Milkotester Master PRO
Milkotester Master PRO
49, Hristo Botev St., 4470 Belovo, BULGARIA, Phone: +359 889 259 646
MILKOTESTER
Milk analyzing device
Model: MASTER Pro
Fat, Solids-Non-Fat (SNF), Protein, Lactose, Water content, Temperature
(°C), Freezing point, Salts, Density, pH
CAUTION!
Safety recommendations :
When plugging the device into the electrical outlet, put away
the power cord so it does not stay in the way when
accessing the device and cannot be stepped on.
Table of contents
2. Close-up view................................................................................. 5
3.1 Front panel .......................................................................... 5
3.2 Display ................................................................................ 6
3.3 Rear panel .......................................................................... 7
6. Troubleshooting ............................................................................. 30
CHAPTER ONE
General description
The factory preset is for cow milk, sheep milk and UHT. Upon user’s
request the device can be calibrated for any of the above mentioned
types of milk.
The device has a compact design with a robust structure and a user-
friendly interface. Most importantly, the one-button operation is
extremely simple – you press only once to start measuring, you press
only once for cleaning. MILKOTESTER MASTER can analyze three
types of milk defined by user. The measurement speed is 50 samples
per hour with cleaning included. The samples are precisely dosed and
small quantities are required – the sample volume is 25 cm³. No
thermal or mechanical treatment of the samples before analysis is
necessary. No use of chemical reagents is required. The working
conditions are as follows – temperatures from 5° to 35 °C, HR from
30% to 80%.
With regard to the “moisture problem” which remains of much
significance the front panel of the device is designed to operate when
with wet hands. In addition to this MILKOTESTER MASTER provides a
one year’s full warranty.
CHAPTER TWO
Close-up view
IN THIS CHAPTER
Front panel
Display
Rear panel
1 – Intake pipette
2 – Sample cup
Close-up view
2.2 Display
7
4
9
8
1 – LCD
2 – Arrow UP button
3 – Arrow DOWN button
4 – ENTER button
5 – EXIT button
6 – MENU button
7 – CLEAN button
8 – pH measurement (active only on request)
9 – Keypad
6 5
2
3
CHAPTER THREE
Samples preparation
In order that most accurate results are obtained it is important that the
milk be kept for 2 hours after milking and stirred well before being used
for sample material. A good way to stir up the milk is to pour it several
times out of one vessel into another and back. Before analysis the
samples have to be filtered in order to be free from foreign fragments.
Milk samples should be 5-35 °C. Full cream samples containing fat
over 10% should be heated up to 42-43 °C and then cooled to 25-30
°C. If the sample temperature is over 36 °C the message “Sample
overheated” will appear on the display. Use milk samples only once
and do not return it to the vessel. Samples can be kept for a maximum
of 2 days if they are stored at a temperature not exceeding 5 °С.
Milk stirring
It is a very important condition for receiving exact results. Before
taking samples from big vessels, the milk (fresh or thermally treated,
whole-milk or whipped) has to be well stirred for no less than 5 min., by
vertical and circular slow movements. Mixing spoon with long handle is
used, allowing the lowest layers of the liquid to be reached. The milk in
the milk-cans is stirred 5 to 8 times from the surface to the bottom and
reverse with slow circular movements.
Sample preservation
The vessels where the samples will be put have to be clean, dry,
glass, metal or from other suitable material, to be tightly closed with
rubber or other stopples. The stopples not to absorb water and fat and
not to influence the analyses sample content.
In summer the sample fills up to the top the vessel, but in winter
– at least 3/4 from the vessel’s volume. Each sample for analyses has
to be labeled and described in a way not allowing to be mixed up.
Do not make analyses if the acidity of the milk is more than 17oT.
CHAPTER FOUR
Working description
IN THIS CHAPTER
Preparing the device
Measuring samples
Corrections
Calibration
1. Unplug the 12V DC power cable from the socket on the rear
panel of the unit and then from the electrical outlet.
NOTE:
If the unit does not start up after it is connected to the power source,
check the fuse inside the connector plugged into the electrical lighter
socket.
