Manual DTX
Manual DTX
User Manual
224311-PTU rev. 15
ENGLISH
Contents
1 Disclaimer.............................................................................................................................. 5
2 Introduction........................................................................................................................... 6
2.1 ORTHOPANTOMOGRAPH™ OP 3D™ LX.................................................................................. 6
2.2 Intended use and intended purpose..........................................................................................7
2.3 Intended user profile................................................................................................................. 7
2.4 Radiation guideline.................................................................................................................... 7
2.5 Contraindications.......................................................................................................................7
2.6 Associated documentation........................................................................................................ 8
2.7 Abbreviations.............................................................................................................................8
2.8 Signal words..............................................................................................................................9
2.9 Disposal and recycling.............................................................................................................. 9
2.10 Warnings and requirements.....................................................................................................9
2.10.1 Warnings and precautions for use............................................................................. 9
2.10.2 Connection requirements.........................................................................................10
2.10.3 Device modifications................................................................................................11
2.10.4 Protection Against Cybersecurity Threats................................................................ 11
3 Overview..............................................................................................................................12
3.1 Main parts............................................................................................................................... 12
3.2 Patient positioning lights..........................................................................................................13
3.2.1 Positioning light locations.......................................................................................... 13
3.2.2 Panoramic lights........................................................................................................ 13
3.2.3 Cephalometric lights.................................................................................................. 14
3.2.4 3D lights.................................................................................................................... 14
3.3 Accessories............................................................................................................................. 15
3.4 Other detachable parts........................................................................................................... 16
3.5 Emergency stop switch........................................................................................................... 17
3.6 Device label locations..............................................................................................................18
4 User interfaces...................................................................................................................20
4.1 Graphical User Interface (GUI)................................................................................................20
4.1.1 Overview................................................................................................................... 20
4.1.2 Panoramic imaging.................................................................................................... 23
4.1.3 Cephalometric imaging.............................................................................................. 25
4.1.4 3D imaging................................................................................................................ 28
4.1.5 Dental cast view........................................................................................................ 34
4.1.6 Device settings.......................................................................................................... 38
4.2 Control panel...........................................................................................................................42
4.3 Status indicator light................................................................................................................43
5 Imaging programs.............................................................................................................44
5.1 Panoramic programs............................................................................................................... 44
5.2 Cephalometric programs......................................................................................................... 46
5.3 3D programs........................................................................................................................... 48
5.3.1 FOV Sizes..................................................................................................................48
5.3.2 3D Resolutions.......................................................................................................... 51
5.4 Dental cast program............................................................................................................... 51
ORTHOPANTOMOGRAPH™ OP 3D LX iii
6 Using the device............................................................................................................... 52
6.1 General imaging workflow....................................................................................................... 52
6.2 Powering on/off the device...................................................................................................... 53
6.3 Preparing the device for imaging.............................................................................................54
6.4 Patient positioning................................................................................................................... 55
6.4.1 Patient positioning for Panoramic imaging................................................................. 55
6.4.2 Patient positioning for Cephalometric imaging........................................................... 59
6.4.3 Patient positioning for Carpus imaging...................................................................... 63
6.4.4 Patient positioning for 3D imaging............................................................................. 65
6.5 Taking an image......................................................................................................................68
6.5.1 Taking Panoramic and Cephalometric images........................................................... 68
6.5.2 Taking Scout and 3D images.................................................................................... 69
6.5.3 Taking Dental cast images........................................................................................ 70
7 Maintenance....................................................................................................................... 72
7.1 Cleaning and decontamination................................................................................................ 72
7.2 Calibrations for the user.......................................................................................................... 73
7.2.1 When to calibrate the device.....................................................................................73
7.2.2 Preparation for calibration......................................................................................... 73
7.2.3 Calibration procedure................................................................................................ 75
7.3 Quality control......................................................................................................................... 79
7.3.1 PAN QC.................................................................................................................... 79
7.3.2 3D QC.......................................................................................................................81
7.3.3 CEPH QC.................................................................................................................. 83
7.4 Annual maintenance................................................................................................................85
8 Troubleshooting.................................................................................................................86
9 Technical data....................................................................................................................87
9.1 Technical specifications...........................................................................................................87
9.2 Imaging program specifications............................................................................................... 91
9.2.1 Panoramic programs................................................................................................. 91
9.2.2 Cephalometric programs........................................................................................... 92
9.2.3 3D programs............................................................................................................. 92
9.2.4 Patient size setting default values..............................................................................94
9.3 Patient contacting parts.......................................................................................................... 95
9.4 Device dimensions.................................................................................................................. 96
9.4.1 Main device dimensions............................................................................................ 96
9.4.2 Cephalometric device dimensions............................................................................. 97
9.5 Symbols that may appear on the device or its parts................................................................98
9.6 Electromagnetic Compatibility (EMC) tables.......................................................................... 100
9.7 X-ray tube assemblies........................................................................................................... 106
9.8 Workstation minimum requirements.......................................................................................108
iv ORTHOPANTOMOGRAPH™ OP 3D LX
1 Disclaimer
1 Disclaimer
ORTHOPANTOMOGRAPH™ OP 3D LX User Manual, 224311-PTU r15.
The manufacturer has no liability for consequential damage, personal injury, loss, damage or expense
directly or indirectly caused by the use of the product. No agent, distributor or other party is authorized to
give warranty or other liability on behalf of the manufacturer with respect to its products.
Manufacturer:
PaloDEx Group Oy
Nahkelantie 160
FI-04300 Tuusula
FINLAND
Download this manual in digital format or order one in printed format for no additional cost from:
http://www.dexis.com
NOTICE! Make sure that you get the manual that applies to your device.
ORTHOPANTOMOGRAPH™ OP 3D LX 5
2 Introduction
2 Introduction
The device can be used for the following procedures depending on the device configuration and country-
specific regulatory approvals:
Panoramic Imaging
• Standard panoramic
• Segmented panoramic
• Pediatric panoramic
• Bitewing
• TMJ, lateral projection
Cephalometric Imaging
• Cephalometric lateral projection
• Cephalometric pediatric lateral projection
• Cephalometric Posterior-Anterior (PA) projection
• Carpus view
3D CBCT Imaging
• FOV (3D Volume Height x Diameter in centimeters)
◦ 5x5
◦ 6x9*
◦ 8x8
◦ 10 x 10 *
◦ 12 x 15 **
◦ 15 x 20 **
• Resolution
◦ Low (using LDT - Low Dose Technology™)
◦ Standard
◦ High
◦ Endo (5 x 5 only)
• Scout images for FOV position and height verification
• Dental cast imaging
* In some countries/areas where 3D volume dimensions are limited by local regulations FOV sizes 6 x 9 and
10 x 10 are replaced with FOV sizes 6 x 8 and 8 x 8.
** Optional FOV sizes available as separately purchasable features.
6 ORTHOPANTOMOGRAPH™ OP 3D LX
2 Introduction
NOTICE! This manual describes the use of a device with all available functions and so may contain
instructions to more functions than what are available with your device configuration.
To function, the device requires a workstation with a dental imaging software. The device produces X-ray
images in the industry standard image formats (DICOM and JPEG).
Any software used to capture and view X-ray images with the device (e.g. DTX Studio Clinic or a third party
dental imaging software supporting TWAIN) must conform to the local regulatory standards, regulations and
approval required to place the device on the market. Any workstation used with the device must conform
to the connection and hardware requirements presented in chapters Connection requirements on page
10 and Workstation minimum requirements on page 108 as well as the dental imaging software system
requirements.
