Lab Activity1
Lab Activity1
Introduction
"Micro" refers to tiny, "scope" refers to view or look at. Microscopes are tools used to enlarge images of
small objects so as they can be studied. The compound light microscope is an instrument containing two lenses,
which magnifies, and a variety of knobs to resolve (focus) the picture. Because it uses more than one lens, it is
sometimes called the compound microscope in addition to being referred to as being a light microscope. In this
lab, we will learn about the proper use and handling of the microscope.
Instructional Objectives
Demonstrate the proper procedures used in correctly using the compound light microscope.
Prepare and use a wet mount.
Determine the total magnification of the microscope.
Explain how to properly handle the microscope.
Describe changes in the field of view and available light when going from low to high power using the
compound light microscope
Explain why objects must be centered in the field of view before going from low to high power using the
compound light microscope.
Explain how to increase the amount of light when going from low to high power using the compound light
microscope.
Explain the proper procedure for focusing under low and high power using the compound light microscope.
Materials
Compound microscope
Glass slides
Cover slips
Eye dropper
Beaker of water
The letter "e" cut from newsprint
Scissors
Pre-Lab
1. Each student is assigned a numbered microscope. That specific microscope must be used every time by
that student.
a. True b. False
2. Each individual student must read and complete a Student Microscope Checklist each time a microscope
is used.
a. True b. False
1. Carry the microscope with both hands --- one on the arm and the other under the base of the microscope.
2. One person from each group will now go over to the microscope storage area and properly transport one
microscope to your working area.
3. The other person in the group will pick up a pair of scissors, newsprint, a slide, and a cover slip.
4. Remove the dust cover and store it properly. Plug in the scope. Do not turn it on until told to do so.
5. Examine the microscope and give the function of each of the parts listed on the right side of the diagram. Use a
separate sheet to list and define the function of each part of the microscope.
3. X-Y mechanical stage control knobs: These knobs, under the left side of the stage, move the slide
horizontally (lower X knob) and vertically (upper Y knob). Never force any knobs.
4. Lamp.
The lamp on/off switch is located in the rear on the base. The light intensity is adjusted with the light
intensity regulator on the right side of the base. The lamp should be adjusted to a medium level at the
start of viewing.
5. Lens systems:
There are three lens systems: the eyepieces (ocular), the objectives (four), and the condenser.
a. Eyepieces (ocular): Magnification of 10X. The eyepieces are held rather loosely in the
eyepiece tubes. Never remove the eyepieces from the eyepiece tubes. A rubber eye
shield should be on the top of each eyepiece. The width between the eyepieces should be
moved until a full circle (the viewing field) is visible with both eyes simultaneously.
Students can record their proper setting by reading the scale between the eyepieces.
b. Objectives: There are four objectives: 4X, 10X, 40X, and the 100X (oil immersion
objective). The objectives are attached to a rotatable nosepiece.
c. Condenser: The condenser is located directly beneath the stage. It gathers and conducts
the light to the specimen. The condenser height adjustment knob (on the right side of the
scope arm below the stage) should be moved clockwise until the condenser is in the
highest position. If the field of view is not uniform enough, it may be improved by
lowering the condenser slightly.
The iris diaphragm controls the amount of light from the condenser and is adjusted
with a thin lever under the stage.
The larger coarse adjustment knob moves the stage up and down much faster and farther than the smaller
fine adjustment knob. The coarse adjustment knob is used ONLY with the lower power (4X, 10X)
objectives. When focusing under the 40X or 100X objective, ONLY use the fine adjustment, never the
coarse adjustment.
Preparation for viewing- Refer to each of the following procedures every time the microscope is
used. Read FIRST, then perform the step.
1. If you have not done so, carefully remove a scope from the cabinet and set it on your lab bench.
Remove the dust cover. Check that no parts are loose or missing). They might be in the dust cover.
Immediately contact the instructor if parts are missing or anything is wrong with the scope. Write
down the number of your scope (found on the front of each scope) on the microscope checklist at the
end of the lab exercise instructions.
2. Plug in the microscope. Be sure the electrical cord is not dangling off of the lab bench and cannot
become entangled.
3. Use the coarse adjustment knob to move the stage to its lowest level.
4. Cleaning the microscope. Clean all objective lenses with swabs and liquid lens cleaner.
Place a small amount of liquid lens cleaner on the swab. Use a fair amount of pressure to clean the
lens in a circular pattern. Dry the lens with the dry, unused end different swab. Use different swabs
for each lens.
Do not use liquid lens cleaner on the eyepieces. Dust on the eyepieces, or elsewhere, may be
removed using the blower brush. NEVER remove any parts.
5. Before use, clean all slides, top and bottom, with Kimwipes. Place a coverslip over the specimen on
the slide. Do not use a coverslip if viewing prepared slides.
