Cadwell Sierra II EMG
Cadwell Sierra II EMG
Cadwell Sierra II EMG
1
Preserving and Restoring Settings ................................................................................. 42
To preserve settings .......................................................................................................................... 42
To restore settings............................................................................................................................. 42
2
For additional information, click . . . ................................................................................................... 53
For procedures, click . . . ................................................................................................................... 53
Selecting a Common Recording Reference ................................................................... 53
To turn common reference on and off ............................................................................................... 54
Working with Latency and Amplitude Markers................................................................ 54
For procedures, click . . . ................................................................................................................... 55
Working with Cursors...................................................................................................... 55
To apply cursors ................................................................................................................................ 55
To move cursors................................................................................................................................ 56
To clear cursors................................................................................................................................. 56
Working with Waveforms, Traces, and Trials ................................................................. 56
To select a waveform, trace, or trial for subsequent action ............................................................... 56
To store ............................................................................................................................................. 56
To move ............................................................................................................................................ 57
To clear and restore .......................................................................................................................... 57
To delete stored................................................................................................................................. 57
To display in raster or overlay format ................................................................................................ 57
To apply a smoothing filter ................................................................................................................ 57
Averaging Waveforms .................................................................................................... 58
To turn averaging on and off ............................................................................................................. 58
To set the number of sweeps included in the average ...................................................................... 59
To start and stop averaging............................................................................................................... 59
To create a grand average ................................................................................................................ 59
To store ............................................................................................................................................. 59
To clear and restore averages........................................................................................................... 59
MENUS ...............................................................................................................60
File Menu ........................................................................................................................ 60
Edit Menu........................................................................................................................ 60
View Menu ...................................................................................................................... 60
Test Protocols Menu....................................................................................................... 61
Window Menu ................................................................................................................. 62
Help Menu ...................................................................................................................... 62
PROTOCOLS .....................................................................................................62
Reviewing Test Studies (TabData) ................................................................................. 62
EMG................................................................................................................................ 62
About the EMG Test ................................................................................................................... 62
For further details, click . . . ........................................................................................................... 63
For test procedures, click . . . ........................................................................................................ 63
For configuration procedures, click . . ........................................................................................... 63
EMG Windows............................................................................................................................ 63
EMG Toolbars ............................................................................................................................ 64
EMG Data ................................................................................................................................... 65
For procedures, click . . . ............................................................................................................... 66
Running the EMG Test ............................................................................................................... 66
To run the test ............................................................................................................................... 66
Setting Capture Trigger Parameters .......................................................................................... 68
To turn the capture trigger on and off ............................................................................................ 68
To set the capture trigger level ...................................................................................................... 68
To set the capture trigger delay ..................................................................................................... 68
Scoring Muscles ......................................................................................................................... 68
To place a muscle in the scored muscles table ............................................................................. 68
To score a muscle ......................................................................................................................... 68
To score multiple muscles ............................................................................................................. 69
To score all muscles in a group ..................................................................................................... 69
To delete a muscle from the scored muscles table........................................................................ 69
3
To sort muscles in the scored muscles table ................................................................................. 69
See also ............................................................................................................................................ 69
Configuring the Muscle List ........................................................................................................ 70
For procedures, click . . . ............................................................................................................... 70
EMG2.............................................................................................................................. 70
About the EMG2 Test ................................................................................................................. 70
SFEMG and SSFEMG.................................................................................................... 70
About the SFEMG and SSFEMG Tests ..................................................................................... 71
For further details, click . . . ........................................................................................................... 71
For test procedures, click . . . ........................................................................................................ 71
For configuration procedures, click . . ........................................................................................... 71
SFEMG and SSFEMG Windows ................................................................................................ 71
See also ............................................................................................................................................ 72
SFEMG and SSFEMG Toolbars ................................................................................................ 72
Primary toolbar (Tool1SFEMG.bmp) ................................................................................................ 72
Additional controls for SSFEMG ........................................................................................................ 72
SFEMG and SSFEMG Data ....................................................................................................... 73
Running the SFEMG and SSFEMG Tests ................................................................................. 73
To run either test ........................................................................................................................... 73
Analyzing Traces ........................................................................................................................ 75
To review traces ............................................................................................................................ 75
To delete traces............................................................................................................................. 75
To position and adjust the analysis marker ................................................................................... 75
To review traces ............................................................................................................................ 76
See also ............................................................................................................................................ 76
Sample Test Screens ................................................................................................................. 76
To view sample test screens, click . . . .......................................................................................... 76
NCV ................................................................................................................................ 76
About the NCV Test.................................................................................................................... 76
For further details, click . . . ........................................................................................................... 76
For test procedures, click . . . ........................................................................................................ 77
For configuration procedures, click . . ........................................................................................... 77
NCV Windows ............................................................................................................................ 77
____________________................................................................................................................... 78
See also ............................................................................................................................................ 78
NCV data .................................................................................................................................... 78
Site table column headings......................................................................................................... 78
Segment table column headings................................................................................................. 78
NCV Toolbars ............................................................................................................................. 78
NCV Data.................................................................................................................................... 79
Running the NCV Test................................................................................................................ 80
To run the test ............................................................................................................................... 80
Showing or Hiding the First Velocity Value for a Nerve Segment .............................................. 81
To show or hide the first velocity value.......................................................................................... 81
Configuring the Nerve List, Site Table, and Segment Table ...................................................... 82
For procedures, click . . . ............................................................................................................... 82
Common NCV Stimulation and Recording Sites ........................................................................ 82
F-Wave and H-Reflex ..................................................................................................... 83
About the F-Wave and H-Reflex Tests....................................................................................... 83
For further details, click . . . ........................................................................................................... 83
For test procedures, click . . . ........................................................................................................ 83
For configuration procedures, click . . ........................................................................................... 83
F-Wave and H-Reflex Windows ................................................................................................. 83
F-Wave and H-Reflex Data ........................................................................................................ 84
See also............................................................................................................................. 85
F-Wave Toolbars ........................................................................................................................ 85
H-Reflex Toolbars....................................................................................................................... 86
Running the F-Wave and H-Reflex Tests................................................................................... 87
4
To run the test ............................................................................................................................... 87
SEP................................................................................................................................. 88
About the SEP Tests .................................................................................................................. 88
For further details, click . . . ........................................................................................................... 89
For test procedures, click . . . ........................................................................................................ 89
For configuration procedures, click . . ........................................................................................... 89
SEP Windows............................................................................................................................. 89
SEP Toolbars ............................................................................................................................. 90
Primary toolbar (Tool1SEP.bmp)..................................................................................................... 90
Secondary toolbar ............................................................................................................................. 91
SEP Data .................................................................................................................................... 91
Running the SEP Tests .............................................................................................................. 93
To run the SEP tests ..................................................................................................................... 93
BAEP .............................................................................................................................. 94
About the BAEP Test.................................................................................................................. 94
For further details, click . . . ........................................................................................................... 94
For test procedures, click . . . ........................................................................................................ 94
For configuration procedures, click . . ........................................................................................... 94
BAEP Windows .......................................................................................................................... 94
See also ............................................................................................................................................ 95
BAEP Toolbars ........................................................................................................................... 95
Primary toolbar (Tool1BAE.bmp)..................................................................................................... 95
Secondary toolbar (Tool2SEP.bmp) ................................................................................................ 96
BAEP Data.................................................................................................................................. 97
Running the BAEP Test.............................................................................................................. 98
To run the BAEP test..................................................................................................................... 98
AEP............................................................................................................................... 100
About the AEP Test .................................................................................................................. 100
For further details, click . . . ......................................................................................................... 100
For test procedures, click . . . ...................................................................................................... 100
For configuration procedures, click . . ......................................................................................... 100
AEP Windows........................................................................................................................... 101
See also .......................................................................................................................................... 101
AEP Toolbars ........................................................................................................................... 101
Primary toolbar (Tool1AEP.bmp).................................................................................................... 101
Secondary toolbar (Tool2SEP.bmp) ............................................................................................... 102
AEP Data .................................................................................................................................. 103
Running the AEP Test .............................................................................................................. 105
To run the AEP test ..................................................................................................................... 105
Editing the audio.cfg File .......................................................................................................... 107
To edit the audio.cfg file .............................................................................................................. 107
Values in the audio.cfg file........................................................................................................... 107
VEP............................................................................................................................... 109
About the VEP Test .................................................................................................................. 109
For further details, click . . . ......................................................................................................... 109
For test procedures, click . . . ...................................................................................................... 109
For configuration procedures, click . . ......................................................................................... 110
VEP Windows........................................................................................................................... 110
See also .......................................................................................................................................... 110
VEP Toolbars ........................................................................................................................... 110
Primary toolbar (Tool1VEP.bmp).................................................................................................... 110
VEP Data .................................................................................................................................. 112
Running the VEP Test .............................................................................................................. 113
To run the test ............................................................................................................................. 113
Selecting the Number of Checks.............................................................................................. 114
To select the number of checks................................................................................................... 115
To determine check size.............................................................................................................. 115
To determine stimulus field size .................................................................................................. 115
Blink Reflex................................................................................................................... 115
5
About the Blink Reflex Test ...................................................................................................... 115
For further details, click . . . ......................................................................................................... 116
For test procedures, click . . . ...................................................................................................... 116
For configuration procedures, click . . ......................................................................................... 116
Blink Reflex Windows ............................................................................................................... 116
Blink Reflex Toolbars................................................................................................................ 116
Blink Reflex Data ...................................................................................................................... 118
Running the Blink Reflex Test .................................................................................................. 118
To run the blink reflex text ........................................................................................................... 119
RNS .............................................................................................................................. 119
About the RNS Test.................................................................................................................. 120
For further details, click . . . ......................................................................................................... 120
For test procedures, click . . . ...................................................................................................... 120
For configuration procedures, click . . ......................................................................................... 120
RNS Windows .......................................................................................................................... 120
RNS Toolbars ........................................................................................................................... 121
RNS Data.................................................................................................................................. 122
Running the RNS Test.............................................................................................................. 123
To run the RNS test..................................................................................................................... 123
QUICKMED UTILITIES.....................................................................................125
About the QuickMed Utilities......................................................................................... 125
For descriptions of the tabs, click . . . .............................................................................................. 125
For Patient procedures, click . . . ..................................................................................................... 125
For Records procedures, click . . . ................................................................................................... 126
For User Databases procedures, click . . ........................................................................................ 126
For Report Templates procedures, click . . ..................................................................................... 126
For Startup procedures, click . . ...................................................................................................... 126
Patient Tab ................................................................................................................... 126
For procedures, click . . . ................................................................................................................. 126
Records Tab ................................................................................................................. 126
For procedures, click . . . ................................................................................................................. 127
User Databases Tab..................................................................................................... 127
For procedures, click . . . ................................................................................................................. 127
Report Templates Tab .................................................................................................. 127
Startup Tab ................................................................................................................... 127
TROUBLESHOOTING .....................................................................................128
Problem Symptoms ...................................................................................................... 128
PATIENT SAFETY ...........................................................................................128
Operating Characteristics ............................................................................................. 128
Current ............................................................................................................................................... 129
Hazardous Operating Conditions.................................................................................. 129
Precautions................................................................................................................... 129
MAINTENANCE AND WARRANTIES .............................................................130
Maintenance Procedures.............................................................................................. 130
Backing Up ............................................................................................................................... 130
For the procedure, click . . . ......................................................................................................... 130
Grounding the Cart ................................................................................................................... 130
To ground the cart ....................................................................................................................... 130
Replacing the System Fuse...................................................................................................... 130
To replace the fuse...................................................................................................................... 131
Removing the Laptop From the Sierra II Wedge...................................................................... 131
To remove the laptop................................................................................................................... 131
6
Cleaning and Sterilizing ................................................................................................ 131
Recommendations.................................................................................................................... 131
Sierra Components................................................................................................................... 131
Stimulators................................................................................................................................ 132
Interface Cables and Power Cords .......................................................................................... 132
Recording Electrodes ............................................................................................................... 132
Warranty Information .................................................................................................... 133
Warranty and Service Contracts .............................................................................................. 133
Limited Warranty ...................................................................................................................... 133
For additional information, click . . . ............................................................................................. 133
7
Clear ............................................................................................................................. 144
Store ............................................................................................................................. 144
Run/Stop....................................................................................................................... 144
Single............................................................................................................................ 144
Volume.......................................................................................................................... 144
LPT 1 ............................................................................................................................ 144
COM 1 .......................................................................................................................... 144
KEYBOARD (or keyboard symbol) ............................................................................... 144
VGA VIDEO .................................................................................................................. 144
Power cord connector................................................................................................... 144
On/Off (l / O) ................................................................................................................. 144
Fuse drawer.................................................................................................................. 144
Power rating label ......................................................................................................... 145
Cart grounding pin/screw.............................................................................................. 145
Caution label................................................................................................................. 145
Network......................................................................................................................... 145
Caution label................................................................................................................. 145
System identification label ............................................................................................ 145
COM 2 .......................................................................................................................... 145
VIDEO (or video symbol) .............................................................................................. 145
Attached LT power cable .............................................................................................. 145
BAEP ............................................................................................................................ 145
Electrical Stimulators .................................................................................................... 145
Floppy disk drive........................................................................................................... 145
Goggles ........................................................................................................................ 145
VEP/AUX ...................................................................................................................... 145
Amplifier........................................................................................................................ 145
FOOTSWITCH.............................................................................................................. 145
AUXILIARY ................................................................................................................... 145
10-BASE T.................................................................................................................... 145
BAEP / VEP MONITOR ................................................................................................ 146
ESTIM........................................................................................................................... 146
TEMP............................................................................................................................ 146
LAPTOP POWER 1A MAX........................................................................................... 146
100-240VAC - 50/60Hz - 2 AMP................................................................................... 146
KEYBOARD.................................................................................................................. 146
COM 1 .......................................................................................................................... 146
COM 2 .......................................................................................................................... 146
PARALLEL.................................................................................................................... 146
10-BASE T.................................................................................................................... 146
Power cord connector................................................................................................... 146
On/off switch ................................................................................................................. 146
Circuit board slots ......................................................................................................... 146
Video card..................................................................................................................... 146
On ................................................................................................................................. 146
EEG .............................................................................................................................. 146
PORT 1......................................................................................................................... 146
PORT 2......................................................................................................................... 147
AUX 1 ........................................................................................................................... 147
AUX 2 ........................................................................................................................... 147
8
THE SYSTEM SETUP MENU...........................................................................147
System Setup/General.................................................................................................. 147
For information about theother types of setup, click . . . ................................................................. 147
System Setup/Test Menu ............................................................................................. 147
For information about the type of setup you want to do, click . . ..................................................... 147
System Setup/Test Setup ............................................................................................. 147
EMG, EMG2, SFEMG, and SSFEMG Setup ................................................................ 148
For information about the type of setup you want to do, click . . ..................................................... 148
Protocol Setup .............................................................................................................. 148
For information about the type of setup you want to do, click . . ..................................................... 148
Configuring the Default Value for Sweep Delay ........................................................... 148
To configure the default value for sweep delay .............................................................................. 148
GAIN................................................................................................................. 149
MARKERS........................................................................................................ 151
RASTER/OVERLAY VIEW...............................................................................152
9
REJECT ON/OFF .............................................................................................152
RUN/STOP .......................................................................................................152
RECORDING SIDE...........................................................................................152
SMOOTH ..........................................................................................................152
STORE TRACES..............................................................................................153
SWEEP SPEED................................................................................................153
IEC....................................................................................................................154
UL .....................................................................................................................154
10
CHANGING THE ORDER OF TESTS IN THE MENU......................................155
To change the order of tests................................................................................................................155
GN ....................................................................................................................165
Description .......................................................................................................................................... 165
How set or adjusted............................................................................................................................. 166
HI, LO ...............................................................................................................166
Description .......................................................................................................................................... 166
How set or adjusted............................................................................................................................. 166
11
SWP..................................................................................................................167
Description ..........................................................................................................................................167
How set or adjusted.............................................................................................................................167
LEFT/RIGHT/BOTH..........................................................................................167
Description ..........................................................................................................................................167
How set or adjusted.............................................................................................................................167
SWPDEL...........................................................................................................168
Description ..........................................................................................................................................168
Where set or adjust ............................................................................................................................168
TRIG LVL..........................................................................................................168
Description ..........................................................................................................................................168
How set or adjusted.............................................................................................................................168
REJ%................................................................................................................168
Description ..........................................................................................................................................168
How set or adjusted.............................................................................................................................168
INTENL, INTENR..............................................................................................169
Description ..........................................................................................................................................169
How set or adjusted.............................................................................................................................169
THRESHL, THRESHR......................................................................................169
Description ..........................................................................................................................................169
How set or adjusted.............................................................................................................................169
MASK ...............................................................................................................169
Description ..........................................................................................................................................169
How set or adjusted.............................................................................................................................169
COND/RARE/ALT ............................................................................................169
Description ..........................................................................................................................................169
How set or adjusted.............................................................................................................................169
PHONES/INSERTS ..........................................................................................170
Description ..........................................................................................................................................170
How set or adjusted.............................................................................................................................170
PW ....................................................................................................................170
Description ..........................................................................................................................................170
How set or adjusted.............................................................................................................................170
MA ....................................................................................................................170
Description ..........................................................................................................................................170
How set or adjusted.............................................................................................................................170
12
RR.....................................................................................................................170
Description ..........................................................................................................................................170
How set or adjusted.............................................................................................................................171
AVG ..................................................................................................................171
Description ..........................................................................................................................................171
How set or adjusted.............................................................................................................................171
AVGCT .............................................................................................................171
Description ..........................................................................................................................................171
How set or adjusted.............................................................................................................................171
RND% ...............................................................................................................171
Description ..........................................................................................................................................171
How set or adjusted.............................................................................................................................171
GOGGLES/CHECKS/GANZFELD/LED/FLASH ..............................................172
Description ..........................................................................................................................................172
How set or adjusted.............................................................................................................................172
FULL/UPPERH/RIGHTH/LOWERH/LEFTH/UPPERL/UPPERR/LOWERR/LOWERL 172
Description ..........................................................................................................................................172
How set or adjusted.............................................................................................................................172
CHECK SZ........................................................................................................172
Description ..........................................................................................................................................172
How set or adjusted.............................................................................................................................172
SPREAD ...........................................................................................................172
Description ..........................................................................................................................................172
How set or adjusted.............................................................................................................................172
TRAIN ...............................................................................................................173
Description ..........................................................................................................................................173
How set or adjusted.............................................................................................................................173
MAX..................................................................................................................173
Description ..........................................................................................................................................173
How set or adjusted.............................................................................................................................173
STRTTRC .........................................................................................................173
Description ..........................................................................................................................................173
How set or adjusted.............................................................................................................................173
TRCCT..............................................................................................................173
Description ..........................................................................................................................................173
How set or adjusted.............................................................................................................................173
13
ADDING A PATIENT FILE ...............................................................................174
To add a patient file .............................................................................................................................174
14
MISCELLANEOUS FUNCTION TOOLBAR.....................................................183
TOOLBARS......................................................................................................190
MENU BAR.......................................................................................................191
STATUS BAR...................................................................................................192
FUNCTION TOOLBAR.....................................................................................192
15
NO VISUAL STIMULUS IS DELIVERED .........................................................194
SOLE PURPOSE..............................................................................................197
INDEMNIFICATION..........................................................................................197
SEVERABILITY................................................................................................197
NO WAIVER .....................................................................................................198
16
MEDIAN F-WAVE.............................................................................................201
PERONEAL MOTOR........................................................................................206
CHANGING REFERENCE SITE, SITES, OR SEGMENTS FOR THE CURRENT TEST 209
To change the reference site...............................................................................................................209
To add entries to the site table ............................................................................................................209
To change the segment table ..............................................................................................................210
ELECTRODE PLACEMENT OF THE TIBIAL NERVE FOR THE H-REFLEX TEST 211
ELECTRODE PLACEMENT OF THE MEDIAN NERVE FOR THE F-WAVE TEST 212
ELECTRODE AND STIMULATOR PLACEMENT FOR MEDIAN NERVE RNS TEST 213
CHECK AND CHECK WIDTH FOR THE VEP MONOCHROME MONITOR ...220
17
CHECK AND DISPLAY WIDTH FOR THE 3-INCH LED STIMULATOR .........220
REVIEW............................................................................................................222
CLOSE..............................................................................................................222
EXIT..................................................................................................................223
COPY................................................................................................................223
USER/PHYS/TECH LIST..................................................................................223
PATIENT LIST..................................................................................................223
TEST SETUP....................................................................................................223
QUICKMED UTILITIES.....................................................................................223
TABDATA.........................................................................................................223
AMPLIFIER BAR..............................................................................................223
ACQUISITION BAR..........................................................................................224
AVERAGER BAR.............................................................................................224
18
MARKERS BAR ...............................................................................................224
POSITION.........................................................................................................226
CURSORS ........................................................................................................226
STATUS BAR...................................................................................................226
SAVE NEW.......................................................................................................226
EMG..................................................................................................................226
NCV/F/H............................................................................................................227
SEPU ................................................................................................................227
SEPL ................................................................................................................227
VEP...................................................................................................................227
BLINK ...............................................................................................................227
RNS ..................................................................................................................227
MEDIAN SEP....................................................................................................227
PERONEAL SEP..............................................................................................227
UPPER DERMATOMES...................................................................................227
19
LOWER DERMATOMES..................................................................................227
AEP ..................................................................................................................227
EMG2................................................................................................................227
SFEMG .............................................................................................................228
SSFEMG...........................................................................................................228
WHAT'S THIS...................................................................................................228
REPORTGEN HELP.........................................................................................228
ABOUT SIERRAWIN........................................................................................228
20
ADDING A COLUMN TO THE SCORED MUSCLES TABLE..........................231
To add a column on the right side of the table.....................................................................................231
To insert a column within the table ......................................................................................................232
CHANGING AN ENTRY IN THE NERVE LIST, SITE TABLE, OR SEGMENT TABLE 235
To change an entry..............................................................................................................................235
DELETING AN ENTRY FROM THE NERVE LIST, SITE TABLE, OR SEGMENT TABLE 236
To delete an entry................................................................................................................................236
21
REVERSING POLARITY..................................................................................238
To reverse polarity...............................................................................................................................238
22
OPEN PATIENT FILE DIALOG BOX ...............................................................243
Selection Criteria ......................................................................................................................243
Type of Record .........................................................................................................................243
PatientID ...................................................................................................................................243
Last Name ................................................................................................................................243
First Name ................................................................................................................................243
Start Date and End Date ..........................................................................................................243
Summary ..................................................................................................................................244
Search ......................................................................................................................................244
List of files.................................................................................................................................244
Print List....................................................................................................................................244
Open .........................................................................................................................................244
Cancel.......................................................................................................................................244
PATIENT LIST DIALOG BOX ..........................................................................245
Summary ..................................................................................................................................245
Enter Selection .........................................................................................................................245
Patient List ................................................................................................................................245
Add New ...................................................................................................................................245
Edit............................................................................................................................................245
Delete .......................................................................................................................................245
Done .........................................................................................................................................245
ADD PATIENT DIALOG BOX ..........................................................................245
23
Start Test ..................................................................................................................................248
Save Only .................................................................................................................................249
Cancel.......................................................................................................................................249
CUSTOM TEST INFO TAB ..............................................................................249
Patient History ..........................................................................................................................249
Medications ..............................................................................................................................249
Findings ....................................................................................................................................249
Conclusions ..............................................................................................................................249
Start Test ..................................................................................................................................249
Save Only .................................................................................................................................249
Cancel.......................................................................................................................................249
EDIT PATIENT DIALOG BOX..........................................................................249
24
Cancel.......................................................................................................................................253
SELECT REPORT DIALOG BOX ....................................................................253
Selection Criteria ......................................................................................................................254
Type of Record .........................................................................................................................254
PatientID ...................................................................................................................................254
Last Name ................................................................................................................................254
First Name ................................................................................................................................254
Start Date and End Date ..........................................................................................................254
Summary ..................................................................................................................................254
Search ......................................................................................................................................254
List of files.................................................................................................................................255
Print List....................................................................................................................................255
Open .........................................................................................................................................255
Cancel.......................................................................................................................................255
SPL...................................................................................................................255
NHL ..................................................................................................................255
25
To delete a custom field ...................................................................................................................... 263
26
Introduction to the Sierra EMG/EP
Overview
General Description
The Cadwell Sierra™ offers a complete range of 2-, or 4-channel electromyography/evoked potential (EMG/EP) testing
and reporting. The system, designed to help diagnose neuro-muscular disorders, is available in two models:
• Sierra II Console
Offers complete, full-color EMG/EP testing and reporting in 2 or 4 channels. An optional laser printer can be placed on
the bottom shelf of the instrument cart.
• Sierra II Wedge
The lightweight version that features a laptop computer with full-color display of EMG/EP test data in 1, 2, or 4
channels.
Both models include a full-color monitor, custom keyboard, and alphanumeric keyboard.
Sierra II Console Sierra II Wedge
27
The Sierra II Console consists of the following components:
• Sierra II custom keyboard
• Third-party personal computer (PC) with Microsoft Windows software and instructions
• Color monitor with cable and power cord
• Serial mouse
• Alphanumeric keyboard with cable
• EMG/EP amplifier with cable
• Instrument cart with monitor bracket
• Cart arm with mounting bracket for amplifier
• Constant-current electrical stimulator
• One auditory stimulator (depending on system configuration)
• One visual stimulator (depending on system configuration)
• Cart grounding cable
• 10BaseT network cable (attaches the custom base to the network card in the PC)
• Printer with cable and power cord (optional)
• Isolation transformer with power cord (optional)—required with the color monitor, laser printer, and monochrome VEP
monitor to meet UL 2601-1, CSA, and IEC 601-1 leakage current requirements
• Footswitch (optional)
28
Connections on laptop computers vary according to model. See the documentation for your particular laptop.
Click a part for a description of the part.
Connections on laptop computers vary according to model. See the documentation for your particular laptop.
Click a part for a description of the part.
Keyboards
Custom Keyboard
Alphanumeric Keyboard
29
The alphanumeric keyboard is the standard one provided with personal computers and laptops. For information about
using the keyboard, see the documentation for the keyboard that came with the Sierra II system.
Accessories
Printers
The printers listed below are available for the Sierra II Console and Sierra II Wedge. Refer to the manufacturer's
documentation for detailed information about the printer used with your system.
International customers requiring 230-volt nominal printers should contact their local Cadwell distributor.
Printer Description
Hewlett-Packard LaserJet Produces 600-dpi printouts. Requires the
isolation transformer to meet UL 2601-1,
CSA, and IEC 601-1 leakage current
requirements.
Hewlett-Packard DeskJet Produces 300-dpi printouts. Also available
in a portable model. Both require the
isolation transformer to meet UL 2601-1,
CSA, and IEC 601-1 current leakage
requirements.
Isolation Transformer
The isolation transformer provides an isolated, AC power source for peripheral system accessories (printers and monitors,
for example).
It is optional for both the Sierra II Console and Sierra II Wedge systems. The six-outlet transformer stores easily beneath
the cart shelf (or on a flat surface) and comes with a hospital-grade power cord and seven adapter cables. The isolation
transformer is available in 110- and 220-volt configurations.
To meet UL 2601-1, CSA, and IEC 601-1 leakage current regulations, the isolation transformer is required when using
any of the following accessories with the Sierra II Console or Sierra II Wedge system:
• Hewlett-Packard LaserJet or DeskJet printer
• 15-inch color monitor
• Monochrome VEP monitor
Amplifiers
30
input, or when positive voltage is applied to the reference input, an upward deflection is produced on the display. Use the
active input for the recording electrode and the reference input for the reference or inactive electrode for a negative up
convention. For a positive up convention, reverse the electrode connections.
