0% found this document useful (0 votes)
481 views50 pages

Service Manual: Martin ME 102 Electrosurgery Unit

Uploaded by

Pepo
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
0% found this document useful (0 votes)
481 views50 pages

Service Manual: Martin ME 102 Electrosurgery Unit

Uploaded by

Pepo
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
Download as pdf or txt
You are on page 1/ 50

Service Manual

Martin ME 102 Electrosurgery Unit


Martin ME 102 Service Manual

© by Martin Medizin-Technik
This documentation is protected by the copyright of the Martin Medizin-Technik
Company.
This documentation may not be changed, expanded or reproduced, or transferred
to third parties without the prior written permission of Martin Medizin-Technik.

Martin Medizin-Technik
Gebrüder Martin GmbH & Co. KG
Ludwigstaler Str. 132
78 532 Tuttlingen

Telephone: + 49 (0) 74 61/7 06-0


Fax: + 49 (0) 74 61/70 61 93
e-mail: info@martin-med.com

If you have technical questions or require replacement parts please consult our Hotline.

Martin Medizin-Technik Hotline:


Telephone: + 49 (0) 74 61/70 63 43
Fax: + 49 (0) 74 61/70 61 90
e-mail: msc@martin-med.com

Replacement Parts:
Telephone: + 49 (0) 74 61/70 61 99
Fax: + 49 (0) 74 61/70 61 90

Our Hotline is available from 8.00 - 17.00 on working days.

Design and device may be subject to changes.

V. 1.0
Martin ME 102 Service Manual

Table of contents

Chapter Subject Page

1. Introduction 1
1.1 Preface 1
1.2 Validity 1
1.3 Guarantee 2
1.4 Recurring technical safety inspections 2
1.5 Symbols 2

2. Safety Instructions 3
2.1 General Information 3
2.2 Areas of Responsibility 3
2.3 Protection of persons 4
2.4 Protection of devices 5

3. Device Description 6
3.1 Front 6
3.2 Back 7
3.3 Overview of components 8
3.4 Wiring diagram 9

4. Switch description 10
4.1 Block switch diagram 10
4.2 Description of the Function Groups 11
4.3 Description of important signals 14

5. Trouble shooting 15

6. Assembling and disassembling modules 16


6.1 Opening and closing the housing 16
6.2 Front, front PCB, sockets 17
6.2.1 Removing the rotary knob 17
6.2.2 Replacing the potentiometer 17
6.2.3 Removing and replacing the front 18
6.2.4 Replacing the mains power switch or front sheeting 20
6.2.5 Removing the NE-connection socket 21
6.2.6 Removing the front PCB 21
6.2.7 Removing the AE Connection Combination 22
6.2.8 Removing the bipolar connection socket 22
6.3 Mains connection socket, power transformer, SMPS
(switched mode power supply) 23

V. 1.0 I
Martin ME 102 Service Manual

Chapter Subject Page

6.3.1 Removing the mains connection socket 23


6.3.2 Removing the power transformer/SMPS
(switched mode power supply) 23
6.4 HF-Output PCB, Foot Switch Sockets, HF-Generator PCB 26
6.4.1 Removing the HF-Output PCB 26
6.4.2 Foot Switch Sockets 26
6.4.3 Removing the HF-Generator PCB 27

7. Adjustment, Modifications 28
7.1 Position of the adjustment points 28
7.2 Fundamental points regarding HF-Output power
measurement 28
7.3 Adjustment of the HF-Output power 29
7.4 Adjustment of the NE Monitor 29
7.5 Adjustment of the switch-on threshold for bipolar-automatic 30
7.6 Changing over the mains voltage 30
7.7 Determination of the HF-Leakage currents 32
7.7.1 Cable connections in the "Monopolar" operating mode 32
7.7.2 Cable connections in the "Bipolar" operating mode 33

8. Recurring technical safety inspections 34


8.1 Overview of the inspection activities to be carried out 34
8.2 Inspection Report 35

9. Technical Data, Diagrams, Replacement Parts 37


9.1 General Data 37
9.2 "Macro Mode" Performance Charts 38
9.3 "Micro Mode" Performance Charts 39
9.4 ME 102 Replacement Parts 40

10. Changes 42
10.1 Procedure 42
10.2 Hardware status 42
10.3 Software status 43
10.4 Configuration ME 102 43
10.4.1 Configuration 0000 43
10.4.2 Configuration 0100 43

11. Appendix 44

II V. 1.0
Martin ME 102 Service Manual

1. Introduction
1.1 Preface
Dear Service Technician,
You are faced with the task of servicing an ME 102 Electrosurgery Unit, analysing
malfunctions or defects and correcting these, as well as carrying out the verification of
operating data.
This Service Manual is intended to support you in the important work for which you are
responsible.
The service work which is described in the following requires good electrical and
mechanical knowledge.
It is for these reasons that this work may only be carried out by service
technicians who are authorised by the manufacturer!
As manufacturer, we urgently recommend that, before commencing work, you become
familiar with
– the functions and operation of the device and

– the service work which has to be carried out.

For this purpose please read the Operating Instructions for the device and this
Service Manual thoroughly and be absolutely certain that you heed the safety
instructions!

We will be pleased to answer any questions you may have regarding this device or
other Martin products.
You will find the telephone number for our Hotline in the imprint at the beginning of this
Service Manual.

Sincerely,
Martin Medizin-Technik
(Martin Medical Technology)

1.2 Validity
The descriptions in this Service Manual refer to the
ME 102 Electrosurgery Unit

from Martin Medizin-Technik.

V. 1.0 1
Martin ME 102 Service Manual

1.3 Guarantee
The ME 102 Electrosurgery Unit is covered by a 12 month guarantee which
commences on the day the device is delivered.
Within this guarantee period our responsible customer service centres or the
manufacturing plant will correct, free of charge, all defects which can be attributed to
manufacturing or material defects.
The general guarantee and warranty provisions of the Martin company apply.

Repairs to the device may only be carried out by the manufacturer or persons or
companies authorised by the manufacturer!