Milk Analyzer
MASTER
Please Wait
System prepare
Followed by:
>Cow Milk
Sheep Milk
UHT Milk
You can choose among three types of milk at a time, (upon user’s
request the device can be calibrated for any of the other types of
milk).
Place the cup with the milk sample
Place a second milk sample under the pH probe (the automatic mode
for pH measurement must be turned on).
When you choose the desired type you use the ↑↓ (UP and DOWN)
arrow buttons, and then press ENTER to activate analyzing.
At this point the sample cup can be removed from under the pipette.
The instrument will continue to show the measured values until a new
analysis is initiated.
WARNING!
Make sure that the instrument is at rest during analysis. Any shaking of
the device will cause inaccurate results.
NOTE:
Due to the possible presence of water inside the flow system left from
the last flushing procedure, we recommend that you do not take into
account the first analysis after flushing if they appear inaccurate. To
prevent or reduce this inaccuracy due to water in the system, please
refer to chapter “Cleaning and Maintenance”, section “Thorough
Cleaning”.
To access the menu, press the MENU button. The range of functions
is grouped into submenus. Scroll through the menu to select the one
that you need and then select the settings you need to make. Press
ENTER to activate them. Press EXIT to keep the previous settings.
This is a basic moment for the correct checking the accuracy of the
analyzer and for making correct and precise correction and calibration.
It is accomplished according Appendix Taking and preparation of
samples for checking correctness of the milk analyzer, making
corrections and recalibration.
Making measurements
Make measurements with different samples (not less than 3) with
known values of a separate parameter (for example fat content),
For example
М% of the reference samples: 2,20 3,00 3,80 4,60 5,20
М% average when measuring
with the analyzer: 2,38 3,17 4,01 4,79 5,42
Difference: 0,18 0,17 0,21 0,19 0,22
For example.
М% of the reference samples: 2,20 3,00 3,80 4,60 5,20
М% when measured with the
analyzer: 2,02 2,93 3,76 4,75 5,44
Difference: -0,18 -0,07 -0,04 0,15 0,24
Conclusion: It is obvious that the difference is variable value and recalibration have to be
done.
4.3.1 Correction
Example:
MENU > Mode > Correction > Cow milk > Fat.
To select Mode, Options or Settings use the ↑↓ (up and down) arrows
and press ENTER.
Mode
Options
Settings
In Submenu Mode Use the ↑↓ (up and down) arrows and press
ENTER to select Correction, Calibration, Back up or Restore.
Correction
Calibration
Back up
Restore
If you choose correction and press ENTER, then use the ↑↓ (up and
down) arrows and press ENTER to select the type of milk.
Correction
Cow Milk
Sheep Milk
UHT Milk
After choosing the type of milk use the ↑↓ (up and down) arrows and
press ENTER to choose a component for correction (mind that there
are 2 (two) pages of components thus scroll up and down).
Fat
SNF
Density
Protein
Lactose
Salts
Water
Press the ↑↓ (up and down) arrows to correct the value with 0.01.
Correction
Of Fat
0.01
4.3.2 Calibration:
Use the ↑↓ (up and down) arrows and press ENTER to select the type
of milk.
The following text will appear on the display:
Use the ↑↓ (up and down) arrows and press ENTER to select a
parameter. Then the cursor will mark the place to enter a known value.
Use the ↑↓ (up and down) arrows to set the desired figure and press
ENTER to confirm. Then the cursor goes to the figure to the right. Use
the same procedure to enter the second and third figure. After
confirmation of the last figure the cursor goes to the next parameter for
entering new values (mind that there are 2 (two) pages of components
thus scroll up and down).
By confirming the last figure of the last parameter the following text
appears on the display:
Put sample
High Milk
1/5 times
and Press ENTER
Put Sample
High Milk
2/5 times
and Press ENTER
Follow the procedure until the 5th measurement. After performing the
last measurement the following text appears on the display:
Put Sample
Low Milk
1/5 times
and Press ENTER
Put Sample
Low Milk
2/5 times
and Press ENTER
Follow the procedure until the 5th measurement. After performing the
last measurement the following text appears on the display:
Put Sample
Water
1/5 times
and Press ENTER
Follow the described above procedure. After the end of the 5th
measurement the calibration procedure is complete.