The Sierra amplifier is isolated for patient safety. Active, reference, and isolated ground electrode inputs are touchproof to
meet FDA recommendations and UL 544 and DIN 42-802 safety regulations. Two 5-pin, DIN round connectors provide
additional electrode inputs.
The amplifier has three functions:
• To provide gain - since the voltages in the tissues are in the millivolt range, amplification is required to boost the
levels.
• To provide an impedance match between the tissue and the instrument. The tissue is not a powerful source of voltage
and care must be taken to not load down the source which could distort the potential being measure.
• To provide for the rejection of common mode noise. To accomplish this, differential amplification is used.
The constant current electrical stimulator is designed for right- or left-hand use and features a built-in, digital stimulus
intensity control dial and single-stimulus trigger. The metal stimulating probes can be removed for cleaning and
sterilization or for inserting other stimulus electrodes.
____________________
See also
Auditory Stimulators
Acoustically shielded
headphones
The headphones feature 10-Ω
speakers, adjustable headband,
and color-coded earphones for easy
right/left identification.
(2-24a.bmp)
Lightweight headphones
Recommended for pediatric use,
these headphones feature 10-Ω
speakers and adjustable headband.
(2-24b.bmp)
Insert earphones
The insert earphones feature 10-Ω
compatible speakers with 1-ms
acoustic delay. They are useful for
intraoperative applications, ECochG
testing, and BAEP testing in infants.
31
(2-24c.bmp)
Auditory bone transducer
The auditory bone transducer is
useful for assessing peripheral
hearing loss and/or head trauma.
(2-24d.bmp)
ECochG headset
Designed for needle-type ECochG
testing, this headset features
removable 10-Ω speakers.
(2-24e.bmp)
____________________
See also
Visual Stimulators
LED goggles
The goggles present LED flash stimuli
to the left, right, or both eyes and are
useful for intraoperative applications.
(2-25c.bmp)
32
Single-intensity flash stimulator
Repetition rate and start and stop
times are controllable from the custom
keyboard. Comes with a 12-volt power
supply.
(2-25d.bmp)
Ganzfeld full-field stimulator
The Ganzfeld stimulator presents flash
stimuli inside an 18-in (46-cm) globe.
Select from adjustable flash intensity,
continuous illumination intensity, and
triggering modes.
(2-25e.bmp)
____________________
See also
Configuration
About Configuration
33
Configuring report key assignments
Configuring protocols
You can add, edit, and delete files for system users (physicians and technicians) and referring physicians. Entering these
names in the system is a convenience for you: once configured, the names appear automatically in drop-down lists when
you are asked for these types of people. You can select from the list instead of typing the information.
You can also configure whether or not user names are required for employees to log in to the Sierra system and the
privileges associated with those user names. You can disable or reinstate the requirement for user name login at any
time. The Sierra system comes with two preconfigured user names (after you enter your own system names, these can be
deleted):
• Sierrauser, A
Has full access privileges (can conduct protocol tests, create new protocol tests, and configure the Sierra system).
• Sierrauser, B
Has no access privileges (can only conduct protocol tests).
The login feature is provided so that users, under their separate user names, can set up their own test protocols,
parameters settings, and report templates.
Use the custom fields procedure to create, change, and delete fields in the Patient Information dialog box. The information
you collect becomes part of each patient file in the QuickMed database.
The standard fields that come with the system are listed below. You can use these labels, or rename them to suit your
needs.
Patient History
Medications
Findings
Conclusions
34
Many of the standard patient reports include information from one or more of the standard fields. You can customize any
report template to include custom fields as well. For information on customizing reports, go to the ReportGen Help file (on
the Help menu, click ReportGen Help).
You can configure the test that opens when you press the TEST 1, TEST 2, and TEST 3 keys on the custom keyboard.
The assignments apply across all protocols.
You can configure which report prints when you press the Report 1 and Report 2 keys on the custom keyboard or select
the Print Report 1 and Print Report 2 commands from the File menu. The assignments are unique for each protocol.
Click here to open a file containing examples of the standard reports.
35
2 For each report key, select one of the following:
Assign To assign a particular report to the key.
Use Menu To cause a menu of all reports to display when you press the
key.
3 If you selected Assign, a list of all reports appears. Select the report from the list and click OK.
4 Click OK to return.
____________________
See also
Printing and saving reports
Configuring Protocols
36
For configuring features, click . . .
Changing the order of tests in the menu
The Sierra system comes with default settings for the parameters in each test protocol. For example, in the RNS test
protocol, the default setting for the gain parameter is 2,000 µV/division and the default setting for the sweep speed
paramter is 5 ms/division.
You can change the settings in several ways:
• For an individual test
You can change the settings while you are conducting a test. Changes affect only that test. Each time you reopen that
particular, theseettings will be in effect.
• For a test protocol
You can change the settings and then apply them to the test protocol. These become the new defaults for that test
protocol, and will be in effect each time you start a new test in the test protocol.
• For a new protocol
You can change the settings for a test protocol and then save the settings as a new test protocol. This test protocol is
added to the list of on the Test Protocols menu . Use this method when you do not want to change the defaults in the
original test protocol.
There are two methods for changing settings. Click below for explanations.
37
To configure settings for an individual test
1 While in the test you want to configure, set the parameters as you want them.
2 When the test is finished, close the test. All current parameter settings are saved. When you open that test, the
changes remain in effect.
This method only covers the parameters controlled in the test setup dialog boxes for protocols.
1 On the Edit menu, click System Setup and then click the Test Setup tab.
2 Select the test protocol you want and click Test Setup.
3 On the tabs, enter the default settings or values that you want.
4 Click OK.
With a test protocol open:
Descriptions of parameters
You can configure the trace labels for the EP and NCV protocols by choosing to display any three of the available
parameters.
Customized trace labeling is not available for the EMG test protocol. In the NCV test protocol, every trace label
automatically includes the site name (for example, Wrist or Elbow).
The following illustration shows a trace label with channel, trial, and side of stimulation configured.
38
1 You can open the dialog box for configuring trace labels in two ways:
• On the Edit menu, click System Setup and then click the Test Setup tab. Select the test protocol you want and
click Test Setup.
• If you are already in a test protocol, on the Edit menu click Test Setup.
2 The setup dialog box appears with the following area for configuring the trace labels:
3 Select from the drop-down list for each trace label field that you want to use. You can select blank as well (use the
blank choice at the bottom of the drop-down list).
Trace label choices
4 If you choose CustomCh for a trace label field, the Ch Left and Ch Right fields become available. These fields let you
configure labels for each channel on a left side/right side basis. Type the labels you want.
5 Use the Display Mode field to control which trials you want trace labels to appear on.
6 Click OK to return to the test.
Configuring Auto Cursors
The EP protocols use auto cursors to mark important waveform features. The measurements that the cursors represent
are used to calculate latency and amplitude values, including left-right side latency differences. The tests display the
calculations in a table on the test screen.
Cursors are also used in the NCV protocol. However, the NCV cursors mark four predetermined waveform
features (onset, peak, trough, and recovery) and cannot be configured.
Each EP protocol comes configured with default auto cursors. However, for each protocol, you can configure the
following:
• Names of the cursors and the amplifier channels they are associated with.
• Specific feature for which each cursor searches (peak or trough).
• Start and end latency search times.
• Which cursors you want to use to calculate latency, interpeak latency, amplitude, and interamplitude measurements;
you can also enter normal values for these measurements and their left-right side differences.
How you configure the cursor table determines the columns in the table of measurements on the right side of the EP test
screen.
When you are setting up auto cursors for the evoked potential tests, you can enter the expected normal values. For
example, the N20 latency cursor for the SEPU response should normally occur somewhere between 18 to 25
milliseconds. So, for the N20 norm, you could enter <25 as the norm value. Norm values are displayed when you
39
review test studies using the TabData function and in printed reports, which lets you compare patient values with the
norm values. You can also enter norms for interpeak latency measurements and amplitude measurements.
See also
Working with cursors
Printing reports
In the EP protocols, you can control how the data window displays (full display, vertical split, horizontal split, or quadruple
split). Traces can then be sorted by side or channel.
3 Click OK to return.
Configuring the Footswitch
A one-position footswitch is an optional accessory for the Sierra system. For the NCV, F-wave, and H-reflex protocols,
you can configure the footswitch to do either of the following:
40
• Store (default)
Pressing the footswitch is equivalent to using the Store control to store traces.
• Stim
Pressing the footswitch delivers electrical stimulation.
If you are using the averaging function, setting the footswitch to Stim causes it to work the same as the Run/Stop
control.
3 Click OK to return.
Restoring Defaults
You can restore the Sierra system to the way it was when you received it--for all protocols or only for the protocols you
select. This restores defaults only for the user you logged in with. It does not affect any other user.
Restoring defaults removes the following kinds of changes you might have made:
• Parameter setups
• Test and report key assignments
• New tests
• Additions to the NCV nerve list and EMG muscle list
To restore defaults
1 On the Edit menu in Sierra, click System Setup.
2 Click Restore Defaults.
3 When the Restore Settings dialog box appears, do one of the following:
• To restore defaults for particular protocols, click the box next to the
protocols you want and then click Restore Checked. If you select a
41
protocol that does not currently exist on your system, it will be added
to the list of tests in the Test Protocols menu.
• To restore the defaults for all protocols, click Restore All. Note that
this restores for all protocols, even if you have selected one or more
of the boxes.
4 When asked to confirm, click Yes.
Configuring Reports
The Sierra system comes with a set of reports designed to fit the basic requirements of all protocols. You can configure
the format and contents of the reports, or create additional reports that fit your needs.
For information on how to configure reports, go to the ReportGen Help file (on the Help menu, click ReportGen Help).
Preserving and Restoring Settings
You can preserve the settings you have established for individual or all protocols. You can then restore the settings at any
time. These procedures apply only to settings for the user you logged in with. They do not affect any other user.
To preserve settings
1 On the Edit menu in Sierra, click System Setup.
2 Click Preserve Settings.
3 In the Save As dialog box, name the settings file, select where you want to save it, and then click Save.
4 In the System Setup dialog box, click Close.
To restore settings
1 On the Edit menu in Sierra, click System Setup.
2 Click Restore Settings.
3 In the Open dialog box, locate the settings file and click Open.
4 In the Restore Settings dialog box, do one of the following:
• To restore settings for particular protocols, click the box next to the
protocols you want and then click Restore Checked. If you select a
42
protocol that does not currently exist on your system, it will be added
to the list of tests in the Test Protocols menu.
• To restore the settings for all protocols, click Restore All. Note that
this restores for all protocols, even if you have selected one or more
of the boxes.
5 When asked to confirm, click Yes.
Basic Operation
Turning Equipment On and Off
If you are using the isolation transformer and cart power switches, you only need to turn on the cart power switch.
Do not turn the Sierra instrument base on or off while a patient is connected.
1 Close all programs that are running on the computer.
2 On the Windows taskbar, click Start, Shut Down, Shut down the computer?, and then Yes.
3 Turn off equipment in the following order:
Sierra instrument base
Computer
Monitor (for Sierra II Console only)
Printer
If you are using the isolation transformer and cart power switches, turn off only the cart power switch. The monitor,
printer, instrument base, and isolation transformer power switches can remain on. The next time you start the
system, turn on only the cart power switch.
You do not need to disconnect the amplifier, stimulators, or accessories. You can leave them connected.
Opening and Closing Sierra
To open Sierra
1 Turn on the Sierra instrument base, computer, monitor (Sierra II Console only), and printer.
Turning equipment on and off
2 Start the Sierra software from your computer in one of the following ways:
43
• Double-click the Sierra icon on the Windows desktop.
• On the Windows taskbar, click Start, Programs, Cadwell, and then
Sierra.
3 If login is required, the Login dialog box opens. Select your user name and click OK.
To close Sierra
When you close a patient file or exit from Sierra while in a patient file, all data is automatically saved to the file.
• To close a patient file without exiting from Sierra: On the File menu, click Close.
• To exit from Sierra: On the File menu, click Exit.
Working with Patient Files
Each test you perform in Sierra is tied to a patient file. When you enter Sierra, you either select an existing patient file or
create a new one. You can edit or delete a patient file at any time.
A field is any place on the screen that contains information you can enter or change. The Sierra system uses the following
methods for entering and changing information in fields:
• Custom keyboard keys and knobs
These keys and knobs control functions specific to test protocols. Examples include keys or knobs for printing reports,
starting and stopping data acquisition, and adjusting gain.
Custom keyboard
• Alphanumeric keyboard keys
Contains standard computer keyboard keys for entering, changing, and deleting information and for moving around.
Examples include the arrow keys, Insert (Ins), Delete (Del), Home, End, Page Up, Page Down, and Tab.
44
• Computer mouse with on-screen pointer
The Sierra system uses the point-and-click method familiar in the Windows operating environment.
• Stimulator dial
Used with protocols that use electrical stimulation to deliver stimulation and control stimulation intensity.
To delete information
You can delete information on the screen in several ways:
• Use the clear functions on the custom keyboard or toolbars (for example, the Clear key on the custom keyboard or
the Clear All control).
• Use the Del (Delete) or Backspace key on the alphanumeric keyboard.
Working with Basic Windows
45
The Sierra computer program works within the Microsoft Windows 95/98 environment. This Help file explains what you
need to know to use Sierra in the Windows 95/98 environment. For more advanced information, use the online Help file
for Windows.
Sierra menus
Moving a window
46
____________________
See also
Working with toolbars
Toolbars on the test protocol screens contain controls for turning features on and off and for adjusting values.
• Primary toolbar (Click for illustration)
Appears by default when a test protocol opens. It contains the most commonly used controls for the protocol.
• Secondary toolbar (Click for illustration)
Appears when you select the More control on the primary toolbar. It contains controls that are less frequently used.
• Function toolbars (Click for illustration)
Contain controls that apply only to a specific function (for example, electrical stimulation). You can show or hide these
toolbars as needed.
You cannot change what appears in any of the toolbars. However, you can preserve the set of toolbars that you have
displayed for a particular protocol by saving the test under a new name. When you select that test in the future, your
toolbar settings appear as the default.
47
All Displays a list of all protocol tests.
1, 2, or 3 Opens the test protocol that you have assigned to the key.
3 If you selected ALL, click the test protocol you want from the list.
____________________
See also
Configuring test key assignments
The Sierra system comes with a set of preconfigured, standard test protocols. You can:
• Add a new test protocol to the list in the Test Protocols menu that is a variation of any of the standard test protocols.
This lets you customize how a test looks and the parameter values that are in effect when it first opens.
• Adjust the settings during a session for a test protocol that you or another user has created and then save the test
protocol with these new settings. These changes remain in effect each time you open that test for any patient. You
can update this test later with any additional adjustments or delete the test.
The tests you create are tied to the user name you logged in with. Other user names do not have access to your tests.
There is no limit to the number of tests you can create.
48
Deleting a Test Protocol
You can only delete a test that has been created with the user name that you logged in with. In other words, you cannot
delete a standard test or a test that has been created under another user name.
When you delete a test, it is removed from the Test Protocols menu. As a result, you can no longer use the Test
Protocols menu to access any files previously created with the test you deleted. However, you can still access
the files using the TabData command.
Reviewing data
You can rearrange the order of tests as they appear in the Test Protocols menu.
49
To save tests to the patient database
When finished with a test session, use either the Close or Exit command on the File menu.
Patient files are identified by an .qmc file extension. You do not need to add the file extension to the file name; this
file extension is automatically assigned to the name.
When a patient file is exported, it is automatically zipped (compressed) to save disk space. When opened again in
Sierra, it is automatically unzipped.
3 Select or enter a location if desired.
4 Click Save.
____________________
See also
Reviewing data
Printing and Saving Reports
The Sierra system comes with reports that meet the basic requirements of the standard test protocols. In the Sierra
system, you choose which report you want to create. A separate Cadwell program handles the printing, saving, and
editing of the reports; this program is called the QM (QuickMed) Report Generator.
Consult the QuickMed Report Generator online help for information about the program (on the Help menu, click
ReportGen Help).
For each test protocol, you can configure which report prints when you press the Report 1 and Report 2 keys on the
custom keyboard or select the Report 1 and Report 2 commands on the File menu. Or, you can configure a menu of
reports to display instead of a particular report.
Before you print a report, Sierra gives you the opportunity to change the format, verify information, and add, change, or
delete information.
50
To obtain information about the QM Report Generator
On the Help menu, click ReportGen Help.
You can capture information from test screens and paste it into another application such as Microsoft’s WordPad,
Notepad, or Word programs.
You can capture the following:
• Information in review windows (such as stored traces).
• Information in tables (such as the NCV nerve list and the EMG scored muscles list).
You cannot capture information from live data acquisition windows.
The Sierra system keeps a record of all the tests performed on the patient. Using the TabData function, you can select
any test and review the test data. You have the ability to sort the data in a different order and change the data.
Using the TabData function is the only way you can review test data for a test protocol that has been deleted.
To review data
1 Open Sierra.
Opening and closing Sierra
2 Use the Review control.
3 In the Open Patient File dialog box, select the Type of Record.
Explanation of the Open Patient File dialog box
4 Select the record you want and click Open.
5 Review the information as needed.
51
You can open the test, edit it, and then save the test with your changes. To do this, click Open Test or double-
click the test.
6 When finished reviewing, click Cancel.
The type of sorting you select is remembered separately for motor, sensory F-wave, and H-reflext test studies.
The selections are remembered from one patient to the next.
To adjust cursors
You cannot adjust any of the cursors while in the TabData view. You must first reopen the test (press the Select/Open
Test knob or the Open Test control) to redisplay the acquisition window. Now you can adjust the cursors.
To change column setup for NCV, F-wave, and H-reflex motor and sensory tables
When you highlight an NCV, F-wave, or H-reflext test in the TabData window, the Column Setup control becomes
available. Clicking this control displays the TabData Column Setup dialog box, which lets you select the columns for the
motor and sensory tables. These settings take effect when you view the All Motor, All Sensory, and Motor & Sensory
tables. The columns you select are also those used for the tables in printed reports.
Impedance is the resistance to alternating current flow. It can affect the signals from the recording electrodes. You should
check impedance before the start of every recording and during the recording if there is reason to suspect poor contact
between the electrode and the patient.
You can check electrode impedance at any time during a test. The impedance of the active, reference, and common
reference (4-channel only) appear on the screen.
Impedance levels below 5 kilohms (kΩ) are acceptable for most tests. The following colors are used to indicate the
impedance level for each electrode pair:
Green < 5 kΩ
Yellow 5 - 15 kΩ
Red > 15 kΩ
52
1 Make sure the amplifier is on and electrode leads are inserted into the amplifier jacks. The ground electrode must be
connected.
2 Make sure the appropriate channels are turned on.
3 Display the impedance box using any of the following methods:
• Imp key on the custom keyboard
• Impedance button on the amplifier
• Impedance control on the Acquisition function toolbar (on the View menu, click
Acquisition Bar)
4 To close the impedance box, click OK in the box, press Impedance on the custom keyboard, or press the impedance
button on the amplifier.
Delivering Electrical Stimulations
The test protocols that use electrical stimulation open with stimulation controls in the primary toolbar, as in the following
example.
In addition, you can open a function toolbar at any time that contains controls associated with electrical stimulation (on the
View menu, click Elec Stim Bar).
Reversing polarity
If you have a 4-channel system, you can select a common recording reference in the EMG and all EP test protocols. This
means that there is one common reference wire. You can turn common reference on or off for each channel or for all
channels.
53
To turn common reference on and off
1 In the test protocol, click the following control to display the secondary toolbar.
The common reference controls also appear on the Amplifier function toolbar (on the View menu, click Amplifier
Bar).
2 Click the common reference control to turn it on or off. Use the channel number controls to select channels. The
common reference setting affects all channels that are active in the primary toolbar.
Latency (time) and amplitude markers can be used in the EMG and all EP protocols.
• Latency markers
These markers show the time in milliseconds on the horizontal axis from the stimulus onset delivery to the marker
along the horizontal axis.
• Amplitude markers
These markers show the height (voltage) of the response in microvolts. Amplitude markers indicate the difference in
voltage between the two markers.
When you turn markers on, differential measurements between the markers are computed automatically and displayed in
the Markers function toolbar (shown below). The latency and amplitude scales are keyed to a particular trace, which might
be active or stored. When more than one trace is displayed, you can select which trace you want to use.
54
Latency markers. The numbers show the values for latency
marker 1 (T1), latency marker 2 (T2), and the difference
between them (T1-T2).
Amplitude markers. The number shows the difference between
the two markers.
You can move the markers on recorded waveforms. When you move the markers, the values in the toolbar change
accordingly.
Moving a marker
Working with Cursors
Cursors are used to mark important waveform features in the following test protocols:
• NCV
The test places four cursors on each waveform as it is collected: O (to mark onset), P (to mark the top of the negative
peak), T (to mark the bottom following the peak), and R (to mark the recovery point where the waveform comes back
to the baseline). These cursors cannot be configured.
• All EP protocols
Auto cursors are used in the EP protocols to label waveforms with latency and amplitude measurements. Each test
protocol has its own auto cursors table that is configured with default auto cursor, latency, and amplitude
measurement parameters. You can change the defaults to meet your individual testing needs. You can have up to
seven individual cursors for each EP test.
You can apply auto cursors during or after waveform acquisition. If placed while data is being acquired, they remain in that
position regardless of changes in the waveform’s shape as additional sweeps are collected. To reposition them, you must
move them manually.
You enter the mode for working with cursors by pressing the Cursor key on the custom keyboard. Then, labels on the
function keys and knobs change to reflect the things you can do.
To apply cursors
Select the waveform and then enter cursor mode by pressing the Cursor key on the custom keyboard. Apply the cursors
using one of the following methods:
• Auto Cursors function key
• Cursors control in the Trace Functions function toolbar (on the View menu, click Trace Functions Bar)
55
To move cursors
Once cursors have been applied, you can move cursors using one of the following methods:
• Mouse pointer
• Move/Sel Cursor knob on the custom keyboard
To clear cursors
• Clear Cursors function key
___________________
See also
You can perform the following actions on waveforms, traces, and trials:
• Store
• Move
• Clear and restore
• Display in raster or overlay format
• Apply and reposition cursors
• Apply a smoothing filter
You can also apply and reposition cursors on waveforms, and average and grand average waveforms. These are
discussed separately.
Working with cursors
Averaging waveforms
To store
Storing waveforms records them in the patient file. If the protocol has a Stored Traces window, storing moves the
waveforms into that window.
There are several methods for storing waveforms:
• Store key on custom keyboard (manual)
• Footswitch, if configured to store data rather than to deliver stimulation (manual)
• Store control on the Acquisition function toolbar (on the View menu, click Acquisition Bar) - manual
56
• Autostore control on the Acquisition toolbar (on the View menu, click Acquisition Bar) - automatic
To move
You can manually position waveforms to any location on the graticule (the exception is in the EMG protocol, where
waveform locations are fixed). Select and move the item with the mouse pointer.
To delete stored
To delete a stored trace:
• Select the waveform with the mouse and press the Clear key on the custom keyboard.
• Press the Posn key on the custom keyboard and then use the Select/Delete knob.
• Raster format
The displayed trace positions are set to spread the displayed traces uniformly from the top to the bottom of the
display. The position of the top trace is set so that its maximum negative deflection remains on the screen. The
bottom trace position is set so the maximum positive deflection remains on the screen (positive down convention).
The remaining traces are positioned evenly between the top and bottom traces without regard to the data values in
the traces. Raster format works best for selecting traces.
• Overlay format
All traces are positioned so that 0 microvolts is at the midpoint of the graticule’s vertical axis.
57
• Smooth function key
• Smooth control on the Trace Functions toolbar (on the View menu, click Trace Functions Bar)
Averaging Waveforms
Averaging is used when the amplitude of the response is small. Signal averaging helps to define the response and pull it
out from the background noise. The epoch of data following each stimulus is averaged point for point to produce an
averaged waveform. Deflections that are time-locked to the stimulus are reinforced, while signals that vary randomly with
respect to the stimulus are reduced.
The final averaged response is based on the number of stimulus presentations that you set. When the test has collected
the given number of sweeps, it automatically stops the averaging and displays the final waveform.
Every sweep of the averager starts with the presentation of a stimulus. However, when artifact reject is being employed,
every sweep might not be included in the actual average. The artifact reject function detects the condition when the input
is larger than an allowable range. Under these circumstances, the input is peak clipped and therefore distorted. When the
artifact reject function detects the input over-scale condition, it does not allow the sweep to be included in the average.
This means that when the average is completed, more stimulations might have been delivered than total sweeps in the
average.
Grand averaging, used only in the evoked potential protocols, is the technique of averaging all waveforms (stored or live)
collected in a single channel and side. If you have acquired several trials but they are a bit noisy, a grand average
provides a smoother trace.
Controls for averaging appear in the default toolbar for the test protocols that use averaging. For example:
Parameters that apply to averaging also appear in the Averager function toolbar (on the View menu, click Averager Bar).
58
• Averager control on the Averager function toolbar (on the View menu, click Averager Bar)
• Footswitch (it operates the same as the Run control during averaging)
• Run/Stop key on the custom keyboard
An Avg counter displays on the toolbar while the averager is running. The averager stops automatically when the AvgCt
count has been reached. You can continue averaging beyond the average number by manually restarting the averager.
• Grand Average control on the Trace Functions function toolbar (on the View menu, click Trace Functions Bar).
To store
You can store traces from the averaging both when the averager is running and when it is stopped. If the averager is
already running when data is stored, it will be cleared and continue to run after the store operation is complete. Averaging
might be temporarily suspended to insure the integrity of the stored data.
Store the averaged waveforms in the following ways:
• Store key on the custom keyboard
• Store control on the Acquisition toolbar (on the View menu, click Acquisition Bar)
59
• Clear key on the custom keyboard (sets the average count back to zero)
Menus
File Menu
Edit Menu
View Menu
60
TabData…
Amplifier Bar
Acquisition Bar
Averager Bar
Markers Bar
Standard Bar
Elec Stim Bar
Audio Stim Bar
Visual Stim Bar
Trace Functions Bar
Miscellaneous Bar
Position
Cursors
Status Bar
61
AEP
EMG2
SFEMG
SSFEMG
Window Menu
Help Menu
Protocols
Reviewing Test Studies (TabData)
EMG
EMG (electromyography) tests detect neuromuscular disorders by measuring the spontaneous and voluntary electrical
activity of muscles. Free-running EMG signals are collected in live mode and displayed as they are collected. You can
stop and review up to 16 previous screens of EMG waveforms at any time, or capture up to 10 sweeps for analysis. When
capturing waveforms, you set the capture delay and trigger level. You decide what waveforms you want to store with the
patient record. The test automatically tabulates muscle scoring results such as spontaneous activity, amplitude, and
duration of waveforms.
62
For further details, click . . .
EMG windows
EMG toolbars
EMG data
Scoring muscles
EMG Windows
The test screen for the EMG protocol features four windows.
• EMG Data
Displays the live or captured waveforms for the muscle. The sweep is continuously updated (not initiated by a
stimulus). Up to 16 screens of live or the last 10 captured waveforms can be collected for review. When capture is on,
the trigger level displays as a dotted line across the window. A small, vertical dash on the trigger line indicates the
trigger delay position.
• Stored EMG Traces
Displays the waveform traces that you have chosen to store from the EMG Data window. You can work with the
traces in all the standard ways. For example, you can move and clear them, and view them in either raster or overlay
format. Stored traces are tagged with the name of the muscle currently highlighted in the muscle list.
• Muscle list
Lists the muscle groups and individual muscles that are available for scoring. You can configure this list by adding,
changing, and deleting muscles and muscle groups. Muscles lists are unique to each user name and to each
individual EMG test created under that user name.
• Scored muscles table
Lists the muscles you have scored from the muscle list for the current test. You can customize the table by changing
the headings and entries and by rearranging, creating, and deleting columns.
