Installation, Operation and Maintenance Manual CD, HL & NC Ranges Dri-Prime Pumps
Installation, Operation and Maintenance Manual CD, HL & NC Ranges Dri-Prime Pumps
Dri-Prime® Pumps
2 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
1 INTRODUCTION
1.1 Manual Purpose
This manual contains useful and important information to allow the pump to be properly installed, operated and maintained.
It also contains important instructions for preventing possible accidents and serious damage whilst carrying out those
activities.
Carefully read these instructions before commencing any of these activities.
The purpose of this Installation, Operating and Maintenance Manual is to provide the installer, owner or user of the
equipment with sufficient information to carry out those tasks for the CD, HL & NC ranges of Dri-Prime® Pumps.
Pump model designations covered by this manual are: -
The information contained in this manual was correct at the time of publication. It is subject to amendment at any time.
Should any doubt exist about the veracity of the information, contact Godwin Pumps Ltd for clarification before proceeding.
The pump may be supplied as a bare shaft pump end or packaged with a driver into a pump set. This manual covers the
pump end only. For packaged pump sets, information on equipment other than the pump end is contained in separate
documentation.
95-0019-0000 Iss 11 3
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
2 WARRANTY
Unless special arrangements have been agreed and signed by both parties Godwin Pumps Ltd. will apply the following
policy over defects found after delivery.
We will make good, by repair or the supply of a replacement, defects which, under proper use, appear in the goods within a
period of twelve calendar months after the goods have been delivered (1) and arise solely from faulty design (other than a
design made, furnished or specified by you for which we have disclaimed responsibility in writing), materials or
workmanship: provided always that defective parts have been returned to us if we shall have so required. We shall refund
the cost of carriage on such returned parts and the repaired or new parts will be delivered by us free of charge.
Our liability under this clause shall be in lieu of any warranty or condition implied by law as to the quality or fitness for any
particular purpose of the goods, and save as provided in this clause we shall not be under any liability, whether in contract,
tort or otherwise, in respect of defects in goods delivered or for any injury (2), damage or loss resulting from such defects as
from any work done in connection therewith.
(1) For export orders, within a period of twelve calendar months after the goods have been delivered or, if delivery is
delayed by reason of customer instructions or lack of instructions, within a period of 18 months after the goods have
been notified as ready for despatch (whichever period expires the earlier).
(2) For UK orders, other than personal injury caused by our negligence as defined in Section 1 of the Unfair Contract
Terms Act, 1977.
CAUTION
This caution symbol draws attention to special instructions or procedures which, if not correctly followed,
may result in damage to, or destruction of equipment.
WARNING
This warning symbol draws attention to special instructions or procedures which, if not strictly observed,
may result in personal injury.
WARNING
A WARNING SYMBOL WITH THIS TYPE OF TEXT DRAWS ATTENTION TO SPECIAL INSTRUCTIONS
OR PROCEDURES WHICH, IF NOT STRICTLY OBSERVED, MAY RESULT IN SEVERE PERSONAL
INJURY, OR LOSS OF LIFE.
4 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
4 SAFETY
WARNING
ALL ITEMS IN THIS SECTION, IF NOT STRICTLY OBSERVED, COULD RESULT IN SEVERE
PERSONAL INJURY OR LOSS OF LIFE.
4.1 Lifting
Use only lifting equipment of suitable capacity for the size and weight of the equipment being lifted
The equipment must always be lifted using safe working practices and in accordance with any local and national guidelines
or statutes. Figure 1, Figure 2, Figure 3 and Figure 4 show possible means of lifting. They are intended for guidance only. If
in doubt, consult Godwin Pumps Ltd or a local lifting expert.
CD225_02
HL100-04
CD100_02
HL100-02
Figure 1 Lifting smaller motor adaptor units Figure 2 Lifting smaller bearing bracket units
Figure 3 Typical lifting methods for close coupled bare shaft units (CD300M shown)
CD400 LIFT
Whilst lifting the unit keep personnel well away and never allow people underneath.
95-0019-0000 Iss 11 5
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
4.2 General
Personnel working on the pump must always wear clean, correctly fitting clothing and safety footwear. Clothing impregnated
with oil or fuel can constitute a health hazard through prolonged contact with the skin and may also constitute a fire hazard.
Signs on the unit (e.g. directional arrows, maintenance and serial number plates & fluid connections, etc.) must be kept
clean and clear.
Check the type of liquid that the pump has been used for before carrying out any maintenance work. Residues could be
hazardous to your health. If in doubt, flush thoroughly with clean water before commencing work.
Rotating equipment presents a hazard in itself. Alert surrounding personnel before starting and post notifications whilst in
operation.
Moving parts are guarded to protect you. Guards removed for maintenance must be replaced before starting the pump.
Never insert anything into the pump body whilst the pump is running and the suction or delivery hoses are disconnected.
Use all flange bolt holes and ensure the correct bolt size and quality is utilised when connecting suction and delivery hoses.
Collapsible hoses must never be used on the suction side of the pump.
Keep the hose end suction area free from debris. Although the pump can handle solids up to the size indicated in the
Technical Data section of this manual, larger or irregular solids may cause blockage with damage to pump components.
Always allow adequate ventilation for the pump driver. Diesel engines require air for both combustion and cooling. Electric
motors require air for cooling purposes. This air must never be allowed to re-circulate.
Be aware of burn and fire risks from items such as exhaust pipes and silencers. Never place flammable items around the
unit.
Diesel engine exhaust and some of its constituents are known in the State of California to cause cancer, birth defects and
other reproductive harm.
