
results will be discarded, and only valid sample results generated
afterwards will be used in the QC calculation.
Print
Click the “Print” button to print the setting information of the current QC file.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
After editing the X-B settings, the system will start the X-B run automatically.
After every 20~200 results (determined by the setting) are obtained, the system will perform
the X-B calculation once automatically. You can review the result in X-B graph or X-B table.
8-173
Using the QC Programs
Graph
Table
Graph Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
You can enter the graph screen in one of the following ways:
Click the “Menu” button on the screen; then select “QC””X-B” on the pop-up menu.
8-174
Using the QC Programs
You can drag the scroll bar down to the graph horizontally to browse all the QC results.
8-175
Using the QC Programs
1- The Mean, SD and CV% of all the QC results of each parameter in the current graph.
2- The saving date and time of the QC points located on the green line
3- The QC points in each graph are displayed from left to right according to the sequence
from the earliest to the latest. The QC points are connected by a line to illustrate the
distribution trend.
4- The QC results of the parameters that correspond to the QC points located on the green
line.
5- When you clicking a QC point in the graph, the QC points of other parameters that saved
together with this one will be marked by a green line.
6- The QC point corresponds to each QC result. Only the selected QC point displays its value
under the parameter. The black QC point indicates the value is within the limit; the red QC
point indicates the value is out of the limit.
7- The relative position of the QC point located on the green line and the total QC points
saved currently.
The value of the outlier is the X-B result of each batch of samples.
The outliers are excluded from the calculation of Mean, SD and CV%.
8-176
Using the QC Programs
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to move the green line to the previous or the
next QC point; you can click the button or button to move the green line to the first
or the last QC point in the graph. When the location of the green line is selected, you can
check the QC results of the QC points located on the green line under each parameter.
If there are 3 or more than 3 QC results within the limits obtained for the parameters, you can
take the following steps to calculate and save the preset value for each parameter:
1. Click the “Calculate Preset Values” button, and then the screen displays two lines for
you to select the range for calculating the preset values.
2. Click and drag the two lines respectively to locate them at the beginning and the ending
of the range for calculating the preset values.
3. The Mean, SD and CV% (on the right of the graph) will change into the new results
which obtained by calculating the selected range.
4. If you wish to save the new results, you can click the “Save Preset Values” button to
save the current Mean, SD and CV% as the preset values for the corresponding level
(high/normal/low). Then, the two selecting lines disappear and the Mean, SD and CV%
return to the calculated results of all QC results.
The calculation and display of the preset values are only available to the
parameter (within the calculation range) which has 3 or more than 3 results
within the limit. Otherwise, the display of the preset values will be empty.
Delete
1. If you wish to delete a single QC result, move the green line to the desired QC result; if
you wish to delete all the data, perform step 2 directly.
8-177
Using the QC Programs
4. Click the “Ok” button and then confirm to delete the selected data.
Click the “Print” button to print all the file information and graphs of the parameters of the
current QC file.
The green line and the corresponding values of the QC points will not be
printed.
8-178
Using the QC Programs
Table Review
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
You can enter the Table screen in one of the following ways:
Click the “Menu” button on the screen; then select “QC””X-B” on the pop-up menu.
8-179
Using the QC Programs
8-180
Using the QC Programs
2. You can drag the scroll bar down to the table horizontally to browse all the QC results.
1- QC Result
2- The No. of the QC result saved in the QC file (arranged from left to right in the order that
from the earliest to the latest)
3- QC parameters (displayed in the same order as the Graph screen)
4- QC flag: The flag “H” or “L” will be used to prompt the result that out of the limits
5- The relative position of the highlighted QC point and the total QC points saved currently.
The value of the QC result is the X-B result of each batch of samples.
You can click the arrow button in the bottom of the screen to browse the QC result saved in
the current QC file.
You can click the button or button to highlight the previous or the next QC result;
8-181
Using the QC Programs
you can click the button or button to highlight the first or the last QC result in the
table.
Delete
1. If you wish to delete a single QC result, click the column contains the desired QC result;
if you wish to delete all the data, perform step 2 directly.
4. Click the “Ok” button and then confirm to delete the selected data.
1. Click the “Print” button, and then you can select “All Data” or “Specified Data” to be
printed.
8-182
Using the QC Programs
2. Click the “All Data” button and then click the “Ok” button to print all the file information
and tables of the parameters of the current QC file; after clicking the “Specified Data”
button, you can select the starting and finishing date (the saved date of the QC results)
for printing, then click the “Ok” button to print the specified data.
Communication
If you wish to transmit the QC data to the external data management software or LIS/HIS, do
as follows:
1. Click the “Communication” button, and then you can select “All Data” or
2. Click the “All Data” button and then click the “Start” button to transmit the
information of the current QC file and QC data. After clicking the “Specified Data”
button, you can specify a date range (the date when the QC result was saved), then
click the “Ok” button to transmit the specified data. The “LIS/HIS connected” icon on
8-183
Using the QC Programs
3. While transmitting, the “Start” button in the pop-up message box will be replaced by
Data Backup
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the data backup function, you
should log on the system as administrator, and then right click the shortcut
icon of the terminal software and run it as administrator.
If you wish to backup the information and the result of the current QC file, do as follows:
1. Click the “Data Backup” button, a message box will pop up.
8-184
Using the QC Programs
2. Select the backup directory (the default directory is the folder of “QC Data” under the
installation location of the terminal software).
3. Enter the name for the backup data (the default name is [X-B_QC_date saved_time
saved]).
5. When the backup is finished, a message box will pop up, and then click “Ok” to exit.
The backup data can not be modified. You can review the data in “History”
screen.
You can click the “Cancel” button to cancel backup when it is in process.
Data Export
If you wish to export the information and the result of the current QC file, do as follows:
1. Click the “Data Export” button, and then a message box will pop up.
8-185
Using the QC Programs
2. Select the export directory (the default directory is the folder of “QC Data” under the
installation location of the terminal software).
3. Enter the name for the export data (the default name is [X-B_QC_date saved_time
saved]).
4. Select the format for the export file. (Default format: “. CSV”)
6. When the export is finished, a message box will pop up, and then click “Ok” to exit.
You can modify the exported data but can not review the data in the
“History” screen.
You can click the “Cancel” button to cancel export when it is in process.
8-186
Using the QC Programs
History
For operating system of Windows Vista, the right of copying database file is
restricted by users' access level. To enable the history function, you should
log on the system as administrator, and then right click the shortcut icon of
the terminal software and run it as administrator.
1. Click the “History” button, and then a message box will pop up.
8-187
Using the QC Programs
4. The display format and the function button are the same as the QC Table review screen.
Browse, print, transmit and export the data as instructed in the section of QC Table”
Review.
8-188
9 Using the Calibration Programs
9.1 Introduction
Calibration is a procedure to standardize the analyzer by determining its deviation, if any,
from calibration references and to apply any necessary correction factors.
There are three calibration programs available on this analyzer: manual calibration, auto
calibration using calibrators and auto calibration using fresh blood samples.
All the parameters or part of the parameters of WBC, RBC, HGB, MCV and PLT can be
calibrated by the calibration procedure.
You should only use the Mindray-specified calibrations and reagents. Store
and use the calibrations and reagents as instructed by instructions for use
of the calibrations and reagents.
9-1
Using the Calibration Programs
it is the first time this analyzer has been used (usually done by a Mindray-authorized
representative when installing the analyzer).
9-2
Using the Calibration Programs
1. Check and make sure enough reagents have been prepared for the calibration. You
need to start over the calibration if the reagents run out during the process.
2. Do the background check. If the analyzer alarms for abnormal background results, see
Chapter 12 Troubleshooting Your Analyzer for solutions.
Run a vial of normal control in the CT-WB-CBC+DIFF mode for 11 consecutive times. Enter
the “TABLE” screen to check the reproducibility of the second to eleventh runs and make sure
they meet the following requirements.
Whole Blood Predilute
Parameter Condition Reproducibility Reproducibility
(CV%) (CV%)
9
WBC (4.0-15.0)×10 /L ≤ 2.0% ≤ 4.0%
12
RBC (3.50-6.00)×10 /L ≤ 1.5% ≤3.0%
HGB (110-180)g/L ≤ 1.5% ≤3.0%
MCV (70-120)fL ≤ 1.0% ≤2.0%
9
PLT (150-500)×10 /L ≤ 4.0% ≤8.0%
3. Run a vial of high control three consecutive times and then immediately run the diluent
three consecutive times. Calculate the carryover per the following equation.
4. It is recommended that you create a log table for your analyzer. This log table should
contain all necessary information that is pertinent to your analyzer. Suggested items that
you may want to include in the log table are: calibration date, supplier of calibrator, lot
9-3
Using the Calibration Programs
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
Keep you clothed, hairs and hands away from the moving parts to avoid
injury.
You should only use the Mindray-specified controls and reagents. Store and
use the controls and reagents as instructed by instructions for use of the
controls and reagents.
9-4
Using the Calibration Programs
1. Click “Menu”, select “Calibration” to enter the ”Calibration Factors” screen. The
calibration factors of whole blood mode and predilute mode are displayed at the
“Calibration Factors” screen.
The login users of common-level can not perform the calibration procedures
but only browse the calibration factors at the current screen. To perform the
calibration, please logout and then login as users of administrator-level.
2. Enter the “Calibration Factors” screen to check the calibration factors and calculate the
new factors per the following equation.
9-5
Using the Calibration Programs
Run the calibrator in the whole blood mode for 11 consecutive times and take the WBC
results of the 2nd to 11th runs (n=10) to calculate: 8.1, 8.0, 8.1, 8.1, 8.3, 8.3, 8.2, 8.0, 8.1, 8.3.
The obtained CV is 1.5% and Mean is 8.16, which meet the requirements.
The new calibration factor is obtained:
98.90% 8.4
New calibration factor= =101.81%
8.16
The calculated calibration factors shall be between 75% - 125%.In case of an invalid
calibration factor, try to find out the reason (e.g. calibration material not thoroughly mixed
misoperation, etc.).Then recalibrate the analyzer and recalculate the calibration factors.
3. Enter the new calibration factors into the factor cell of the parameter that requires
calibration.
4. After the entry, click the “Save” button at the bottom of the screen. If the new calibration
factors are valid and different from the originals, a message box shown below will pop
up.
Click “Yes” to save the news calibration factors and the calibration date of the corresponding
parameter changes to the current system date. Then, close the message box and return to
the “Calibration Factors” screen without any cell being highlighted.
If the new calibration factors are invalid, the message box will pop up.
Click “Ok” to close the message box and the cell of the first invalid calibration factors is
9-6
Using the Calibration Programs
5. After the calibration factors are modified, a prompt will show if you switch to another
screen without clicking the “Save” button.
If the entered calibration factors are valid, the message box will pop up when you exiting
the screen.
Click “Yes” to save the news calibration factors and the calibration date of the corresponding
parameter changes to the current system date and be recorded in the history; then, close the
message box and switch to another screen.
If the entered calibration factors are invalid, the message box will pop up when you
switching to another the screen.
Click “Yes” to close the message box and switch to another screen without saving; keep the
original calibration factors and date.
Other Operations
Restore
Click the “Restore” button to restore the calibration factors to the values displayed when you
entering the “Calibration Factors” screen.
If the calibration factors have not been changed, click the ”Print” button to print the current
9-7
Using the Calibration Programs
calibration factors.
If the changed calibration factors are invalid, then a message box will pop up when you
clicking the “Print” button.
Click “Ok”, then the cell of the first invalid calibration factors will be highlighted and the data in
the cell will not be cleared.
If the changed calibration factors are valid but have not been saved, then a message box will
pop up when you clicking the “Print” button.
Click “Yes” to close the message box and save the new calibration factors and date, and then
print the new calibration factors; click “No” to close the message box without saving the
calibration factors and date, and then print the saved calibration factors before editing.
1. Click the “Menu” button, and then select “Calibration” to enter the “Calibration
Factors” screen.
2. At the “Calibration Factors” screen, click the “Calibrator” tab to enter the “Calibrator”
screen.
9-8
Using the Calibration Programs
Only in the Whole Blood Mode can the calibration using calibrators be ran.
3. Enter the lot No. of the calibrator into the “Lot No.” box.
4. Enter the expiration date. The default “Exp. Date” is the current system date. You can
click the “Exp. Date” box, and then edit the date.
5. Select the parameter to be calibrated from the check box on the first line of the list.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
9-9
Using the Calibration Programs
See the instructions for use of the calibrators for the lot No., expiration date
and the target.
The expiration date can not be earlier than the current system date.