The device produces X-ray beams to produce digital images. Reduce the risk of excessive radiation
exposure by following the As Low As Reasonably Achievable (ALARA) principle and try to reduce the
radiation dose to the level where obtained images are still clinically adequate.
The justification for X-ray exposure and the balance between intensity of radiation exposure and image
quality should always be considered.
The X-ray beam intensity is a product of beam quality (kVp), beam quantity (mAs) and the distance between
the X-ray tube and the exposed area. The radiation intensity is provided as Dose Area Product value
2
(mGycm ) on the GUI.
2.5 Contraindications
X-rays are known to be potentially harmful and the decision if the risk outweighs the benefit is always done
by the professional user. No contraindications are identified for this device.
ORTHOPANTOMOGRAPH™ OP 3D LX 7
2 Introduction
2.7 Abbreviations
3D Three Dimensional
PAN Panoramic
CEPH Cephalometric
L Left
R Right
H Head
F Foot
P Posterior
A Anterior
PA Posterior-Anterior
FOV Field Of View. The cylindrical 3D volume that is reconstructed by the device. The FOV
sizes are presented as the Height x Diameter of the 3D volume in centimeters.
ROI Region Of Interest. The anatomical area or region that you are interested to examine.
FH Frankfort-Horizontal
MAR Metal Artifact Reduction. Reduces the effect of metal and other dense radiopaque
objects, that typically create artifacts which are seen as stripes and shadows.
QC Quality Control
8 ORTHOPANTOMOGRAPH™ OP 3D LX
2 Introduction
At least the following parts of the device should be recycled according to local and national regulations
regarding disposal of non-environmentally friendly materials:
• Tubehead (oil)
• All electronic circuits and electronic boards
• Plastic parts
ORTHOPANTOMOGRAPH™ OP 3D LX 9
2 Introduction
• When taking an X-ray image of a patient who is feeling insecure or has exceptional anatomy (typically
very tall or large patient) use the Test-mode first to demonstrate the device movements and to make sure
that the rotating device doesn't collide with the patient during the imaging process.
• If the patient is using a pacemaker, consult the manufacturer of the pacemaker before taking an X-ray
image to confirm that the device will not interfere with the operation of the pacemaker.
• Avoid taking X-ray images of pregnant women.
• Switch the device off after use.
• This device complies with the EMC (Electromagnetic Compatibility) according to IEC 60601-1-2. Radio
transmitting equipment, cellular phones etc. shall not be used in proximity of the device as they could
influence the performance of the device.
• Where it is likely that evaluation of soft tissues will be required as part of the patient’s radiological
assessment, the imaging should be done using conventional medical computed tomography or magnetic
resonance, rather than 3D imaging using Cone Beam technology.
• The device must not be used for screening examinations.
• Do not move the upper shelf of the device by hand as it may hit a wall or cause the device to go to an
error state.
• Report any incident, related to use of this device, having a serious effect on the health of a patient, user
or other person to the manufacturer and to the local competent authority.
• USA only: Federal law restricts this device to sale by or on the order of a dentist or other qualified
professional.
10 ORTHOPANTOMOGRAPH™ OP 3D LX
2 Introduction
• Connecting the device to an IT-network that includes other equipment or changing the IT-network
can cause unidentified safety or security risks to patients or operators. It is the responsibility of the
organization controlling the IT-network to identify, analyze, evaluate and control these possible risks.
• The device must be connected to a private, firewall protected local area network to ensure proper
data security. All connections from outside of the local area network to the device must be blocked.
Connections between the device and the workstation in the local area network must be allowed.
• Network elements used in the IT-network must not use their default passwords and wireless connections
must be encrypted.
• The workstation connected to the device must use an anti-virus software and have individual password
protected user accounts to prevent unauthorized access to patient data.
CAUTION! Failure to maintain cybersecurity can result in compromised device functionality, loss
of data (medical or personal) availability or integrity, or exposure of other connected devices or
networks to security threats.
• Workstation should have anti-virus software running on it.
• Access to the workstation should be protected by strong passwords.
• Passwords should not be shared or reused.
ORTHOPANTOMOGRAPH™ OP 3D LX 11
3 Overview
3 Overview
MAIN DEVICE
1. Column 2. Carriage 3. Lower shelf
4. Rotator 5. Upper shelf 6. PAN/3D Tubehead
7. X-ray sensor 8. Exposure switch 9. Patient handles
10. Emergency stop switch 11. Control panel 12. Head support
13. Mirror 14. Status indication light
CEPHALOMETRIC UNIT
15. CEPH tubehead 16. CEPH arm 17. CEPH head support
18. Nasion support 19. Ear rods
12 ORTHOPANTOMOGRAPH™ OP 3D LX
3 Overview
1 - Tilt light
2 - Midsagittal light
3 - FH (Frankfort-Horizontal) light
NOTICE! The position of FH light is automatically adjusted according to the selected imaging
program.
ORTHOPANTOMOGRAPH™ OP 3D LX 13
3 Overview
5 - FH (Frankfort-Horizontal) light
NOTICE! Used only in lateral imaging programs.
3.2.4 3D lights
2 - Midsagittal light
3 - Horizontal light, top of FOV
4 - Horizontal light, bottom of FOV
NOTICE! The position of the top of FOV light is automatically adjusted according to the selected
FOV.
14 ORTHOPANTOMOGRAPH™ OP 3D LX
3 Overview
3.3 Accessories
Chin rest Bite block Lip support
ORTHOPANTOMOGRAPH™ OP 3D LX 15
3 Overview
16 ORTHOPANTOMOGRAPH™ OP 3D LX
3 Overview
Pressing the emergency stop switch immediately terminates the imaging and all device movements.
NOTICE! An interrupted imaging process cannot be resumed. A new image needs to be taken.
ORTHOPANTOMOGRAPH™ OP 3D LX 17
3 Overview
18 ORTHOPANTOMOGRAPH™ OP 3D LX
3 Overview
NOTICE! These labels are for illustrative purposes only. Texts and images may vary.
ORTHOPANTOMOGRAPH™ OP 3D LX 19
4 User interfaces
4 User interfaces
20 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
Test mode
Indicates if Test mode is active. Test mode disables device radiation
production.
Test mode can be used for example to demonstrate the device movements.
You can also use the control panel to enable/disable the Test mode.
Patient size
Indicates the currently selected patient size preset.
• Press the Patient size selection icon to open a list of patient sizes;
Imaging parameters
Indicates the currently selected kV and mA values. By default the selection is
based on the imaging program and patient size selections.
• Press the Imaging parameters to open an adjustment menu.
• If needed, adjust the kV and mA values manually with the + and - buttons
ORTHOPANTOMOGRAPH™ OP 3D LX 21
4 User interfaces
Message bar
The message bar provides messages and guidance to the user. Press on the
bar to show messages if there are more than the ones shown.
22 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
A sample image for the selected imaging program is shown on the GUI. If
you take a segmented image, the unselected segments are grayed out in the
sample image.
ORTHOPANTOMOGRAPH™ OP 3D LX 23
4 User interfaces
Zoom slider
You can zoom the preview image in/out using the zoom slider.
While zoomed in, you can click and drag to scroll the image.
Brightness slider
You can adjust the brightness of the shown preview image using the brightness
slider.
Contrast slider
You can adjust the contrast of the shown preview image using the contrast
slider.