Liquid pressure may still be present even after shutdown of the pump. Particular attention should be paid to delivery lines
that are long, or rise through any height, as these can contain large volumes of liquid. These lines must be isolated and
drained down before commencing work. Sudden release of this liquid can cause serious injury to an operator either directly
or indirectly through the rotational motion it can induce.
6 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
5.1 Spares
Use only genuine parts from your local Godwin Pumps Ltd supplier or distributor. Failure to do so may invalidate warranty
and/or reduce the pump’s working life.
6 INSTALLATION
All pumps in the range are designed to be run in one direction only. This is anti-clockwise when viewed from the suction
flange end or clockwise when viewed from the driving shaft end.
Rotation on electric motor driven pumps should be checked carefully – preferably with the coupling disconnected. On
installations where coupling disconnection is not possible apply power only for a brief instance, preferably with a soft start, in
order to determine rotation. A fast start or prolonged running in the wrong direction could cause parts to come loose.
CAUTION.
Failure to provide the correct rotation will give rise to poor performance, vibration and possible severe
damage to the pump.
6.1 Versions
Larger pumps (CD300/400/500 or HL130/160/225/250/260) can be supplied with or without an SAE adaptor fitted. With the
adaptor fitted they are suitable for direct coupling, without they are suitable for open coupling. The bearing bracket must be
adequately supported when used as open coupled units. If any doubt exists over the suitability of the support consult
Godwin Pumps Ltd.
Smaller pumps in the range are supplied as either motor adaptor or bearing bracket versions: -
95-0019-0000 Iss 11 7
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
6.2 Alignment
The following notes provide general guidance on alignment and its elimination.
It is normal practice to carry out any adjustment to alignment by moving the driver – not the pump. In designing any
structure to support the pump set allow an additional 0.25mm (0.010”) extra height difference between pump and driver
support for shimming purposes.
There are two forms of misalignment – parallel and angular. Both must be less than the coupling manufacturers’ guidelines
in order for the coupling and bearings to give long and trouble free operation.
It is essential that the baseplate or support structure be fully tightened down before commencing any alignment procedures.
This will ensure that no misalignment is introduced by support structure deformation.
Alignment must always be carried out before any pipework is attached to the pump. Keep a note of the actual figures. Carry
out a check after the pipework has been attached and compare the figures. Any discrepancy must have been introduced by
the pipework attachment and must be corrected by refitting the pipework before putting the pump into service.
CAUTION.
Incorrect coupling alignment will lead to premature bearing and /or coupling failure.
diameter of the pump coupling (see Figure 6). If the driver shaft is not accessible then the dial
gauge can be mounted on the driver coupling.
Turn the driver shaft, note the total indicator reading, and adjust the driver position
accordingly. Recheck.
Parallel alignment.wmf
b) Angular alignment
Mount the dial gauge on the driver shaft or coupling (as for step a)) and adjust the dial Pump Driver
gauge to run on a mating face of the coupling as close to the outer diameter as
possible (see Figure 7).
Turn the driver shaft, note the total indicator reading, and adjust the driver in the
direction required. Recheck.
Angular alignment.wmf
the dial gauge can be mounted on the pump coupling provided the mounting face is
concentric with the shaft.
Turn the pump shaft and note the total indicator reading. Adjust the driver in the
direction required and recheck.
After alignment the pump and driver should be fully tightened down and the alignment rechecked to ensure that nothing has
moved in the tightening down procedure.
If a dial gauge is not available, then vernier callipers (or a taper gauge) and a straight edge can be used, providing the
coupling is of a type that permits it.
Measure the gap between coupling faces with the vernier callipers (or taper gauge) at four points equidistant around the
circumference of the coupling. Adjust until all are equal.
Use a straight edge across both machined outer diameters of the coupling at four points equidistant around the
circumference to confirm parallel alignment.
8 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
6.3 General Installation Notes
Godwin Pumps Ltd. may refute warranty liability if the installation does not meet the requirements of the pump. Should any
doubt exist as to the suitability of an installation, then Godwin Pumps Ltd. should be consulted.
Only suitably qualified personnel (both mechanical and electrical) should carry out the installation. All local and national
regulations in force must be observed.
The coupling used to connect the driver and pump must have the correct characteristics for the purpose. If in doubt –consult
Godwin Pumps Ltd.
The completed pump set must be mounted on a firm level surface.
On bare shaft pumps guarding of the belts, pulleys and couplings is the responsibility of the installer.
Pump sets are supplied with guarding to meet general applicable standards but site conditions may necessitate further
measures such as railings or screens. These are also the responsibility of the installer.
The Godwin Dri-Prime pump range is designed to operate on long lengths of suction pipe but it is always advisable to
position the pump as close to the source as possible.
The Godwin Dri-Prime product is designed to handle small amounts of air leakage, but to ensure optimum pump
performance, it is advisable to ensure that the suction pipe work is airtight and the end is sufficiently submerged to prevent
air being drawn into the pipe.
A general-purpose strainer should be fitted to the end of the suction pipe work and the total area of the openings should be
at least three times the normal area of the suction pipe work. Strainer apertures should be no greater than the maximum
recommended solids handling size for the pump.
The suction and delivery pipe work should be kept as short as possible with a minimum number of large radii bends to
minimise pipe friction losses. To maximise flow rates, it is encouraged to fit larger diameter pipe work than the pump
connection sizes.
Lay out piping runs before connecting to the pump to ensure that tight bends and other flow restrictions are not included.
Figure 9 shows some common pipeline installation problems and their recommended solutions.
Figure 9 Common pipeline installation problems and their solutions
It is important that the suction and delivery pipe work is supported immediately after the pump connections to prevent
straining and possible misalignment of the equipment.