1.4 Recurring technical safety inspections


The device has to be inspected every 24 months, as described in Chapter 8. This
inspection has to be made by persons who are capable of correctly implementing such
an inspection as a result of their training, their knowledge and their practical experience
and are not in any way restricted in their capacity to carry out this inspection.

1.5 Symbols
Texts which are marked with this symbol and printed in bold contain extremely
important information which may also include information regarding the
prevention of health hazards!
It is essential that you heed this information!

Texts which are marked with this symbol and printed in bold contain important
information which may also include information regarding the prevention of
damage to property!
It is essential that you heed this information!

Texts which are marked with this symbol contain comments, instructions or tips.

This dot marks instructions for direct action.

– This dash marks parts of an enumeration.

2 V. 1.0
Martin ME 102 Service Manual

2. Safety Instructions
2.1 General Information
The ME 102 Electrosurgery Unit and the accompanying components and individual
parts, both individually and as a whole, comply with the currently applicable safety
standards and corresponding guidelines of the European Union.
The device and its external components (accessory parts) are safe when used for their
intended purpose and in accordance with the descriptions and instructions in this
documentation.
Nevertheless, hazards may occur in connection with the device or its external
components!
For this reason we expressly advise all persons who work with the
Electrosurgery Unit to obtain information about the hazards related to the
device and its components before carrying out any work using the device!
In addition, read and observe all of the safety instructions in this Service Manual
and in the Operating Instructions for the device!

2.2 Areas of Responsibility


– All service work (maintenance, repairs and technical safety inspections) may
only be carried out by service technicians who have been authorised by the
manufacturer!
– After carrying out service work the service technician has to check that the
device as a whole is functioning perfectly and is safe for operation!
– All work carried out has to be certified by the service technician.

For your own safety, read the safety instructions which follow carefully and
heed the information provided while carrying out your work!
Carry out all service work
– in accordance with the descriptions in this Service Manual,
– with technical correctness and the greatest possible precision and
– adhering to the relevant safety and accident prevention regulations!

V. 1.0 3
Martin ME 102 Service Manual

2.3 Protection of persons


When service work is carried out on a device which is operational there is a risk
for the service personnel of a potentially fatal electrical shock, as well as danger
of pacemaker malfunctions and the risk of burns caused by the HF-Currents
produced by the device!
For all work for which it is not absolutely essential that the device be operational,
separate the device from the electrical power supply and remove the electrodes and
foot switch connected to the device!
Under no circumstances should personnel who have an implanted pacemaker carry
out service work on an operational device!
Remove jewellery which could conduct electricity, e.g. chains, bracelets or rings,
before commencing work!
Use sufficiently insulated tools for all work!
Do not reach into an operational device with bare hands!
If appropriate, inform employees that service work is being carried out and of the
accompanying danger!

If the device receives insufficient maintenance or repairs there is a risk for


patients, employees or third parties of a potentially fatal electrical shock, as well
as a danger of pacemaker malfunctions and the risk of unintentional burns!

Carry out all work with extreme care!


Do not repair damaged components or individual parts under any circumstances.
Replace them with original replacement parts from the manufacturer!
Carry out all work in accordance with the descriptions in this Service Manual!
Only release the device for operation if it is undamaged and fully functional, all
recurring technical safety inspections can and have been carried out and you have
completely filled out the inspection certificate and signed it!

During all service and maintenance work there is a risk of infection for the
service personnel resulting from the possible adherence of pathogens to the
device or its accessories!
If it has not already been done, disinfect the device in accordance with the operating
instructions before starting work.

4 V. 1.0
Martin ME 102 Service Manual

2.4 Protection of devices


During service work on operational devices there is a danger of causing device
defects if you bridge contacts with measuring equipment or tools, when plug
connections are inserted or removed , or if materials capable of conducting
electricity, which are not part of the device, come into contact with it!

For all work for which it is not absolutely essential that the device be operational,
separate the device from the electrical power supply and remove the electrodes and
foot switch connected to the device!
Remove jewellery which could conduct electricity, e.g. chains, bracelets or rings,
before commencing work!
If appropriate, inform employees that service work is being carried out and of the
accompanying danger!

As with all CMOS products, even when they are not operational, there is a risk
that semi-conductor components may be damaged by the discharge of electrical
currents!

During all work – in particular when replacing circuit boards – wear a CMOS
grounding arm band which is connected to the earth via a high megohm resistor
matrix!
Do not remove CMOS replacement parts from the anti-static packaging until
immediately before installation and avoid touching them in conductive areas!

If replacement parts are used which have not been inspected by the
manufacturer there is a risk of device defects which may only occur after the
servicing!

Do not repair damaged components or individual parts under any circumstances.


Replace them with original replacement parts from the manufacturer!

V. 1.0 5
Martin ME 102 Service Manual

3. Device Description
3.1 Front

1 MICRO key
2 Signal light for monopolar cutting
3 Power output controller for monopolar cutting
4 Signal light for monopolar coagulation
5 Power output controller for monopolar coagulation
6 Signal light for bipolar coagulation
7 Power output controller for bipolar coagulation
8 Connection for bipolar instruments
9 Change-over key for foot switch activation (bipolar)
10 Change-over key for bipolar automatic
11 Connection for monopolar handpiece
12 Socket for 3-Pin handpiece
13 Connection for neutral electrode (NE)
14 Signal light for the neutral electrode alarm
15 Change-over key for monopolar PURE CUT/BLEND CUT
16 Power switch

Fig. 3-1

6 V. 1.0
Martin ME 102 Service Manual

3.2 Back

1 Connection for monopolar dual pedal foot switch


2 Primary fuse holder
3 Potential equalisation bolt
4 Connection for mains power cable
5 Sticker with details of the input voltage range
6 Connection for bipolar foot switch
7 CE designation
8 Rating plate

Fig. 3-2

V. 1.0 7
Martin ME 102 Service Manual

3.3 Overview of components

1 Front with front PCB A4, jacks and potentiometers


2 HF-Generator PCB A2
3 Mains transformer
4 Mains connection socket with fuse holder
5 Potential equalisation bolt
6 Foot switch sockets
7 HF-Output PCB A3
8 SMPS (switch mode power supply) PCB A1

Fig. 3-3

8 V. 1.0
Martin ME 102 Service Manual

3.4 Wiring diagram

Fig. 3-4

V. 1.0 9
Martin ME 102 Service Manual

4. Switch description
4.1 Block switch diagram

Fig. 4-1

10 V. 1.0
Martin ME 102 Service Manual

4.2 Description of the Function Groups


The block diagram illustrates the individual function groups and how they are
connected. For the description of the function groups you should also refer to the
wiring plan in Chapter 3.4 of this Service Manual.