OK button
Press OK button to close the image preview.
24 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
The cephalometric view shows which parts of the skull are imaged with the
selected imaging program.
In lateral imaging, the image field width can be manually adjusted by dragging
the slider control. The available adjustment range is indicated with a scale.
ORTHOPANTOMOGRAPH™ OP 3D LX 25
4 User interfaces
A sample image for the selected imaging program is shown on the bottom of
the GUI. The grayed out area shown in the sample image is an approximation of
the unexposed region.
26 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
Zoom slider
You can zoom the preview image in/out using the zoom slider.
While zoomed in, you can click and drag to scroll the image.
Brightness slider
You can adjust the brightness of the shown preview image using the brightness
slider.
Contrast slider
You can adjust the contrast of the shown preview image using the contrast
slider.
OK button
Press OK button to close the image preview.
ORTHOPANTOMOGRAPH™ OP 3D LX 27
4 User interfaces
4.1.4 3D imaging
4.1.4.1 3D imaging program settings
Dental chart is used to select the Region of Interest (ROI) for the 3D scan. The FOV changes automatically
according to the selections made in the dental chart.
• To individually select the studied parts of dentition, press teeth, jaw and TMJ icons:
◦ A single tooth (1)
◦ Range of teeth (2)
◦ Jaw with all teeth (3)
◦ TMJ (4)
◦ Maxillary Sinus (5)
◦ Frontal Sinus (6)
◦ Whole dentition
NOTICE! Both TMJs are selectable only with 12 x 15 and 15 x 20 FOV sizes.
• To select all parts of the dentition, double-click on any part of the dentition.
• To clear all selections, click on the gray area (7) on the lower corners of the dental chart.
• To undo the latest action and return to previous selection, double-click on any part of the dentition.
28 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
FOV size
Indicates the currently selected FOV size (H x D).
The FOV changes automatically according to the selections made in the dental
chart but can be changed manually also:
• Press the FOV size icon to open a list of the available 3D FOV sizes.
NOTICE! Use the ORTHOselect™ dental chart as primary method for FOV
size selection.
NOTICE! The list of available FOV sizes depend on the device configuration
and country specification.
Image resolution
Indicates the currently selected resolution. The resolution affects the image
quality and radiation dose to the patient.
• Press the image resolution selection icon to open a list of available image
resolutions;
ORTHOPANTOMOGRAPH™ OP 3D LX 29
4 User interfaces
A Scout image is taken to verify and perform adjustment to the FOV position
and height before initiating a full 3D scan. By default, the Scout image is always
active when taking 3D images.
When the Scout image program is active, the FOV coverage illustration turns
green.
30 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
Brightness slider
You can adjust the brightness of the shown image preview using the brightness
slider.
NOTICE! Adjustment is not saved to the image.
Contrast slider
You can adjust the contrast of the shown image preview using the contrast
slider.
NOTICE! Adjustment is not saved to the image.
ORTHOPANTOMOGRAPH™ OP 3D LX 31
4 User interfaces
OK button
Press OK button to approve the changes made to the FOV location and size
using the Scout image and to proceed to the 3D exposure.
NOTICE! The image has indications from which direction the image is shown; A/P (Anterior/
Posterior), L/R (Left/Right) and H/F (Head/Foot).
Slice preview
Shows a preview of 3D slices. You can scroll through the preview using the
slider on the right side of the image.
32 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
Volume preview
Shows a preview of the 3D volume. You can rotate the preview using the slider
on the right side of the image.
Brightness slider
You can adjust the brightness of the shown image preview using the brightness
slider.
NOTICE! Adjustment is not saved to the image.
Contrast slider
You can adjust the contrast of the shown image preview using the contrast
slider.
NOTICE! Adjustment is not saved to the image.
OK button
Press OK button to close the image preview.
ORTHOPANTOMOGRAPH™ OP 3D LX 33
4 User interfaces
Image resolution
Indicates the currently selected resolution. The resolution affects the image
quality and the amount of image layers.
• Press the image resolution selection icon to open a list of available image
resolutions;
34 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
Brightness slider
You can adjust the brightness of the shown image preview using the brightness
slider.
NOTICE! Adjustment is not saved to the image.
Contrast slider
You can adjust the contrast of the shown image preview using the contrast
slider.
NOTICE! Adjustment is not saved to the image.
ORTHOPANTOMOGRAPH™ OP 3D LX 35
4 User interfaces
OK button
Press OK button to approve the changes made to the FOV location and size
using the Scout image and to proceed to the 3D exposure.
NOTICE! The image has indications from which direction the image is shown; A/P (Anterior/
Posterior), L/R (Left/Right) and H/F (Head/Foot).
Slice preview
Shows a preview of 3D slices. You can scroll through the preview using the
slider on the right side of the image.
36 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
Volume preview
Shows a preview of the 3D volume. You can rotate the preview using the slider
on the right side of the image.
Brightness slider
You can adjust the brightness of the shown image preview using the brightness
slider.
NOTICE! Adjustment is not saved to the image.
Contrast slider
You can adjust the contrast of the shown image preview using the contrast
slider.
NOTICE! Adjustment is not saved to the image.
OK button
Press OK button to close the image preview.
ORTHOPANTOMOGRAPH™ OP 3D LX 37
4 User interfaces
1. QUALITY CONTROL
Shows a list of Quality Control (QC) programs, their completion status and last completion date.
The Quality Control (QC) programs are used to ensure that the technical performance and the image
quality of the device remains constant and valid for clinical use. Quality Control should be performed at
regular intervals, preferably at least once a month and always after calibration.
See chapter Quality control on page 79 for more information on how to take quality control images.
2. CALIBRATIONS
Shows a list of user performable device calibrations, their completion status and last completion date.
The calibrations are carried out through this menu.
See chapter Calibrations for the user on page 73 for more detailed information on calibration
programs and how to perform them.
NOTICE! The available calibration programs depend on the device configuration.
38 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
3. SETTINGS
Device settings:
Friendly name Set a name for the device, which is shown next to the device
settings icon and on the imaging software.
3D Metal Artefact Enable or disable MAR, Metal Artefact Reduction. MAR is used to
Reduction reduce the effect of metals and other dense radiopaque objects on
the 3D image.
It is recommended to keep MAR activated.
Force device language Change the language of the device GUI if needed.
It is recommended not to force the language of the GUI but change
the language profile of the workstation.
Default to Scout before 3D Enable or disable automatic Scout image activation when a 3D
imaging modality is selected. It is recommended to have the Scout
image taking enabled.
NOTICE! If the selection is set to "disabled", a Scout image
mode can still be activated manually in the 3D imaging
program selection view.
Volume level Adjust the volume level of the audible signals played by the device.
NOTICE! Does not affect exposure warning signal.
Button volume level Adjust the volume level of the audible signal played when a button
on the control panel is pressed.
The default volume level setting is 8.
Button sound Select the sound that is played when HOME, Test mode or
Patient positioning lights buttons are pressed.
ORTHOPANTOMOGRAPH™ OP 3D LX 39
4 User interfaces
Time zone Set the timezone to match the installation location. To change the
time zone, start to write the continent, capitol or state (US) and
select the correct selection from the drop down list.
Adjust the default mA and kV values for medium size patient preset. Other patient size selections are
changed relatively to the adjustment made.
Adjustments made to the 3D imaging program with standard resolution are applied relatively to other
resolution selections as well.
The default values for each program are listed in chapter Patient size setting default values on page
94.
