370a550_rigel_bp_sim_manual_v2
370a550_rigel_bp_sim_manual_v2
Instruction Manual
370A550 Issue 2.0
July 2009
© 2009 Seaward Group
Issue 2.0
Limited Warranty & Limitation of Liability
Rigel Medical, part of the SEAWARD GROUP guarantees this product for a period of 1
year. The period of warranty will be effective at the day of delivery.
Calibration Statement
The Rigel BP-SIM NIBP simulator is fully calibrated and found to be within the specified
performance and accuracy at the time of production. The Seaward Group provides its
products through a variety of channels, therefore it may be possible that the calibration
date on the provided certificate may not represent the actual date of first use.
Experience has indicated that the calibration of this instrument in not effected by
storage prior to receipt by the user. We therefore recommend that the recalibration
period be based on a 12 month interval from the first date the unit is placed in to
service.
© Copyright 2009
All rights reserved. Nothing from this edition may be multiplied, or made public in any
form or manner, either electronically, mechanically, by photocopying, recording, or in
any manner, without prior written consent from the SEAWARD GROUP. This also
applies to accompanying drawings and diagrams.
Due to a policy of continuous development the SEAWARD GROUP reserves the right to
alter the equipment specification and description outlined in this publication without prior
notice and no part of this publication shall be deemed to be part of any contract for the
equipment unless specifically referred to as an inclusion within such contract.
Index
1. INTRODUCTION.......................................................................................... 7
1.1. Rigel BP-SIM Design Philosophy ............................................................................... 7
1.2. Key Features .............................................................................................................. 8
1.3. Interfaces .................................................................................................................... 9
............................................................................................................................................... 9
1.4. Rigel BP-SIM includes:............................................................................................. 10
1.5. Optional Accessories ................................................................................................ 10
1.6. Test ’n Tag Compatible ............................................................................................ 10
1.7. Unique use of ICONS ............................................................................................... 11
2. LOGGING ON ........................................................................................- 12 -
8. ABOUT...................................................................................................- 64 -
10 TECHNICAL SPECIFICATIONS............................................................- 68 -
Manufactured by:
Millennium Statement
This product is Millennium compliant, and conforms fully to the document BSI DISC
PD2000-1.
Statement of Conformity
Based on test results using appropriate standards, the product conforms with
Electromagnetic Compatibility Directive 89/336/EEC and Low Voltage Directive
73/23/EEC
Standards used:
WEEE Directive
The Rigel BP-SIM has been designed and manufactured with
high quality materials and components, which can be recycled
and reused.
The following symbols are used in these operating instructions and on the Rigel BP-
SIM.
If the BP-SIM is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
The BP-SIM is supplied with a mains power supply unit. The use of any other power
supply unit may cause damage to the equipment and may result in a hazard to the user.
The new simulator incorporates a range of custom settings that include a variety of
paediatric and adult NIBP pressures, pulse volume adjustments, heart rate and
manufacturer-specific O-curves. The Rigel BP-SIM is the first hand-held NIBP simulator
combining full performance features with a large capacity internal memory for the data
capture, storage and subsequent downloading of test results for record keeping
purposes.
Other special features of the Rigel BP-SIM include an in-built pump to generate user
specific pressure settings for comprehensive system leak tests and over pressure tests.
An integral digital manometer also enables highly accurate static pressure
measurements to be taken using a wide range of selectable pressure units.
The new Rigel BP-SIM forms part of a comprehensive range of high performance
specialist biomedical test equipment supplied by Rigel Medical, part of the Seaward
Group.
The challenge was to combine the benefits of the size and weight of a smaller hand-
held NIBP Simulator with the test versatility and convenience of a larger bench-mounted
simulator. Today’s industry demands test equipment that can save time and cost, thus
offering greater flexibility. These factors have all been taken into account during the
development of the BP-SIM.
The test capabilities and functionality exceed that of most common NIBP Simulator yet
the instrument is only a fraction of the size and weight.
No other NIBP simulator on the market offers a hand-held enclosure with the test
capabilities equal to most bench marked version interface, large internal memory, Blue
Tooth communication, asset management facilities, user configurable Performance
Tests and more.