Mains Power Supply and Power Output Controller


Two separate switching circuits are installed together on the SMPS (switched mode
power supply) PCB A1. There is a switching controller which operates on the principle
of a step-down regulator. It supplies the 15 V-supply for the signal and control
electronics and is protected at the input side by the F2 (M1A) fuse.
The power output controller is the same type of circuit based on the principle of a step-
down regulator. Its output voltage P1 forms the energy source for the HF-Generator
and is controlled by the analogue signal U1. P1 has double the voltage level of U1.
The power output controller is protected at the input side by the F1 (M4A) fuse.
The power supply and the power output controller are supplied by two separate
secondary windings of the mains transformer T1.

HF-Generator and Modulator


These are located in the middle in the right third of the HF-Generator PCB A2 and,
together, they form the HF-Current generator. The modulator affects the supply voltage
of the HF-Generator P1. The symmetrical outputs PH are conducted over the respec-
tive active output transformers. The voltage P1 is conducted to the HF-Generator via
the middle point of the output transformers.

Monopolar Output Circuit


This constitutes an interface to the monopolar application part and contains the neutral
electrode monitor, the auxiliary supply for the hand piece control, the driver circuit for
the monopolar HF-Indicator lights, the relays in the HF-Output and a load switch for the
generator. All of the parts of the monopolar output circuit, as well as the transformer to
the application part, are located at the back on the right of the HF-Generator PCB A2
under the (monopolar) HF-Output PCB A3.
The neutral electrode monitor operates with a harmonic oscillator, the resonant circuit
of which is conducted via the transformer T300 to the two partial connections of the
NE-connection socket. When a neutral electrode with a bridged double cable on the
end (one-piece NE) is connected, or the correct application of a two-piece NE, the
oscillator is attenuated by the shunt. If the electrode is missing, the cable breaks, or
the application is made incorrectly, this attenuation decreases.
The auxiliary supply for the handpiece control consists of an oscillator, the voltage of
which is transferred to the application piece by the transformer T302.
The driver circuit for the HF-Indicator lights is controlled by an auxiliary winding of the
monopolar HF-Output transformer T301.

V. 1.0 11
Martin ME 102 Service Manual

The HF-Generator has the unusual characteristic, that it stimulates oscillations in its
auxiliary power voltage, in doing so, however, it cannot give off any power. If the
attenuation is too slight (idle operation) and the setting is low (micro) it is possible for
the voltage P1 to increase to values higher than the set values in reaction to the
resonant circuit. If this happens the fault recognition would respond. For this reason,
for low settings, a constant load is switched to the generator.

Monopolar Application Part


This constitutes the monopolar patient interface and is located on the HF-Output
PCB A3. The transformers which constitute the interface to the monopolar output
circuit are located on the HF-Generator PCB A2. The output socket is located on the
front PCB A4.
The handpiece control decodes the signals of the finger switch on the electrode
handpiece and transfers them via opto-electronic couplers to the monopolar switch-on
control. These opto-electronic couplers are constructed of separate transmitters and
receivers which are connected by a plastic fibre-optical waveguide. The receivers are
located at the front on the HF-Generator PCB A2. The active electrode connections are
designed in such a way that the MARTIN-coaxial plug or the three terminal flat plug
found in the North American market can be connected. The handpiece control is
supplied via the transformer T302 from the monopolar output circuit.

Bipolar Output Circuit


This constitutes the interface to the bipolar application part, contains the automatic
bipolar activation, a driver circuit for the bipolar HF-Indicator and is located at the front
right on the HF-Generator PCB A2.
The automatic bipolar activation enables the bipolar HF-Power to be switched on
automatically when applying the bipolar forceps. It functions according to the principle
of a attenuated harmonic oscillator. Its resonant circuit voltage is laid via a HF-Output
transformer on the shank of the bipolar forceps which are connected. The shunt
through the tissue between the forceps attenuates the oscillator which serves to
generate a switch-on signal.
The driver for the bipolar HF-Indicator light is controlled by an auxiliary winding of the
bipolar HF-Output transformer and produces a signal which indicates the presence of
a bipolar HF-Current.

Bipolar Application Part


This constitutes the bipolar patient interface and, together with the bipolar HF-Output
transformer, it is located at the front right on the HF-Generator PCB A2.

12 V. 1.0
Martin ME 102 Service Manual

Switch-on Control
This translates the signals from the finger switch and the foot switch into an appropriate
selection of the operational mode with the use of corresponding relays and selection of
the path for the setpoint values. It is located at the back on the left and in the front in
the middle (fibre optic receiver) on the HF-Generator PCB A2.

Setpoint Generator
This is located at the front left on the HF-Generator PCB A2 and consists of two
separate circuits for the generation of the setpoint value (W1) and the reference
setpoint value (G1).
By means of various control signals, one of the three potentiometers in the front is
selected and, depending on the mode of operation, (cutting I, II and micro-modus) a
setpoint value adjustment is carried out.

Fault Monitor
This is located at the back on the left of the HF-Generator PCB A2 and releases the
HF-Generator if no fault exists.
In the monopolar operation mode a neutral electrode fault causes a block in the
HF-Generator. The bipolar operation mode can be activated even if a connected
neutral electrode is defective due to the fact that, for bipolar coagulation, a neutral
electrode fault is insignificant.

