BIOBASE Discrete Automatic Biochemistry Analyzer - Operator Manual
BIOBASE Discrete Automatic Biochemistry Analyzer - Operator Manual
, LTD
BIOBASE Discrete Automatic Biochemistry Analyzer
Operation Manual
Please read this operation manual before using the automatic biochemistry analyzer.
Keep the manual for accessing information.
Warning: biohazard
2.Overview
Automatic Biochemistry Analyzer BIOBASE-Crystal is the hospital’s routine in vitro
diagnostic equipment, and it is an instrument. Using in testing a variety of biochemistry
index in the body fluids of patients.
After having some knowledge of the components of BIOBASE- Crystal, we will teach
you step by step how to use these modules to complete routine tasks. After reading this
manual thoroughly, I believe you can operate BIOBASE- Crystal well.
We will use some warnings, hints and very useful comments emphasizing the key
chapters, which have a striking "Note" mark.
2.3 Users
BIOBASE-CRYSTAL is operated by professional people in accordance with the
manual.
2.4 Security
The following sections will introduce the note in terms of security, electronics and
other relevant potential dangers.
2.5General Prevention
When operating BIOBASE-CRYSTAL, the user should be strictly in compliance with
laboratory practice, and must wear suitable protective clothing and rubber gloves.
During the operation of BIOBASE-CRYSTAL, the sample arm moves quickly, so the
user should not place the hand in the work area of the moving parts so as to avoid
accidental injury. Meanwhile, the user should not rely on BIOBASE-CRYSTAL or tilt
it to avoid potential personal injury or equipment damage.
When disassembling the equipment components or circuit boards, the user should first
turn off the power and disconnect the power.
3. Introduction
3.1 Basic Principles
According to the Lambert - Beer law, when a bunch of incident light shoots through the
colorimetric solution, there is the law concerning the incident light, transmitted light
and the solution of light absorption.
A=-Log10T=KCL
T=I/I0
I0---- Incident light intensity
I----- Transmitted light intensity
A---- Absorbance
C----- Concentration
L----- Optical path
k----- Coefficient
The testing principle of biochemistry analyzer mainly applies light absorption: the
Lambert - Beer law, by measuring the absorbance of the monochromatic light which
shoots through the liquid mixture of the serum or urine and other body fluids and
reagents to determine whether human physiological indicators are normal.
3.2 Hardware
In order to auto-complete specimen test, hardware support is indispensable, which
includes the following aspects:
1 2 3
8 spectral groups
Light of light source shoots upon the fiber end and then is divided into eight beams
which act as the incident light shooting upon the cuvette. The beams become
monochromatic light after shooting through the solution and go into interference filter
spectrophotometric and then shoot a photoelectric receiver.
We choose the quartz glass halogen lamp of 12V 20W as thr light source, ant the light
source collimates into a parallel beam of 2.5mm diameter. We should use rectangular
discrete plexiglass cuvette of 6mm optical diameter through which UV can shoot
through cuvettes. Cuvettes should be uniform in the shape and there should be no
scratches on the surface of light transmission hole. Light transmittance slot of precise
120 attainment yards disk align at the middle of the cuvette so as to ensure the readings
sampling point is in the middle of the cuvette; 120 cuvettes are placed in a circle of the
reaction plate with the smallest volume of 180ul. The advantage is that there is no cross
contamination. Taking into account the speed of rotating colorimetric turntable, we use
static eight-way splitter system, and integrated components package consisting of
import filters and optical receiver to test wavelength of 340 ,
405,450,510,546,578,630,700 with spectral half width of less than 10nm ± 2nm, so that
we can collect eight wavelength data at once without any mechanical movement, and
overcome the defect that the reaction disk of some of the imported products and the
filter halo also rotate at the same time so as to ensure measurement accuracy.