The entered expiration date should be either the expiration date printed on
the labeling or the open-container expiration date, whichever is earlier. The
open-container expiration date is calculated as follows: the date that
container is opened + the open-container stability days.
8. Click the “Start” button, and then a message box will pop up.
Press the [RUN] key or click the “Run” button to start the calibration and the message box will
close automatically, and then a progress bar will pop up.
Once you click the “Start” button and press the [RUN] key to start the first
run, the “Start” button will be displayed in gray. Then, you can directly press
the [RUN] key to continue the calibration.
9-10
Using the Calibration Programs
9. After every calibration run, the progress bar will close automatically and the analyzer will
have different responses according to different analysis results.
When the current running is done, if there is a parameter whose calibration data is out of
its linear range but still within the display range, then the calibration data will be
displayed in the list and a message box will also pop up.
Click “Ok” to close the message box and delete the data from the table without saving.
When the running is done, if there is a parameter whose calibration data is out of the
display rage, then the non-numeric parameter values “***” will be displayed in the list and
a message box will pop up.
Click “Ok” to close the message box and delete the data from the table without saving.
The valid results within the linear range will be displayed directly.
10. If the calibration factors have not been calculated but you switch to another screen, then
a message box will pop up.
9-11
Using the Calibration Programs
Click “Yes” to switch to another screen while aborting the calibration data and closing the
message box. The original calibration factors remain.
11. When the amount of the valid calibration reaches N (N ≥6), the analyzer will
automatically calculate the mean, CV% and new calibration factors with all the selected
data ( the first data is excluded).
You can also select the desired data (5 at least) to calculate the calibration factors. Every time
when you select or de-select a data by clicking the check box, the calibration factors will be
refreshed immediately.
The out-of-range CV% does not influence the display of the calibration
factors.
When the amount of the valid calibration data in the list reaches 11, a
message box of “Calibrator calibration done!” will pop up. Then, if you
press the [RUN] key again, the analyzer will beep and does not response.
12. There may be two cases when you switching to another screen:
If the calculated calibration factor of any parameter is out of the range [75%-125%] or the
CV% of any parameter exceeds the reproducibility standard, then the calculated
calibration factors of all parameters will not be saved and a message box will also pop
up.
9-12
Using the Calibration Programs
Click “Yes” to close the message box and switch to another screen without changing the
original calibration factors and the calibration date.
If the calculated calibration factors of all parameter are within the range [75%-125%] and
the CV% of all parameter are also within the reproducibility standard, then a message
box will pop up.
Click “Yes” to save the new calibration factors while closing the message box and switching
to another screen.
Other Operations
Print
If the calibration factors are invalid, then a message box will pop up when you clicking the
“Print” button.
Click “Ok”, then the cell of the first invalid calibration factors will be highlighted and the data in
the cell will not be cleared.
If the calibration factors are valid, then a message box will pop up when you clicking the
“Print” button.
9-13
Using the Calibration Programs
Click “Yes” to close the message box and save the calibration results and the calibration date,
and then print the contents of the current calibration screen; click “No” to close the message
box without saving.
2. At the “Calibration Factors” screen, click the “Fresh Blood” tab to enter the “Fresh
Blood” screen.
9-14
Using the Calibration Programs
Analyzer.
4. Run each of the prepared samples on the reference instrument (or by the reference
method) three times at least. Average the results for your reference values
5. Click the radio button “CT-WB” or “CT-PD” on the screen to select the desired calibration
mode.
If you run the sample in the “Predilute” mode, and then the blood mode is
switched from the “Predilute” to “Whole Blood”, the analyzer will switch
modes automatically and a progress bar will be displayed on the screen.
Be sure to use the adapter of Ф11x40 (mm) model when calibrating in CT-PD
mode. For details to install the adapter, see Chapter 6 Operating Your
Analyzer.
6. Select the sample ID of the current sample from the “Current sample ID” pull-down list.
7. Select the parameter to be calibrated from the check box on the first line of the list.
9. Prepare the whole blood or predilute fresh blood sample ready for calibration.
10. Click the “Start” button, a message box will pop up.
Press the [Run] key or click the “Run” button, and then the message box will close
automatically and a process bar will pop up.
9-15
Using the Calibration Programs
After you click the “Start” button and press the [RUN] key to start the first
run, the “Start” button will display in gray. Then, you can press the [RUN]
key to continue the calibration.
11. After the analysis, the progress bar will close and the compartment door will open
automatically. The analyzer will have different responses to different analysis results.
If the results are out of the linear range but still within the display range, the message
box will pop up at the same time the results are displayed in the table.
Click “Ok” to close the message box and delete the results from the table without saving.
If the results are out of the display rage, the non-numeric parameter values “***” are
obtained and the message box will pop up.
9-16
Using the Calibration Programs
Click “Ok” to close the message box and delete the results from the table without saving.
The valid results within the linear range will be displayed directly.
12. When the amount of the valid calibration reaches N (N ≥6), the analyzer will calculate the
Mean, CV% and Calibration Factors of the data selected with “√” automatically ( the first
data is excluded.)
You can select several data to calculate the calibration factors, but after 5 groups of the data
are selected at least can you get the calibration factors. Every time when you select or cancel
a data by clicking the check box, the calibration factors will refresh and display immediately.
The exceeded CV% does not influence the display of the calibration factors.
When the amount of calibration data in the table reaches 11, a prompt of
“Fresh blood calibration done!” will pop up; if you press the [RUN] key
again, the analyzer will beep and does not response.
13. Select other calibration samples from the “Current sample ID” pull-down list, run the
samples as instructed in steps 8 to 12 to obtain the calibration factors of each sample.
9-17
Using the Calibration Programs
14. There may be several cases when switching to another blood sample:
If the calibration factors of the blood sample are invalid or the CV% of any parameter
exceeds the reproducibility standard, a message box will pop up when you switching to
another blood sample.
Click “Yes” to clear the entered target of the current sample, all the calibration data obtained
and each calculated value including calibration factors, then close the message box and
switch to another blood sample.
If the calibration factors have not been calculated, the message box will pop up.
Click “Yes” to empty the entered target of the current sample and all the calibration data
obtained, then close the message box and switch to another blood sample.
If the calibration factors of the sample are valid and the CV% of all the parameters do not
exceed the reproducibility standard, you can switch to another blood sample directly.
15. After calibration factors of at least 3 fresh blood samples are obtained, click the
“Calculate” button to enter the screen of calibration calculation.
9-18
Using the Calibration Programs
Select or cancel the calibration factors of a blood sample to calculate for the Mean calibration
factors by click the relevant check box.
When the selected calibration factors reaches 3 or more than 3, the CV% will be calculated
over again according to the selected calibration factors.
The exceeded CV% does not influence the display of the calibration factors.
When the selected calibration factors reaches 3 or more than 3, the mean calibration factors
will be calculated over again according to the selected calibration factors. If the deviation of
the calibration factor involved in the mean calibration factor calculation from the previous
calibration factor equals to or is greater than 5%, this calibration factor will be considered as
invalid, and the following message box will pop up when you try to exit the fresh blood
calibration screen.
Click "Yes" to close the message box, clear current calibration data, and switch to the
corresponding screen.
Click "No" to go back to the current screen. The invalid calibration factor(s) will be marked
with "?" and highlighted in red.
9-19
Using the Calibration Programs
When the calculated mean calibration factor is invalid, you can perform
manual calibration at the calibration factor screen.
16. If the mean calibration factors have not been calculated, when you switch to the fresh
blood screen or switch to another calibration mode, a message box will pop up.
Click “Yes” to abort the calibration data and close the message box, switching to the
corresponding screen or other calibration mode. The original calibration factors and date
remain the same.
17. If the calculated mean calibration factors are valid, when you switch to the fresh blood
screen or switch to another calibration mode, a message box will pop up.
Click “Yes” to save the current mean calibration factors and refresh the calibration factors and
date in the table at the “Calibration Factors” screen. Then, you can switch to another screen
or calibration mode. Click “No” to close the message box and switch to another screen or
calibration mode without saving the mean calibration factors and all the calibration data.
Other Operations
Print
9-20
Using the Calibration Programs
If the mean calibration factors are invalid, then a message box will pop up when you clicking
the “Print” button.
Click “OK”, then the cell of the first invalid calibration factor will be highlighted and the data in
the cell will not be cleared.
If the mean calibration factors are valid, click the “Print” button to print the following data in
the form of list, namely, the calibration factors of the sample in the “Calculated Result” table,
the results included in calculating the calibration factors and the mean calibration factors.
1. Run the calibrator at least three times and check whether the means of the obtained
results are within the expected ranges.
2. Run the low, normal and high level controls each for three times at least, and check
whether the means of the obtained results are within the expected ranges.
3. Run at least three fresh blood samples with known reference values, each for six times
at least, and check whether the means of the obtained results are within the expected
ranges.
9-21
Using the Calibration Programs
The history list displays the calibration information for the latest 80 calibrations; you can view
the contents in the list, but you are not allowed to modify or delete any content.
1. If the calibration mode of the selected record is “Adjust Manually”, the new calibration
factor and the original calibration factor are displayed in grey edit box.
2. If the calibration mode of the selected record is “Calibrator”, the new calibration factor,
the original calibration factor and all medium data are displayed in grey edit box.
9-22
Using the Calibration Programs
The calibration data marked with “√” are used for calculation of calibration
factors.
3. If the calibration mode of the selected record is “Fresh Blood”, the calibration factor,
mean calibration factor and original calibration factor of each sample are displayed in
9-23
Using the Calibration Programs
The calibration factors marked with “√” are used for calculation of mean
calibration factor.
If the calibration factor of a sample is displayed, you can click the “Detail…” button to display
all medium data.
9-24
Using the Calibration Programs
Click “Close” to close the dialog box and return to the “History” screen.
The calibration data marked with “√” are used for calculation of sample
calibration factors.
Click the “Print” button down to the screen to print all calibration history records in table
format.
9-25
10Maintaining Your Analyzer
10.1 Introduction
Preventive and corrective maintenance procedures are required to keep the analyzer in a
good operating condition. This analyzer provides multiple maintenance functions for this
purpose.
This chapter introduces how to use the provided functions to maintain and troubleshoot your
analyzer.
All the analyzer components and surfaces are potentially infectious, take
proper protective measures for operation or maintenance.
Exercise caution to avoid contact with the sharp sample probe when
performing maintenance.
10-1
Maintaining Your Analyzer
10.2 Maintenance
At the “Motor” and “Valve” tabs of the “Self-test” screen and the “Status”
screen, the analyzer can not sleep.
Click the “Menu” button on the screen, then select “Shutdown””Sleep” on the pop-up
menu.
10-2
Maintaining Your Analyzer
After the preparation is complete, the progress bar closes automatically and the analyzer
enters the sleep status.
When the analyzer is sleeping, the analysis status icon at the screen
displays in red. The indicator on the analyzer displays in red at the same
time.
You can not run any sample when the analyzer is sleeping.
You can perform the operations without the cooperation of the analyzer
when it is sleeping, namely, communication and print etc.
If any error happens during the process of entering the sleep status, the
analyzer will not sleep but alarm for the error. See Chapter 11
Troubleshooting Your Analyzer for solutions.
Click the “Menu” button on the screen, then select “Shutdown””Cancel” on the pop-up
menu.
10-3
Maintaining Your Analyzer
After the exiting is complete, the progress bar closes automatically and the analyzer exits the
sleep mode.
[RUN] key
Press the [RUN] key on the analyzer to wakeup the analyzer from sleeping.
10-4
Maintaining Your Analyzer
After the exiting is complete, the progress bar closes automatically and the analyzer exits the
sleep mode.
If any error happens during the process of exiting the sleep status, see
Chapter 11 Troubleshooting Your Analyzer for details to remove the error.
After waking up from sleeping successfully, the analyzer will return to the
certain status before sleeping. The analysis status icon at the screen
displays in green. The indicator on the analyzer displays in green at the
same time.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
After installing a new container of reagent, keep it still for a while before use.
When you have changed the diluent, cleansers or lyses, run a background
to see if the results meet the requirement.
10-5
Maintaining Your Analyzer
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
Click the “Replace Reagent” tab to enter the “Replace Reagent” screen.
10-6
Maintaining Your Analyzer
Diluent
LEO(I) lyse
LEO(II) lyse
LH lyse
Cleanser
Keep the diluent container from any strong vibration or collision with other
objects. Otherwise unreliable error messages may be reported.
While replacing the diluent container, be sure to follow the following steps: 1)
install the supporting board as shown below; 2) insert the cap assembly
(shown in the figure below) into the diluent container vertically, and then
secure the cap. Otherwise unreliable error messages may be reported.
