MPTEE
MPTEE
User Manual
3 Transducer Use.....................................................................9
6 Specifications .....................................................................18
User Manual
1 General Information
1
User Manual
WARNING
The transducer should be used only by a qualified
physician who has received appropriate training in
proper operation of the transducer and in endoscopic
techniques as dictated by current relevant medical
practices.
WARNING
The transducer is not intended for fetal imaging.
2
User Manual
3
User Manual
4
User Manual
Electrical Hazard
5
User Manual
6
User Manual
CAUTION
Ensure that the sheath’s sterile surface is maintained.
7
User Manual
WARNING
CAUTION
Only use water-soluble acoustic coupling gel. Other
coupling gels containing ingredients like ethanol,
mineral oil, Iodine, lotions, lanolin, aloe vera, methyl
alcohol or ethyl parabenzoic acid can cause
transducer damage.
8
User Manual
3 Transducer Use
CAUTION
9
User Manual
10
User Manual
WARNING
Check if the maximum deflection of the tip is 90°to 120°
upward and 60° to 90° downwards. If the up/down
deflection shows an unwanted amount of free play or
exceeds the maximal deflection angles given above, do
not use the transducer. Contact the service organization
to re-adjust the steering of the transducer. In this way, the
risk of "buckling" or "U-turning" of the transducer in the
esophagus is minimized.
Unlock
Red
Lock
11
User Manual
CAUTION
12
User Manual
3.3 Examination
The actual examination with the transducer is beyond the
scope of this manual. There are many medical articles and
books, which very thoroughly address this topic. There are
however specific cautions that should be considered.
CAUTION
Long-term exposure to ultrasound should be minimized.
Although there have been no confirmed adverse effects
produced by diagnostic levels of ultrasound, unnecessary
patient exposure to ultrasound energy should be avoided,
especially in the Doppler mode.
Prolonged pressure on the esophagus by the tip of the
transducer may lead to a Pressure Necrosis phenomenon.
Thus in monitoring applications the tip should be removed
from the esophagus wall when you are not scanning by
releasing it in the neutral position. If continuous monitoring
is required the transducer tip should often be re-positioned.
Whenever active scanning is not desired ‘freeze’ the image
and release the deflection control.
Position the array such, that good acoustical contact is
made. Before rotating the array, actuate the ratchet ('brake')
on the up/down movement. If this is not done, rotating the
array can mean loss of acoustical contact!
13
User Manual
Biological Hazard
Adequate cleaning and, if necessary disinfection are
carried out to Prevent disease transmission. It is the
responsibility of the user to verify and maintain the
effectiveness of the procedure used. A single-use,
sterile disposable sheath can be used.
CAUTION
Keep the control handle and system connector out of any
cleaning or disinfection solutions. The control handle and cable
may be cleaned with a damp cloth, but only the distal end of the
transducer up to the 100cm marker on the shaft (70cm for
SonoScape’s mini TEE transducers) may be placed into a
disinfection solution.
Daily
Weekly
Monthly
Other
After each use
14
User Manual
Transducer Cleaning
CAUTION
Do not clean any portion of the transducer with methanol or
ethanol. Such substance can cause irreparable damage to the
transducer.
d) Air-dry or dry with a soft cloth.
15
User Manual
Transducer Disinfection
CAUTION
The transducer should never be left in the disinfection
solution for more than 1hour.
Do not forget to rinse the transducer directly after
disinfection.
WARNING
16
User Manual
5 Transducer Storage
The transducer is a sensitive and expensive instrument.
Use great care when storing the transducer. Make sure
the transducer is adequately cleaned and disinfected prior
to storage.
WARNING
17
User Manual
6 Specifications
The transducer is in conformity with the relevant
provisions of the European Medical Device Directive
93/42/EEC. The transducer is categorized in Class IIa.
The transducer is classified according to IEC 601-1 as
class I, type BF, defibrillator-proof.
Transducer Type
MPTEE, MPTEE mini
Mechanical Dimensions
Transducer
18
User Manual
19