EVS 2430Wi User’s Manual To Customers
To Customers
Thank you for purchasing the DRTECH Radiography EVS 2430Wi (hereinafter, this Product). This User’s
Manual explains how to use the detector, x-ray interface unit, and other peripheral equipments. Before
using this product, be sure to read this manual thoroughly in order to utilize it more effectively. Also, please
read the Operation Manual for EVS Calibration and configuration Software (hereinafter, ECali1).
Disclaimer
1. In no event shall DRTECH be liable for any damage or loss arising from fire, earthquake, any action or accident by
a third party, any intentional negligent action by users, any trial usage, or other usage under abnormal conditions.
2. Roentgenography, image processing, image reading, and image data storage must be performed in accordance
with the laws of the country or region in which the product is being used. The user is resposible for maintaining the
privacy of image data.
3. In no event shall DRTECH be liable for personal physical harm or property damage that is sustained when the
instructions are not followed or the product is misused.
4. It is the resposibility of the attending physicians to provide medical care services. DRTECH will not be liable for
faulty diagnoses.
5. In no event shall DRTECH be liable for direct or indirect consequential damages arising from the use or
unavailbility of this product. DRTECH shall not be liable for loss of image data for any reason.
6. In no event shall DRTECH be liable for any damage arising from moving, alteration, inspection or repair by a
person other than authorized service engineers.
7. Specifications, compositions, and appearance of this product may change without prior notice.
DRT-MAN-092 2
EVS 2430Wi User’s Manual Contents
Contents
To Customers .......................................................................................................................................... 2
Contents ................................................................................................................................................... 3
Safety notices .......................................................................................................................................... 7
2. Introduction ............................................................................................ 16
2.1. Features ....................................................................................................... 16
2.2. System Configuration................................................................................ 17
2.2.1. Basic Configuration .......................................................................................................................... 17
DRT-MAN-092 3
EVS 2430Wi User’s Manual Contents
4.2.2. Battery Pack .......................................................................................................................................... 29
4.2.2.1. Battery Pack Specification ........................................................................................................................................... 29
4.2.2.2. Battery Pack Components ............................................................................................................................................ 29
4.2.2.3. Charging Battery Pack .................................................................................................................................................... 30
4.2.3. Adaptor .................................................................................................................................................... 30
4.3. Wireless Charging System ....................................................................... 31
4.3.1. EVS-WPCS ............................................................................................................................................. 31
4.3.1.1. EVS-WPCS Specifications ............................................................................................................................................ 31
4.3.1.2. EVS-WPCS Components ............................................................................................................................................... 31
4.3.1.3. TX / RX Module Specification ...................................................................................................................................... 32
4.3.1.4. RX Module Components ................................................................................................................................................ 32
4.3.1.5. EVS-WPCS Operations ................................................................................................................................................... 33
4.3.1.6. Attach Direction .................................................................................................................................................................. 33
4.3.1.7. Indicator LED Connector Description .................................................................................................................... 34
4.3.1.8. Indicator LED Connector Description (Isolation) ............................................................................................. 34
8. Troubleshooting .................................................................................... 68
8.1. Failed to Turn the Detector On ................................................................. 68
8.2. Errors in Detector LED .............................................................................. 68
8.3. The LINK LED does Not Turn on .............................................................. 69
8.4. Rapid Consumption of Battery ................................................................ 69
8.5. Battery Pack or Installation Part of Battery is Getting Hot ................. 70
8.6. The Power Switch of SSU or Status LED is not working ..................... 70
DRT-MAN-092 5
EVS 2430Wi User’s Manual Contents
10.1.2. Battery Charger System .............................................................................................................. 73
10.1.3. Battery Pack ...................................................................................................................................... 74
10.1.4. EVS-WPCS ......................................................................................................................................... 75
10.2. Charateristics ........................................................................................... 76
10.3. Packing...................................................................................................... 77
10.3.1. Product Configuration List ........................................................................................................ 77
10.3.2. Assemble Package ........................................................................................................................ 78
10.3.3. Detector Panel Package .............................................................................................................. 79
10.3.4. Component Box Assemble Package .................................................................................... 80
10.3.4.1. AP Box Component ....................................................................................................................................................... 80
10.3.4.2. Battery Box Component .............................................................................................................................................. 80
DRT-MAN-092 6
EVS 2430Wi User’s Manual Conventions
Safety notices
The following safety notices are used to emphasize certain safety instructions. Follow the safety instructions in this
User’s Manual along with warning and cautions symbols. Ignoring such warnings or cautions while handling the
product may results in serious injury or accient. It is important for you to read and understand the contents of this
User’s Manual before attemting to use the product.
This notice used to idendifiy conditions under which improper use of the
WARNING product may cause death or serious personal injury.
This notice used to idendifi\y conditions under which improper use of the
CAUTION product may cause minor personal injury.
This notice used to idendifiy conditions under which improper use of the
CAUTION product may cause property damage.
Prohibited
DRT-MAN-092 7
EVS 2430Wi User’s Manual 1. Safety Information
1. Safety Information
1.1. Safety precautions
Follow these saftey guides and properly use the equipment to prevent injury and damage to any equipment/data.
WARNING
Installation and environment of use
- Do not use or store the equipment near flammable chemical such as alcohol, thinner,
benzine, etc.
If chemicals are spilled or evaporated, it may result in fire or electric shock through contact with
electric parts inside the equipment. Also, some disinfectants are flammable. Be sure to take care
when using them.
Prohibited - Do not connect the equipment to anything other than specified connectoins.
Doing so may result in fire or electric shock.
Power supply
- Do not operate the equipment using any type of power supply other than the one
indicated on the rating label.
Otherwise, it may result in fire or electric shock.
- Do not handle the equipment with wet hands.
You may experience an electric shock that could result in death or serious injury.
- Do not place heavy object such as medical equipments on cables and cords, or do not
pull, bend, bundle, or step on them. These precautions are required to be followed to
prevent sheathes of cables and cords from being peeled. Do not alter the cables and
cords.
Doing so may damage the cords which could result in fire or electric shock.
Prohibited - Do not supply power to more than one of equipment using the same AC outlet.
Doing so may result in fire or electric shock.
- Do not turn on the system power when condensation has formed on the equipment.
Doing so may result in fire or electric shock.
- Do not connect multiple portable socket-outlets or extension cords to the system.
Doing so may result in fire or electric shock.
-
DRT-MAN-092 8
EVS 2430Wi User’s Manual 1. Safety Information
WARNING
Handling
The system, in whole or in parts, cannot be modified in any ways without any written approval from DRTECH.
- Have the patient take a fixed posture and do not let patient touch the parts unnecessarily.
If a patient touches connectors or switches, it may result in electric shock or malfunction of the
equipment.