10-7
Maintaining Your Analyzer
1. Double click the icon of the desired reagent, and then enter the lot No. and expiration
date of the new reagent.
The check box “Change New Vial” should be selected if you wish to change
a container of reagent. Then, the edit box of “Lot No.” and “Exp. Date” will
be activated for you to enter the new lot No. and expiration date. When
finishing the replacement, the analyzer will save the new lot No. and
expiration date and then automatically modify the open-container expiration
date. The “Change New Vial” check box is selected as default.
The check box “Change New Vial” should not be selected if you only replace
the reagent in the container. Then, the edit box “Lot No.” and “Exp. Date”
will be unavailable. When finishing the replacement, the analyzer will not
modify the open-container expiration date.
1 to 16 digits can be entered into the box of “Lot No.” and an empty entry is
allowed.
After the “Use barcode scanner” is selected, you can enter the expiration
date of the reagents by the barcode scanner.
2. Click the “Replace” button to save the entered expiration date and lot No. and start
replacing.
3. After the replacing is completed, the following prompt will pop up.
10-8
Maintaining Your Analyzer
10.2.4 Cleaning
You should clean the corresponding components under the following circumstances:
When the background of WBC and/or HGB relative parameters exceeds the Ref. Range,
you should clean the WBC bath.
When the background of RBC and (or) PLT relative parameters exceeds the Ref. Range,
you should clean the RBC bath.
When the background of the scattergram has abnormal excessive cells, you should
clean the DIFF Bath.
When the background of the scattergram has abnormal excessive cells, or bad
differential of WBC, you should clean the flow cell.
When the sample probe is dirty, you should clean the sample probe.
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-9
Maintaining Your Analyzer
WBC bath
RBC bath
DIFF bath
Flow cell
Sample probe
Do as follows to clean:
10-10
Maintaining Your Analyzer
10.2.5 Unclogging
When clogging happens, you should perform the unclogging procedure.
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-11
Maintaining Your Analyzer
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-12
Maintaining Your Analyzer
10-13
Maintaining Your Analyzer
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-14
Maintaining Your Analyzer
When the problems including the background results exceeds the Ref. Range, bad
differential of scattergram and clogging still exist after other maintenances have been
10-15
Maintaining Your Analyzer
adopted.
If your analyzer is to run few samples, you should perform this procedure every two
weeks.
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
1. Double click the icon of “Probe Cleanser Soak”, and then a message box will pop up.
10-16
Maintaining Your Analyzer
2. Click “Yes”, and then the following progress bar will pop up and the analyzer is preparing
to soak.
3. After the preparation is done, the message box will pop up.
4. Perform the aspiration of the cleanser as instructed. Then, the priming process starts
automatically after the aspiration.
10-17
Maintaining Your Analyzer
5. When the priming is completed, the progress bar closes and a count-down box will pop
up. The soaking process starts.
6. The soaking process will last about 20 minutes. You may click the “Stop Soaking”
button in the message box to stop it. If you stop soaking in less than 5 minutes, the
following message box will pop up.
10-18
Maintaining Your Analyzer
10. Do the above procedures to perform the probe cleanser soak if necessary.
10-19
Maintaining Your Analyzer
Take the following steps to perform the probe cleanser soak (DIFF bath):
1. Double click the “Soak DIFF Bath” icon, and then a message box will pop up.
2. Click “Yes”, and then the progress bar shown below will pop up and the analyzer is
preparing.
10-20
Maintaining Your Analyzer
4. After aspirating the probe cleanser as instructed, the following progress bar will pop up and
the analyzer starts priming automatically.
5. When the priming is completed, the progress bar closes and a count-down box will pop up.
The soaking process starts.
6. The soaking process will last about 20 minutes. You may click the “Stop Soaking” button
in the dialog box to stop it. If you stop soaking in less than 5 minutes, the following message
box will pop up.
10-21
Maintaining Your Analyzer
Do the above procedures to perform the probe cleanser soaking for WBC bath and RBC bath
if necessary.
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-22
Maintaining Your Analyzer
10-23
Maintaining Your Analyzer
2. Click “Ok” and then the progress bar shown below will pop up and the analyzer is
preparing.
3. When the preparation is done, the progress bar closes and a count-down box will pop up.
The soaking process starts.
4. You may click the “Stop Soaking” button in the dialog box to stop the soaking process. If
you stop soaking in less than 4 hours, the following message box will pop up.
10-24
Maintaining Your Analyzer
5. The cleaning process starts automatically after the soaking progress is done
You should perform the cleanser soak regularly to ensure the analyzer
works in good condition.
10-25
Maintaining Your Analyzer
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
Then, click the “Maintain the whole device” tab to enter the screen.
1. Double click the icon of “Fluidics initialization”, and then a message box will pop up.
10-26
Maintaining Your Analyzer
2. Click the “Ok” button to start initialization and “Fluidics initializing…” will be displayed in
the information area down to the screen.
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-27
Maintaining Your Analyzer
Then, click the “Maintain the whole device” tab to enter the screen.
1. Double click the icon of “Clean Fluidics”, and then a message box will pop up.
10-28
Maintaining Your Analyzer
2. Click the “Yes” button to start cleaning and “Fluidics cleaning…” will be displayed in the
information area down to the screen.
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
10-29
Maintaining Your Analyzer
Then, click the “Maintain the whole device” tab to enter the screen.
1. Double click the icon of “Empty Fluidics”, and then a message box will pop up.
10-30
Maintaining Your Analyzer
2. Click the “Yes” button to start emptying, and a message box shown below will display.
3. Remove all reagent pickup tube assemblies according to the prompt, and then click “OK”
to start emptying the fluidic system.
4. After the emptying is complete, a message box will display. You should turn off the power
switch according to the prompt displayed on the screen.
After the fluidic system is emptied, you can still use the software.
In the message box of “Please power off the analyzer!”, you can click the
“Restart” button to restart the analyzer.
10-31
Maintaining Your Analyzer
Click the “Menu” button on the screen, and then select “Service””Maintenance” on the
pop-up menu.
Then, click the “Maintain the whole device” tab to enter the screen.
10-32
Maintaining Your Analyzer
1. Double click the icon of “Prepare to Ship”, and then a message box will pop up.
2. Click the “Yes” button to perform the packing up and a message box shown below will
display.
3. Remove all reagent pickup tube assemblies according to the prompt, and then click the
“OK” button to start emptying the fluidic system.
5. Place all reagent pickup tube assemblies into the distilled water, and then click the “Ok”
button to start priming.
10-33
Maintaining Your Analyzer
7. Remove all reagent pickup tube assemblies from the distilled water according to the
prompt, then click “Ok” to start emptying the fluidic system.
8. After the emptying is complete, a message box will display. You should turn off the
power switch according to the prompt displayed on the screen.
After the prepared to ship procedure is done, you can still use the software.
In the message box of “Please power off the analyzer”, you can click the
“Restart” button to restart the analyzer.
10.2.15 Auto-clean
When the sample count times reach or over 100, the analyzer will perform the cleaning
procedure automatically once, and a prompt will be displayed on the screen.
10-34
Maintaining Your Analyzer
Click “Yes”, and then the progress bar shown below will pop up and the analyzer is preparing.
10-35
Maintaining Your Analyzer
After aspirating the probe cleanser as instructed, the following progress bar will pop up and
the analyzer starts the priming automatically.
When the priming is complete, the progress bar closes and a count-down box will pop up.
The soaking process starts.
The soaking process will last about 20 minutes. You may click the “Stop Soaking” button in
the dialog box to stop it after five minutes. The cleaning process starts after the soaking
progress is done.
10-36
Maintaining Your Analyzer
At the “Self-test” or “Status” screen, the analyzer does not ask for
confirmation to perform the probe cleanser soak.
If it is the time to perform the auto prompt for probe cleanser soak but the
analyzer is running or error happened; only after the running is completed
or the error is removed will the auto prompt starts.
After the probe cleanser soaking is complete, the sample count times will
reset to zero automatically.
10-37
Maintaining Your Analyzer
Do as the prompt instructed, and then click the “Ok” button to close the message box.
If it is the time to perform the auto-prompt for sample probe replacing but
the analyzer is in running status, then only after the running is completed
will the auto-prompt starts.
If you do not replace the sample probe after the prompt is popped up, the
prompt will pop up again every time when you finish running the samples
for 100 times.
After you confirmed, the following message box will pop up.
10-38
Maintaining Your Analyzer
Click “Ok” and then the progress bar shown below will pop up and the soaking preparation
starts.
After the preparation is finished, the progress bar will be closed automatically and a
count-down box will pop up. The soaking process starts.
You may click the “Stop Soaking” button in the message box to stop the soaking process. If
you stop soaking in less than 4 hours, the following message box will pop up.
10-39
Maintaining Your Analyzer
The cleaning process starts automatically after the soaking progress is done.
After the cleaning is complete, a message box will pop up. Click the “OK” button to close the
message box.
10-40
Maintaining Your Analyzer
At the “Self-test” or “Status” screen, the analyzer does not ask for
confirmation to perform the preset cleanser soak.
Only when the connecting time of the analyzer and the computer reaches or
over 24hours will the preset cleanser soaking be prompted.
For details to preset the cleanser soaking time and the prompting time, see
chapter 5 Customizing the Analyzer Software.
If it is the time to prompt for cleanser soaking but the analyzer is in running
or error status, then only after the running is completed or the error is
removed will the prompt starts accordingly.
When it reaches the preset time to perform the cleanser soaking but the
analyzer is in running or error status, only after the running is completed or
the error is removed will the cleanser soaking starts accordingly.
When it reaches the preset time to perform the cleanser soaking but the
analyzer is sleeping, and then the analyzer will wake up automatically and
then proceed to perform the cleanser soaking.
10.2.19 Auto-sleep
When the fluidics system stop working for 15 minutes (default), then the analyzer will enter
the sleeping status automatically.
When the analyzer is in sleeping status, a prompt will display on the screen.
10-41
Maintaining Your Analyzer
You can set the waiting time for auto-sleeping, see chapter 5 Customizing
the Analyzer Software for details.
If it is the time to auto-sleep but the analyzer is error status, then only after
the error is removed will the auto-sleeping starts accordingly.
You can perform the operations without the cooperation of the analyzer
when it is sleeping, namely, communication and print etc.
10-42
Maintaining Your Analyzer
If the results of the status testing exceed the normal range, they will be
highlighted by the red background.
Then, click the “Temperature&Pressure” tab and a message box will pop up.
When the sequence is complete, the message box closes automatically and you will enter the
following screen.
10-43
Maintaining Your Analyzer
You can check the information about the temperature and pressure, and also export or print
the information.
Export
1. Click the “Export” button at the bottom of the screen and then select the desired
information from the pop-up message box.
2. Click the “Browse” button, and then a message box will pop up.
10-44
Maintaining Your Analyzer
3. Select the directory and format for the exported file and enter the file name.
4. Click the “Ok” button to save the selected information to the specified location.
If the exporting failed, you should click the “Ok” button and then try again or
change another exporting directory; if it does not help, please contact
Mindray customer service department or your local distributor.
The default format of the exported information is “.txt” and you can also
choose the “.csv” format.
5. Click the “Print” button at the bottom of the screen and then select the desired
information from the pop-up message box.
10-45
Maintaining Your Analyzer
The user of common level can not see the “Analyzer Information” option in
the “Print” message box. If you wish to print the analyzer information,
please log out and then log in as user of administrator level.
10-46
Maintaining Your Analyzer
You can check the information about the voltage and current, and also export or print the
information.
Export
1. Click the “Export” button at the bottom of the screen and then select the desired
information from the pop-up message box.
10-47
Maintaining Your Analyzer
3. Select the directory and format for the exported file and enter the file name.
4. Click the “Ok” button to save the selected information to the specified location.
If the exporting failed, you should click the “Ok” button and then try again or
change another exporting directory; if it does not help, please contact
Mindray customer service department or your local distributor.
The default format of the exported information is “.txt” and you can also
choose the “.csv” format.
1. Click the “Print” button at the bottom of the screen and then select the desired
information from the pop-up message box.
10-48
Maintaining Your Analyzer
The user of common level can not see the “Analyzer Information” option in
the “Print” message box. If you wish to print the analyzer information,
please log out and then log in as user of administrator level.
10-49
Maintaining Your Analyzer
You can check the information about the version and configuration, and export or print them.