Use the View key on the custom keyboard to cycle through the windows.
View 1 The opening view. All four windows appear.
View 2 EMG Data and muscle list only.
View 3 Stored EMG Traces and muscle list only (same view reached
using the Stored View control).
View 4 Muscle list and scored muscles table only (same view reached
63
using the Score control).
____________________
See also
EMG data
EMG toolbars
EMG Toolbars
Primary toolbar
Secondary toolbar
64
Turns the indicated channel (1, 2, 3, or 4) on and
off and sets the gain, hicut filter, and locut filter
values for that channel.
Turns common reference on and off for a channel.
Selecting a common recording reference
A quick way of selecting all channels and setting
the same gain, hicut filter, and lowcut filter values
for them all. The values here override the values
above.
Sets the gain.
EMG Data
• Channels
One channel is typically used, although more than one be can be turned on.
• Filters
Three filters can be used: hicut, locut, and notch.
• Live data
Live data appears in the EMG Data window in either run or stop mode. When in run mode, the EMG Data window
shows a continuous, non-triggered sweep for the muscle being tested. Up to 16 screens of data are continuously
stored in a memory buffer. Using the Run/Stop control, you can switch to stop mode at any time (which stops live
data acquisition) and review these 16 screens of data. Use the horizontal scroll bar or the Review knob to scroll
forward or backward through the screens.
• Triggered/captured data
Triggered/captured data also appears in the EMG Data window in either run or stop mode. Up to 10 captured traces
65
based on the trigger level can be stored in a memory buffer. You set the capture trigger level (in the range 5% to 90%
of the maximum negative deflection) and trigger delay (from one to nine divisions). The trigger level is set interactively
with an amplitude marker that extends horizontally across the window. You can switch to stop mode at any time
(which stops the triggered display) and review the last 10 captured traces.
The trigger appears as a dashed purple line in the EMG Data window. The delay shows as a small horizontal marker
on the trigger line.
• Stored data
You can store the following:
• The live trace that is currently displayed when you stop data acquisition in the EMG Data window (not all 16
traces that have been stored in the memory buffer).
• Triggered traces you have captured in the EMG Data window (the 10 most recent).
The traces you decide to store are moved to the Stored EMG Traces window. They also become a part of the muscle
record for the patient, which you can recall at any time. You can switch between raster and overlay views of the traces
using either the Raster/Overlay control in the Trace functions toolbar (on the View menu, click Trace Functions
Bar).
During the test, you can swap between two gain and sweep settings (Fibs or Recruit). The settings are useful for
viewing the different waveforms generated by low-to-moderate as opposed to maximal effort. Use FIBS/Recruit
control on the Miscellaneous function toolbar (on the View menu, click Miscellaneous Bar).
Fibs The fibrillations test parameters default to a 100 µV/div gain and
10 ms/div sweep setting. This setting can be used to record
spontaneous activity or activity during low-to-moderate volitional
effort.
Recruit The recruitment test parameters default to a 1,000 µV/div gain
and 20 ms/div sweep setting. This setting can be used to record
activity during maximal voluntary effort (interference pattern).
Scoring muscles
66
Do not turn the Sierra on or off while a patient is connected.
2 Prepare the patient by inserting the needle electrode and connecting the leads to the amplifier.
To reduce unwanted noise while applying electrodes, turn the speaker volume down with the Volume knob on the
custom keyboard.
3 On the File menu, click one of the following:
• Select Patient
To start a new test using patient information that is already in
the system.
• New Patient
To create a file for a new patient and start a test for that
patient.
4 On the Test Protocols menu, select the EMG test.
5 In the muscle list, select the muscle you want to examine.
6 Select the side you want to test using either of the following controls on the toolbar:
All captured data is erased when you return to run mode. Use the Store control if you want to keep the data.
You can use the Review control (knob 1) to review the last 16 sweeps of data until the activity of interet is
displayed.
9 Stop data acquisition with either the Run/Stop control on the custom keyboard or the Run/Stop control on the
Acquisition function toolbar (on the View menu, click Acquisition Bar).
10 Optional. Store the traces that you want to keep with the record of the muscle. To store traces, make sure the muscle
is selected in the muscle list (click with the cursor or use the Musc Sel/Score knob) and then use the Store key on
the custom keyboard.
Storing EMG traces is optional. You can skip this step and go directly to scoring the muscle if you want.
11 Score the muscle.
Scoring muscles
12 Repeat for each muscle you want to investigate.
13 Print the results, if desired.
Printing and saving reports
14 Save the test. To do this, click the File menu and then click either Close or Exit.
67
To turn the capture trigger on and off
Turn the capture trigger on and off using the following control:
• Live/CAPTURE control on the Miscellaneous function toolbar (on the View menu, click Miscellaneous Bar)
Scoring Muscles
The muscle list for the EMG test contains muscles for the upper and lower extremities. You can customize the list to fit
your needs. Muscles lists are unique to each user name as well as to each EMG protocol test.
Selecting a muscle from the muscle list places it in the scored muscles table, which is where you enter the scoring
information. The table displays the muscle name, innervating nerve, and root level along with other scoring data such as
insertional activity, fibrillations, positive sharp waves, amplitude, duration, phases, recruitment, and interference pattern.
You can customize the scored muscles table to contain other information.
For your convenience in scoring, use the Score control to display an enlarged view of just the muscle list and scored
muscles.
To score a muscle
1 In the scored muscles table, move the cursor to the cells in which you want to enter information.
2 Either type the information in the cells or click the cells to select from a list.
68
To score multiple muscles
1 Hold down the Ctrl key on the alphanumeric keyboard while single-clicking the mouse on each muscle in the muscle
list that you want to score.
2 Click the “S in the upper left corner to score all the selected muscles.
3 In the scored muscles table, move the cursor to the cells in which you want to enter information.
4 Either type the information in the cells or click the cells to select from a list.
If you select the wrong group by mistake, just turn the knob and the muscles will be unhighlighted.
2 Click the “S in the upper left corner to score all the selected muscles, or press the Muscle Sel/Score knob again.
3 In the scored muscles table, move the cursor to the cells in which you want to enter information.
4 Either type the information in the cells or click the cells to select from a list.
See also
Cnfiguring the muscle list
69
The default muscle list for the EMG, EMG2, SFEMG, and SSFEMG test protocols contains a common set of muscles for
the upper and lower extremities. You can customize the list by adding, changing, rearranging, and deleting entries.
Muscles lists are unique to each user name as well as to each test protocol.
You configure the muscle list in setup on the Muscle List tab. Following is an illustration for EMG and EMG2.
The EMG2 test protocol is a variation of the EMG test protocol. The only difference between them is that the EMG2 test
protocol has a different set of windows for the views.
• EMG
Opens with all four windows (View 1): EMG Data, Stored EMG Traces, muscle list, and scored muscles table.
Pressing View once takes you to to View 2 (EMG Data and muscle list). Pressing View again takes you to View 3
(Stored EMG Traces and muscle list). Pressing View yet again takes you to View 4 (muscle list and scored muscles
table). The Score function is a shortcut to View 4.
• EMG2
Opens with only two windows (View 1): EMG Data and muscle list. Pressing View once takes you to View 2 (Stored
EMG Traces and muscle list). Pressing View again takes you to View 3 (muscle list and scored muscles table). The
Score function is a shortcut to View 3.
Both the EMG and EMG2 test protocols share the same muscle list, but each test protocol can have its own unique
muscles scoring definitions.
Refer to the EMG test protocol for information about using the EMG2 test protocol.
SFEMG and SSFEMG
70
The SFEMG and SSFEMG test protocols are specialized electromyography tests.
• SFEMG (single fiber EMG)
Measures the activity of single muscle fibers in a motor unit and summarizes the findings by calculating standard
response statistics.
• SSFEMG (stimulated single fiber EMG)
Operates like the SFEMG test, except that it records muscle fiber activity triggered by an electrical stimulator.
These protocols have their own version of the muscle list that appears for the EMG test protocol.
The test screen for the SFEMG and SSFEMG protocols features three windows.
• SFEMG Data
Displays the live waveforms for the muscle. The trigger level displays as a dotted line across the window. A small,
vertical dash on the trigger line indicates the trigger delay position.
• Muscle list
Displays the individual muscles that are available for examination. Normative data for MCD (mean consecutive
difference) or “jitter is provided for each muscle. You can configure this list by adding, changing, and deleting
muscles. Muscles lists are unique to each user name and to each individual SFEMG or SSFEMG test created under
that user name.
• Analysis table
Displays the results of the analysis for a muscle.
Use the View key on the custom keyboard to cycle through the windows.
View 1 The opening view. SFEMG Data and muscle list only.
View 2 All three windows.
71
____________________
See also
Sample test screens
Analysis table
72
Sets the repetition rate for electrical stimulations.
• Channels
One channel is typically used.
• Filters
Three filters can be used: hicut, locut, and notch.
• Acquisition mode
The SFEMG and SSFEMG test protocols open in the data acquisition mode. Each time a potential exceeds the trigger
level, it is stored in a memory buffer for review in analysis mode. Triggered responses continue to be stored until you
press the Run/Stop control. You set the number of samples you want to collect in a run. During data acquisition, you
can reset the sample count (shown in the toolbar) to zero by pressing Clear.
• Analysis mode
You enter analysis mode after you have acquired the number of samples you want. Analysis mode opens with the
captured traces displayed in raster format. You can view the traces in overlayed format using the RSTR/Ovly control.
In analysis mode, you can divide the number of traces displayed into smaller groups for easier review. You can also
delete unwanted traces and position the analysis marker. The analyzed results appear in an analysis table on the test
screen.
• Runs and samples
Up to 20 runs can be acquired and stored for each muscle. The maximum number of samples is 100. All traces for
each run can be saved for later review.
• Mean latency and mean consecutive difference
The SSFEMG test protocol automatically calculates and displays the mean latency (Lat) and mean consecutive
difference (MCD). The mean latency is measured from the trigger (stimulus artifact) to the single fiber potential.
73
the system.
• New Patient
To create a file for a new patient and start a test for that
patient.
3 On the Test Protocols menu, select the SFEMG or SSFEMG test.
4 Set the number of samples you want to collect in a single run using the following control on the toolbar (usually set
between 50-100).
7 Prepare the patient by positioning the needle electrode so that at least two single fiber potentials can be seen on the
test screen. Use the Gain/Sweep controls to adjust the gain and sweep. Connect the leads to the amplifier.
To reduce unwanted noise while applying electrodes, turn the speaker volume down with the Volume knob on the
custom keyboard.
8 Use the Trig Lvl knob to adjust the trigger level to the position you want in the graticule.
Each time a triggering potential crosses the trigger level, the number of samples collected increases by one. The
potential is stored in a memory buffer for review in the analysis mode. Triggered responses continue to be stored
until you press the Run/Stop control.
9 For SSFEMG only. To begin electrical stimulation, pull back on the knob on the handheld electrical stimulator or use
the STIM OFF/Stim On control. You can adjust stimulus intensity by turning the knob on the hand-held stimulator or
by using the intensity control in the toolbar. Repetition rate and pulse width are also controlled in the toolbar.
During data acquisition, you can reset the sample count (shown by Smpls in the toolbar) to zero by pressing
Clear.
11 Analyze the traces.
Analyzing traces
Analysis table
12 Store the results.
Storing results in the run table
Run table
13 Repeat for each additional run you want to make on the muscle.
14 Print the results, if desired.
Printing and saving reports
15 Save the test. To do this, click the File menu and then click either Close or Exit.
74
Analyzing Traces
Analysis mode opens with the captured traces displayed in raster format. Use the RSTR/Ovly control to switch to overlay
format.
In analyze mode you can:
• Review traces
• Delete traces
• Position and adjust the analysis marker
• Analyze traces
To review traces
The test protocol lets you divide the traces into smaller groups for easier review. To enter the number of traces you want
to display at one time, change TrcCt in the toolbar. Then enter the trace number you want to start with in StrtTrc. The
range of traces you selected is displayed.
Example: 100 traces are stored in the memory buffer. To review them in groups of 20, enter 20 for TrcCt. To display the
first 20 responses, enter 1 in StrtTrc. Traces 1 through 20 appear in the window. To view the next 20 traces, enter 21 in
StrtTrc.
To delete traces
To delete an unwanted trace, turn the Trace Sel/Delete knob to highlight the trace (it becomes yellow). Press the knob to
delete it.
To review traces
After you have positioned the analysis marker, use the Analyze control. The number of Blocks and the MCD, MSCD, and
IPI values are calculated and displayed in the analysis table. The analysis is performed only on the displayed traces.
If you deleted any traces before analyzing, the sample count is updated to reflect the remaining number of traces.
If you reposition the analysis marker, use Analyze again to update the values in the analysis table.
75
You can enter the fiber density in the Fib Dens field in the analysis table. Click this field and type the value.
____________________
See also
Analysis table
Analysis mode
NCV
Nerve conduction velocity (NCV) tests measure the speed and intensity of electrical signals that travel along nerves. The
time relationship between the stimulus and the response can be displayed, measured, and tabulated.
The NCV test protocol provides basic nerve conduction velocity testing for the motor and sensory nerves. The tests
operate in single sweep run mode or averaging mode. Up to eight traces can be stored. Nerve conduction velocity is
automatically calculated.
Two specialized NCV tests are also available. Both feature a split gain control to allow F-wave and H-reflex responses to
be displayed at a more sensitive gain than the initial motor response. They do not feature averaging or nerve conduction
calculation.
• F-wave
For collecting and analyzing the backfiring of reexcited motor neurons.
• H-reflex
For collecting and analyzing the H-reflex response.
All three NCV test protocols open to a nerve list that contains commonly tested motor, sensory, F-wave, and H-reflex
nerves. The type of nerve you select determines the test screen that appears and the default parameters that are in effect.
Nerve lists are unique to each user name, but are global to all NCV-type tests.
NCV toolbars
NCV data
76
For configuration procedures, click . . .
Configuring the NCV, F-wave, and H-reflex protocols
NCV Windows
You can change the reference site, sites, and segments for a nerve while you are conducting a test. The changes
apply only to the test you are conducting and do not affect the configured settings.
Changing reference site, sites, or segments for the current test
Use the View key on the custom keyboard to cycle through the windows.
View 1 The opening view. All five windows appear.
View 2 NCV/F/H Data and all tables.
View 3 Stored NCV/F/H Traces and all tables
____________________
See also
Working with waveforms
77
NCV data
NCV toolbars
NCV Toolbars
Primary toolbar
78
Secondary toolbar
NCV Data
• Channels
One or two channels can be used for the NCV test protocol.
• Filters
Two filters are available: hicut and locut.
• Waveforms
Up to 20 traces can be stored.
• Averaging
You can run in either live or averaging mode. In averaging mode, the waveform that appears in the Live Window
represents the average of the total number of responses collected. You can set the total number of responses for
each average.
• Cursors
The test automatically places cursors on each waveform as it is collected. The cursors mark onset (O), peak (P),
trough (T), and recovery (R). Once the cursors have been applied, you can reposition them as needed.
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation). The site name is automatically included.
• Footswitch
You can configure the footswitch to either store data or deliver stimulations.
• Temperature probe
The temperature probe is sampled at least once every 5 seconds and displayed on the test screen. The value is
displayed in degrees Celsius.
____________________
See also
Averaging waveforms
79
Configuring the footswitch
You can change the reference site, sites, and segments for a nerve while you are conducting the test. The
changes apply only to the test you are conducting and do not affect the configured settings.
Changing reference site, sites, or segments for the current test
You can swap sides at any time using either of the following controls on the toolbar:
6 If you want to average waveforms, click the Averager On/Off control in the toolbar. Additional controls appear on the
Averager function toolbar to let you adjust parameters that pertain to averaging – for example, AvgCt (on the View
menu, click Averager Bar).
Averaging waveforms
7 Adjust other parameters as needed.
8 Deliver the stimulation.
If averaging, stimulus stops automatically when the number of stimulations for AvgCt has been delivered. To stop
stimulus at any time, use the Run/Stop control or footswitch.
9 Gradually adjust the stimulus intensity until you achieve a satisfactory supramaximal response waveform.
80
Optional. Use the NR column in the site table to record no response. Click in the field to display “NR.
10 Store the response. Storing moves the active waveform into the Stored NCV/F/H Traces window.
11 The system advances you to the next site (if there is one). Repeat the steps for stimulating and storing.
12 The test places onset (O), peak (P), trough (T), and recovery (R) cursors on each waveform as it is collected.
Reposition the cursors if desired using the Onset/Peak knob. You might want to expand the view of the window to
make this easier (use the View key on the custom keyboard).
Working with cursors
Adjusting cursors manually in the NCV test
13 Press or click Distance and enter the distance (in centimeters) of the first nerve segment in the segment table. The
test calculates the conduction velocity and places it in the Vel column. Repeat for each segment.
If you want to change the distance, press Distance again until you return to the appropriate line. Always make
sure you are measuring the correct segment by reviewing the segment sites in the lower right corner of the
screen.
You can restimulate a site if needed. To do this, select the site using the Next Site function key and repeat the
stimulation. When you store the restimulation, it replaces the previous response. If you decide not to replace the
original response, press Clear.
14 Click Next Nerve to return to the nerve list.
15 Select the next nerve and repeat the test steps until you have tested all the nerves that you want for the patient.
If you stimulate and store traces out of order, you can sort them by increasing latency using the Sort control on the
toolbar.
The nerve list is used in the NCV, F-wave, and H-reflex protocols and contains commonly tested nerves for four types of
nerve groups: motor, sensory, F-wave, and H-reflex. You can configure the list by adding, changing, rearranging, and
deleting entries. Nerve lists are unique to each user name, but are global to all NCV-type tests.
Each nerve has two tables associated with it, which you can also configure:
81
• Site table
Lists the stimulation sites, which are used to record nerve measurements.
• Segment table
Lists the nerve segments over which conduction velocities are calculated.
You can change the reference site, sites, and segments for a nerve while you are conducting a test. The changes
apply only to the test you are conducting and do not affect the configured settings.
Changing reference site, sites, or segments for the current test
Deleting an entry from the nerve list, site table, or segment table
Select the nerve you want for NCV simulation and recording sites.
Median motor
Median F-wave
Ulnar motor
Ulnar F-wave
Tibial motor
Tibial F-wave
Peroneal motor
Sural sensory
The F-wave and H-reflex test protocols are specialized NCV protocols. Like the NCV test protocol, they measure the
speed and intensity of electrical signals that travel along nerves. However, they have slightly different default settings for
82
parameters and in addition have split gain controls to allow the F-wave and H-reflex responses to be displayed at a more
sensitive gain than the motor response.
There are two types of F-wave tests:
• Markers mode
Uses time and amplitude markers.
• Cursors mode
Uses autocursors to measure onset and amplitude of both motor and F-wave responses, and calculates values for
Mean-F, Min-F, Max-F, Min-Max, and F-M ratio.
Both the F-wave and the H-reflex test protocols open to the same nerve list, which contains motor, sensory, F-wave
(markers), F-wave (cursors), and H-reflex nerves. The type of nerve you select determines the test screen that appears
and the default parameters that are in effect.
F-wave toolbars
H-reflex toolbars
83
You can work with the waveforms in all the standard ways. For example, you can move and clear them, and view
them in either raster or overlay format.
• Nerve table
Contains information about the nerve selected from the nerve list. The columns contain the name and reference or
recording site. You cannot change this information on the test screen; however, you can configure the information on
the nerve list from which it comes.
• Cursor table
Contains information reported by the F-wave cursors.
• Measurements table
Shows various calculations based on the latency and amplitude markers.
Use the View key on the custom keyboard to cycle through the windows.
View 1 The opening view. All five windows appear.
View 2 NCV/F/H Data and all tables.
View 3 Stored NCV/F/H Traces and all tables
____________________
See also
F-wave toolbars
H-reflex toolbars
• Channels
The F-wave and H-reflex test protocols use one channel only. Data is always input on channel 1.
• Waveforms
Averaging is not used for the F-wave and H-reflex test protocols. Approximately 20 sweeps should be collected for
adequate evaluation.
• Manual or automatic store
The standard test defaults to manual storage (you use the Store control each time you want to store data). You can
change to automatic storage (the test automatically stores the waveforms following the delivery of a single stimulus).
• Split gain
F-wave and H-reflex responses are much lower in amplitude than motor responses and have a longer latency. When
the gain is set to properly view the F-wave or H-reflex response, the motor response is normally clipped. The split gain
feature of the Stored Traces window helps you view the data better. Using the Split control in the toolbar, you can set
the split gain separately from the M-wave gain. This lets you enlarge the F-wave or H-reflex so you can evaluate it
properly.
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
84
• Footswitch
You can configure the footswitch to either store data or deliver stimulations.
• Temperature probe
The temperature probe is sampled at least once every 5 seconds and displayed on the test screen. The value is
displayed in degrees Celsius.
____________________
See also
Averaging waveforms
F-Wave Toolbars
Primary toolbar
85
Secondary toolbar
H-Reflex Toolbars
Primary toolbar
Secondary toolbar
86
Sets the hicut filter value for the current channel.
87
• H-reflex: Increase stimulus intensity gradually, observing the H-
reflex and M-wave on the display. The maximal amplitude of the H-
reflex is normally obtained with low intensity stimulation. With
supramaximal stimulation, the H-reflex is no longer obtainable and
is sometimes replaced by an F-wave.
9 Continue with the stimulations until you get a satisfactory waveform for the first trial.
Optional. Use the NR column to record no response at a specific stimulation site. Click in the field to display “NR.
10 If storing manually, store the responses after each stimulation (press Store).
11 Apply markers as needed using the marker controls on the Markers function toolbar (on the View menu, click
Markers Bar). Adjust markers using the mouse.
12 Repeat until you have collected 20 trials.
13 Repeat the process for each site.
14 Click Next Nerve to return to the nerve list.
15 Select the next nerve and repeat the test steps until you have tested all the nerves that you want for the patient.
16 Print the results, if desired.
Printing and saving reports
17 Save the test. To do this, click the File menu and then click either Close or Exit.
SEP
SEP toolbars
88
SEP data
SEP Windows
SEP data
SEP toolbars
SEP Toolbars
89
The four controls for channel 1, 2, 3, and 4 set the
gain for the channel.
Sets the sweep speed for all channels.
Secondary toolbar
90
Turns the indicated channel (1, 2, 3, or 4) on and
off and sets the gain, hicut filter, and locut filter
values for that channel.
Turns common reference on and off for a channel.
Selecting a common recording reference
A quick way of selecting all channels and setting
the same gain, hicut filter, and lowcut filter values
for them all. The values here override the values
above.
Sets the gain.
SEP Data
• Channels
The SEP tests feature a full range of testing in two or four channels. When in live mode, the live data displays for each
channel that you have made active.
• Filters
Two filters can be used: hicut and locut.
• Waveforms
You can save up to 32 waveforms or 16 trials in each SEP test (whichever comes first)—32 single-channel trials, 16
two-channel trials, or 8 four-channel trials. You can work with the waveforms or trials in all the standard ways. For
example, you can move and clear them, and view them in either raster or overlay format.
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
91
• Averaging
You can view either live or averaged waveforms. However, data for EP protocols is usually acquired in trials by
averaging waveforms. In averaging mode, the waveforms that appear in the Live window represent the average of the
total number of responses collected. You set the number of responses for each average. Averaging is on by default
when the standard SEP test protocols open, but you turn it off if desired.
You can configure the trace labels on the right side of the EP Traces window to show whether the waveform is “Live
or “Avg.
• Grand averaging
Grand averaging, used only in the evoked potential protocols, is the technique of averaging all waveforms (stored or
live) collected in a single channel and side. If you have acquired several trials but they are a bit noisy, a grand
average provides a smoother trace.
• Gain
In evoked potential work, the input sensitivity is usually 10 microvolts/vertical division. The evoked potentials
themselves, however, are commonly in the range of one microvolt or less. If the averaged EP were to be displayed at
the same factor as the input, its morphology barely would be appreciated. Therefore, two scale factors are defined:
one for the input (amplifier gain) and one for averaging (display gain).Thus it becomes necessary to define a scale
factor separately from that of the input to be used for the displayed average.
When you adjust the gain in the averaging mode, you are adjusting the gain on the display only. When you adjust the
gain in live mode, you are adjusting the amplifier gain and the display gain at the same time.
• Repetition rate
Using the RR control on the toolbar, you can set the time interval between each stimulation when averaging is being
used. Note that this rate affects the total time required to obtain a single averaged response.
• Artifact reject function
By default, the artifact reject function is on when the standard SEP test protocols open. You can turn it on and off with
the Reject On/Off control in the toolbar. When on, the test automatically rejects any response exceeding 95% of full
amplifier scale. The Rej% field in the toolbar shows the percentage of responses that have been rejected.
• Auto cursors
Cursors are used in the SEP test protocols to mark important amplitude and latency features on averaged waveforms
(they cannot be applied to live waveforms). Once cursors have been applied, you can reposition them as needed. The
default cursors are set up with defaults appropriate for SEP responses, but you can configure them according to your
needs.
When you are setting up auto cursors, you can enter the expected normal values. For example, the N20 latency cursor for
the SEPU response should normally occur somewhere between 18 to 25 milliseconds. So, for the N20 norm, you
could enter <25 as the norm value. Norm values are displayed when you review test studies using the TabData
function and in printed reports, which lets you compare patient values with the norm values. You can also enter norms
for interpeak latency measurements and amplitude measurements.
____________________
See also
Averaging waveforms
92
To run the SEP tests
1 Open Sierra.
Opening and closing Sierra
If you use the Store control while the averager is running, the active waveform is stored, the averager resets to
zero, and stimulation continues.
12 For this test, you will usually want to replicate results twice and evaluate them for consistency. So, clear the averager
using the Clear key on the custom keyboard and repeat data acquisition to collect a second trial.
You can create a grand average of all waveforms (stored or live) for a smoother overall trace. Use the Grand
Average control on the Trace Functions function toolbar (on the View menu, click Trace Functions Bar).
13 Press the Cursor key on the custom keyboard and use the Trial Sel knob to highlight the trial that will receive the
cursors. Then press the Auto Cursors function key. Adjust the cursors if necessary.
Working with cursors
93
14 Print the results, if desired.
Printing and saving reports
15 Save the test. To do this, click the File menu and then click either Close or Exit.
BAEP
The BAEP (brainstem auditory evoked potential) test records brain stem evoked potentials using click stimuli. In this test,
headphones or ear inserts can be used for stimulation. By checking the pathway from the ear to the brain, the test helps
to uncover the cause of hearing and balance problems and to evaluate auditory thresholds.
Responses are usually averaged for this type of test in order to unmask the tiny evoked potentials from background noise
that might obscure them. Cursors mark important latency and amplitude features on waveforms; the default cursors are
set up with defaults appropriate for BAEP responses, but you can configure them according to your needs.
BAEP toolbars
BAEP data
BAEP Windows
94
View 3 Measurements window only.
____________________
See also
Working with waveforms
BAEP data
BAEP toolbars
BAEP Toolbars
95
Sets the right-side behavioral threshold.
96
BAEP Data
• Channels
The BAEP test protocol features testing in the full range of one to four channels. When in live mode, the live data
displays for each channel that is turned on.
• Filters
Two filters can be used: hicut and locut.
• Waveforms
You can save up to 32 waveforms or 16 trials in each BAEP test (whichever comes first)—32 single-channel trials, 16
two-channel trials, or 8 four-channel trials. You can work with the waveforms or trials in all the standard ways. For
example, you can move and clear them, and view them in either raster or overlay format.
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
• Averaging
You can run in either live or averaging mode. However, data for EP protocols is usually acquired in trials by averaging
waveforms. In averaging mode, the waveform that appears in the Live window represents the average of the total
number of responses collected. You can set the total number of responses for each average. Averaging is on by
default when the standard BAEP test protocol opens, but you can turn it off if desired.
