BS3000M
BS3000M
User’s Manual
Built-in printer
Touch screen
PUSH Button
1
Stability of absorbency
Less than 0.002A within 20minsat 340nm
Repetition of absorbency
Coefficient variance(CV) ≤1.0%
Cross Contamination Rate:
When reaction liquid volume is 1ml, the cross contamination rate should be less than 1% by
flow cell test; no cross contamination by using separated cuvette inspection.
2. Installations
In order to ensure the normal operation of equipment, it must be installed and debugged by the
engineers who are authorized by Sinnowa or Sinnowa authorized organization. For any reason to
reinstall, debug the analyzer, only the authorized engineer can do the works.
Attention :
No authorized installation may make wrong or damage to the equipment, the problem or
damage is not in term of free warranty.
Before the installation, user and engineer must check and confirm that the lab meets the
requirements of space, power supply and working environment, etc.
Attention :
The power supply socket should be within 1m from the analyzer in order to pull out the plug
2
timely when accident happens.
Check if network voltage is the same to the equipment voltage
2.2.1 Steps:
Before opening the package, please carefully check the package. If the package is broken or wet
or polluted, please do not open it and contact immediately with the carrier and our local dealer. If
no outer damage, please open it by following steps:
-- unpack the package and check whether the packing contents are complete or not referring to
packing list
-- check whether the outer appearance is damage or not
-- check whether the serial number is in accord with packing list, outer package.
3
Load the new printing paper into the paper slot
Put the paper to the Feed form and press” FEED”
Cover printer cover
Attention :
Installing a thermal printing paper, pay attention to the direction of the paper
Before installing the printing paper, do not print, or else it will cost system crash
The probe or waste fluid joint may carry some serum, control, calibrator and reagent, which
is of potential biological risk. Therefore, it is dangerous to touch probe directly.
3.2 Operation
4
print it automatically
Edit: To add, modify, delete and print test items.
Result: Result print, print, delete, QC management, QC statistics and printing general report, etc.
Wash: The shortcut key for cleaning, aspiration volume is 1.5ml per keystroke, used to clean flow
cell.
Feed: Used for printer to load papers, pass papers and cut papers
Pump: Calibration for Peristaltic pump aspiration volume
Gain: AD Auto Zero to confirm whether the instrument is in the regular test status.
Filter: Measure and adjust each filter’s AD value, blank value and absorbance
Setup: Show cuvette temperature; set filter number, cuvette diameter, language, aspiration mode,
sleep mode; Hospital and instrument model title Settings; time format and date/time Settings;
screen brightness settings, etc.
Use to choose Temperature, click main menu to save. This function can only provide the
confirmation on whether temperature is right or not. When testing sample, the temperature will be
shifted according to the test program. Input the password and then input calibration temperature (for
professionals only). Password: 123456.
5
Attention:
Filter wheel, cuvette, aspiration mode, language setting, screen brightness settings, the password are all
123456.
The default values of the instrument is that 3000ul aspiration sample volume corresponds to 20000
motor steps. When the aspiration volume is not correct, adjusting is necessary. Enter into the calibrate
interface, input pump calibrate volume, then pour corresponding distilled water into the tube, insert
pipette into water, press PUSH to aspirate. After the distilled water was sucked done. Press PUSH
again, the instrument will display motor step and exit pump calibration to save step. If the aspiration
volume is still not right, repeat the above operation or input moter steps directly.
Attention:
● Pay attention to use environmental dustproof, moisture proof, and installation air conditioning is
better, environment temperature in 18-25 ° C is the best working temperature.
● When external power is not stable, instruments must connect regulated power supply
● Instruments in the process of operation shall not open the cover, so as not to cause damage to
equipment or operator
● Leakage and electrostatic prevention, the instrument should be in good grounding. Line power
socket must have a reliable grounding line to guarantee in a steady state and security
● After work, wash the instrument 3 times at least immediately to keep the cuvette and pipeline from
6
liquid waste.
● After finish the testing, the used should be collected and disposed according to the medical waste.
● The waste pipe end should not be dipped into the waste to avoid poor drainage.
● Use qualified reagent within the period of validity.
Attention:
AD auto zero is that the instrument is based on water to measure the initial light intensity of
different wavelength. The gain coverage, AD value and offset value are calculated together
to obtain absorbance. This step is very important; users should do this step at each starting
the instrument.
7
AD auto zero for flow cell need more distilled water, it is recommended that the aspiration
pipe should insert into distilled water to ensure no bubble when AD auto zero.
