0% found this document useful (0 votes)
586 views4 pages

Precimeasure: MODEL - 1007 SERIES...

The document summarizes specifications for Precimeasure model 1007 series dial type thermometers. Key features include fully sealed liquid expansion for self-powered operation, ambient temperature compensation, optional built-in PT-100 sensor and current converter unit (CCU) for hybrid models, accuracy of ±1.5% of full scale, up to 4 potential free microswitch contacts, and IP55 protection. Optional components include remote indicators with 4-20mA or Modbus outputs, and external power supply units.

Uploaded by

Prem Pagol
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
0% found this document useful (0 votes)
586 views4 pages

Precimeasure: MODEL - 1007 SERIES...

The document summarizes specifications for Precimeasure model 1007 series dial type thermometers. Key features include fully sealed liquid expansion for self-powered operation, ambient temperature compensation, optional built-in PT-100 sensor and current converter unit (CCU) for hybrid models, accuracy of ±1.5% of full scale, up to 4 potential free microswitch contacts, and IP55 protection. Optional components include remote indicators with 4-20mA or Modbus outputs, and external power supply units.

Uploaded by

Prem Pagol
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
Download as pdf or txt
You are on page 1/ 4

Precimeasure MODEL - 1007 SERIES....

DIAL TYPE THERMOMETER


General Features
 The fully sealed liquid expansion in the system provides a self powered indication and operation to the instruments,
and does not require any trimming or other periodic adjustments.

 Instruments are fully compensated for ambient temperature.

 The WTI instruments have a built in Thermal imaging system for transformer’s hot spot indication.

 WTIs are calibrated & supplied suitable for range of CT secondary input & temperature rise.

 Instrument cases are aluminium die casted and polyester powder coated, and provided with projection mounting clamps.
As per enclosure part no. EN7.

 A maximum pointer facility is provided, can be reset from outside.

 All switch contacts are potential free micro switches.

Hybrid Models Contain


 Embedded PT-100 sensor in the sensing bulb of the local OTI/WTI for master indication.
 Current Converter Unit (CCU) is mounted inside the Local OTI/WTI, CCU output to be connected to our External PSU or Remote indicator unit.

General Standard Specifications MODEL 1007/1007H


Temperature Range 0-1500C, Standard (other ranges on request)

Dial Size 115mm, Segment Dial

Dial angular Sweep 600

Dial Graduation 20C

Accuracy Class ±1.5% of FSD

Contacts Upto 4 Normally open Contacts

Contact Rating Standard Micro Switch15A @ 250V AC or 0.5A @ 250V DC

Switching Range 15% - 85% of FSD

Switch Differential S1& S2: 3°C to 9°C (6±3°C)Fixed S3 to S4: 8°C to 40°C Adjust-
able.

Switching Accuracy ±2.50C of Set Value

Max. Capillary Length Up to 25 Mtrs (Bottom Entry)

Electrical Connection Press Fit WagoType

Window Glass UV Resistant Polycarbonate

Degree of Protection IP – 55 (IEC - 60529 Cat -II)

Electrical entry M20 Threaded Holes - 3Nos

Hybrid Models Specification


Embedded Pt100 sensor Calibration as per IEC 60751-1985/DIN 43760
Built in CCU CCU is mounted inside the Local OTI/WTI
OTI Signal input Pt 100
Supply input 24v Dc-through our make PSU/Remote
indicator
For WTI Additional Winding temperature simulator is provided inside the
local WTI.

1/4
Optional Spec
Temperature Range 0-1600C / 30-1500C / -20-1400C / other range on request

Contact Rating Magnetic blowout Micro Switch 15A @ 250V AC or 5A @ 250V DC

Switch Configaration Change over Contacts

Cable Gland Brass / polyamide / others on request

Thermometer Pocket On request

Connector Type ½” / ¾” / 1” BSP M/F Brass / Aluminum Alloy

Capillary Entry Top Entry

Remote Indicator Options

Dial Type Thermometer Models Suitable Digital Remote


Remote Indicator Signal Inputs
Indicator Models

1007H Built in CCU Model 2245H Suitable Hybrid Model Connection from built-in
CCU inside Local OTI/WTI