Export
1. Click the “Export” button, and then the following message box will pop up.
10-50
Maintaining Your Analyzer
2. Select the directory and format for the exported file and enter the file name.
3. Click the “Ok” button to save the selected information to the specified location.
If the exporting failed, you should click the “Ok” button and then try again or
change another exporting directory; if it does not help, please contact
Mindray customer service department or your local distributor.
The default format of the exported information is “.txt” and you can also
choose the “.cvs” format.
Click the “Print” button at the bottom of the screen to print the information.
10-51
Maintaining Your Analyzer
10.5 Self-test
Then, click the “Syringe and Sampling Mechanism Self-test” tab to enter the following
screen.
10-52
Maintaining Your Analyzer
You can check the status of all items and print the results.
Self-test
2. When the self-testing is finished, a message box will pop up to inform you the normal
testing results. Then, click the “Ok” button to close the message box.
If the testing result is abnormal, you should click the “Ok” button and try
again for several times; if it does not help, please contact Mindray customer
service department or your local distributor.
Click the “Print” button at the bottom of the screen to print the latest testing results of all
items.
10-53
Maintaining Your Analyzer
Click the “Autoloader Assembly Self-test” tab to enter the following screen.
You can check the status of all the items and print the results.
10-54
Maintaining Your Analyzer
Self-test
2. When the self-testing is finished, a message box will pop up to inform you the normal
testing results. Then, click the “Ok” button to close the message box.
If the testing result is abnormal, you should click the “Ok” button and try
again for several times; if it does not help, please contact Mindray customer
service department or your local distributor.
Click the “Print” button at the button of the screen to print the latest testing results of all items.
10.5.3 Valve
Click the “Menu” button on the screen, and then select “Service””Self-test” on the pop-up
menu.
10-55
Maintaining Your Analyzer
Single valve
Click the desired Valve No. (e.g. “1”), then identify whether it works well by judging its sound
when opening and closing.
All valves
After clicking the “All Valves” button, all valves will be tested according to their No. one by
10-56
Maintaining Your Analyzer
Whether the valve works well or not is identified by judging its sound when
opening and closing
The current status (open/close) of all valves is displayed in the “Status” box
of the screen. When testing the valves, they open first and then close.
You can click the “Cancel” button on the progress bar to stop the testing for
all valves.
10.5.4 Others
Click the “Menu” button on the screen, and then select “Service””Self-test” on the pop-up
menu.
10-57
Maintaining Your Analyzer
You can test all above items and print the result.
Self-test
Do as follows to test the WBC volumetric tube filter, RBC volumetric tube filter, WBC aperture
voltage, RBC aperture voltage and count time.
2. When the self-testing is finished, a message box will pop up to inform you the result.
Then, you can click the “Ok” button to close the message box.
10-58
Maintaining Your Analyzer
If the testing result is abnormal, you should click the “Ok” button and try
again for several times; if it does not help, please contact Mindray customer
service department or your local distributor.
1. Double-click the “Built-in Barcode Scanner” button, and a message box will pop up.
2. The test starts and a progress bar will appear. When the test is completed, a message
will pop up.
3. Check the sample ID as prompted in the message box. Then, click the “Ok” button to
close the box.
Click the “Print” button at the button of the screen to print the latest testing results of the
items.
10-59
Maintaining Your Analyzer
10.6 Counter
Click the “Menu” button on the screen, and then select “Service””Counter” on the pop-up
menu.
You can check the statistic information of all the above items and the detail statistic
information of some items.
10-60
Maintaining Your Analyzer
You can check the detail information for the sample count times, QC times and calibration
times.
You can click the “Detail…” button next to the “Sample Count Times” to display the detail
statistic information about the sample count times.
You can click the “Detail…” button next to the “QC Times” to display the detail statistic
information about the QC times.
You can click the “Detail…” button next to the “Calibration Times” to display the detail
10-61
Maintaining Your Analyzer
Click the “Print” button at the bottom of the screen to print all the statistic information of the
current screen.
10-62
Maintaining Your Analyzer
10.7 Log
If you add a new record when the log is full, the newest record will overwrite
the oldest automatically.
Then, click the “Set Paras” tab to enter the following screen..
10-63
Maintaining Your Analyzer
You can check the log information, enter remark information and also export and print the
information.
Remark
1. Enter the remark information in the “Remark” box of the desired log record.
2. Click the “Save” button at the bottom of the screen to save the remark.
Print
Click the “Print” button at the bottom of the screen. You can select “Date Range” or “No.
Range” to determine the print range.
10-64
Maintaining Your Analyzer
1) Enter the starting date and finishing date of the records you want to print.
2) Click “Ok” button to print the selected log.
1) Enter the starting date and finishing date of the log you want to print.
2) Click the “Ok” button to print the selected log.
Detail
Click the “Detail…” button to check the details of the highlighted record.
10-65
Maintaining Your Analyzer
Then, click the “Other Logs” tab to enter the following screen.
You can check the log information, enter remark information and also print the information.
Remark
1. Enter the remark information in the “Remark” box of the desired log record.
2. Click the “Save” button at the bottom of the screen to save the remark.
10-66
Maintaining Your Analyzer
Print
Click the “Print” button at the bottom of the screen. You can select “Date Range” or “No.
Range” to determine the print range.
1) Enter the starting date and finishing date of the records you want to print.
2) Click “Ok” button to print the selected log.
1) Enter the starting date and finishing date of the log you want to print.
2) Click the “Ok” button to print the selected log.
Detail
Click the “Detail…” button to check the details of the highlighted record.
Click the “Menu” button on the screen, and then select “Service””Log” on the pop-up
10-67
Maintaining Your Analyzer
menu.
Then, click the “Error Info.” tab to enter the following screen.
You can check the log information, enter remark information and also print the information.
Remark
1. Enter the remark information in the “Remark” box of the desired log record.
2. Click the “Save” button at the bottom of the screen to save the remark.
10-68
Maintaining Your Analyzer
Print
Click the “Print” button at the bottom of the screen. You can select “Date Range” or “No.
Range” to determine the print range.
1) Enter the starting date and finishing date of the records you want to print.
2) Click “Ok” button to print the selected log.
1) Enter the starting date and finishing date of the log you want to print.
2) Click the “Ok” button to print the selected log.
Detail
Click the “Detail…” button to check the details of the highlighted record.
The “All Logs” tab displays all the available log information for the current
user.
Click the “Menu” button on the screen, and then select “Service””Log” on the pop-up
10-69
Maintaining Your Analyzer
menu.
Then, click the “All Logs” tab to enter the following screen.
You can check the log information, enter remark information and also print the information.
Remark
1. Enter the remark information in the “Remark” box of the desired log record.
2. Click the “Save” button at the bottom of the screen to save the remark.
10-70
Maintaining Your Analyzer
Print
Click the “Print” button at the bottom of the screen. You can select “Date Range” or “No.
Range” to determine the print range.
1) Enter the starting date and finishing date of the records you want to print.
2) Click “Ok” button to print the selected log.
1) Enter the starting date and finishing date of the log you want to print.
2) Click the “Ok” button to print the selected log.
Detail
Click the “Detail…” button to check the details of the highlighted record.
10-71
11Troubleshooting Your Analyzer
11.1 Introduction
This chapter contains information that is helpful in locating and correcting problems that may
occur during operation of your analyzer.
11-1
Troubleshooting Your Analyzer
The red error message means that the analyzer will terminate the current action
immediately and you can not perform any operation.
The orange error message means that the analyzer will terminate the current action
immediately.
The blue error message means that the analyzer can still proceed with the current action,
but other operations related to the error(s) of this kind will be restricted.
The green error message means that the analyzer can still proceed with the current
action and other operations will not be restricted.
You can see the error name(s) and the corresponding troubleshooting information in the
pop-up message box. The error names are displayed in order.
11-2
Troubleshooting Your Analyzer
You can click the error name in the message box to select (highlight) it and check the
corresponding troubleshooting information in the “Troubleshooting” list under the message
box. The troubleshooting information of the first error will display (default). Follow the
instructions in the message box to remove the error(s)
Remove Error
Press the “Remove error” button, then the system will remove the error automatically if
possible. If the error(s) still exists, you should follow the instructions of the troubleshooting to
remove the error(s).
Silent
Click the “Close” button to close the “Error” message box, but the corresponding error
message will display in the error message area. If you click the error message again, the
“Error” message box will be re-opened.
The possible error(s) and the corresponding troubleshooting information are listed below:
Error Name Troubleshooting Information
1. Please turn off the analyzer power directly and restart the
analyzer later.
Voltage error
2. If the error still exists, contact our customer service
department.
Laser diode current abnormal 1. Please turn off the analyzer power directly and restart the
analyzer later.
2. If the error still exists, contact our customer service
department.
1. Click the “Remove error” button to remove this error.
Drive board communication
2. If the error still exists, contact our customer service
error
department.
1. Click the “Remove error” button to remove this error.
Sheath fluid channel clog 2. If the error still exists, contact our customer service
department.
1. Click the “Remove error” button to remove this error.
Syringe action error 2. If the error still exists, contact our customer service
department.
1. Click the “Remove error” button to remove this error.
Sample probe action error 2. If the error still exists, contact our customer service
department.
11-3
Troubleshooting Your Analyzer
11-4
Troubleshooting Your Analyzer
11-5
Troubleshooting Your Analyzer
11-6
Troubleshooting Your Analyzer
11-7
Troubleshooting Your Analyzer
cable is damaged.
3. If the cable is not damaged, click the “Remove error”
button to remove the error.
4. If the error still exists, contact our customer service
department.
1. Click the “Remove error” button to remove this error.
Autoloader board
2. If the error still exists, contact our customer service
communication error
department.
1. Click the “Remove error” button to remove this error.
Mix mechanism action error 2. If the error still exists, contact our customer service
department.
1. Click the “Remove error” button to remove this error.
2. If the error still exists, please try to remove the tube rack
Autoloader mechanism action from the autoloader and then click the “Remove error” button
error again.
3. If the error still exists, contact our customer service
department.
1. Remove the rack(s) from the unloading tray.
Unloading tray is full 2. If the error still exists, contact our customer service
department.
1. Check whether the barcode is pasted incorrectly,
damaged, or illegible.
2. If there is nothing wrong with the barcode, click the
Scanner error
“Remove error” button to check the built-in barcode scanner.
3. If the error still exists, contact our customer service
department.
1. Check if the sample compartment door is unexpectedly
open.
Closed-tube sampler door
2. Click the “Remove error” button to remove this error.
open failed
3. If the error still exists, contact our customer service
department.
11-8
12Customizing the Print Template
12.1 Introduction
You can modify the print template based on the default one provided by the software in order
to customize the format of the report.
After customizing and saving a template, you can select the newly customized one in the print
setup. And then, the report will be printed in the customized template.
12-1
Customizing the Print Template
1. Click the “Customize” button to enter the “PrintTemplate” screen, and the following
message box will pop up.
2. Enter the correct user name and password in the message box, and go to the print
12-2
Customizing the Print Template
12-3
Customizing the Print Template
Click on the "Report" tab in the "ProjectProperty" area to display all existing templates
in the current template library. Click one of the template names and the corresponding
template will be displayed in the working area.
Click "File" "Open" on the menu bar or the button on the toolbar, and then
specify the directory and select the template file. Click "Open" to open the template.
You can click the button (line control) to draw a straight line or an oblique line in the
working area.
You can click the button (title control) to add a title in the working area.
You can click the button (label control) to add the fixed text information in the
working area.
You can click the button (edit control) to add details associated with the print
You can click the button (picture control) to arrange the location and size of the
12-4
Customizing the Print Template
You can click the button (table control) to add a table in the working area.
Inserting a Business
A business is a set of controls which can be inserted in the template to facilitate the editing
process. Do as follows to insert a business:
1. Click "Insert" on the menu bar and choose "Head", "Body" or "Tail". The following
message box will pop up.
2. Select the desired business name in the pull-down list. Click the "Ok" button to close the
message box and insert the selected business.
Press and hold the "Ctrl" key on the keyboard, and at the same time, click on the
controls you want to select.
Click on the template in the working area and drag the mouse to enclose the controls
you want to select in the rectangular box displayed.
Select the control(s) you want to move. Left click and hold the mouse, and then move the
control to the destination and release.
Select the control(s) you want to move. Press and hold the "Ctrl" key, and move the
control using the arrow keys on the keyboard.
12-5
Customizing the Print Template
12-6
Customizing the Print Template
1. On the menu bar, click "File" "Import", and the following message box will pop up.
2. Choose the import type and click "Ok". Then the following message box will pop up.
3. Select the template file you want to import and click "Open" to import the template into the
current library. The name of the imported template will be displayed under the "Report" tab,
shown as follows.
