Manual Do or

The document provides assembly and operating instructions for the Distinction Water Still D4000 & D4000/EURO. It describes how to assemble the various components of the still including attaching the boiler, heater, thermostat tubes and connecting water and electrical sources. It also provides details on operating the still such as turning it on, allowing it to cool and collecting distilled water. Electrical installation instructions specify using a qualified electrician and recommended safety devices like an isolation switch and overcurrent protection.

Uploaded by

Rosa Maria Pais
Copyright
© Attribution Non-Commercial (BY-NC)
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
0% found this document useful (0 votes)
278 views16 pages

Manual Do or

The document provides assembly and operating instructions for the Distinction Water Still D4000 & D4000/EURO. It describes how to assemble the various components of the still including attaching the boiler, heater, thermostat tubes and connecting water and electrical sources. It also provides details on operating the still such as turning it on, allowing it to cool and collecting distilled water. Electrical installation instructions specify using a qualified electrician and recommended safety devices like an isolation switch and overcurrent protection.

Uploaded by

Rosa Maria Pais
Copyright
© Attribution Non-Commercial (BY-NC)
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
Download as pdf or txt
You are on page 1/ 16

Distinction

D4000 & D4000/EURO Water Still

Distinction

Assembly and operating instructions Instructions de montage et de fonctionnement

Distinction Water Still D4000 & D4000/Euro

Condenser Thermostat reset buttons Condenser thermostat

Condenser outlet Condenser thermostat tube

Boiler thermostat tube

Boiler thermostat On/Off switch

Cooling water inlet Wall mounting screw holes

Clip

Constant level control To drain

Distinction
Reservoir level control

Stopcock

Heater

Gaskets and flange

Distillate outlet tube

Boiler

Straps

Stand

Figure 1 General assembly

Before use
If the equipment is not used in the manner described in this manual the protection provided by the equipment may be impaired. N.B. The Distinction Water Still is classified as Permanently Connected Equipment and should be connected to the electricity supply by a qualified electrician in the manner described in the electrical installation section of this manual. Limitations of use The Distinction Water Still is designed to operate under the following conditions: For indoor use only Ambient temperature range +5C to +40C Altitude up to 2000m Relative humidity not exceeding 80% Mains supply voltage fluctuations not greater than 10% of nominal Overvoltage category II IEC60364-4-443 Pollution degree 2 IEC664 This product meets the applicable E.C. harmonised standards for radio frequency interference and may be expected not to interfere with, or be affected by other equipment with similar qualifications. We cannot be sure that other equipment used in its vicinity will meet these standards and we cannot guarantee that interference will not occur in practice. Where there is a possibility that injury, damage or loss might occur if equipment malfunctions due to radio frequency interference, or for general advice before use, please contact the Technical Service Department of Bibby Scientific Ltd.
The Distinction Water Still complies with the following European Directives: 89/336/EEC - E.M.C. DIRECTIVE Amended by 92/31/EEC and 93/68/EEC 73/23/EEC - L.V.D. Amended by 93/68/EEC

Location
The Distinction Water Still can be wall or bench mounted. Select a convenient location which has access to the following services:Electricity Supply Before connection please ensure that the line supply is suitable. The Distinction Water Still is available in two models, model no. D4000 for supplies rated at 3kW, 220-240V 50/60 Hz single phase and model no. D4000/EURO for supplies rated at 3kW, 200-230V, 50/60Hz, single phase. Water Supply A cold water supply capable of providing a minimum flow rate of 60 l/hr. Drain A waste water drain located below the level of the Still so that the drain pipe can fall away straight without kinks or bends, to allow an unimpeded flow. Reservoir A distillate collection reservoir should be located beneath the Still.

2.

Remove the end cover from the electrical compartment by undoing the retaining screws in the left hand panel. Take the metal stand and place in the desired location. Note that 2 screw holes are provided for wall mounting. If wall mounted the unit should be secured using 38mm (1.5), No. 10 screws. Take the boiler, the heater and the gasket kit and assemble as shown in figure 2. i) Place the bolts through the metal flange in the manner shown and then place the metal flange over the tapered glass tube of the boiler. Take the plastic insert and bend it round the glass tube and into the metal flange. Pull the flange and insert towards the open end of the glass tube until the insert presses on the glass. Place the aluminium plate, followed by the silicone ring over the heating element and locate approximately 4045mm from the end with the electric cable. Insert the heating element through the tapered glass tube and into the boiler ensure that there is clearance at the inside end and no danger of the end of the element pressing against the end of the boiler. Now secure the element using the 3 bolts provided. Do not overtighten.