95-0019-0000 Iss 11 9
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
7 OPERATION
7.1 General
Before attempting to start the unit ensure that you are familiar with the controls and any local or national safety regulations.
If the unit is engine driven refer to the engine manufacturer’s manual.
If the unit is driven by an electric motor refer to the relevant control requirements.
Once the unit is started, the air compressor or vacuum pump will start to prime the pump and the unit will require no further
attention.
CAUTION.
Failure to disconnect the pipeline will result in serious compressor or vacuum pump damage and
consequential pump failure.
Disconnect
compressor
drive
Disconnect
vacuum pump
Remove priming tower drive Remove ejector
complete and replace package and
with blanking plate
blank off
HL100-05
To disconnect: -
a) Ensure that the driver is isolated and the unit cannot be started whilst the pump is being worked on.
b) Compressor units: -Disconnect the air delivery pipe to the ejector.
Vacuum pump units: - Disconnect the air suction pipe from the priming tower.
Compressor units: -Remove the ejector housing complete and replace with a plain blanking plate (see
c) Figure 11).
Vacuum pump units: -Remove the priming tower complete and replace with a plain
blanking plate.
Blanking plates are available from Godwin Pumps Ltd or may be fabricated on site.
d) Ensure that the blanking plate has a leak free joint to the pump
e) Remove the isolation means from the driver and restart the pump.
HL100_09
Remove the blanking plate, replace the ejector housing or priming tower, reconnect the air hose and reinstall the
compressor/vacuum pump drive when the pump is next required to self-prime.
10 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
7.4 Draining
In cold weather when the slightest possibility of frost exists, drain the pump and the non return valve (if fitted). The drain
plug is either on the bottom or side of the volute (see Figure 12), or on the underside of the discharge flange (see
Figure 13).
CAUTION.
Failure to drain the pump and non return valve in these conditions could result in the pumped product
residue freezing and cracking the volute or non return valve.
Side/bottom drain if
volute is turned for
vertical discharge
Discharge flange drain
95-0019-0000 Iss 11 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8 MAINTENANCE
8.1 Bearings
All CD, HL & NC range pump shafts are supported from two bearing positions. The types of bearings and configurations
vary across the ranges.
CAUTION.
Failure to re-grease the bearings correctly with the correct type of grease at suitable intervals could result in
premature bearing failure.
Remove any excess grease from the unit, particularly in the area of any drive belt. Grease on the drive belt will shorten its
life and lead to premature failure.
Pump
Adaptor Pump
CD100-03
CD80-05
HL100-07
Grease nipple
accessed through
adaptor
Grease
relief
valves
Pump
body
Adaptor
Pump
Adaptor
HL200-01
CD80-02 CD225-03
Grease
nipples
Adaptor
Pump
HL250-04
12 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.2 Seal coolant level
All pumps are fitted with mechanical seals cooled by either oil or a glycol mixture. The HL260M & CD500M are glycol
cooled, all others use oil. Seal types and configurations vary across the ranges. Check the coolant level in the seal housing
at least once a week – preferably daily.
Please not the seal oil may become emulsified after a short period of operation, this is normal and not cause for concern.
The extent and timing depends on the conditions under which the pump is operated. This is not detrimental to the operation
of the seal, but the oil must be changed at the first sign of sludging.
Coolant level can vary depending on the conditions under which the pump is operated. A slight loss of coolant under high
suction lift conditions or slight emulsification of oil under low conditions is normal.
Oil and glycol specifications: - see Technical Data section
CAUTION.
Dry running caused by failure to initially fill, subsequently maintain the correct coolant level, or use the wrong
coolant in the seal housing will result in seal failure.
Seals that have failed due to lack of operator provision of the correct and sufficient coolant are not covered by the warranty.
The various configurations for filling and draining are shown in Figure 15, Figure 16 or Figure 17.
Fill
Fill
Pump
body
The seal coolant level for pumps of the style Pump
shown in Figure 15 is 10 –15mm below the
bottom of the plug. If no coolant is visible
Adaptor
then it must be topped up before starting the
pump.
CD80-06
CD80-03
Drain
Drain
Pump Fill
Adaptor
Adaptor
The seal coolant level for pumps of the style Pump
shown in Figure 16 is 35-40mm below the
filler plug seating face (10 –15mm below the Drain
(diametrically
bottom of the filling hole) opposite to fill)
CD225-04
Drain
CD100-04
Adaptor
Larger pumps (Figure 17) have a filler plug
combined with a dipstick. The coolant level Pump
must be maintained between the maximum
and minimum marks.
HL250-03
Drain
95-0019-0000 Iss 11 13
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.3 Seal removal
WARNING.
This procedure requires access to operating and rotational parts of the pump. Accidental start up could
cause injury.
1) Isolate the driver and ensure the unit cannot be started whilst the pump is being worked on.
2) Drain down the delivery line (see warning note in Section 4). Disconnect both suction and delivery lines. Ensure that
the non-return valve is empty.
3) Drain water from the pump (see Section 7.4).
4) Drain the coolant oil from the mechanical seal cavity (see Section 8.2).
5) For units fitted with a compressor disconnect the air supply hose to the ejector head. For units fitted with a vacuum
pump disconnect the suction hose from the priming tower.
WARNING.
Dismantling requires adequate support of both the parts being removed and the remaining parts of the
pump. Those being removed must be supported by suitable lifting gear whilst fasteners are undone. The
lifting gear can then be used to lift the part clear. Remaining parts must also be adequately supported.
Failure to hold or support the parts adequately could cause them to fall and cause injury.