NOTICE! The available programs and patient size presets depend on the device configuration.
40 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
4. ABOUT
Exposure counters Shows the amount of exposures taken with the device.
5. SERVICE MODE
NOTICE! This menu is intended for authorized service personnel only.
This menu is used to unlock Installation and Service specific functions on the GUI.
• Additional device settings
• Calibrations for installation and service of the device
• Optional imaging programs activation
• Demonstration mode activation for exhibition use
• Verification programs for radiation related testing
To access these functions, a PIN code is required.
NOTICE! The functions unlocked by the PIN code are highlighted with coloured indicators.
ORTHOPANTOMOGRAPH™ OP 3D LX 41
4 User interfaces
1. Device height controller. Move the device up/down by pressing on the device height controller. The
device can move in two speeds, slow and fast.
MOVE DEVICE UP
1. Press the underside of the controller to move 2. Press also the top of the controller to move
slowly. faster.
1. Press the top of the controller to move slowly. 2. Press also the underside of the controller to
move faster.
2. Patient positioning lights button. Turn the patient positioning lights on/off.
3. Test mode button. Disable radiation production of the device. You can also use the GUI to activate the
Test mode.
42 ORTHOPANTOMOGRAPH™ OP 3D LX
4 User interfaces
4. Device status indicators. Device status indicators light up according to the device status.
Ready
Device is ready for imaging.
Error state
Device is in error state or pending user action. Check the GUI for details.
Exposure
Device is generating X-rays.
5. HOME button. Move the device to HOME (Patient-In) and scan start positions.
ORTHOPANTOMOGRAPH™ OP 3D LX 43
5 Imaging programs
5 Imaging programs
The Standard Panoramic imaging program provides general view of dental and facial anatomy
based on panoramic imaging technique.
You can choose which segments of the dentition are imaged if imaging of the whole dentition is
not required. Press the segments on the dental chart to deselect and select them.
The Pediatric Panoramic imaging program provides a general view of the dental and facial
anatomy based on panoramic imaging technique for pediatric patients while using smaller
radiation dose and smaller imaging area.
Adult patients with exceptionally narrow jaw can also be imaged with this program but note
that the image height is limited.
You can choose which segments of the dentition are imaged if imaging of the whole dentition is
not required. Press the segments on the dental chart to deselect and select them.
44 ORTHOPANTOMOGRAPH™ OP 3D LX
5 Imaging programs
Bitewing program
Lateral TMJ program provides a lateral view of the patient's left and right temporomandibular
joints.
You can choose to image both or only other segment in single scan. Press the segments on
the dental chart to deselect and select them.
ORTHOPANTOMOGRAPH™ OP 3D LX 45
5 Imaging programs
Cephalometric programs provide projection images of patient skull and dental anatomy. Images are utilized
in orthodontics and general diagnostics.
NOTICE! The image field dimensions shown here are for illustrative purposes only.
The lateral projection uses a full height image field that provides a nearly full skull image.
You can adjust the width of the image field by dragging the slider control on the cephalometric
view. The available image field widths, represented at the image plane, are 16-26 cm. The
image field starts at the center of the ear rods with the smallest image width selection.
The exposure time depends on the width of the selected image field.
The pediatric lateral projection is a lateral projection with a reduced image height, optimized
for pediatric patients.
The program can also be used for adult patients to reduce the radiation dose when imaging
of the upper part of the skull is not necessary. The pediatric lateral projection covers all typical
cephalometric landmarks from the glabella to the spine.
You can adjust the width of the image field by dragging the slider control on the cephalometric
view. The available image field widths, represented at the image plane, are 16-26 cm. The
image field starts at the center of the ear rods with the smallest image width selection.
The exposure time depends on the width of the selected image field.
46 ORTHOPANTOMOGRAPH™ OP 3D LX
5 Imaging programs
PA projection imaging can be used, for example, to identify any facial asymmetries,
dentoalveolar asymmetries, dental cross-bite and mandibular displacement.
In PA projections, the patient faces away from the tubehead to keep the radiation dose to
patient's eyes in minimal level.
PA projection uses a full height image field.
ORTHOPANTOMOGRAPH™ OP 3D LX 47
5 Imaging programs
5.3 3D programs
NOTICE! These programs are available only for 3D devices.
Always select the smallest feasible FOV size, resolution and imaging parameters for the 3D image in order to
follow the ALARA (As Low As Reasonably Achievable) principle.
NOTICE! It is always up to the dental professional to select the appropriate FOV, resolution and
imaging parameters.
NOTICE! FOV sizes 12x15 and 15x20 are optional programs and available only with a separately
purchased license.
FOV 5 x 5
Optimized for single site implants or localized diagnostics, for example, 3rd molar
extractions, impacted teeth, periodontal cases, root fractures, single TMJ analysis,
endodontics and for pediatric imaging.
Available resolutions:
Low (LDT)
Standard
High
Endo
48 ORTHOPANTOMOGRAPH™ OP 3D LX
5 Imaging programs
FOV 6 x 9 / 6 x 8
Optimized for multiple implant placement using surgical guides, imaging
the whole dental arc of one jaw, 3rd molar visualization, pathology (bilateral
analysis) and periodontal cases.
Available resolutions:
Low (LDT)
Standard
High
FOV 10 x 10 / 8 x 8
Optimized for imaging the entire dentition, both mandibula and maxilla with
portion of maxillary sinuses.
Available resolutions:
Low (LDT)
Standard
High
ORTHOPANTOMOGRAPH™ OP 3D LX 49
5 Imaging programs
FOV 12 x 15
Optimized for imaging the entire dentition, both mandibula and maxilla, including airway
and upper cervical spine or the sinus, maxillary sinuses, jaws with bilateral joints, jaws with
airways, analyses of both TMJs and for maxillofacial surgeries.
Available resolutions:
Low (LDT)
Standard
High
FOV 15 x 20
Optimized for imaging the maxillofacial complex, both mandibula and maxilla, including
airway and upper cervical spine or the sinus, maxillary sinuses, jaws with bilateral joints,
jaws with airways, analyses of both TMJs, maxillofacial surgeries and orthodontic analyses.
Available resolutions:
Low (LDT)
Standard
High
50 ORTHOPANTOMOGRAPH™ OP 3D LX
5 Imaging programs
5.3.2 3D Resolutions
The resolution affects the image quality and radiation dose to the patient. For example, high resolution will
produce more details than standard resolution, but the radiation dose is also higher. The unit offers a LDT
(Low Dose Technology™) resolution which can be used for example in treatment follow-up cases. The LDT
resolution will result in images proportional to the low dose and it is up to the healthcare professional to
decide when it is sufficient to be utilized.
Available resolutions:
NOTICE! The resolution selection affects the amount of
image layers produced. Dose production is not affected.
Low resolution
Standard resolution
High resolution
ORTHOPANTOMOGRAPH™ OP 3D LX 51
6 Using the device
DTX
STUDIO
20.3 cm
66 kV 66 kV
10.0 mA
10 mA
3D 2D
Release patient
Take Scout Take X-ray and end study
image image
Adjust FOV
52 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
NOTICE! The device enters a power save mode after 30 minutes of inactivity. While in the
power save mode, the device cooling fan is not active, but the control panel lights are lit. The
device wakes up from the power save mode when connection to it is established or when the
HOME button is pressed.
ORTHOPANTOMOGRAPH™ OP 3D LX 53
6 Using the device
5. The device GUI opens on the workstation when the connection to the device is established.
NOTICE! The device must be powered on and initialized before the connection is established.