1007H Built in CCU and external PSU Model 2245 Suitable for 4 - 20mA input from External PSU

Note: Refer Ordering information for various types of remote indicators

Remote Indicator General Specification


Type Quad Seven segment LED display

Resolution 0.10C

Auxiliary Supply input 90-260V AC/DC

Degree of protection IP 20

Operating temperature 0-600C

Mounting Cutout 92 sq.mm in panel, Mounting side clamp

Signal input Suitable for connection from built-in CCU inside the Local OTI/WTI / Ex-
ternal PSU

Signal output Refer table below (Remote Indicator Output Options) Remote Indicator 96 Sq

Power Cunsumption 5 Watts

PSU Spec PSU

Signal input Suitable for connection from built-in CCU inside Local OTI/WTI

Auxiliary supply input 90-260v AC/DC

Signal Output 2Nos. 4-20mA DC

Burden on signal outputs Maximum 600Ω including loop resistance

Mounting Din rail or Screw fixing

Degree of protection IP 40 for housing & IP 10 for terminals

Power Consumption 3Watts

Note: Available Scheme options referenced below. Please contact us for details.
1. Standard (1007)
2. Hybrid and External Remote Indicator (1007H + RI)
3. Hybrid and External PSU (1007H + PSU)
4. Hybrid & External PSU + Remote Indicator (1007H + PSU + RI)

2/4
Routine and Type Tests Basic Ordering Information
Conducted Select Your Model  Standard -1007
1 Routine test : HV Test and IR Test and Calibra-  with 1 inbuilt PT 100 Sensor (Simplex) & CCU - 1007H
tion Test
Switch Contact Rating OP1 - Option 1 - Standard Micro Switch
OP 2 - Option 2 - Magnetic Blowout Micro Switch
2 Protection class:
IP55 as per IEC 60529 CAT II. Type of Instrument OTI/WTI
Range R1.0 - 1500C
R2.0 - 1600C
3 Cold Test: -400C as per IEC 60068-2-1
R3.30 - 1500C
Dry Heat Test: + 700C as per IEC 60068-2-2
R4. -20 to 1400C
(other ranges on request)
4 Vibration Test: as per IEC 60068-2-6
No. of Switches S2 -2switches
Displacement : 0.15mm,
Frequency: 10-55Hz, S3 -3switches.........upto S4 switches
Duration: 1hr. 20min/X,Y,& Z axis
Switch Configuration L- Normally open contact
L2 - Changeover contact
5 Salt Spray Test: Capillary Length C3 - 3 mtrs C5 - 5 mtrs
For polyester powder coating and C8 - 8 mtrs C10 - 10 mtrs
SS 304 Armour for 500 Hrs as per C12 - 12 mtrs C15 - 15 mtrs
DIN 50021 Standard
C20 - 20 mtrs C25 - 25 mtrs
Union Type U1 - ½” BSP Male
6 Micro Switch rating Test: U2 - ¾” BSP Male
a) 15A @ 250V AC & U3 - 1” BSP Male
0.5A @ 250V DC U4 - ½”BSP Female
U5 - ¾” BSP Female
b) 15A @ 250V AC &
Other threads on request
5A @ 250V DC
Indicator RI - Remote Indicator (for options please see Note below)
If WTI please specify Transformer CT Secondary output ..............
Simulated Temperature rise over top oil ..........
Contact Precimeasure for gradient graph

Note:
Remote Indicator Options
T1 - Indication only.
T2 - 2Nos, 4-20mA DC outputs.
T3 - one 4-20mA DC output + 2Nos. Change over relay contacts
T4 - one 4-20mA DC output + RS 485 Modbus RTU protocol
T5 - 4Nos. of 4-20mA DC outputs.
Note: WTI Gradient calibration is factory set as per customer specification. If required,
a provision is made for small adjustment at site.
Sample Ordering Information : 1007\OP1\WTI\R1\S4\L2\C6\U1\RI-T1\........... Amps .......... oC

EN7 ENCLOSURE

3/4
Desclaimer : Precimeasure reserves the right to make any changes in the specification without prior notice.
4/4