14 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Pump
body
Adaptor
CD80-07
CD80-04
Figure 18 Typical pump body attachment to bearing Figure 19 CD75 & CD80D Pump body attachment to motor
bracket (CD80D shown) adaptor
Pump types CD103M/ CD140M/ CD150M CD160M (Mk 1 & 2)/ CD180M/ CD225/ CD200M/ CD300M/ CD400M/
CD500M
HL80/ HL100/ HL110M/ HL125MS/ HL150M/ HL150MHD/ HL200M/ HL130M/ HL160M/ HL225M/
HL250M/ HL260M
NC80/ NC100/ NC150
Due to the weight and size of the parts on these larger pumps and lifting gear limitations it is recommended that items such
as separation tanks, priming towers and non return valves are removed before proceeding further.
Release and remove the fasteners holding the front cover to the pump body (see Figure 20). These are on the periphery of
the front cover. At this stage do not disturb any other fasteners, as these hold the front wearplate to the front cover. Remove
the front cover and wearplate complete. Larger pumps are provided with pre-tapped holes around the periphery to enable
the use of jacking screws to help in removal.
NOTE: - The CD400M and CD500M have an adaptor plate fitted between the front cover and pump body. It carries the
front wearplate and is located by a spigot on the front cover. It normally detaches with the front cover, but may occasionally
require separate removal.
Jacking
Screws
CAUTION.
Failure to support the motor adaptor once the pump body is removed could distort the adaptor and displace
the bearings leading to pump failure.
95-0019-0000 Iss 11 15
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.3.2 Removing the impeller
Lock the pump shaft to prevent rotation. Tools are available from Godwin Pumps Ltd for this purpose. See Section 9.
WARNING.
Failure to lock the pump shaft could lead to unexpected rotation of the impeller causing entrapment and the
possibility of serious injury.
Pump types CD100M/ CD103M/ CD140M/ CD150M/ CD160M (Mk 1 & 2)/ CD180M/ CD200M/ CD225/ CD250M
HL80/ HL100/ HL110M/ HL125MS/ HL50M/ HL150MHD/ HL200M
The impeller is screwed (right hand thread) directly onto the shaft and retained by a screw in the centre of the boss. Remove
the screw. (NOTE: - this screw incorporates a self locking device and must be replaced and not reused). Unscrew the
impeller in an anti clockwise direction.
If the impeller is reluctant to rotate, place a block of wood against one of the vanes and tap lightly with a hammer to
overcome the initial resistance.
Shims will have been fitted on the shaft behind the impeller. Remove these carefully and retain for reuse unless damaged.
16 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
95-0019-0000 Iss 11 17
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Pump types CD140M/ CD160M (Mk 1 & 2)/ CD180M/ CD200M/ CD225/ CD250M/ CD300M/ CD400M
HL100/ HL110M/ HL125MS/ HL150M/ HL150MHD/ HL200M/ HL130M/ HL160M/ HL225M/ HL250M
Removal of the wear plate exposes the outer seal seat carrier retained by four off cap head socket screws. Removing the
screws and carrier exposes the remainder of the mechanical seal on the shaft. The inside face of the seal carrier holds the
outer seal seat (item 12 in Figure 25 or Figure 26, item 1 in Figure 27) retained by a clamping ring held in place by screws.
Remove the screws from the clamping ring. Remove the clamping ring and press out the outer seal seat.
18 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Lapped
seal faces
Pump types CD160M Mk 2/ CD180M/ CD300M
(Figure 25) HL130M/ HL160M/ HL250M
12 1 2 3 4 5 13 11
Remove the rotating seal face (1) and O-ring (2) from the shaft.
Unscrew the grub screws (5) holding the stationary seat (3) to the
shaft. Remove the stationary seat.
Remove the second rotating seal face (13).
The seal seat (11) is held in a carrier retained by 4 off cap head
socket screws. Undo the screws, remove the carrier and seat.
Press the seat out of the carrier.
Figure 25 Type 2 two part double mechanical seal
Lapped Lapped
seal faces seal faces
Pump types CD160M Mk 1/ CD400M
(Figure 26) HL110M/ HL200M
12 1 2 3 4 5 6 7 8 9 10 11
Remove the rotating seal face (1) and O-ring (2) from the
shaft.
Unscrew the grub screws (5) holding the stationary seat (3)
to the shaft. Remove the stationary seat.
Unscrew the grub screws (6) holding the second stationary
seat (8) to the shaft. Remove the stationary seat.
Figure 26 Type 3 double mechanical seal
Ø90double mechanical seal
Remove the second rotating seal face (10) and O-ring (9) Lapped Lapped
from the shaft. seal faces seal faces
The seal seat (11) is held in a carrier retained by 4 off cap head socket screws. Undo the screws, remove the carrier and
seat. Press the seat out of the carrier.
8.4.2 Inspection
Inspect all items for wear. The seal face for type 1, 2 & 3 seals (item 6 in Figure 24; item
12 in Figure 25 or Figure 26) is double sided. It may be reversed when refitted to provide
a second wearing face. Replace any damaged parts and renew all O-rings.
95-0019-0000 Iss 11 19
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.5 Seal fitting
CAUTION.
Mechanical Seals are precision engineered devices. Extreme care must be taken to ensure that no damage
occurs to the lapped faces. These faces must be kept absolutely clean throughout the entire installation. Do
not touch them or allow any contaminant to come into contact with them. Soiled faces will have to be
cleaned with appropriate alcohol based degreasing cleaner and soft tissue. Failure to observe these
precautions will lead to premature seal failure.
8.5.1 Preparation
Ensure that the parts and edges the seal will pass over or through during the assembly process are free from any burrs or
sharp edges and are scrupulously clean. Any of these could damage the seal during assembly and cause premature failure.