6. GUI: Select the imaging modality, imaging program and patient size to set up the device. If the
preset selections are not suitable for the patient in question, adjust the imaging parameters manually.
Imaging modality
Imaging program
Patient size
Imaging parameters
NOTICE! Exercise special care when imaging patients outside the typical adult size range,
especially smaller pediatric patients. When imaging pediatric patients follow the ALARA (As
Low As Reasonably Achievable) principle and seek to reduce the radiation dose to the amount
necessary to acquire images that are clinically adequate. This can be achieved for example
through utilizing the pediatric imaging programs and adjusting the imaging parameters for the
patient in question.
54 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
NOTICE! Make sure that the device is set up correctly for the intended examination.
2. The device goes to ready state and the status indicator light turns green.
NOTICE! If the status indicator light does not turn green, look for more information on the GUI.
3. Select the patient positioning accessory according to the image to be taken, attach it to the chin rest
and place them on the lower shelf of the device, as shown below.
NOTICE! The Chin Rest needs to be attached to the device in the correct orientation for the
selected imaging program before the device allows images to be taken.
ORTHOPANTOMOGRAPH™ OP 3D LX 55
6 Using the device
7. Tell the patient to remove their glasses, hearing aids, removable dentures, jewellery, hair clips and all
other things that can cause artifacts to the image.
8. Protect the patient from radiation according to the local regulations, for example with a lead apron or a
thyroid shield.
9. Guide the patient to the device and instruct to stand as straight and as tall as possible.
10. Tell the patient to hold the patient handles, place their chin on the chin rest and to bite the notches on
the bite block or to push their lip against the lip support if it is used.
CAUTION! Evaluate the condition of the patient’s dentition before using the Bite Block. If the
patient's teeth are more fragile than a normal healthy patient's, use the Lip Support and a cotton
roll between the front teeth instead.
NOTICE! If the patient has wide shoulders that could collide with the rotating unit, ask the
patient to cross their arms when holding the handles in order to contract the shoulders.
11. Ask the patient to take a step forward to straighten their spine.
56 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
12. Fine adjust the device height and adjust patient's head position/orientation using the patient positioning
lights as guides.
NOTICE! To ensure optimal image quality, pay attention to the correct patient positioning.
NOTICE! You can open the mirror and use it to help in positioning the patient. Close the mirror
before you start the imaging.
• Align the patient so that the root tips of lower and upper incisors are vertically aligned and parallel
to the Tilt light. The ORTHOfocus™ feature automatically finds the optimum panoramic image layer,
enabling forgiving patient positioning.
NOTICE! When you align the incisors to the same vertical level, the patient's Frankfort-
Horizontal plane is also in the necessary orientation and close to parallel to the FH light.
NOTICE! In rare cases, typically caused by abnormal anatomy, if the ORTHOfocus™
feature cannot find the optimal image layer, the device provides 5 different image layers to
use in diagnosis.
BITEWING IMAGING:
• The patient’s occlusal plane should be horizontal and parallel to the FH light.
ORTHOPANTOMOGRAPH™ OP 3D LX 57
6 Using the device
TMJ IMAGING:
• The patient’s Frankfort-Horizontal plane should be close to parallel to the horizontal light. The FH
light indicates the center of the field. You can take a TMJ image with the patient's mouth closed or
open.
13. Check that the patient’s head is straight and not rotated and the patient’s mid-line coincides with the
midsagittal light.
16. Ask the patient to close their lips and press their tongue against the palate if possible.
58 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
1. Press HOME button. The device moves to CEPH HOME (Patient-In) position.
NOTICE! Ensure that the device has been set up correctly for the required examination.
2. The device goes to ready state and the status indicator light turns green.
NOTICE! If the status indicator light does not turn green, look for more information on the GUI.
3. Turn the CEPH head support to Lateral or PA position, according to the required examination.
4. To ease the patient entry you may turn the nasion support an the ear rods aside and pull them to their
out-most position.
ORTHOPANTOMOGRAPH™ OP 3D LX 59
6 Using the device
7. Tell the patient to remove their glasses, hearing aids, removable dentures, jewellery, hair clips and all
other things that can cause artifacts to the image.
8. Protect the patient from radiation according to the local regulations, for example with a lead apron or a
thyroid shield.
9. Guide the patient to the device and instruct to stand as straight and as tall as possible.
10. Turn the ear rods down if they were previously turned aside.
11. Fine adjust the device height so that the ear rods are on the level of the patient's ear canals.
12. Slide the ear rods gently to the patient's ears so that the tips hit the patient's external auditory canals.
60 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
LATERAL PROJECTION:
• The patient's Frankfort-Horizontal plane should be horizontal. Use the FH light as a guide.
PA PROJECTION:
• Turn the patient's head downwards until the angle between the patient’s Frankfort-Horizontal plane
and the horizontal light is around 30°.
• The patient's mouth should be opened fully and pressed against the neck/upper chest.
ORTHOPANTOMOGRAPH™ OP 3D LX 61
6 Using the device
PA WATERS PROJECTION:
• Turn the patient's head upwards until the angle between the patient’s Frankfort-Horizontal plane
and the horizontal plane is around 35-40°.
14. Turn the nasion support down and slide it against the patient's nasion when taking Lateral projections.
NOTICE! If the patient has wide shoulders that could collide with the rotating unit, ask the
patient to cross their arms in front of them in order to contract the shoulders.
15. Ask the patient to gently bite their molars together, if possible.
16. When taking a Lateral projection, ask the patient to close their lips.
62 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
NOTICE! Ensure that the device has been set up correctly for the required examination.
2. The device goes to ready state and the status indicator light turns green.
NOTICE! If the status indicator light does not turn green, look for more information on the GUI.
4. Move the nasion support to its out-most position and turn it aside, to the direction shown in the image.
You may also turn the ear rods aside for easier positioning.
5. Attach the carpus holder by sliding it to the base of the nasion support.
ORTHOPANTOMOGRAPH™ OP 3D LX 63
6 Using the device
NOTICE! Only the center area of the carpus holder is exposed. Ensure that the patient's hand
is positioned so that the imaged parts are on the center area of the carpus holder (white area in
the image above).
NOTICE! You may place the patient's hand diagonally on the carpus holder if it doesn't fit
properly otherwise.
NOTICE! To ensure optimal image quality, pay attention to the correct patient positioning.
64 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
NOTICE! Make sure that the device is set up correctly for the intended examination.
2. The device goes to ready state and the status indicator light turns green.
NOTICE! If the status indicator light does not turn green, look for more information on the GUI.
3. Select the patient positioning accessory according to the image to be taken, attach it to the chin rest
and attach them on the lower shelf of the device, as shown below.
NOTICE! It is possible to take 3D images with FOV 15 x 20 without chin rest if diagnostic need
requires it and reduced patient support is justified.
4. Place disposable covers on the patient positioning accessories.
ORTHOPANTOMOGRAPH™ OP 3D LX 65
6 Using the device
7. Tell the patient to remove their glasses, hearing aids, removable dentures, jewellery, hair clips and all
other things that can cause artifacts to the image.
8. Protect the patient from radiation according to the local regulations, for example with a lead apron or a
thyroid shield.
9. Guide the patient to the device and instruct to stand as straight and as tall as possible.
10. Ask the patient to grab the patient handles and place their chin on the Chin Rest and press their lip
against the Lip Support.