3.2.2 Circuit
Cuir
pipetting arm control reagent disc control
-cuit
-con
syringe control circuit co
necti RS-232
data acquisition mp
ng
ute
com
colorimetric turntable sample disk control r
-put
sy
er
The interface circuit consists of two 8051 single chip microcomputers and some
peripheral circuits and is used to the convert the command sent by the computer into
executable command for various moving parts so as to coordinate the work of the
various components; meanwhile it sends the A / D conversed digital signal and
instrument status information to the computer
routine test calibrated quality control tests parameters of tested query of patient’s result routine maintenance
repeated test test of reagent amount item; parameters of QC queries calibrate the instrument
test test of absorbance special item statistical result detect the instrument
Routine testing menu is mainly to complete the daily work and it consists of several
sub- menus: the conventional test is carried out in entry of the day-to-day specimens
information of the patients, in the choice of test items, and in testing samples, and it is
the most commonly used software module of the instrument; emergency test is carried
out in emergency specimens entry and in emergency insertion test of emergency
specimens, priority having been given to some choreographed specimens; real-time
results show that real-time display of test results for each specimen; response curve
display module can display the reaction curve of a sample in any testitem and at any
time so as to easily understand the details of the test; instrument status submenu can
show real-time display of each work status of the instrument; Printing the result means
to verify the detected samples results , print test results reports and give it back to
patients. Thus a complete cycle is completed
The parameter setting menu is to complete the module produced by various parameters
and it consists of two sub-menus: the sub-menu of parameters of tested item mainly sets
the sample volume, reagent volume, the analysis wavelength, test methods and other
parameters of each tested items:; parameter settings of special items is to set a formula,
based on which test results has been calculated and form a new item results;
combinated items settings is to facilitate the item input Based on clinical requirements,
routine testing items can be combined into a new item , and we can enter the new
combinated item name while inputting these items.
Special inspection is the exclusive test in addition to routine test, and it can test the
standard samples, draw standard curves, test quality control specimen, describe quality
control diagram and other work; it can also test the instrument performance, such as
linear measurement, absorbance measurements, the repeatability measurements ,which
has facilitated the identification of the measurement work and special needs of the users;
The layout of each test reagents used in the test means the position of reagents on
turntable is set so that the instrument can accurately find the required reagents.
Instruments can measure the amount of each reagent before the item is tested and tell
the operator whether he needs to add reagents. Maintenance and repair of sub-menu
facilitates maintenance engineers to test various components of the instrument so as to
identify failures.
Data processing module can check the patient's test results, QC results etc,, and carry
out mathematical statistics, calculate daily average, and draw trend mapping to provide
valuable statistical information to the Clinical Laboratory; records of equipment failure
can record the fault information of the instrument while working and provide
maintenance information to maintenance personnel.
Instrument maintenance and maintenance module are to provide staff with a means of
testing instrument. Routine maintenance can provide routine maintenance of the
instrument and calibrating the instrument can test the instrument's moving parts and
optical signal.
3.4 Structure
Biochemistry analyzer consists of the main unit of analyzer, computer (including the
main unit and monitor), the Windows 2000 or more superior operating system,
exclusive software and printers, etc.
It is suggested to buy the computer and printer which meet the national standard.
!! ! Note: The specimens may carry infectious bacteria. Protection is needed when
handling specimens. Operator should wear a mask and put on work clothes to
avoid being infected, and in some cases protective glasses is a necessity.
When handling waste, operator should wear a mask and put on work clothes to
prevent infection, and in some cases protective glasses is a necessity. Please observe
the local emission standards, and consult the reagent manufacturer or distributor.
30cm
switch computer printer
83cm 60cm
1 Main unit / 1
2 Manual / 1
Warranty card
3 / 1
Certificate
4 / 1
Accessories box
5
Sample cup
5.1 / 1 Quantity:500
Cuvette
5.2 / 2
Positioning block of
5.3 reagent bottle / 45
Bucket
5.4 / 2
RS232 communication
5.5 cable / 1
Power cord
5.6 / 1
Operating software CD
5.7 / 1
(CD-ROM)
External pipes
5.8 / 1 4mm×6mm
External pipes
5.9 / 1 6.5mm×10mm
Simple tools
5.10 / 1
Spare halogen light
5.11 / 1
Fuse
5.12 / 2
Water falling head
5.13 / 2
As the instrument is relatively heavy, two or more staff is needed to lift the instrument.