Clean the shaft thoroughly and lubricate with clean water or a diluted soft soap solution. Do not use heavy grease, silicone
or PTFE based lubricants, as these would prevent the seal bellows from gripping the shaft.
It is recommended that the seal components be laid out in the order of assembly before starting to enable easy identification
during the procedure.
The retaining screws for the seal shaft sleeve should be fitted using a thread locking compound. See Technical Data section
for details.
Pump types CD75/ CD80D/ CD80M/ CD100M/ CD103M/ CD140M/ CD150M/ CD200M/ CD225/ CD250M
HL80/ HL100M/ HL125MS/ HL150M/ HL150MHD
(Figure 24)
NC80/ NC100/ NC150
CD140M, CD200M, CD225M, CD250M, HL100M, HL125MS, HL150M & HL150MHD only. Fit the seal spacer onto the
shaft ensuring it seats against the shaft shoulder.
Carefully remove the rotary head (1) and O-ring (2) from the assembly. Do not place the rotary head (1) face down as this
may cause damage to the lapped seal face.
Take the sleeve assembly (3) and ensure that the drive screws (4) are fully retracted to prevent scratching of the shaft
during assembly. Carefully slide the sleeve assembly along the shaft until it butts up against the shaft shoulder (CD75 &
CD80D), collar (CD80M, CD100M, CD103M, CD150M) or spacer (CD140M, CD200M, CD225 CD250M, HL100M,
HL125MS, HL150M & HL150MHD).
Apply thread locking compound (see Technical Data Section) and lightly tighten the drive screws to centralise the sleeve on
the shaft. Ensure that the sleeve is still abutted to the shoulder, collar or spacer, and then tighten the drive screws fully.
Lightly smear the O-ring (2) with a silicon based grease lubricant. Place the O-ring inside the rotary head (1). Avoid getting
any grease onto the lapped seal face.
Gently push the rotary head (1) onto the sleeve assembly (3) ensuring that the drive pins (5) are engaged in the slots of the
rotary head.
Carefully clean the lapped seal face with a suitable degreasing agent and soft tissue.
Fit the seal seat (6) to the wear plate and retain with either the circlip or retaining clamp. Ensure a new O-ring is fitted.
20 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
95-0019-0000 Iss 11 21
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
CAUTION.
Failure to ensure the pin locates in the groove of the seal for CD500M & HL260M units will result in seal
seat breakage.
Carefully clean the lapped seal face with a suitable degreasing agent and soft tissue.
Ensure the O-ring(s) is/are in place inside the sleeve (2) carrying the double seal faces and carefully slide the whole onto
the shaft.
Fit the seal seat (1) to the carrier (CD300M, HL130M & HL225M), or to the wear
plate (CD500M & HL260M) and retain with the retaining clamp. Ensure a new O-
ring is fitted.
22 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Pump types CD140M/ CD160M (Mk 1 & 2)/ CD180M/ CD200M/ CD225/ CD250M/ CD300M/ CD400M
HL100/ HL110M/ HL125MS/ HL150M/ HL150MHD/ HL200M/ HL130M/ HL160M/ HL225M/ HL250M
Slide the carrier over the shaft and attach with four off cap head socket screws. Position the rear wear plate and attach with
fasteners from the motor adaptor/bearing bracket/seal housing adaptor side.
95-0019-0000 Iss 11 23
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Pump types CD100M/ CD103M/ CD140M/ CD150M/ CD160M (Mk 1 & 2)/ / CD180M/ CD200M/ CD225/ CD250M
HL80/ HL100/ HL110M/ HL125MS/ HL150M/ HL150MHD/ HL200M
Refit the shims (or new equivalents) removed during strip down.
Screw the impeller (right hand thread) directly onto the shaft. Do not fit the self locking retaining screw at this stage. Check
there is still a clearance and torque down fully.
Check the rear impeller clearance. If within the permitted tolerance secure the impeller with the self locking retaining screw.
24 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Pump types CD103M/ CD140M/ CD150M/ CD160M (Mk 1 & 2)/ CD180M
CD225/ CD200M/ CD250M/ CD300M/ CD400M/ CD500M
HL80/ HL100/ HL110M/ HL125MS/ HL150M/ HL150MHD/ HL200M/ HL130M/ HL160M/ HL225M/
HL250M/ HL260M
Larger pump bodies (HL80, CD300M, CD400M, CD500M, HL160M, HL225M, HL250M but not HL130M) were probably not
removed as they have sufficient room to allow the seal to be removed.
For other units or if the larger pump body has been removed, fit the body, guiding the body studs into the adaptor holes.
Secure the peripheral fasteners.
Attach the front cover and wearplate assembly to the pump body (see Figure 20) using the peripheral fasteners.
NOTE: - The CD400M and CD500M have an adaptor plate fitted between the front cover and pump body. It carries the
front wearplate and is located by a spigot on the front cover. It normally remains with the front cover, but may have been
dismantled and require refitting.
Check the front impeller clearance. Adjust the shims between pump body and front wear plate to achieve the correct figure.
95-0019-0000 Iss 11 25
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.6.2 Dismantling
Pump types CD75/ CD80D/ CD80M/ CD100M/ CD103M/ CD140M/ CD150M/ CD160M/ CD180M/ CD200M/ CD225/
CD250M
NC80/ NC100/ NC150
1) CD160M, CD180M
Undo the fasteners holding the seal-housing adaptor to the bearing housing and remove the seal housing adaptor.
Undo the fasteners holding the SAE adaptor to the bearing housing and remove the SAE adaptor.