- If any of the following problems occur, immediately turn OFF the power to each piece of
equipment, unplug the power cord from the AC outlet, and contact your sales
representative or local DRTECH dealer:
When there is smoke, when an odd smell or abnormal sound occurs.
When liquid has been spilled into equipment or metal object has entered through an opening.
When the equipment was dropped and damaged.
- Clean the plug of the power cord periodically by unplugging it from the AC outlet and
removing dust or dirt from the plug. Clean the peripherals and AC oulet with a dry cloth.
If the cord is kept plugged in for a long time in a dusty, humid or a sooty place, objects around the
plug will attract moisture, and this could cause insulation failure that could result in a fire.
- For safety reasons, be sure to turn OFF the power to each piece of equipment when the
inspections indicated in this manual are going to be performed.
Otherwise, electric shock may occur.
DRT-MAN-092 9
EVS 2430Wi User’s Manual 1. Safety Information
WARNING
Battery pack and charger
- Do not use the battery pack as a power source for equipment other than EVS 2430Wi detectors.
Be sure to use only the dedicated battery pack for the EVS 2430Wi detector.
- The battery charger is designed for the dedicated battery pack. Do not use the battery charger
other than the dedicated one. Otherwise, a battery explosion or a battery leak may occur,
resulting in fire or electrical shock.
- Do not operate the battery charger using any type of power supply other than the one indicated
on the rating label.
- Do not handle the product with wet hands.
- Do not attempt to disassemble, alter, or apply heat to the product.
- Avoid dropping or subjecting the product to severe impacts. To avoid the risk of injury, do not
touch the internal parts of the battery if it has been cracked.
- Stop using the battery pack immediately if it emits smoke, a strange smell, or otherwise behaves
abnormally.
- Do not let the battery pack and battery charger come into contact with water or other liquids and
do not allow them to get wet.
- Do not clean with substances containing organic solvents such as alcohol, benzene, thinner, or
other chemicals. Otherwise, fire or electrical shock may occur.
- Do not allow dirt or metal objects (such as hair pins, clips, staples or keys) to contact the
terminals. Otherwise, battery explosion or leakage of electrolyte may occur, resulting in fire,
injury or pollution of surrounding area. If the battery leaks and the electrolytes come into contact
with your eyes, mouth, skin or clothing, immediately wash it away with running water and seek
medical attention.
- Do not leave, store, or place the product in a location near heat sources, or in a place subject to
direct sunlight, high temperature, high humidity, excessive dust, or mechanical shock. Otherwise,
battery leakage, overheating or damage to the product may occur, resulting in electrical shock,
burns, injury or fire.
- Do not attempt to use a battery pack that has deteriorated. Using a battery pack that has
exceeded its life cycle may lead to overheating, fire or explosion.
- The Lithium ion/polymer battery is recyclable.
- Battery slowly discharges even if not in use.
- The useful battery life can expire if it discharges immediately after being fully charged. You can
purchase an optional battery pack to replace an exhausted one.
- The battery pack is a consumable item. If a fully charged battery is consumed quickly, use a new
and fully charged battery pack.
- Be sure to charge the battery periodically (once a year) if it is not used for an extended period of
time. The battery pack cannot be charged if it has been over discharged.
- Before discarding the battery pack, cover the terminals with adhesive tape or other insulators.
Contact with other metal materials may cause fire or explosion.
- Remove battery pack from equipment which is unused for a few minutes to prevent discharge .
DRT-MAN-092 10
EVS 2430Wi User’s Manual 1. Safety Information
WARNING
Installation and environment of use
- Do not install the equipment in any of the locations listed below. Doing so may result in
malfuction, equipment faillng, fire or injury.
ㆍ Close to facilities where water is used.
ㆍ Where it may be exposed to direct sunlight.
ㆍ Close to the air outlet of an air-conditioner or ventilation equipment.
ㆍ Close to a heat source such as a heater.
ㆍ Where the power supply is unstable.
ㆍ In a saline or sulfurous environment.
ㆍ Where temperature or humidity is high.
ㆍ Where there is freezing or condensation.
ㆍ In areas prone to vibration.
ㆍ On an inclined surface or in an unstable area.
- Because the equipment cable is long, take care that cables do not become tangled during
use. Also, be careful not to get your feet caught in the cable.
Otherwise, it may cause a malfunction of the equipment or injury of the user due to tripping over
the cable.
Power supply
- Always connect the three-core power cord plug to a grounded AC power outlet.
- To avoid the risk of electric shock, this equipment must only be connected to a power
supply that maintains protective earth”.
- To make it easy to disconnect the plug at any time, avoid putting any obstacles near the
outlet. Otherwise, it may not be possible to disconnect the plug in an emergency.
- Be sure to ground the equipment to an indoor grounded connector. Also, be sure to
connect all the earth connections for the system to a common ground.
- Do not use any power source other than the one provided with this equipment.
Otherwise, fire or electric shock may be caused due to leakage.
- The product has lower breaking capacity type. So do not intall at the building power
system prospective short-circuit current exceeding 35A
Handling
- Do not spill liquid or chemicals onto the equipment. In cases when the patient is injured,
do not allow liquid or chemicals to come in contact with blood or body fluids.
Doing so may result in fire or electric shock.
In such a situation, protect the equipment with a disposable covering as necessary.
- Turn OFF the power to each piece of equipment for safety when not being used.
DRT-MAN-092 11
EVS 2430Wi User’s Manual 1. Safety Information
CAUTION
Handling the equipment
The Equipment must be handled with care to avoid personal injury or damage to the internal image sensor.
<Load Limit>
Uniform load: 150 kg over the whole Local load: 100 kg on an area
area of the detector of 40 mm in diameter
- Be sure to use the detector on a flat surface that does not bend.
Otherwise, the internal image sensor can be damaged.
- Be sure to securely hold the detector while using it in upright positions.
Otherwise, the detector may fall over, resulting in injury to the user or patient, or
may flip over, resulting in damage to the inner device.
DRT-MAN-092 12
EVS 2430Wi User’s Manual 1. Safety Information
System Diagnostic
• Run Ecali1 software after installing the system, at least once a year. If an error occurs, report the detailed error
information to DRTECH local dealer or distributor.
Calibration
• To ensure optimal performance of the system, it is important to verify that the system is calibrated.
Before exposure
• Be sure to check the equipment on a daily basis and confirm that it works properly.
• Suddenly heating the room in clod area will cause condensation to form on the equipment. In this case, wait until
the condensation evaproates before performing an exposure. If the equipment is used while the condensation
forms, problems may occur in the quality of captured images. When an air-conditioner is used, be sure to
raise/lower the temperature gradually so that a significant change in temperature in the room and in the
equipment does not occur, to prevent condensation.