6. Rotate the nosepiece until the 4X (10X with bacteria) objective clicks into place.
Turn on the lamp. Adjust the on/off switch to a medium intensity.
7. Gently place the slide on the stage so that it is held within the mechanical stage clamping device. The
bow-shaped lever of the mechanical stage clamping device should be pulled outward and released gently
after the slide is placed on the stage within the clamping device.
The lever should not be released in the middle or the slide could be broken. The slide must lie flat on the
stage.
8. Using the mechanical stage X-Y movement control knobs, position the slide so that the specimen is in
the exact center of the light coming through the condenser. Do not use your hands to move the slide; only
use the X-Y control knobs.
Viewing and focusing the specimen under the 4X, 10X, and 40X objectives
1. Always start with a low power objective (4X or 10X when viewing bacteria) clicked into place.
The lower power objectives have the largest field of view (a larger portion of the slide can be seen),
making it easier to initially find the specimen.
2. Only now, with a lower power objective in place, move the stage to its highest point by rotating the
course adjustment know clockwise. In other words, the stage should be moved as close to the objective as
possible.
a. Inter-pupillary distance: While looking through the eyepieces, adjust the width between the eyepieces
(inter-pupillary distance) until a single image is seen simultaneously with both eyes. The left and right
fields of view should coincide completely. The position of the dot indicates the inter-pupillary value.
Note your inter-pupillary value.
b. Diopter adjustment: When you look through a microscope with two eyepiece lenses, you must be
able to change the focus on one eyepiece to compensate for the difference in vision between your two
eyes. The diopter adjustment does this.
1) Rotate the right eyepiece to match your inter-pupillary distance value.
2) Close the eye over the eyepiece with the diopter adjustment (left eye), and normally focus
the microscope so that the open right eye sees the image in focus.
3) Next, switch eyes (close the right eye, open the left eye) and without changing the main
focus knobs, focus on the image by turning the diopter adjustment ring on the left eyepiece
only. Now with both eyes open, the image should be clear with both eyes.
c. Eye Shades
When wearing eyeglasses: Use the eye shades in the normal, folded-
down position. This will prevent the eyeglasses from being scratched.
When not wearing eyeglasses: Extend the folded eye shades outwards
(direction of the arrow) to prevent extraneous light from entering into
your line of vision.
Light intensity is a essential aspect of microscopy. Perform the following steps to adjust the light. For
optimal viewing, the light must be adjusted several times while focusing. Always adjust the light while
looking through the eyepieces.
a. Adjust the light intensity switch that turns the lamp on. Begin at a medium level. Adjust
again while focusing.
b. The entire viewing area (field) must be filled with light. If this is not the case, contact the
instructor.
c. Locate the thin, iris diaphragm lever under the stage condenser. Adjust this lever to a
medium/low light level. The iris diaphragm will need to be adjusted as magnifications increase
and focusing continues.
5. Under low power, SLOWLY focus with the coarse adjustment knob counterclockwise, moving the
stage down, until the specimen comes into view. Adjust the light as instructed in step 3 above.
Many specimens, especially bacteria, are very small and may look like specks at this magnification. If the
stage moves too quickly, you may go past the specimen without seeing it. Look for ‘color’ since the
specimens are stained.
6. Refine the image with the fine adjustment knob and by adjusting the light.
7. Important! Before switching to the next objective, move the slide so that the desired specimen is
located in the center of the field (circular viewing area).
8. SIGNIFICANT POINTS WHEN FOCUSING UNDER THE 4X, 10X, AND 40X OBJECTIVES!
a. Always start with a low power objective (4X or 10X). After focusing and making
appropriate observations, rotate the nosepiece until the next objective clicks into place. Do
not skip objectives.
b. Do not move the stage up or down before rotating the nosepiece to the next objective. When
properly focused, there is no need to adjust the objective's distance from the stage before
increasing the magnification.
d. Before increasing the magnification, always use the X-Y control knobs to move the slide so
that the desired specimen is located in the center of the field.
e. Remember to adjust the light each time the magnification is changed and while
focusing.
1. Be sure that the specimen has been optimally focused using the 40X objective and is in the EXACT
CENTER of the viewing field under the 40X objective.
2. NOTE! Again, do not move the stage up or down before rotating the nosepiece to the 100X
objective.
3. Rotate the 40X objective away from the slide but do not yet click the 100X objective into place.
4. Put a small drop of immersion oil on the slide/coverslip directly over the light.
5. Rotate the nosepiece until the 100X oil immersion objective is clicked into place. If there are oil
bubbles, revolve the nosepiece slightly to remove the bubbles.
6. DO NOT USE THE COARSE ADJUSTMENT KNOB WHEN FOCUSING UNDER THE 100X
OBJECTIVE! ONLY USE THE FINE ADJUSTMENT KNOB!
7. If problems are encountered during viewing, re-clean the objectives and repeat the procedure. If
problems persist, review the Common Problems section at the end of this document.