2) Unscrew the drive end bearing cover fasteners and remove the drive end bearing cover.
3) CD75, CD80D, CD80M & CD160M motor adaptor units.
Release the tab washer and remove the bearing locknut.
4) Extract the shaft and bearing assembly.
5) Remove pump end bearing cover on bearing bracket units where the cover is retained by screws. CD100, CD103M,
CD150M, CD250M, NC80, NC100, & NC150 motor adaptor units have their pump end bearing cover retained by a
circlip. Do not remove the clip, cover or pump end outer bearing race unless required for cleaning purposes or because
of damage.
6) If necessary, remove the bearing inner races from the shaft.
7) Inspect all parts for damage or wear and replace as necessary. All lip seals and O-rings must be replaced.
1) HL80
Undo and remove the nuts and spring washers from the studs holding the pump body to the bearing bracket or motor
adaptor.
Undo and remove the pump body holding down bolts (if fitted). Remove the pump body.
Pull the collar and tolerance ring off the pump end of the shaft.
HL100M, HL125MS, HL150M, HL150MHD, HL110M & HL200M
Release the nuts on the bolts holding the seal housing adaptor to the bearing bracket and remove the fasteners.
Remove the adaptor.
For pumps other than the HL110M or HL200M this will include the shaft lip seal.
For the HL110M or HL200M this will include the rear mechanical seal seat if it has not already been removed.
2) Unscrew both bearing cover fasteners and remove the bearing covers.
3) HL100M, HL125MS, HL150M, HL150MHD, HL110M & HL200M
Release the pump end tab washer, lock the shaft against rotation and remove the bearing locknut.
4) Extract the shaft and bearing assembly from the drive end, leaving the pump end outer bearing race in place.
5) HL80
Motor adaptor units have their pump end bearing cover retained by a circlip. Do not remove the clip, cover or pump end
outer bearing race unless required for cleaning purposes or because of damage.
6) Remove pump end bearing cover.
7) If necessary, remove the bearing inner races from the shaft.
8) Inspect all parts for damage or wear and replace as necessary. All lip seals and O-rings must be replaced.
26 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
95-0019-0000 Iss 11 27
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.7 Bearing Bracket & Motor Adaptor Assembly
CAUTION.
Bearing races must be pressed into position
and not hammered either directly or by drift. A
purpose designed ‘top-hat’ type of assembly
tool as acceptable. Direct hammering will
damage the bearing or rollers. Drift hammering
will introduce swarf into the assembly. Either
will result in early bearing failure.
8.7.2 Assembly
Pump types CD75/ CD80D MA Units
1) Heat the roller bearing. Once up to temperature, slide it hard against the shaft shoulder, holding it there for a minimum
of 30 seconds. This allows the race to grip the shaft and prevent it from creeping away from the shoulder during
cooling. Let the assembly cool completely.
2) Fit the inner and outer bearing spacers.
3) Heat the ball bearing. Once up to temperature, slide it hard against the inner spacer, holding it there for a minimum of
30 seconds. This allows the race to grip the shaft and prevent it from creeping away from the spacer during cooling. Let
the assembly cool completely.
4) Fit the tab washer and locknut. Prise up at least two tabs to lock the nut in position.
5) Pack the bearings with grease allowing sufficient excess to one-third fill the cavities at each side of the bearings.
6) Position the motor adaptor with the shaft axis vertical (pump end down). Press the shaft assembly into the motor
adaptor.
7) Grease the pump end lip seal and press it into the motor adaptor body (lip facing in).
8) Grease the drive end lip seal and press it into the bearing cover (lip facing in).
9) Fit the bearing cover and tighten down the fasteners evenly to the recommended torque.
10) Fit ancillary components.
28 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
95-0019-0000 Iss 11 29
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
30 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
1) Heat the single roller bearing inner race and the two angular contact bearings. Once up
to temperature, take the two angular contact bearings and slide them up hard against
the shaft shoulder at the drive end.
The bearing orientation is in ‘O’ arrangement, i.e. with the two outer races ‘trapped’
between the balls.
2) Take the single roller bearing inner race and slide it up hard against the shaft shoulder.
3) Use the lock nuts without tab washers to hole them hard against the shoulders for a
minimum of 30 seconds to allow the races to grip the shaft and prevent them creeping
away from the shoulders during cooling. Let the assembly cool completely. Remove both
locknuts.
4) Fit the drive end tab washer and locknut. Tighten the locknut securely with a C spanner. Prise up at least two tabs to
lock the nut in position.
5) Locate and secure the bearing bracket on an assembly jig or bench.
6) Take the single roller bearing outer race and press it into the bearing
bracket at the pump end. Use the pump end bearing cover, without the lip
seal fitted, to press it into its final position. Remove the cover.
7) Secure the shaft aligning jig to the pump end of the bearing bracket. Place
the two bearing inner covers in the bracket. Lift the shaft horizontally using
suitable over head lifting equipment and slide the shaft assembly in to the
bracket.
8) Fit the lip seal to the drive end bearing cover (seal lip facing in). Smear the
lip seal lightly with oil and fit the cover over the shaft and secure it to the
bearing bracket and the inner cover.
9) Remove the shaft-aligning jig. Fit the non-drive end bearing tab washer and
locknut. Tighten the lock nut securely with a C spanner. Prise up at least two
tabs to lock the nut in position.
10) Fit the lip seal to the pump end bearing cover (seal lip facing in). Smear the
lip seal lightly with oil and fit the cover over the shaft and secure it to the
bearing bracket and the inner cover.
11) Fit the bearing cover extension pipes and grease nipples. Charge the
bearings with grease until it is witnessed just emerging from the inner
covers. During this process it is important to continually rotate the shaft.