NOTICE! If the patient has wide shoulders that could collide with the rotating unit, ask the
patient to cross their arms when holding the handles in order to contract the shoulders.
NOTICE! If you are taking FOV 15 x 20 image without a chin rest, use the horizontal lights to
determine the correct height for the head support and patient head.
11. Ask the patient to take a step forward to straighten their spine.
66 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
12. Fine adjust the device height and adjust patient's head position/orientation using the patient positioning
lights as guides.
NOTICE! To ensure optimal image quality, pay attention to the correct patient positioning.
NOTICE! You can open the mirror and use it to help in positioning the patient. Close the mirror
before you start the imaging.
• The patient's occlusal plane should be horizontal and the ROI is between the top and bottom FOV
lights.
NOTICE! You may place a cotton roll or similar between the patient's front teeth to separate the
jaws.
13. Check that the patient’s head is straight and not rotated and the patient’s mid-line coincides with the
midsagittal light.
ORTHOPANTOMOGRAPH™ OP 3D LX 67
6 Using the device
NOTICE! If the patient is feeling insecure or has an exceptional anatomy, use the Test mode to
demonstrate the unit movements and to make sure that the rotating unit does not collide with the
patient during the imaging process. Activate the Test mode from the GUI or the control panel and
then press and hold the exposure button. The unit will complete the imaging movements without
generating X-rays.
NOTICE! This is an optional step but decreases the imaging process time and the time that the
patient must stay still.
NOTICE! The device stays in the IMAGING START position for 1 minute.
3. Ask the patient to stay still during the whole imaging process.
4. Protect yourself from radiation.
5. Press and hold the exposure button down to take the image. The device starts to move and plays an
audible exposure warning signal.
68 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
6. Release the exposure button after all device movements stop and the device plays an audible program
end signal.
7. A preview image appears on the GUI for a quick review (no visible artifacts etc.).
8. Acknowledge the preview by pressing the OK button.
11. Remove all disposable covers and decontaminate the device and the patient positioning accessories.
NOTICE! It's recommended to have Scout mode enabled by default. If you do not
wish to take a Scout image, deactivate the selection from the GUI.
1. Ensure the correct patient positioning, imaging program selection and that the device is in Ready state
with the indicator light turned green.
2. Ask the patient to stay still during the whole imaging process.
3. Protect yourself from radiation.
4. Press and hold the exposure button down.
5. The device starts to move and an audible exposure warning signal is played.
6. Hold the exposure button down until all movements have stopped.
NOTICE! The device also plays a program end tone when the imaging process is complete.
ORTHOPANTOMOGRAPH™ OP 3D LX 69
6 Using the device
10. Press HOME button. The device moves to IMAGING START position and locks the device in place to
prevent unwanted movement.
NOTICE! This is an optional step but decreases the imaging process time and the time that the
patient must stay still.
NOTICE! The device stays in the IMAGING START position for 1 minute.
11. Ask the patient to stay still until the 3D image has been taken.
12. Protect yourself from radiation.
13. Press and hold the exposure button down to take the image. The device starts to move and plays an
audible exposure warning signal.
14. Release the exposure button after all device movements stop and the device plays an audible program
end signal.
15. Image preview appears on the GUI for a quick review (no visible artifacts etc.).
16. Acknowledge the preview by pressing the OK button.
17. Continue to take the next image, if multiple images need to be taken.
18. Release the patient from the device.
19. Remove all disposable covers and decontaminate the device and the patient positioning accessories.
NOTICE! You can take a Scout image of the dental cast before the actual 3D image
by activating the Scout selection from the GUI.
70 ORTHOPANTOMOGRAPH™ OP 3D LX
6 Using the device
NOTICE! The device also plays a program end tone when the imaging process is complete.
8. A preview image appears on the GUI for a quick review (no visible artifacts etc.).
9. Acknowledge the preview by pressing the OK button.
ORTHOPANTOMOGRAPH™ OP 3D LX 71
7 Maintenance
7 Maintenance
The maintenance and calibration procedure intervals described here are minimum requirements and
recommendations. The maintenance and calibration procedures can be made more frequent and stringent
to comply with local regulations regarding the use and maintenance of dental X-ray devices.
NOTICE! Bite block and ear rods must always be used with new disposable covers.
NOTICE! Wear gloves and other protective equipment during the decontamination process.
CAUTION! Do not allow water or other cleaning liquids to enter the device's interior since it might
cause short-circuits or corrosion. If you use a spray cleaner do not spray into any ventilation grills.
CAUTION! Clean the dust off the device regularly. The device might overheat if excess dust is
gathered on the ventilation grills.
Device surfaces
All device surfaces can be wiped clean with a soft cloth dampened with a mild detergent, for example soapy
water. DO NOT use abrasive cleaning agents or polishes.
Control panel
Wait for the control panel to dry or wipe it dry before use. Moisture can affect the usability of the panel.
72 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
NOTICE! The device will automatically remind of the re-calibrations (2 times a year for User
calibrations and once a year for service calibrations by default). If you want to change the frequency
of the recalibration reminders, contact service for assistance.
ORTHOPANTOMOGRAPH™ OP 3D LX 73
7 Maintenance
3. A list of available device calibrations is shown with the status of the calibration.
NOTICE! The calibration programs listed on the GUI depend on the device configuration.
NOTICE! Performed calibrations are indicated with green colour and the completion date of the
calibration. All calibrations need to be successfully performed before using the device.
NOTICE! Calibration images are shown on the GUI as viewed from the sensor.
74 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
NOTICE! Make sure that PAN/3D tubehead front cover and sensor covers are installed. Calibrating
without the covers may have an effect on the image quality.
4. When the exposure warning stops and the program end tone is played, the program is complete.
5. The calibration image appears on the GUI.
ORTHOPANTOMOGRAPH™ OP 3D LX 75
7 Maintenance
NOTICE! Make sure that PAN/3D tubehead front cover and sensor covers are installed. Calibrating
without the covers may have an effect on the image quality.
4. When the exposure warning stops and the program end tone is played, the program is complete.
5. The calibration image appears on the GUI.
76 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
NOTICE! The list of available calibrations depend on the device configuration and country
specification.
Perform the calibrations in this order, the instructions are the same for all programs:
• 3D geometry calibration, 10x10 FOV LS
• 3D geometry calibration, 10x10 FOV H
• 3D geometry calibration, 12x15 FOV
• 3D geometry calibration, 15x20 FOV
ORTHOPANTOMOGRAPH™ OP 3D LX 77
7 Maintenance
6. When the exposure warning stops and the program end tone is played, the program is complete.
7. The calibration image appears on the GUI.
78 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
7.3.1 PAN QC
NOTICE! The PAN QC is a recommended procedure but some local regulations/authorities may
require it to be performed.
NOTICE! The PAN QC requires separately purchasable 2D Quality Control tools.
3. Attach the PAN QC phantom holder and 2D QC test phantom to the lower shelf.
4. Attach the copper filter in front of the radiation window on the PAN/3D tubehead. The filter attaches in
place with magnets.
ORTHOPANTOMOGRAPH™ OP 3D LX 79
7 Maintenance
6. Press and hold the exposure button down to take the QC image.
9. Workstation: Visually evaluate the image using the dental imaging software:
80 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
7.3.2 3D QC
1. GUI: Go to device settings.
3. Attach the 3D QC phantom holder and the 3D QC phantom to the lower shelf.
6. The QC image preview appears on the GUI, showing the result of the check.
ORTHOPANTOMOGRAPH™ OP 3D LX 81
7 Maintenance
8. Workstation: Visually evaluate and inspect the 3D image for visual defects such as artifacts using the
3D imaging software.