Sound Generator
This is located on the left of the HF-Generator PCB A2 and consists of a voltage
controlled oscillator, an indicator and an interval switch. The indicator, which is
controlled by the neutral electrode fault signal, activates the NE fault indicator light
on the front and activates the audio signal if there is a fault. When the audio signal
has been active for approx. 15 seconds the interval switch increases the volume.

Key Interface
This is located on the front left of the HF-Generator PCB A2. It stores the (short time)
activation of the keys and acknowledges activation by illuminating the keys. This takes
place via bistable relays.

V. 1.0 13
Martin ME 102 Service Manual

4.3 Description of important signals


Manipulated variable U1
The analogue signal "manipulated variable" is generated by the fault monitor and
constitutes the control voltage for the power output controller.
The value range of the manipulated variable U1 is between 0 and 12 V when one
of the operation modes is activated.

Input direct current voltage P1


The voltage "input direct current voltage" is generated by the power output controller
and constitutes the source of energy for the HF-Generator, as well as the actual value
for the fault monitor.
Influencing this voltage cause a modulation.
Average range: 0 - 24 V.

14 V. 1.0
Martin ME 102 Service Manual

5. Trouble shooting

Fault Possible cause Fault correction


– Mains switch light No mains voltage Check the power supply
does not light up, Power cable not plugged Check the connection of the power cable
– indicator elements in or incorrectly plugged
on the front PCB do into the power outlet or
not light up, the device
– device does not Device not switched on Switch on the device
function
Mains power fuse – Check the fuses in the power supply
defective connection of the device and replace if
necessary
– Check the mains voltage details on
the back of the device and change-
over the mains voltage, if necessary,
following the instructions in
Chapter 7.6
– If the mains fuse immediately blows
again there is an fault in the mains
transformer (short)
In spite of intact mains Fault (interruption) in the mains switch or
fuse, no mains voltage in the mains socket
on connections 2 and 6
of the mains transformer
– Mains switch light Fuse F2 of the SMPS Replace SMPS
lights up, defective
– indicator elements Fuse F2 of the SMPS Check the secondary voltage on the
on the front do not not defective mains transformer connections 12 and 14
light up, Is there approx. 21 V AC?
– device does not No: mains transformer defective
function (interruption)
Yes: check cooling unit on the right on
the outside of the SMPS for
warmth
Hot: short circuit on the HF-Generator
PCB
Cold: SMPS defective
– Mains switch light Defect in the power path Check secondary voltage on the mains
lights up, of the HF-Generator transformer connections 8 and 10. Is
– indicator elements there approx. 33 V AC ?
on the front PCB and No: mains transformer defective
the activation sound (interruption)
function normally Yes: fuse F1 in SMPS defective?
but there is no Yes: SMPS defective
HF-Power output No: HF-Generator defective

HF-Output power Fault in the setpoint Replace setpoint value potentiometer


failure only for certain value potentiometer for (connect a potentiometer for a different
types of power this type of power or its type of power for testing purposes)
connection

V. 1.0 15
Martin ME 102 Service Manual

6. Assembling and disassembling modules


6.1 Opening and closing the housing

Fig. 6-1

Separate the device from the power supply and remove all connected accessories.
Remove the fastening screws (1) on the back of the device and pull off the upper
part of the housing towards the back.

Fig. 6-2

To assemble the housing push the upper part of the housing onto the lower part so
that the guide strips (3) run along the lower side. If the upper part of the housing
does not fit easily into the groove of the front (2), adjust it. Secure the housing with
the three fastening screws (1).

16 V. 1.0
Martin ME 102 Service Manual

6.2 Front, front PCB, sockets


6.2.1 Removing the rotary knob

Fig. 6-3

Stick a piece of tape on the rotary knob as illustrated in Fig. 6-3.


Remove the rotary knob with flat nosed pliers parallel to the axis.

To avoid damaging the knob, grip only the tape on the rotary knob with the
pliers!

Remove the tape and, if necessary, any remnants of tape.

6.2.2 Replacing the potentiometer


Pull off the rotary knob parallel to the axis, as described above.
Loosen the corresponding plug connection on the HF-Generator PCB.
Remove the potentiometer nut with a suitable socket wrench or pliers.
Replace the potentiometer. Hold the potentiometer in place from the back while
tightening the nut.
Place the rotary knob on the potentiometer axis and check that the limit stops of
the potentiometer coincide with the printed scale.
Plug the plug connection into the appropriate counterpart on the HF-Generator PCB.

V. 1.0 17
Martin ME 102 Service Manual

6.2.3 Removing and replacing the front

Fig. 6-4

Loosen the four fastening screws (1) of the front.


Remove the ribbon cable retaining bracket (2) with suitable tweezers.

Fig. 6-5

Lift the front part until it can be removed from the front.
Remove the plug-in connections (3, 4 and 5).

18 V. 1.0
Martin ME 102 Service Manual

Fig. 6-6

Make a sketch of the connection assignment of the mains power switch.


Remove the mains power switch connections by gently moving them back and forth
with pointed pliers.

Fig. 6-7

Remove the protective earth conductor connection (6).

To replace the front proceed in the reverse order.

V. 1.0 19
Martin ME 102 Service Manual

6.2.4 Replacing the mains power switch or front sheeting

Fig. 6-8

To remove the mains switch press it out with both thumbs.


In order to replace the front sheeting, also remove the rotary knob as described in
Chapter 6.2.1 and carefully pull off the front sheet(s).

Fig. 6-9

Place the new front sheet(s) on the upper edge of the recess and check that the
sheet is positioned correctly on the edges of the rotary knob before pressing down.
Insert the rotary knob and the mains power switch (1 facing upwards) into the front
from the front.

20 V. 1.0
Martin ME 102 Service Manual

6.2.5 Removing the NE-connection socket

Fig. 6-10

Insert a 6 mm-Allen key in the NE connection socket from the front and unscrew the
socket ! clockwise !.