It is also available to use cuvette for AD auto zero, the distilled water in the cuvette should
reach more than 10mm away from the bottom of the cuvette.
4. Device Maintenance
4.1 Daily Maintenance
Daily maintenance is mainly on washing the flowing cuvette, keep it clean. Before testing, 10 times
washing are necessary. After each test, wash 4-5 times is essential. If there are bubbles in flowing
cuvette, you may draw ethanol to soak and flush firstly, then wash it by distilled water. When all tests
done, please use distilled water flush.
Daily maintenance is mainly about cleaning dust and stains on the casing, correct sample absorption
amount of peristaltic pump.
5 Troubleshooting
This chapter explains all kind of malfunctions, which often happen in the routine operations. Besides, it
analyzes the related reasons about malfunctions and supplies some methods against the malfunctions.
Warnings:
● You must turn off the analyzer, cut off the power, and then remove the power plug from the socket.
The repair work must be taken by SINNOWA professional trained men.
● The analyzer must use suited power supply and voltage. Or else, the damage which is caused against
this order is out of SINNOWA’s responsibility.
Caution:
● Analysis of samples may give incorrect test results in the case of instrument malfunction. If there is a
fault detected in the sample, be sure to troubleshoot before use.
8
● Sample, quality control samples, calibration samples, wasted liquid and so on have potential
biochemistry risk. The operator must comply with the laboratory regulations about the safety operator
to wear personal protective device (like: laboratory protective clothing, gloves etc.), and accordance
with local government regulations to dispose the waste materials generated by the instrument
detection.
Cause: Due to the local ac voltage is different; Led display voltage is not the same, but in general all in the
visible range
Maintenance: Open the instrument, find the cable of main board connecting with the display, you can see a
blue 203 potentiometer; adjust the potentiometer to change the definition.
Cause: The cable of printer head connecting with control panel is loose.
Maintenance: the cable of printer head connecting with control panel is white, open the instrument, take
off the printer head, and compress the interface.
9
Attention: Be careful about the white cable and the printer head cable do not drag too hard.
Cause:
a) Effect of ambient temperature (Especially in winter), increase indoor environment temperature, keep the
environment temperature to 10℃~30℃;
b) The heating voltage is insufficient could lead to a longer heating time. Please open the instrument, use
the digital multimeter to test the voltage of incubator heating rod, if there is no voltage, and replace the
heating rod.
Maintenance: Against the possible reasons check and maintenance individually.
Cause: Serum and reagent do not have enough incubate time. Or Instrument needs calibration again.
Maintenance: Serum and reagent incubate more than 3 minutes in winter it should around 5 minutes. Or
reference to the manual, run the calibration procedure to calibration.
In order to make the analyzer’s running reliable, it’s necessary to proofread or replace some parts of
analyzer and take effective maintenance.
Attention:
● The user must be trained by our professional engineers before they take maintenance and replacements
alone.
1. Turn off the power of analyzer and pull out the power cord.
2. Pull the power wire from the power socket of mainframe and elicit the fuse housing.
3. Take out the fuse housing and install a new fuse into fuse housing.
4. Plug the fuse housing into original position
Warning:
10
● The operator must use fuse of appointed specification.
The standard configuration of the instrument has a halogen lamp set in the opposite position of detector;
Two LED light sources are disposed on two adjacent sides of the detection position and the detector.
It needs replacing while the lamp is damaged or has been working for 2 years.
The replacement to LED: when instrument scattering detection is in abnormal, the LED light source should
be checked or replaced.
Attention:
● After change the lamp and LED, Please check and confirm the light source and cable are secure
connected, and after calibration, then it can put into normal use.
Warnings:
Figure 4
Attention:
● Unplug the pump pipe should be very careful, to prevent break incoming point and outlet point
● In order to guarantee the reliability of the test, need to inspect the peristaltic pump pipe every month
● Replacement of the pump pipe must be provided by SINNOWA, do not use other types of pump pipe
to replace
12
6. TRANSPORTATION AND STORAGE
6.1 Transportation
Transport should be in accordance with the regulations implementing of order contract, Away from the
toxic, harmful, corrosive substances
It should be to prevent severe shocks, rain and exposure, overturned not be permitted in transportation.
6.2 Storage
It should be stored in environment temperature −5℃~50℃, Relative humidity no more than 80%,
well-ventilated indoor. It shouldn’t storage with toxic, harmful, corrosive materials stored.
13