We believe that the new BP-SIM is set to become the new standard in NIBP
performance Testing.
• Portable and Lightweight - Ideal for field service technicians and in-situ
testing of NIBP performance
• Battery Operated - For convenient and fast testing without mains power
• Web Updates - Enjoy easy updates through our website to benefit from
the latest features
4
5
3
2
6 8
1) Full graphic Monochrone LCD, blue with white backlit, 1/4 VGA
2) Larger Red key Stop/Off Key
3) Larger Green Start/On key
4) 4 Programmable soft keys below display
5) Alpha Numeric keyboard, Up / Down and Left / Right cursor control.
6) Power Supply inlet
7) Pneumatic interface connector
8) Input/Output connector (future use only)
• Battery charger
• Carry case
• 5-pc tubing adaptor set
• Instruction manual
• Application disc
• BlueTooth USB adaptor
• Calibration certificate
The Rigel BP-SIM is compatible with the Seaward Test ‘n Tag and Elite system. Our
Test ‘n Tag systems allows you to print customised Thermal PASS / FAIL labels.
The unique Test ‘n’ Tag label provides test status and retest due date, barcode and
asset ID number, and person conducting the test. For more information on using the
Test ‘n Tag printer, see section 4.
The Rigel BP-SIM features a hi resolution graphic back lit display which not only
provides highly visible and easy to follow menu structures but also allows the user to
operate the tester using intuitive icons to speed up their test routines.
Escape OK – PASS
Print
Add
Less
Copy
Repeat
Delete
Save
Edit
Search
Help
Settings
Menu – List
Home Shift
New Mute
Sound
When switching on the Rigel BP-SIM, the user will be able to log-in to allow for specific
default user preferences and to provide traceability of test results. To increase the
security and protection of the user, a password can be set prior to using the Rigel BP-
SIM. See 3.5.3 for further details. The default password setting is OFF and the login
screen will not require a password. In case a user and password are set, the tester will
default to the last user / password.
To change the user, use the drop down menu provided [<>] and select the new user. In
case a password is set for the new user, please enter the password (case sensitive).
In case a password is lost and the unit is blocked, please call our helpdesk on:
+44 (0) 191 5878701 or contact your local distributor and quote the 8-digit code (shown
below the password field) and the serial number of the tester. A new password will then
be created. For security reasons, this new password can only be provided to the email
address of the original purchaser of the device.
The Rigel BP-SIM is designed to allow a user to customise test sequences and allow
default settings. This customisation can speed up testing by providing the user with
default manufacturer lists, model numbers, user text protocols, automatic printing post-
text and fault menus.
All custom facilities can be found under the SETUP menu. Simply press the [F4]
from the home screen and select Setup from the menu.
Hint: The underlined character acts as a short key to allow swift navigation through the
menu structure.
Options available
The Rigel BP-SIM can be set-up to create new test sequences to meet monitor specific
requirements or to modify existing test sequences to meet personal preference.
Preset test sequences cannot be changed by the user, however alterations are possible
by producing a copy of the default test sequence setting.
The total number of possible preset test sequences is 50 including 12 default test
sequences.
This feature gives the Rigel BP-SIM the capability of recording user defined visual
inspections, checks or tests including measurements from SPO2, ECG, NIBP, Defib,
Infusion, Ventilation, Pressure etc.
The input is text only as no measurements are performed by the Rigel BP-SIM during
these tests. The user can enter questions or instructions followed by either a
PASS/FAIL result or alpha numeric input. Preset engineering units e.g. %, Joules,
mmHg, PSI, CmH2O etc are available.
Preset test sequences cannot be deleted or changed by the user however, alterations
are possible by producing a copy of the default test sequence.
To enter the Test Sequences menu, click on , followed by Setup. Select the Test
The following overview is provided and shows the list of default and customised test
sequences availble. Copied Test Sequences will appear in the list with an *pre-fix.
From this menu, the user is able to View a test sequence by pressing the View button
(F4). Default test settings cannot be altered at any stage.
keys to highlight the test sequence and press the button (F2).
