Inspiration

People with Parkinson's can suddenly get "stuck" while walking. The feet feel glued to the floor, often at a doorway or when turning. This is called freezing of gait (FoG), and it is a big cause of falls. A simple rhythmic beat can help them start walking again, but only if it plays at the right moment. Nobody can watch a patient all day, so I wanted a wearable that notices the freeze by itself.

My final year project detects machine faults from vibration sensors. Freezing also shows up as a change in frequency: normal steps are around 0.5–3 Hz, and freezing looks like trembling at 3–8 Hz. So I reused my FYP pipeline on human gait.

What it does

ThawStep takes accelerometer data from the ankle, thigh and trunk. Every 0.5 s it looks at the last 4 s and decides if the patient is freezing. If it is, it raises an alarm so a cue (a beat) can be played. It has two modes:

  • Detect: is the patient freezing now?
  • Warn: will the patient freeze in the next 3 s?

The live demo replays real patients the model never saw and shows the alarm as it happens.

How I built it

  • Data: Daphnet Freezing of Gait dataset. 10 patients, 3 accelerometers, 64 Hz, 237 freeze episodes labelled from video.
  • Features: 12 features per axis (RMS, peak, kurtosis, spectral centroid, band powers, ...), including the Freeze Index:

$$\text{FI} = \frac{P_{3\text{–}8\,\text{Hz}}}{P_{0.5\text{–}3\,\text{Hz}}}$$

  • Model: Random Forest. An alarm needs 2 positive windows in a row, and alarms less than 5 s apart count as one cue.
  • Explainability: SHAP shows the Freeze Index features from the ankle drive most decisions.
  • Demo: Streamlit app with a replay, a clinician summary and all the evaluation results.

Evaluation

  • Split: leave-one-subject-out. Train on 9 patients, test on the 10th, and repeat for all 10. Windows overlap, so a random split would leak the same seconds into both training and test.
  • Threshold: picked only on the training patients (inner 3-fold CV), with a clinical rule: catch as many freezes as possible while at least 70% of alarms are real freezes. A false cue is a few seconds of beat, but a missed freeze can be a fall.
  • Baseline: the Freeze Index threshold method from Bächlin et al. (2010), tuned with the same rule.
  • Metrics: share of freeze episodes caught, share of alarms that were real, alarm time vs freeze onset (s), false alarms per hour, and a 95% CI by bootstrapping over patients.
Setup Freezes caught Alarms real Alarm vs onset
ThawStep, detect, 3 sensors 176/237 (74%, CI 55–85%) 72% +0.6 s
ThawStep, detect, ankle only 130/237 (55%) 73% +0.9 s
Freeze Index baseline 27/237 (11%) 53% +4.5 s
ThawStep, warn 3 s ahead 87/237 (37%) 36% 5.6 s before

AUROC per window was 0.90 for ThawStep vs 0.76 for the baseline.

What it means: ThawStep catches about 3 out of 4 freezes within a second of them starting, and 7× more than the standard method. Early warning is not good enough yet: only about a third of warnings were real.

Where it stops being useful:

  • Short freezes under 3 s: only 44% caught, compared with 85–91% for longer ones.
  • Some patients: S02 had only 25% caught.
  • Anything outside a lab, which it was never tested on.

Who benefits, and who could be harmed

  • Benefits: patients (fewer falls, more confidence walking), carers, and physios, who get a real count of freezes instead of guessing.
  • Missed freeze: a patient who trusts the device and still freezes could fall. 26% of freezes were missed.
  • False alarms: about 13 per hour in this lab setting. That could be annoying enough that people stop wearing it.
  • Not representative: only 10 patients, so it may work worse for people unlike them (different age, disease stage, walking aids).

Challenges

  • My first alarm rule (at most 2 false alarms per hour) only caught 13% of freezes. The lab protocol triggers about 40 freezes per hour, so I switched to the "70% of alarms must be real" rule.
  • I didn't want leakage. Overlapping windows plus a random split would have given fake high scores, so everything is per patient.
  • Early warning was much harder than detection. I kept it in and reported it honestly.

What I learned

  • How a good evaluation changes the story: per-window accuracy looked fine, but per-episode results showed what actually matters.
  • Signal-processing ideas from machines carry over to the human body surprisingly well.
  • Being honest about failure cases is part of the result.

Limitations

  • Only 10 patients, recorded in a lab with set walking tasks. It is not tested at home.
  • Freeze start and end come from video labels, which are a bit subjective.
  • Compute is 14 ms per decision on a laptop, but the full model is about 20 MB, so it needs shrinking for a real wearable.
  • Research prototype, not a medical device.

With another month

  • Test on a bigger dataset recorded at home.
  • Shrink the ankle-only model and run it on an ESP32 + MPU-6050 (my FYP hardware), with a real vibration or audio cue.
  • Personalise the threshold per patient after a short calibration walk.
  • Talk to physiotherapists about what false-alarm rate patients would accept.