3.

4.

ii)

iii)

iv)

Assembly
1. Qty. 1 1 1 1 1 1 1 1 2 2 1 1 Unpack the Still and identify the following components:Component Support Stand Boiler Condenser Heating element Hose kit Gasket kit Boiler thermostat tube Condenser thermostat tube Plastic connecting caps Silicone rubber rings P.T.F.E washer Reservoir level control Catalogue No. D4000/S D4000/B D4000/M2 D240H for D4000 D220H for D4000/EURO W4000/HK D4000/GK WTT48 A515414 QC18/11 QR18/7 and QR18/10 QW18/7 WLS v)

Screwcap 40-45mm Boiler thermostat tube Boiler

Heater

Silicone Aluminium ring Plastic plate insert

Tapered glass tube Metal flange Bolts

Figure 2 Heater and thermostat tube assembly

5.

Take the two metal straps and attach the springs to the rear of the cradles on the stand. Place the assembled boiler and element onto the cradle of the stand. Slide the assembly along until the protruding end of the element is located inside the electrical compartment and at the same time ensure that the boiler thermostat probe is located in the thermostat tube. Attach the free ends of the straps to the front of the cradle to secure the boiler in place. N.B. A pip on the bottom of the boiler locates in a hole in the right-hand cradle.

8. From the hose kit take the length of 9mm bore hose fitted with the plastic screwthread connectors at either end. Referring to figure 1, screw one end of the hose to the upper outlet of the condenser and the other end to the glass thread on the constant level control. 09. From the hose kit take the 1000mm length of 16mm bore hose and carefully push on to the outlet of the constant level control. For ease of connection, first warm the end of the hose with hot water. Secure with the tie strap provided. 10. Lead the free end of the hose to drain, ensuring it falls away from the Still with no kinks or bends to impede the water flow. 11. Ensure that the stopcock on the constant level control is closed.

6.

Connect the heater cable as shown in figure 3a. - Ensure screw terminations are tight. - Push wires fully into block.

Heater electrical connections Brown Blue

12. Assemble the condenser thermostat tube A515414 connecting cap QC18/11, silicone rubber ring QR18/7 as shown in figure 4. If the cap is left loose the assembly may now be carefully slid over the air vent on the distillate outlet - see figure 4. Front panel Tighten up the cap once the tube is in place. Identify the condenser thermostat and place the free end of its probe into the thermostat tube. Carefully bend the capillary to ensure that it does not obstruct access to the thermostat reset buttons. - see figure 1.

Heater 3kW 220/240V

Brown

Blue

Assembly of condenser thermostat tube Condenser thermostat

Heater cable Figure 3a

Condenser thermostat tube Mains electrical connections Screwcap Yellow/green Mains cable Figure 4 Blue N L Brown 13. Connect the lower inlet of the condenser to the cold water feed. A plastic screwthread connector is provided for easy attachment. Select good quality tubing and secure all connections with hose clips. Figure 3b Air vent Distillate outlet

7.

Fit the condenser WC48/M2 to the vapour tube of the boiler, ensuring that the distillate outlet faces forward.

14. Connect the distillate outlet of the condenser to a suitable collection reservoir, using a suitable length of 9mm tubing. A glass reservoir system is available from Bibby Scientific Ltd. (Catalogue No. WR20). Please contact the sales office for details. 15. Replace the end panel on to the unit. 16. Take the reservoir control WLS and push the free end of the rubber tubing onto the nozzle on the left hand end of the unit - see figure 1. Use the clip provided to mount the glass pressure bell in the neck of the distillate reservoir. Adjust the height of the pressure bell so that the open end is approx. 140mm below the maximum distillate level.

Operation
1. Turn on the cold water supply and adjust the flow to approx. 60 l/hr. Observe that the water flows via the condenser and into the boiler. Wait until the boiler has attained its correct operating level and make sure that the excess water is flowing freely to drain. Switch on the heating element. The on/off switch is located on the front panel - see figure 1. After a few minutes the water will start to boil and distillate will emerge from the condenser. With new glassware, or after cleaning, it is advisable to allow this to run to drain for approximately 30 minutes before beginning collection. To turn off the still, first turn off the heating element but allow cooling water to continue for a further 10 minutes to allow the Still to cool.

2. 3.

4.