Please note:
Before each sample measurement stir the milk by pouring it from one
vessel to another.
The temperature of the samples should be in the range of 15 to 25
degrees.
Before starting calibration it is recommendable to back up the
parameters of the existing calibration:
Cow – calibr. 1
Sheep – calibr.2
UHT – Calibr. 3
Choose the type of milk calibration for back up and press ENTER
Choose System info, Language or Time and Date, and press ENTER.
MILKOTESTER
Model: MASTER Pro
Ver: [Link]
SN: XXXXXX
Language:
English
Portuguese
Use ↑↓ (Up and Down) arrows to choose a type of milk and press
ENTER. The following text appears on the display:
Use ↑↓ (Up and Down) arrows to enter the new value and Press Enter
to confirm. After dialing the last symbol press ENTER to confirm the
new value.
In order to calibrate the device you will need two buffer solutions with
known values, for example – one with value of pH 3 /low value/ and
the other – pH 7 /high value/.
After choosing Set pH the following text will appear on the display:
Set
pH low buffer:
[Link]
Setting
Wait until the electrode and the buffer stabilize /the reference value
should change the least possible/. Press Enter to set the value.
Dip the electrode into the high value buffer solution and make sure that
the junction zone of the electrode is at least at 30 mm below the
surface of the buffer solution and leave it at rest. Now you have to
enter the known value of the “high” buffer solution. Use ↑↓ (Up and
Down) arrows to enter the new value and Press Enter to confirm. After
dialing the last symbol press ENTER to confirm the new value.
The following text will appear on the display:
Setting
Wait until the electrode and the buffer stabilize /the reference value
should change the least possible/. Press Enter to set the value
Please note:
Press Enter and the following text will appear on the display:
Calibration
Measure
Scale ON/OFF
Printer auto En
Printer auto Dis
4.6 Measurement of pH
(active only on request)
The milk analyzer has two modes of work – manual and automatic.
They are selected from MENU > Options > pH mode
Choose pH mode and press ENTER.
System Info
Language
Time and Date
pH mode
Under the manual mode, the milk sample is put under the pH probe,
then you press the pH button and the measurement will last without
interruption (until the pH button is pressed again). In order to measure
pH the device must be in a working condition.
Measurement of pH is not possible when the device is in the following
mode:
MEASURING…
Please Wait
pH result:
[Link]
Press the EXIT button to stop measurement of pH and enter the mode:
Cow Milk
Sheep Milk
UHT Milk
45 113 0,30 0,24 0,18 0,12 0,06 0 0,06 0,12 0,18 0,24 0,30
55 131 0,45 0,36 0,27 0,18 0,09 0 0,09 0,18 0,27 0,36 0,45
65 149 0,60 0,48 0,36 0,24 0,12 0 0,12 0,24 0,36 0,48 0,60
75 167 0,75 0,60 0,45 0,30 0,15 0 0,15 0,30 0,45 0,60 0,75
85 185 0,90 0,72 0,54 0,36 0,18 0 0,18 0,36 0,54 0,72 0,90
Put Light Weight Using the buttons Up, Down, Enter (as it is done
10 kg when calibrating the rest of the milk analyzer’s
Set and press
ENTER
options) enter the weight of the lighter standard
of weight, in our case 10.00kg.
Put Heavy Weight The same procedure is repeated with the heavy
50 kg standard weight, for example 50.000kg.
Set and press
ENTER
CHAPTER FIVE
IN THIS CHAPTER
Routine cleaning
Complete flushing
Weekly cleaning
For high accuracy and reliability of the device, the following cleaning
procedures should be performed on due times.
1. Fill up a sample cup with clean warm water (45 – 50°С) and
place it on the sample shelf under the pipette.
After the end of the procedure the device is ready for measurement.