6.2.6 Removing the front PCB

Fig. 6-11

Lay the front on a soft surface and remove the seven fastening screws from the front
PCB with a Phillips screwdriver.

V. 1.0 21
Martin ME 102 Service Manual

6.2.7 Removing the AE Connection Combination

Fig. 6-12

Press the shanks of the circlip ring apart with snap ring pliers (5) and remove the
circlip ring towards the back.
Remove the AE-socket from the recess.
Press the two sockets for the 3-pin-handpiece (1) out of the board and slide their
connection cables (4) through the slots (2) of the board in the recess of the AE-
socket.
Remove the AE-connection combination by pressing the plug (3) through the
recess.

6.2.8 Removing the bipolar connection socket


Press the shanks of the circlip ring apart with snap ring pliers and remove the
circlip ring towards the back.
remove the bipolar-connection socket (Fig. 6-12, Pos. 6) by pressing the plugs
through the recess.

The installation of the front PCB and the sockets which are attached to it takes place in
the reverse order.

22 V. 1.0
Martin ME 102 Service Manual

6.3 Mains connection socket, power transformer, SMPS (switched mode


power supply)
6.3.1 Removing the mains connection socket

Fig. 6-13

If necessary, remove the mains connection cable from the mains connection
socket!
Remove the two cable lugs (1) by carefully moving them back and forth by hand.
Remove the two fastening screws (4) e.g. with a socket wrench.
Press the connection socket slightly outwards and remove the protective conductor
connection (2).

6.3.2 Removing the power transformer/SMPS (switched mode power


supply)
If necessary, remove the mains connection cable from the mains connection
socket!
Remove the two cable lugs (Fig. 6-13, Pos. 1) on the mains connection socket by
carefully moving them back and forth.
Remove the two fastening screws (Fig. 6-13, Pos. 3) and the washers from the
SMPS, e.g. with a socket wrench.
Remove the SMPS from the fastening screws and turn it 90 º upwards
(see Fig. 6-14).

V. 1.0 23
Martin ME 102 Service Manual

Fig. 6-14

Remove the plug-in connector (1) and take out the SMPS.
In order to remove the transformer; disassemble the HF-Output PCB (2) as
described in Chapter 6.4.1.

Fig. 6-15

If necessary, loosen the front and make a sketch of the connection assignment of
the mains power switch.
Remove the mains power switch connections by moving them gently from side to
side with pointed pliers.

24 V. 1.0
Martin ME 102 Service Manual

Fig. 6-16

Remove the two plastic fastening nuts (1) of the HF-Generator PCB and pull out the
mains cables (2) carefully.
Place the device on its side and remove the transformer fastening screws with a
Philips screwdriver(see Fig. 6-17).

Fig. 6-17

To install the transformer, SMPS and mains connection socket proceed in the reverse
order.
When screwing down the transformer take care that the tooth lock washer is in
the correct position!

V. 1.0 25
Martin ME 102 Service Manual

6.4 HF-Output PCB, Foot Switch Sockets, HF-Generator PCB


6.4.1 Removing the HF-Output PCB

Fig. 6-18

Remove the plug-in connection (2), by pressing the circlips on the under side
downwards and pulling out the plug-in connections.
Loosen the retaining nuts of the optical waveguide transmitter (5) and pull out the
plastic optical waveguide.
Remove the plastic nuts (1).
Take out the HF-Output PCB (3) and its safety cover (6) from above.
In order to disassemble the HF-Generator PCB also remove the three plastic
spacers (4).

6.4.2 Foot Switch Sockets


There is a special tool for removing the foot switch sockets but this can only be used
when the HF-Output PCB and transformer have been disassembled. With a bit of skill,
however, the retaining nuts can also be loosened with a fine screwdriver.

26 V. 1.0
Martin ME 102 Service Manual

6.4.3 Removing the HF-Generator PCB

Fig. 6-19

The HF-Generator PCB is fitted with CMOS components!

When replacing the HF-Generator PCB heed the safety instructions in


Chapter 2.4.
Remove the front and the HF-Output PCB.
Loosen the retaining nuts of the optical waveguide receiver (1) and pull out the
plastic optical waveguide.
Remove all plastic fastening nuts or spacers.
Take out the HF-Generator PCB from above.

To install the HF-Output PCB, foot switch sockets and HF-Generator PCB proceed in
the reverse order.

V. 1.0 27
Martin ME 102 Service Manual

7. Adjustment, Modifications
7.1 Position of the adjustment points

Fig. 7-1

All alignment positions are located on the HF-Generator PCB.


These are:
– SK: Setting for the output power of the operation mode (spray) coagulation
– S2: Setting for the output power of the current type cutting II
– NE: Setting for the deployment point of the neutral electrode monitor
– BI: Setting for the deployment point of the bipolar automatic activation

7.2 Fundamental points regarding HF-Output power measurement


In contrast to other electrical devices, the measurement of electric high frequency
output is subject to large tolerances. As a result, the regulations treat measurement
values which deviate as much as +/- 20 % from the nominal value as correct.
For the ME 102 device the following tolerances in the measurement are permissible:
Cutting I 100 W at 600 ohm -20/+20 % (80 - 120 W)
Cutting II 90 W at 600 ohm -20/+20 % (72 - 108 W)
Coagulation 70 W at 600 ohm -20/+20 % (56 - 84 W)
Bipolar Coagulation 80 W at 100 ohm -20/+20 % (64 - 96 W)

The devices are set in the plant with the MARTIN-HF-Output measurement devices
EPM 2 and FM2097.
In addition, please note the output diagrams in Chapter 9 Technical Data.

28 V. 1.0
Martin ME 102 Service Manual

7.3 Adjustment of the HF-Output power


On the ME 102 device there is no opportunity to adjust the operating modes "Cutting I"
and "Bipolar Coagulation". The output power for the operating modes "Cutting II" and
"(Spray-)Coagulation" are adjusted by setting the degree of modulation.