Electrical installation
THIS EQUIPMENT MUST BE EARTHED! The electrical installation should be carried out by a qualified electrician. The following method is recommended. The equipment is supplied with 1.5m of PVC/PVC flexible, triple core, circular cable with the following specification, 1.5mm 2 to BS 6500 or equivalent and <HAR> or BASEC approved. The cable gland is designed to accept cable whose diameter is >7.0mm and <10.5mm. It is recommended that, at the connection to the mains electrical supply, there is fitted a double pole isolation switch with a continuous current carrying capacity of 15 amps at 250V, and that overcurrent protection is provided by either a double pole approved fuse rated at 15A, 250V, or a double pole approved circuit breaker of similar rating. These devices should be sited near to the equipment and be clearly marked Disconnect device for Distinction Water Still. NOTE: Do not connect to electricity supply with a 13 amp plug. Cable connections are shown in figures 3a and 3b. Connect to the line supply noting that the wires in the instrument lead are coloured in accordance with the following code: Blue - Neutral Brown - Live Green & Yellow - Earth Mains Cable Replacement If the mains cable requires replacement, only cable with the specification given above should be used. For live and neutral leads use bootlace crimps, size 1.5mm, with a performance conforming to BS 4G178 part 1, 1984 and part 2, 1986. For the earth lead use 1.5mm, M4 eyelet crimps with a performance conforming to BS 4G178 part 1, 1984 and part 2, 1986. A suitable crimp tool should be used.

WARNING! Do not use this equipment to distil any liquid other than water.

Safety cut-outs
The Distinction Water Still is protected by the following safety cutouts: 1. Reservoir level control This should be positioned in the collecting reservoir (see Assembly) and turns off the heating element when the reservoir is full. 2. Condenser thermostat Mounted in the condensers distillate outlet. Should the water supply be interrupted steam will begin to emerge from the condenser causing the thermostat to operate and turn off the electricity supply to the heating element. 3. Boiler thermostat Mounted in the boiler above the heating element. Should the boiler water level fall and expose the element the thermostat will operate and turn off the electricity supply to the element. After operation of either of the thermostats, normal operation may be resumed by resetting the thermostat by means of their respective reset buttons mounted on the right hand end panel of the electrical compartment. Remove the black plastic cover and then press the red button - a slight click will be heard if the thermostat had operated. Before resetting either thermostat the still should be allowed to cool completely and the cause of the cutout operation identified and rectified. If the thermostats continue to operate, consult a qualified electrician or the Technical Service Department of Bibby Scientific Ltd.

Care and maintenance


N.B. Before commencing any maintenance, cleaning or fault finding the equipment should be isolated from the mains electricity supply. These operations should only be carried out by suitably qualified personnel. Only spare parts supplied or approved by Bibby Scientific Ltd or its agent should be used. Fitting of non-approved parts may affect the performance or safety of the equipment.

9.

The Distinction Water Still may now be restarted by referring to the instructions given under Operation in this manual. Note: The stand and outer surfaces of the glassware should be cleaned using a damp cloth and a mild detergent solution.

Heater replacement*
1. 2. Isolate the Still from the electricity supply. Clean the heater and boiler if they are heavily scaled by following the cleaning instructions given in this manual. It will be found easier to replace the heater by working with the boiler on the bench. Therefore: Drain the boiler via the stopcock. Remove the end panel. Disconnect the heater cable. Disconnect all water hoses. Remove the condenser. Remove the boiler, complete with heater, from the stand. Unscrew the 3 bolts securing the heater and withdraw the heater from the boiler. Remove the aluminium plate and silicone rubber ring from the heater. 4. Now fit a new heater as described in assembly instruction number 4. N.B. Examine the silicone ring and replace if not in good condition. 5. Re-assemble the Still by following the assembly instructions.

Cleaning
Over a period of operation scale will build up inside the boiler. To obtain optimum performance this scale should be removed on a regular basis. Please note that scale can build up both in the neck and main body of the boiling chamber. The time span between cleaning depends greatly on the hardness of the water and the amount of use. Frequently used stills in hard water areas may need descaling once a week whereas in a soft water area several weeks may elapse before descaling is necessary. N.B. Heavy scaling will reduce distilled water quality and can shorten the life of the heating element. It is possible to descale the Distinction Water Still without dismantling the glassware by following these instructions in conjunction with Control of Substances Hazardous to Health regulations (COSHH) 1988: 1. 2. 3. 4. Switch off the electricity supply to the still and allow it to cool completely. Turn off the cooling water supply. Open the stopcock on the constant level control and allow the boiler to drain completely. Close stopcock. Turn on the cooling water supply and allow the boiler to fill to approximately half way to its normal operating level. Turn off the water supply. Into the open funnel of the constant level control carefully add about 100ml of concentrated hydrochloric acid. WARNING! ALWAYS HANDLE STRONG ACIDS WITH GREAT CARE. PROTECTIVE CLOTHING, GLOVES AND FACE-MASKS SHOULD BE WORN DURING THE DESCALING OPERATION. REMOVE ANY ACID SPILLS IMMEDIATELY. Turn on the water supply and fill the boiler to the normal operating level. The water will flush the acid into the boiler. The water supply should be turned off when the level in the boiler is slightly below the overflow. 6. 7. Leave the acid in the boiler to dissolve the scale. This may take some time depending on the severity of the build-up. Open the stopcock and allow the boiler to drain. Note: If the acid in the boiler has not been completely neutralised the liquid flowing to drain may be highly acidic. All necessary safety precautions should be observed around the drain and any effluent control procedures followed. 8. Close the stopcock, turn on the water and allow the boiler to fill with cold water. Turn off the water, re-open the stopcock and allow the boiler to drain. Repeat this procedure three times.