DRT-MAN-092 13
EVS 2430Wi User’s Manual 1. Safety Information
During exposure
• Do not use the selected frequency chanel (2.4GHz and 5GHz dual band) for other wireless devices.
Mutual interference may affect the image data transmission rate.
• Do not use the detector near devices generating a strong magnetic field. Doing so may produce image noise or
artifacts.
Image Backup
• To avoid missing images which might result in patient being exposed to additional doses of radiation, it is
important to send the images to PACS or back up the images by filming or by using external storage devices
such as CD, DVD, HDD, USB, etc.. This should be done as a routine operation for every patient. If the hard disk
of your workstation is about to full, the operator should backup images and manually delete the images as
administrator to make room on the hard disk for new images.
User Limitation
• The Ecali1 software has the technician mode which could only be operated by inputting the correct PASSWORD.
The technician mode should be operated by the personnel who are qualified by DRTECH.
• To reduce electric shock hazards, the system must be connected to an electrical ground.
• A three-contact conductor AC power cable is supplied with this system to provide the proper electrical grounding.
The power cable must be plugged into an UL-approved three-contact electrical outlet.
• Do not disassemble or modify the product as it may result in fire or electric There are no serviceable parts inside
equipments and adjustments should not be made. Only trained and qualified personnel should be permitted
access to the internal parts of the system.
• To reduce electric shock hazards, product is required to be well insulated with the use of appliance coupler, mains
plug, and other seperable connections.
DRT-MAN-092 14
EVS 2430Wi User’s Manual 1. Safety Information
• Do not use solvents such as alcohol, thinner or benzene. Doing so may damge the surface of the equipment.
Operating/storage environment
• Be sure to use and store this equipment under the conditions described below:
Transportation &
-10 to 40 ℃ 5 to 95 % RH 500 to 1060 hPa
Storage environment
• Do not expose this equipment to high temperatures and/or high humidity. Malfunction can occur.
• When not in use, keep the detector, handle unit, and grid in a designated location or in a location where they are
safe and cannot fall down.
Notes on disposal
• Disposal of this product in an unlawful manner may have a negative impact on health and on the environment.
Therefore, when disposing this product, be absolutely sure to follow the procedure which complies with the laws
and regulations applicable in your area.
• The equipment must be handled with care to avoid personel injury or damage to the internal image sensor.
• The EVS 2430Wi Wireless is an advanced wireless digital radiographic equipment in the DRTECH Exprimer
series. This equipment is designed to provide the highest resolution and sensitivity in the series. In addition, the
wireless LAN (IEEE 802.11n*) communication feature improves the operability, and high-speed processing.
DRT-MAN-092 15
EVS 2430Wi User’s Manual 2. Introduction
2. Introduction
2.1. Features
• Wireless LAN communication (IEEE 802.11n*) includes a lightweight and thin detector that is easy to handle.
• The shape of the detector, which is identical to that of a conventional film cassette complying with ISO4090,
enables digital radiography to apply to the existing analog radiography configuration.
• The new sensor with 76 μm of pixel pitch and CsI (Cesium Iodide), Gadox (Gadolinium Oxysulfide) used for the
scintillator produces high resolution (approx. 11.8 Mega pixels) digital images within the effective imaging area
(223 x 291 mm), with low doses of X-rays.
• Depending on the operating conditions at each site, the wiring unit (optional) enables the equipment to be used
through expansion to a wired connection.
At the time of installation, set a specific channel in the frequency band of 5.0 GHz before using the LAN.
Note that the available frequency band for this standard varies, depending on the local radio laws,
regulations and system requirements.
Intended Use
The EVS 2430Wi of Flat Panel Digital X-ray detector for industry was designed for industrial use.
Considerations Condition
Operating conditions
- Temperature: 10 ºC to +35 ºC
- Barometric Pressure: 700 hPa to 1060 hPa
- Humidity: 30 % R.H. to 85 % R.H.
Storage and delivery conditions
- Temperature: -10 ºC to +40 ºC
Environment including hygienic requirements
- Barometric Pressure: 500 hPa to 1060 hPa
- Humidity: 5 % R.H. to 95 % R.H.
Non-sterile
Less than ten minute contact
Ambient luminance range: 100 lx to 1500 lx
Viewing angle: normal to the display ± 20°
DRT-MAN-092 16
EVS 2430Wi User’s Manual 2. Introduction
Generally, the EVS 2430Wi detector is used in system configuration as illustrated below:
DRT-MAN-092 17
EVS 2430Wi User’s Manual 2. Introduction
Wireless Connection
• EVS 2430Wi wireless detector transmits images and data by wireless communication.
• A battery pack should be installed in the detector to use it under the wireless configuration.
- Use of multiple WLAN devices within the same frequency band may cause interference within
each wireless communication and slow down the transmission speed
- Do not cover or block the wireless module of the detector. Otherwise, the transmission speed or
operable distance may reduce.
- Recommended maximum operating distance of wireless communication, between the detector
and Access Point (AP), is 8 meters.
Wired Connection
• Connect EVS 2430Wi wireless detector and SSU / or Functional cable with a tether interface cable to make a
wired configuration.
• As the tether interface cable supplies power, a battery pack is not needed to be inserted in the detector.
The wired connection is more suitable for stabilized communication when the detector is set in a
bucky stand or on a table.
enables the battery pack to be continously supplied with power while using the detector.
drastically.
DRT-MAN-092 18
EVS 2430Wi User’s Manual 2. Introduction
EVS 2430Wi Wireless system consists of detector, system synchronization unit (SSU), CDs and relevant accessories.
(Refer to chapter 3-1 “Product Components” for CD information)
Wireless Charging System Power adaptor (12V, 7.08A) LAN Card(PCIe Giga-bit LAN)
: Optional + AC Power Cable(2m) LAN Cable (CAT.7, 15m)
DRT-MAN-092 19
EVS 2430Wi User’s Manual 2. Introduction
DRT-MAN-092 20
EVS 2430Wi User’s Manual 3. Production Description
3. Product Description
3.1. Product Components
3.1.1. Auto Trigger Mode (AT Mode)
WARNING
The use of accessories and cables other than those specified, with the exception of EVS 2430Wi Wireless
accessories and cables sold by DRTECH Co., LTD. as replacement parts for internal components, may result in
increased emissions or decreased immunity of the equipment. Accessory equipment connected to the analog and
digital interfaces must be certified according to the respective IEC standards. All combinations of equipment must be
in compliance with IEC 60601-1-1 system requirements. Any person who connects additional equipment to the signal
input or signal output ports configures a medical system, and is therefore responsible for ensuring that the system
complies with the requirements of the system standard IEC 60601-1. If in doubt, consult DRTECH technical support
representative.
