Basic IR Knowledge
Basic IR Knowledge
Spectral Rang: There are two main bandwidths used in infrared testing. In general the
8-13 µm band is preferred for high performance thermal detectors because of the greater
sensitivity to ambient temperature objects and good transmission through smoke. The 3-5
µm band may be more appropriate for hotter objects or if sensitivity is less important than
contrast. For certain optical resolution it can use smaller optics, which may be useful in
some circumstances.
Black Body: An ideal thermal radiator emitting and absorbing all possible thermal radiation
at a given temperature, hence having an emissivity of 1.
Emissivity: The ratio of the radiance of a body at a given temperature to that of a black
body at the same temperature. Instruments will require emissivity compensation to permit
accurate temperature measurement across a range of materials and surface conditions.
Field of View (FOV): This is a function of the system optics and usually described in
degrees of arc in the vertical and horizontal planes. Some systems include changeable
lenses which will therefore change the FOV in the system.
Fill Factor: The ratio of the active area of each detector in an array to the inactive space
which surrounds each detector.
Filters: Spectral filters are used to adapt the infrared system response to objects which
have special spectral characteristics, such as measuring objects through flames,
measuring energy in the CO2 absorption band, or suppressing certain wavelengths.
Focal Plane Array (FPA): An array is any grouping of detectors which has more than one
vertical row and one horizontal line of detectors together. It could be as small as two
detectors a line and two rows, four detector elements in all. More usually such arrays are
256 lines and 256 rows. The term focal plane refers to the location of the detector array in
the optical path. The focal plane is that point where the image is focused.
Instantaneous Field of View (IFOV): This describes the optical resolution of the system
and may be expressed in milliradians (mRAD) or minutes of arc. It is equivalent to the
horizontal and vertical fields of view of an individual detector and as part of the overall
resolution of the system and can be calculated using IFOV = A1/2/FND, where A1/2 is the
square root of the linear dimension of the detector, FN is the f number of the optics (focal
length), and D is the diameter of the front objective.
Microbolometer Detector: These are thermal detectors (very small bolometers) and not
photon detectors. The detector (thermistor) actually heats up when exposed to infrared
energy, changing its electrical resistance proportionally, which can then be measured. No
cryogenic cooling devices are required, but images using this detector type are generally
less sensitive.
Modulation Transfer Function (MTF). MTF is a function of spatial frequency and in infrared
systems MTF is the mathematical description of the spatial distribution of amplitude
attenuation. The ability of an infrared system to transmit the spatial frequency of a scene
is described in terms of the MTF. The overall system MTF is obtained from the product of
the MTFs of its subsystems.
Optics: The lens system which focuses the scene on the detector. Transmission through
the optics can be as low as 60 percent and this has a direct affect on the NETD.
Staring Arrays: The pyroelectric vidicon is an example of a staring array in that the
thermally sensitive target is exposed to the scene for one TV field time, producing a low
NETD value with no cooling required. The staring array can have limitations in spatial
resolution and amplifier noise.
What is a pixel?
A pixel is the smallest single individual image element of detection on the thermal imaging
sensor.
What is a microbolometer?
A microbolometer is the latest type of thermal imaging FPA, which consists of materials
that measure heat by changing resistance at each pixel. The most common
microbolometer material is amorphous silicon. Although microbolometers do not require a
chopper to refresh the image, they must occasionally be recalibrated for the pixels to
provide a consistent output and to avoid oversaturation. The device that occasionally
(every 30 seconds to 5 minutes) and automatically recalibrates the FPA is called a
"shutter" (see definition).
What is a shutter?
A shutter is a mechanical device, generally shaped like a flag, which closes in front of the
detector to activate the calibration for a uniform temperature (or black body). This
automatic, periodic calibration is necessary because pixels in microbolometers drift and
cause image degradation.