3.

*Note: Heater is not covered by the 3 year warranty. It is only guaranteed for up to 1000 hours in use.

5.

List of spare parts


The following components are available from most laboratory suppliers. In case of difficulty please contact Bibby Scientific Ltd. Component Support stand Boiler Condenser Heating element 240V Heating element 220V Hose kit Gasket kit Boiler thermostat tube Condenser thermostat tube Plastic connecting caps Silicone rubber rings Catalogue No. D4000/S D4000/B WC48/M2 D240H D220H W4000/HK D4000/GK WTT48 A515414 QC18/11 QR18/7 and QR18/10 Heater sealing ring Component Reservoir level control Boiler thermostat Condenser thermostat Reservoir level control pressure switch On/Off switch Boiler O rings Catalogue No. WLS WT4 W14TS DPS D4000/SW Top Bottom QR38/24 WB48/01 QR38/24

Fault finding
In the event of operating difficulties with your Distinction Water Still, please consult the following notes. If these fail to identify and remedy the fault you are advised to contact your supplier or the Technical Service Department of Bibby Scientific Ltd. Symptom Heater not working Cause a) Burnt out heater b) Failed electricity supply c) Thermostat operated Remedy Replace heater Restore supply Identify and rectify the cause and reset thermostat Adjust the supply to approximately 60 l/hr Close stopcock Adjust the supply to approximately 60 l/hr Ensure drain pipe falls freely from Still with no kinks Adjust the supply to approx 60 l/hr Ensure tubing falls freely from Still with no kinks Remove obstruction Adjust the supply to approximately 60 l/hr Clean boiler

Water level in boiler is too low - heater exposed

a) Supply of cooling water is insufficient b) Boiler drain stopcock left open a) Flow of cooling water is too high b) Drain is restricted

Water level in boiler is too high - e.g. water surging into condenser

Distillate temperature is too high - e.g. >50C Water in boiler is pumped out of the boiler to drain

Supply of cooling water is insufficient a) Tubing from condenser distillate outlet to the reservoir is constricted b) Mounting for condenser thermostat is blocked c) Supply of cooling water is insufficient Boiler heavily scaled Mains electricity supply below nominal voltage

Distillate quality is poor* Output less than 4 l/hr

*Distillate quality when measured by pH or conductivity is greatly affected by temperature and the presence of absorbed carbon dioxide.

Distillateur deau D4000/D4000-Euro

Boutons de rarmement des thermostats

Rfrigrant Thermostat du rfrigrant

Sortie du rfrigrant

Tube du thermostat bouilleur Thermostat du bouilleur Interrupteur Marche/Arrt

Tube du thermostat rfrigrant Entre deau de refroidissement Orifice pour fixation murale

Patte

Controleur de niveau Evacuation

Distinction
Contrleur de niveau de recette Elment chauffant Figure 1 Vue densemble Joint dtanchit Tube de sortie du distillat Bouilleur Collier mtallique de fixation

Robinet

Support

Avant utilisation
Si lappareil nest pas utilis dans les conditions dcrites ciaprs, la scurit dutilisation pourrait sen trouver altre. N.B. Le distillateur deau D4000 est class dans la catgorie des appareils branchs en permanence. Il doit tre reli lalimentation letrique par un lectricien qualifi selons les instructions de ce manuel. Il ne doit pas tre reli un prise de courant classique, pur rester en conformit avec le norm IEC 1010. Le distillateur deau D4000 est conu pour fonctionner dans les conditions suivants: A lintrieur dune pice uniquement Temprature ambiante comprise entre +5C et +40C Altitude infrieur 2000m Taux dhumidit relative infrieur 80% Variation de voltage de lalimentation lectrique infrieur 10% de la valeur nominale Catgorie de survoltage II IEC 60364-4-443 Degr de pollution II IEC 664
Le distillateur deau est en conformit avec les normes Europennes: 89/336/EEC - E.M.C. DIRECTIVE Amendes par 92/31/EEC & 93/68/EEC 73/23/EEC - L.V.D. Amende par 93/68/EEC