Use the Up & Down Arrow keys to select the required action and press the
(F4) to confirm. Press Escape (F3) to cancel and return to previous screen.
Copied Test Sequences will appear in the list with an *pre-fix. All Copied test
Sequences can be edited. See 3.1.3 for more on editing test sequences.
Print Test Sequence will provide an overview of the Test Sequence on the Favourite
Blue Tooth Printer. See 3.4 for help on setting up Blue Tooth devices.
Delete Sequence will remove the highlighted test sequence from the memory of the
Rigel BP-SIM.
Preset test sequences cannot be modified by the user, however alterations are possible
by producing a copy of the default test setting (see 3.1.2).
Non-default Test Sequences can be modified from the Test Sequences menu. When
non-default Test Sequences are highlighted, the (F4) will appear in the menu
screen.
To change the Name and or Class of the Test Sequence, press the Button (F4)
and enter the Name and or Class required.
To change to upper or lower case or use language specific characters, press the
button (F1) and select the required entry.
To return to the previous menu without changes, press the button (F3)
To Insert a new test, highlight the test using the Up & Down arrow keys and press the
Insert button (F1)
To Delete an individual test, highlight the test using the Up & Down arrow keys and
To go back to the previous menu, highlight the test using the Up & Down arrow keys
To modify an individual test, highlight the test using the Up & Down arrow keys and
Press the button (F4) as shown above. The Leak Test sub menu will allow you
to alter the settings for; Test duration and Pressure set point.
keypad to enter the correct data. To Delete the field content, press the button as
shown below.
When the right settings have been achieved, press the button (F4) to save the
Repeat this action for every test that requires editing. Once all the requires tests are
To insert an individual test we use the up and down keys to highlight the position where
a new test needs inserting and press the Insert button (F1) from the menu below.
Note; Tests will be inserted before the highlighted position not after.
Once the desired test has been highlighted, confirm using the OK button (F4) and finally
confirm to insert into the test Sequence. Pressing the button (F3) will return to
the previous screen without changes being made.
The individual test has now been inserted and can be edited as described in 3.1.4
Further tests can be inserted or deleted by following the steps described herein.
Custom tests or Instructions are intended to allow the user to capture additional
information either prior to or after the main NIBP performance tests. Such information
could indicate the condition such as Labels, software version, certain damage or
upgrades. To create customised inspections, use the insert Custom Test function using
the instructions below and set the engineering units to blank.
Hint: Create a unique range of visual tests or instructions by creating a new test
sequence (see 3.1.6) and select Custom Test as the nature of the test. This will
create a unique test sequence.
To insert a Custom Test, highlight the place where a new test needs inserting using the
up and down keys and press the Insert button (F1) from the menu below.
Note; Tests will be inserted before the highlighted position not after.
Once the Insert button (F1) has been pressed, a drop down box will appear with all
available tests and inspection available. Activate the drop down box using the left key
and select the Custom Test option.
Confirm using the button (F4). The following menu will define the nature of the
test (eg test Instruction prior to leak test)
To insert the new Custom Test, press the button (F4) or exit using the
button (F3) and return to the previous screen without changes being made.
Further tests can be inserted or deleted by following the steps described herein.
To exit and save the changes for future press the button followed by pressing the
button. If the Escape button (F3) is pressed, the Rigel BP-SIM will return to the
previous meny without changes being made.
To create a new Test Sequence, enter the Menu button (F4) from the home screen,
select Setup followed by Test Sequences from the menu.
The following screen displays the list of default and customised tests available. Copied
Test Sequences will appear in the list with an *pre-fix.
From this menu, the user is able to create a new Test Sequence by pressing the
button (F1).
To change to upper or lower case or use language specific charaters, press the
button (F1) and select the required entry.
To return to the previous menu without changes, press the button (F3)
Next step is to save the name of the newly create Test Name, by pressing the OK
button (F4). Pressing the key (F3) would return to the Test Sequence menu
without changes.
The newly created Test Name now appears in the Test Sequence menu and can be
modified to include the required individual tests. Press the button (F4) to open
the sequence for editing.