The daily cleaning of the milk analyzer is done with water, heated to
50° C or with cleaning powder “Master Daily”, by pressing the button
“Clean”. The liquid in the cup must be changed each time until all of
the contamination is removed. At the end the system must be washed
out with water.
We recommend that you carry out the following procedure at the end
of every week.
The weekly cleaning of the milk analyzer is done with water, heated to
50° C or with cleaning powder “Master Weekly”, by pressing the button
“Clean”. The liquid in the cup must be changed each time until all of
the contamination is removed. At the end the system must be washed
out with water.
CHAPTER SIX
Troubleshooting
CHAPTER SEVEN
Technical specification
1. Measured components:
Fat
Solids-Non-Fat (SNF)
Protein
Lactose
Water content
Temperature (°C)
Freezing point
Salts
Density
2. Ranges of measurement:
* To determine the milk density, add 1000 to the result from the
display.
Example: Result = 22; density = 22+1000 =1022.
3. Accuracy:
Fat ± 0.08%
SNF ± 0.1%
3
Density ± 0.3 kg/m
Proteins ± 0.1%
Lactose ± 0.1%
ADDED Water ± 3.0%
o
Sample Temperature ±1 C
o
Freezing point ± 0.005 C
Salts ± 0.05%
Air temperature from 10˚C to 40˚C
Relative humidity from 30% to 80%
Electrical parameters:
AC Power supply 95÷250V
DC Power supply 12V
3. Dimensions:
W/D/H 110/260/205 mm
5. Weight – 3.1 kg
3
4. Milk sample volume per 1 measurement: 25 cm (=25 ml)
The milk analyzer determines the freezing point of each sample and
the quantity of added water. The milk analyzer does not measure the
freezing point, but calculates it from the components it depends on.
The basic components in the milk are water, Salts, lactose, FAT,
proteins, minerals (salts) and acids. The freezing point depends only
on the diluted in the milk components and quantity of the solvent (in
the milk it is water). The ultrasonic technology allows direct
measurement of FAT, proteins, lactose + salts (the soluble
components, only influencing the freezing point), and the quantity of
the solvent in % is determined by 100 % – total solids %, total solids =
lactose % + FAT % + proteins % + salts % + acids %.
Without understanding the meaning of the freezing point – determined
or shown from the milk analyzer added water result easily may lead to
a mistake for the value of this parameter.
Where:
FrPointBase is the basic freezing point
FrPointCalc is measured freezing point
Sample:
First variant
If you’ve entered for milk analyzer basic freezing point -0.520C
(according article 5.9 of the EU Milk Hygiene Directive 92/46/ЕЕС),
measured freezing point –0.540C, using the above pointed formula
you’ll receive –3,8%. Because it is not possible the added water to be
negative value, the milk analyzer indicates 0% added water. The
reason for this is the tolerance in the basic freezing point, reasons for
which are described below.
If in the same milk we add 3,8% water, and the basic freezing point is
the same, the milk analyzer will measure freezing point –0.520C, and
will indicate again 0% added water.
Second variant
The device’s results for added water may give information about doubt
of added water in the milk and the exact value of this added water may
be determined after a “cowshed sample” is taken and the result for the
freezing point, measured by the milk analyzer of the “cowshed sample”
is entered as basic freezing point in the formula for calculation of
added water.
Then the result from this formula will give us the absolute value of the
added water for the corresponding milk supplier.
Method of determination.
Conductivity (or Electrolytic Conductivity) is defined as the ability of a
substance to conduct electrical current. It is the reciprocal of the resistance.
*These values depend on the geographical region, the breed and on other
factors.
Conductivity calibration
1/5 times
And press Enter
Milkotester Ltd.
49, Hristo Botev St. 4470 Belovo, BULGARIA
MILKOTESTER MASTER
Guarantee period is 1 (one) year after purchasing date.
Improper handling, transport and storage will invalidate the guarantee.
Guarantee is void if warranty labels are removed.
Distributor: ......................
Signature: ......................
Stamp: ......................