If the instrument is too wide and can not enter the door, the instrument can be tilted
slightly, but can not be inverted; instrument is required to be placed in the level firmly,
and 30cmwith of space at the back of the instrument should be left for maintenance;
Remove the moisture-proof plastic sheeting of the instrument;
Remove the apparatus used for t fixed attachment of the instrument during the transport,
see Figure as follow
Panel and side panel
made of wooden box
2, Instrument installation
Install cleaning bucket;
Install the waste buckets;
Intake Intake
OFF
Refrigeration Output
ON
RS-232 RS-232
OFF
Main power
! ! !Note: The parts of the outlet pipe inserting into the waste container should not
be too long, 10 cm in length is appropriate, so as to prevent the excessive
wastewater immersing outlet pipe which will result in poor drainage. !
Turn on the computer power switch and the printer power switch, and the main
computer Windows interface appears; double-click the shortcut
"BIOBASE-CRYSTAL" and enter the “log in” interface; enter a user name and
password (the initial user name and password are number "1000"); if they are
confirmed, yon can go to the "BIOBASE-CRYSTAL" main test interface
Please be ready to detect the specimens after the machine has warmed up for 30
minutes.
Repeat the above steps. When the entry is completed, please exit the routine test login
screen. Click "Testing" to enter the following test interface:
Check reagents and samples in accordance with the place order of the reagents and
samples in the interface. Enter the number of the first cuvette and click the “Start”
button. Hint bar prompts information. After confirmation, please click "OK" button to
start the test;
Instrument starts to test the sample, and the information of the sample is revealed in the
instrument column.
4. 2. 5 Processing and Printing of the Result
In the sample test menu, click “Report”, the following interface appears:
Select the number of samples in Sample ID list box, and the results are displayed in the
results column. If you need patient information, please enter patient information in
patient information column. If you need to print other test results, some other relevant
information can be input to the adding column. The results can also be edited in the
interface;
After the test result is confirmed, please select to print the current report or the whole
report, or choose to print a range of the report;
After finishing the tests, please turn off the system, exit the test interface, and turn off
the computer.
NO
Calibration
YES
Test Item calibration
是
QC NO
YES
QC 是
Detect specimen
Result
Click "Add" button and Input the information: input the number of samples (sample
number can only have a unique number within a day without repetitive use);
The patient's name, gender, age can be entered in the patient information bar, patient
information can be input while the test report is being printed;
Combination login: If you have a number of specimens, , and each specimen will be
tested in terms of the same item, in order to facilitate rapid input , you can place them
continuously on the specimen ring and use the combination login: firstly enter the
starting specimen number, and then enter the start cup No. Finally, enter the quantity of
specimen. Then the number of sample specimens and sample number starting Cup are
continuously choreographed automatically by a computer;
Item selection: You can two input methods----individual item input and item portfolio
input. Portfolio items are defined in later chapters; If you select the individual item, you
can click in the name of the test items (press Ctrl to make jumping options, or press
Shift to make consecutive multiple-choice) the item is displayed as selected, click "=>"
(you can double-click to add) to enter them to the column of test items. You also can
select other items, and enter "OK" button to confirm;
If you select the input of the item portfolio, you can select the name of the item
portfolio to be tested in the item portfolio bar. Click on the item, the item is displayed
as selected, click the "=>" and add it to Selected Item column; if you want to delete a
test item , please click the test item in the Selected Item column, click the" <= " to
delete the test item;
Repeat the above operation to input the specimens; you can edit the information entered.
In the Login Column, you can select a sample number for inquiries and it will reveal
the information of patient and test items. You can use "Edit" or "Delete" button to edit
and delete the displayed content. If the information is correct, please enter “OK” button
for conformation;
Note: The patient information can be entered here and can also be input while
printing a test report;
When the input is completed, please exit the Sample Input interface.