Note; Tests will be inserted before the highlighted position not after.
The Asset Trace Variables allow the user to include valuable data to the test results to
enhance the traceability and to provide increased search criteria when using database
software. Asset trace variables may consist of a maximum of 25 characters and can be
selected prior to each test when using the Rigel BP-SIM in automatic mode.
The default settings on the Rigel BP-SIM include the Site and Location variables. Both
are activated. The drop down box is set to [Yes].
To activate or de-activate each variable, use the up and down keys to select the
variable and the left and right keys to select yes or no.
Once a variable has been activated, the user is given the ability to enter Trace
Variables prior to a safety test by selecting from or adding to the list of default items.
Trace Variables that are de-activated (set to NO) will not appear during the automatic
tests.
Note that a maximum of 40 individual entries are available for each Trace Variable.
Additional entries will overwrite on a first in first out basis.
This feature can setup the Rigel BP-SIM to automatically perform certain features
during or after a test as well as allowing the user to set Time and Date preference.
To select System Config, press the button (F4) followed by Setup in the menu and
System Config from the list as shown below;
3.3.1. Asset ID
Provides automatic Asset ID configuration. Use the Left & Right Keys to select between
Provides automatic action after a test is completed. Use the left and right keys to select
from the following
3.3.3. Comments
Provides the ability to enter further comments after successful or failed tests. Use the
left and right keys to select from the following
3.3.4. Language
Provides the ability to change the default language in the Rigel BP-SIM. A number of
languages are available and will be updated which ach firmware revision. Use the up,
down, left and right keys to select from the available languages. Additional languages
(firmware upgrades) can be downloaded from www.rigelmedical.com or contact your
local Rigel sales office (UK ; +44 (0) 191 5878701)
Allows the user to set the current date / time and preferred formats.
Press the Date / Time button (F2). The following menu will be displayed;
Ones all System Configurations have been completed, press the button (F4).
Changes are saved automatically.
In order for the Rigel BP-SIM to automatically connect to the correct Bluetooth
accessory, details of that accessory must be entered into the Bluetooth Favourites list.
The Bluetooth Favourites list comprises of three sub-lists (device types), each of which
can store up to 3 accessories of the same type. The three sub-lists are shown below:
• Barcode (Scanner)
• Printer
• Computer
To select Bluetooth Favourites, press the button (F4) followed by Setup in the
Menu and Bluetooth Favourites from the list as shown below. Press (F4) to
select;
Use the up/down arrow keys to highlight the device type (in this case Computer) then
press (F3).
Make sure that the device you want to add to your new favourites is switched on.
Press (F1). The Rigel BP-SIM will search for all discoverable Bluetooth devices
within a 10m range. The time taken to perform the search will vary depending on the
number of Bluetooth devices in range. A progress bar will show during the search.
When the search is completed, the progress bar will disappear.
Use the arrow keys to highlight the field labelled “New” and press the left or right keys
to open the list of devices which have been found during the search.
In the following example, the search has identified Bluetooth devices named
“SELLT023”, “SELPC045” and “Elite 36-5011”
Note: If the required device is not shown, check that it is powered on with the
Bluetooth function active then repeat the search.
Note: Some Bluetooth devices do not report a recognisable name and may cause
confusion when setting up Favourites. Turn off or disable unused Bluetooth
devices when configuring Favourites.
Use the up/down keys to highlight the required device (in this case SELLT023i) and
press the button (F4). SELLT023 is now placed in the “New” field as shown
below.
To add further devices to the list of available computers select the new field again, use
the up, down, left and right keys to select the required device and press (F3) to
add the chosen device.
“SELLT023” has been added to the Computer field with Bluetooth ID.
Press (F4) to store the changes and return to the previous screen.
Note: Seaward and Rigel accessories do not require a PIN. For other Bluetooth
devices refer to the manufacturer’s documentation.
Note: Devices that are not required can easily be removed by selecting the
Computer field, pressing the left of right keys to open the list, highlighting a
device using the up/down keys and pressing (F2). Confirm the
All Bluetooth devices have a unique ID. The Bluetooth ID is shown when a
device is added to the Favourites list.