You can configure the trace labels on the right side of the EP Traces window to show whether the waveform is “Live or
“Avg.
• Grand averaging
Grand averaging, used only in the evoked potential protocols, is the technique of averaging all waveforms (stored or
live) collected in a single channel and side. If you have acquired several trials but they are a bit noisy, a grand
average provides a smoother trace.
• Gain
In evoked potential work, the input sensitivity is usually 10 microvolts/vertical division. The evoked potentials
themselves, however, are commonly in the range of one microvolt or less. If the averaged EP were to be displayed at
the same factor as the input, its morphology barely would be appreciated. Therefore, two scale factors are defined:
one for the input (amplifier gain) and one for averaging (display gain).Thus it becomes necessary to define a scale
factor separately from that of the input to be used for the displayed average.
When you adjust the gain in the averaging mode, you are adjusting the gain on the display only. When you adjust the gain
in live mode, you are adjusting the amplifier gain and the display gain at the same time.
• Repetition rate
Using the RR control on the toolbar, you can set the time interval between each stimulation when averaging is being
used. Note that this rate affects the total time required to obtain a single averaged response.
• Artifact reject function
By default, the artifact reject function is on when the standard BAEP test protocols open. You can turn it on and off
with the Reject On/Off control in the BAEP Average window. When on, the test automatically rejects any response
exceeding 95% of full amplifier scale. The Rej% field in the toolbar shows the percentage of responses that have
been rejected.
• Responses
The initial movement of the headphone transducer away from and toward the tympanic membrane produces
rarefaction and condensation clicks, respectively. Rarefaction clicks are usually used in routine BAEP testing.
Whatever the polarity, the testing protocol should be the same as that used in establishing normative values.
97
• Auto cursors
Cursors are used in the BAEP test protocols to mark important amplitude and latency features on averaged
waveforms (they cannot be applied to live waveforms). Once cursors have been applied, you can reposition them as
needed. The default cursors are set up with defaults appropriate for BAEP responses, but you can configure them
according to your needs.
When you are setting up auto cursors, you can enter the expected normal values. Norm values are displayed when
you review test studies using the TabData function and in printed reports, which lets you compare patient values with
the norm values. You can also enter norms for interpeak latency measurements and amplitude measurements.
____________________
See also
Averaging waveforms
Selecting Inserts automatically subtracts one millisecond from all cursor values to offset the stimulus delay
produced by the insert earphones.
98
8 Select the side of stimulation on the toolbar.
9 Adjust the stimulus intensity levels (IntenL and IntenR).
The stimulus intensity level is measured in dB nHL (normalized hearing level). For the approximate sPL (sound
pressure level) equivalent, add 33 dB to the displayed value.
10 Optional. Set behavioral threshold intensities in the toolbar (ThreshL and ThreshR). If you do not wish to use the
threshold parameter, leave these set to zero. Turn off the clicks after determining thresholds using the Clicks On/Off
control on the toolbar.
The threshold stimulus intensity level is added to the stimulus intensity level, determining the total stimulus
intensity. Intensity + threshold = output intensity (for example, 70 dB stimulus intensity + 10 dB threshold stimulus
intensity = 80 dB click).
11 Adjust the masking level on the toolbar with the Mask control.
The masking value is automatically subtracted from the stimulus intensity level to determine the masking intensity
level. The masking intensity level is adjusted in increments as the stimulus intensity level is adjusted. For
example, if the stimulus intensity is set at 80 dB and the differential masking is set at 30 dB, the masking level is
50 dB.
12 Select the click polarity on the toolbar (Rare, Cond, or All).
13 Adjust the number of stimulations that will be averaged for the trial using the AvgCt control in the BAEP Average
window.
14 Set or adjust other parameters as needed (for example, sweep and repetition rate).
15 Deliver the stimulation using the Run/Stop control.
If you are averaging, stimulation stops automatically when the number of stimulations in AvgCt has been collected. To
stop stimulating at any time, use the Run/Stop control or footswitch.
16 Store the responses.
If you use the Store control while the averager is running, the active waveform is stored, the averager resets to
zero, and stimulation continues.
17 For this test, you will usually want to replicate results twice and evaluate them for consistency. So, clear the averager
and repeat data acquisition to obtain a second trial.
You can create a grand average of all waveforms (stored or live) for a smoother overall trace. Use the Grand
Average control on the Trace Functions function toolbar (on the View menu, click Trace Functions Bar).
18 Press the Cursor key on the custom keyboard and use the Trace Sel knob to highlight the trial that will receive the
cursors. Then use the Auto Cursors function key. Adjust the cursors if necessary.
Working with cursors
19 Print the results, if desired.
Printing and saving reports
20 Save the test. To do this, click the File menu and then click either Close or Exit.
AEP
99
About the AEP Test
The AEP (auditory evoked potential) test evaluates the state of auditory nerve and brainstem auditory sensory pathways.
It records short, middle, and long latency auditory responses. The stimulator can generate click, tone, and pip stimuli.
Responses are usually averaged for this type of test in order to unmask the tiny evoked potentials from background noise
that might obscure them. However, you have the option of switching to live (single) stimulation. Cursors mark important
latency and amplitude features on waveforms; the default cursors are set up with defaults appropriate for AEP responses,
but you can configure them according to your needs.
The AEP protocol is similar to the BAEP test, except that it provides three additional stimulus types. In addition to click
stimuli, the AEP protocol includes a tone burst (10-30-10) and two tone pips (2-0-2 and 2-1-2). The intensity units can be
switched from nHL (the defaults) to SPL units.
All EP protocols default to parameter values specified in the American Electroencephalographic Society’s Guidelines for
Clinical Evoked Potential Studies (1992). To order a copy of the guidelines, contact the Society at 203-243-3977.
AEP data
AEP Windows
See also
100
Working with waveforms
AEP data
AEP toolbars
AEP Toolbars
101
side of stimulation.
Turns the audio masking on and off.
102
Sets the gain.
AEP Data
• Channels
The AEP test protocol features testing in the full range of one to four channels. When in live mode, the live data
displays for each channel that is turned on.
• Filters
Two filters can be used: hicut and locut.
• Waveforms
You can save up to 32 waveforms or 16 trials in each AEP test (whichever comes first)—32 single-channel trials, 16
two-channel trials, or 8 four-channel trials. You can work with the waveforms or trials in all the standard ways. For
example, you can move and clear them, and view them in either raster or overlay format.
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
• Averaging
You can run in either live or averaging mode. However, data for EP protocols is usually acquired in trials by averaging
waveforms. In averaging mode, the waveform that appears in the Live window represents the average of the total
number of responses collected. You can set the total number of responses for each average. Averaging is on by
default when the standard AEP test protocol opens, but you can turn it off if desired.
You can configure the trace labels on the right side of the EP Traces window to show whether the waveform is “Live or
“Avg.
• Grand averaging
Grand averaging, used only in the evoked potential protocols, is the technique of averaging all waveforms (stored or
live) collected in a single channel and side. If you have acquired several trials but they are a bit noisy, a grand
average provides a smoother trace.
• Gain
In evoked potential work, the input sensitivity is usually 10 microvolts/vertical division. The evoked potentials
themselves, however, are commonly in the range of one microvolt or less. If the averaged EP were to be displayed at
the same factor as the input, its morphology barely would be appreciated. Therefore, two scale factors are defined:
one for the input (amplifier gain) and one for averaging (display gain).Thus it becomes necessary to define a scale
factor separately from that of the input to be used for the displayed average.
When you adjust the gain in the averaging mode, you are adjusting the gain on the display only. When you adjust the gain
in live mode, you are adjusting the amplifier gain and the display gain at the same time.
• Repetition rate
Using the RR control on the toolbar, you can set the time interval between each stimulation when averaging is being
used. Note that this rate affects the total time required to obtain a single averaged response.
103
• Artifact reject function
By default, the artifact reject function is on when the standard AEP test protocols open. You can turn it on and off with
the Reject On/Off control in the AEP Average window. When on, the test automatically rejects any response
exceeding 95% of full amplifier scale. The Rej% field in the toolbar shows the percentage of responses that have
been rejected.
• Type of audio stimulator
You can set the stimulator type to phones or inserts.
Phones
The test uses the head phone stimulator types normally used with BAEP tests.
Inserts
Insert ear phones have a 1-millisecond acoustic delay line. With this stimulator type, 1 millisecond is automatically
subtracted from all cursor or marker latency measurements. This setting has no effect on the start of sweep and is
used only to compensate the latency measurements. The sweep delay is set independently and is properly
accounted for in all latency measures.
• Type of stimulation
You can set the type of stimulation to click, tone, or pips.
Clicks
This type of stimulation uses 50 to 500 µs square-wave clicks. The impulse is generated using a fixed pulse width
of 100 µs. You can set the phase for rarefaction, condensation, or alternating phase clicks.
Tones/pips
This type of stimulation uses frequencies from 250 to 8000 Hz in 10 steps. You can set the polarity as either
positive, negative, or alternating phase. The rise, plateau, and fall times for the tone burst are 10-30-10 ms. The
rise, plateau, and fall times for the tone pips are 2-0-2 and 2-1-2 cycles. Frequency stability is better than 0.1%.
• Auto cursors
Cursors are used in the AEP test protocol to mark important amplitude and latency features on averaged waveforms
(they cannot be applied to live waveforms). Once cursors have been applied, you can reposition them as needed. The
default cursors are set up with defaults appropriate for AEP responses, but you can configure them according to your
needs.
When you are setting up auto cursors, you can enter the expected normal values. Norm values are displayed when
you review test studies using the TabData function and in printed reports, which lets you compare patient values with
the norm values. You can also enter norms for interpeak latency measurements and amplitude measurements.
____________________
See also
Averaging waveforms
104
1 Open Sierra.
Opening and closing Sierra
Selecting Inserts automatically subtracts one millisecond from cursor and marker latency values to offset the
stimulus delay produced by the insert earphones.
8 Select the side of stimulation on the toolbar.
9 Adjust the stimulus intensity levels (IntenL and IntenR).
The stimulus intensity level is measured in dB nHL (normalized hearing level). For the approximate sPL (sound
pressure level) equivalent, add 33 dB to the displayed value.
10 Optional. Set behavioral threshold intensities in the toolbar (ThreshL and ThreshR). If you do not wish to use the
threshold parameter, leave these set to zero. Turn off the clicks after determining thresholds using the Clicks On/Off
control on the toolbar.
The threshold stimulus intensity level is added to the stimulus intensity level, determining the total stimulus
intensity. Intensity + threshold = output intensity (for example, 70 dB stimulus intensity + 10 dB threshold stimulus
intensity = 80 dB click).
11 Adjust the masking level on the toolbar with the Mask control.
The masking value is automatically subtracted from the stimulus intensity level to determine the masking intensity
level. The masking intensity level is adjusted in increments as the stimulus intensity level is adjusted. For
example, if the stimulus intensity is set at 80 dB and the differential masking is set at 30 dB, the masking level is
50 dB.
12 Select the type of stimulation from the toolbar (Click, Tone, Pip202, or Pip212). Stimulus characteristics are listed
below.
• Click: 100-µs square wave click.
105
• Tone: 50-ms duration with 10-ms rise, 30-ms plateau, 10-ms fall.
• Pip202: 2 cycles rise, 0 plateau, 2 cycles fall.
• Pip212: 2 cycles rise, 1 cycle plateau, and 2 cycles fall.
13 Select the units on the toolbar that you want to use for measuring auditory intensity—nHL (normal hearing level) or
SPL (sound pressure level) intensity units.
SPL values are based on audio calibration tables in the audio.cfg file, which you can adjust.
Editing the audio.cfg file
14 Select the stimulus polarity from the toolbar.
• For Click stimulator type: Cond (condensating), Rare (rarefaction), or Alt
(alternating).
• For Tone or Pip stimulator types: Pos (positive), Neg (negative), or Alt (alternating).
15 If you selected Tone or Pip, you can set the stimulus frequency level on the toolbar. Select from the Freq list. If you
enter a value outside of this list, the test automatically defaults to the closest value in the list.
16 Adjust the number of stimulations that will be averaged for the trial using the AvgCt control in the BAEP Average
window.
17 Set or adjust other parameters as needed (for example, sweep and repetition rate).
18 Deliver the stimulation using the Run/Stop control.
If you are averaging, stimulation stops automatically when the number of stimulations in AvgCt has been collected. To
stop stimulating at any time, use the Run/Stop control or footswitch.
19 Store the responses.
If you use the Store control while the averager is running, the active waveform is stored, the averager resets to
zero, and stimulation continues.
20 For this test, you will usually want to replicate results twice and evaluate them for consistency. So, clear the averager
and repeat data acquisition to obtain a second trial.
You can create a grand average of all waveforms (stored or live) for a smoother overall trace. Use the Grand
Average control on the Trace Functions function toolbar (on the View menu, click Trace Functions Bar).
21 Press the Cursor key on the custom keyboard and use the Trace Sel knob to highlight the trial that will receive the
cursors. Then use the Auto Cursors function key. Adjust the cursors if necessary.
Working with cursors
22 Print the results, if desired.
Printing and saving reports
23 Save the test. To do this, click the File menu and then click either Close or Exit.
SPL values are based on audio calibration tables in the audio.cfg file. See the values below. You can change the values
in any text editor program.
106
To edit the audio.cfg file
1 Open a text editor program on your computer (for example, Notepad or WordPad).
2 Open the audio.cfg file (the default installed path is C:\Cadwell\SierraWin\audio.cfg).
3 Change the values in the audio.cfg file.
4 Save your changes and exit the file.
107
4KHz 3 4 3 13 16 21
6KHz 10 8 8 13 24 19
8KHz 16 14 14 13 30 29
108
VEP
The VEP (visual evoked potential) test protocol performs a full range of visual evoked potential testing. Visual evoked
potentials are elicited in response to light stimulation. In addition to detecting various diseases of the optic nerve, the VEP
test protocol can be used to detect macular degeneration, glaucoma, field defects, and visual acuity.
For this test protocol, you can use a goggle stimulator, Ganzfeld stimulator, LED stimulator, or video display. Responses
are usually averaged for this type of test in order to unmask the tiny evoked potentials from background noise that might
obscure them. Cursors mark important latency and amplitude features on waveforms; the default cursors are set up with
defaults appropriate for VEP responses, but you can configure them according to your needs.
VEP data
VEP Windows
See also
109
Working with waveforms
VEP data
VEP toolbars
VEP Toolbars
Secondary toolbar
110
Turns the indicated channel (1, 2, 3, or 4) on and
off and sets the gain, hicut filter, and locut filter
values for that channel.
Turns common reference on and off for a channel.
Selecting a common recording reference
A quick way of selecting all channels and setting
the same gain, hicut filter, and lowcut filter values
for them all. The values here override the values
above.
Sets the gain.
VEP Data
• Channels
The VEP test protocol features testing in the full range of one to four channels. When in live mode, the live data
appears for each channel that is turned on.
• Filters
Two filters can be used: hicut and locut.
• Waveforms
You can save up to 32 waveforms or 16 trials in each VEP test (whichever comes first)—32 single-channel trials, 16
two-channel trials, or 8 four-channel trials. You can work with the waveforms or trials in all the standard ways. For
example, you can move and clear them, and view them in either raster or overlay format.
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
111
• Averaging
You can run in either live or averaging mode. However, data for EP protocols is usually acquired in trials by averaging
waveforms. In averaging mode, the waveform that appears in the Live window represents the average of the total
number of responses collected. You can set the total number of responses for each average. Averaging is on by
default when the standard VEP test protocol opens, but you can turn it off if desired.
You can configure the trace labels on the right side of the EP Traces window to show whether the waveform is “Live
or “Avg.
• Grand averaging
Grand averaging, used only in the evoked potential protocols, is the technique of averaging all waveforms (stored or
live) collected in a single channel and side. If you have acquired several trials but they are a bit noisy, a grand
average provides a smoother trace.
• Gain
In evoked potential work, the input sensitivity is usually 10 microvolts/vertical division. The evoked potentials
themselves, however, are commonly in the range of one microvolt or less. If the averaged EP were to be displayed at
the same factor as the input, its morphology barely would be appreciated. Therefore, two scale factors are defined:
one for the input (amplifier gain) and one for averaging (display gain). Thus it becomes necessary to define a scale
factor separately from that of the input to be used for the displayed average.
When you adjust the gain in the averaging mode, you are adjusting the gain on the display only. When you adjust the
gain in live mode, you are adjusting the amplifier gain and the display gain at the same time.
• Repetition rate
Using the RR control on the toolbar, you can set the time interval between each stimulation when averaging is being
used. Note that this rate affects the total time required to obtain a single averaged response.
• Artifact reject function
By default, the artifact reject function is on when the standard VEP test protocol opens. You can turn it on and off with
the Reject On/Off control in the VEP Average window. When on, the test automatically rejects any response
exceeding 95% of full amplifier scale. The Rej% field in the toolbar shows the percentage of responses that have
been rejected.
• Auto cursors
Cursors are used in the VEP test protocol to mark important amplitude and latency features on averaged waveforms
(they cannot be applied to live waveforms). Once cursors have been applied, you can reposition them as needed. The
default cursors are set up with defaults appropriate for VEP responses, but you can configure them according to your
needs.
When you are setting up auto cursors, you can enter the expected normal values. Norm values are displayed when
you review test studies using the TabData function and in printed reports, which lets you compare patient values with
the norm values. You can also enter norms for interpeak latency measurements and amplitude measurements.
_____________________
See also
Averaging waveforms
112
To run the test
1 Open Sierra.
Opening and closing Sierra
If you use the Store control while the averager is running, the active waveform is stored, the averager resets to
zero, and stimulation continues.
113
13 For this test, you will usually want to replicate results twice and evaluate them for consistency. So, clear the averager
and repeat data acquisition to collect a second trial.
You can create a grand average of all waveforms (stored or live) for a smoother overall trace. Use the Grand
Average control on the Trace Functions function toolbar (on the View menu, click Trace Functions Bar).
14 Press the Cursor key on the custom keyboard and use the Trace Sel knob to highlight the trial that will receive the
cursors. Then use the Auto Cursors function key. Adjust the cursors if necessary.
Working with cursors
15 Print the results, if desired.
Printing and saving reports
16 Save the test. To do this, click the File menu and then click either Close or Exit.
114
Check and display width for 3-inch LED stimulator
F = 57.3 x W/D
where:
F = field size in degrees of arc
W = width of the monitor screen in centimeters
D = distance from the screen to the eye in centimeters
Blink Reflex
Blink reflex studies are performed when abnormalities of the 5th (trigeminal) or the 7th (facial) cranial nerves or their central
connections are suspected. The Sierra blink reflex protocol is optimized for performing these studies by the following:
• Markers have been configured specifically for the measurements required in a blink study.
• Both left and right responses and difference comparisons can be recorded with one electrode setup. This is because
the test automatically switches channel definitions when you select a different side.
The blink reflex protocol features two windows. Use the View key on the custom keyboard to cycle through them.
• Blink Traces
The graticule is split into two areas. All left-side trials display in the top half of the window and all right-side trials
display in the bottom half. The left-side trials are set to record with channel 1 (ipsilateral) on top and channel 2
(contralateral) below. The right-side trials display with channel 2 (ipsilateral) on top and channel 1 (contralateral)
below.
Vertical dotted lines in both areas show the position of the R1, R2i, and R2c markers.
You can work with the trials in all the standard ways. For example, you can move and clear them and view them in either
raster or overlay format.
• Measurements table
As you position markers for each side, their corresponding latency measurements display in this table. The table
115
provides left-side (L) and right-side (R) latency measurements for R1, R2i, R2c, as well as the latency difference
between R2i and R2c on the left and right sides.
Use the View key on the custom keyboard to cycle through the windows.
View 1 The opening view. Both windows appear.
View 2 Blink Traces window only.
View 3 Measurements window only.
____________________
See also
Primary toolbar
The two controls for channel 1 and 2 set the gain for
the indicated channel.
Sets the sweep speed for all channels.
Secondary toolbar
116
Turns the indicated channel (1, 2, 3, or 4) on and
off and sets the gain, hicut filter, and locut filter
values for that channel.
Turns common reference on and off for a channel.
Selecting a common recording reference
A quick way of selecting all channels and setting
the same gain, hicut filter, and lowcut filter values
for them all. The values here override the values
above.
Sets the gain.
• Channels
Normally two channels are used for the blink reflex test protocol.
• Waveforms
Single sweeps of data are acquired in trials. Up to 16 trials can be stored.
• Markers
Three markers per side display in the Blink Traces window. They extend for only half of the graticule (left markers on
the top and right markers on the bottom). These markers represent the three desired latency measurements of a blink
study, and by default appear at latency locations consistent with a normal blink response. You can adjust the markers
if desired using cursor or knob.
R1 Onset on channel 1 (default start at 10 ms).
R2i Onset on channel 1 (default start at 30 ms).
117
R2c Onset on channel 2 (default start at 31 ms).
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
• Footswitch
You can configure the footswitch to either store data or deliver stimulations.
____________________
See also
Connect the electrode leads from the patient’s left side (orbicularis oculi) to the Ch 1 inputs on the amplifier, and
connect the electrode leads from the right side to the Ch 2 inputs. Electrodes must be connected to the amplifier
in this configuration to ensure proper display of test data.
Active Belly of the orbicularis oculi below the canthus (ipsi
and contra sides)
Reference Side of nose (ipsi and contra sides)
Ground Forehead or cheek
Selecting a different side automatically stores the currently active trial on the screen. The newly stored traces
overlay (per side) by default.
6 Adjust parameters as needed.
118
7 Position the stimulator at the stimulation site (supraorbital nerve) and stimulate.
8 Repeat stimulation using the Single control and store using the Store control.
9 Review the trials in the Blink Traces window.
Trials collected during left-side stimulation are displayed in the top half of the screen with the response recorded on
the ipsilateral side (e.g., 1:L) positioned above the response recorded on the contralateral side (2:L).
Responses collected during right side stimulation display in the bottom half of the screen. Eliminating the need to
change electrodes at the amplifier, the Sierra automatically inverts the order in which trials collected during right side
stimulation are displayed so that the ipsilateral response (2:R) appears above the contralateral response (1:R).
10 Use the knobs to adjust the position of markers if needed.
11 Use the Chg Side function key to change side and repeat the test steps.
12 Print the results, if desired.
Printing and saving reports
13 Save the test. To do this, click the File menu and then click either Close or Exit.
RNS
RNS (repetitive nerve stimulation) is the technique of repeated supramaximal stimulations of a nerve while recording the
electrical activity from a muscle innervated by the nerve. The test is used primarily for diagnosing myasthenia gravis.
In this protocol, a trial is recorded by presenting a train of stimuli and recording the compound muscle action potentials.
RNS toolbars
RNS data
RNS Windows
The RNS protocol features two windows. Use the View key on the custom keyboard to cycle through them.
• RNS Traces
Displays the traces and trials resulting from stimulation. You can work with the trials in all the standard ways. For
119
example, you can move and clear them, and view them in either raster or overlay format. Note that these functions
apply only to complete trials; you cannot apply them to individual traces within a trial.
• RNS table
Summarizes amplitude and area measurements for two responses that you select. The table automatically updates
each time you select a different response for comparison.
Use the View key on the custom keyboard to cycle through the windows.
View 1 The opening view. Both windows appear.
View 2 RNS Traces window only.
View 3 Measurements window only.
____________________
See also
RNS data
RNS toolbars
RNS Toolbars
Primary toolbar
120
Sets the horizontal distance in divisions between
successive samples in the RNS protocol.
Resets the timer to zero. The timer automatically
starts at the end of each train to time the interval to
the next train.
Sets the side of stimulation (Right, Left, Both).
Secondary toolbar
121
Sets the gain.
RNS Data
• Channels
Typically only one channel is used with the RNS test protocol.
• Filters
Two filters can be used: hicut and locut.
• Trial waveforms
During data acquisition, the motor response to each stimulus in the stimulus train is recorded and stored separately in
trace memory. Collectively, the set of responses to the stimulus train composes a trial. All of the traces in the trial are
locked together in a superimposed position with the vertical baseline at the same screen position. Adjustment to the
vertical position moves all the traces as a group. Effectively they behave as a single compound waveform. Traces are
spread in time so that the individual responses can be viewed as an overlapping sequence of responses. This allows
amplitude decrements and response quality to be viewed and still maintains a compact display of the information.
You collect a trial by starting stimulation. Stimulation continues until the entire pulse train has been delivered or you
stop data acquisition.
• Stimulus train
The stimulus train length is the number of stimuli to be presented per trial (from 1 to 150). Repetitive stimulation stops
automatically when the specified number of stimuli have been delivered.
• Cursors
Cursors for onset (O) and peak (P) latency are automatically placed on responses 1 and 5 in a train (default setting).
• Timer
When train stimulation is being used, a timer on the toolbar shows the elapsed time in minutes and seconds between
stimulus trains. It has 1-second resolution and is updated every second. The timer starts immediately after the first
train is stored. It automatically resets to zero after each subsequent train is stored. You can reset the timer back to
zero by clicking the Reset Timer function key or the Timer control on the toolbar.
• Trace spread
The trace spread is the number of divisions to be applied between each successive trace to spread them in the
window. You can adjust the number of divisions between each response in a stimulus train with the Spread control.
Use it to move responses closer together or farther apart.
• Manual or automatic store
The standard test defaults to manual storage (you use the Store control each time you want to store a trial). You can
change to automatic storage (the test automatically stores the trial as soon as acquisition of the data is complete).
• Amplitude and area ratios
The following equation is used to calculate the amplitude and area ratios in the RNS table:
D = (VR2 – VR1 / VR1) x 100%
where:
D = decrement/increment; a negative sign (-) indicates a decrement and a positive sign (+) indicates an increment
122
V = value (amplitude or area)
R1 = response 1
R2 = response 2
• Trace labels
Trace labels provide information about the traces being displayed. You can configure the information reported by the
label (such as channel number, gain setting, and side of stimulation).
____________________
See also
6 Decide whether you want single or train stimulation (train is the default). To turn train stimulation on or off, use the
Single/TRAIN function key or the Single/Train control on the toolbar.
Set the number of stimuli you want to present in a single stimulus train with the Train control.
Use single stimulation to establish supramaximal stimulation, then switch to train stimulation. The control is
pushed in for train stimulation.
7 Adjust other parameters as needed.
8 Deliver the stimulation.
Train stimulation: Press the Single or Run/Stop key on the custom keyboard, the stimulator dial, or the footswitch to
present a train of stimuli up to the train number.
123
Single stimulation: Press the Single key on the custom keyboard, the stimulator dial, or the footswitch to present a
single stimulus. (Pressing the Run/Stop control while in the single stimulation mode starts repetitive stimulation.)
You can stimulate at up to 50 Hz, but at that level the sweep must be set to 2.0 ms/div.
For train stimulation, a timer on the toolbar shows the time elapsed between stimulus trains. The timer starts
immediately after the first train is stored. It automatically resets to zero after each subsequent train is stored. You
can reset the timer back to zero by using the Reset Timer function key or the Timer control in the toolbar.
9 If using single stimulation and manual storage, store responses with the Store control.
10 Adjust the spread between responses using the Spread control on the toolbar. You can adjust this after you collect
responses.
11 Cursors for onset (O) and peak (P) latency are automatically placed on responses 1 and 5 in a train (default setting).
Use the Rsp1 and Rsp2 controls on the toolbar to select the two responses within a train that you want to compare.
Selected responses are automatically marked with onset (O) and peak (P) latency cursors. Once you have selected
the responses for analysis, the amplitude (onset to peak) and area (negative peak) measurements are automatically
calculated and displayed. The amplitude and area ratios between the selected responses in each train are also
calculated and displayed.