DRT-MAN-092 21
EVS 2430Wi User’s Manual 3. Production Description
DRT-MAN-092 22
EVS 2430Wi User’s Manual 3. Production Description
EVS-USB01
USB Switch Box
Supported in USB Mode
Document : User ‘s Manual (PDF)
CD (Software / Calibration) Calibration Data : MAP, PMP, GMP
Software : Econsole1, Ecali1
DRT-MAN-092 23
EVS 2430Wi User’s Manual 3. Production Description
Item Specification
Item Description
Ratio 10 : 1
Inter spacer Al
DRT-MAN-092 24
EVS 2430Wi User’s Manual 3. Production Description
DRT-MAN-092 25
EVS 2430Wi User’s Manual 4. Components and Specifications
Item Description
Pixel Pitch 76 ɥm
Dimensions (㎜) [±0.5 mm] 267.5 (H) × 327.5 (V) × 14.9 (D)
Weight 1.8 kg
Environmental Requirements
Temperature: 0 to +40℃
Storage and Transportation(unpacked) Humidity: 10 to 90% (Without Condensing)
Atmospheric pressure: 500 to 1060 hPa
†Tether Interface:
Allows the detector to communicate with SSU via Ethernet cabling when wireless communications is not available or
when higher speed data transfer is necessary
DRT-MAN-092 26
EVS 2430Wi User’s Manual 4. Components and Specifications
Captured images are transmitted to the EVS 2430Wi image-capture computer using the wireless/wired data transfer
B. Battery Pack : Supplies electrical power to the detector while using wireless communication
C. Satatus indicators
• Power : Shows power on/off status of the detector.
• Ready : Shows data communication status and ready status of detector
(Lights on indicates that detector is busy, lights off means detector is ready)
• Link : Shows detector’s registraion and connection status.
• AP : Lamp indicating Wired/ Wireless mode (2.4 GHz / 5 Hz)
G. Debugging Cover
Don’t remove G and H. if metal objects come into case, it can lead to
WARNING product malfunction..
The operator should not contact the inside of the access cover(G and H)
DRT-MAN-092 27
EVS 2430Wi User’s Manual 4. Components and Specifications
Item Description
Model EVS-BCS
Weight 0.5 kg
A D
D. DC Input : Connect the DC adapter to supply electrical power to the battery charger
DRT-MAN-092 28
EVS 2430Wi User’s Manual 4. Components and Specifications
Item Description
Model EVS-MBP-Y
Weight 0.24 kg
DRT-MAN-092 29
EVS 2430Wi User’s Manual 4. Components and Specifications
• Connect the power adapter to the DC Input port of the battery charger. The power LED lights in blue indicates the
presence of direct current (DC) power.
• Insert the battery pack into the battery charger. Charging starts automatically. The charge LED lights appear
green when the battery pack is being charged. When battery pack is completely charged, all levels of chare LEDs
will illuminate.
• Gently pull the charged battery pack to remove from the battery charger.
Securely plug the power cord into the power source. If contact failure
WARNING occurs, or if dust or metal objects come into contact with the exposed metal
prongs of the plug, fire or electrical shock may occur.
Be sure to stop charging the battery pack when the charge LED lights
CAUTION appear in green beyond the specified charging time. Not doing so may result
in battery pack overheating or smoke emission, or battery explosion or fire.
You must use the power adaptor that is certified with IEC 60950 or IEC
CAUTION 60601-1.
It takes approximately two hours to fully charge a battery pack. The required
charging time may vary depending on the temperature and remaining
battery level.
4.2.3. Adaptor
Item Description
Model AHM85PS12
Maker XP POWER
DRT-MAN-092 30
EVS 2430Wi User’s Manual 4. Components and Specifications
DRT-MAN-092 31
EVS 2430Wi User’s Manual 4. Components and Specifications
1. EVS-WPCS Base
+ Tx Module + Rx Module
- PCB size : 55(mm)x45.5(mm) - PCB size : 65(mm)x47(mm)
- Coil size : 52.5(mm)x52.5(mm) - Coil size : 45.6(mm)x45.6(mm)
- Coil inductance : 11.7uH (MP A2) - Coil inductance : 7.3uH
1. RX Module Base
2. Ferrite Sheet
3. Coil
DRT-MAN-092 32
EVS 2430Wi User’s Manual 4. Components and Specifications
5. Connector : Power Output Connector.
Cradle Carging
Charger
(Wall or Stand
Air State
(No Charging /
Using)
Stand Charging
Mobile
DRT-MAN-092 33
EVS 2430Wi User’s Manual 4. Components and Specifications
DRT-MAN-092 34
EVS 2430Wi User’s Manual 5. Operating Procedure
5. Operating Procedure
General Workflow
The following workflow indicates the procedures after startup of EConsole1 and other system equipments
3)
:ㆍ Selection of EVS 3643 wireless
2. Register1) the detector and make connection2) to the from the Exprimer series detector
EConsole1 power supply to the detector ㆍ Selection of wireless/wired
data transfer
3. Conduct Examination
DRT-MAN-092 35
EVS 2430Wi User’s Manual 5. Operating Procedure
Be sure to use only the dedicated power supply for the EVS-2430WI
CAUTION detector
A. Wireless Connection
- EVS 2430Wi transmits images and data by wireless communication.
- A battery pack should be installed in the detector to use it under the wireless configuration.
- Up to 2 battery packs can be charged simultaneously from a battery charger.
B. Wired Connection
- Connect EVS 2430Wi and PC with a tether interface & Functional cable to make a wired
configuration.
- As the Functional cable supplies power, a battery pack is not needed to be installed in the
detector.
- Data communications are faster than wireless connection.
- It is able to keep charging a battery pack while using the detector.
DRT-MAN-092 36
EVS 2430Wi User’s Manual 5. Operating Procedure
Securely attach the battery into the detector or charger. If contact failure
WARNING occurs, or if dust or metal objects come into contact with the exposed
connector pins of the detector or charger, fire or electrical shock may occur
Make sure to turn off the detector before detaching a battery pack. Press
WARNING and hold the power button for about 2 seconds. All status LED lamps turned
off indicates the detector is turned off.
DRT-MAN-092 37
EVS 2430Wi User’s Manual 5. Operating Procedure
5.1.2.3.1.1. Attachment
5.1.2.3.1.2. Detachment
1) Put the finger into the groove on the charger and grab the battery pack.
2) Pull up the battery pack.
DRT-MAN-092 38
EVS 2430Wi User’s Manual 5. Operating Procedure
5.1.2.3.2.1. Attaching
5.1.2.3.2.2. Detaching
DRT-MAN-092 39
EVS 2430Wi User’s Manual 5. Operating Procedure
This section describes how to connect the EVS 2430Wi system (Detector)
5.2.1.1. Operating AP
1) Connect the LAN cable to ethernet port (not internet) of AP, and the other to the LAN Card connector of
workstation assigned for the data transfer.
