Procedure - Testing and Calibration For Process Instrument
Procedure - Testing and Calibration For Process Instrument
Contents
1.0 PURPOSE..............................................................................................................................
1.0 PURPOSE
2.0 PROCEDURE
2.1 General
All instruments and instrument systems should be inspected
immediately upon receipt from the manufacturer/supplier to ensure
they are free from defects.
Any defects or deficiencies shall be brought to the attention of the
principal.
However, certain instruments and instrument systems will still need
to be checked, tested and/or calibrated, either before installation in
the instrument workshop or in the field during the pre -commissioning
inspection
Q-PR-00-059-01
Procedure No. GS-MO-Q-PR-050-01 Page 3 of 9
Testing and Calibration of the Process Instruments
Revision 01, dated 15.12.2003
Carry out the procedure given above for pressure transmitters and in
addition:
Apply a static pressure, corresponding to the maximum operating pressure as
indicated in the requisition, to both sides of the differential pressure cell and
record the variation of zero output on the calibration sheet.
2.3.4 Controllers
Alignment check:
- with the controller in manual operation, apply 50% of the measured
variable signal to input and move set point to 50% of scale. Reset adjustment
to minimum setting and rate adjustment to maximum setting. Set output to 50%
and switch to auto. Adjust the proportional band setting to maximum and
observe the deviation of output.
If the results are not within the manufacturer's tolerance re-align the controller
in accordance with the manufacturer's recommendations.
After re-alignment, repeat the manual adjustment of output and record the
results on the calibration sheet.
Reset test:
- with the controller in auto and the proportional band at 100%, apply a
50% measured variable signal to input, with reset adjustment at the minimum
setting and rate adjustment at maximum. Adjust the set point until output is
Q-PR-00-059-01
Procedure No. GS-MO-Q-PR-050-01 Page 4 of 9
Testing and Calibration of the Process Instruments
Revision 01, dated 15.12.2003
Rate test:
- with the controller in auto and the proportional band at 100%, apply a
50% approximately, measured variable signal to input, set reset adjustment to
minimum setting and rate adjustment to middle of range, adjust set point to
50%. Change input signal rapidly to 25% of scale.
Output should fall to zero or near zero before increasing again to approx-
imately 25% over a period of time. If the controller does not act as above, refer
to manufacturer's data for the corrective actions to be taken.
Manual/auto transfer test:
perform an auto/manual transfer test by switching repeatedly between the two
controller modes after balancing the output.
Apply a pressure to the switch input and record the pressure which activates
the switch.
If the switch does not operate at the pressure stated in the requisition, adjust
the switch until correct operation is obtained.
Increase the pressure and record the result in the calibration report for the
instrument.
Decrease the pressure and record the result.
Q-PR-00-059-01
Procedure No. GS-MO-Q-PR-050-01 Page 5 of 9
Testing and Calibration of the Process Instruments
Revision 01, dated 15.12.2003
If the results are not within the manufacturer's tolerance, re-calibrate the
gauge in accordance with the ma nufacturer's recommendations.
After re-calibration, re-apply pressures and record the results on the
calibration sheet.
Mount the gauge with its stem in a temperature bath and apply temperatures
at 0%, 50% and 100% of full scale and record the results.
If the results are not within manufacturer's tolerance, re-calibrate the gauge in
accordance with the manufacturer's recommendations.
After re-calibration, re-apply temperatures and note the results on the
calibration sheet.
2.3.10 Thermocouples
Continuity test:
Connect the thermocouple to a digital temperature indicator (Calibrator), allow
the thermocouple to stabilize at ambient temperature and note the reading.
The thermocouple reading should be within +3°C of ambient temperature as
indicated by the calibration standard.
Dimensional check:
Measure and record the length and diameter of the thermocouple probe.
Compare with the requisition data sheet and the corresponding thermowell
dimensions and note any discrepancy.
Record all data on the calibration check sheet. Thermocouples not meeting
the criteria given above shall not be installed.
Continuity test:
allow the instrument to stabilize at ambient temperature then measure and
note the resistance of the sensor with an ohmmeter. Also note ambient
temperature as indicated by the calibrating standard. The resistance of the
sensor shall be within +1 ohm of the value specified by the manufacturer.
Dimensional check:
measu re and record the length and diameter of the sensor. Compare with the
requisition data sheet and the correspond-ing thermowell dimensions, note
any discrepancy.
Q-PR-00-059-01
Procedure No. GS-MO-Q-PR-050-01 Page 6 of 9
Testing and Calibration of the Process Instruments
Revision 01, dated 15.12.2003
Record all data on the calibration check sheet. RTDs not meeting the criteria
given above, shall not be installed.
2.3.12 Recorders
Apply the correct power supplies for the chart mechanism and recorder.
Apply an input signal corresponding to 0%, 50%, 100%, 50%, 0% of full scale
to each recorder channel and note the indicated values.
If the results are not within the manufacturer's tolerance, re - calibrate the
recorder in accordance with the manufacturer's recommendations.
After re-calibration, re -apply input signals and record the results on the
calibration sheet.
2.3.13 Indicators
2.3.14 Converters
Positioner calibration
Q-PR-00-059-01
Procedure No. GS-MO-Q-PR-050-01 Page 9 of 9
Testing and Calibration of the Process Instruments
Revision 01, dated 15.12.2003
4.0 ATTACHMENT
Test Sheets:
Q-PR-00-059-01