5.2.2 Testing
Click "Testing" to enter into the following interface:
Check reagent and sample in accordance with the placement order of the reagents and
samples in the interface. Enter the number of the first and last cuvette, click the "Start"
button and pop-up the hint box:
This interface is to remind users to check whether the reaction cup is washed clean and
whether the number of selected initial and last reaction cup is correct;
After confirmation, please click "OK (the first one)", the instrument begins test; if you
cancel the above input, please click the "Cancel (the second one)";
If you want to suspend operations in the testing process, please press the "Pause" button
(Note: pause time in the testing process can not be too long) If you want to continue the
test, please press the "Continue", the instruments will continue to test. If you want to
cease the operation you can click the “Stop” button to cancel all tests and retreat to the
main interface;
After confirmation, please click "OK". The status bar will have real-time display of the
sample state in order to observe the work of the instrument;
When the instrument is to show the test results show, it will have real-time display of
Put the specimens on the sample ring (specimens can be placed on any vacancy in the
sample circle. Action: should be quick, do not place while the sample circle is still
rotating);
Click "Sample Order" menu interface is as follows:
Emergency interface and routine test interface is roughly the same, and the difference is
that when entering the emergency sample, one should select the emergency check box”
Stat” (see above).The other input is the same with a common sample. After the input is
completed and the current sample is finished, the instrument will test emergency
specimens.
5.4 Repeat-test Sample Inputting
If the item of some specimens was found too high in concentration or other bad results,
you can re-test it. The test results will overwrite the original data. We can enter the
“Repeat Test”, and the interface is as follows:
Click”Edit”, Select the number of specimen which will be retested, edit it. Select the
items which will be retested and add them to the selected item. Repeat the above
process multiple input samples. Make sure click the” Repeat Test” check box .Click
“OK” to enter the test interface for test.
5.5 Registration and Printing of the Results
Click “Report" to enter the following interface:
The specimen test results of the day can all be found in the results sample box. First,
select the number of samples, and test results are displayed in the results column; If it is
needed to enter patient information, please enter the patient information in patient
information column; If need to print other test results in the report, you can also add
item name in addition column and enter data in the results column; If you want to edit
the test results, move the cursor to results column in which the results are needed to be
replaced. Enter the new results and confirm.
If all the samples have not been tested, and you want to print part of the results report
(release the report while testing the sample), you have to click the button of “Calculate”.
Once the specimen is tested, the calculation results will appear immediately; otherwise
the calculation results will appear after all the tests end.
If there is the "√" mark in “Print Order” box, it means the printing order of the data
conforms to the setting of date printing order described by the user in data maintenance
(printing order maintenance is defined in "Data Management" of "Data Maintenance" ).
After confirming test result, please select “All” “Current” or “Range” of report for
printing.
‘
××××××test report
Item NO.:0764
NAME:______ SEX___ AGE___ No ITEM RESULT 单
1 ALT 13.00 U/L 0.00—45.00 9 TG 1.46 mmol/l 0.48—1.88
2 AST 25.00 U/L 0.00—45.00 10 Crea 70.0 umol/l 36.0—144.0
3 AKP 83.00 U/L 15.0—150.0
4 TBIL 16.80 umol/l 6.0—22.0
5 IBIL 6.00 umol/l 0.00—6.0
6 TP 74.50 g/L 60.0—85.0
7 ALB 48.4 g/L 35.0—55.0
8 A/G 1.9 1.00—2.00
DATE:200*/*/*
When inputting a new test item or replace reagents, one needs to change test parameters.
Parameter setting process is as follows:
Click the “Parameter” menu of "Item parameters" in the main interface and enter the
following
Click the item name in the item column, such as "ALT", the background of the item
turns blue. At this point, item number, English name, item name, Chinese name, the test
item parameters and other information of the tested item are revealed.
Sample parameters: “Sample volume”, “R1 volume” o, “R2 volume”. These data will
limit the volume of solution extracted by sample syringe, reagent syringe; “Sample
volume” range is between 2-50ul, “R1 volume” range is between 25 -300μl, “R2
volume” range is between 10-150ul. If you do not set the second reagent, you can enter
"0" or not input anything in the “R2 volume”.