The setting for the degree of modulation, in particular the setting "SK" for the
(Spray-)Coagulation, should only be moved from its original position very slowly,
otherwise, there is a risk that the degree of modulation will be set too high,
which could result in damage to the SMPS or the HF-Generator!
Heed the calibration breaks on the MARTIN output measurement devices!

To set "Cutting II" proceed as follows:


Connect the output measurement device and set it to 600 ohm.
Select the current type "Cutting II" (shut off "Micro" mode) and set the left power
output controller to 10.
Activate the operating mode "Cutting" and, by carefully turning the trimmer "S2", set
90 W output power.
(For more power turn to the left, for less to the right.)

To set "(Spray-)Coagulation" proceed as follows:


Connect the output measurement device and set it to 600 ohm.
Switch off the "Micro" mode and set the middle power output controller to 10.
Activate the operating mode "Coagulation" and, by carefully turning the trimmer
"SK", set 70 W output power.
(For more power turn to the left, for less to the right.)

7.4 Adjustment of the NE Monitor


Connect a 230-ohm-resistor to the contacts of the NE-connection socket.
(solder 2 x 470-ohm-resistors in customary 6.3 mm jack plugs).
Move the trimmer "NE" clockwise until the NE fault indicator light blinks.
Turn the trimmer "NE" slowly anti-clockwise until the blinking stops.
Check that it is functioning correctly by unplugging the plug and plugging it in again.

V. 1.0 29
Martin ME 102 Service Manual

7.5 Adjustment of the switch-on threshold for bipolar-automatic


Set the right power output controller to 1 and activate the "Micro" and "Auto-Bipolar"
modes.
Connect a resistor with 2000 ohm to the bipolar connection socket. (Wind the ends
of the resistor wires around the branches with the forceps removed).
Be certain to set the power to a minimum, otherwise there is a risk of burns!
Turn the trimmer "BI" clockwise until the activation shuts off, then slowly counter
clockwise until the activation restarts.
Check that it is functioning correctly by unplugging the bipolar plug and plugging it in
again.

7.6 Changing over the mains voltage

Wiring for 220 V - 240 V


8
7

Wire link:
9
6

Pin 3 - 4
10
5

11
4

Fuses:
12
3

2 x 1.0 AT
13
2

14
1

Wiring for 100 V - 127 V


8
7

Wire link:
9
6

Pin 2 - 3 and 4 - 5
10
5

11
4

Fuses:
12
3

2 x 2.0 AT
13
2

14
1

Fig. 7-2

30 V. 1.0
Martin ME 102 Service Manual

The ME 102 device is set up for the voltage of the market for which it is intended. If a
change-over to a different voltage becomes necessary the connections on the primary
side of the mains power transformer have to be desoldered and the power fuses
replaced.

The change-over from 230 V to 115 V is made as follows:


Remove the mains power cable and open the housing.
Desolder the wire link between the transformer connections 3 and 4.
Connect the transformer connections 2 and 3, as well as 4 and 5, with wire links.
Change the mains power fuses in the mains connection combination on the back of
the device (use 2 x T 2.0 A ).
Replace or cover the stick-on label with the label "Voltage and Fuse Selection" and
tick the new input voltage.
Close the housing.

The change-over from 115 V to 230 V is made as follows:


Remove the mains power cable and open the housing.
Desolder the wire link between the transformer connections 2 and 3, as well as
4 and 5.
Connect the transformer connections 3 and 4 with a wire link.
Change the mains power fuses in the mains connection combination on the back of
the device (use 2 x T 1.0 A ).
Replace or cover the stick-on label with the label "Voltage and Fuse Selection" and
tick the new input voltage.
Close the housing.

V. 1.0 31
Martin ME 102 Service Manual

7.7 Determination of the HF-Leakage currents


7.7.1 Cable connections in the "Monopolar" operating mode

ME 102

NE AE

EPM

Foot-Switch Power-Meter-Var. Load Fixed Load


out 10R...3175R 200R

mA

( R=200R )

Fig. 7-3

32 V. 1.0
Martin ME 102 Service Manual

7.7.2 Cable connections in the "Bipolar" operating mode

ME 102

NE AE

EPM

Foot-Switch Power-Meter-Var. Load Fixed Load


out 10R...3175R 200R

mA

( R=200R )

Fig. 7-4

V. 1.0 33
Martin ME 102 Service Manual

8. Recurring technical safety inspections


8.1 Overview of the inspection activities to be carried out
The following inspections of this device have to be made every 24 months by persons
who are capable of correctly implementing such inspections, as a result of their
training, their knowledge and their practical experience, and are not in any way
restricted in their capacity to carry out these inspections.
– Make a visual inspection of the device and accessories for mechanical damage
which could influence functions. Check that security relevant markings are readable.
– Check the fuse units of the device protection fuses for nominal current and current
characteristic.
– Check the functions according to the operating instructions.

– Check that the increase or decrease in the energy output corresponds correctly to
the direction the potentiometer is turned.
– Check the acoustic signal, as well as the HF-Power shutoff, when the neutral
electrode connection is interrupted.
– Compare the setpoint value and actual values of the maximum output power for all
four types of current on their respective nominal resistors in accordance with
Chapter 7.
– Compare the setpoint value and actual values of the maximum output power in the
"Cutting I“ mode on the monopolar output with 300 ohm load. The output power has
to be less than for a termination with 600 ohm.
– Electrical inspection in accordance with the inspection report for recurring technical
inspections.

The standby leakage currents may not be greater than 1.5 times the first
measured values and, at the same time, no greater than the limit values.
An entry has to be made in the device book that the technical safety inspection
has been carried out and the results of the inspection recorded.

34 V. 1.0
Martin ME 102 Service Manual

8.2 Inspection Report

Inspection Report Inventory No.:

Inspector: Device No.: ME 102

Manufacturer: Martin Medizin-Technik

Certificate holder: Kind of unit: Electrosurgery Unit

Device Type: ME 102

Year of Production:

Inspection Standard: EN60 601-1

Inspection Result:

1. See reverse side of this report for measurement values.

2. Inspection points not fulfilled: _______

3. No faults or only faults which are not relevant for safety.


The device may continue to be used.