Ensure when selecting the printer, that the Printer type is selected corresponding with
the print media:
Press (F4) to store the changes and return to the previous screen.
When all your favourite devices have been set up, press (F4) and they will be
saved.
The User Admin feature allows the user to personalise the way in which the tester
behaves during normal use and include features such as:
To enter the User Admin feature, press the button (F1) from the home screen,
select Setup followed be User Admin.
3.5.1. Preferences
Provides means of configuring the default settings of the tester’s behaviour during
normal use. All settings are user specific and stored against the logged-in user :
• Set Contrast – Use the up, down, left and right keys to set the contrast
• Auto Off Time (mins) - Use the up, down, left and right keys to set the
power off time on the tester. Settings between 0- 10 minutes. Note 0 minutes will
deactivate the AutoSwitch Off feature.
• Data Entry – Sets the tester up to take data entry from Barcode or Keypad only.
Use the up, down, left and right keys to select Barcode or Keypad only.
Note: In data entry mode the unit will automatically try to connect to a barcode
scanner via the internal Bluetooth module – if the scanner is available this feature
will drain the batteries u-necessarily. In this instance it is recommended that the
unit be set to Keypad only.
• Backlight Mode – Chose between Always Off, Always On, Power Save (Back
light on for 30 secs after each key press). Use the up, down, left and right keys
to select.
The (F2) button allows the user to set the audible warnings and beep on
key presses. See menu below:
Navigate using the Up & Down Keys and change between Yes / No using the
left key.
This feature allows an operator to default to a different existing user or setup a new
user. Admin rights are required for this feature. Alternatively, new users can be created
in the User Profile menu, see below.
Note A brand new Rigel BP-SIM will automatically default to admin rights so new
users can be created.
To create a new password, highlight the New Password field using the Up & Down Keys
and enter a new password followed by confirmation of the new password. Use the
key (F4) to finish and press the button (F4) to confirm and save the
new password.
To change an existing password, enter current password and repeat the steps above.
See menu below for an overview of the Change Password screen.
This feature allows the administrator to setup new and existing users and inhibit or
assign certain feautures to individual users.
Note that the User Profile is only valid when the ADMIN user has been setup with
a PASSWORD. See item above. If no ADMIN password has been set, all users
have FULL User Rights.
Select the applicable user name, use the Up & Down Keys to navigate through the
menu, use the Left & Right Keys to activiate (yes) or de-activate (no) certain features.
See menu below.
This feature is used to view information about the Rigel BP-SIM memory status. Press
(F4), select setup using the up/down keys and select Memory Options by pressing
The nature of Flash memory is such that when records are deleted only the reference to
the data is removed. The data still remains and will use memory space. The memory
must be erased in order to release memory used by deleted records.
If the Erase (F2) key is pressed, a prompt is shown below. To erase the memory press
button (F4).
The factory settings can be restored at any time using this function. Press button
(F4). Select Setup followed by Restore Factory Settings.
All the settings provided within the Setup menu will default to Factory Settings including;
All Asset Trace Variables, Test Sequences, User Admin, Test Codes and System
Config.
Warning: Restore Factory Settings cannot be undone and will remove all the
customised items described above.
Hint; All customised settings within Setup can be cloned to a PC. We advise that this is
done on a regular basis to ensure custom settings are saved for back-up. See 7.4 for
further details on Cloning.
The Automatic mode provides the user with the ability to run a preset test sequence and
store the data afterwards.
Press (F4), select Auto Mode using the up/down keys, Select by pressing the
Asset ID
Each test record is stored using the Asset ID number (25 character max) and date /
time of test. Multiple entries can be applied using the same Asset ID number. In this
case, the memory will hold all entries and differentiate using the test date.
Test Sequence
The Rigel BP-SIM comes standard with preset test routines to meet the requirements
of:
• Default Adult Sequence
• Default Pediatric Sequence
In addition, the user can create a total of 50 preset test sequences to meet specific
requirements. (See 3.1 more information on creating new test sequences)
• Semi-Automatic allows the user the enter data after each individual test that
forms a test sequence.