9. In order to determine the size of your specimen, you will need to estimate utilizing the size of the field
(circular viewing area). When you observe your specimen, estimate its size by comparing it with the size
of the field.
When finished viewing and drawing your specimens, complete the Student Microscope
Checklist. Turn it in with your lab report.
Common Problems
2. You are not sure if you are looking at dirt on the objective lens or the specimen.
Use the mechanical stage knobs to move the slide slightly while looking through the eyepieces. If what
you are looking at does not move, it is probably dust or dirt on the objective. If it does move, it is on the
slide.
Rotate the ocular gently between your fingers. If what you are looking at rotates, it is probably dirt on the
ocular.
The stage may be too far from the objective. Did you start with the stage as close as possible to the low
power objective?
Is the specimen directly over the light?
Is the slide secure and flat in the mechanical stage?
Did you start with a low power objective and focus on the lower objectives first?
Have you adjusted the light?
Are you moving the adjustment knobs too quickly? Work slowly so you do not miss the specimen.
Remember, bacteria look like specks at low magnifications.
5. You lose the specimen when switching from the 40X objective to the oil immersion objective.
Was the specimen in the exact center of the field before switching to the 100X objective?
Is the 100X objective lens clean?
Have you adjusted the light?
Have you refined the image with the fine adjustment?
b. Using your left eye only, use the 10X objective and focus on the specimen by adjusting the
coarse adjustment knob.
c. When the image is in view, refine the image to its sharpest focus by turning the fine adjustment
knob.
d. Using both eyes, rotate the right eyetube collar below the eyepiece (dioptric adjustment) until
the sharpest image appears.
8. One of the most common problems when switching objectives several times, especially after
using oil, is that the objectives need RE-CLEANING, and the student must refocus
beginning with the 10X objective again.
Call the instructor after you have focused at 100X on each slide.
Remember to complete the Student Microscope Checklist. Be sure to write the microscope number of the
scope you used on the Checklist. The Checklist and the Lab Report will be returned to the instructor.
1. With your scissors cut out the letter "e" from the newspaper.
2. Place it on the glass slide so as to look like (e).
3. Cover it with a clean cover slip. See the figure below.
4. Using your eyedropper, place a drop of water on the edge of the cover slip where it touches the glass slide. The
water should be sucked under the slide if done properly.
6. Describe the relationship between what you see through the eyepiece and what you see on the stage.
___________________________________________________________________________________________
___________________________________________________________________________________________
____________________________________________________
7. Looking through the eyepiece, move the slide to the upper right area of the stage. What direction does
the image move?
____________________________________________________________________________________
____________________________________________________________________________________
__________________________________________________________________
8. Now, move it to the lower left side of the stage. What direction does the image move?
____________________________________________________________________________________
____________________________________________________________________________________
__________________________________________________________________
9. Re-center the slide and change the scope to high power. You will notice the "e" is out of focus. DO
NOT touch the coarse focus knob, instead use the fine focus to resolve the picture. Draw the image you
see of the letter e (or part of it) on high power.
10. Locate the diaphragm under the stage. Move it and record the changes in light intensity as you do so.
____________________________________________________________________________________
____________________________________________________________________________________
____________________________________________________________________________________
____________________________________________________________
1. Use a clear ruler to determine the width of the viewing field under the low power
objective. Position the ruler so that the millimeter marks are visible in your viewing
field. Remember that there are 1000 micrometers in a millimeter.
Name:
Microscopy
1. Draw and label the letter "e" after focusing with 10X and 40X objectives.
2. Draw and label the crossed hairs after focusing with 10X and 40X objectives.
3.
Draw and label the cheek cells (squamous cells) after focusing with 10X and 40X objectives.
4. If you had enough time to look at a prepared Human Blood Smear, what cells did you
identify?
2. Explain why the light microscope is also called the compound microscope.
3. Images observed under the light microscope are reversed and inverted. Explain what this means.
4. Explain why the specimen must be centered in the field of view on low power before going to high
power.
a) Calculate the low power magnification of this microscope. Show your formula and all work.
b) Calculate the high power magnification of this microscope. Show your formula and all work.
6. In three steps using complete sentences, describe how to make a proper wet mount of the letter e.
7. Describe the changes in the field of view and the amount of available light when going from low to high
power using the compound microscope.
8. Explain what the microscope user may have to do to combat the problems incurred in question # 7.
9. How does the procedure for using the microscope differ under high power as opposed to low power?
Student Microscope Checklist
When finished using the microscope, complete the following.
Check off each step when completed.
Turn the light switch OFF. Do not move the microscope for a few minutes before putting it
away,
Rotate the head so that the eyepieces are facing forward (away from the arm).
Replace the dust cover.
Return the microscope to the appropriate location in the cabinet in the appropriate postion.
Check to be sure all waste is removed from the sink and floor.