12-7
Customizing the Print Template
2. Double click the template you want to export to open it in the working area.
3. On the menu bar, click "File" "Export", and the following message box will pop up.
12-8
Customizing the Print Template
4. Specify the directory you want to save the template and enter the file name. Click "Save" to
save the template.
Click "File" "Preview" on the menu bar or the button on the toolbar to preview
the current template.
Click "File" "Print" on the menu bar or the button on the toolbar to print the current
template.
Click "File" "Delete" on the menu bar or the button on the toolbar to delete the
current template.
12-9
Customizing the Print Template
3. Click "Business" "Save" on the menu bar, and the following message box will pop up.
Enter the information of the business in corresponding boxes and click "Ok" to save the
business.
2. Select the right machine model and click "Ok" to load the template library for this model.
3. When the loading is completed, all the templates in the loaded library will be displayed
under the "Report" tab.
12-10
13Appendices
A Index
A D
B
E
Background, 11-7
barcode, 5-5, 6-49 Electrical, 3-6
Bas# Eos#
definition, 3-7 definition, 3-8
formula, 3-7 formula, 3-8
Bas% Eos%
definition, 3-7 definition, 3-8
formula, 3-7 formula, 3-8
error
drive, 11-3
C
optical, 11-4
manual, 9-5
Calibrators, 2-32 F
Cleaning, 10-9
Cleanser, 2-31, 11-5 flags, 6-25
Closed-tube, 6-12, 11-8 Flow, 3-5
Compare, 7-39 Flushing, 10-14
controls, 2-32
CT-PD, 6-21
CT-WB, 6-17
CV, 7-26
A-1
Appendices
Mon%
H definition, 3-7
formula, 3-7
HCT
formula, 3-11
Help, 2-28 N
HGB, 3-9
formula, 3-9 Network, 11-8
Neu#
definition, 3-8
I formula, 3-8
Neu%
Initialization, 6-5
definition, 3-7
Installation, 4-2
formula, 3-7
No, 11-5, 11-6
L
Laser, 1-13 O
LEO, 2-31
Optical, 11-4
LEO(I), 11-4
LEO(II), 11-4
LH, 2-31, 11-5 P
Lym#
definition, 3-8 parameter
A-2
Appendices
Q
T
Quality, 8-2, 8-72, 8-122, 8-163
Tab, 2-17
Table, 7-13
R
Temperature, 10-43
RDV-CV
definition, 3-11 U
RDW-SD
definition, 3-11 Unclogging, 10-11
Reading, 8-11, 8-81 User, 2-12
Reagents, 2-31 User/Lab, 5-9, 5-42
Record, 2-27
Ref., 5-17
V
Remove, 11-3
Replacing, 10-5 Vacuum, 11-4
reviewing Validate
graph, 7-2 graph, 7-10
table, 7-13 table, 7-22
Right, 11-7 Verifying, 9-21
Running, 8-16, 8-86, 8-126, 8-173
W
S
Wash, 3-13
Scanner, 11-8 Waste, 11-7
Self-test, 10-52 WBC, 3-5, 11-7
settings definition, 3-7
auto, 5-32 Worklist, 6-51
communication, 5-26
gain, 5-31
Z
Para., 5-15
print, 5-22
Zapping, 10-12
ref., 5-17
RUO, 5-29
Sheath, 11-3
A-3
B Specifications
B.1 Classification
According to the CE classification, the BC-5180 belongs to In vitro diagnostic medical devices
other than those covered by Annex II and devices for performance evaluation.
B.2 Reagents
M-53 Diluent M-53D Diluent
M-53 Lyse M-53LEO(I) Lyse
M-53LEO(II) Lyse
M-53LH Lyse
Cleanser M-53 Cleanser
M-53P Probe Cleanser
Ф13X75 (mm) (without the cap) evacuated blood collection tube, used for Whole Blood
Mode
Ф12X75 (mm) (without the cap) evacuated blood collection tube, used for Whole Blood
Mode
Ф14X75 (mm) (without the cap) evacuated blood collection tube, used for Whole Blood
Mode
Ф15X75 (mm) (without the cap) evacuated blood collection tube, used for Whole Blood
Mode
Note: the height of the evacuated blood collection tube with cap can not be higher than
83mm.
The tubes of the model that are available in autoloader mode can also be used in CT-WB
mode.
B-1
Appendices
B.4 Parameters
Parameter Abbreviation Default Unit
White Blood Cell count WBC 109/L
Neutrophils number Neu# 109/L
Lymphocytes number Lym# 109/L
Monocytes number Mon# 109/L
Eosinophils number Eos# 109/L
Basophils number Bas# 109/L
Abnormal Lymphocytes number ALY# (RUO) 109/L
Large Immature Cells number LIC# (RUO) 109/L
Neutrophils percentage Neu% %
Lymphocytes percentage Lym% %
Monocytes percentage Mon% %
Eosinophils percentage Eos% %
Basophils percentage Bas% %
Abnormal Lymphocytes percentage ALY% (RUO) %
Large Immature Cells percentage LIC% (RUO) %
Red Blood Cell count RBC 1012/L
Hemoglobin Concentration HGB g/L
Hematocrit HCT %
Mean Corpuscular Volume MCV fL
Mean Corpuscular Hemoglobin MCH pg
Mean Corpuscular Hemoglobin MCHC g/L
Concentration
Red Blood Cell Distribution Width RDW-SD fL
Standard Deviation
Red Blood Cell Distribution Width RDW-CV %
Coefficient of Variation
Platelet count PLT 109 / L
Mean Platelet Volume MPV fL
Platelet Distribution Width PDW None
Plateletcrit PCT %
Red Blood Cell Histogram RBC Histogram None
Platelet Histogram PLT Histogram None
White Blood Cell/Basophils WBC/BASO Histogram None
Scattergram
White Blood Cell Histogram WBC Histogram None
4 differential Scattergram Diff Scattergram None
B-2
Appendices
B.5.2 Throughput
Autoloader mode ≥ 60 samples/ 1 hour
Closed-tube Mode ≥ 50 samples/ 1 hour
B-3
Appendices
B.6.4 Reproducibility
These reproducibility requirements apply only to the situation in which a qualified sample has
been run for 11 times and the results of the 2nd to 11th runs are used to calculate the
reproducibilities.
Parameter Condition Whole Blood Predilute
Reproducibility(CV% / Reproducibility(CV% /
absolute deviation d※) absolute deviation d※)
9
WBC (4.0-15.0)×10 /L ≤2.0% ≤4.0%
Neu% 50.0%-60.0% ±4.0(absolute deviation) ±8.0(absolute deviation)
Lym% 25.0%-35.0% ±3.0(absolute deviation) ±6.0(absolute deviation)
Mon% 5.0%-10.0% ±2.0(absolute deviation) ±4.0(absolute deviation)
Eos% 2.0%-5.0% ±1.5(absolute deviation) ±2.5(absolute deviation)
Bas% 0.5%-1.5% ±0.8(absolute deviation) ±1.2(absolute deviation)
12
RBC (3.50-6.00)×10 /L ≤1.5% ≤3.0%
HGB (110-180) g/L ≤1.5% ≤3.0%
MCV (70-120) fL ≤1.0% ≤2.0%
9
PLT (150-500)×10 /L ≤4.0% ≤8.0%
MPV / ≤4.0% ≤8.0%
※: Absolute deviation d = analysis result – average of analysis results
B.6.5Carryover
Parameter Carryover
WBC ≤ 0.5 %
RBC ≤ 0.5 %
HGB ≤ 0.6 %
HCT ≤ 0.5 %
PLT ≤ 1.0 %
B-4
Appendices
RAM: ≥256 MB
B.7.5 Printer
B.8 Interfaces
One LAN interface
B-5
Appendices
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference, in which case, you may need to take
measures to mitigate the interference.
B.11 Sound
Maximal sound: 67.7 dBA
Relative humidity: 10 % - 90 %
Relative humidity: 10 % - 90 %
B-6
Appendices
Height
Depth
Width
Analyzer
Width(mm) ≤590
Height(mm) ≤525
Depth(mm) ≤570
Weight(Kg) ≤58
B-7
Appendices
B.17 Contraindications
None
B-8
C Communication
C.1 Introduction of communication protocol of the auto
hematology analyzers
Connecting
After starting up, the IPU software connects the LIS server actively according to the settings. If
the connecting is failed, it retries; if the connecting is successful, it keeps the connection to
make sure the data can be sent at any time. If the connection is found disconnected during
operating, it retries to connect.
Data transmitting
Besides batch sending the data at the List Review and QC screen, if auto-communication is
enabled, the IPU software will send the message while the new sample results are obtained.
Sending and receiving the message are synchronous both for batch communication and
auto-communication. i.e. when every message is sent, it will wait for the confirmation. If the
confirmation is received within 10s, then a complete message is sent and the next message
will be sent; if the confirmation is not received within the 10s, then it is regarded that the
sending is failed and it will skip to the sending of the next message.
The communication of QC data records is similar to that of the analysis results: send
messages at the QC screen or QC History screen. Wait for the confirmation after sending each
QC data. If the confirmation is received within 10s, then the message has been sent
successfully; if the confirmation is not received within 10s, then it is regarded that the sending
is failed and it will skip to the sending of the next message.
The bidirectional LIS inquiry communication is different from the processes mentioned above.
The IPU software will send an inquiry (including the sample ID) every time it opens the
C-1
Appendices
bidirectional LIS communication, saves worklists or before counting. The LIS will respond with
a HL7 message based on the message it received, and then IPU will fill in the worklist or
perform counting according to the response. If there is no response within 10s after the inquiry
was sent, it is regarded that the inquiry is failed.
Disconnecting
When exiting the IPU software, the connection will be closed actively. When changing the
communication settings, the connection will also be disconnected and then re-connect
according to the new settings.
The data of sample results etc. are transmitted in the form of UTF-8 coding strings.
The message strings are composed as per the HL7 standard. A message consists of several
segments, each segment consists of several fields, a field consists of several components,
and component consists of several sub components. The segment, field, component and sub
component are divided by separators. The structure of the message is shown in Figure 1.
MSH|^~\&|BC-5380|Mindray|||20080617143943||ORU^R01|1|P|2.3.1||||||UNICODE
C-2
Appendices
PID|1||7393670^^^^MR||Joan^JIang||19900804000000|Female
PV1|1||nk^^001
OBR|1||20071207011|00001^Automated
Count^99MRC||20080508140600|20080508150616|||John||||20080508150000||||||||||HM||||||||
Mindray
OBX|1|IS|08001^Take Mode^99MRC||O||||||F
OBX|2|IS|08002^Blood Mode^99MRC||W||||||F
OBX|3|IS|08003^Test Mode^99MRC||CBC||||||F
OBX|4|IS|01002^Ref Group^99MRC||Woman||||||F
……
TCP/IP is a protocol of byte stream. It doesn’t provide the message boundary.HL7 of top
protocol is based on messages. The function of terminating the message is not provided. In
order to determine the message boundary, the bottom protocol of MLLP is used (such
descriptions are also included in HL7 Interface Standards Version 2.3.1.).
Communication level
Messages are transmitted in the following format:
<SB> ddddd <EB><CR>
Among them:
<SB> = Start Block character (1 byte)
ASCII <VT>,i.e., <0x0B>.Do not confuse with the SOH or STX character in ASCII.
C-3
Appendices
Every HL7 message consists of several segments and ends up with the <CR> character.
Each segment consists of the segment name of three characters and field of changeable
characters, and each field consists of the component and subcomponent. For each message,
the separators of the field, component and subcomponent are defined in the MSH segment.
For example:
MSH|^~\&|BC-5380|Mindray|||20080617143943||ORU^R01|1|P|2.3.1||||||UNICODE
In this message:
The five characters following MSH define the separators to distinguish each field, component
and subcomponent. Although they can be any non-text characters, HL7 standard recommends
the characters in the table below:
Character Meaning
| Field separator
^ Component separator
& Subcomponent separator
~ Repetition separator
\ ESC
The first field of MSH includes every separator. Some field behind are empty because they are
optional and not used by Mindray HL7 interface. Detailed field definition and selection will be
stated in the following contents.
For message of any type, the segments behind MSH appear in the fixed order. The order will
be described in the following contents and the grammar is used to organize the segments
order.
The segment appeared in [] is optional.
The segment appeared in {} can be repeated once or more.