12 Print the results, if desired.
Printing and saving reports
13 Save the test. To do this, click the File menu and then click either Close or Exit.
QuickMed Utilities
About the QuickMed Utilities
The QuickMed Utilities, available from the Edit menu, provide useful tools that are grouped under the following
categories:
124
Patient tab
Records tab
Startup tab
Use the procedures on the Patient tab to work with entire patient charts. A chart contains all the data records and reports
collected for an individual patient.
Procedures on the Records tab are used to work with selected parts of the patient chart, such as individual test data
records and reports.
Records tab
Example: If you tested the same patient on two separate occasions and saved two reports, the patient chart would
consist of the information you had entered on the patient (name, address, and so on), two test data records, and two
reports. Using the delete function on the Patient tab would remove all information collected for the patient, as well as all
125
test data records and reports. You would use the delete function on the Records tab if you want to delete only one test
data record or only one report from the chart.
Use the procedures on the Records tab to work with selected records in patient charts, such as individual test data
records and reports.
Procedures on the Patient tab are used to work with entire patient charts. A charts contains all the data records and
reports collected for an individual patient..
Patient tab
Example: If you tested the same patient on two separate occasions and saved two reports, the patient chart would
consist of the information you had collected on the patient (name, address, and so on), two test data records, and two
reports. Using the delete function on the Patient tab would remove all information collected for the patient, as well as all
test data records and reports. You would use the delete function on the Records tab if you want to delete only one test
data record or only one report from the chart.
Use the procedures on the User Databases tab to back up and restore information in the user database and to compress
the user database file.
User database information consists of the following:
• Profiles of all users, physicians, referring physicians, and technicians.
• All user settings (such as default settings for protocols).
• Report templates.
126
Use the procedures on the Report Templates tab to do the following
• Open the report template editor so that you can add or change a report template.
• Delete report templates.
• Back up report templates.
• Restore report templates from a backup.
For information on how to work with report templates, go to the ReportGen Help file (on the Help menu, click ReportGen
Help).
Startup Tab
Troubleshooting
Problem Symptoms
Printer problems
127
Patient Safety
Operating Characteristics
The Sierra II Wedge contains one constant-current electrical stimulator. The output generates a unipolar output. The
maximum outputs (which cannot occur simultaneously) are:
380 volts
100 mA
1000 microseconds pulse width
100 pulses per second
0.3 Watts r.m.s.
1.6 mA r.m.s.
Continuous, fast repetition rates and high currents will cause the output voltage to droop as a safety measure, stabilizing
at the smaller of the rated r.m.s. wattage or r.m.s. current.
The stimulator contains fault detectors that monitor for current delivered when no stimulus is requested, for current that is
higher than requested, and for other conditions. The fault detector will disable the high voltage, and/or clamp the outputs
until reset.
The maximum deliverable current for a single stimulus is a function of electrode impedance, and is calculated as:
I (ma) = 380 volts / R (kOhms), or R = 380 / I
• If needle electrodes are used, the current at the tip may be high enough to cause arcing and possible local tissue
damage.
• Patients who are unresponsive or cannot communicate, lack sensation, or are anesthetized may be injured by
prolonged high-intensity stimulation.
• Injury may occur if the electrode is directly applied to muscle, nerve, brain, spinal cord, or other exposed tissue.
Precautions
128
• Use a low-current stimulator with needle electrodes and in surgery to avoid injury
• Permanently applied surface electrodes should be at least 0.5 cm in diameter.
• Current densities for any electrode exceeding 2 mA r.m.s./cm2 may require the special attention of the operator.
• Avoid prolonged, high-current, high-repetition-rate stimulation.
• Avoid stimulation near implanted electronic devices (e.g., pacemakers).
• Do not stimulate a patient with an implanted electronic device unless a specialist medical opinion has first been
obtained.
• Electrodes should be considered a possible return path for electro-cautery under certain fault conditions, and should
be positioned to avoid burns or damage to the equipment.
• This equipment should not be used within 5 meters of therapeutic microwave or shortwave equipment.
• A video VEP monitor which does not meet IEC 60601-1-1 should have a protective earth connection.
• Avoid accidental electrical contact between stimulators, amplifiers, sensors and other isolated components and the
computer, cart, and other grounded or non-isolated components.
• Disconnect isolated components for storage or when contact with non-isolated components is possible.
• Do not wrap amplifiers in blankets or other insulators that impede cooling.
• Do not place amplifiers or stimulators against the patient for prolonged periods.
Backing Up
If you use the cart with your Sierra II Console or Sierra II Wedge and do not use the isolation transformer, you need to
ground the cart to the instrument.
129
To ground the cart
1 Remove the grounding pin (SG) from the Sierra back panel.
2 Insert the pin through the end of the grounding wire, and reattach it to the back panel.
3 Tighten the pin to secure the connection.
4 Insert the provided screw into the other end of the wire, and attach it to the back of the cart.
A blown fuse usually results from an overload condition, excessive vibration, or failure in the main module. If a fuse is
blown, the Sierra II Console or Sierra II Wedge will not power up.
Recommendations
Perform cleaning and sterilization procedures on a periodic basis to ensure the safe operation of your Sierra system. As
you clean, visually inspect the instrument and its components for damage or wear. Contact Cadwell if you notice damage
to the exterior of the instrument.
Customer support
Cadwell recommends that you refer to the AAMI Standards and Recommended Practices for Sterilization, Volume 1:
Good Hospital Practices, and Volume 2: Hospital Equipment and Industrial Process Control or equivalent standard text for
130
detailed sterilization instructions. These texts may be ordered by calling the Association for Advancement of Medical
Instrumentation at 1-800-332-2264, ext. 217.
Turn off and disconnect all Sierra components before cleaning. Do not attempt to service or repair
damaged or inoperable equipment.
Sierra Components
Use a soft cloth to wipe away any dust that has collected on the instrument casing. You can dampen the cloth slightly in a
mild soap-and-water solution or cold sterilization fluid such as Cidex Plus™ or Omnicide 28™ (active ingredient-
glutaraldehyde, 2.0%, inert ingredients 98.0%).
A soft brush or pressurized air duster can also be used to remove dust from hard-to-access areas such as the custom
keyboard, alphanumeric keyboard, and between connectors on the side and rear panels. To clean dust from the printer or
the monitor’s screen, wipe it with a soft cloth.
Turn off and disconnect the amplifier before cleaning. Keep all cleaning fluids away from electrical
connectors. Do not let solutions seep inside the amplifier. Do not use abrasive cleansers.
Stimulators
Visually inspect each stimulator for unusual wear or damage prior to each use. If you notice damage, disconnect the
stimulator immediately and contact Cadwell.
Customer support
Use a soft cloth to clean stimulator parts that come in contact with the patient after each use (e.g., electrical stimulator
probes, earphones, and LED goggles). Properly dispose of the foam tips used with the insert earphones after each use.
You can dampen the cloth slightly in a mild soap-and-water solution or cold sterilization fluid such as Cidex Plus™ or
Omnicide 28™ (active ingredient-glutaraldehyde, 2.0%, inert ingredients 98.0%).
In cases requiring the full sterilization regimen, the electrical stimulator probes (removed from the stimulator) can be
steam autoclaved at 250 °F (121 °C) and 15 psi for 15 minutes, or the standard recommended temperature, pressure, and
time for rubber products.
Turn off and disconnect stimulators before cleaning. Keep all cleaning fluids away from electrical
connectors. Do not let solutions seep inside the stimulators. Do not use abrasive cleansers.
Visually inspect the interface cables and power cords that are used with components, stimulators, and accessories. If you
notice unusual wear or breakage, disconnect the cable or cord immediately and contact Cadwell.
Customer support
Recording Electrodes
The quality of signal acquisition is affected by the condition of your recording electrodes. Routine inspection, careful
handling, frequent cleaning, and proper storage are essential practices for properly maintaining them.
The following maintenance procedures apply to nonspecialized electrodes:
131
• Inspect electrode leads for breaks, rust, corrosion, and debris prior to use.
• Handle electrodes carefully, and do not pull on the leads.
• Clean electrodes in warm water, using a toothbrush to scrub the surface of the electrode. If necessary, soak
electrodes to loosen debris. (Using a file or steel wool to scrub electrodes may lead to contamination of electrode
metal. This can result in signal artifact.)
• For proper sterilization techniques refer to the electrode manufacturer's instructions.
• If necessary, dry electrodes thoroughly to prevent corrosion from hard water deposits.
• Store electrodes in a clean, dry place.
• Do not tangle or kink electrode leads.
• Properly discard disposable electrodes after each use (e.g., disposable needle electrodes and disposable surface
electrodes).
Specialized electrodes are often more complex and require special care. Anyone using them should consult the
manufacturer for proper maintenance and handling techniques.
Warranty Information
Your Cadwell Sierra is covered by a one-year limited warranty at the time of purchase. Additional one-year service
contracts may be purchased from Cadwell Laboratories before the warranty period or previous service contract expires.
All warranties and service contracts are nontransferable.
Limited Warranty
(a) Cadwell warrants for a period of one year from the date of purchase, the system listed against defects in material and
workmanship. Notice of a defect and an explanation of circumstances concerning any claim that the system has
proven defective in material or workmanship shall be given to Cadwell Laboratories, Inc. within two (2) working days
of the discovery of the defect in the system. Upon such notice, Cadwell will follow the company service procedures
outlined on page 1 (of the warranty).
(b) Cadwell agrees to retrofit equipment. If the company makes standard improvements to components of the equipment
covered under this agreement during the term of the warranty and if such improvements are compatible with the
equipment located on the customer’s premises, then Cadwell will retrofit such physical equipment at no additional
charge. Cadwell reserves the right to make changes in specifications, construction or design of its products at any
time in such a manner as it may consider necessary or advisable.
Warranty exclusion
Sole purpose
132
Indemnification
Business use
Severability
No waiver
Label Symbols
Label Symbols
The following table explains the symbols that appear on equipment labels.
Symbol Title/Meaning Reference
Attention, consult accompanying IEC
documents. UL
Attention, consulter les documents
d’accompagnement.
Caution. To reduce the risk of electric UL
shock, do not remove cover or back.
Power ON UL
IEC
Marche (mise sous tension)
Power OFF UL
IEC
Arrêt (mise hors tension)
Equipotentiality IEC
UL
Equipotentiálite
133
Warning. Replace fuse as marked. UL
Customer Support
Sierra Customer Support
Contact numbers
In the U.S. Voice: 1-800-245-3001
134
Fax: 1-509-783-6503
Outside the U.S. Local distributor or 1-509-735-6481
Support hours
Application support 8 am to 5 pm PST (Pacific Standard Time)
Service department support 7 am to 5 pm PST
Shipping equipment
Servicing Sierra Systems
Custom keyboard
If the service department determines that the problem lies in the custom keyboard, you will be asked to ship the base of
the unit (or the entire unit for Sierra II Wedge) to Cadwell Laboratories for service or repair.
Refer to the manufacturer’s instructions for general maintenance of items not manufactured by Cadwell
Laboratories.
Sierra Console
If the service department determines that the problem lies with the custom keyboard or internal computer processor, you
may be asked to ship the entire base of your system to Cadwell Laboratories for service or repair.
135
Refer all service of internal system components to authorized Cadwell service personnel.
If the service department determines that the problem lies in one of the modules located on the right-side panel, you might
be asked to remove and ship the module to Cadwell Laboratories for service or repair.
Sierra II Wedge
If the service department determines that the problem lies at the component level or is in the custom keyboard or
computer, you may be asked to ship the base of your system or the laptop computer to Cadwell Laboratories for service
or repair.
____________________
See also
Shipping equipment
Exchanging and Returning Parts for Repair
Most problems can be diagnosed over the telephone. Those that cannot may require that you ship the defective part to
Cadwell Laboratories for service or repair.
Repairs are usually made by sending an exchange/repair part. Every effort is made to ship the part the same day the
problem is identified. Parts are sent by second-day air service. Overnight and Saturday services are available for an
additional charge.
If you are returning any item for repair, Cadwell issues a return merchandise authorization (RMA) number. Refer to this
number when requesting information about the repair.
____________________
See also
Shipping equipment
Shipping Equipment
Federal Express is the authorized shipping agent for returning equipment to Cadwell Laboratories.
Contact Cadwell if you have questions about packing and shipping equipment.
Customer support
To return equipment
1 Place the item in the original shipping box or other appropriate shipping container.
2 Write the Return Merchandise Authorization (RMA) number on the left corner of the box. The RMA number is on the
shipping paperwork you received with the new part, as well as on the Federal Express airbill.
3 Complete the return address information (section 1) on the airbill. Keep all copies together.
4 Call Federal Express to arrange for pickup.
5 Use the plastic pouch that the Federal Express driver gives you to enclose the airbill form and attach it to the sealed
box. Be sure to note the Federal Express airbill number for your records.
136
Cadwell International Distributors
North America
South America
Argentina Brazil
J.C. GUZMAN & CIA, S.A MGC. MACHADO & W. MACHADO IND. LTDA.
Div. Cadwell Equipment Rua Manoel da Nobrega, 595
Estados Unidos 3833 6º andar, cj. 63
1228 Buenos Aires Bairro Paraiso, Sao Paulo (SP)
ARGENTINA BRAZIL CEP 04001-083
Phone/Fax: 541-932-5586 Phone: 55-11-889-0904
E-mail: [email protected] Fax: 55-11-884-9864
E-mail: [email protected]
Chile Colombia
CADWELL CHILE A & B EQUIPOS MEDICOS
5 Oriente 2249, Penalolen Kra 7, No. 84-86, Apto 702
Santiago De CHILE Bogota, COLOMBIA
Nicosia, CYPRUS
Phone: 562-333-0750
Fax: 562-333-0751 Phone: 357-2-435-503
E-mail: [email protected] Fax: 357-2-435504
E-mail: [email protected]
Ecuador Venezuela
EDUARDO CASTRO TECNOMED J. TRAPP, C.A.
Hospital Metropolitano Peligro a Puente Republica
Consolteria No. 302 Edif. 142 Este 2. La Canderaria
Quito, ECUADOR Caracas, VENEZUELA
Fax: 593-226-9247 Phone: 582-575-2422
Fax: 582-576-2444
E-mail: [email protected]
137
Europe
Belgium Denmark
CIMA+ B.V.B.A. CLINILAB DENMARK
Leuvensesteenweg 282/B1 Engvej 13
B-3190 Boortmeerbeek DK-2625 Vallensbeak
BELGIUM Copenhagen, DENMARK
Phone: 32-15-528-408 Phone: 45-43-625100
Fax: 32-15-528-457 Fax: 45-43-625-323
E-mail: [email protected]
Finland Germany
CLINILAB OY SCHWARZER GMBH
Itatuulenkuja 1A Barmannstr. 38
02100 Espoo, FINLAND D-81245 Munchen
GERMANY
Phone: 3589-525-9260
Fax: 3589-525-92627 Phone: 49-89-839420
E-mail: [email protected] Fax: 49-89-83942186
E-mail: [email protected]
Greece Hungary
DAMPLAID NEUROTREND
3, Pontou St. Horvat u. 11-17
115 28 Athens 1027 Budapest
GREECE HUNGARY
Phone: 30-1-779-2193 Phone: 3630-446-175
Fax: 30-1-775-3627 Fax: 361-202-6674
E-mail: [email protected]
Italy Netherlands
S.P.M. SISTEMI E PRODOTTI FLEXCARE MEDICAL
MEDICALI Oude Kruisweg 25
Via A Borrelli, 3 2145 EE Cruquius
90139 Palermo, ITALY HOLLAND
Phone: 39-91-62-62629 Phone/Fax: 31-23-563-3015
Fax: 39-91-62-62661 E-mail: [email protected]
E-mail: [email protected]
Spain Sweden
NEUROLINE, S.L. CLINILAB SVERIGE
Antonio Palacios No. 20-8 P.O. Box 40
36400 Porrino S. 86521 Alno
Pontevedra, SPAIN SWEDEN
Phone:34-986-336697 Phone: 46-60-904529
Fax: 34-986-339154 Fax: 46-60-557043
E-mail: [email protected]
138
United Kingdom
NEUROLEC, LTD.
Unit 2 Kenway, Brook Street
SHEPSHED LE12 9RG
ENGLAND
Phone: 44-15-09-650890
Fax: 44-15-09-650508
Middle East
Cyprus Egypt
L.T.C. - X. LYSSIOTIS TRADING CO. BEHMAN
4 Pindarou Street 80, El Goumhouria St.
Ayios Antonios, CY-1060 Cairo, EGYPT
Nicosia, CYPRUS
Phone: 20-2-591-9340
Phone: 357-2-435-503 Fax: 20-2-588-2787
Fax: 357-2-435504 E-mail: [email protected]
E-mail: [email protected]
Israel Turkey
MIROM MEDICAL & RESEARCH DRAGER MEDICAL TRADE & SERVICE
EQUIPMENT LTD. CORP.
58 Moria Avenue, P.O. Box 6222 Eston Camlievler A7 Blok D:1
31061 Haifa 81120 Icerenekoy
ISRAEL Istanbul, TURKEY
Phone: 972-4-834-2281 Phone: 216-469-0555
Fax: 972-4-834-1025 Fax: 216-469-0560
139
Asia
China India
E.C. RYAN INTERNATIONAL BIOMED SYSTEMS
(U.S. OFFICE) 412 Hillview Industrial Estate
1921 Alicante Street off L.B.S. Marg., Ghatkopar (W)
Davis, CA 95616, U.S.A. Mumbai, INDIA
Phone: 530-765-4081 Phone: 91-22-517-0552
Fax: 530-756-4083 Fax: 91-22-517-0560
E-mail:
E.C. RYAN INTERNATIONAL [email protected]
(CHINA OFFICE)
BeijingXinQiao Hotel MEDCHEM TECH
Rm. 360, 2 Dong Jiaominxiang St. 60, Shantisadan Estate
Beijing, CHINA Opp. Dinbai Tower, Lai Darwaja
Ahmedabad 380 001
Phone: 86106-513-3366 INDIA
Fax: 86106-524-1965
E-mail: [email protected] Phone/Fax: 9179-550-6419
E-mail: [email protected]
Indonesia Japan
PT. PUTRIA PRATAMA HAYU CELCOM, INC.
Jalan Lapangan Roos Raya No. 26 2-13 Tokoji-Machi
Jakarta 12840 2-chome, Hakata-Ku
INDONESIA Fukuoka, 816, JAPAN
Phone: 62-21-830-8781 Phone: 81-92-471-8863
Fax: 62-21-830-5886 Fax: 81-92-472-8573
Korea Philippines
SAEHAN SALES CORP. FAIRBRIGHT ENTERPRISES
216-18 Nonhyun-Dong Lg 28-32 Alfaro Place Bldg
Kangnam-Ku 146 Alfaro Street, Salcedo Village
Seoul 135-010 1227 Makati, Metro Manila
KOREA THE PHILIPPINES
Phone: 822-234-0051 Phone: 632-815-9331
Fax: 822-234-3800 Fax: 632-892-7362
E-mail: [email protected] E-mail: [email protected]
Taiwan Thailand
MERIDIANS INSTRUMENTS CO., MAROONGROGE CO.
LTD. 219-221 Maitrichit Road
2nd Floor, No. 289, Sec. 4 Promprab, Bangkok 10100
Chen Tei Road, Shyh Lin THAILAND
111 Taipei, Taiwan
REPUBLIC OF CHINA Phone: 662-225-23257
Fax: 662-225-2328
Phone: 88622-880-2720 E-mail: [email protected]
140
Fax: 88622-883-6326
E-mail: [email protected]
Australia
ELECTROLAB PTY/LTD
Unit 10, 9 Foundry Road
Seven Hills NSW 2147
AUSTRALIA
Phone: 612-9620-8491
Fax: 612-9620-8492
E-mail: [email protected]
Clinical References
Clinical References
The references listed below provide clinical information on electromyography, nerve conduction, and evoked potential
testing. The list contains several publications commonly cited in electrodiagnostic testing literature.
• American Electroencephalographic Society: American Electroencephalographic Society Guidelines for Clinical
Evoked Potential Studies. American Electroencephalographic Society, 1992.
• Aminoff MJ: Electrodiagnosis in Clinical Neurology, ed 3. New York, Churchill Livingstone, 1992.
• Blackhouse KM, Hutchings RT: Color Atlas of Surface Anatomy: Clinical and Applied. Baltimore, Williams & Wilkins,
1986.
• Chiappa KH: Evoked Potentials in Clinical Medicine, ed 2. New York, Raven Press, 1990.
• Chu-Andrews J, Johnson RJ, Bruyninckx FL, et al.: Electrodiagnosis: An Anatomical and Clinical Approach.
Philadelphia, Lippincott, 1986.
• Curtis BA: Neurosciences: The Basics. Philadelphia, Lea & Febiger, 1990.
• deGroot J: Correlative Neuroanatomy, ed 21. Norwalk, Appleton & Lange, 1991.
• Delisa JA, Lee HJ, Baran EM, Lai K, Spielholz N: Manual of Nerve Conduction Velocity and Clinical Neurophysiology,
ed 3. New York, Raven Press, 1994.
• Delisa JA, Mackenzie K, Baran EM: Manual of Nerve Conduction Velocity and Somatosensory Evoked Potentials, ed
2. New York, Raven Press, 1987.
• Johnson EW: Practical Electromyography, ed 2. Baltimore, Williams & Wilkins, 1988.
• Kimura J: Electrodiagnosis in Diseases of Nerve and Muscle: Principles and Practice, ed 2. Philadelphia, F.A. Davis
Company, 1989.
• Liveson JA, Ma DM: Laboratory Reference for Clinical Neurophysiology. Philadelphia, F.A. Davis Company, 1992.
• Liveson JA: Peripheral Neurology: Case Studies in Electrodiagnosis, ed 2. Philadelphia, F.A. Davis Company, 1991.
141
• Ma DM, Liveson JA: Nerve Conduction Handbook. Philadelphia, F.A. Davis Company, 1983.
• McMinn RMH, Hutchings RT: Color Atlas of Human Anatomy: Human Anatomy. Chicago, Year Book Medical
Publishers, 1988.
• Misulis, KE: Spehlmann's Evoked Potential Primer, ed 2. Boston, Butterworth-Heinemann, 1994.
• Nuwer, MR: Evoked Potential Monitoring in the Operating Room. New York, Raven Press, 1986.
• Oh SJ: Clinical Electromyography: Nerve Conduction Studies, ed 2. Baltimore, Williams & Wilkins, 1993.
• Pansky B: Review of Gross Anatomy. New York, Macmillan Publishing Company, 1984.
• Sethi RK, Thompson LL: The Electromyographer’s Handbook. Boston, Little, Brown, 1989.
Welcome to Sierra II
Online Help
142
SEPU function keys in cursor mode:
1, 2, 3, 4 (knobs)
Use these four knobs to perform different functions. They correspond to labels at the bottom of the test display. Functions assigned to
the knobs are specific to each test, and all knobs might not be used in a particular test.
Functions change according to the mode of operation you are in: normal acquisition mode (when you first enter a protocol test), cursor
mode (when you press the Cursor key on the custom keyboard to work with cursors), or position mode (when you press the Posn key
on the custom keyboard to position waveforms).
SEPU knobs in normal acquisition mode:
TEST
Use the TEST keys to open other tests. All displays a list of all tests, from which you can select the one you want. You configure keys
1, 2, and 3 to open the particular test you want (see the configuration help topic for information on configuring key assignments).
Numeric keypad
Use the keys on the keypad as follows:
0 through 9 Enter numeric values on the screen
← Move the cursor to a previous field
• Insert a decimal
Report 1, Report 2
Press these keys to print reports. You can configure the type of report that prints for each key, or you can configure either key to display
a list of all reports, from which you can select the one you want (see the configuration help topic for information on configuring key
assignments).
Imp
Press this key to check the impedance of patient connections.
Patient
143
Press this key to display the Edit Patient dialog box.
View
Press this key to cycle through the windows associated with a test protocol.
Posn
Press this key in certain tests to change the settings for the F1, F2, F3, and F4 keys and 1, 2, 3, and 4 knobs to functions for positioning
waveforms.
Cursor
Press this key in certain tests to change the settings for the F1, F2, F3, and F4 keys and 1, 2, 3, and 4 knobs to functions for working
with latency and amplitude markers and auto cursors.
OK
Press this key as an alternative to clicking the OK button on the display.
Also use it to return you from cursor or position mode back to normal knob assignments.
Clear
Press this key to clear waveforms from the display. The averager automatically resets to zero. Press a second time to restore the
waveforms to the display.
Store
Press this key to store waveform traces from the active window in the database.
Run/Stop
Press this key to start and stop live data acquisition.
Single
In tests that use stimulators, press this key to deliver a single stimulus.
Volume
Turn this knob to adjust speaker loudness.
LPT 1
Connector for the peripheral printer.
COM 1
Connector for the serial mouse.
KEYBOARD (or keyboard symbol)
Five-position, circular DIN connector for a separate alphanumeric keyboard.
VGA VIDEO
Connector for the Sierra II Console VGA monitor.
On/Off (l / O)
Power switch.
Fuse drawer
110-volt systems use one 2-amp fuse. 220-volt systems use two 2-amp fuses.
144
Power rating label
Lists the instrument voltage, frequency, and current ratings.
Caution label
Danger-Explosion hazard. Do not use in the presence of flammable anesthetics. Replace fuse as marked.
Network
Network connector for optional equipment.
Caution label
Hazardous electrical output. This equipment is for use by qualified medical personnel. Grounding reliability can only be achieved when
the equipment is connected to an equivalent receptacle marked Hospital Only or Hospital Grade.
COM 2
Not used with Sierra software.
BAEP
Connector for the acoustically shielded headphones, lightweight headphones, insert earphones, auditory bone transducer, or ECochG
headset cables.
Electrical Stimulators
Electrically isolated connectors for the constant current electrical stimulator.
Goggles
Connector for the LED goggles, Ganzfeld stimulator, or single-intensity flash stimulator.
VEP/AUX
Connector for the 3-inch LED stimulator or monochrome VEP monitor.
Amplifier
Connector for the 2-channel and 4-channel amplifiers.
FOOTSWITCH
Connector for the footswitch.
AUXILIARY
Auxiliary connector for the 3-inch LED stimulator, flash stimulator, LED goggles, Ganzfeld stimulator, or Trigger In/Out.
10-BASE T
145
Connection between the custom base of the Sierra II Wedge and the PC. Communication between the two uses the Ethernet 10-Base
T connection.
ESTIM
Connector for the electrical stimulator.
TEMP
Connector for the temperature probe.
KEYBOARD
Connector for the keyboard.
COM 1
Connector for the serial mouse.
COM 2
Not used for the LPT printer.
PARALLEL
Connector for the LPT printer.
10-BASE T
Not used for the Sierra II Console. Instead, a board inside the unit is used to communicate with the custom keyboard.
Power cord connector
Connector for the power cord.
On/off switch
Power switch.
On
An LED that lights green when the unit is on.
EEG
Connector for the Easy II EEG amplifier. The Easy II and Sierra amplifiers can both be connected at the same time.
PORT 1
146
10-BASE T connector for the network card in the computer.
PORT 2
Extra network connector. Not used for normal operation.
AUX 1
Auxiliary connector for the 3-inch LED stimulator, flash stimulator, LED goggles, Ganzfeld stimulator, or Trigger In/Out.
AUX 2
Extra auxiliary connector for the 3-inch LED stimulator, flash stimulator, LED goggles, Ganzfeld stimulator, or Trigger
In/Out.
System Setup/General
Restoring defaults
Use the Test Menu tab to rearrange the order of tests in the Test Protocols menu and to delete tests that have been
created by users.
For information about the type of setup you want to do, click . . .