DRT-MAN-092 40
EVS 2430Wi User’s Manual 5. Operating Procedure
DRT-MAN-092 41
EVS 2430Wi User’s Manual 5. Operating Procedure
5.2.1.2. Functional Cable
This section describes how to connect the EVS 2430Wi system (Detector) without SSU by using functional cable.
LAN
Power
Tether
Tether
Tether
DRT-MAN-092 42
EVS 2430Wi User’s Manual 5. Operating Procedure
2) Connect the LAN cable to the LAN Card connector of workstation assigned for the data transfer
3) Connect the power cable to other side of functional cable to supply power
DRT-MAN-092 43
EVS 2430Wi User’s Manual 5. Operating Procedure
2) Register the detector and make connection to the EVS control system
i. Registration
When the AP lamp is blinking 1 time in 2seconds, system is in wired mode status.
When the AP lamp is blinking 2 times in 2seconds, system is in wireless mode (AP_1) status.
When the AP lamp is blinking 3 times in 2seconds, system is in wireless mode (AP_1) status.
When the AP lamp is blinking 4 times in 2seconds, system is in wireless mode (detector AP) status.
When the AP lamp is blinking 6 times in 2sec, system is in wireless mode (portable mode) status.
**User can set value of AP_1, AP_2, in Ecali Program. Please refer to Operation Manual for Ecali1
(Calibration tool).
ii. Connection
The LINK lamp does not light up when the detector is not registered or the communication connection is
not established.
When the READY and AP lamp is blinking and LINK lamp does not light up, a communication error has
occurred.
Please refer to troubleshooting.
DRT-MAN-092 44
EVS 2430Wi User’s Manual 5. Operating Procedure
3) Conducting Examination
For details about operation, refer to the Operation Manual for the EConsole1.
• Arrange the patient in the correct posture and position the detector aligning it with the target body part.
Make sure that two lamps (POWER and LINK) are lit and AP lamp is blinking. This means that the system
is ready to start an examination.
A communication error has occurred when LINK lamp lights are off.
When the READY lamp (orange color) blinks slowly, the detector is in detector selection status (Sleep).
The detector enters detector selection status automatically when it has not been used for a certain
period of time.
DRT-MAN-092 45
EVS 2430Wi User’s Manual 5. Operating Procedure
Mode Description
AED Auto Trigger Mode.
USB hand switch USB Switch Mode.
Sterilize the portion of the detector that has been in contact with a patient to
IMPORTANT prevent infection
DRT-MAN-092 46
EVS 2430Wi User’s Manual 5. Operating Procedure
A signal strength indicator appears on the screen of the ECali1 computer. It shows the wireless
communication level between the detector and ECali1.
Keep the wireless communication level stable on capturing or transmitting images.
Wireless, high
Normal
(Stable)
Wireless, Normal
Normal
(Stable)
Check whether there is any obstacle (e.q., your
hands) between the wireless module and the
Unstable wireless access point.
Wireless, Low communication.
If there is any obstacle, remove it.
(Unstable) Communication
speed is lowered If the problem cannot be resolved, ask for
consultation to your sales representative or local
DRTECH dealer.
Confirms that detector and the access point are
turned on.
No signal or No Link Disconnected
If the problem cannot be resolved, ask for
(Communication failed) communications
consultation to your sales representative or local
DRTECH dealer.
Active
DRT-MAN-092 47
EVS 2430Wi User’s Manual 5. Operating Procedure
DRT-MAN-092 48
EVS 2430Wi User’s Manual 5. Operating Procedure
If user does not use acquisition mode after reconnection, image data is saved as file. Image file cannot be
retransmitted automatically even if acquisition mode is activated.
For details about operation, refer to the Operation Manual for the EConsole1.
DRT-MAN-092 49
EVS 2430Wi User’s Manual 5. Operating Procedure
Power ON ○ X X X X
Communication established ○ X X ○ ★
Sleep Mode ○ X ☆ ○ ★
Power OFF X X X X X
DRT-MAN-092 50
EVS 2430Wi User’s Manual 6. Extension Facility
6. Extension Facility
6.1. X-ray Generator Interface
6.1.1. X-ray Exposure Mode
Mode Description
1. The detector detects actual amount of X-rays without any connection to the X-ray
generator, and then performs image acquiring to the extent of image acquisition time
and transmits the image data.
Auto Trigger Mode (AT)
2. No signal used (No need of connector interface cable)
• You can use AT mode without connecting the generator with USB SW Box or SSU
physically.
1. The detector receives EXP_REQ signal that X-ray generator is prepared to generate
X-rays.
2. The detector prepares image acquiring and then responds EXP_OK signal to the X-
ray generator.
Sync. Trigger Mode
3. The X-ray generator confirms EXP_OK signal and generates X-ray, then the detector
performs image acquiring, according to image acquisition time and transmits the
image data.
• EXP_REQ (Generator→ Detector), EXP_OK (Detector →Generator)
1. EXP_IN signal generates by USB SW Box then the detector receives Ready signal.
And simultaneously, X-ray generator have ready status in
2. The detector prepares image acquiring
USB SW Mode 3. EXP_IN signal generates by USB SW Box and generates X-ray, then the detector
performs image acquiring, according to image acquisition time and transmits the
image data.
• Ready_IN (USB SW Box -> PC -> Detector and X-ray Generator)
• EXP_IN (USB SW Box -> PC -> Detector and X-ray Generator)
DRT-MAN-092 51
EVS 2430Wi User’s Manual 6. Extension Facility
DRT-MAN-092 52
EVS 2430Wi User’s Manual 6. Extension Facility
DRT-MAN-092 53
EVS 2430Wi User’s Manual 6. Extension Facility
- The USB interface mode is a wired connection of wireless detector with SSU.
- PC and detector are connected by tether interface cable.
(consistent power supplying by adoptor).
- It is used in case when faster image transmission is needed, compared to a wireless connection.
- A wireless module in the detector is deactivated. (The equipped battery is not consumed)
- Connect the generator with USB SW Box physically to make an exposure in USB SW mode.
- You can use AT mode without connecting the generator with USB SW Box physically.
- Multiple detectors can be configured with one USB Box.
DRT-MAN-092 54
EVS 2430Wi User’s Manual 6. Extension Facility
Adapter
(Optional)
Hand
Switch
1m
3m
X-Ray
Consol
e
Figure 6.13. Wiring USB Switch box
DRT-MAN-092 55
EVS 2430Wi User’s Manual 6. Extension Facility
6.1.3.4. X-ray_UNIT
6.1.3.6. PC USB
DRT-MAN-092 56
EVS 2430Wi User’s Manual 6. Extension Facility
Software Description
Software development kit for EVS 2430Wi detector only, provided by DRTECH
EVS SDK
You can develop your own software dedicated to EVS 2430Wi by using this kit.