For the field data of ST, TX, FT, and CF, etc., separators may be contained in the string data
like remark, clinical diagnosis and customized gender etc. When coding, the separators in the
original strings shall be transferred into transferred character sequence; then, restore them
when decoding. The transferring principles are shown in the table:
Transferred character Original character
\F\ Field separator
\S\ Component separator
\T\ Subcomponent separator
\R\ Repetition separator
\E\ Transferred separator
\.br\ <CR>,i.e. end character of segment
C-4
Appendices
Note: “\” in the transferred character sequence represents the transferred separator. Its value
is defined in MSH segment.
All the data information can be expressed by different types of HL7 fields. Only part of the HL7
standard is used in the communication protocol, see D4 Appendix for details.
1. The main unit directly sends the test results (or QC data) to LIS as Figure 2 shows.
C-5
Appendices
ORU^R01 message: it is mostly used for the transmission of the test results and QC data.
ORU Observational Results (Unsolicited) Description
MSHMessage header, necessary, including the communication information of message No.,
sending time, message separator and coding method, etc
{
PID Patient basic information, including patient name, gender, patient ID and birthday, etc
[PV1] Patient visit information, including patient type, department, bed No. and charge, etc
{
OBRsample information, including sample No., operator and run time, etc
{[OBX]} test data, including test results and work mode, etc
}
}
ORM^O01 message: Common order message, all the actions related to order basically use
the message of this type. For example, create a new order or cancel an order. Here, the main
unit requests LIS to re-fill the order message.
ORM General Order Message Description
MSH Message header
{ORC} Common message of Order, including the No. information of the sample searched
ORR^O02 message: affirming of the ORM^O01 message. Here, returning the completed
information of order (i.e. worklist).
ORR^O02 General Order Response Message Description
MSH Message header
MSAMessage affirm
[PIDPatient basic information
C-6
Appendices
MSH
C-7
Appendices
C-8
Appendices
received message.
12 Version ID VID 60 HL7 version information; the 2.3.1
value is “2.3.1”.
18 Character ID 10 Character set. UNICODE
Set The value is “UNICODE”, and
the message is expressed by
Unicode string.
MSA
C-9
Appendices
102 Data type error Segment data type error, e.g. numbers are replaced
by characters
103 Table value not Table value is not found; not used temporarily
found
Rejection status AR
code:
200 Unsupported Message type is not supported
message type
201 Unsupported event Event code is not supported
code
202 Unsupported Processing ID is not supported
processing id
203 Unsupported Version ID is not supported
version id
204 Unknown key Unknown key identifier, e.g. transmitting the patient
identifier information that is not exited
205 Duplicate key Repeated key words existed
identifier
206 Application record Issues can not be executed at application program
locked saving level, e.g. database is locked
207 Application internal Other interior errors of application program
error
PID
C-10
Appendices
PV1
OBR
C-11
Appendices
C-12
Appendices
the clinical
diagnostic
information in the
patient info.
14 Specimen TS 26 Sample received 20080508150000
Received time
Date/Time * To express the
delivery time.
15 Specimen CM 300 Sample source
Source * Its value in HL7
message on the
auto hematology
analyzers:
“BLDV”- Venous
blood
“BLDC”- Capillary
blood
22 Results TS 26 Results
Rpt/Status report/Status
Chng - Change -
Date/Time + Date/Time
To be used as
validating time.
24 Diagnostic ID 10 Diagnostic ID, the HM
Serv Sect ID value is “HM”,
means Hematology.
28 Result Copies XCN 60 Result copies to
To To indicate the
validater.
32 Principal CM 200 Principal result Mindray
Result interpreter
Interpreter + To be used as
tester in the sample
message
To be used as “set
by” in the QC
message
To be used as
“Operator” in the
QC run message
C-13
Appendices
OBX
The OBX(Observation/Result) segment contains the parameter information of each test result.
Message used for example:
OBX|6|NM|6690-2^WBC^LN||9.81|10*9/L|4.00-10.00|N|||F||E
See Table 7 for definition of the fields used.
C-14
Appendices
C-15
Appendices
O – Expired reagent
E – Active editing
e – Passive editing
ORC
Sample message
C-16
Appendices
MSH|^~\&|BC-5380|Mindray|||20080617143943||ORU^R01|1|P|2.3.1||||||UNICODE
PID|1||7393670^^^^MR||Joan^JIang||19900804000000|Female
PV1|1||nk^^001
OBR|1||20071207011|00001^Automated
Count^99MRC||20080508140600|20080508150616|||John||||20080508150000||||||||||HM||||||||
Mindray
OBX|1|IS|08001^Take Mode^99MRC||O||||||F
OBX|2|IS|08002^Blood Mode^99MRC||W||||||F
OBX|3|IS|08003^Test Mode^99MRC||CBC||||||F
OBX|4|IS|01002^Ref Group^99MRC||Woman||||||F
OBX|5|NM|30525-0^Age^LN||18|yr|||||F
OBX|6|NM|6690-2^WBC^LN||9.81|10*9/L|4.00-10.00|N|||F||E
OBX|7|NM|704-7^BAS#^LN|||10*9/L|0.00-0.10||||F
OBX|8|NM|706-2^BAS%^LN||||0.000-0.010||||F
OBX|9|NM|751-8^NEU#^LN|||10*9/L|2.00-7.00||||F
OBX|10|NM|770-8^NEU%^LN||||0.500-0.700||||F
OBX|11|NM|711-2^EOS#^LN|||10*9/L|0.02-0.50||||F
OBX|12|NM|713-8^EOS%^LN||||0.005-0.050||||F
OBX|13|NM|731-0^LYM#^LN|||10*9/L|0.80-4.00||||F
OBX|14|NM|736-9^LYM%^LN||||0.200-0.400||||F
OBX|15|NM|742-7^MON#^LN|||10*9/L|0.12-0.80||||F
OBX|16|NM|5905-5^MON%^LN||||0.030-0.080||||F
OBX|17|NM|26477-0^*ALY#^LN|||10*9/L|0.00-0.20||||F
OBX|18|NM|13046-8^*ALY%^LN||||0.000-0.020||||F
OBX|19|NM|10000^*LIC#^99MRC|||10*9/L|0.00-0.20||||F
OBX|20|NM|10001^*LIC%^99MRC||||0.000-0.025||||F
OBX|21|NM|789-8^RBC^LN||4.53|10*12/L|3.50-5.00|N|||F
OBX|22|NM|718-7^HGB^LN||65|g/L|110-150|L|||F
OBX|23|NM|787-2^MCV^LN||89.5|fL|80.0-100.0|N|||F
OBX|24|NM|785-6^MCH^LN||14.4|pg|27.0-31.0|L|||F
OBX|25|NM|786-4^MCHC^LN||160|g/L|320-360|L|||F
OBX|26|NM|788-0^RDW-CV^LN||0.133||0.115-0.145|N|||F
OBX|27|NM|21000-5^RDW-SD^LN||50.9|fL|35.0-56.0|N|||F
OBX|28|NM|4544-3^HCT^LN||0.405||0.370-0.480|N|||F
OBX|29|NM|777-3^PLT^LN||212|10*9/L|100-300|N|||F
OBX|30|NM|32623-1^MPV^LN||6.6|fL|7.0-11.0|L|||F
OBX|31|NM|32207-3^PDW^LN||15.4||15.0-17.0|N|||F
OBX|32|NM|10002^PCT^99MRC||1.40|mL/L|1.08-2.82|N|||F
OBX|33|IS|12014^Anemia^99MRC||T||||||F
C-17
Appendices
OBX|34|IS|15180-3^Hypochromia^LN||T||||||F
OBX|35|NM|15001^WBC Histogram. Left Line^99MRC||7||||||F
OBX|36|NM|15002^WBC Histogram. Right Line^99MRC||65||||||F
OBX|37|NM|15003^WBC Histogram. Middle Line^99MRC||30||||||F
OBX|38|ED|15008^WBC Histogram. BMP^99MRC||^Image^BMP^Base64^ … … WBC
Histogram bmp data……||||||F
OBX|39|NM|15051^RBC Histogram. Left Line^99MRC||26||||||F
OBX|40|NM|15052^RBC Histogram. Right Line^99MRC||164||||||F
OBX|41|ED|15056^RBC Histogram. BMP^99MRC||^Image^BMP^Base64^ … … RBC
Histogram bmp data……||||||F
OBX|42|NM|15111^PLT Histogram. Left Line^99MRC||3||||||F
OBX|43|NM|15112^PLT Histogram. Right Line^99MRC||43||||||F
OBX|44|ED|15116^PLT Histogram. BMP^99MRC||^Image^BMP^Base64^……PLT Histogram
bmp data……||||||F
OBX|45|ED|15200^WBC DIFF Scattergram. BMP^99MRC||^Image^BMP^Base64^……WBC
Diff Scattergram bmp data……||||||F
OBR|2||20071207011|00002^Manual Count^99MRC|||||||||||BLDV
OBX|46|NM|747-6^Myeloblasts%. Manual^LN||0.0|%|||||F
OBX|47|NM|783-1^Promyelocytes%. Manual^LN||0.0|%|||||F
OBX|48|NM|749-2^Myelocytes%. Manual^LN||0.0|%|||||F
OBX|49|NM|740-1^Metamyelocyte%. Manual^LN||0.0|%|||||F
OBX|50|NM|764-1^Neuts Band%. Manual^LN||0.0|%|||||F
OBX|51|NM|769-0^Neuts Seg%. Manual^LN||0.0|%|||||F
OBX|52|NM|714-6^Eosinophils%. Manual^LN||0.0|%|||||F
OBX|53|NM|707-0^Basophils%. Manual^LN||0.0|%|||||F
OBX|54|NM|33831-9^Lymphoblasts%. Manual^LN||0.0|%|||||F
OBX|55|NM|6746-2^Prolymphocytes%. Manual^LN||0.0|%|||||F
OBX|56|NM|737-7^Lymphocytes%. Manual^LN||0.0|%|||||F
OBX|57|NM|29261-5^Abnormal Lymphs%. Manual^LN||0.0|%|||||F
OBX|58|NM|33840-0^Monoblasts%. Manual^LN||0.0|%|||||F
OBX|59|NM|13599-6^Promonocytes%. Manual^LN||0.0|%|||||F
OBX|60|NM|744-3^Monocytes%. Manual^LN||0.0|%|||||F
OBX|61|NM|18309-5^NRBCs%. Manual^LN||0.0|%|||||F
OBX|62|NM|31112-6^Reticulocytes%. Manual^LN||0.0|%|||||F
OBX|63|NM|11000^Undefined Cells%. Manual^99MRC||0.0|%|||||F
OBX|64|NM|11001^Other Abnormal Cells%. Manual^99MRC||0.0|%|||||F
Every time a sample result is received, a sample response message composed of two
C-18
Appendices
message segments (MSH and MSA) will be sent. To send a correct response message, take
into consideration that: the MSH-9 field should be ACK^R01 which indicates that it is a sample
response message; If the value in the MSA-2 field is the same with the MSH-10 value of the
analysis result, it indicates that this response message is corresponding to the sent analysis
result. The MSA-2 value in the following example is 1
MSH|^~\&|LIS||||20080617143944||ACK^R01|1|P|2.3.1||||||UNICODE
MSA|AA|1
QC message
The content of the QC message differs from that of the sample analysis result: the MSH-11
value of the QC message is Q which indicates that it is a QC message; each QC message is
corresponding to one QC point in the IPU software which may contain several analysis results.
For example, there is one analysis result in an L-J QC message, while there are two analysis
results and one mean calculation result in an X-R QC message.
A QC messasge is composed of an MSH message head and several analysis results, each of
which contains the PID and OBR segments as the head of the sample message, as well as
several OBX segments to carry parameters and other information. The OBR-4 field of each
analysis result indicates the type of the result (X-R analysis result, X-R mean or L-J analysis
result). See Appendix: Message coding definition for details.