Limites dutilisation Cet appareil est conforme aux directives Europennes en vigeur pour les interfrences de frquences radio. Il ne doit donc pas interfrer avec dautres appareils rpondant aux mmes directives. Ne pouvant tre sr que dautres appareils en fonctionment autour du distillateur soint conformes ces normes, nous ne pouvons garantir quacune interfrence ne se produira en ralit. En cas de mauvais fonctionment d aux interfrences de frquence radio, prire de contacter le service technique de Bibby Scientific France ou votre revender.

Emplacement
Le distillateur D4000 peut tre install sur une paillasse ou fix au mur. Choisir un emplacement proximit des dispositifs suivants: Alimentation lectrique: Avant de connecter lappareil, vrifier que lalimentation lectrique est correcte. Deux modles de distillateurs DISTINCTION sont disponibles. Modle D4000 pour alimentation lectrique 3KW 220240 V, 50/60 Hz monophas et modle D4000/EURO pour alimentation lectrique 3KW, 210-230 V, 50/60 Hz monophas. Alimentation en eau: Source deau froid capable de fournir un dbit minimum de 60 l/hr. Evacuation: Dispositif dvacuation ou vier situ en dessous de lappareil de faon ce que le tuyau dvacuation deau ne prsente aucun coude. Rservoir de recette: Un rservoir de recette du distillat pourra tre install sous lappareil.

2.

Retirer le capot latral du compartment lectrique enlevant les vis de fixation. Placer le support mtallique lemplacement choisi. Deux orifices de fixation sont prvus sur le support pour un montage mural. Dans ce cas utiliser des vis de 38mm. Prendre le bouilleur, llment chauffant et le jeu de colliers, et assembler selon la figure 2. i) Insrer les boulons au travers du collier mtallique et le placer autour du tube vas du bouilleur. Enrouler le collier en fibre et linsrer entre le tube vas et le collier mtallique. Le pliage du collier en fibre est plus facile aprs trempage dans leau. Tirer le collier mtallique et le collier en fibre vers lextrmit du bouilleur de faon ce quils exercent une pression sur le tube vas. Enfiler la plaque aluminium et lanneau silicone sur llment chauffant jusqu 40-45 mm de lextrmit munie du cable lectrique. Insrer llment chauffant dans le bouilleur au travers du tube vas. Verifier que lextrmit de llment chauffant nexerce aucune pression sur la paroi latrale interne du bouilleur. Visser sans trop serrer les boulons sur la plaque daluminium.

3.

4.

ii)

iii)

Montage
1. Dballe le distillateur et identifier les composants suivants: iv) Qte Designation 1 1 1 1 Support Bouilleur Rfrigrant Elment chauffant pour D4000/EURO (ou Elment chauffant pour D4000) 1 1 1 1 2 2 1 1 Jeu de tuyaux Jeu de colliers por lment chauffant Tube de thermostat du bouilleur Tube de thermostat du rfrigrant Capuchons plastiques Anneaux silicones Rondelle PTFE Contrleur de niveau de recette 753.36 756.42 701.83 701.84 701.85 753.75 754.32 754.34 Reference 754.31 753.61 v)

Capuchon Tube du thermostat bouilleur Bouilleur

40-45mm

Elment chauffant

Anneau silicone Plaque daluminium Collier Collier mtallique

Tube vas Boulon

Figure 2 Montage de llment chauffant

5.

Fixer les 2 colliers mtalliques de maintien du bouilleur sur les encoches arrires des pattes arrondies supportant le bouilleur. Placer le bouilleur sur les pattes arrondies du support. Insrer la sonde du thermostat du bouilleur dans le tube en verre. Faire glisser le bouilleur latralement de faon ce que lextrmit de llment chauffant se trouve lintrieur du compartment lectrique. N.B. Un pointeau en verre situ sous le bouilleur pntre dans une encoche situe sur la patte arrondie droite, quand le bouilleur est en bonne position. Fixer les 2 colliers mtalliques de maintien du bouilleur sur les encoches avant des pattes arrondies.

8.