NOTICE! The device determines if the QC image is PASSED or FAILED according to measured
data, not based on the visible image quality.
NOTICE! You should also compare the new QC image to the reference image taken during the
installation or the latest service. Doing this helps you ensure that the image quality has remained
constant.
9. If the image fails or any visual defects are apparent, redo the QC program. If it fails again, recalibrate
the device according to Calibration procedure on page 75 or contact service.
10. Detach the 3D QC phantom and the 3D QC phantom holder from the device.
82 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
7.3.3 CEPH QC
NOTICE! This program is available only for devices with cephalometric (CEPH) modality.
NOTICE! The CEPH QC is a recommended procedure but some local regulations/authorities may
require it to be performed.
NOTICE! The CEPH QC requires separately purchasable 2D Quality Control tools.
ORTHOPANTOMOGRAPH™ OP 3D LX 83
7 Maintenance
7. If required by the local regulations, attach the copper filter in front of the radiation window on the CEPH
tubehead. The filter attaches in place with magnets.
12. Workstation: Visually evaluate the image using the dental imaging software:
84 ORTHOPANTOMOGRAPH™ OP 3D LX
7 Maintenance
14. Detach the CEPH QC phantom holder and 2D QC test phantom from the device.
15. Detach the copper filter from the CEPH tubehead if it was used.
ORTHOPANTOMOGRAPH™ OP 3D LX 85
8 Troubleshooting
8 Troubleshooting
No connection to the device. • The device is not powered on. 1. Power the device on.
• Problem with the local area 2. Check the Ethernet
network. cable connections on the
• Check the device workstation and the device.
configuration. 3. Restart the device and the
workstation.
Device shows error message Device has detected an issue Contact service. The device can
about disabled modalities. with one or more of its functions be used normally with the enabled
and has disabled their use. After functions in the meantime.
the next device boot the faulty
functions are removed from the
GUI.
Calibration and QC program date- The device receives the system 1. Restart the device and redo a
and timestamps are incorrect. time from the network it is calibration or QC program to
connected to during start-up. check if the device receives
If the time stamps in the the time correctly.
Calibration and Quality control 2. Check that the system time
menus are incorrect, the system is correct on the acquisition
time has not been received workstation, or any other
correctly. computer that may operate
as network time server in the
local network.
3. Check that the acquisition
workstation has UDP port
123 opened in the firewall
and the NetTime service
installed and running.
GUI shows a message "Detach Issue in CEPH head support Recalibrate the CEPH head
carpus holder" even though the calibration. support (service technician only).
carpus holder is not attached.
86 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
9 Technical data
Model: PCX–1
Protection against electric shock Class I
Degree of protection Type B applied with no conductive connection to the patient
Protection against the ingress of liquids IP20
Cleaning agents and protection against • Distilled water
cross contamination • Ethanol 96%
• Isopropyl alcohol
• Soapy water
• Cavicide™ disinfectant
• CaviWipes™ disinfectant
• Metasys™ disinfectant
• Disposable plastic covers for bite block, chin rest, lip
support
• Disposable plastic covers for ear rods and nasion support
(CEPH only)
ORTHOPANTOMOGRAPH™ OP 3D LX 87
9 Technical data
Tubehead assembly
Tubehead assembly THA/HVGEN PCX
Tube type D–054S or equivalent
Stationary anode
Tube voltage 60 – 95 kV
Max. tube current 16 mA
Target angle 5°
Focal spot 0.5 (IEC 60336/2020)
Total filtration min. 3.4 mm Al @ 95 kV
Additional filtration of cover materials max. 1.6 mm Al @ 95 kV
(CEPH only)
Leakage technique factors PAN: 4613 mAs/h @ 90 kV/12.5 mA
3D: 2880 mAs/h @ 95 kV/4 mA
CEPH: 4736 mAs/h @ 95 kV/11 mA
NOTICE! Leakage technique factors may be impacted by ambient temperature or used workstation.
Electrical connections
Nominal mains voltage 100 - 240 VAC
Tolerance: ± 10%
100–120 VAC:
Littelfuse 326 (SlowBlow) 15 A
Cooper Bussman (Time Delay) MDA–15
External warning light fuse (F3) Cooper Bussmann (Time Delay) S506–2–R 2 A
Power consumption 2.3 kVA @ 220–240 VAC, 1.65 kVA @ 100–120 VAC
Maximum impedance of mains 0.2 Ω
Mains cord EU 230V: 16 A / 250 VAC; 50/60 Hz H05VV–F3G1.5
US 115V: 15 A / 125 VAC 50/60 Hz SJT 3 x AWG 14
US 230V: 15 A / 250 VAC 50/60 Hz SJT 3 x AWG 14
Exposure controller cable Remote exposure switch: DINFLEX–YY 4 x AWG 26, max. 15
m
Data communication cable CAT6 UTP Ethernet cable
88 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
X–ray generator
Generator High frequency DC generator
Anode voltage 60 – 95 kV (± 5 kV)
Anode current 2 – 16 mA (± 20%)
Exposure time accuracy ± 5%
DAP display accuracy ± 25%
Spine compensation kV / mA compensated
Accuracy of radiation output > 95%
max. 240 mm
max. 147 mm centra
l beam
central beam
ORTHOPANTOMOGRAPH™ OP 3D LX 89
9 Technical data
NOTICE! The device package and packaging material weights may vary depending on the device
configuration.
Ambient temperatures
Transportation and storage -25 – +55°C
RH 0 – 90%
Atmospheric pressure 70 – 108 kPa
90 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
ORTHOPANTOMOGRAPH™ OP 3D LX 91
9 Technical data
95 kV 4.0 - 11.0 mA
9.2.3 3D programs
NOTICE! Available imaging programs and 3D FOV sizes depend on the device configuration and
country specification.
NOTICE! Voltage is fixed to 95 kV
10 x 10 Scout 0.04 s
12 x 15 Scout 0.07 s
15 x 20 Scout 0.07 s
92 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
FOV Resolution Voxel size Beam size at mA range Exposure Exposure Duty cycle
sensor time * pulse
2
5x5 Low 280 μm 65 cm 2.0 - 12.5 mA 0.9 s 5 ms 5:24
Standard 300 μm
High 200 μm
ORTHOPANTOMOGRAPH™ OP 3D LX 93
9 Technical data
Panoramic programs
Patient size
Cephalometric programs
Patient size
94 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
3D programs
Patient size
Standard
High 5 mA
Endo 6.3 mA
Standard
High 3.2 mA
Standard
High 3.2 mA
Standard
High 3.2 mA
Standard
High 3.2 mA
12 x 15 Low 4 mA 5 mA 8 mA
Standard 6.3 mA 7 mA 10 mA
High 8 mA
Standard 4 mA 6.3 mA 8 mA
High 6.3 mA 8 mA 10 mA
ORTHOPANTOMOGRAPH™ OP 3D LX 95
9 Technical data
400 ± 20 mm
15.7 ± 0.8 in
953 - 1718 mm / 37.5 - 67.6 in
320 ± 20 mm
12.6 ± 0.7 in
902 - 1667 (nominal 940) mm
35.5 - 65.6 (nominal 37) in
100 mm
3.9 in
A= 0.05m2
198 mm
7.8 in
NOTICE! The maximum height of the device is adjustable during the installation calibration.