An example of the X-R QC message is shown as follows:
MSH|^~\&|BC-5380|Mindray|||20081120171602||ORU^R01|1|Q|2.3.1||||||UNICODE
PID|1||6666666||||20080807235959
OBR|1||6|00006^XR QCR^99MRC|||20080807142518|||||||||||||||||HM||||||||R&D Engineer
OBX|1|IS|05001^Qc Level^99MRC||M||||||F
OBX|2|IS|08001^Take Mode^99MRC||C||||||F
OBX|3|IS|08002^Blood Mode^99MRC||W||||||F
OBX|4|NM|6690-2^WBC^LN||0.00|10*9/L|||||F
OBX|5|NM|704-7^BAS#^LN||***.**|10*9/L|||||F
OBX|6|NM|706-2^BAS%^LN||**.*|%|||||F
OBX|7|NM|751-8^NEU#^LN||***.**|10*9/L|||||F
OBX|8|NM|770-8^NEU%^LN||**.*|%|||||F
OBX|9|NM|711-2^EOS#^LN||***.**|10*9/L|||||F
OBX|10|NM|713-8^EOS%^LN||**.*|%|||||F
OBX|11|NM|731-0^LYM#^LN||***.**|10*9/L|||||F
OBX|12|NM|736-9^LYM%^LN||**.*|%|||||F
OBX|13|NM|742-7^MON#^LN||***.**|10*9/L|||||F
OBX|14|NM|5905-5^MON%^LN||**.*|%|||||F
OBX|15|NM|789-8^RBC^LN||0.02|10*12/L|||||F
C-19
Appendices
OBX|16|NM|718-7^HGB^LN||0|g/L|||||F
OBX|17|NM|787-2^MCV^LN||***.*|fL|||||F
OBX|18|NM|785-6^MCH^LN||***.*|pg|||||F
OBX|19|NM|786-4^MCHC^LN||****|g/L|||||F
OBX|20|NM|788-0^RDW-CV^LN||**.*|%|||||F
OBX|21|NM|21000-5^RDW-SD^LN||***.*|fL|||||F
OBX|22|NM|4544-3^HCT^LN||0.0|%|||||F
OBX|23|NM|777-3^PLT^LN||4|10*9/L|||||F
OBX|24|NM|32623-1^MPV^LN||**.*|fL|||||F
OBX|25|NM|32207-3^PDW^LN||**.*||||||F
OBX|26|NM|10002^PCT^99MRC||.***|%|||||F
OBX|27|NM|10003^GRAN-X^99MRC||6||||||F
OBX|28|NM|10004^GRAN-Y^99MRC||32||||||F
OBX|29|NM|10005^GRAN-Y(W)^99MRC||20||||||F
OBX|30|NM|10006^WBC-MCV^99MRC||83||||||F
OBX|31|NM|15001^WBC Histogram. Left Line^99MRC||10||||||F
OBX|32|NM|15002^WBC Histogram. Right Line^99MRC||72||||||F
OBX|33|NM|15003^WBC Histogram. Middle Line^99MRC||37||||||F
OBX|34|ED|15008^WBC Histogram. BMP^99MRC||^Image^BMP^Base64^……WBC
histogram data……||||||F
OBX|35|NM|15051^RBC Histogram. Left Line^99MRC||10||||||F
OBX|36|NM|15052^RBC Histogram. Right Line^99MRC||250||||||F
OBX|37|ED|15056^RBC Histogram. BMP^99MRC||^Image^BMP^Base64^……RBC
histogram data……||||||F
OBX|38|NM|15111^PLT Histogram. Left Line^99MRC||3||||||F
OBX|39|NM|15112^PLT Histogram. Right Line^99MRC||24||||||F
OBX|40|ED|15116^PLT Histogram. BMP^99MRC||^Image^BMP^Base64^……PLT histogram
data……||||||F
OBX|41|ED|15200^WBC DIFF Scattergram. BMP^99MRC||^Image^BMP^Base64^……DIFF
scattergram data……||||||F
PID|2||6666666||||20080807235959
OBR|2||6|00006^XR QCR^99MRC|||20080807142640|||||||||||||||||HM||||||||R&D Engineer
OBX|42|IS|05001^Qc Level^99MRC||M||||||F
OBX|43|IS|08001^Take Mode^99MRC||C||||||F
OBX|44|IS|08002^Blood Mode^99MRC||W||||||F
OBX|45|NM|6690-2^WBC^LN||0.00|10*9/L|||||F
OBX|46|NM|704-7^BAS#^LN||***.**|10*9/L|||||F
OBX|47|NM|706-2^BAS%^LN||**.*|%|||||F
OBX|48|NM|751-8^NEU#^LN||***.**|10*9/L|||||F
OBX|49|NM|770-8^NEU%^LN||**.*|%|||||F
C-20
Appendices
OBX|50|NM|711-2^EOS#^LN||***.**|10*9/L|||||F
OBX|51|NM|713-8^EOS%^LN||**.*|%|||||F
OBX|52|NM|731-0^LYM#^LN||***.**|10*9/L|||||F
OBX|53|NM|736-9^LYM%^LN||**.*|%|||||F
OBX|54|NM|742-7^MON#^LN||***.**|10*9/L|||||F
OBX|55|NM|5905-5^MON%^LN||**.*|%|||||F
OBX|56|NM|789-8^RBC^LN||0.02|10*12/L|||||F
OBX|57|NM|718-7^HGB^LN||0|g/L|||||F
OBX|58|NM|787-2^MCV^LN||***.*|fL|||||F
OBX|59|NM|785-6^MCH^LN||***.*|pg|||||F
OBX|60|NM|786-4^MCHC^LN||****|g/L|||||F
OBX|61|NM|788-0^RDW-CV^LN||**.*|%|||||F
OBX|62|NM|21000-5^RDW-SD^LN||***.*|fL|||||F
OBX|63|NM|4544-3^HCT^LN||0.0|%|||||F
OBX|64|NM|777-3^PLT^LN||5|10*9/L|||||F
OBX|65|NM|32623-1^MPV^LN||**.*|fL|||||F
OBX|66|NM|32207-3^PDW^LN||**.*||||||F
OBX|67|NM|10002^PCT^99MRC||.***|%|||||F
OBX|68|NM|10003^GRAN-X^99MRC||28||||||F
OBX|69|NM|10004^GRAN-Y^99MRC||19||||||F
OBX|70|NM|10005^GRAN-Y(W)^99MRC||20||||||F
OBX|71|NM|10006^WBC-MCV^99MRC||60||||||F
OBX|72|NM|15001^WBC Histogram. Left Line^99MRC||10||||||F
OBX|73|NM|15002^WBC Histogram. Right Line^99MRC||72||||||F
OBX|74|NM|15003^WBC Histogram. Middle Line^99MRC||37||||||F
OBX|75|ED|15008^WBC Histogram. BMP^99MRC||^Image^BMP^Base64^……WBC
histogram data……||||||F
OBX|76|NM|15051^RBC Histogram. Left Line^99MRC||10||||||F
OBX|77|NM|15052^RBC Histogram. Right Line^99MRC||250||||||F
OBX|78|ED|15056^RBC Histogram. BMP^99MRC||^Image^BMP^Base64^ ……RBC
histogram data……||||||F
OBX|79|NM|15111^PLT Histogram. Left Line^99MRC||3||||||F
OBX|80|NM|15112^PLT Histogram. Right Line^99MRC||24||||||F
OBX|81|ED|15116^PLT Histogram. BMP^99MRC||^Image^BMP^Base64^ ……PLT histogram
data……||||||F
OBX|82|ED|15200^WBC DIFF Scattergram. BMP^99MRC||^Image^BMP^Base64^ ……DIFF
scattergram data……||||||F
PID|3||6666666
OBR|3||6|00008^XR QCR Mean^99MRC||||||||||||||||||||HM
OBX|83|NM|6690-2^WBC^LN||0.00|10*9/L|||||F
C-21
Appendices
OBX|84|NM|704-7^BAS#^LN||***.**|10*9/L|||||F
OBX|85|NM|706-2^BAS%^LN||**.*|%|||||F
OBX|86|NM|751-8^NEU#^LN||***.**|10*9/L|||||F
OBX|87|NM|770-8^NEU%^LN||**.*|%|||||F
OBX|88|NM|711-2^EOS#^LN||***.**|10*9/L|||||F
OBX|89|NM|713-8^EOS%^LN||**.*|%|||||F
OBX|90|NM|731-0^LYM#^LN||***.**|10*9/L|||||F
OBX|91|NM|736-9^LYM%^LN||**.*|%|||||F
OBX|92|NM|742-7^MON#^LN||***.**|10*9/L|||||F
OBX|93|NM|5905-5^MON%^LN||**.*|%|||||F
OBX|94|NM|789-8^RBC^LN||0.02|10*12/L|||||F
OBX|95|NM|718-7^HGB^LN||0|g/L|||||F
OBX|96|NM|787-2^MCV^LN||***.*|fL|||||F
OBX|97|NM|785-6^MCH^LN||***.*|pg|||||F
OBX|98|NM|786-4^MCHC^LN||****|g/L|||||F
OBX|99|NM|788-0^RDW-CV^LN||**.*|%|||||F
OBX|100|NM|21000-5^RDW-SD^LN||***.*|fL|||||F
OBX|101|NM|4544-3^HCT^LN||0.0|%|||||F
OBX|102|NM|777-3^PLT^LN||5|10*9/L|||||F
OBX|103|NM|32623-1^MPV^LN||**.*|fL|||||F
OBX|104|NM|32207-3^PDW^LN||**.*||||||F
OBX|105|NM|10002^PCT^99MRC||.***|%|||||F
OBX|106|NM|10003^GRAN-X^99MRC||17||||||F
OBX|107|NM|10004^GRAN-Y^99MRC||26||||||F
OBX|108|NM|10005^GRAN-Y(W)^99MRC||20||||||F
OBX|109|NM|10006^WBC-MCV^99MRC||72||||||F
QC response message
The only difference between the QC response message and the analysis result response
message is that the MSH-11 value of the QC response message is Q.
An example of the ACK X-R QC message is shown as follows:
MSH|^~\&|LIS||||20081120171602||ACK^R01|1|Q|2.3.1||||||UNICODE
MSA|AA|1
A bidirectional LIS inquiry message contains a sample ID. After the LIS received the inquiry
message, it will search for the corresponding patient and sample information to provide a
response.
C-22
Appendices
The inquiry message is composed of two message segments: MSH and ORC. The MSH
segment is almost the same with that of the analysis result, except that the MSH-9 value is
ORM^O01. The ORC-3 field should be filled with the receiver code (in this case, the sample ID;
where in the following sample, it is SampleID1). Note that in the autoloading analysis, if there
is a barcode scan error while sending an inquiry message, the sample ID will be “Invalid”.
An example of the inquiry message is shown as follows:
MSH|^~\&|BC-5380|Mindray|||20081120174836||ORM^O01|4|P|2.3.1||||||UNICODE
ORC|RF||SampleID1||IP
When the LIS received an inquiry message, it needs to send back an inquiry response
message. The first two message segments of the inquiry response message are MSH and
MSA. The MSH-9 field (indicating the type of the segment) is filled with ORR^O02, while the
MSA segment should be filled up as shown in the following example of the inquiry response
message. If the LIS gets searching results for the inquiry, there will be PID, PV1, ORC, OBR
and OBX message segments after the two heading segments to provide the patient and
sample information, in the same way as the sample data message does. The ORC segment is
indispensable for an inquiry response message with searching results, in which the ORC-1
value is AF, and ORC-2 is the filter (the sample ID). Note that the OBR-2 field indicates the
sample ID, which should be the same value as in the ORC-2 field; otherwise, the message will
be regarded as incorrect.
An example of the inquiry response message with searching results is shown as follows:
MSH|^~\&|LIS||||20081120174836||ORR^O02|1|P|2.3.1||||||UNICODE
MSA|AA|4
PID|1||ChartNo^^^^MR||^FName||19810506|NT
PV1|1|nk^^Bn4|||||||||||||||||NewCharge
ORC|AF|SampleID1|||
OBR|1|SampleID1||||20060506||||tester|||Diagnose
content....|20060504||||||||20080821||HM||||Validator||||Operator
OBX|1|IS|08001^Take Mode^99MRC||A||||||F
OBX|2|IS|08002^Blood Mode^99MRC||W||||||F
OBX|3|IS|08003^Test Mode^99MRC||CBC||||||F
OBX|4|IS|01002^Ref Group^99MRC||XXXX||||||F
OBX|5|NM|30525-0^Age^LN||1|hr|||||F
OBX|6|ST|01001^Remark^99MRC||remark content....||||||F
An example of the inquiry response message with no search result is shown as follows, in
which the MSA-2 field indicates the result of the response. In this example, the MSA-2 value is
C-23
Appendices
“AR”, indicating the inquiry was rejected; if it is “AE", then there is an error in the inquiry
process.