Dans le jeu tuyaux prendre le tuyau de diamtre 9 mm quip de connecteurs plastiques ses deux extrmits. Selon la figure 1, visser une extrmit sur la sortie du rfrigrant et lautre extrmit sur le contrleur de niveau. Prendre le tuyau de 1000mm de long et 16mm de diamtre. Enfiler avec prcaution son extrmit en sortie du contrleur de niveau. Pour faciliter cette opration, passer pralablement lextrmit du tuyau sous leau chaude. Fixer avec le collier plastique fourni.

9.

10. Placer lextrmit libre du tuyau vers lvacuation, en sassurant quil ne prsente aucun coude pouvant perturber lcoulement de leau. 11. Vrifier que le robinet du cntroleur de niveau est ferm. 12. Assembler le tube du thermostat du rfrigrant, le capuchon plastique et lanneau silicone selon la figure 4.

6.

Brancher les cbles de llment chauffant selon la figure 3a.

Raccordement lectrique de llment chauffant Le capuchon ntant pas compltement viss, glisser dlicatement le tube sur lvent de la sortie du distillat. Visser alors le capuchon. Insrer la sonde du thermostat du rfrigrant dans le tube de verre. Tordre dlicatement le fil du thermostat de faon ce quil nobstrue pas laccs aux boutons de rarmement des thermostats. Voir figure 1. Panneau frontal
Elment chauffant 3kW 220/240V

Marron

Bleu

Montage du tube du thermostat rfrigrant Thermostat du rfrigrant

Marron

Bleu Cble de llment chauffant Tube du thermostat

Figure 3a

Capuchon

Raccordement lalimentation lectrique

Event Sortie du distillat Figure 4

Vert/jaune Cble

13. Fixer le tuyau dalimentation en eau sur lentre du rfrigrant. Un connecteur vis en plastique est fourni cet effet. Bleu N P Marron Choisir un tuyau de bonne qualit et renforcer toutes les connexions par des colliers.

Figure 3b

7.

Enfiler le rfrigrant sur le tube vapeur du bouilleur. Lorifice de sortie du distillat doit faire face lutilisateur.

10

14. Relier par un tuyau de diamtre 9 mm la sortie du distillat au rservoir de recette. Un rservoir en verre PYREX de 20 litres (Rf. 75354) est disponible en accessoire. Contacter Bibby Scientific France ou votre revendeur pour plus dinformation. 15. Remonter le panneau latral. 16. Prendre le contrleur du niveau de recette et enfiler lextrmit du tuyau caoutchouc sur la ttine situe sur le panneau latral. Voir figure 1. Utiliser ventuellement la patte pour fixer la sonde au bord du rservoir de recette. Rgler la hauter de la sonde pour que son extrmit libre se situe environ 140 mm en dessous du niveau maximum de distillat.

Mise en route
1. Ouvrir le robinet dalimentation en eau froide et rgler le dbit 1 l/mn environ. Leau doit circuler travers le rfrigrant. Attendre que le niveau deau dans le bouilleur soit stable. Vrifier que leau scoule normalement vers lvacuation. Allumer llment chauffant via linterrupteur Marche/Arrt situe sur le panneau frontal. Aprs quelques minutes, leau commence bouillir et le distillat scoule du rfrigrant. A la mise en route ou aprs nettoyage, nous vous conseillons dattendre 30 mn avant de rcouprer le distillat. Pour arrter lappareil, couper dabord llment chauffant via linterrupteur. Attendre ensuite 10 mn avant de couper lalimentation en eau, afin de permettre lappareil de refroidir.

2. 3.

4.

Installation lectrique
CET APPAREIL DOIT ETRE RELIE A LA TERRE! Linstallation lectrique doit tre effectue par un lectricien qualifi. Nous vous recommandons de suivre la procdure suivante: Lappareil est livr avec un cble PVC/PVC flexible de 1,5 m, de section 3 x 1,5 mm rpondant aux normes BS 6500 <HAR> et BASEC. Le presse toupe peut accepter des cbles de diamtres compris entre 7 mm et 10,5 mm. Il est recommand de protger lappareil par un fusible de 16A/250V, sur la phase et ne neutre ou un disjoncteur bipolaire 16A/250V. Il est galement conseill de connecter lappareil sur un interrupteur bipolaire 16A/250V. Ces protections doivent tre places prs de lappareil et clairement signalises. Voir figure 3a et 3b pour la connection du cble lectrique. Connecter le cble lectrique de la manire suivante: bleu - neutre marron - phase jaune/vert - terre

ATTENTION! CET APPAREIL NE DOIT PAS ETRE UTILISE POUR DISTILLER DAUTRES LIQUIDES QUE LEAU.