• For a client who uses EVS SDK, Install the Setup program after reading EVS SDK Developer’s Manual
DRT-MAN-092 57
EVS 2430Wi User’s Manual 6. Extension Facility
Communication disruption between the detector (Or SSU) and workstation occurs unless the
network adaptor is set properly. It may cause serious repercussion to the product and image
quality.
CAUTION
Network Adaptor Selection
1) Click Start → Control Panel → Network and Internet → Network and Sharing Center → Change
Adapter Setting.
2) Choose the networks adaptor for communicating with the detector and SSU, and then rename it.
3) Click the chosen network adaptor with the right mouse button and click Properties to display the Properties
window.
DRT-MAN-092 58
EVS 2430Wi User’s Manual 6. Extension Facility
1) Click Configure button to open the following dialog box, and then go to the Advanced tab.
2) Choose Flow Control in the list of Properties and click Value button on the right.
- Click the Power Management tab and check on Allow the computer to turn off this device to save
power.
- Click OK button
DRT-MAN-092 59
EVS 2430Wi User’s Manual 6. Extension Facility
- Input the IP address and subnet mask as shown below, and then click OK button.
- It is recommended to set IP address and Subnet mask in the presented range of this
document.
- If you use IP address and Subnet mask out of the presented range, it would be difficult to
CAUTION identify the cause or resolve in case of communication disorder.
DRT-MAN-092 60
EVS 2430Wi User’s Manual 6. Extension Facility
1) Click Start → Control Panel → Power Options and then move to the Choose when to turn off the display
tab.
2) Set Put the computer to sleep to Never to disable the sleep mode.
DRT-MAN-092 61
EVS 2430Wi User’s Manual 7. Device Setting
7. Device Setting
7.1. AP Setting
7.1.1. AP Configuration
Normally, AP setting does not need to be changed by user, because AP is set to match the use environment when the
product is inspected for shipping.
1) Check IP address of AP
- Register wireless Network Name same with the detector, and select Region, Channel Width, Channel
depending on the use enviroment.
DRT-MAN-092 62
EVS 2430Wi User’s Manual 7. Device Setting
DRT-MAN-092 63
EVS 2430Wi User’s Manual 7. Device Setting
Menu Description
Path to save map files Register each map file that corresponds to the product
IP address of AP.
Router IP
The detector can not be connected to AP if router IP is changed.
Grid selection Select type of the grid and register the path of grid filter.
Waiting time for sleep mode can be adjusted by setting parameters below.
1) Acquisition sleep time
Power management Mode is changed from acquisition mode to sleep mode, after the set waiting time.
2) Deep sleep time
Mode is changed from sleep mode to deep sleep mode, after the set waiting time.
DRT-MAN-092 64
EVS 2430Wi User’s Manual 7. Device Setting
Image TimeOut
Item Description
After starting image transmission, the detector ignores received information of image transmission request, if the
following conditions are all met.
Sleep Detector can be operated and take an exposure at any time within 2 seconds.
Deep Sleep Detector can not be operated. User can take an exposure only if Deep Sleep Mode is canceled.
Power Off Detector has been turned off. User can take an exposure after detector is rebooted.
- You can prevent unnecessary battery consumption by using the Deep Sleep function.
- When SSU power is supplied to the detector by connecting a tether interface cable, the Deep Sleep function
cannot be operated.
- When using the Deep Sleep function, be sure to check if the detector is in Deep Sleep mode before making an
exposure. You cannot acquire images when the detector is in Deep Sleep mode.
- When the Deep Sleep mode is disabled, the detector needs maximum of 10seconds to wake up. It may not be
available to acquire images during this time.
DRT-MAN-092 65
EVS 2430Wi User’s Manual 7. Device Setting
Meaning
Sleep Detector can be operated and take an exposure at any time within 2 seconds.
Deep Sleep Detector can not be operated. User can take an exposure only if Deep Sleep Mode is canceled.
Power Off Detector has been turned off. User can take an exposure after detector is rebooted.
Mode Description
Normal -
Sleep Activates when the detector has not been used for a certain time.
Time can be set to automatically go into Deep Sleep mode, after a certain period of time passed
Deep Sleep
under sleep mode.
Detector is turned off if detector has not been used for a set period of time (Power off) from the
Power Off entry time of Deep Sleep Mode. However, in case of not using Deep Sleep mode function,
Detector is turned off if detector has not been used for certain time (Power Off After).
Mode Description
Normal -
Power Off Reboot the detector with pressing power button on the detector.
DRT-MAN-092 66
EVS 2430Wi User’s Manual 7. Device Setting
Other Information
Normal -
- When using the Sleep function, be sure to check if the detector is in Sleep mode before
making an exposure. You cannot acquire images when the detector is in Sleep mode.
- When the Sleep mode is disabled, the detector needs max. 1 to 2 seconds to wake up. It
CAUTION may not be available to acquire images during this time.
DRT-MAN-092 67
EVS 2430Wi User’s Manual 8. Troubleshooting
8. Troubleshooting
8.1. Failed to Turn the Detector On
Symptom
Possible Causes
Solutions
3) Check the result after getting rid of the battery pack and connecting the tether interface cable.
Symptom
Possible Cause
Solutions
DRT-MAN-092 68
EVS 2430Wi User’s Manual 8. Troubleshooting
Symptom
Possible Cause
Solutions
4) Check the windows and network environment again such as firewall and IP address setting.
5) Check if the SSID of Detector is same as SSID of AP by using tether interface cable.
Symptom
Possible Causes
Solutions
1) Replace to new battery pack if the battery pack has been used for a long time. (Battery pack is a consumable)
2) Use battery pack in normal room temperature environment. Charging capacity of battery pack in low
temperature environment will decrease the capacity.
DRT-MAN-092 69
EVS 2430Wi User’s Manual 8. Troubleshooting
Symptom
Possible Causes
- Detector Failure
Solutions
Symptom
Possible Causes
Solutions
DRT-MAN-092 70
EVS 2430Wi User’s Manual 9. Maintenance and Inspection
Daily Inspection
For safety reasons, be sure to turn OFF the power to each piece of
WARNING equipment before the following procedures. Otherwise, an electric shock
may result.
Cable
1) Ensure that cables are not damaged and cable jackets are not torn.
2) Ensure that the power cord plugs are securely connected to both the equipment AC inlet and the AC outlet.
Detector
1) Ensure that there are no loose screws or broken parts.
3) Ensure that there are no broken parts or short-circuits in the power supply conector.
Monthly Inspection
1) Ensure that there are no loose screws or broken parts.
Yearly Inspection
1) Perform a Performance Test or Self-diagnosis using a phantom or resolution chart, etc.