MSH|^~\&|LIS||||20081120175238||ORR^O02|1|P|2.3.1||||||UNICODE
MSA|AR|9
CE - Code Element
<identifier (ST)> ^ <text (ST)> ^ <name of coding system (ST)> ^ <alternate identifier (ST)> ^
<alternate text (ST)> ^ <name of alternate coding system (ST)>
CM - Composite
<ID (ST)> ^ <check digit (ST)> ^ <code identifying the check digit scheme employed (ID)> ^ <
assigning authority (HD)> ^ <identifier type code (IS)> ^ < assigning facility (HD)>
ED – Encapsulate Data
EI - Entity Identifier
<entity identifier (ST)> ^ <namespace ID (IS)> ^ <universal ID (ST)> ^ <universal ID type (ID)>
FC – Financial Class
HD - Hierarchic designator
FT - Formatted text
This data type is derived from the string data type by allowing the addition of embedded
formatting instructions. These instructions are limited to those that are intrinsic and
independent of the circumstances under which the field is being used.
The value of such a field follows the formatting rules for an ST field except that it is drawn from
a site-defined (or user-defined) table of legal values. There shall be an HL7 table number
associated with IS data types.
C-24
Appendices
The value of such a field follows the formatting rules for an ST field except that it is drawn from
a table of legal values. There shall be an HL7 table number associated with ID data types.
NM - Numeric
PL - Person location
<point of care (IS )> ^ <room (IS )> ^ <bed (IS)> ^ <facility (HD)> ^ < location status (IS )> ^
<person location type (IS)> ^ <building (IS )> ^ <floor (IS )> ^ <location description (ST)>
PT - Processing type
SI - Sequence ID
A non-negative integer in the form of an NM field. The uses of this data type are defined in the
chapters defining the segments and messages in which it appears.
ST – String
TS - Time stamp
In Version 2.3, use instead of the CN data type. <ID number (ST)> ^ <family name (ST)> &
<last_name_prefix (ST) ^ <given name (ST)> ^ <middle initial or name (ST)> ^ <suffix (e.g., JR
or III) (ST)> ^ <prefix (e.g., DR) (ST)> ^ <degree (e.g., MD) (ST)> ^ <source table (IS)> ^
<assigning authority (HD)> ^ <name type code (ID)> ^ <identifier check digit (ST)> ^ <code
identifying the check digit scheme employed (ID)> ^ <identifier type code (IS)> ^ <assigning
facility (HD)> ^ <name representation code (ID)>
In Version 2.3, replaces the PN data type. <family name (ST)> ^ <given name (ST)> &
<last_name_prefix (ST)> ^ <middle initial or name (ST)> ^ <suffix (e.g., JR or III) (ST)> ^
<prefix (e.g., DR) (ST)> ^ <degree (e.g., MD) (IS)> ^ <name type code (ID) > ^ <name
representation code (ID)>
1. In the HL7 message, the OBR-4(Universal Serview ID) field is used for identifying the type
C-25
Appendices
of the test results, for example, to identify the results as sample results, microscope exam
results or QC results in the form of “ID^Name^EncodeSys”. The coding values of the field are
listed in the table below.
2. Each OBX segment contains one test parameter or information of other data and consists of
the following fields: OBX-2, it indicates the HL7 type of the data contained; OBX-3, it is the
mark of the data in the form of “ID^Name^EncodeSys”; OBX-5, it contains the value of the data;
OBX-6, it contains the unit for the parameter, expressing in the ISO standard.
The HL7 types and coding marks of all the communication data are listed in Table 10. The
units of all the communication data are listed in table 11.
C-26
Appendices
05001^Qc
QC Level IS 05001 Qc Level 99MRC
Level^99MRC
Analysis results data
WBC NM 6690-2 WBC LN 6690-2^WBC^LN
BAS NM 704-7 BAS# LN 704-7^BAS#^LN
BAS_PER NM 706-2 BAS% LN 706-2^BAS%^LN
NEU NM 751-8 NEU# LN 751-8^NEU#^LN
NEU_PER NM 770-8 NEU% LN 770-8^NEU%^LN
EOS NM 711-2 EOS# LN 711-2^EOS#^LN
EOS_PER NM 713-8 EOS% LN 713-8^EOS%^LN
LYM NM 731-0 LYM# LN 731-0^LYM#^LN
LYM_PER NM 736-9 LYM% LN 736-9^LYM%^LN
MON NM 742-7 MON# LN 742-7^MON#^LN
MON_PER NM 5905-5 MON% LN 5905-5^MON%^LN
26477- 26477-0^*ALY#^LN
ALY NM *ALY# LN
0
13046- 13046-8^*ALY%^LN
ALY_PER NM *ALY% LN
8
LIC NM 10000 *LIC# 99MRC 10000^*LIC#^99MRC
10001^*LIC%^99MR
LIC_PER NM 10001 *LIC% 99MRC
C
RBC NM 789-8 RBC LN 789-8^RBC^LN
HGB NM 718-7 HGB LN 718-7^HGB^LN
MCV NM 787-2 MCV LN 787-2^MCV^LN
MCH NM 785-6 MCH LN 785-6^MCH^LN
MCHC NM 786-4 MCHC LN 786-4^MCHC^LN
RDW_CV NM 788-0 RDW-CV LN 788-0^RDW-CV^LN
21000- 21000-5^RDW-SD^L
RDW_SD NM RDW-SD LN
5 N
HCT NM 4544-3 HCT LN 4544-3^HCT^LN
PLT NM 777-3 PLT LN 777-3^PLT^LN
32623- 32623-1^MPV^LN
MPV NM MPV LN
1
32207- 32207-3^PDW^LN
PDW NM PDW LN
3
PCT NM 10002 PCT 99MRC 10002^PCT^99MRC
10003^GRAN-X^99M
GRAN-X NM 10003 GRAN-X 99MRC
RC
GRAN-Y NM 10004 GRAN-Y 99MRC 10004^GRAN-Y^99M
C-27
Appendices
RC
10005^GRAN-Y(W)^
GRAN-Y(W) NM 10005 GRAN-Y(W) 99MRC
99MRC
10006^WBC-MCV^99
WBCMCV NM 10006 WBC-MCV 99MRC
MRC
Microscope exam data
882-1^Blood
Blood Type ST 882-1 Blood Type LN
Type^LN
WBC 11156- 11156-7^WBC
ST WBC Morphology LN
Morphology 7 Morphology^LN
RBC 6742-1^RBC
ST 6742-1 RBC Morphology LN
Morphology Morphology^LN
11125- 11125-2^PLT
PLT Morphology ST PLT Morphology LN
2 Morphology^LN
Myeloblasts%. 747-6^Myeloblasts%.
Myeloblast NM 747-6 LN
Manual Manual^LN
Promyelocytes%. 783-1^Promyelocytes
Promyelocyte NM 783-1 LN
Manual %. Manual^LN
Myelocytes%. 749-2^Myelocytes%.
Myelocyte NM 749-2 LN
Manual Manual^LN
Metamyelocyte%. 740-1^Metamyelocyt
MetaMyelocyte NM 740-1 LN
Manual e%. Manual^LN
Neuts Band%. 764-1^Neuts Band%.
BandFormNeut NM 764-1 LN
Manual Manual^LN
Neuts Seg%. 769-0^Neuts Seg%.
SegmentNeut NM 769-0 LN
Manual Manual^LN
Eosinophils%. 714-6^Eosinophils%.
Eosinophils NM 714-6 LN
Manual Manual^LN
Basophils%. 707-0^Basophils%.
Basophils NM 707-0 LN
Manual Manual^LN
33831- Lymphoblasts%. 33831-9^Lymphoblas
Lymphoblast NM LN
9 Manual ts%. Manual^LN
Prolymphocytes 6746-2^Prolymphocyt
Prolymphocytes NM 6746-2 LN
%. Manual es%. Manual^LN
Lymphocytes%. 737-7^Lymphocytes
Lymphocytes NM 737-7 LN
Manual %. Manual^LN
Abnormal 29261-5^Abnormal
29261-
AbnLymph NM Lymphs%. LN Lymphs%.
5
Manual Manual^LN
C-28
Appendices
C-29
Appendices
C-30
Appendices
C-31
Appendices
99MRC
12009^Eosinophilia^9
Eosinophilia IS 12009 Eosinophilia 99MRC
9MRC
12010^Basophilia^99
Basophilia IS 12010 Basophilia 99MRC
MRC
12011^WBC
WBC abnormal IS 12011 WBC Abnormal 99MRC
Abnormal^99MRC
17790- 17790-7^WBC Left
Left Shift? IS WBC Left Shift? LN
7 Shift?^LN
Immature 34165- Imm 34165-1^Imm
IS LN
Granulocyte? 1 Granulocytes? Granulocytes?^LN
Abnormal/Atypic 15192- 15192-8^Atypical
IS Atypical Lymphs? LN
al Lymphocyte? 8 Lymphs?^LN
RBC Lyse 34525- 34525-6^rstRBC^LN
IS rstRBC LN
Resist? 6
12012^Erythrocytosis
Erythrocytosis IS 12012 Erythrocytosis 99MRC
^99MRC
12013^RBC
RBC abnormal RBC Abnormal
IS 12013 99MRC Abnormal
distribution distribution
distribution^99MRC
15150- 15150-6^Anisocytosis
Anisocytosis IS Anisocytosis LN
6 ^LN
15198- 15198-5^Macrocytes
Macrocytosis IS Macrocytes LN
5 ^LN
15199- 15199-3^Microcytes^
Microcytosis IS Microcytes LN
3 LN
10379- 10379-6^RBC Dual
Dimorphologic IS RBC Dual Pop LN
6 Pop^LN
12014^Anemia^99M
Anemia IS 12014 Anemia 99MRC
RC
15180- 15180-3^Hypochromi
Hypochromia IS Hypochromia LN
3 a^LN
HGB 12015^HGB
IS 12015 HGB Interfere 99MRC
Abn/Interfere? Interfere^99MRC
PLT Abnormal PLT Abnormal 12016^PLT Abnormal
IS 12016 99MRC
Distribution Distribution Distribution^99MRC
12017^Thrombocytos
Thrombocytosis IS 12017 Thrombocytosis 99MRC
is^99MRC
Thrombopenia IS 12018 Thrombopenia 99MRC 12018^Thrombopenia
C-32
Appendices
^99MRC
7796-6^Platelet
PLT Clump? IS 7796-6 Platelet Clump? LN
Clump?^LN
3. Part of the OBX messages adopt the customized enumeration values listed in the following
table.
C-33
Appendices
4. Histogram data: according to the software setup, there are several cases for the
communication of the histogram data.
(1)Do not transmit the histogram data.
(2)Transmit histogram data in the form of bitmap. In the OBX segment, the value of the data
type field is “ED”; the value of data is in the form of “^Image^BMP^Base64^……histogram data
in the form of bitmap……”, the “image” herein indicates the image data is transmitted; the
“BMP” is the customized sub-data type, it indicates the BMP bitmap is transmitted; “Base64”
indicates the coding method of the data.
(3)Transmit binary histogram data. In the OBX segment, the value of the data type field is “ED”;
the value of data is in the form of “^Application^Octet-stream^Base64^……histogram
data……”; “Application^Octer-stream” herein is the sub-data of HL7 standard, indicating the
C-34
Appendices
binary data type defined by the application program; “Base64” indicates the coding method of
the data.
Note: to transmit the histogram data in the form of bitmap or binary is determined by the ID
field in the OBX segment.
5. Scattergram data: when transmitting bitmap data, in the OBX segment, the value of the data
type field is “ED”; the value of data is in the form of “^Image^BMP^Base64^……scattergram
data in the form of bitmap……”. The “Image^BMP^Base64” indicates the bitmap data is of
BMP type and coded by Base64.
6. Age communication: the age in the patient information will be transferred as an OBX
message segment composed of an integer and the age unit. If the age in the IPU software is
displayed as “<1”, then the age value in the communication is “0”.
(1) Select the 3 adjacent bytes (i.e. 24 bit) from the data stream to be coded; from left to right,
divide them into 4 groups of 6-bit; then, ASCII string is obtained by mapping as per the Table
12.
Initial data 15H A3H 4BH
Binary data 00010101 10100011 01001011
6-bit group obtained after dividing 000101 011010 001101 001011
Corresponding coding value 5H 1AH 0DH 0BH
Corresponding character F a N L
C-35
Appendices
11 L 28 c 45 t 62 +
12 M 29 d 46 u 63 /
13 N 30 e 47 v
14 O 31 f 48 w (pad) =
15 P 32 g 49 x
16 Q 33 h 50 y
(2) Repeat the coding of procedure (1) continuously till finish coding the data stream.
When the data left is less than 3 bytes, 0 is used to complement to the right. If the whole 6-bit
group obtained is composed of 0, then it is mapped to the “=” character. When one byte is left,
then the obtained coding string consists of two “=” characters; when two bytes are left, then the
obtained coding string consists of one “=” character. The two cases are demonstrated below:
C-36
P/N: 046-001574-00 (2.0)