Securits
Le distillateur deau D4000 est protg par les dispositifs de scurit suivants: 1. Contrleur du niveau de recette Il doit tre plac dans le rservoir de recette (voir Montage). Il coupe llment chauffant quand le rservoir est plein. 2. Thermostat du rfrigrant Il est mont sur le tube de sortie du distillat. En cas de rupture de lalimentation en eau, de la vapeur est gnre au niveau du rfrigrant. Le thermostat se met alors en fonctionnement et coupe lalimentation lectrique de llment chauffant. 3. Thermostat du bouilleur Il est situ dans le bouilleur, au-dessus de llment chauffant. En cas de baisse du niveau deau dans le bouilleur, le thermostat se met en fonctionnement et coupe lalimentation lectrique de llment chauffant. Aprs fonctionnement de lun ou lautre des thermostats, la remise en route de lappareil seffectue en appuyant sur les boutons de rarmement situs sur le panneau latral droit du compartiment lectrique. Enlever le capuchon plastique noir et appuyer sur le bouton rouge. Vous pourrez entendre alors un dclic tmoignant que le rarmement est effectu. AVANT DE REARMER UN THERMOSTAT, LAISSER REFROIDIR LAPPAREIL ET IDENTIFIER LA CAUSE DE SA MISE EN FONCTIONNEMENT. Si le thermostat continue de se mettre en route, consulter un lectricien qualifi ou le service technique de Bibby Scientific France.

Remplacement du Cble DAlimentation Electrique Si le cble dalimentation lectrique ncessite un remplacement, nutiliser quun cble ayant les mmes spcifications que celles cites ci-dessus, et fixer le cble de la manire suivante: 1) Pour la phase et le neutre, utiliser des cosses sertir de diamtre 1,5 mm rpondant aux normes BS 4 G 178 point 1, 1984 et point 2, 1986. Pour la terre, utiliser un oeillet sertir de taille M 4 / 1,5 mm rpondant aux normes BS 4 G 178 point 1, 1984 et point 2, 1986.

2)

Une pince sertir est ncessaire.

11

Entretien et maintenance
NB : Avant toute opration dentretien ou de maintenance, isoler lappareil de lalimentation lectrique. Cette opration doit tre effectue par un personnel qualifi. Seules les pices dtaches fournies ou approuves par Bibby Scientific France ou ses revendeurs devront tre utilises. Lutilisation de pices dtaches non approuves peut diminuer les performances ou la scurit de lappareil.

Remplacement de llment chauffant


1. 2. 3. Isoler lappareil de lalimentation lectrique. En cas dentartrage, nettoyer llment chauffant et le bouilleur selon les instructions du chapitre Nettoyage. Il est plus facile de remplacer llment chauffant aprs dpose du bouilleur. Pour cela : Vider le bouilleur. Retirer le panneau latral. Dbrancher les cbles de llment chauffant. Retirer le rfrigrant. Retirer le bouilleur complet du support. Dvisser les 3 boulons et retirer llment chauffant du bouilleur. Retirer la plaque daluminium et le joint silicone du bouilleur.

Nettoyage
Aprs un certain temps de fonctionnement, du tartre peut se dposer lintrieur du bouilleur. Afin dassurer le bon fonctionnement de lappareil, le tartre devra tre limin rgulirement. Lintervalle de temps entre deux nettoyages dpend de la duret de leau dalimentation. Dans les rgions o leau est trs dure, il peut savrer ncessaire de nettoyer le distillateur toutes les semaines. Dans les rgions ou leau est plus douce, plusiers semaines peuvent scouler entre deux oprations de nettoyage. NB : Un entartrage important rduira la qualit de leau distille et la dure de vie de llment chauffant. Il est possible de dtartrer lappareil sans dmonter la verrerie, en respectant les instructions suivantes. 1. 2. 3. 4. Arrter lappareil avec linterrupteur et laisser le refroidir compltement. Arrter lalimentation en eau. Ouvrir le robinet du contrleur de niveau pour vider compltement le bouilleur. Fermer ensuite le robinet. Ouvrir lalimentation en eau pour remplir le bouilleur la moiti de son niveau normal de fonctionnement. Arrter leau. 5. Ajouter 100ml dacide chlorhydrique concentr par lentonnoir du contrleur de niveau du bouilleur. 5. 4.

Monter un nouvel lment chauffant selon le paragraphe 4 du chapitre Montage. NB : Vrifier ltat du joint silicone et remplacer si ncessaire. Remonter lappareil selon les instructions du chapitre Montage.

