0% found this document useful (0 votes)
5 views3 pages

silhouette-vr error 90

The document outlines the error code structure for the JEDI error detection and logging system used in the Silhouette VR System. It details the simplified error codes, their classifications, and potential causes for various application errors. The error log file can contain up to sixty-four entries, each providing essential information about the detected errors.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
0% found this document useful (0 votes)
5 views3 pages

silhouette-vr error 90

The document outlines the error code structure for the JEDI error detection and logging system used in the Silhouette VR System. It details the simplified error codes, their classifications, and potential causes for various application errors. The error log file can contain up to sixty-four entries, each providing essential information about the detected errors.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
Download as pdf or txt
Download as pdf or txt
You are on page 1/ 3

GE MEDICAL SYSTEMS SILHOUETTE VR SYSTEM

REV 18 DIRECTION 2229351-100

1-4 Error Code Structure

The error code structure described in this chapter applies to JEDI error detection
and logging. The JEDI error log file can be accessed from the system through the
system console or a laptop (JEDI error log upload functionality).

When an error is detected, it is logged in the JEDI error log file. The file contains
a maximum of sixty four error logs.

Each log shows the following structure:

Simplified Generator Error Class Error Code Data Number of Date & time
error code Phase associated occurrences
with the error
code

The fields are described in the following paragraphs.

1-5 Simplified Error Code Definition

The simplified error code is a grouping of the JEDI error codes. This field gives a
rapid understanding of which part of JEDI is faulty.

Simplified Error Code Description


30 Tube spits errors
40 Rotation errors
50 Heater errors
60 Exposure errors (HV inverter + mA measure + exposure control
70 Power supply errors (low voltage + DC bus)
80 Hardware errors (internal communications + cables)
90 Application errors (saved RAM + software)
100 External communication errors
110 Thermal errors
10 Rotation warnings (engineering use)
20 Heater warnings (engineering use)
25 LVPS warnings (engineering use)
27 Application warnings (mainly saved RAM battery change)

6-2
GE MEDICAL SYSTEMS SILHOUETTE VR SYSTEM
REV 18 DIRECTION 2229351-100

simplified Description Error Error Description


error Class
code
80 Hardware error 4 0180H Rotor board communication problem
4 0181H Rotor board has reset
4 0280H Heater board communication problem
4 0281H Heater board has reset
4 0321H kV conversion error
4 0322H kV ref ADC / DAC failed
4 0601H RTL error
4 0602H External CAN bus off
4 0902H tube Fan supply error
4 1402H Internal CAN bus off
4 1403H Connectic Fault
4 1404H FPGA configuration problem
4 1405H Tank sensor problem

90 Application 4 0701H Saved RAM checksum pb


errors
4 0702H Software problem
4 0703H Watchdog reset has just occurred
4 0704H Rotor/Heater hold too long

100 Communication 4 0603H Debug screen com error


errors
4 0604H Database download error
4 0605H TAV communication error
4 0606H MPC/Madrid communication error
4 1301H AEC communication error

110 Thermal error 5 0804H Tank Thermal Error


5 0903H Tube exceeded 70degC

10 Rotor Warning 1 0151H CAN Domain command number error


1 0152H CAN Domain request with no transfer init
1 0153H CAN Domain Toggle bit error
1 0154H CAN Domain : less than 2 data to download
1 0155H CAN Domain Abort received & applied
1 0156H Bad index in config upload
1 0157H Tube switch while rotor not off
1 0158H Acceleration cmd while no tube selected
1 0159H Acceleration cmd while database not OK
1 0160H Database download while rotor speeding
1 0161H Acceleration command not OK
1 0162H Rotor acceleration while in error
1 0163H No CAN message received within 4 secs
1 0164H Rotation Inverter overcurrent (< 3 times)
1 0199H Unknown rotation warning

6-9
GE MEDICAL SYSTEMS SILHOUETTE VR SYSTEM
REV 18 DIRECTION 2229351-100

Application errors (Code 90)

Class 4 Message/explanation Potential cause


Error code

90- NV RAM checksum pb. 1. Database problem


0701H Generator kV control board has 2. External cause (Spikes)
detected corruption in the NV Ram 3. kV control
verification (After power up)
90- Software problem. 1. Software or Database problem.
0702H 2. kV control board failure.
90- Watchdog reset has just occurred. 1. Software or Database problem.
0703H This error is logged when the CPU of 2. kV control board failure.
the kV control board is reset by it’s
Watch Dog timer.
90- Rotation/Heater hold too long. Software problem.
0704H Will pop up if preparation command
from the system is maintained longer
than 3 minutes.

6-26