• Full Automatic will run through the test sequence and provides an opportunity to
enter data at the end of the complete sequence.
Test Period
The re-test period allows the user to include the next test date on printed labels and PC
downloads that make it easier to schedule future work. More details on printing labels
see 4.2)
Trace Variables
To increase the traceability of the safety test, the user can include valuable data with
the test record such as Re-test period and Asset variables (see 3.2 for more
information). Trace variables can be added once the above data has been entered and
confirmed (See next chapters for more details
The automatic mode provides an option to test all available tests (NIBP simulations,
Leak Test, Overpressure test) in an automatic sequence.
When a specific test fails, the user will be provided with a number of options depending
on the nature of the fault. The Failure Menu will enable the user to:
The following procedures describe the setup and completion of an automatic test
sequence. The steps are applicable to all available test sequences. As example we
describe a default test sequence.
Select the right criteria for the Test from the Asset Details screen;
5. Returning to the Asset Details screen, the button (F2) provides the user
with the option to modify or create a new Test Sequence from this menu. This
feature is identical to that of the Test Sequence feature in the Setup - Test
Sequence menu. (See 3.1 for more information)
When all parameters are set, press the button (4) to enter the Asset Trace
Variables screen (when applicable). In this example, all Asset Variables are activated.
button (F4). Each variable entered during the automatic test will
automatically be added to the list for future use. Trace Variables that are de-
activated (set to NO in the Setup menu, section 3.2) will not appear during the
automatic tests.
Note that a maximum of 40 individual entries are available for each Trace
Variable. Additional entries will overwrite on a first in first out basis.
When all information is provided, press the button (F4) to start the safety test.
The first test of the Default test setting is the the NIBP Leak Test
Following the Overpressure valve test, the BP-SIM is ready to make NIBP simulations
as per the test routine;
Pressing the (STOP) button (F3) during any test will bring up the failure menu
(see 4.1.1)
At the end of the test, the user has the ability to enter comments. See 3.3.3 for more
information on how to activiate comments after a test.
Depending on the settings in Menu - Setup - System Config (see 3.3.2) , a number of
After Test options are possible;
• Download – Automatically down loads the test results to the PC (see 7.1)
No further options or viewing of results available. The tester will return to the
home screen.
• Options Menu – Provides a menu to allow further choices after a test. (see 4.3)
The Rigel BP-SIM is compatible with the Seaward Test ‘n Tag and Test ‘n tag Elite
system To use the Test ‘n Tag printer, add your T’nT printer to the BlueTooth Favorites.
See 3.4
A detailed application note is available from the Rigel website on the used of Test’ n
Tag printers with the Rigel BP-SIM.
When Options has been selected in the Setup - System Config menu, the Rigel BP-SIM
will show a Test Details screen, see below.
To view the options menu, select the button (F4) from the test details screen;
The options menu provides the following sub menu:
View Results will display the results of the test (see section 4.4)
User Comment will allow the user to enter comments if required
Print Label will print a pass/fail label on the thermal printer (50 mm roll)
Print Result will print the result on the thermal printer (50 mm roll)
Press the View result Button to review the data. Use the Up and Down keys to scroll
through the screen.
The button (F4) provides extra option to print results, label or Test ‘n tag
• NIBP tests
• Leak tests
• Static pressure calibration
• Pop-off valve test
• Manometer function
Navigate using the Up & Down Keys and change each setting by using the left key.
Systolic / Diastolic; Select a preset range of NIBP setting from the drop down
list
Note that the settings Low / Medium / High are not directly linked to any
physiological condition
Once the selections are complete, select the (F4), to confirm the
settings;
Note when the valve is not closed, the pressure within the cuff will not be
contained and the monitor might go into alarm.
When the simulation is present (ie the internal BP-SIM valve is closed), a “pulse”
symbol will be shown;
In the NIBP leak test, the user is able to test the pneumatic leakage within the
system.
• Start Pressure
• Timer (seconds)
Note: Prior to initiating the test, some NIBP monitors might have to be set
to a service mode to ensure that the internal monitor valves are shut for
incoming pressures. If the monitor’s internal valve is open, the leak test
will not be able to pressurise the system.