ATTENTION! IL FAUT TOUJOURS MANIPULER LES ACIDES AVEC PRUDENCE. DES VETEMENTS DE PROTECTION, DES GANTS ET DES MASQUES FACIAUX DEVRONT ETRE UTILISES LORS DES OPERATIONS DE NETTOYAGE. NETTOYER IMMEDIATEMENT TOUTE ECLABOUSSURE DACIDE. Ouvrir lalimentation deau pour remplir le bouilleur jusqu son niveau normal de fonctionnement. Fermer lalimentation deau avant que le mlange acide ne se dverse lgout par le trop-plein. 6. 7. Laisser agir lacide qui doit dissoudre les dpts de tartre. Le temps ncessaire dpend du degr dentartrage. Ouvrir le robinet de vidange afin de vider compltement le bouilleur. NOTE : Si lacide dans le bouilleur na pas t compltement neutralis, leau de vidange peut savrer tre fortement acide. Prendre les prcautions ncessaires au niveau de lvacuation. 8. Fermer le robinet de vidange et remplir le bouilleur deau. Vidanger le bouilleur en ouvrant le robinet. Rpter 2 3 fois cette opration de rinage. Le distillateur deau D4000 peut alors tre rutilis selon les instructions du chapitre Mise en route de ce manuel. Note : Le support et les surfaces externes de verrerie doivent tre nettoyes avec un chiffon humide et un dtergent doux.

9.

12

Liste des pices dtachees


Les pices dtaches suivantes sont disponibles auprs de votre revendeur. Designation Support Bouilleur Rfrigrant Elment chauffant 220V pour D4000/EURO Elment chauffant 240V pour D4000 Jeu de tuyaux Jeu de colliers de fixation Tube du thermostat bouilleur Tube du thermostat rfrigrant Capuchons plastiques Anneaux silicone Reference 754.31 753.61 754.32 754.34 753.36 756.42 701.83 701.84 Designation Contrleur de niveau de recette Thermostat du bouilleur Thermostat du rfrigrant Pressostat du cntoleur de niveau de recette Interrupteur Marche/Arrt Joint haut du tube vapeur Joint bas du tube vapeur Anneau dtachit de llment chauffant Reference 753.75 753.85 756.43 754.33 754.35 753.68 756.75 756.75

Incidents de fonctionnement
En cas dincidents de fonctionnement consulter le tableau ci-dessous. Dans le cas o ces informations ne permettraient pas didentifier et de rsoudre le problme constat, consulter votre fournisseur ou les services techniques de Bibby Scientific France. Nature de Lincident Llment chauffant ne chauffe pas Cause a) Elment chauffant dfectueux b) Rupture dalimentation lectrique c) Mise en route du thermostat Solution Remplacer llment chauffant Rtablir lalimentation Identifier la cause et rarmer le thermostat Rgler de dbit 60 l/h environ Fermer le robinet Rgler le dbit 60 l/h environ Vrifier que le tuyau dvacuation ne prsente aucun coude Rgler le dbit 60 l/h environ

Niveau deau dans le bouilleur trop bas. Elment chauffant non immerg Niveau deau dans le bouilleur trop haut ex : leau chaude monte vers le rfrigrant

a) Dbit deau de refroidissement trop faible b) Robinet de vidange ouvert a) Dbit deau de refroidissement trop lev b) lvacuation deau se fait mal

La temprature du distillat est trop leve (ex :/ 50C) Leau du bouilleur est pompe vers lvacuation

Dbit deau de refroidissement trop faible

a) Lvacuation du distillat vers le rservoir se fait mal b) Event obstru par la sonde du thermostat rfrigrant c) Dbit deau de refroidissement insuffisant

Vrifier que le tuyau de sortie du distillat ne prsente aucun coude Dgager lvent

Rgler le dbit 60 l/h environ Nettoyer le bouilleur

Mauvalse qualit du distillat* Dbit infrieur 4 l/h

Bouilleur entartr Alimentation lectrique infrieure au voltage requis

*Lorsque la qualit du distilat est contrle par les mesures de PH ou de conductivit, les valeurs mesures sont dpendantes de la temprature et de la prsence doxyde de carbone.

13

Notes

Bibby Scientific France SAS


ZI du Rocher Vert - BP 79 77793 Nemours Cedex France Tel: +33 1 64 45 13 13 Fax: +33 1 64 45 13 00 e-mail: bsf@bibby-scientific.fr www.bibby-scientific.com

Bibby Scientific Italia Srl


Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel: +39 (0)2 98230679 Fax: +39 (0)2 98230211 e-mail: marketing@bibby-scientific.it www.bibby-scientific.it

Bibby Scientific Ltd