Test method: the end-point method, the two-point method, dynamic method,
turbidimetry, select according to the test reagents and methods; typical reaction curves
are as follows.
OD OD
2.0 2.0
1.0 1.0
0.0 0.0
0 10 20 30 40 50 0 10 20 30 40 50
Figure 10. Endpoint method response curve diagram Figure 11 dynamic method reaction
curve diagram
Slope: If the absorbance readings are becoming bigger, that is a positive slope.
If the absorbance readings are becoming smaller, that is a negative lope
Maximum slope: it represents the maximum allowable amount of absorbance change. If
it is larger than the set value, indicating that there is error the sample which is being
analyzed. Instrument gives failure code, suggesting that the specimen need to be
retested
Reading point: the main reading segment is beginning time of reading and the period of
reading time;
There are 40 points of the instrument in the whole process of reading, interval of each
point is about 18 seconds, and the entire reading time is 40 × 18 = 720 seconds (12
minutes).
If you choose the third minutes as the starting point, the reading interval is three
minutes. Then the starting value is 10, the end value is 20.
If you need to adopt the blank sample method during the test, you should read blank
sample while reading. You should apply the method of Sec readings, meaning reading
the blank sample data in the auxiliary reading section, read the response data in the
main reading section, and process data according to the need. Auxiliary reading section
is located before main reading section.
Please set the data according to the need.
OD Second reagent
1.5
1.0
Secondary Readings section Primary reading section
0.5
0 5 10 15 20 25 30 35 40 45
Linear correlation: If the machine's test results are different from expectations, you can
be modify the result according to the formula Y = AX + B, change the coefficient A
and B to amend the test results.
Unit: Concentration units can be selected in the drop-down list.
Accuracy: the number of decimal places of the test results can be selected in the
drop-down list.
If you want to modify the item parameters, select the item in the list of items column,
click "Edit" button, and then the cursor can be moved to necessary parts to modify, if
one wants to cancel the change; click "Delete" button. If you want to save your changes,
then click the "Ok" button.
If you want to delete test item parameters, select the item in the list of items column,
click the "Delete" button, and then remove the item parameters.
If you want to add the test item, click the "Add" button to enter the test parameters, and
save the input.
You can edit the items portfolio with the help of "Delete" button and "Edit" button.
6.4 Parameter Settings of Items for Calculation
Please click the "Add" button in the column of Calculated Items. Please enter Number,
Item Name, Full Name, Item Name, and the Item Name must be entered. Enter the
value in Low, High, and Decimal. If during specimen test, all the test items contained in
the Formula. After the test is completed, the value of items for calculation will be
automatically calculated and results will be printed.
When edit the Formula, you can directly enter the item name (item must be in the Item
List), or select an item in the optional item column. There can be only the item name
and calculation symbols in the Formula. No other non-standard symbols are allowed.
Please click "OK" button after setting the item.
!!! After deleting one item, you must also delete this item in Print item order
(“Data Management”, “Data Maintenance”), otherwise will have mistakes.
You can edit the item for calculation with the help of button "Delete" and "Edit"
Note: Input item in the “Report” must be maintained in this column. If there is no
item in this column, the entered data can not be printed.
7 Specific Tests
7.1 Calibration and QC
Some test items need calibration.
Click the “special test” and choose “Calibration And QC” to display the following
Interface:
Click the “Item List” to list the item parameter in the screen
Item name and test parameter shows the same. The calibration mode can have a variety
of options.
Calibration mode: Calibration Type for CRYSTAL
1、 One Point
2、 Two Point
3、 Multi-Point 1 (passing the zero point)
4、 Multi-Point 2(without passing the zero point)
Formula: the formula is as follows:
1、 Y=AX
2、 Y=AX+B
3、 Y=A0X^n+A1X^(n-1)+……..+AnX
4、 Y=A0X^n+A1X^(n-1)+……..+AnX+B
Concentration Concentration
40.0 40.0
20.0 20.0
0.00 0.00
0.4 0.8 1.2 1.6 1.8 OD 0.4 0.8 1.2 1.6 1.8 OD
Concentration Concentration
40.0 40.0
20.0 20.0
0.00 0.00
0.4 0.8 1.2 1.6 1.8 OD 0.4 0.8 1.2 1.6 1.8 OD
Reagent blank absorbance (sample): reagents (samples) and blank absorbance value,
this data is automatically measured in instrument calibration testing process (it needs
not to be input).
