0% found this document useful (0 votes)
224 views24 pages

Lecture No. 9 Basic Principles of CT Scan

Copyright
© © All Rights Reserved
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
Download as pptx, pdf, or txt
0% found this document useful (0 votes)
224 views24 pages

Lecture No. 9 Basic Principles of CT Scan

Copyright
© © All Rights Reserved
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1/ 24

Lecture notes

No. 9

Physical Principles(CT) (MDR 522)


ADVANCE PHYSICS
&INSTRUMENTATION Basic principles of CT Scan
OF CT SCAN
INTRODUCTION

 Clinical CT was introduced in 1971 - limited to axial imaging of the brain in


neuroradiology

 It developed into a versatile 3D whole body imaging modality for a wide range of
applications for example

• oncology, vascular radiology, cardiology, traumatology and interventional


radiology.
INTRODUCTION

 CT scan be used for

• diagnosis and follow-up studies of patients

• planning of radiotherapy treatment

• screening of healthy subpopulations with specific risk factors.


INTRODUCTION

 Nowadays dedicated CT scanners are available for special

clinical applications, such as

• For radiotherapy planning - these CT scanners offer an extra wide bore, allowing the CT
scans to be made with a large field of view.

• The integration of CT scanners in multi modality imaging applications, for

example by integration of a CT scanner with a PET scanner or a SPECT

scanner.
INTRODUCTION

 Other new achievements for dedicated diagnostic imaging new achievements


concerns for example the development of a dual source CT scanner (a CT scanner that
is equipped with two X-ray tubes), and a volumetric CT scanner (a 320 detector row
CT scanner that allows for

scanning entire organs within one rotation).


INTRODUCTION

 CT scanning is perfectly suited for 3D imaging and used


in, for example, brain, cardiac, musculoskeletal, and whole body
CT imaging
CT PRINCIPLES

 X-ray projection, attenuation and acquisition of transmission profiles

The purpose of a CT acquisition is to measure x-ray transmission through a patient


for a large number of views.
 Different views are achieved in CT primarily by using:
• detectors with hundreds of detector elements along the detector
arc (generally 800-900 detector elements)
• by rotation of the x ray tube around the patient, taking about 1000
angular measurements
• by tens or even hundreds of detector rows aligned next to each other
along the axis of rotation
CT PRINCIPLES

 The values that are assigned to the pixels in a CT image are associated with the
average linear attenuation coefficient μ (m-1) of the tissue represented within that
pixel.

 The linear attenuation coefficient (μ) depends on the composition of the


material, the density of the material, and the photon energy as seen in Beer’s law:

• I(x) is the intensity of the attenuated X ray beam


• I0 the un attenuated X ray beam
• x the thickness of the material.
 As an X ray beam is transmitted through the patient, different tissues are
encountered with different linear attenuation coefficients.

 The intensity of the attenuated X ray beam, transmitted a distance d, can be


expressed as:
 A CT image is composed of a matrix of pixels representing the average
linear attenuation co-efficient in the associated volume elements (voxels).
 Illustration: a simplified 4 x 4 matrix representing the
measurement of transmission along one line.
 Each element in the matrix can in principle have a different
value of the associated linear attenuation coefficient.
 The equation for the attenuation can be expressed as:
 From the above it can be seen that the basic data needed

for CT is the intensity of the attenuated and un attenuated

X ray beam, respectively I(d) and I0, and that this can be

measured.

 Image reconstruction techniques can then be applied to

derive the matrix of linear attenuation coefficients, which is

the basis of the CT image.


Hounsfield Units
 In CT the matrix of reconstructed linear attenuation
coefficients (μ material) is transformed into a corresponding
matrix of Hounsfield units (HU material), where the HU scale
is expressed relative to the linear attenuation coefficient of
water at room temperature (μwater):

• HU water = 0 as (μ material = μ water)


• HUair = -1000 as (μmaterial = 0)
• HU=1 is associated with 0.1% of the linear attenuation coefficient of water.
HOUNSFIELD UNITS

 Typical values for body tissues.

• The actual value of the HU depends on:

 Composition of the tissue or material

 Tube voltage

 Temperature
CT NUMBER WINDOW

 Hounsfield units are usually visualized in an eight bit grey scale offering
only 128 grey values.

 The display is defined using

• Window level (WL) as CT number

of mid-grey

• Window width (WW) as the number

of HU from black -> white


HOUNSFIELD UNITS

 The choice of WW and WL is dictated by clinical need

 Optimal visualization of the tissues of interest in the CT image can only be


achieved by selecting the most appropriate window width and window level.

 Different settings of the WW and WL are used to visualize

for example soft tissue, lung tissue or bone.


HOUNSFIELD UNITS
 Same image data at different WL and WW

WL -593, WW 529 WL -12, WW 400


HOUNSFIELD UNITS

 The minimum bit depth that should be assigned to the Hounsfield unit is 12,
this enables creating a Hounsfield scale that runs from –1024 HU to +3071
HU, thus covering most clinically relevant tissues.

 A wider Hounsfield scale with a bit depth of 14 is useful for extending the
Hounsfield unit scale upwards to +15359 HU thus making it compatible with
materials that have a high density and high linear attenuation coefficient.

 An extended Hounsfield scale allows for better visualization of body parts


with implanted metal objects such as stents, orthopedic prosthesis’s, dental- or
cochlear implants.
HOUNSFIELD UNITS

 From the definition of the Hounsfield unit

• for substances and tissues, except for water and air, variations of the Hounsfield
units occur when they are derived at different tube voltages.

 The reason is that as a function of photon energy different substances and tissues
exhibit a non linear relationship of their linear attenuation coefficient relative to
water.

 This effect is most notable for substances and tissues that have a relatively high
(effective) atom number such as contrast enhanced blood (iodine) and bone
(calcium).
HOUNSFIELD UNITS

 In clinical practice, considerable deviations between the expected and the


actually observed HU may occur.

 Causes for such inaccuracies may be the dependence of the HU for example on
the:

 Reconstruction filter

Size of the scanned field of view

 Position within the scanned field of view.

 Image artefacts.
HOUNSFIELD UNITS

 When performing clinical studies over time, one should take into account
that even on the same scanner, with time, a certain drift of the HU may occur.

 In multicenter studies that involve different CT scanners, significant


variations in the observed HU may also occur between centers.

 Therefore, quantitative imaging in CT requires special attention and often


additional calibrations of the CT scanners are needed.
THANK YOU

You might also like