When completed remove the grease extension pipes temporarily to avoid
damage during further assembly. Plug the bearing cover holes to avoid
grease contamination.
95-0019-0000 Iss 11 31
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
8.8 Air Compressor
Smaller pumps in the range are fitted with an air compressor. The following instructions apply to those units only.
32 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Adjusting
studs
3 off
Clamping
screws
2 off
Figure 29 Link belt driven compressor (shown on a CD150 bearing bracket - guards and belt removed)
A) Checking / Tensioning
Link belt tension must be checked and, if necessary, adjusted weekly or after every 100 hrs operation.
See Section 11 for force/deflection details.
CAUTION.
Link belt tension must be maintained within the specified limits. Belts that are too slack or too tight will wear
out quickly, damage the pulleys and will result in loss of pump performance.
To check/adjust tension:-
1) Remove at least one half of the guard.
2) Check the belt condition. Look for signs of dust and/or pulley wear (belt over tensioned) or smearing of the belt sides
(belt under tensioned and slipping).
3) Using a spring balance and rule (or a belt tension indicator) check the force and deflection of both belts.
4) If adjustment is required, slacken the two clamping screws (see Figure 29) enough to allow the compressor carrying
frame to slide on the supporting angle frame.
5) Slacken the 3 off adjusting stud nuts on top of the carrying frame (see Figure 29).
6) Evenly adjust the nuts under the carrying frame to raise or lower the compressor.
CAUTION.
Ensure the compressor drive shaft remains parallel to the pump shaft.
Check visually that the drive belts are perpendicular to the shaft axes.
Use a straight edge across the pulley faces for confirmation.
Failure to ensure that the drive is perpendicular will result in premature belt wear and early failure.
7) Check the tension and deflection are within the specified range and that it is the same within 2 or 3% on both belts.
8) Correct tension: - Tighten the 3 off adjusting stud nuts (standard torque for nuts on studs – see Section 11) above the
carrying frame. Go to instruction 20).
9) Tension not achievable because the limit of adjustment has been reached:- A link must be removed from each belt.
Lower the carrying frame to the bottom of its adjustment and remove the belts from the pulleys.
CAUTION.
The drive must be slack enough to allow the belt to be removed without forcing it over the pulley rim which
will damage the belt and lead to premature failure.
95-0019-0000 Iss 11 33
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
10) Each belt link is tabbed to the next two. To disconnect turn the belt inside out so the tabs point outwards.
11) Bend the belt backwards as far as possible so that the tabs protrude. Hold the belt in this position with one hand and
twist one tab so that it is parallel with the slot of the underlying link.
12) Pull the underlying link over the twisted tab.
13) The belt is now held together by only one tab. Rotate the two parts 90° and remove this remaining tab from the
underlying link.
14) Remove one link from the belt by repeating instructions 11) to 13).
15) To rejoin the belt, ensure it is still threaded around the pump shaft and push one end tab through two links at once on
the other end.
16) Flex the belt until the second tab can be inserted by twisting into place.
17) Ensure both tabs have returned to their normal orientation. Reverse the belt so that the tabs are inside.
18) Refit the belt over both pulleys. Go to instruction 6).
19) Tighten the 2 off clamping screws (standard torque for nuts and bolts – see Section 11).
20) Recheck the force/deflection to ensure no movement has taken place whilst tightening up.
21) Replace the guard.
22) Run the pump for thirty minutes; recheck the tension adjusting if necessary.
23) The pump is now ready for service.
34 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
9 TOOLS
9.1 Tool Kits
CD75 ACC-PT-01 √
CD80D ACC-PT-03
CD103M ACC-PT-31*
CD140M
ACC-PT-07 ACC-PT-08
CD150M
ACC-PT-27* ACC-PT-28*
ACC-PT-33 (Mk1)
CD160M
ACC-PT-36 (Mk2)
ACC-PT-09 ACC-PT-20
CD200M
ACC-PT-29* ACC-PT-22*
CD225 ACC-PT-29* ACC-PT-22*
CD300M ACC-PT-30
CD400M ACC-PT-38
CD500M ACC-PT-40
HL260M ACC-PT-41
The kits generally comprise of: - a special spanner and locating tool to aid in impeller removal/fitting.
* These tool sets have an impeller spanner designed for use with a 1” square drive torque multiplier
a locking bar to prevent shaft rotation
a sleeve to aid in fitting the mechanical seal
a sleeve to aid in pressing the bearing home on the shaft
a tool to aid in assembly/removal of the rear wear plate
a tool to aid correct location of the lip seals in the bearing covers
a tool to position the bearing spacer (where applicable) in the bracket
a seal assembly and setting tool.