Blank Category: reagent blank and sample blank value. If you choose reagent blank,
you can get the final absorbance test value by deducting in the test the he reagent blank
of total absorbance. If you choose sample blank, firstly test sample blank in each
sample test, and then test the reactant absorbance, the absorbance of the reactants minus
sample blank absorbance will get the final absorbance data.
The graphic box showed the calibration curve after calibration, the horizontal axis is the
absorbance OD value and the vertical axis is the concentration value.
Change the scale of the abscissa scale; you can change the horizontal aspect ratio.
Change the scale of the ordinate scale can change the horizontal aspect ratio.
Current Data: select accredited calibration results in date list of all calibration and
display it in the text box. If you want to delete a calibration data in the list, you can
select the data; click “Delete” button.
If you want to modify the calibration parameters, please press "Change" button,.
If the calibration results are right, you can click the "OK" button to save the calibration
results.
Click the "Exit" button, and returned to “Calibration And QC” interface.
Select the calibration item in the Item Lists column, method of selection is the same as
that described earlier.
If the data is entered correctly, click the "Start" button, the calibration test begins.
In the test process, you can click the "Stop" button, to stop the test.
After the test is completed, you can click the “Calibration Parameter” to see results of
the calibration.
7.2 QC Test
In order to monitor the accuracy of test results, it is sometimes necessary to test quality
control of some specimens. There are a variety of ways of doing QC test, the first
method is in ordinary sample test, you can observe whether the test results is within the
quality control range, the second method is to enter the “quality control test” interface
Please firstly set the quality control parameters before doing quality control, click “QC
Parameter" button to display the interface above;
Click the “Add” button:
Enter the QC Lot:
Select test items in Item column, add it to the "QC Item" column, steps are the same as
describe above;
In the "QC Item" column, click an item and enter the Mean Standard Deviation and QC
Lot, repeat the above steps, you can enter quality control data for each item;
Press the "Delete" and "Edit" button to edit the input values;
In "QC" column, press "Add" button to enter QC Order No, QC Cup No. (Indicating
the quality control specimens were placed in the placing number of sample disk) and
QC Lot. In “Items Lists" column, select an item and add it to “Items" column; Repeat
the above steps and enter multiple quality control specimens. Use "Edit" and "Delete"
button to modify the input data; entry information can be shown in the QC Order List;
If the data is entered correctly, click the "Start" button, the quality control test begins;
In the test process, you can click the "Stop" button to stop the test;
The test results are shown in real time in the results column, and the results are stored
in the QC database.
7.4 Regents
The instrument has a reagent disk in which 45 reagent boxes can be placed; order of
reagents shall be set in the “Reagent Status” interface. The interface can estimate how
many tests can the remaining reagents in the reagent boxes can do so as to avoid the
situation that reagent is in shortage in the test process.
On the main interface, click the "Reagent Status" menu under the "Specific Test" menu
to appear the following interface:
No.: reagent box is placed on corresponding location of number in the reagent disk
(range: No.1-No.45);
Amounts: maximum volume of reagent box (ml); The height of the reagent box: net
height of the reagent box (mm);
If you want to test the remaining reagent volume, click the "Start" button, the
instrument automatically detects the reagent stock and calculates the number of items
that can be tested.
!! ! NOTE: The reagents location can not be misplaced, before the testing, please
make sure there is sufficient reagent for testing items.
In the main interface, click the "Reading Absorbance” under “Specific Test" menu to
enter the above interface.
Select the desired wavelength in Wavelength drop-down list, after sample in cuvette is
ready, click the button "Start", and the instrument begins reading; After the reading is
completed, the table shows the number of the cuvette and corresponding absorbance
values, select the number of the cuvette to find the corresponding measured absorbance
value.