96 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
773 mm / 30.4 in
1228 -1500 mm / 48.3 - 58.7 in
340 mm / 13.4 in
1690 - 2455 mm / 66.5 - 96.6 in
400 ± 20 mm
15.7 ± 0.8 in
953 - 1718 mm / 37.5 - 67.6 in
320 ± 20 mm
12.6 ± 0.8 in
902 - 1667 (nominal 940) mm
35.5 - 65.6 (nominal 37) in
A= 0.05m2
198 mm
7.8 in
* 500 ± 10 mm
* 19.7 - 0.3 in
ORTHOPANTOMOGRAPH™ OP 3D LX 97
9 Technical data
Manufacturer
Date of manufacture
Medical device
Serial number
Lot number
Caution
General warning
Radiation warning
Dangerous voltage
On or enabled
Off or disabled
Exposure switch
Ethernet
98 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
Focal spot
Do not reuse
Recyclable
This symbol indicates that the waste of electrical and electronic equipment
must not be disposed as unsorted municipal waste and must be
collected separately. Please contact an authorized representative of the
manufacturer for information concerning the decommissioning of your
equipment.
CE Mark
Marking for devices to be sold in European Economic Area with Notified
Body number
NRTL Mark
Conforms to U.S. and Canada national safety standards
Importer
ORTHOPANTOMOGRAPH™ OP 3D LX 99
9 Technical data
The device is suitable for use in both professional healthcare (hospitals/large clinics) facility environment and
home healthcare (clinics in domestic establishments and those directly connected to the public low-voltage
power supply) environment.
Exceptions for professional healthcare facility environment: Not to be used or installed near active HF
SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance imaging,
where the intensity of EM DISTURBANCES is high.
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of PCX-1 should assure
that it is used in an electromagnetic environment as described below:
Radio-Frequency Emissions Group 1 PCX-1 uses RF energy only for its internal function. Therefore,
CISPR11 the RF emission is very low and not likely to cause any
interference in nearby electronic equipment.
Radio-Frequency Emissions Class B PCX-1 is suitable for use in both at professional healthcare
CISPR11 (hospitals/large clinics) facility environment and home healthcare
(clinics in domestic establishments and those directly connected
Harmonic emissions IEC61000-3-2 to the public low-voltage power supply) environment.
IEC61000-3-2 Class A Exceptions for professional healthcare facility environment : Not
to be used or installed near active HF SURGICAL EQUIPMENT
Voltage fluctuations/ flicker Complies
and the RF shielded room of an ME SYSTEM for magnetic
emissions resonance imaging, where the intensity of EM DISTURBANCES is
IEC61000-3-3 high.
100 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of PCX-1 should assure
that it is used in an electromagnetic environment as described below:
Electrostatic discharge (ESD) ± 8 kV for contact discharge ± 8 kV for contact discharge Floors are wood, concrete,
IEC61000-4-2 or ceramic tile, or floors are
±2, 4, 8, 15 kV for air discharge ±2, 4, 8,15 kV for air discharge covered with synthetic material
and the relative humidity is at
least 30 percent.
Electrical fast transient/burst ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality is that of
IEC 61000-4-4 (100 kHz) a typical commercial and/or
±1 kV for input/output lines hospital environment
±1 kV for input/output lines
(100kHz)
Voltage dips, short 0 % UT; 0,5 cycle 0 % UT; 0,5 cycle Mains power quality is that
interruptions and voltage of a typical commercial and/
At 0°, 45°, 90°, 135°, 180°, At 0°, 45°, 90°, 135°, 180°, or hospital environment. If
variations on power supply
225°, 270° and 315° 225°, 270° and 315° the user of PCX-1 requires
input lines
continued operation during
IEC61000-4-11 0 % UT; 1 cycle 0 % UT; 1 cycle power mains interruptions,
At 0° At 0° it is recommended that
PCX-1 be powered from an
70 % UT; 25/30 cycles 70 % UT; 25/30 cycles uninterruptible power supply.
At 0° At 0°
NOTICE! UT is the a.c. mains voltage prior to application of the test level.
ORTHOPANTOMOGRAPH™ OP 3D LX 101
9 Technical data
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of PCX-1 should assure
that it is used in an electromagnetic environment as described below:
Immunity Test IEC 60601-1-2 Test Level Compliance Level Electromagnetic Environment
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be performed. If the measured field strength exceeds the RF compliance level
above, observe PCX-1 to verify normal operation in each use location. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating PCX-1.
Guidance for actions taken can be found from AAMI TIR 18:2010, Guidance on electromagnetic compatibility of medical devices in
healthcare facilities.
NOTICE! Precautions to take if the use location is near (e.g. less than 1,5 km from) AM, FM or TV broadcast antennas.
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m. The Recommended Separation Distances
are listed in the next table.
NOTICE! These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
102 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
Test specifications for enclosure port immunity to RF wireless communications equipment IEC
60601-1-2 ed4.1 table 9
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of PCX-1 should assure
that it is used in an electromagnetic environment as described below:
745 217 Hz
780
5500 217 Hz
5785
If necessary to achieve the immunity test level, the distance between the transmitting antenna and the ME equipment or ME system
may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
* For some services, only the uplink frequencies are included.
** The carrier shall be modulated using a 50% duty cycle square wave signal.
*** As an alternative to FM modulation, the carrier may be pulse modulated using a 50% duty cycle wave signal at 18 Hz. While it
does not represent actual modulation, it would be worst case.
ORTHOPANTOMOGRAPH™ OP 3D LX 103
9 Technical data
Separation distances
Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC 60601-1-2
The PCX-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user
of PCX-1 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the PCX-1 as recommended below, according to the maximum output power of the
communications equipment.
Frequency of Transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.7 GHz
Equation
Rated Maximum Output Power Separation Distance (meters) Separation Distance (meters) Separation Distance(meters)
of Transmitter (watts)
1 2 0,6 0,6
100 20 6 6
For transmitters rated at maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
WARNING! *** Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12") to any part of the device, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result. See Separation distances on
page 104.
NOTICE! At 80 MHz, the separation distance for the higher frequency range applies.
NOTICE! These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
USE LIMITATION:
External components
WARNING! Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this device and result in improper operation.
104 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
ORTHOPANTOMOGRAPH™ OP 3D LX 105
9 Technical data
106 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
ORTHOPANTOMOGRAPH™ OP 3D LX 107
9 Technical data
108 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
ORTHOPANTOMOGRAPH™ OP 3D LX 109
9 Technical data
Device Driver DTX Studio Driver 23.1 or newer (latest version is always preferred)
Network • Supported interface chipsets: Intel i210, i219, Intel i225 or Broadcom
• Gigabit Ethernet 1000Base-T
• The device must be connected to a private, firewall protected local area
network to ensure proper data security. All connections from outside of the
local area network to the device must be blocked. Connections between
the device and the workstation in the local area network must be allowed.
110 ORTHOPANTOMOGRAPH™ OP 3D LX
9 Technical data
ORTHOPANTOMOGRAPH™ OP 3D LX 111
PaloDEx Group Oy reserves the right to make changes in specification and features shown herein, or discontinue the product
described at any time without notice or obligation.
Contact your PaloDEx Group Oy representative for the most current information.
Headquarters
PaloDEx Group Oy
Nahkelantie 160 | FI-04300 Tuusula | FINLAND
Tel. +358 10 270 2000
https://dexis.com/contact-us