95-0019-0000 Iss 11 35
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
10 FAULT FINDING
If possible fit a suction and pressure gauge to assist fault finding and check pump rating
FAULT
Excessive power
Insufficient liquid
consumption
overheating
Vibration or
emulsified
delivered
prime
POSSIBLE CAUSE
Suction lift too great √ √ √
Insufficient water at suction inlet √ √
Suction inlet or strainer blocked √ √ √
Separation tank filter blocked √ √ √
Suction line not air tight √ √
Suction hose collapsed √ √ √
Non return valve not seating √
Mechanical seal drawing air into pump √ √ √
Ejector jet or nozzle blocked or badly worn √ √
Compressor
√
units only
36 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
11 TECHNICAL DATA
11.1 Designations, Sizes & Capacities
11.1.1 CD range
(1)
Weight (kg) Speed Range
Max solids
Max. head
Max work
temp (°C)
bore(ins)
Max flow
Suction /
pressure
handling
Delivery
working
CD
(m /hr)
(mm)
(bar)
Max
Engine
(m)
driven
driven
Motor
3
Range Min Norm Max
95-0019-0000 Iss 11 37
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
11.1.2 HL Range
(1)
Weight (kg) Speed Range
Max solids
Max. head
Max work
bore (ins)
temp (°C)
Max flow
Suction /
pressure
handling
Delivery
working
HL
(m /hr)
(mm)
(bar)
Max
Engine
(m)
driven
driven
Motor
3
Range Min Norm Max
11.1.3 NC Range
(1)
Weight (kg) Speed Range
Max solids
Max. head
Max work
temp (°C)
bore(ins)
Max flow
Suction /
pressure
handling
Delivery
working
NC
(m /hr)
(mm)
(bar)
Max
Engine
(m)
driven
driven
Motor
3
38 95-0019-0000 Iss 11
Pump Type
CD75
CD225
CD80D
CD80M
CD100M
CD103M
CD150M
CD140M
CD180M
CD200M
CD250M
CD300M
CD400M
CD500M
Detail
CD160M Mk1
CD160M Mk 2
95-0019-0000 Iss 11
MA Flange SAE 5 4 3 2 or 3 4 3 1 (2) N/A 0 or 1
0.127 to 0.127 to
11.2 CD Range data
Shaft end float 0.33mm Nominal bearing clearance only 0.33mm Nominal bearing clearance only
0.075 to 0.10mm (0.003” to 0.004”)
mm (ins) (0.005” to 0.05mm (0.002”) max (0.005” to 0.05mm (0.002”) max
0.013”) 0.013”)
Impeller clearances front & rear (rear Cast Iron Imp/Wear Plate = 0.50 to 0.625mm (0.020” to 0.025”)
only on CD75)mm(ins) Hard Iron & Stainless Steel Imp/Wear Plate = 0.625 to 0.762mm (0.025” to 0.030”)
Inboard
Inboard 37.0±0.3
37.0±0.3 32.0±0. Outboard 50.0 N/A
Seal working length (mm) 32.0±0.3 34.0±0.3 39.0±0.3 34.0±0.3
Outboard 3 45.5±0.5 ±0.5 (1)
45.5±0.5
N/A (1)
Grease Texaco Starplex EP2, Shell Nertia HV, Mobil SHV 46, Total Multis Complex EP2, Belesta XPG
CD, HL & NC Ranges of Dri-Prime® Pumps
50/50
Type Oil. Any SAE 20/20 or BIOPUS 46 Biodegradable water/
Installation, Operation and Maintenance Manual
glycol
Seal
Coolant Capacity (litres) 0.5 1.75 3.8 1 6 6.8 7
Vol – Max to Min (litres) N/A 0.35 0.5 N/A 0.6 0.6 1.3
NOTES:-
(1) Seal length is not applicable for units with a one piece seal. The seal length is set automatically on installation. Where other figures are given these are for units with a multi piece seal.
(2) Only for close coupled units.
39
40
Pump Type
HL80
HL100M
HL110M
HL150M
HL200M
HL130M
HL160M
HL225M
HL250M
HL260M
Hl125MS
HL150MHD
Detail
MA Flange SAE 4 3 2 or 3 3 2 0 or 1
50/50
Type Oil. Any SAE 20/20 or BIOPUS 46 Biodegradable
water/glycol
5.5 litres initial fill.
Seal
Capacity (litres) 5.25 5.25 Top up to full mark on 7.3
Coolant dipstick after 1 hr running
95-0019-0000 Iss 11
Pump Type
NC80
NC100
NC150
Detail
95-0019-0000 Iss 11
MA Flange SAE 4
11.4 NC Range Data
Seal
Coolant Capacity (litres) 0.5
Installation, Operation and Maintenance Manual
41
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
Steel Stainless
Torque (N-m)
Thread Size Nuts & Bolts Nuts on Studs All
M6 x 1.0 11.7 4.6 9
M8 x 1.25 28 11 22
M10 x 1.5 56 22 44
M12 x 1.75 98 38 76
M16 x 2.0 244 95 187
M20 x 2.5 476 185 364
M24 x 3.0 822 320 629
M30 x 3.5 1633 633 1240
Steel Stainless
Torque (N-m)
Thread Size Nuts & Bolts Nuts on Studs All
¼” - 20 13.5 5.4 10.5
5/16” – 18 27.1 10.5 21.0
3/8” – 16 48.8 20 37.8
7/16” – 14 74.6 30 57.8
½” – 13 122 48 94.6
5/8” – 11 237.3 95 183.9
¾” – 10 420.4 167 325.8
7/8” – 9 664.4 266 514.9
1” - 8 1003 401 777.3
Table 1 figures are for Metric fasteners to BS3692. Steel fasteners: - Grade 8.8 for bolts, Grade 4.6 for studs, Grade 10 for
nuts; Stainless Steel fasteners: - Grade A2.
Table 2 figures are for UNC fasteners to BS1768 Steel fasteners: -Grade S and Grade 3 for nuts; Stainless Steel fasteners:
- Grade A2.
If steel fasteners are of dissimilar condition then the figures Table 1 or Table 2 must be multiplied by those in Table 3.
Table 3 Correction factors
42 95-0019-0000 Iss 11
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
95-0019-0000 Iss 11 43
Installation, Operation and Maintenance Manual
CD, HL & NC Ranges of Dri-Prime® Pumps
NOTES
E-mail: sales@godwinpumps.co.uk
Website: www.godwinpumps.co.uk
E-mail: sales@godwinpumps.com
Website: www.godwinpumps.com
44 95-0019-0000 Iss 11