The Pop-off test allows the user to verify the correct operation of the protection
valve within the NIBP monitor.
Note: Prior to initiating the test, some NIBP monitors might have to be set
to a service mode to ensure that the internal monitor valves are shut for
incoming pressures. If the monitor’s internal valve is open, the leak test
will not be able to pressurise the system.
To initiate the test, press START (F4). The BP-SIM will inflate the system until
the protection valve is activated;
At the end of the test, the BP-SIM will show the set-point of the protection valve
(ie 331 mmHg)
The BP-SIM has a built-in manometer function which allows the user to measure
the pressure in a number of engineering units;
Select the engineering unit from the drop-down box using the left and right keys:
The manometer will automatically convert the pressure to the desired pressure
unit.
The Set Zero function is required to ensure the manometer is pressurised to the
ambient pressure.
To view the stored data, press the button (F4) and select View Data from the
menu;
The following screen provides an overview of all available records in the memory,
identified by Asset ID, Site and Pass or Fail.
Use the up and down arrow keys to scroll through the database. When the desired
record is highlighted, press the button (F4) to enter the record and view the
data.
Use the button (F2) to return to the main View Data screen.
In the Test Details screen, press the button (F4) to view the following options:
− View Results will display the results of the test (see section 6.1)
− View Sequence will display the test Sequence
− User Comment will allow the user to enter comments if required
Use the up and down keys to select the view results option. Press the button
(F4) to display the view text results screen (see below).
Press the button (F4) to provide the options to print the results;
From the main View Data screen, press the button (F1).
Use the up, down, left and right keys to navigate this screen.
The default setting is ‘*’ which includes all records that Passed.
To search for more specific criteria, fill in the required fields by either using the
keyboard or drop down boxes.
For Date entry use format DDMMYYYY or DD MM YYYY. For example for
31st October 2008 use 31102008.
Test Status Allows the user to select search criteria based on PASSED, FAILED or
UPLOADED items.
To start the search, press the button (F4) or the button (F3) to leave this
screen.
From the main View Data screen, press the button (F2).
• Print All Labels – Will download all records within the search criteria to the Test
‘n Tag printer, setup in Bluetooth favourites.
• Download to PC – Will download all records within the search criteria to the PC,
setup in Bluetooth favourites.
• Delete Single Asset – Will delete the single selected record.
• Delete All Assets – Will delete all records from the memory (this action is
irreversible).
The Data Transfer functions are accessed by pressing the button (F4), selecting
Data Transfer using the up and down keys and pressing the button (F4).
7.1. Download to PC
This feature is used to download records to a PC. Select Download to PC and press the
button (F4) to accept. The Rigel BP-SIM will now attempt to make a Bluetooth
connection with the Computer that has been configured in the Bluetooth Favourites.
Initially, the Bluetooth status icon will display the egg timer for a few seconds, as shown
on the left, while the connection is established. When the connection is established and
ready to communicate the Bluetooth status icon will change to show the Bluetooth icon
as shown on the right.
Toggle between the options by using the left and right arrow keys.
The Rigel BP-SIM is now ready to transfer data. The No. of assets field shows the
number of records that will be downloaded. Ensure that the PC application is ready to
receive data then press the Send button (F4).
A progress bar, as shown below, will now follow the data transfer until download is
complete.
This feature is only available when used in combination with the new Med-eBase PC
software and when the SSS format is activated on your device.
To activate the .SSS format on your Rigel BP-SIM tester, please download the SSS
Application note from our knowledge base.
This feature can be used to transfer store or modify the current Trace Variables, User
Profiles and Comment lists to/from a PC. It can be used to configure a number of Rigel
BP-SIM units with identical Trace Variable, User Profile and Comment lists.
Once the BP-SIM has connected to the PC, the BlueTooth icon will appear. Capture the
data using the datatransfer.exe program or Windows hyperterminal.
In the application, select Start Capture from the Capture menu in the task bar.
Enter a valid file name for the text file containing the download information; see below.
Once the text file has been created, press the Send button (F4) to transfer the Asset
Trace variables and User names from the Rigel BP-SIM to the PC.