Rearranging the order of tests in the menu
Deleting tests
Use the Test Setup tab to move directly to the setup dialog box for a particular protocol. Simply select the protocol you
want and click Test Setup.
You can also access the setup dialog box for a protocol while you are in that protocol. However, when you change
settings for a test while in a protocol, the changes affect only that test.
147
When you access the test setup for a protocol from the System Setup dialog box, the changes become the defaults for
that protocol.
Use the test setup dialog box to configure settings for EMG tests.
For information about the type of setup you want to do, click . . .
Configuring report key assignments
Use the setup dialog box to configure settings for test protocols.
For information about the type of setup you want to do, click . . .
Configuring report key assignments
Sweep delay determines the start of a sweep relative to the onset of the stimulus. A prestimulus sweep delay (negative
value) is useful for acquiring pre-stimulus or baseline-level activity. A post-stimulus sweep delay (positive value) can be
used to remove stimulus artifact from the display.
Average Count
AvgCt (average count) is used in the protocols with waveform averaging. Use it to set the number of sweeps to be
collected in a single averaging run. Averaging stops when AvgCt has been reached. The value can be set from 1 to
10,000.
148
Averager On/Off
Turns the averager on and off.
Channel Gain
Controls the channels on which the gain control works.
Clear Waveforms
Clears active waveforms from the display. This automatically sets the averager (if being used) to zero. Click a second time
to restore the waveforms to the display.
Equivalent to the Clear key on the custom keyboard.
Clear Averager
Clears and restarts the count for the averager. You can clear the averager while it is running or stopped. If you clear the
average while it is running, the active waveform is cleared, and the averager resets to zero.
Function Keys
The F1, F2, F3, and F4 function keys each have different, preset functions assigned to them. Functions assigned to the
keys are specific to each test, and all keys might not be used in a particular test.
Functions change according to the mode of operation you are in: normal acquisition mode (when you first enter a protocol
test), cursor mode (when you press the Cursor key on the custom keyboard), or position mode (when you press the Posn
key on the custom keyboard to position waveforms).
SEPU function keys in normal acquisition mode:
You can activate a function assigned to F1, F2, F3, or F4 in three ways:
• Click the F1, F2, F3, or F4 button at the bottom of the test screen.
• Press the F1, F2, F3, or F4 key on the custom keyboard.
• Press the F1, F2, F3, or F4 key on the alphanumeric keyboard.
Single Stimulus
Delivers a single electrical stimulus.
Equivalent to the Single key on the custom keyboard.
Gain
Gain is the input sensitivity of the instrument when data is being collected and is measured in microvolts per division
(µV/div). This sensitivity determines the vertical scale of the input as displayed on the test screen when set to view the
input signal during acquisition. You can adjust gain settings (as well as sweep speed) to ensure proper display of the
signal. You can adjust the gain setting independently in each channel.
149
When you adjust the gain setting while collecting live data in the EP, EMG, NCV, F-wave, H-reflex, RNS, and Blink test
protocols, you are adjusting the amplifier gain and the display gain at the same time.
When you adjust the gain setting while averaging in the EP or NCV protocol, you are adjusting the display gain only. The
display gain can also be adjusted after averaging. (To adjust the amplifier gain, you must be in the live mode.)
Hicut Filter
Adjusts the hicut filter, which removes unwanted higher-frequency electrical interference (noise) or biological signals that
might interfere with the biological signals of interest. Frequencies are eliminated before they reach the computer. Values
are in hertz (Hz). The filter is 2 pole. You can adjust the filter independently in each channel. The setting affects only the
incoming live signal and has no effect on stored waveforms.
You can set this filter to any preset value in the range 30 to 15,000 Hz. You cannot attain hicut filter settings below 1000
Hz at all sweep speeds, as indicated by the table below.
Sweep Speed Minimum Hicut Frequency
0.5, 1.0 ms/div 1000 Hz
1 to 30 ms/div (1000/sweep speed) = min. Hz
≥ 50 ms/div 30 Hz
Impedance
Displays the impedance of patient connections (the opposition to current flow in the alternating current circuit).
Impedance levels below 5 kilohms (kΩ) are acceptable for most tests. The following colors are used to indicate the
impedance level for each electrode pair:
Green < 5 kΩ
Yellow 5 - 15 kΩ
Red > 15 kΩ
Equivalent to the Imp key on the custom keyboard.
Knobs
The four knobs on the custom keyboard each have different, preset functions assigned to them. Functions assigned to the
knobs are specific to each test, and all knobs might not be used in a particular test. You can identify the assignments by
the labeling at the bottom of the test screen.
Functions change according to the mode of operation you are in: normal acquisition mode (when you first enter a protocol
test), cursor mode (when you press the Cursor key on the custom keyboard), or position mode (when you press the Posn
key on the custom keyboard to position waveforms).
SEPU knobs in normal acquisition mode:
150
Locut Filter
Adjusts the locut filter, which removes unwanted lower-frequency electrical interference (noise) or biological signals that
might interfere with the biological signals of interest. Frequencies are eliminated before they reach the computer. Values
are in hertz (Hz). You can adjust the filter independently in each channel. The setting affects only the incoming live signal
and has no effect on stored waveforms.
A 2-pole locut filter is provided for all test protocols except NCV, where a 1-pole locut filter is used. You can set the filter to
any preset value in the range 0.04 to 500 Hz.
Markers
Turns latency (vertical) markers on and off.
Next Window
Cycles from one window to the next, making each one active in turn.
Notch Filter
Used only in the EMG protocol. The notch filter is always off in all other protocols. It reduces unwanted interference from
50- or 60-Hz power lines. Set it for the AC line frequency in your country.
When on, the notch filter affects both channels. You cannot set it selectively for individual channels..
Channels On/Off
Turns a channel on or off. Up to four channels can be used, depending on the amplifier configuration and type of test.
Pulse Width
Sets the pulse width of the electrical stimulus, measured in microseconds (µs).
Pulse width in the BAEP test protocol is 100 µs and cannot be modified. For electrical stimulation, the pulse width can be
set from 50 to 1000 µs.
151
Fmax = maximum frequency
Fo = nominal repetition rate frequency
RV = random variation from 0 to 100%
Raster/Overlay View
Switches between raster and overlay views of traces in the stored traces window.
View Description
Raster Stacks traces by channel.
Overlay Superimposes each channel’s traces.
Reject On/Off
Turns the reject function on and off. If turned on, the Sierra rejects sweeps that exceed the preset threshold value of 95%
of full amplifier scale.
The reject function is used in all EP protocols and is automatically enabled when you select an EP test.
Reject Percentage
Shows the percentage of total sweeps that have been rejected during a single run. Sweeps are rejected when they
exceed the preset threshold value of 95% of full amplifier scale.
Run/Stop
Stops or starts live data acquisition.
Equivalent to the Run/Stop key on the custom keyboard.
Secondary Toolbar
Opens or closes the secondary toolbar, which contains additional amplifier controls.
Recording Side
Sets the side of stimulation. This sets the appropriate labels for the trace/trial and data tables; or, in the EMG protocol, it
labels scored muscles with the appropriate side.
Smooth
Smoothing is used to reduce high frequencies from active or stored waveforms. It is available in all protocols except EMG
and RNS.
Single Stimulus
Delivers a single stimulus.
152
Equivalent to the Single key on the custom keyboard.
Stimulus Level
Sets the stimulus intensity level, in milliamperes from 0 to 100.
Current Stimulator
When more than one stimulator is being used, selects the one to be active.
Store Traces
Stores waveforms as traces in the Stored Traces window. The traces are also stored in the database for the patient.
Equivalent to the Store key on the custom keyboard.
Sweep Speed
Sets the sweep speed time base in milliseconds per division (ms/div). There are always 10 divisions across the display.
Therefore:
sweep speed x 10 = total timebase
You can set sweep speed only at one of the preset values. If you enter a value not included in the list, the closest preset
value is used.
Trigger Delay
Trigger delay is used in capture mode. Use it to move the trigger point of the trigger level threshold line. You can set the
trigger delay from 1 to 9 divisions. Move the trigger delay out to 8 divisions if you want to see more data before the trigger
point; move it to 1 division if you want to see more data after the trigger point.
The trigger delay shows as a small vertical line on the trigger level line at the location of the division mark.
Trigger Level
Trigger level is used in capture mode. It sets the voltage level at which trace data triggers a sweep in the Capture EMG
data acquisition window. When the rising/falling edge of the action potential exceeds the threshold, a data trigger is
generated to capture the sweep. The level can be set in the range ± 90% of the maximum negative deflection.
The trigger level displays as a horizontal dashed line in the live data acquisition window. Changing the level moves the
trigger line up and down on the graticule.
153
Trigger On/Off
In capture mode only, turns the capture trigger on or off.
IEC
International Electrotechnical Commission
UL
Underwriter's Laboratories
154
Disabling and Reinstating the Login Requirement
To disable the login requirement, you simply hide the login window. Anyone who then opens Sierra has the privileges of
the last user in effect.
You can change the order in which test protocols are listed in the Test Protocols menu.
The primary and secondary toolbars display all parameter settings. You can change the settings in three ways. Choose
the method most convenient for you.
• Change parameters directly in the toolbar using the mouse or computer keyboard
The following EMG toolbar shows the parameter settings for gain, highcut filter, locut filter, sweep speed, notch filter,
and scored muscle side. The way you enter the information depends on the kind of field; you might, for example,
select from a drop-down list or type the information from the computer keyboard.
155
• Change parameters using the knobs
Some parameters can be controlled with knobs. Check the lower part of the test display to see which ones. For
example, in the SEPU test protocol, you can use knobs to adjust the Intensity, Pulse W (pulse width), and Gain
parameters.
There are two methods for reaching the test setup dialog box for a test protocol:
• When a test protocol is open
On the Edit menu, click Test Setup. The setup dialog box for that test protocol appears. Changes you make affect
only the test you are conducting, unless you use the Save Parameters on the Test Protocols menu to save them, in
which case they become the default for that test protocol.
• When no test protocol is open
On the Edit menu, click System Test and then click the Test Setup tab. Select the protocol you want and then click
Test Setup. Changes you make with this method become the default for that test protocol.
The following table shows the parameters controlled in test setup boxes:
Test Protocol Parameters Other
EMG Trigger Delay Report keys
EMG2 Confirm “Storecommands
Scored muscles table definition
Muscle list
SFEMG Trigger Delay Report keys
SSFEMG Save trace with patient data
Muscle list
NCV Start Intensity Report keys
F-wave Sweep Delay Footswitch
H-Reflex Amp Cutoff Trace labels
Nerve list
Nerve/nerve group settings
Site table
Segment table
156
Show Norm Velocity value
SEPL Amp Cutoff Report keys
SEPU Display Mode Trace labels
Split Screen Auto-cursors
Sweep Delay
BAEP Click Display Mode Report keys
Pulse Width Trace labels
Split Screen Auto-cursors
Sweep
Sweep Delay
AEP Display Mode Report keys
VEP Split Screen Trace labels
Sweep Delay Auto-cursors
BLINK Amp Cutoff Report keys
Split Screen Tracel labels
Sweep Delay Footswitch
RNS Amp Cutoff Report keys
Sweep Delay Trace labels
Descriptions of Parameters
157
owerL
(area where checks display)
Check Sz (size of checks on video screen)
Miscellaneous Max (maximum count of samples in SFEMG/SSFEMG)
Rsp1 (first waveform for decrement measurement)
Rsp2 (second waveform for decrement measurement)
Split (split gain to the right of the split point)
SplitPoint (division on the screen for gain readings)
Spread (distance between samples)
StrtTrc (trace to start reviewing with in SFEMG/SSFEMG
Train (number of responses in train stimulation)
TrcCt (number of traces to review in SFEMG/SSFEMG)
Trig Lvl (trigger level)
Trig Del (trigger delay)
158
Parameter Fibs Default Recruit Default
Channel 1 On On
Channel 2 Off Off
Common reference Off Off
Gain 100 µV/division 1,000 µV/division
Sweep speed 10 ms/division 20 ms/division
Hicut filter 10,000 Hz 10,000 Hz
Locut filter 10 Hz 10 Hz
Notch filter Off Off
159
Hicut filter 10,000 Hz
Locut filter 1,000 Hz
Notch filter Off
160
Pulse width 100 µs 500 µs
Store Manual Manual
Sweep delay 0 div 0 div
Footswitch Store Store
Amp cutoff On On
Start Intensity 0 mA 0 mA
Parameter Default
Channels 1, 2 On
Common reference Off
Gain 0.5 µV/division
Hicut filter 3,000 Hz
Locut filter 100 Hz
Sweep speed 1 ms/division
Repetition rate 11.1 stimuli/s
161
Averages 2,000
Reject On
Pulse width 100 µs
Stimulator type Phones
Polarity Rarefaction
Side Right
Intensity (L/R) 80 dB nHL
Threshold (L/R) 0 dB
Masking On, 30 dB Diff
Sweep delay 0 divisions
Parameter Default
Channels 1, 2 On
Common reference Off
Gain 0.5 µV/division
Hicut filter 3,000 Hz
Locut filter 100 Hz
Sweep speed 1 ms/division
Repetition rate 11.1 stimuli/s
Averages 1,000
Reject On
Pulse width 100 µs
Stimulator type Phones
Stimulus type Click
Polarity Rarefaction
Side Right
Intensity (L/R) 80 dB nHL
Threshold (L/R) 0 dB
Masking On, 30 dB Diff
Sweep delay 0 divisions
162
VEP Parameter Defaults
Parameter Default
Channel 1 On
Common reference Off
Gain 5 µV/division
Hicut filter 100 Hz
Locut filter 1 Hz
Sweep speed 25 ms/division
Repetition rate 2.11 stimuli/s
Averages 100
Reject On
Side Right
Stimulator Video checks
Target On
Check size 32
Visual field Full
Sweep delay 0 divisions
Parameter Default
Channels 1 and 2 On
Gain 100 µV/division
Hicut filter 5,000 Hz
Locut filter 10 Hz
Sweep speed 10 ms/division
Repetition rate 1.0 stimuli/s
Side of stimulation Left
Elec 1 On/single stim
Pulse width 100 ms
Sweep delay 0 division
Footswitch Store
Amp cutoff Off
163
RNS Parameter Defaults
Parameter Default
Channel 1 On
Gain 2,000 µV/division
Hicut filter 10,000 Hz
Locut filter 10 Hz
Sweep speed 5 ms/division
Repetition rate 3 stimuli/s
Side Right
Elec 1 On/single stim
Pulse width 100 µs
Train 8
Spread 1 division
Sweep delay 0 divisions
Amp cutoff On
164
InterAmplitude Type of calculation for interamplitude:
Diff—To show the difference between two
amplitudes, calculated by subtracting the second
Amplitude cursor from the first.
Ratio—To show the ratio of two amplitudes,
calculated by dividing the first Amplitude cursor by
the second cursor.
You can also enter a normal (Norm) value for the
measurement and its left-right difference.
Latency Up to five cursors that you want to use for latency
calculations. For each cursor, you can enter a normal
value for the measurement and its left-right difference.
InterLatency Up to three sets of cursors that you want to use for
interpeak latency calculations. For each pair, you can
enter a normal value for the measurement and its left-
right difference.
Gn
Description
Gain is the input sensitivity of the instrument when data is being collected. This sensitivity determines the vertical scale of
the input as displayed on the test screen when set to view the input signal during acquisition. You can adjust gain settings
(as well as sweep speed) to ensure proper display of the signal. You can adjust the gain setting independently in each
channel. Gain is measured in microvolts per division (µV/div).
When you adjust the gain setting while collecting live data in the EP, EMG, NCV, F-wave, H-reflex, RNS, and Blink
protocols, you are adjusting the amplifier gain and the display gain at the same time.
165
When you adjust the gain setting while averaging in the EP or NCV protocol, you are adjusting the display gain only. The
display gain can also be adjusted after averaging. (To adjust the amplifier gain, you must be in the live mode.)
Hi, Lo
Description
The Sierra hicut (Hi) and locut (Lo) filters remove unwanted electrical interference (noise) or biological signals that might
interfere with the biological signals of interest. The frequencies are eliminated before they reach the computer. Values are
in hertz (Hz).
Each protocol comes configured with default filter settings. You can adjust the settings as needed, according to the
situation and type of signal you are measuring. You can set the filters independently in each channel.
The setting affects only the incoming live signal and has no effect on stored waveforms.
• Hicut filter
Rejects higher frequencies. You can set this filter to any preset value in the range 30 to 15,000 Hz. The filters are 2
pole.
Hicut filter settings below 1000 Hz are not attainable at all sweep speeds (see below).
Sweep Speed Minimum Hicut Frequency
0.5, 1.0 ms/div 1000 Hz
1 to 30 ms/div (1000/sweep speed) = minimum Hz
50 ms/div or greater 30 Hz
• Locut filter
Rejects lower frequencies. A 2-pole locut filter is provided for all tests except NCV, where a 1-pole locut filter is used.
You can set this filter to any preset value in the range 0.04 to 500 Hz.
Notch Off/50/60
Description
The notch filter reduces unwanted interference from 50- or 60-Hz power lines. In the EMG protocol, you can turn the filter
on or off, and select between 50-Hz or 60-Hz. The notch filter is always off in all other protocols.
When on, the notch filter affects both channels. You cannot set it selectively for individual channels.
166
• Default toolbar for the test protocol
• Amplifier function toolbar (on the View menu, click Amplifier Bar)
Swp
Description
Sweep speed sets the time base in milliseconds per division (ms/div). There are always 10 divisions across the display.
Therefore:
sweep speed x 10 = total timebase
You can set sweep speed only at one of the preset values. If you enter a value not included in the list, the closest preset
value is used.
Left/Right/Both
Description
Sets the side of stimulation. It is important to select this before delivering stimulation, as it sets the appropriate labels for
the trace/trial and data tables.
167
• Mouse cursor
SwpDel
Description
Sweep delay determines the start of the sweep relative to the onset of the stimulus. A pre-stimulus sweep delay (negative
value) is useful for acquiring pre-stimulus or baseline-level activity. A post-stimulus sweep delay (positive value) can be
used to remove stimulus artifact from the display. The value can be set to -1.0, 0.0, or 1.0 division.
Sweep duration is always 10 divisions long.
Trig Lvl
Description
Trigger level is used in capture mode. It sets the voltage level at which trace data triggers a sweep in the Capture EMG
data acquisition window. When the rising/falling edge of the action potential exceeds the threshold, a data trigger is
generated to capture the sweep. The level can be set in the range ± 90% of the maximum negative deflection.
The trigger level displays as a horizontal dashed line in the live data acquisition window. Changing the level moves the
trigger line up and down on the graticule.
Trig Del
Description
The trigger delay parameter is used in capture mode. Use it to move the trigger point of the trigger level threshold line.
You can set the trigger delay from 1 to 9 divisions. For example, move the Trig Del out to 8 divisions if you want to see
more data before the trigger point; move it to 1 division if you want to see more data after the trigger point.
The trigger delay shows as a small vertical line on the trigger level line at the location of the division mark.
Rej%
Description
The artifact reject function is used to remove signals (excessive noise or movement artifact) from the running average
when they exceed the preset maximum (95% of full amplifier scale). The Rej% (reject percentage) parameter shows the
percentage of total sweeps that have been rejected during a single run.
You can turn the reject function on and off with the artifact reject control on the toolbar.
168
• Default toolbar for the test protocol
• Averager function toolbar (on the View menu, click Averager Bar)
IntenL, IntenR
Description
Intensity of auditory stimulation on the left and right sides, measured in units of dB nHL (decibel normalized hearing level).
The maximum is 100dB nHL.
ThreshL, ThreshR
Description
The ThreshL (threshold left) and ThreshR (threshold right) parameters are used to set patient behavioral threshold for
stimulation. Values are in dB SL (decibel sensation level), where sensation level is the softest sound that the patient can
hear. The total intensity output to the stimulator is the intensity value plus the threshold value for that side.
Mask
Description
Adjusts the differential mask intensity opposite the side of stimulation in the BAEP test protocol. Masking shields the ear
opposite the stimulated one from receiving sound waves conducted through the head. The masking intensity is the mask
value subtracted from the stimulation intensity value. For example, if Mask is 40 and the click intensity is 70, the masking
intensity is 30 (70 minus 40).
Cond/Rare/Alt
Description
Sets the polarity for auditory click stimulation.
• Cond
Condensation clicks produce an initial movement of air toward the eardrum.
• Rare
Rarefaction clicks produce an initial movement of air away from the eardrum.
• Alt
Alternating condensation and rarefaction clicks.
169
• Default toolbar for the test protocol
• Audio Stim function toolbar (on the View menu, click Audio Stim Bar)
Phones/Inserts
Description
Type of auditory stimulation:
• Phones
Head phone stimulator (normally used).
• Inserts
Insert ear phones with a 1-millisecond acoustic delay line. When set to this stimulator type, 1 millisecond is
automatically subtracted from all cursor or marker latency measurements. This setting has no effect on the start of
sweep and is used only to compensate for the latency measurements. The sweep delay is set independently and is
properly accounted for in all latency measures.
PW
Description
Pulse width determines the duration of the electrical stimulus. It is measured in microseconds (µs). Pulse width in the
BAEP test is 100 µs and cannot be modified. For electrical stimulation, the pulse width can be set from 50 to 1000 µs.
mA
Description
Level of stimulation, in milliamperes from 0 to 100.
RR
Description
170
Repetition rate, expressed as stimuli per second, determines the rate of stimulus presentation. Stimulus repetition rate
can be set in the range 0.01 to 90 Hz.
Multiple stimuli must be presented to obtain a final averaged waveform. Under most circumstances, these stimuli are
presented at a fixed rate with a constant time interval between stimuli. Note also that the rate at which stimuli are
presented affects the total time required to obtain a single averaged response.
Avg
Description
A counter that shows the number of sweeps that have been collected during waveform averaging.
AvgCt
Description
Sets the number of sweeps to be collected in a single run for averaging. Averaging stops when the average count has
been reached. The value can be set from 1 to 10,000.
Rnd%
Description
The random repetition rate parameter is used when random variation is enabled. You can set a random variation in the
range 0 (no variation) to 100%. Use random variation to prevent unwanted synchronization with noise sources.
The random variation affects the repetition rate (RR) by setting a range of frequencies over which the repetition rate will
vary. The frequencies vary randomly within a range. The variation approximates a uniform distribution. The following
formulas specify the frequency range:
Fmax = Fo * 100/(100 – RV/2)
Fmin = Fmin * 100/(100 + RV/2)
where
Fmax = minimum frequency
Fmin = maximum frequency
Fo = nominal repetition rate frequency
RV = random variation from 0 to 100%
171
• Default toolbar for the test protocol
• Averager function toolbar (on the View menu, click Averager Bar)
Goggles/Checks/Ganzfeld/Led/Flash
Description
Used in the VEP test protocol to set the type of visual stimulation.
Full/UpperH/RightH/LowerH/LeftH/UpperL/UpperR/LowerR/LowerL
Description
Used in the VEP test protocol when Checks is the type of visual stimulation. It determines the area where checks will be
displayed on the checkerboard monitor.
Check Sz
Description
Used in the VEP test protocol when Checks is the type of visual stimulation. The size sets the number of checks
displayed on the video screen measured along the horizontal axis.
Rsp1, Rsp2
Description
Used in the RNS protocol to set the waveforms to be used for decrement measurement.
Spread
Description
Trace spread, used in the RNS protocol, is the horizontal distance (in divisions) between responses. Use this parameter
to move responses closer together or further apart.
172
• Default toolbar for the RNS test protocol
• Miscellaneous function toolbar (on the View menu, click Miscellaneous Bar)
Train
Description
Sets the number of traces to be contained in a train of stimuli. The train is presented each time you press the footswitch or
press the Run/Stop key on the custom keyboard. The value can be set from 1 to 150.
Max
Description
In the SFEMG and SSFEMG test protocols, sets the maximum number of samples you want to collect in a single run. The
value can be set at 10, 20, 50, or 100 (the maximum).
StrtTrc
Description
In the SFEMG and SSFEMG test protocols, sets the specific number of the trace that you want to start reviewing.
TrcCt
Description
In the SFEMG and SSFEMG test protocols, sets the number of traces that you want to review at one time. This lets you
divide the traces into smaller groups for easier review.
173
Review
To open a patient test file for review or additional testing. In the Open Patient File dialog box, select the file and click
Open.
Explanation of searches
Open from File
To open a patient test file that has previously been saved using the Export command. In the Open Record dialog box,
locate the file and click Open.
Entering the the information on this tab as a part of the patient file is a convenience for you. This information is
included automatically in test reports for the patient, so you do not have to enter the information each time you
create a report. You can change the information if you need to.
5 Click the Custom Test Info tab to enter custom information about the (patient history, medications, findings, and
conclusions).
Explanation of the Custom Test Info tab
The note just above applies to the Custom Test Info tab also.
6 Click Start Test to save the patient file and then begin a test.
- Or -
Click Save Only to save the file without beginning a test.
174
Deleting a Patient File
To delete a patient file
1 Open Sierra.
Opening and closing Sierra
2 On the Edit menu, click Patient List.
3 In the Patient List dialog box, select the patient and then click Delete.
Explanation of the Patient List dialog box
4 Click Yes to confirm the deletion.
5 In the Patient List dialog box, click Done.
Parts of a Window
The entire surface of the screen is called the desktop. It is made up of three basic parts: menu bar, windows, and status
bar.
• Menu bar (Click for illustration)
The bar that extends across the top of the desktop. The items in the menu bar are names of menus. When you click a
menu name, a list of commands appears.
• Windows (Click for illustration)
The individual rectangular areas on the desktop. Every window has a title bar with the name of the window. Each
window is self-contained and serves a different purpose. You can have several windows open on the desktop at the
same time. The actions you take in one window do not affect other windows. You can maximize (enlarge) a window
so that it fills the entire desktop and then minimize (restore) it to its former size and position. You can move windows
around on the desktop and change their size and shape.
You work in only one window at a time, the active window. You make the window active by clicking the title bar or
anywhere in the window. You know when a window is the active one because the title bar appears highlighted (darker
than the title bars of the other windows).
Sometimes there is more information than can fit in a window (either horizontally or vertically). When this is the case,
scroll bars let you the move the view right and left or up and down to see the hidden information.
• Status bar (Click for illustration)
A bar at the bottom of the desktop that contains status messages.
____________________
See also
Sierra menus
Moving a window
175
Maximizing and Minimizing a Window
You can maximize (enlarge) a window so that it fills the entire desktop and then minimize (restore) it to its former size and
position. You use two different buttons on the window title bar to maximize and minimize.
To maximize (enlarge) a window
Click the maximize button on the window title bar (the middle button in the following illustration).
Moving a Window
To move a window
1 Point the cursor to the title bar of the window you want to move.
2 Hold the left mouse button down and use the mouse to drag the window to the position you want.
3 Release the mouse button.
When there is more information than can fit in a window (either horizontally or vertically), a scroll bar lets you move the
view right and left or up and down to see the hidden information. For example, the Muscle List window from the EMG
protocol contains a vertical scroll bar because the window cannot contain the full list of muscles.
176
To use a scroll bar
There are several ways to use the scroll bar:
• Click the cursor on the position in the scroll bar that corresponds to the general location of the information you want
(beginning, middle, or end).
• Click the up or down arrow (in the vertical scroll bar) or the left or right arrow (in the horizontal scroll bar) to move one
line at a time.
• Point to the up or down arrow (in the vertical scroll bar) or the left or right arrow (in the horizontal scroll bar) and hold
the cursor down to move continuously in that direction.
• Point to the slider box, hold the left mouse button down, and slide the box in the direction you want.
Traces (individual collected waveforms of a defined duration) are displayed on graticule—a two-dimensional graph with
time on the horizontal axis and voltage on the vertical axis. The trace deflects up or down according to the relative
polarities of the preamplifier inputs.
