UltraMaxO2 Oxygen Analyzer Overview

The UltraMaxO2 analyzer was designed to verify oxygen concentration, flow, and outlet pressure from oxygen concentrators. It has an LCD display and measures oxygen concentration from 20.9% to 96% with ±1.5% accuracy, flow from 0 to 10 lpm with ±0.2 lpm accuracy, and pressure from 0.5 to 50 psi with ±0.5% accuracy. It operates for over 1,100 hours on two AA batteries and has an impact resistant, ergonomic case for portability and reliability in verifying oxygen equipment performance.

Uploaded by

Mehmet Çakır
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
93 views2 pages

UltraMaxO2 Oxygen Analyzer Overview

The UltraMaxO2 analyzer was designed to verify oxygen concentration, flow, and outlet pressure from oxygen concentrators. It has an LCD display and measures oxygen concentration from 20.9% to 96% with ±1.5% accuracy, flow from 0 to 10 lpm with ±0.2 lpm accuracy, and pressure from 0.5 to 50 psi with ±0.5% accuracy. It operates for over 1,100 hours on two AA batteries and has an impact resistant, ergonomic case for portability and reliability in verifying oxygen equipment performance.

Uploaded by

Mehmet Çakır
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

SENSING. ANALYSIS. DELIVERY.

UltraMaxO2
Verify oxygen concentration, accuracy
of flow, and outlet pressure.

LCD Display Gas Sample


Inlet

Gas Sample
Outlet

Impact Resistant
Ergonomic Case

ULTRA PERFORMANCE! PART NUMBERS

UltraMaxO2. . . . . . . . . . . . . . . . . . R221P11
The UltraMaxO2 analyzer was designed to verify oxygen UltraMaxO2 (Int'l). . . . . . . . . R221P11-001
concentration at a glance, including the flow and outlet pressure UltraMaxO2 (Turkey). . . . . . R221P11-002

of oxygen concentrators. Its advanced design provides high-level


performance and reliability.
4

2 5
3

1. LCD Display 3. On/Off Button 5. Gas Sample Outlet


Large, easy-to-read liquid crystal display One-touch on and off power An outlet for the gas sample; also used as a
trigger for pressure measurement when occluded
2. Mode Button 4. Gas Sample Inlet 6. Battery Door
No in-field calibration required Used to receive the gas sample Long battery life with two (2) AA batteries

Specifications

Oxygen Measurement Range 20.9% to 96%.


(from concentrator)
Oxygen Measurement Accuracy ±1.5% of full-scale at constant temperature and optimal flow
Oxygen Measurement Resolution 0.1% oxygen
Flow Measurement Range 0 to 10 lpm
Flow Measurement Accuracy ±0.2 lpm
Flow Measurement Resolution 0.1 lpm
Pressure Measurement Range 0.5 to 50 psi (3.4 to 344 kPa)
Pressure Measurement Accuracy ±0.5% psi (±0.5% kPa)
Pressure Measurement Resolution  0.1 psi (0.1 up to 199, 1 from 200 to 344 kPa)
Response Time ≤ 17 seconds
Warm-up Time < 1 second
Operating Temperature 15°C to 40°C (59°F to 104°F)
Storage Temperature -15°C to 60°C (5°F to 140°F)
Atmospheric Pressure 800 to 1000 mbars
Humidity 0% to 95% (non-condensing)
Power Requirement Two (2) AA alkaline batteries (2 x 1.5 volts)
Battery Life 1,100 hours (16,500 read cycles)
Low Battery Indication ”Low Battery” icon displayed on LCD
Dimensions 3.39” x 5.10” x 1.00” (85.98mm x 129.54mm x 25.27mm)
Weight 0.4 lbs (181 grams)

ML# 230 Rev.D