Close down the Application and open the text file using word pad or equivalent.
The downloaded information can be used to ‘clone’ other Rigel BP-SIM’s or modified to
add additional trace variables. Again these can be uploaded to any Rigel BP-SIM for
convenience.
Note: The file transfer can be made using any suitable PC application such as
the Seaward Bluetooth Downloader (Data Transfer.exe) or Hyperterminal
capable of sending and receiving a text file.
Note: The maximum number of entries for each field is SiteName (40),
LocationName (50), UserName (40) and Comment (80).
Follow the steps above in Send Config Data to run the Bluetooth Downloader and
connect the Rigel BP-SIM to your PC.
Click on File on the Bluetooth Downloader menu and select Send File. This will open a
window allowing you to select the text file containing your Trace Variables, User Profiles
If the transfer is not successful, check that the format of the configuration file exactly
matches that specified.
Note: The Config file must be ASCII text and the last entry must be the [END]
statement.
A bitmap logo can be transferred to the Rigel BP-SIM for use with the Seaward Test ‘n’
Tag print system.
The logo must be in .BMP format with a maximum size of 320 pixels by 240 pixels.
Press button (F4), select Data Transfer using the up/down keys, select load T ‘n’ T
The Rigel BP-SIM is now ready to receive data. The amount of data received is shown
numerically on the top left of the display (0 in the screen shot above).
When using the BlueTooth Downloader application, select Send File from the File menu
in the task bar. This will open a window allowing you to select the Bitmap file.
Select the required Bitmap and click on Open. The Bluetooth Downloader will now
transfer the file to the Rigel BP-SIM.
When the transfer is complete the Rigel BP-SIM screen will display a message as
shown below.
Press the OK button (F4) to save the logo. The logo is now available for
printing using the Seaward Test ’nTag printer.
Note: The file transfer can be made using any suitable PC application such as
the Seaward Bluetooth Downloader (data transfer.exe) or Hyperterminal.
• Firmware version
• Hardware version
• Calibration status
• Last Calibration Date
• Serial Number
• Blue Tooth ID
Use the scoll keys (F3) to toggle between the different information screens;
The Rigel BP-SIM case can be cleaned with a damp cloth with, if necessary, a small
amount of mild detergent. Prevent excessive moisture around the socket panel or in the
lead storage area.
Do not allow liquid inside the Rigel BP-SIM or near the sockets. Do not use abrasives,
solvents or alcohol.
If any liquid is spilt into the Rigel BP-SIM case, the simulator should be returned for
repair, stating the cause of the defect.
The Rigel BP-SIM is a rugged quality instrument. However, care should always be
taken when using, transporting and storing this type of equipment. Failure to treat the
product with care will reduce both the life of the instrument and its reliability.
If the Rigel BP-SIM is subject to condensation, allow the Analyser to completely dry
before use.
• Always check the Rigel BP-SIM and supplied accessories for signs of damage and
wear before use.
• Do not open the Rigel BP-SIM under any circumstances.
• Keep the instrument clean and dry.
• Avoid testing in conditions of high electrostatic or electromagnetic fields.
• Maintenance should only be performed by authorised personnel.
• There are no user replaceable parts in the Rigel BP-SIM.
• The unit should be regularly calibrated (at least annually).
Rigel Medical
Service Dept.,
p.a. Seaward Electronic Limited.
Bracken Hill
South West Industrial Estate
Peterlee, Co. Durham
SR8 2SW, England
sales@rigelmedical.com
calibration@seaward.co.uk
www.seaward.co.uk
www.rigelmedical.com
Prior to returning your unit for service, please contact our service department to obtain a
Returns Number.
By obtaining a Returns Number, your service request can be booked in advance thus
reducing the down time of your equipment.
Simulation
Waveform Oscillometric
Pressure Units mmHg, inHg, kg/cm2, cmH2O, mBar, PSI, inH2O and kPa
General
Keypad Alpha-numeric
Weight <1 kg
Environmental Protection IP 40
Auto Mode
View Data
Memory Options
Restore Factory Settings
About