The width of the waveform display represents the maximum latency that can be assigned to a pictured waveform. The
sweep is all the data collected after the delivery of any single stimulus and presented on the display.
Following are common characteristics of the data windows for test protocols:
• The stimulator, when used, triggers the sweep of the visual display. You control the sweep speed. You can also delay
the sweep, which permits full visualization of the pre-stimulus activity.
• When you want to save a trace (or trial), you use the Store control. This saves the trace to the patient file. In some
protocols (EMG, for example), the traces are moved into a stored traces window. In some protocols, you can
configure the footswitch to store data.
• In some test protocols, waveforms can be averaged. The waveform that appears in the live window represents the
average of the total number of responses collected. You can set the number of responses collected for the average.
• In some test protocols, cursors can be applied to waveforms and then repositioned as needed. Cursors mark
important features on the waveforms, such as amplitude and latency measurements. You can configure the cursors to
mark the features important to you.
• In some test protocols, trace labels are used to provide information about the traces labeled. You can configure the
information reported by the label (such as channel number, gain setting, and side of stimulation).
177
• In some test protocols, amplitude and latency markers are used. Latency markers indicate the time from the onset of
the trace. Amplitude markers indicate the difference in voltage between the two markers. You can show or hide the
markers, and reposition them as needed.
• In the EP protocols, you can control how the data window is split (full, vertical split, horizontal split, or quadrulple split).
Traces in the window can then be sorted by side or channel.
For procedures, click . . .
Working with traces
Averaging waveforms
178
Once you have placed a function toolbar on the screen, you can move it to any position on the screen. To do this,
move the mouse pointer to the toolbar title and then drag the toolbar to the position you want.
If you moved the toolbar to the top of the screen, the title bar does not show. Move the function toolbar down so
that the title bar reappears.
You can choose any of the following function toolbars to display on a test screen. Click for illustrations and descriptions.
179
Amplifier Bar
Acquisition Bar
Averager Bar
Markers Bar
Standard Bar
Elec Stim Bar
Audio Stim Bar
Visual Stim Bar
Trace Functions Bar
Miscellaneous Bar
180
override the values above.
Sets the gain.
181
each stimulation.
182
Turns latency markers on and off. The three numbers show
the values for latency marker 1 (T1), latency marker 2 (T2),
and the difference between them (T1-T2).
Working with latency and amplitude markers
Turns amplitude markers on and off. The number shows the
difference between the two amplitude markers.
Opens the Add Patient dialog box so that you can create a file
for a new patient.
Opens the Open Patient File dialog box so that you can open
the file for an existing patient from the database.
Lets you open a patient file located on your system.
Opens the Select Report Template dialog box so that you can
select a report template and generate a report of your
choosing.
Displays information about the SierraWin program.
183
different waveforms generated by low and
maximal efforts.
EMG data
Turns the capture trigger on and off.
Setting capture trigger parameters
Sets the trigger level.
Setting capture trigger parameters
Sets the trigger delay.
Setting capture trigger parameters
Changes the gain to the right of the split point in
the Stored Traces window.
Sets the split point in the Stored Traces window
for the F-wave and H-reflex test protocols.
Sets the waveform to be used for the first sample
in the RNS test protocol.
Sets the waveform to be used for the second
sample in the RNS test protocol.
Sets single or train stimulation.
184
Audio Stim Function Toolbar
185
Selects the side of stimulation; or, for muscle
scoring, labels the muscle side (Right, Left, Both).
Sets the stimulator type (Goggles, Checks,
Ganzfeld, Led, Flash).
Sets the visual field type (Full, UpperH,
RightH, LeftH, UpperL, UpperR, LowerR,
LowerL).
Sets the visual stimulator check size.
Sorts traces.
186
NCV/F/H TabData: Motor & Sensory
187
NCV/F/H TabData: Left Median-Motor
188
Desktop with Several Windows
This view of the desktop (from the EMG protocol) contains four windows: Live EMG Data, Muscle List, Scored
Muscles, and Stored Traces. The title bar for the Live EMG Data window is highlighted (darker than the title bars for the
other windows), which means that it is the active window.
189
Toolbars
Following is an example of the primary toolbar for the EMG protocol.
190
Knob Labels
Following is an example of custom keyboard knob labels at the bottom of the test screen.
Protocol Window
Following is an illustration of a window designed to display live data for the EMG protocol.
Menu bar
Following is an illustration of the Sierra menu bar.
Following is an example of the list of commands that appears when you click a menu name.
191
Status bar
Following is an illustration of the Windows status bar at the bottom of the screen that shows a system message.
Function Toolbar
Following is an illustration of the function toolbar for trace functions.
Secondary Toolbar
Following is an illustration of the secondary toolbar for the EMG test protocol.
• Make sure the monitor is turned on (Sierra II Console). The power light illuminates.
192
• Make sure the laptop is turned on (Sierra II Wedge).
• Check the monitor power cord connections (Sierra II Console).
• Check the monitor interface cable connections (Sierra II Console).
• Adjust the brightness on the monitor.
Press Ctrl + Alt + Del on the alphanumeric keyboard to open the Close Program dialog box. Click End Task to end the
highlighted task that is locked up. You might need to shut Windows down and reopen Sierra.
No electrical Actions:
stimulus delivered
• Make sure the electrical stimulator is activated
(stimulus intensity level appears in the toolbar).
• Check the stimulus intensity level, adjusting it as
necessary.
• Check the stimulator cable connections.
• Close and then reopen the test.
The electrical stimulator cannot deliver the current that
has been specified. This happens when:
• The impedance level between the stimulator and
the stimulation site is too high.
• There is a problem in the connection to the
patient.
Action:
• Check the patient connection. Make sure adequate
conductive gel is being used and that each tip is
making contact.
Unexpected The system detects that current was delivered even
output message though it was not supposed to have been generated.
Actions: Immediately disconnect the patient and
contact Cadwell.
Customer support
Over current An over current message appears on the display when
193
message the system detects more current being delivered than
has been specified.
Actions: Immediately disconnect the patient and
contact Cadwell.
Customer support
Other fault If a fault condition occurs, power to the electrical
condition stimulator is shut off so that energy output stops
messages immediately without operator intervention. A message
describing the condition appears on the display.
Action: Press the OK key to exit out of the fault
detection window. Immediately disconnect the patient
and contact Cadwell.
Customer support
____________________
See also
Stimulator diagnostics
Stimulator Diagnostics
An electrical stimulator diagnostic occurs automatically whenever a test protocol is selected that uses the electrical
stimulator.
When the electrical stimulator is operating, diagnostic checks run continuously to verify the general integrity of the
electrical stimulator and in particular the current output. The checks verify that the repetition rate and pulse width do not
exceed the requested values by more than 10%. The actual current output is sensed to verify that it does not exceed the
requested output current.
____________________
See also
Electrical stimulation problems
• Make sure the stimulator is turned on (monochrome VEP monitor, Ganzfeld stimulator).
• Check the stimulator power cord connections (monochrome VEP monitor, Ganzfeld stimulator).
194
• Check the stimulator interface cable connections.
• Close and then reopen the test.
• Adjust the brightness and contrast on the monochrome VEP monitor.
195
Custom or Alphanumeric Keyboard Is Not Functioning
Custom keyboard:
• Make sure no keys are stuck in the depressed position.
Alphanumeric keyboard:
• Check the keyboard’s cable connections.
Printer Problems
Preventative Maintenance
When equipment is delivered at customer’s expense during the term of the warranty to Cadwell Laboratories, Cadwell will
take the action necessary to insure performance to specifications, including cleaning as necessary, checking, calibrating,
and verifying operation. Preventative maintenance must be scheduled with Cadwell.
Warranty Exclusion
(a) Except as expressly set forth herein, there is no warranty, expressed or implied as to fitness for any particular
purpose, nor to anything else unless contained on the face of this Agreement. All other warranties, expressed or
196
implied, including, but not limited to, any IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR ANY
PARTICULAR USE, ARE HEREBY DISCLAIMED.
(b) Cadwell Laboratories shall be under no obligation to furnish, repair or replace any part which has been damaged,
abused or misused through no fault of Cadwell Laboratories, or if unauthorized attempts to modify, repair or service
equipment have impaired the performance of the equipment.
Representatives of Cadwell are not authorized to make warranties about the merchandise described in this Agreement.
These representatives’ ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES, shall not be relied upon by the
customer and are not part of the contract. The entire contract is embodied in this writing and NO OTHER WARRANTIES
are given beyond those set forth in this contract. This writing constitutes the final expression of the parties’ agreement,
and it is a complete and exclusive statement of the terms of that agreement.
Sole Purpose
The sole purpose of the exclusive remedy shall be to provide the customer with free repair and/or replacement of
defective parts in the manner provided herein. This exclusive remedy shall not be deemed to have failed of its essential
purpose so long as the company is willing to repair or replace defective parts in the prescribed manner.
Indemnification
Customer agrees to indemnify Cadwell and hold it harmless from any and all demands or claims of any nature whatsoever
arising out of the use of the equipment, including product liability or physical injuries to person or property, loss of life, or
other claims or damages.
Business Use
Except as expressly provided by Cadwell in writing, all goods sold hereunder are intended for the ultimate use by a
commercial user and for operation by persons trained and experienced in the use and maintenance of such equipment.
Cadwell’s warranty does not extend to any consumer. Customer acknowledges that it is purchasing the equipment for a
business use and for no other.
The parties agree that the customer’s SOLE AND EXCLUSIVE REMEDY against Cadwell shall be for repair and/or
replacement of defective parts as provided herein. Customer agrees that NO OTHER REMEDY SHALL BE AVAILABLE
to them and that the company shall not, in any event, be liable for incidental damages or consequential damages,
including loss of income, loss of time, lost sales, injury to person or personal property, liability customer had with respect
to any other person, or for any other type of formal consequential damage or economic loss.
Severability
If any provision or clause of this Contract or the application thereof to any person or circumstance is held invalid or
unconscionable, such invalidity or unconscionability shall not affect other provisions or applications of the Contract which
can be given without the invalid or unconscionable provision or application, and to this end the provisions of this Contract
are declared to be severable.
197
Cost of Legal Action
In the event any action shall be instituted for the failure to perform the terms and conditions of this Agreement, the
prevailing party shall be entitled to recover costs, including attorney’s fees.
An action by the customer for breach of this Agreement or any other action by customer otherwise arising out of this
transaction must be commenced within one year from the date the right, claim, demand or cause of action shall first
accrue, or be barred forever.
No Waiver
If Cadwell at its option agrees to a waiver of any of the terms and conditions recited herein, such waiver shall not for any
purpose be construed as a waiver of any succeeding breach of the same or any other terms or conditions of this
Agreement; nor shall such a waiver be viewed as a course of performance.
CUSTOMER AGREES TO BE BOUND BY ALL OF THE ABOVE TERMS AND FURTHER AGREES THAT THEY
CONSTITUTE THE COMPLETE AND EXCLUSIVE STATEMENT OF THE AGREEMENT BETWEEN THEM WHICH
SUPERSEDES ALL PROPOSALS, ORAL OR WRITTEN AND ALL OTHER COMMUNICATIONS BETWEEN THEM
RELATING TO THE SYSTEM AND WARRANTY THEREON.
THIS AGREEMENT WILL BE GOVERNED BY THE STATE OF WASHINGTON AND SHALL NOT BE MODIFIED,
AMENDED OR OTHERWISE AFFECTED EXCEPT BY A WRITING SIGNED BY BOTH PARTIES.
Heading Description
Side Side of recording. Default choices: R, L, L/R.
Muscle Name of the muscle.
Nerve Name of the innervating nerve.
Root Nerve root level(s).
Ins Act Insertional activity. Default choices: Nml (normal), Inc
(increased).
Fibs/Psw Fibrillation potential and positive sharp wave. Default
choices: Nml, 1+, 2+, 3+, 4+.
Amp Motor unit action potential (MUAP) amplitude. Default
choices: Nml, ++, - -.
Dur MUAP duration. Default choices: Nml, >12.
Phases Number of MUAP phases. Default choices: 2, 3, 4, 5, >5.
Recrt MUAP recruitment. Default choices: Nml, Early,
Reduced.
Int Pat Interference pattern. Default choices: Nml, 75%, 50%,
198
25%.
Comment Muscle scoring comments. Default field size: 25
characters.
____________________
See also
onfiguring the scored muscles table
Site 1, Site 2 Used to specify the sites at the end points of the nerve
segment. The velocity will be calculated for this
segment. A site can be specified by either the site
name or reference site. The reference site can be
entered only in the Site 2 input field.
Delta-O This is a calculated field and is displayed anytime
latency can be determined for the trace or traces
specified in the Site 1 and Site 2 columns. The field
will be blank if cursors have not been placed on the
traces. Delta is computed as
Delta = abs(L1 – L2)
where L1 is the latency of trace 1 and L2 is the
latency of trace 2. If R is specified for the second end
point, L2 is set to 0.0. The latencies are derived from
either the onset (O) or peak (P) cursors (whichever
you have configured to show).
Dist The distance (in centimeters) from the stimulus
cathode to the recording active electrode. You must
type this value in.
Vel Nerve conduction velocity (in meters per second),
automatically calculated anytime the Delta-O and Dist
fields contain values.
Norm Vel The normal conduction velocity for the nerve segment.
The value appears only if you have configured it to
display.
____________________
See also
Configuring nerve/group settings
Supramaximal stimulus
Response on the test screen increases in size as the stimulus strength is raised until it remains stable. All the fibers that
respond to the nerve stimulation will have done so.
199
Nerve conduction velocities
Nerve conduction velocities generally refer to the recording and measurement of a compound nerve action potential. They
are calculated by dividing the length of a particular nerve segment by the amount of time it takes the generated action
potentials to travel from the stimulation to recording sites, referred to as the latency.
Tibial Motor
200
It is convenient and efficient to place the ground on the top of the foot, and leave it there for all lower motor and sensory
studies. Different authors, however, suggest several possible locations for the ground on various studies.
Median F-wave
201
Active Motor point of abductor pollicis brevis
(APB)
Reference Distal phalanx of thumb
Ground Dorsum of wrist or hand
Ulnar F-wave
202
Active Motor point of abductor digiti quinti (ADQ)
Reference Fifth digit
Ground Dorsum of the wrist or hand
Median Motor
203
Active Motor point of abductor pollicis brevis
(APB)
Reference Distal phalanx of thumb
Ground Dorsum of wrist or hand
204
Active Proximal at base of second or third digit
Reference 4 cm distal to active
Ground Dorsum of wrist or hand
Tibial F-wave
205
Active Motor point of abductor hallucis
Reference Medial big toe
Ground Top of foot
Peroneal Motor
206
Active Motor point of EDB
Reference Lateral small toe
Ground Top of foot
Sural Sensory
207
Active Over the sural nerve posterior and inferior
to the lateral malleolus
Reference 3 cm distal
Ground Ankle
Ulnar Motor
208
Active Motor point of abductor digiti quinti (ADQ)
Reference Fifth digit
Ground Dorsum of the wrist or hand
You can change the reference site, sites, and segments for a nerve while you are conducting a test. The changes apply
only to the test you are conducting and do not affect the configured settings.
209
Enter information for the new site in the bottom room (marked with the *).
Click the empty cells in the Site 1 and Site 1 columns and use the drop-down boxes that appear to select the entries you
want to add to the segment table. The choices reflect any changes you made to the site table.
You can adjust the onset latency of a motor nerve or the peak latency of a sensory nerve (for example, if the auto cursors
did not pick accurately). You can do this either while the trace is still in the active grid or after the trace has been stored.
If you use the cursor key method, press OK on the custom keyboard when you have finished moving the cursors.
This will reset the knb and function key labels to their normal acquisition labels.
210
Electrode Placement of the Tibial Nerve for the H-reflex Test
211
Electrode Placement of the Median Nerve for the F-wave Test
Trial Each row in the RNS table is labeled with a sequential trial
number.
Amp1 These two columns show the amplitude of response 1
Amp2 (Rsp1) and the amplitude of response 2 (Rsp2). The peak-
to-trough amplitude is used. Measurements are in millivolts
(mV).
Amp % The amplitude ratio between the two responses. A negative
sign (-) indicates a decrement and a positive sign (+)
indicates an increment.
Area1 Absolute area of the two responses, computed by summing
Area2 the absolute value of the difference between the baseline
and the trace. The interval is from the onset of the response
to the completion of the negative phase. The trace value at
the onset is used as the baseline. Measurements are in
millivolts millisecond (mV ms).
Area % Area ratio between the two responses. Negative values
constitute decrement and positive values constitute
212
increment.
Other commonly evaluated muscles and their corresponding nerves/stimulation sites are listed below.
Muscle Nerve
Nasalis Facial
Deltoid Erb’s point (upper trunk brachial
plexus)
Biceps brachii Musculocutaneous
Vastus medialis Femoral
Trapezius Spinal accessory
Abductor digiti Ulnar
quanti (ADQ)
Abductor pollicis Median
brevis
213
Evoked Potential Montages Used
Evoked potential montages used reflect the modified combinatorial nomenclature. This system is a slight modification of
the international 10-20 system that designates the 10% positions which are currently unnamed.
214
International 10-20 system
215
Pg Nasopharyngeal
A Auricular
Fp Frontal Polar
F Frontal
C Central
T Temporal
P Pareital
O Occipital
216
Right Posterior Tibial SEP
Additional recording channels and their corresponding potentials for lower extremity SEP testing are listed below.
ACT REF Potential
CPZ FPZ P37 – primary cortical potential
T12S IC LP – stationary lumbar potential
CPi FPZ P37 – primary cortical potential
FPZ C5S P31, N34 – subcortical far-field potentials
217
Right Median SEP
Additional recording channels and their corresponding potentials for upper extremity SEP testing are listed below.
ACT REF Potential
CPc CPi N20 – cortical potential
EPi EPc EP – brachial plexus potential
C5S EPc N13 – cervical potential
CPi EPc P14, N18 – subcortical far-field potentials
218
BAEP Two-channel Protocol
219
Check and Field Sizes for the 3-inch LED Stimulator
220
Electrode Placement for Single-channel Visual Evoked Potential
Additional recording channels and their corresponding potentials for VEP testing are listed below.
ACT REF Potential
MO MF N75, P100, N145
LO MF N75, P100, N145
RO MF N75, P100, N145
MF A1 N100
Sierra Menus
The Sierra system features six menus across the top of the screen.
File menu
Edit menu
View menu
221
Window menu
Help menu
New Patient
Opens the dialog box for adding a new patient file.
Select Patient
Opens the Start New Test dialog box, which lists existing patients. Selecting a patient from the list lets you start a new test
using patient information that is already in the system.
Review
Opens the Open Patient File dialog box, which lists all tests conducted for all patients. Selecting a test from the list opens
the TabData dialog box, which displays information about that test.
Export File
Opens the Select Export Destination dialog box, which allows you to save a patient test in the folder of your choice in the
computer. Files are saved with the extension .qmc (QuickMed compressed). The files that you save in this way do not
become part of the QuickMed database.
When a patient file is exported, it is automatically zipped (compressed) to save disk space. When opened again in Sierra,
the file is automatically unzipped.
Close
Saves all data in the current test protocol and closes the patient file. You remain in the Sierra program.
Print Report 1
Prints a copy of the report that has been configured for the Report 1 key on the custom keyboard. You can configure
which report prints when you use this command.
Print Report 2
Prints a copy of the report that has been configured for the Report 2 key on the custom keyboard. You can configure
which report prints when you use this command.
222
Exit
Saves all data in the current test protocol, closes the patient file, and exits from the Sierra application.
Copy
Copies information from the active window to the Clipboard. You can then open another application (such as Microsoft’s
WordPad, Notepad, or Word program) and paste the information from the clipboard into the application.
User/Phys/Tech List
Opens the list of users with authorization to use Sierra so that you can edit, add, or delete an existing user profile
(physician, referring physician, technician, or other user).
Patient List
Opens the Select Patient dialog box that lists all patients. You can add a patient file, or select an existing patient and edit
or delete the patient’s file.
Current Patient
Selecting this while conducting a test for a patient opens the Edit Patient dialog box and lets you edit the file for that
patient.
Test Setup...
Opens the setup dialog box for the test protocol you are currently using. You use setup dialog boxes to configure such
things as report key assignments, trace labels, and parameter defaults.
System Setup...
Opens the System Setup dialog box, which is used for system-wide configuration. You use the System Setup dialog box,
for example, to configure the test keys on the custom keyboard and to enable and disable the login requirement.
Quickmed Utilities
Opens the Quickmed Utilities dialog box, which provides tools for working with patient charts, test data records, stored
reports, user profiles, and report templates.
TabData...
Opens a window that lists all the tests for a patient. Click a test and the test data appears.
Amplifier Bar
Opens and closes the function toolbar that contains controls for the amplifier.
223
Acquisition Bar
Opens and closes the function toolbar that contains controls for data acquisition.
Averager Bar
Opens and closes the function toolbar that contains controls for averaging waveforms.
Markers Bar
Opens and closes the function toolbar for the latency and amplitude markers.
Standard Bar
Opens and closes the function toolbar that contains standard controls for opening, closing, saving, and printing files, and
for obtaining help.
224
Elec Stim Bar
Opens and closes the function toolbar that contains controls for electrical stimulation.
Miscellaneous Bar
Opens and closes the function toolbar that contains miscellaneous controls.
225
Position
In certain tests, changes the settings for the F1, F2, F3, and F4 keys and 1, 2, 3, and 4 knobs to functions for positioning
waveforms.
The command is particularly useful if you are using the SierraWin software without the custom base attached (that is, as a
review station), because you can enter and exit Position mode using the software alone.
Cursors
In certain tests, changes the settings for the F1, F2, F3, and F4 keys and 1, 2, 3, and 4 knobs to functions for working with
latency and amplitude markers and auto cursors.
The command is particularly useful if you are using the SierraWin software without the custom base attached (that is, as a
review station), because you can enter and exit Cursor mode using the software alone.
Status Bar
Shows or hides the status bar at the bottom of the screen. A check mark appears next to this command when the status
bar is active.
The computer uses the status bar to display information messages, such as the "Ready" message shown below.
Save New
Saves the current test protocol under a new name. The new test retains all of the settings that were in effect when you
saved it (such as parameter settings, window sizes, and window locations).
The new test can be opened only by users who log in under the same user name that you used when you created the
test.
Save Parameters
Updates the current test protocol with the settings that are currently in effect. You can update only tests that you or
another user have previously saved under unique names (that is, you cannot update any of Sierra’s standard test
protocols).
The test protocol opens with the saved settings the next time you select it.
EMG
Opens the standard EMG (electromyography) test.
226
NCV/F/H
Opens the standard NCV (nerve conduction velocity) protocol. The type of test you perform (NCV, F-wave, or H-reflex)
depends on the kind of nerve you select.
SEPU
Opens the standard SEPU (somatosensory evoked potential - upper) test.
SEPL
Opens the standard SEPL (somatosensory evoked potential - lower) test.
BAEP Click
Opens the standard BAEP (brainstem auditory evoked potential) test.
VEP
Opens the standard VEP (visual evoked potential) test.
BLINK
Opens the standard blink reflex test.
RNS
Opens the standard RNS (repetitive nerve stimulation) test.
Median SEP
Opens the standard Median SEP (somatosensory evoked potential) test.
Ulnar SEP
Opens the standard Ulnar SEP (somatosensory evoked potential) test.
Tibial SEP
Opens the standard Tibial SEP (somatosensory evoked potential) test.
Peroneal SEP
Opens the standard Peroneal SEP (somatosensory evoked potential) test.
Upper Dermatomes
Opens the standard Upper Dermatomes SEP (somatosensory evoked potential) test.
Lower Dermatomes
Opens the standard Lower Dermatomes SEP (somatosensory evoked potential) test.
AEP
Opens the standard AEP (auditory evoked potential) test.
EMG2
Opens the standard EMG2 (electromyography) test, which is a variation of the EMG test. The only difference between
them is that the EMG2 test protocol has a different set of windows for the views.
227
SFEMG
Opens the standard SFEMG (single fiber electromygraphy) test.
SSFEMG
Opens the standard SSFEMG (simulated single fiber electromygraphy) test.
Next View
Cycles the display through full-screen views of each window for the test protocol.
Live, Grid
List of all windows for the test protocol currently in use. A check mark appears next to the window that is currently active.
Help Topics
Opens the table of contents for the Sierra Help file.
Protocol Help
Opens the Sierra Help file for the test protocol you are currently using.
What's This
Activates the Windows “What’s This Help feature. The cursor changes into a cursor coupled with a question mark. Click
this on an item on the screen a pop-up appears with information about that item.
ReportGen Help
Opens the Help file for the QuickMed Report Generator program.
About SierraWin
Opens a dialog box with information about the version of Sierra software that is currently installed on the computer and
the software copyright notice.
In the EMG and EMG2 protocols, you can configure the following:
• Report key assignments
• Default settings for EMG parameters
• Muscle list
• Scored muscles table
For additional information, click . . .
About the EMG test
Descriptions of parameters
228
Configuring settings for parameters
229
3 With the cursor, drag the column marker to the new position and then release the cursor.
4 Click Close to return to the test.
You use the scored muscles table in the EMG and EMG2 test protocols to enter values for the muscles that you are
testing. You can customize the table by changing entries, and by adding, changing, rearranging, and deleting columns.
Each test protocol can have its unique scored muscles table.
You configure the scored muscles table in EMG or EMG2 setup on the EMG tab. Each row in the Scored Muscles Table
Definition (shown below) corresponds to a column in the scored muscles table.
There are four types of entry possibilities for each column in the scored muscles table (see the examples in the illustration
below).
230
ReadOnly You cannot enter or change the text.
Text You can enter or change text.
List You must select from the list or choices that appears.
The first item in the list is always the default choice.
Note the following when you create the list:
• Enter a space between each List choice.
• When the text for a List choice includes a space,
use an underscore ( _ ) to indicate the space
(spaces are not allowed within a list choice).
• There is no limit to the number of choices you can
have in a list.
ListEdit You can select from the list of choices that appears or
enter something that is not on the list. The same notes
for creating a list apply to creating this entry type.
The Side, Muscle, Nerve, and Root columns must exist for the program to run. They can be in any order but must all be
included in the Scored Muscles Table Definition.
The Scored Muscles table adjusts to accommodate the combined width of all columns. It widens when you add columns.
The wider the table, the more scrolling that is required to see all of the information in the table.
231
2 Go to the last (blank) row in the table.
3 Click each column in turn and enter the information for the new column.
4 Click OK to return to the test.
The Side, Muscle, Nerve, and Root columns must exist for the program to run. They can be in any order but must all be
included in the Scored Muscles Table Definition.
To delete a column
1 On the Edit menu in the test, click Test Setup. You are positioned at the EMG tab.
2 Right-click the left column of the row you want to delete and then click Delete Row.
3 Click OK to return to the test.
In the NCV, F-wave, and H-reflex protocols, you can configure the following:
• Report key assignments
• Footswitch
• Default settings for NCV, F-wave, and H-reflex parameters
• Trace label format
• Default settings for nerves and nerve groups
232
• Nerve list, and associated site table and segment table
Descriptions of parameters
You can configure the following for individual nerves or for nerve groups (motor, sensory, F-wave, or H-reflex).
• Averaging
Whether averaging will be on or off when the test opens; if on, the default settings for the Avg Count and RR
(repetition rate) parameters.
• Channels
What channels are active when the test opens; for each channel, the gain, hicut filter, and locut filter values.
• Parameters
Default values for Pulse Width and Sweep.
• Site table
Columns that display for the site table. You can configure up to five columns; each can have one of 11 possible
headings.
233
You can change the reference site, sites, and segments for a nerve while you are conducting a test. The changes
apply only to the test you are conducting and do not affect the configured settings.