4. The device may be returned to operation when the points listed above have
been corrected.

5. Faults which require that the device be repaired or maintained before being
used again because they present a hazard for patients, employees or third
parties.

Date: Signature: Next inspection due on:


____________ ____________________ _____________________

V. 1.0 35
Martin ME 102 Service Manual

Inspector: Consecutive No:

Inspection Report: Type: ME 102 Device No.:

i.o. n. i. o. N/A Fault


1. Rating plate
2. Operating Instructions
3. Labelling
4. Operation components
5. Equipotential Connection
6. Original-Accessories? (Otherwise, name of manufacturer)
7. Visual inspection of HF-Cable
8. Foot switch in OP waterproof and AP-protected
9. Operation mode of neutral electrode marked
10. If neutral electrode missing, no output power
(for devices with more than 50 W HF-Output)
11. Monitoring circuit of the neutral electrode (acoustic
signal, for devices with more than 50 W HF-Output)
12. Function check of hand and foot switch
13. Check of the optical and acoustic signal function for
HF-Output
14. HF-Output measurement Cutting I at 600 ohm: ______ watt
(maximum at nominal resistance): Cutting II at 600 ohm: ______ watt
Coagulation at 600 ohm: ______ watt
Bipolar Coagulation at 100 ohm: ______ watt

Comments:

Electrical Measurement: Inspected according to EN60 601-1


15. Insulation resistance between power supply and housing: ______ MΩ
16. Insulation resistance of application part: ______ MΩ
17. Equipment grounding conductor resistance measurement: ______ Ω
18. Low frequency leakage current, normal operation ______ µA
19. Low frequency leakage current, initial fault: ______ µA
20. Housing leakage current, normal operation: ______ µA
21. Housing leakage current, initial fault grounding conductor: ______ µA
22. Housing leakage current, initial fault power supply: ______ µA
23. Patient leakage current, normal operation: ______ µA
24. Patient leakage current, single fault grounding conductor: ______ µA
25. Patient leakage current, single fault power supply: ______ µA
26. Patient auxiliary current, normal operation: ______ µA
27. Patient auxiliary current, initial fault grounding conductor: ______ µA
28. Patient auxiliary current, initial fault mains power supply: ______ µA
29. Patient leakage current, supply-voltage on application part parallel: ______ µA
30. dto., interchanged phases: ______ µA

Unit inspected on: by:

36 V. 1.0
Martin ME 102 Service Manual

9. Technical Data, Diagrams, Replacement Parts


9.1 General Data

mains power connection 100 - 127 V AC/220 - 240 V AC; 50/60 Hz


can be set with solder bridge inside device
power consumption without HF-Output approx. 16 VA
at max. output power approx. 220 VA
protection class I
classification according to MPG II b
leak currents LF and HF in accordance with IEC 601, Part 2-2
type CF; defibrillation proof
nominal frequency 500 kHz
modulation frequency 30 kHz
operation mode intermittent INT 10 s/30 s corresp. 25 % ED
mains supply fuses 100 - 127 V: T 2,0 A
220 - 240 V: T 1,0 A
signal level HF-Display: 55 dB (A)
alarm: 65 dB (A)
weight approx. 4.6 kg
interference suppression limit values accord. to EN 55011 and IEC 601-2-2
interference resistant accord. to IEC 801
dimensions width 256 mm
height 97 mm
depth 320 mm
identification conforms to 93/42/EEC

V. 1.0 37
Martin ME 102 Service Manual

9.2 "Macro Mode" Performance Charts

ME 102 Cutting I ME 102 Cutting II

100 100
90 90
80 80

HF-Power in watt
70 70
HF-Power in watt

pos. 10 pos. 10
60 60
50 50
40 40
30 30
pos. 5 pos. 5
20 20
10 10
0 0
0 200 400 600 800 1000 1200 1400 1600 1800 2000 0 200 400 600 800 1000 1200 1400 1600 1800 2000
Connection resistance in ohm Connection resistance in ohm

ME 102 Coagulation ME 102 Bipolar Coagulation

100 80
90 70
80
60
HF-Power in watt

70
HF-Power in watt

60 50
pos. 10
50 40
40
30
30
pos. 5 pos 10
20 20
10
10
0
0 pos 5
0 200 400 600 800 1000 1200 1400 1600 1800 2000
0 100 200 300 400 600 700 800 900 1000
Connection resistance in ohm
Connection resistance in ohm

Crest-
Macro Mode Output Voltage VSS *
Factor
Cutting I max. 100 W ± 20 % at 600 Ω 1200 V ± 10 % 1.8
Cutting II max. 90 W ± 20 % at 600 Ω 1600 V ± 10 % 2.3
Coagulation max. 70 W ± 20 % at 600 Ω 2000 V ± 10 % 3.0
Bipolar Coagulation max. 80 W ± 20 % at 100 Ω 440 V ± 10 % 1.9

Pos. 5 or pos. 10 within the chart indicates the position of the power output controller
(Potentiometer) of the corresponding mode of operation.

* in idle operation

38 V. 1.0
Martin ME 102 Service Manual

9.3 "Micro Mode" Performance Charts

ME 102 Cutting I ME 102 Cutting II

30 30

25 25

HF-Power in watt
HF-Power in watt

20 20
pos. 10 pos. 10

15 15

10 10
pos. 5 pos. 5
5 5

0 0
0 200 400 600 800 1000 1200 1400 1600 1800 2000 0 200 400 600 800 1000 1200 1400 1600 1800 2000

Connection resistance in ohm Connection resistance in ohm

ME 102 Coagulation ME 102 Bipolar Coagulation

30 30

25 25
HF-Power in watt

HF-Power in watt

20 20
pos. 10

15 15

10 10
pos. 5
pos. 10
5 5

0 pos. 5
0
0 200 400 600 800 1000 1200 1400 1600 1800 2000 0 100 200 300 400 600 700 800 900 1000
Connection resistance in ohm Connection resistance in ohm

Crest-
Mico Mode Output Voltage VSS *
Factor
Cutting I max. 30 W ± 20 % at 600 Ω 600 V ± 10 % 1.8
Cutting II max. 30 W ± 20 % at 600 Ω 800 V ± 10 % 2.3
Coagulation max. 25 W ± 20 % at 600 Ω 1000 V ± 10 % 3.0
Bipolar Coagulation max. 25 W ± 20 % at 100 Ω 240 V ± 10 % 1.9

Pos. 5 or pos. 10 within the chart indicates the position of the power output controller
(Potentiometer) of the corresponding mode of operation.