Irregular Inspection
Calibration
DRT-MAN-092 71
EVS 2430Wi User’s Manual 10. Specification
10. Specification
10.1. Main Specifications
10.1.1. EVS 2430Wi / EVS 2430GWi X-ray Detector
DRT-MAN-092 72
EVS 2430Wi User’s Manual 10. Specification
10.1.2. Battery Charger System
DRT-MAN-092 73
EVS 2430Wi User’s Manual 10. Specification
DRT-MAN-092 74
EVS 2430Wi User’s Manual 10. Specification
10.1.4. EVS-WPCS
DRT-MAN-092 75
EVS 2430Wi User’s Manual 10. Specification
10.2. Charateristics
- EVS 2430Wi Wireless responds linearly against the exposure range for 500-1000 speed film/screen where
it can depict the clinical information. It means that EVS 2430Wi fully covers a dynamic range of 0.2-20 μGy
at least.
- EVS 2430Wi
A typical MTF value at 2 lp/mm, than 40%.
- EVS 2430GWI
A typical MTF value at 2 lp/mm, than 25%.
DQE
- EVS 2430Wi
A typical DQE value at 1 lp/mm is 55%.
- EVS 2430GWI
A typical DQE value at 1 lp/mm, than 50%.
The product safety standards that apply to the EVS 2430Wi, which includes the following equipments, are as
below.
• Detector
• Battery Charger
• Battery Pack
• WPCS Module
DRT-MAN-092 76
EVS 2430Wi User’s Manual 10. Specification
10.3. Packing
Note
Figures and Illustrations in this Technical Manual are provided for reference only and may differ from the actual
product appearance.
Note
If you find any items missing from the list above upon unpacking, please contact your dealer.
DRT-MAN-092 77
EVS 2430Wi User’s Manual 10. Specification
Cushion
Cushion
Detector Panel Package
Warning
Operational issues may occur if inappropriate force is applied to the product during unboxing. Please handle the box
containing the product with care.
Note
The packaging that came with the product should not be damaged or discarded as it is needed for after-sales service
or for exchanging the product with a new product. If the packaging or any component is missing, the damaged product
cannot be refunded or exchanged with a new product.
DRT-MAN-092 78
EVS 2430Wi User’s Manual 10. Specification
Top cushion
PCM Pack
Inner form
600 Detector
PCM Pack
Bottom cushion
Note
• The PCM pack can explode from a hard impact or a contact with a sharp object.
• It is non-toxic and may not cause serious harm. However, please take extra precaution as it is a chemical
substance, and remember to carefully follow the instructions.
• If the inner chemicals get on your body, immediately wash it off thoroughly with clean water.
• If the inner chemicals get in your eyes, wash your eyes with tap water for more than 15 minutes and consult your
doctor.
• If the contents are accidentally ingested, drink two glasses of water (500ml) and consult your doctor.
• How to discard the content: Please discard used PCM packs using one of the following methods.
• Collect a certain number of packs and send them to the manufacturer.
• Treat them in a nearby chemical waste processing facility.
Note
• PCM Pack [Phase Change Material Pack]
• The PCM pack placed in the Bottom cushion is intended to maintain mild temperature over a certain period of time
in order to protect Detector from cold weather.
• The PCM is originally a liquid, and it begins to emit heat when the box is left for a long period at low temperature
below zero Celsius. Its phase changes to a solid state gradually during the heat radiation.
• The PCM is a reusable material, which means the solid state PCM returns to its liquid state within approximately 2
hours of heating over 35 degrees Celsius. Utilizing this method, the PCM can be used repeatedly.
• The liquid PCM should be stored around room temperature, and can be reused in the same way as the original
packing when any transportation is needed.
DRT-MAN-092 79
EVS 2430Wi User’s Manual 10. Specification
1 Access Point 1
3 Functional Cable 1
5 USB Cable 1
Optional
6 Hand Switch 1
7 X-Ray Cable 1
1 Battery Pack 2
2 Battery Charger 1
3 Software CD 1
4 Power Adaptor 1
Default
5 Power Code 1
6 LAN Card 1
7 LAN Cable 1
8 Adaptor Cable 1
DRT-MAN-092 80
EVS 2430Wi User’s Manual 11. Regulatory Information
Declaration of Conformity
U.S.A. FCC Part 15 Subpart B Class A and Part 15 Subpart C (RF Exposure)
Canada RSS-210
• FCC/IC Compliance
This device complies with Part 15 of the FCC Rules and RSS-Gen of IC Rules.
ii. This device must accept any interference received, including interference that may cause undesired
operations.
• FCC ID
Note :
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of
the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular Installation.
If this equipment dose cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encourage to try to correct the interference by one or more of the following
measures:
DRT-MAN-092 81
EVS 2430Wi User’s Manual 11. Regulatory Information
• FCC WARNING:
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
This transmitter must not be co-located or operated in conjunction with any other antennas or transmitters.
• RF Exposure Compliance
The available scientific evidence does not show that any health problems are associated with using low power
wireless devices. There is no proof, however, that these low power wireless devices are absolutely safe. Low
power wireless devices emit low levels radio frequency energy (RF) in the microwave range while being used.
Whereas high levels of RF can produce health effects (by heating tissue), exposure of low-level RF that does
not produce heating effects causes no known adverse health effects. Many studies of low-level RF exposures
have not found any biological effects. Some studies have suggested that some biological effects might occur,
but such findings have not been confirmed by additional research. EVS 2430Wi Wireless has been tested and
found to comply with FCC/IC radiation exposure limits set forth for an uncontrolled equipment and meets the
FCC radio frequency (RF) Exposure Guidelines in Supplement C to OET65 and RSS-102 of the IC radio
frequency (RF) Exposure rules.
• Disposal
Disposal of this product in an unlawful manner may have negative effects on health and on the environment.
When disposing this product, therefore, be absolutely sure to follow the procedure which is in conformity with
the laws and regulations applicable in your area.
DRT-MAN-092 82
EVS 2430Wi User’s Manual 11. Regulatory Information
11.2.1. Detector
DRT-MAN-092 83
EVS 2430Wi User’s Manual 11. Regulatory Information
Detector Case
(257 x 336.5 x 14.9 mm)
Detector Case
(257 x 336.5 x 14.9 mm)
DRT-MAN-092 84
EVS 2430Wi User’s Manual 11. Regulatory Information
DRT-MAN-092 85
EVS 2430Wi User’s Manual 11. Regulatory Information
11.2.3. EVS-WPCS
DRT-MAN-092 86
EVS 2430Wi User’s Manual 11. Regulatory Information
Non-ionized radiation
The Waste Electrical and Electronic Equipment Regulations indicates separate collection for
electrical and electronic equipments.
Certification mark that indicates the product complies UL 61010-1 (2012), CSA C22.2 No.