Changing reference site, sites, or segments for the current test
234
of the trace deviation from the baseline over the
interval from the onset of the response to the
recovery.
Norm Onset The normal values for onset latency of the response.
Norm Peak The normal values for peak cursor latency of the
response.
Norm Amp The normal values for the amplitude of the response.
To insert an entry within the list: Right-click the row above which you want to insert the new nerve and then click
Insert Row.
3 Click each column in turn and enter the information for the new nerve.
The nerve group you choose from the drop-down list for the Type column determines the default parameters
assigned to that nerve (for example, settings for gain, hicut filter, and locut filter). You configure these defaults on
the Nerve/Group Settings tab.
4 Configure sites for the nerve in the Site Table.
5 Configure nerve segments for recording conduction velocities in the Segment Table.
6 Click Close to return to the test.
____________________
See also
235
2 Click the cursor where you want to make the change.
3 Using the cursor and alphanumeric keyboard, delete the information you do not want and type the new information.
4 Click OK to return to the test.
Deleting an Entry from the Nerve List, Site Table, or Segment Table
To delete an entry
1 On the Edit menu in the NCV, F-wave, or H-reflex test, click Test Setup and then select the Nerve List tab.
2 Right-click the row you want to delete and then click Delete Row.
3 Click OK to return to the test.
A red line indicates where the roll will be positioned when you release the cursor. The row will be above the red
line.
4 Click Close to return to the test.
Descriptions of parameters
236
Configuring auto cursors
Descriptions of parameters
Descriptions of parameters
237
You can adjust stimulus intensity in several ways:
• Electrical stimulator
To increase the intensity level, turn the dial in the direction of the upward slope of the ramp label. To decrease the
intensity level, turn the dial in the direction of the downward slope of the ramp label. To adjust in small increments,
turn the dial slowly. Turning the dial quickly changes the intensity level in larger increments.
• Custom keyboard
Use the knob labeled for stimulation intensity.
• Toolbar
Use the stimulus intensity control.
The Stim indicator in the toolbar flashes yellow each time a stimulus is delivered.
Reversing Polarity
To reverse polarity
Turn the stimulator over.
Amplitude markers
238
Moving a Marker
To move a marker
Click the marker and drag it to the position you want.
The values in the markers toolbar change to reflect the new position.
With the User/Physician/Technician dialog box, you can add user, physician, and technician profiles to the QuickMed
database or choose a profile already stored in the QuickMed database. After you select a profile from the list, you can edit
the profile or delete it from the database.
Summary
The primary information for the currently selected user, physician, or technician, such as name, ID code, and contact
information.
Seek
Lets you quickly find a specific profile if the list is very large. As you type the last name in the box, the list scrolls to
show the names that match the letters you have typed.
List of profiles
The profiles currently entered in the QuickMed database.
To sort the list by a column, click a column heading. To change the sort direction of the column, click the column
heading again.
To change the width of the column, hold the pointer over the column boundary until it turns into a double arrow. Then
drag the column boundary to the desired width.
Add New
Opens the Add New Employee Data dialog box, allowing you to create new profiles in the QuickMed database.
Edit
Opens the Edit Employee Data dialog box, allowing you to change the information in the currently selected profile.
Delete
Removes the currently selected profile from the QuickMed database.
Done
Closes the dialog box.
239
Add Employee Data Dialog Box
With the Add Employee Data dialog box, you can create new profiles in the QuickMed database for users, physicians, and
referring physicians. Entering these names in the system is a convenience for you: once configured, the names appear
automatically in drop-down lists when you are asked for these types of people. You can select from the list instead of
typing the information.
To successfully add a new profile to the QuickMed database, the Last Name, First, and ID boxes must be
completed.
Last Name
The individual's last name.
First
The individual's first name.
MI
The individual's middle initial.
Title
The individual's professional or courtesy title (Dr. Mrs., Mr.).
Degree
The type of academic degree held by the individual (MD, PhD).
Specialty
The field in which the individual focuses his or her work.
Position
The type of job held by the individual. You can type whatever you want in this field. However, the names to which you
assign either Physician or Technician appear during the report generation process as physician and technician names
you can select for inclusion on reports.
ID
The individual's user ID, which is used to track the profile in the QuickMed database.
This user ID is automatically created when you type the individual's last name, first name, and middle initial. However,
you can customize the user ID, if desired.
240
eMail
The individual's e-mail address.
Works Here
Select this check box if the individual works on-site (that is, if the individual is a technician or physician who uses the
instrument). If the individual is a referring physician, do not select this box.
Work Phone
The individual's work telephone number.
Home Phone
The individual's home telephone number.
Security
The level of access privileges that the individual has, which determines what information the individual can view or
change.
QuickMed uses two types of security:
• NoAccess
Privileges only to conduct protocol tests.
• FullAccess
Privileges to conduct protocol tests, create new protocol tests, and configure the system.
Password
This feature will be used in future versions of the software.
OK
Stores the profile in the QuickMed database.
Cancel
Closes the dialog box without saving the information you have entered.
With the Edit Employee Data dialog box, you can change existing user, physician, and referring physician profiles in the
QuickMed database.
To maintain a valid profile in the QuickMed database, the Last Name, First, and ID boxes must contain entries.
Last Name
The individual’s last name.
241
First
The individual’s first name.
MI
The individual's middle initial.
Title
The individual's professional or courtesy title (Dr., Mrs., Mr.).
Degree
The type of academic degree held by the individual (MD, PhD).
Specialty
The field in which the individual focuses his or her work.
Position
The type of job held by the individual. You can type whatever you want in this field. However, the names to which you
assign either Physician or Technician appear during the report generation process as physician and technician names
you can select for inclusion on reports.
ID
The individual's user ID, which is used by the QuickMed database to track the profile.
This code is automatically created when the profile is created.
eMail
The individual's e-mail address.
Works Here
Select this check box if the individual works on-site (that is, if the individual is a technician or physician that uses the
instrument). If the individual is a referring physician, do not select this box.
Work Phone
The individual's work telephone number.
Home Phone
The individual's home telephone number.
242
Security
The level of access privileges that the individual has, which determines what information the individual can view or
change.
QuickMed uses two types of security:
• NoAccess
Privileges only to conduct protocol tests.
• FullAccess
Privileges to conduct protocol tests, create new protocol tests, and configure the Sierra system.
Password
This feature will be used in future versions of the software.
OK
Stores the changes in the individual profile in the QuickMed database.
Cancel
Closes the dialog box without saving the information you have entered.
The Open Patient File dialog box lets you search for and open patient files. You can change the search variables to
control which files display in the list of files. For example, you can display the files for all patients whose last name starts
with a particular letter.
Selection Criteria
Type of Record
The type of file you want QuickMed to display (for example, Sierra patient files).
PatientID
The ID code for the patient in the QuickMed database.
Last Name
The last name of the patient.
First Name
The first name of the patient.
243
The date the file you are searching for was saved in the QuickMed database. You can use these criteria to search
for files that have been saved:
• Since a specific date
(For example, all files saved after 5/1/99)
• Up to a certain date
(For example, all files saved before 7/25/99)
• Between a range of dates
(For example, all files saved between 5/1/99 and 7/25/99)
Summary
Any words or characters included in the file summary. After you type this information in the Summary box, select
Starts With, Contains, or Exact.
Search
Begins the search.
List of files
The list of files that match the criteria set in the Selection Criteria section.
To sort the list by a column, click a column heading. To change the sort direction of the column, click the column
heading again.
To change the width of the column, hold the pointer over the column boundary until it turns into a double arrow. Then
drag the column boundary to the desired width.
Print List
Prints a copy of the list of files.
Open
Opens the selected file.
Cancel
Closes the dialog box.
244
Patient List Dialog Box
With the Patient List dialog box, you can add patient files to the QuickMed database or choose a file already stored in the
QuickMed database. After you select a file from the list, you can edit the file or delete the file from the database.
Summary
The primary information for the currently selected patient, such as name, ID code, and contact information.
Enter Selection
Lets you quickly find a specific patient file if the list is very large. As you type the patient's last name in the box, the list
scrolls to show the names that match the letters you have typed.
Patient List
The patient files currently entered in the QuickMed database.
To sort the list by a column, click a column heading. To change the sort direction of the column, click the column
heading again.
To change the width of the column, hold the pointer over the column boundary until it turns into a double arrow. Then
drag the column boundary to the desired width.
Add New
Opens the Add Patient dialog box, which lets you create new patient files in the QuickMed database.
Edit
Opens the Edit Patient dialog box, which lets you change the information in the currently selected patient's file.
Delete
Removes the currently selected patient file from the QuickMed database.
Done
Closes the dialog box.
With the Add Patient dialog box, you can create new patient files in the QuickMed database. The dialog box has three
tabs:
Patient Entry
Use this tab to enter basic information about the patient.
Test Info
Use this tab to to set up additional information for the patient (staff associated with the patient, and patient height and
weight).
245
Custom Test Info
Use this tab to enter custom information about the (patient history, medications, findings, and conclusions).
Entering information on the Test Info and Custom Test Info tabs is a convenience for you. This information is included
automatically in test reports for the patient , so you do not have to enter the information each time you create a report.
Use the Patient Entry tab to enter basic information about the patient.
To successfully add a new patient file to the QuickMed database, you must complete the Last Name, First/MI,
and ID boxes.
Last Name
The patient's last name.
First/MI
The patient's first name and middle initial.
Title
The patient's professional or courtesy title.
Birthdate
The day the patient was born.
Type the date using an MM/DD/YY format. (For example, type the birthday of a patient born on February 1, 1958 as
02/01/58.)
SSN
The patient's Social Security Number.
Marital
The patient's marital status.
Race
The patient's race or ethnicity.
246
Sex
The patient's sex.
Address
Line 1
The building number and street where the patient would like correspondence to be sent.
Line 2
Additional street address information, such as a building name or the letter or number of a suite or apartment.
City/St
The city and state for the street address in Line 1 and Line 2.
Zip (+4)
The postal code for the street address in Line 1 and Line 2.
Type the first five numbers of the Zip code in the first box. Type the last four numbers of the Zip code in the
second box, if available.
Phone
Home
The patient's home telephone number.
Work
The patient's work telephone number.
Fax
The patient's fax number.
ID
The patient's ID code. This code is used by the QuickMed database to track the patient file and reports attached to
the patient file.
This ID code is automatically generated using a combination of the patient's last name, first name, and date.
However, you can change the ID before exiting the dialog box, if desired.
247
After the ID code is set in this dialog box, it cannot be changed.
eMail
The patient's e-mail address.
Start Test
Stores the patient file in the QuickMed database and allows you to start a test for the patient.
Save Only
Stores the patient file in the QuickMed database and closes the file.
Cancel
Closes the dialog box without saving the information you have entered.
Use the Test Info tab to to set up additional information for the patient. Entering this information is a convenience for you.
Once configured, the names appear automatically in drop-down lists when you are asked for these types of people. You
can select from the list instead of typing the information This information is also included automatically in test reports for
the patient , so you do not have to enter the information each time you create a report.
Physician
The physician associated with the patient. Click Select and select from the drop-down list or type a name.
A drop-down list of physicians appears only if you have previously configured physician profiles in the system.
Adding a Physician/Technician/Referring Physician profile
Referring Physician
The referring physician associated with the patient.
A drop-down list of referring physicians appears only if you have previously configured referring physician profiles
in the system.
Adding a Physician/Technician/Referring Physician profile
Technician
The technician associated with the patient.
A drop-down list of technicians appears only if you have previously configured technician profiles in the system.
Adding a Physician/Technician/Referring Physician profile
Start Test
248
Stores the patient file in the QuickMed database and allows you to start a test for the patient.
Save Only
Stores the patient file in the QuickMed database and closes the file.
Cancel
Closes the dialog box without saving the information you have entered.
Use the Custom Test Info tab to enter custom information about the patient. Entering this information is a convenience
for you. This information is included automatically in test reports for the patient, so you do not have to enter the
information each time you create a report.
Patient History
The patient’s history.
Medications
Medications taken by the patient.
Findings
Description of findings regarding the patient.
Conclusions
Description of conclusions regarding the patient.
Start Test
Stores the patient file in the QuickMed database and allows you to start a test for the patient.
Save Only
Stores the patient file in the QuickMed database and closes the file.
Cancel
Closes the dialog box without saving the information you have entered.
With the Edit Patient dialog box, you can change existing patient files in the QuickMed database. The dialog box has three
tabs:
Patient Entry
Use this tab to enter basic information about the patient.
Test Info
249
Use this tab to to set up additional information for the patient (staff associated with the patient, and patient height and
weight).
Use the Patient Entry tab for basic information about the patient.
To maintain a valid patient file in the QuickMed database, you must complete the Last Name, First/MI, and ID
boxes.
Last Name
The patient's last name.
First/MI
The patient's first name and middle initial.
Title
The patient's professional or courtesy title.
Birthdate
The day the patient was born.
Type the date using an MM/DD/YY format. (For example, type the birthday of a patient born on February 1, 1958 as
02/01/58.)
SSN
The patient's Social Security Number.
Marital
The patient's marital status.
250
Race
The patient's race or ethnicity.
Sex
The patient's sex.
Address
Line 1
The building number and street where the patient would like correspondence to be sent.
Line 2
Additional street address information, such as a building name or the letter or number of a suite or apartment.
City/St
The city and state for the street address in Line 1 and Line 2.
Zip (+4)
The postal code for the street address in Line 1 and Line 2.
Type the first five numbers of the Zip code in the first box. Type the last four numbers of the Zip code in the
second box, if available.
Phone
Home
The patient's home telephone number.
Work
The patient's work telephone number.
Fax
The patient's fax number.
ID
251
The patient's ID code. This code is used by the QuickMed database to track the patient file and reports attached to
the patient file.
This ID code is automatically generated using a combination of the patient's last name, first name, and date.
However, you can change the ID before exiting the dialog box, if desired.
eMail
The patient's e-mail address.
Start Test
Stores the patient file in the QuickMed database and allows you to start a test for the patient.
Save Only
Stores the patient file in the QuickMed database and closes the file.
Cancel
Closes the dialog box without saving the information you have entered.
Use the Test Info tab to maintain additional information for the patient. Entering this information is a convenience for you.
Once configured, the names appear automatically in drop-down lists when you are asked for these types of people. You
can select from the list instead of typing the information This information is also included automatically in test reports for
the patient , so you do not have to enter the information each time you create a report.
Physician
The physician associated with the patient. Click Select and select from the drop-down list or type a name.
A drop-down list of physicians appears only if you have previously configured physician profiles in the system.
Adding a Physician/Technician/Referring Physician profile
Referring Physician
The referring physician associated with the patient.
A drop-down list of referring physicians appears only if you have previously configured referring physician profiles
in the system.
Adding a Physician/Technician/Referring Physician profile
Technician
The technician associated with the patient.
A drop-down list of technicians appears only if you have previously configured technician profiles in the system.
Adding a Physician/Technician/Referring Physician profile
252
Height and Weight
The patient’s height and weight.
Start Test
Stores the patient file in the QuickMed database and allows you to start a test for the patient.
Save Only
Stores the patient file in the QuickMed database and closes the file.
Cancel
Closes the dialog box without saving the information you have entered.
Use the Custom Test Info tab to maintain custom information about the patient. Entering this information is a
convenience for you. This information is included automatically in test reports for the patient, so you do not have to enter
the information each time you create a report.
Patient History
The patient’s history.
Medications
Medications taken by the patient.
Findings
Description of findings regarding the patient.
Conclusions
Description of conclusions regarding the patient.
Start Test
Stores the patient file in the QuickMed database and allows you to start a test for the patient.
Save Only
Stores the patient file in the QuickMed database and closes the file.
Cancel
Closes the dialog box without saving the information you have entered.
253
The Select Report dialog box lets you search for and open report files. You can change the search variables to control
which files display in the list of files. For example, you can display the files for all patients whose last name starts with a
particular letter.
Selection Criteria
Type of Record
The type of file you want QuickMed to display.
PatientID
The ID code for the patient in the QuickMed database.
Last Name
The last name of the patient.
First Name
The first name of the patient.
Summary
Any words or characters included in the file summary. After you type this information in the Summary box, select
Starts With, Contains, or Exact.
Search
Begins the search.
254
List of files
The list of files that match the criteria set in the Selection Criteria section.
To sort the list by a column, click a column heading. To change the sort direction of the column, click the column
heading again.
To change the width of the column, hold the pointer over the column boundary until it turns into a double arrow. Then
drag the column boundary to the desired width.
Print List
Prints a copy of the list of files.
Open
Opens the selected file.
Cancel
Closes the dialog box.
SPL
Sound pressure level. Used in acoustics for continuous tones.
nHL
Normal hearing level. The softest intensity that an average, normal adult listener can hear.
To store the results of the current run, press the Store key or use the Run Table control. The run table appears with the
muscle name and side designation appearing at the top. A separate line of test data appears in the table for each run. The
mean and standard deviation are automatically calculated and displayed at the bottom of the table. The run table can
include up to 20 runs per muscle.
Once a run has been added to the run table, use the OK control in the run table or on the custom keyboard to return to
analysis mode. The Stored status for the run will now be set to Yes. To return to acquisition mode, press the Run/Stop
key.
Data from the current run will be lost if you use Run/Stop key storing the run to the run table.
___________________
See also
Run table
Analysis Table
255
Run Number of the current run for the selected muscle. Up to
20 runs can be acquired for each muscle.
Side Muscle side. If necessary use the Param Sel/Modify
knob to change.
Muscle Name of the selected muscle. Before analyzing, you can
change the selected muscle using Thresh/Musc Sel.
Samples Total number of traces in the current run.
Blocks Number of traces in the current run that do not exceed
the analysis marker level.
MCD Norm MCD normal value for the selected muscle. You can
configure this value in the muscle list.
Configuring the muscle list
MCD Mean consecutive difference measurement for the
current run (the jitter value, in microseconds).
MSCD Mean sorted consecutive difference measurement for
the current run (in microseconds).
IPI Mean interpotential interval measurement for the current
run (in milliseconds).
Freq Mean firing frequency of the triggering potential (in
Hertz).
Fib Dens Fiber density value. You can manually enter information
in this field.
Stored Shows whether the run has been stored (No - run has
not been stored, Yes - run has been stored).
Run Table
Delete All Use this control to delete all runs for the selected
muscle.
256
SFEMG Data Acquisition Mode
257
SFEMG Analysis Marker
258
Use the archive procedure to back up entire patient charts from the QuickMed database onto the backup medium of your
choice. You can either copy or move the charts. Archiving zips (compresses) all the patient charts you select into one
archive file.
• Copy
Makes a copy of the selected charts, but leaves the original charts in the QuickMed database.
• Move
Makes a copy of the selected charts, and deletes the original charts from the QuickMed database.
Use the restore procedure to open an archive file and import all or selected charts back into the QuickMed database.
It’s a good idea to create a “backup folder on your hard disk drive where you can save archive files. Then, from
there, copy or move the files to a more permanent archive medium, such as a Zip drive, tape drive, or recordable
CD.
Use the search feature to locate specific patients. Enter a character or set of characters in any or all of the search
fields. When you click Search, you get a list of all patients who match your search criteria. Click Clear to clear the
results of the search and start another search. To get the complete, unfiltered list back, clear all fields and then
click Search.
8 Click OK. The single archive file is created, incorporating all the patient charts you selected.
9 Click OK to leave the Quickmed Utilities dialog box.
259
5 Select the patients you want to restore.
• One patient: Click the patient name.
• More than one patient: Hold the Ctrl key down on the computer keyboard and select the patients.
6 Click OK. The patient charts you selected are imported back into the QuickMed database.
7 Click OK to leave the Quickmed Utilities dialog box.
You can use the Review Menu to locate and review the data you restored.
Use the delete procedure to remove entire patient charts from the QuickMed database. You have two choices in deleting:
• Delete the patient chart from the system permanently (no recovery unless you have previously archived the chart).
• Delete the patient chart from the QuickMed database, but copy it into another folder outside the database. You can
then use the undelete function to move these charts back into the QuickMed database (for example, if you deleted the
charts by mistake).
It’s a good idea to archive before you delete any patient charts.
Archiving patient charts
Use the search feature to locate specific patients. Enter a character or set of characters in any or all of the search
fields. When you click Search, you get a list of all patients who match your search criteria. Click Clear to clear the
results of the search and start another search. To get the complete, unfiltered list back, clear all fields and then
click Search.
4 Click OK.
5 When asked to confirm the delete, do one of the following:
• To delete permanently: Place a checkmark in the Delete Permanently box and click Yes.
• To delete from the QuickMed Database, but not from the system. Click Yes, but do not place a checkmark in the
Delete Permanently box. The patient charts are copied into a folder located at C:\Quickmed\QMDeletedCharts.
You can use the Undelete function to restore these patients (explained just below).
6 Click OK to leave the Quickmed Utilities dialog box.
260
2 Click Undelete. A list of the deleted patient charts that have been placed in the C:\Quickmed\QMDeletedCharts folder
appears.
3 Select the patient charts you want to move back into the QuickMed database.
• One patient: Click the patient name.
• More than one patient: Hold the Ctrl key down on the computer keyboard and select the patients.
4 Click OK. The patient charts you selected are moved out of the QMDeletedCharts folder and back into the QuickMed
database.
5 Click OK to leave the Quickmed Utilities dialog box.
Use the export and import procedures to exchange patient charts with Sierra users on other computers. You can export
and import only one patient chart at a time.
The system automatically zips (compresses) exported charts to save disk space. When you import, the charts are
unzipped.
The import procedure is also used to import individual test data records or reports that were exported from the
system using the export feature of the Records tab. (Note that there is no Import button on the Records tab.)
It’s a good idea to create an “export folder on your hard disk drive where you can save exported charts. Then,
from there, copy or move the files to a transportable medium, such as a floppy diskette, Zip drive, or recordable
CD.
You can use the Review Menu to locate and review the data you imported.
261
Adding a Custom Field
Use this procedure to add a custom field to the Patient Information dialog box in order to collect data about patients for the
QuickMed database.
In reports, you can include information that you enter in a custom field. To do this, you configure the custom field in the
report templates. For information on customizing reports, go to the ReportGen Help file (on the Help menu, click
ReportGen Help).
You cannot include any spaces in what you type. Use an underscore character (_) for any spaces you want in the
name.
The name that you assign here becomes the name of the “token you use when configuring report templates to
include this custom field.
5 In Field Size, select a size for the field.
Short: A small box of 23 characters maximum.
Medium: A single-line box, about 65 characters.
Long: A multiple-line box spanning the entire width of the Patient Information dialog box.
6 If desired, change the order of the fields. To do this, select the field and use the up and down buttons.
7 Click OK to save the new custom field.
8 Click OK to leave the Quickmed Utilities dialog box.
Check out the field you just added. On the File menu, click New Patient and then click the Custom Test Info tab.
Use this procedure to change fields in the Patient Information dialog box that are used to enter data about patients for the
QuickMed database.
If you change the name of a field, you must make the same change in report templates that include the field. For
information on customizing reports, go to the ReportGen Help file (on the Help menu, click ReportGen Help).
If you change the name of a field, you also delete all the data collected by means of the field under its previous
name. For example, suppose you used the “Conclusions field for 10 patients and then change the name to
“Recommendations. If you recall the stored data records for any of those previous patients, the records will no
longer contain any “Conclusions information and the “Recommendations field will be blank.
262
3 In the Form Name field, select Sierra. A list of all current fields appears.
4 Click the field you want to change and make the change.
• To change the name: Type the new name in FieldName. You cannot include spaces in what you type; use an
underscore character (_) for any spaces you want in the name.
• To change the size of a field: Select from the options in FieldSize:
Short: A small box of 23 characters maximum.
Medium: A single-line box, about 65 characters.
Long: A multiple-line box spanning the entire width of the Patient Information dialog box.
5 Click OK to save your changes.
6 Click OK to leave the Quickmed Utilities dialog box.
Use this procedure to delete a field from the Patient Information dialog box that you no longer need.
If you delete a field, you also delete all the data collected by means of the field. For example, suppose you used
the “Conclusions field for 10 patients and then delete the field. If you recall the stored data records for any of
those previous patients, the records will no longer contain any “Conclusions information.
Use the export procedure to export an individual record (test data or report) from a patient chart to a file that can be
transferred to another Sierra II.
It’s a good idea to create an “export folder on your hard disk drive where you can save exported records. Then,
from there, copy or move the files to a transportable medium, such as a floppy diskette, Zip drive, or recordable
CD.
To export a record
1 On the Edit menu, click Quickmed Utilities.
2 Click the Records tab and then click Export. The Export Records dialog box appears.
3 Change the Type of Record field to Any (for a list of all records), Sierra (for data file records), or Report (for report
records).
4 Click the record you want to export.
263
Use the search feature to locate a specific record. You can enter one or more characters for patient ID or name, or
search by dates. The Summary field lets you search for records that start with, contain, or match characters
exactly. When you click New Search, you get a list of all records that match your search criteria. To get the
complete, unfiltered list back, clear all search fields and then click New Search.
5 Click Export. The Select Export Destination dialog box appears.
6 In the Save in field, enter where you want the export file placed. You can use any location or drive on the system.
7 In the File name field, change the name of the export file, if desired. The system assigns an extension of “.qmc to the
file, which you cannot change.
8 Click Save.
9 Click OK to leave the Quickmed Utilities dialog box.
Use the delete procedure to remove a specific record (data file or report) from a patient chart in the QuickMed database.
You can use the undelete function to move these records back into the QuickMed database (for example, if you deleted
the records by mistake).
It’s a good idea to back up before you delete any patient records.
Archiving patient records
To delete a record
1 On the Edit menu, click Quickmed Utilities.
2 Click the Records tab and then click Delete. The Delete Records dialog box appears.
3 Change the Type of Record field to Any (for a list of all records), Sierra (for data file records), or Report (for report
records).
Use the search feature to locate a specific record. You can enter one or more characters for patient ID or name, or
search by dates. The Summary field lets you search for records that start with, contain, or match characters
exactly. When you click New Search, you get a list of all records that match your search criteria. To get the
complete, unfiltered list back, clear all search fields and then click New Search.
4 Click the record you want to delete.
5 Click Delete.
6 Click OK to leave the Quickmed Utilities dialog box.
To undelete a record
1 On the Edit menu, click Quickmed Utilities.
2 Click the Records tab and then click Undelete. The Restore Deleted Records dialog box appears.
3 Change the Type of Record field to Any (for a list of all records), Sierra (for data file records), or Report (for report
records).
4 Click the record you want to restore.
5 Click Undelete.
6 Click OK to leave the Quickmed Utilities dialog box.
264
Backing Up the User Database
Use the backup procedure to back up all of the following in a single operation:
• Profiles of all users, physicians, referring physicians, and technicians.
• All user settings (such as default settings for protocols).
• Report templates.
To preserve individual user’s settings in the user database, use the Preserve Settings procedure, which is a
feature of system setup.
Preserving and restoring settings
It’s a good idea to create a “backup folder on your hard disk drive where you can save backup files. Then, from
there, copy or move the files to a more permanent backup medium, such as a Zip drive, tape drive, or recordable
CD.
Use the restore procedure to copy a previous (backup) copy of the user database back into the system. This overwrites all
current user data with the previous data.
Use the compress procedure to “clean up the user database. The procedure removes information that is no longer
necessary and compacts the user database file into a smaller size.
265
1 On the Edit menu, click Quickmed Utilities.
2 Click the User Databases tab and then click Compress.
3 Click OK to leave the Quickmed Utilities dialog box.
In the SFEMG and SSFEMG protocols, you can configure the following:
• Report key assignments
• Default settings for SFEMG and SSFEMG parameters
• Muscle list
For additional information, click . . .
About the SFEMG and SSFEMG tests
Descriptions of parameters
266