* in idle operation

V. 1.0 39
Martin ME 102 Service Manual

9.4 ME 102 Replacement Parts

Item Description Order No.


1 Lower part of housing 08-012-00-15
2 Upper part of housing 08-012-00-02
3 Front 08-010-00-01
4 Front sheeting set ME 102 08-010-00-70
5 Rotary knob D 30 08-006-00-01
10 Mains power connection combination 08-024-00-01
11 Mains power fuse holder 08-034-00-01
12 Mains power transformer with cable 08-024-00-50
13 Connection for potential equalisation 08-027-00-01
14 Mains power switch 08-042-00-02
20 Setpoint value controller with cable 08-034-00-02
21 NE-connection socket with cable 08-024-00-04
24 AE-connection combination with cable 08-001-00-01
25 Bipolar-connection socket with cable 08-003-00-01
26 Monopolar foot switch socket with cable 08-022-00-01
27 Bipolar foot switch socket with cable 08-003-00-02
30 Front PCB with cable 08-010-00-21
31a HF-Generator PCB ME 102 08-014-00-30
31b HF-Generator PCB ME 102/2 08-014-00-33
32 HF-Output PCB ME 102 08-014-00-29
33 SMPS (switched mode power supply) PCB 08-032-00-01
40 Connector 5-pole power supply/HF-Generator 08-041-00-01
41 PE-connection cable set 08-027-00-02
42 Plastic optical guide 08-020-00-01
50 Upper part safety cover 08-034-00-03
51 Output PCB safety cover 08-034-00-04
52 Front safety cover 08-034-00-05
53 Locking bracket for DIL-plug 18-pole 08-041-00-02
–- Set of mechanical small parts ME 50/80/81/102 08-018-00-01
–- Set of adhesive labels ME-devices 08-018-00-02
–- Light bulb for front PCB 08-012-00-03
–- Fuse T 2.0 A ME 102 (100 V - 127 V) 08-084-00-13
–- Fuse T 1.0 A ME 102 (220 V - 240 V) 08-084-00-21
–- Fuse M 1.0 A auxiliary supply 08-034-00-10
–- Fuse M 4.0 A power output control 08-034-00-11

40 V. 1.0
Martin ME 102 Service Manual

Fig. 9-1

V. 1.0 41
Martin ME 102 Service Manual

10. Changes
10.1 Procedure
Experience has shown that during their period of operation technical products are
subject to repeated technical changes which result from improvements in the product.
This creates the problem for servicing of there being several versions of the same type
of device. Before a repair can be made the version of the device has to be established.

As a result, it is absolutely essential to achieve clarity about the version status


before undertaking any work on the device!

This is illustrated in the device number:

ME 102 XX XX XX XXXX

consecutive production number

software status

hardware status

device type

10.2 Hardware status


The hardware status is determined by:
– Revision status of the PCBs (printed circuit boards)
– Revision status of the mechanical configuration

– Revision status of the adjustment

There is a sticker on the PCBs with the production identification number C 40-XXXX
and the version status Rev XX or V XX
If it has the same MARTIN order number (in accordance with Chapter 9.4), a PCB with
a higher revision status can be used to replace the older original part (downward
compatibility). In general, this does not apply in the reverse!
Incompatible PCBs are assigned a new identification number C 40-XXXX and a new
MARTIN-order number.

42 V. 1.0
Martin ME 102 Service Manual

10.3 Software status


The software status is always 00 as the ME 102 is not equipped with software.

10.4 Configuration ME 102


10.4.1 Configuration 0000
Serial Number: ME 102 0000 99 0100 up to ME 102 0000 00 0999
SMPS PCB: C40-1046 Version 00 (Order No. 08-032-00-01)
Front PCB: C40-1360 Version 00 (Order No. 08-010-00-21)
HF-Output PCB : C40-1701 Version 00 (Order No. 08-014-00-29)
HF-Generator PCB: C40-1700 Version 01 (Order No. 08-014-00-30)

10.4.2 Configuration 0100


Serial Number: from ME 102 0100 01 1000 onwards
SMPS PCB: C40-1046 Version 00 (Order No. 08-032-00-01)
Front PCB: C40-1360 Version 00 (Order No. 08-010-00-21)
HF-Output PCB: C40-1701 Version 00 (Order No. 08-014-00-29)
HF-Generator PCB: C40-1755 Version 00 (Order No. 08-014-00-33)
(ME 102/2)

V. 1.0 43
Martin ME 102 Service Manual

11. Appendix

Comments on change Page No.. Rev-old Rev-new Date Symbol

New HF-Generator PCB


C 40-1755 (Bipol-Automatic 0000 0100 10/2000 rk/AS
is not saved)

44 V. 1.0
Gebrüder Martin GmbH & Co. KG 01.02 · 90-788-62-21 · Printed in Germany
Copyright by Gebrüder Martin GmbH & Co. KG
Ludwigstaler Straße 132 · D-78532 Tuttlingen Alle Rechte vorbehalten.
Postfach 60 · D-78501 Tuttlingen · Germany Technische Änderungen vorbehalten.
Telefon (0 74 61) 7 06-0 · Telefax (0 74 61) 70 61 93 We reserve the right to make alterations.
Cambios técnicos reservados.
E-mail: info@martin-med.com Sous réserve de modifications techniques.
Internet: www.martin-med.com Ci riserviamo il diritto di modifiche tecniche.

You might also like