61010-1-12-CAN/CSA (2012) that specifies protection against fire, electric shock, and
mechanical hazards.
This mark shows compliance with the equipment with R&TTE Directive 1999/5/EC
Direct Current
Alternating Current
Equipotentially.
Stand-by
Read and understand all instructions and warning labels in the product documentation before
using the equipment.
Keep manual for future reference.
Splashing of water
IPX4 This applies to EVS 2430Wi only.
DRT-MAN-092 87
EVS 2430Wi User’s Manual 11. Regulatory Information
The EVS 2430Wi System is used in the following electromagnetic settings. Users of the EVS 2430Wi System should
check whether their systems are used in these settings.
Since the EVS 2430Wi System only uses RF energy for internal
RF emission
Group 1 functions, it has very low RF emissions and normally causes no
CISPR 11
interference to neighboring electronic devices.
RF emission
Class A
CISPR 11
Voltage changes
/flicker emission Compliance
CISPR 11
DRT-MAN-092 88
EVS 2430Wi User’s Manual 11. Regulatory Information
Full Compliance to the IEC 60601-1-2:2007 and the System’s Tolerance to EM Waves
Users of the EVS 2430Wi System should check whether their systems are used in these settings.
<5% UT <5% UT
(<95%Dip at the UT), (<95%Dip at the UT),
0.5 cycles 0.5 cycles The main power’s quality should be equal to
general commercial or user settings.
40% UT 40% UT
Voltage loss in the
(60% Dip at the UT), (60% Dip at the UT),
power supply, short Note :
5 cycles 5 cycles
intermittence and Most components in the EVS 2430Wi
voltage changes 70% UT 70% UT System have their power supplied from the
(30% Dip at the UT), (30% Dip at the UT), uninterrupted power supply.
IEC 61000-4-11
25 cycles 25 cycles
The IEC61000-4-11 only applies to the EVS
<5% UT <5% UT 2430WI System Power Box.
(>95% Dip at the UT), (>95% Dip at the UT),
5 seconds 5 seconds
Note :
The UT is the main AC voltage before the test standards have been applied.
DRT-MAN-092 89
EVS 2430Wi User’s Manual 11. Regulatory Information
Users of the EVS 2430Wi System should check whether their systems are used in these settings.
Recommended intervals :
d = 1.17√p
d = 1.17√p80MHz to 800MHz
Conductive RF
3Vrms 150kHz-80MH 3Vrms d = 2.33√p800MHz to 2.5GHz,
IEC61000-4-6
Radioactive RF where p is the transmitter-receiver’s maximum power
3v/m 80MHz-2.5GHz 3v/m
IEC61000-4-3 rating in watts (W) and d is the recommended interval.
Note 1 :
The high-frequency range is applied at 80MHz to 800MHz.
Note 2 :
This guideline does not apply in all situations. Electromagnetic waves may be affected
through absorption and reflection from structures, objects and people.
DRT-MAN-092 90
EVS 2430Wi User’s Manual 11. Regulatory Information
It is very difficult to accurately predict the magnetic field strength of wireless (mobile/wireless) telephones, land
mobile radio base stations, amateur wireless channels, AM, FM wireless and TV broadcasting systems. To assess
electromagnetic settings using fixed RF receivers, area walk-down is needed. If the magnetic field strength
measured at the point, where the EVS 2430Wi System is used, exceeds the applicable RF compliance level, you
should check whether the EVS 2430Wi System is operating normally. If any performance abnormality is observed,
additional action may be needed such as changing the EVS 2430Wi System’s direction or location. At the frequency
range between 150kHz and 80MHz, the magnetic field strength should be less than 3v/m.
Recommended Intervals between the EVS 2430Wi System and the Portable
or Mobile RF Communications Equipment
The EVS 2430Wi System should be used in an electromagnetic setting where RF communication interferences are
controlled. Users of the EVS 2430Wi System should maintain the minimum intervals between the System and the
portable or mobile RF communications equipments to prevent electromagnetic interferences more effectively.
For maximum power voltages of receivers are not shown in the above list, the recommended interval, d(m), can be
calculated by using the equation used for the receiver’s frequency. The p is the transmitter-receiver’s maximum
power rating in watts (W).
Note 1 :
The high-frequency range is applied at 80MHz to 800MHz.
Note 2 :
This guideline does not apply in all situations. Electromagnetic waves may be affected through absorption and
reflection from structures, objects and people.
DRT-MAN-092 91
EVS 2430Wi User’s Manual 12. Warranty
12. Warranty
DRTECH Corporation warrants that this product will be free from defects in materials and workmanship for a period of
twelve (12) months from the date of delivery. If any such product proves to be defective during this warranty period,
DRTECH Corporation at its options, will repair the defective product without charge for parts and labor, or will provide
a replacement in exchange for the defective product. In order to obtain service under this warranty, customer must
notify DRTECH Corporation of the defect before the expiration of the warranty period and make suitable arrangements
for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the
service center designated by DRTECH Corporation with shipping charges prepaid. DRTECH Corporation shall pay for
the return of the product to customer if the shipment is to a location within the country in which the DRTECH
Corporation designated service center is located. Customer shall be responsible for paying all shipping charges,
duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure, or damage caused by improper or inadequate maintenance and
care. DRTECH Corporation shall not be obligated to furnish service under this warranty to repair damage resulting
from attempts by personnel other than DRTECH Corporation or its representatives to install, repair, or service this
product, to repair damage resulting from improper use or connection to incompatible equipment or power source; or to
service a product that has been modified or integrated with other products when the effect of such modification or
integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY DRTECH Corporation WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. DRTECH Corporation AND ITS VENDOR DISCLAIM ANY IMPLIED
WARRANTIES OF MERCHANTABLILITY OR FITNESS FOR A PARTICULAR PURPOSE. DRTECH Corporation
RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE REMEDY PROVIDED TO
THE CUSTOMER FOR BREACH OF THIS WARRANTY. DRTECH AND ITS VENDORS WILL NOT BE LIABLE FOR
ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER
DRTECH Corporation OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
There are no warranties which extend beyond the description mentioned in this document.
DRT-MAN-092 92
EVS 2430Wi User’s Manual 12. Warranty
Revision History
DRT-MAN-092 93
EVS 2430Wi User’s Manual 12. Warranty
DRTECH Corporation.
Suite No.2 3 Floor, 29
Dunchon-daero541beon -gil,
Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea
Web Site : http://www.drtech.co.kr
Customer Support Team
E-mail : drtech@drtech.co.kr
Tel : +82 31 730 6805 / Fax : +82 31 730 6899
EXAMION GmbH
Erich-Hherion Stuttgart. 37 in 70736 Fellbach. Germany
DRT-MAN-092 94