!!!Note: In the process of the instrument performance tests, make sure 120 cuvette
was added deionized water as water blank.
8. Data Management
Queries can be carried out by a variety of ways, such as Date, Name, Department, and
Patient ID. There are three kinds of query: Date, Name and Sample No. If according to
the Date, select the correct date, confirm the registration number of the day to display in
the Sample No. column, select the appropriate number of samples, the results are
displayed in the results column, click the button "Print", then print a report. Also can
use a combination of query, enter a few queries, such as at the same time to enter the
date and name.
8.1.2 QC Query
In the data processing interface, click the “Data Management - Query” menu, click “QC
Query” menu, appear the following interface.
Select the test item in the drop-down list, enter the From and To dates.
Data and QC Lot and Value within a period of time are displayed in the QC value
column. If several quality controls are made in the same day, we need to choose an
appropriate data as a quality control chart parameters (the default value is the first data),
the statistics column automatically calculates their average value, SD, CV. Quality
control chart shows quality control graphics, click “Print” to print.
8.2 Reaction Curve Monitoring
During test, each test item, the instrument read a set of data on eight wavelengths every
circle for each test item. There is a total of 40 point value for each tested item; we can
observe the reaction of the whole process to understand the correctness of the results.
Click the “Data Management- Reaction Curve” to enter the following interface:
In Date column, listed all stored dates data. Click the "Delete” button to delete all the
data of selected date
!! ! Note: Be careful to use the “Delete” button. After delete, data can not be
recovered.
Select the types of specimens that you want to query in Sample, Calibration and QC.
If you choose Sample, sample number displayed in the Sample No. column within that
day, select the correct number of samples, the samples tested in the project displays on
the Item list;
If you want to observe the test data, click the "Data" button; the interface displays all
the absorbance value of main wavelength auxiliary wavelength and the difference
between the two, click the "Data" button again to exit,. Repeat the above process, you
can query, calibration and quality control; click upper right corner "x" to exit the
response curve monitoring interface.
In the Report Format column, you can enter the name of the hospital which produced
the report, place print order. In the Data column, you can input Units, Departments,
Diagnose, Doctor and Sample Type.
!! ! Note:
1. When edit the Item Order, it should be arranged all the items (contain
computational item and manual item); otherwise the item data will not be printed.
2. the item has been deleted in” Parameter- Item parameters”, it should be also
deleted here, otherwise it will go wrong!
9. Maintenance
9.3 Adjustment
!! ! Note: The instrument adjustment procedures must be carried out by
professionals. To avoid damage is done to the machine non-professionals, we set
the password: 666666.
It’s mainly to adjust the position of the “Sample Arm” needle, “Reagents Arm” needle,
Pipetting needle is better to be adjusted to the bottom position of the sample cup.
!!! Note: Sample cup and test tube could select one only during adjustment.
9.3.3 Horizontal Adjustment of Reagent Arm
This adjustment corresponds to “Reagent Arm >> Horizontal”. Place1=washing
position, Place2=No.1 reagent position, Place3=No.1 reaction position. We can adjust
the horizontal position of the reagent arm through “+1 Step” and “-1 Step” button, save
the changes through "Ok" button, "Cancel" button to abandon changes.
9.3.4 Vertical Adjustment of Reagent Arm
! ! ! Note: When replace lamp, power must be cut off to avoid electrical shock; and
!! ! Note: Please use the fuse with a specification of φ5 × 20. Please replace the fuse
according to the specification marked on the instrument.
!! ! Note: If the fault can not be ruled out, please contact maintenance engineers of
the OEM Company.
Remarks:
The accuracy of test results of automatic biochemistry analyzer is mainly decided by
two factors:
First, the reliable analytical instruments. Second, the reagents of stable performance.
Both are indispensable. In the basis of ensuring high quality instrument performance, we
suggest users buy trustworthy quality product for the accuracy of your test results.
Firstly, the reagent is approved by the related institutions. Secondly, the products are
verified by the majority of users.
Web: www.biobase.cc